The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

Aveir DR Real-World Evidence Study


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Ongoing
Application Number /
Requirement Number
P150035 S003/ PAS001
Date Original Protocol Accepted 06/29/2023
Date Current Protocol Accepted 11/22/2024
Study Name Aveir DR Real-World Evidence Study
Device Name Aveir DR Leadless System
Clinical Trial Number(s) NCT05252702  
General Study Protocol Parameters
Study Design Comprehensive/Linked/RegistryBased Surveillance
Data Source Sponsor Registry
Comparison Group No Control
Analysis Type Descriptive
Study Population Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives The purpose of the post-approval study (PAS) is to evaluate the long-term safety of the dual-chamber Aveir Leadless Pacemaker device (aka DR LP) using real-world evidence (RWE) methods. The study utilizes a real-world evidence method merging multiple real-world datasets from Abbott and CMS to assess the Aveir DR LP safety in a large patient population.

Study Population All de novo patients implanted with the Aveir DR LP who successfully link to Medicare FFS data will be included in the analysis, starting at market approval and continuing for approximately one year or longer, if necessary, to include a minimum of 1,805 subjects. Patients implanted with an Aveir atrial LP only (i.e., without a ventricular LP) will be analyzed in additional analysis and not in the primary and secondary analyses.
Sample Size 1805
Key Study Endpoints Primary Endpoints:
1. Acute complications through 30 days post implant procedure
2. Freedom from chronic complications from 31 days through 5 years post implant procedure

Secondary Endpoints:
1. Complication rate of the Aveir DR LP for key individual acute and chronic complications
2. Care of subjects at the end of device service

Descriptive Endpoints and Analyses:
1. Rates of other procedure-related acute complications
2. Survival at 5 years
3. Device longevity
4. Device longevity by device (atrial, ventricular)
5. Acute complications by device (atrial, ventricular)
6. Chronic complications by device (atrial, ventricular)

Additional analyses include:
1. Cohort demographics and comorbidities
2. Device-device interaction
3. Effect of implanter experience with DR leadless pacemakers
4. Acute and chronic complications in patients with an atrial LP only
5. Battery performance
6. i2i performance
7. Acute and chronic complications in all Aveir patients (de novo and non de novo)

Follow-up Visits and Length of Follow-up The study completion is 5 years after the final implant.
Interim or Final Data Summary
Interim Results Safety Results:
The key acute complication rate through 30 days post implant was 6.6% (95% CI 5.7%-7.6%) as of the data cut-off date for 2,793 implants in the primary de novo
analysis group. The primary de novo cohort includes subjects implanted with both AVEIR DR devices (atrial and ventricular LPs) that meet the de novo
criteria defined as subjects who do not have evidence of a prior Cardiac Implantable Electronic Device (CIED). A CIED includes a leadless single-chamber
pacemaker, transvenous single-chamber pacemaker, dual-chamber leadless pacemaker, dual-chamber transvenous pacemaker, implantable cardioverter
defibrillator (ICD), or cardiac resynchronization therapy pacemaker (CRT-P) or defibrillator (CRTD).
Effectiveness Results:
Freedom from key chronic complications from 31 days to 5 years post implant was evaluated through 24-months based on data available through the data
cut-off date for this 36-month report. The event-free survival rate was 99.4% at 24-months post implant for 2,793 implants in the primary de novo analysis
group.
Actual Number of Patients Enrolled 5169
Actual Number of Sites Enrolled 589 enrolled
Patient Follow-up Rate N/A


Aveir DR Real-World Evidence Study Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
6 month report 12/28/2023 12/15/2023 On Time
1 year report 06/28/2024 07/01/2024 Overdue/Received
18 month report 02/27/2025 02/27/2025 On Time
2 year report 08/28/2025 08/28/2025 On Time
3 year report 08/28/2026 08/28/2026 On Time
4 year report 08/28/2027    