The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

NSE Alpha New Enrollment PAS


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Study Pending
Application Number /
Requirement Number
P230026 / PAS001
Date Original Protocol Accepted 08/02/2024
Date Current Protocol Accepted  
Study Name NSE Alpha New Enrollment PAS
Device Name Lacrosse NSE ALPHA Coronary Dilatation Catheter
Clinical Trial Number(s) NCT04985773 
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group No Control
Analysis Type Descriptive
Study Population Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives This new enrollment study is a prospective, multicenter, single-arm, post-approval study to further evaluate the Lacrosse NSE ALPHA coronary dilation catheter in patients with stenotic coronary artery disease who are suitable candidates for percutaneous transluminal coronary angioplasty. The purpose of the study is to further evaluate the performance of Lacrosse NSE ALPHA for crossing and scoring stenotic lesions in complex lesions, with specific focus on understanding success of lesion crossing after encouraging physicians to plan for the use of extra delivery support in complex (moderate to severely calcified and/or tortuous) lesions. Patients will be consecutively enrolled and at least 30 patients will have both severe calcification and moderate/severe vessel tortuosity. The remaining patients will have lesions with complex characteristics (i.e., lesions characterized on a spectrum of moderate to severe calcification or vessel tortuosity).
Study Population Patients with stenotic coronary artery disease who are suitable candidates for percutaneous transluminal coronary angioplasty. At least 30 patients with complex lesion characteristics will be enrolled. Outside of the minimum 30 patients enrolled for severe calcification and moderate/severe vessel tortuosity, the remaining patients will maintain complex lesion characteristics i.e., lesions characterized on a spectrum of moderate to severe calcification or vessel tortuosity.
Sample Size Number of subjects:
At least 100 subjects, including at least 30 subjects with complex lesion characteristics (i.e., lesions with both severe calcification and moderate/severe vessel tortuosity), will be consecutively enrolled and followed through discharge. Outside of the minimum 30 subjects with both severe calcification and moderate/severe vessel tortuosity, the remaining patients will have lesions with complex characteristics (i.e., lesions characterized on a spectrum of moderate to severe calcification or vessel tortuosity).
Assumptions for sample size estimation: N/A
Number of sites: Subjects will be enrolled at up to 10 sites, with no single sire enrolling more than 20% of the total enrollment.
Sites location: U.S.
Key Study Endpoints Primary Endpoints:
Device procedural success, defined as:
- Successful delivery, inflation, deflation, and withdrawal of the study balloon; and
- No evidence of device-related vessel perforation, flow limiting dissection (grade C or higher, per Clinical Events Committee [CEC] adjudication) or reduction in thrombolysis in myocardial infarction (TIMI) flow from baseline (per core laboratory assessment); and
- Final TIMI flow grade of 3 at the conclusion of the percutaneous coronary intervention (PCI) procedure (per core laboratory assessment).
This endpoint will be presented as the proportion of subjects experiencing device procedural success. The proportion of target lesions meeting the primary endpoint will be concurrently calculated and presented as a secondary endpoint, as noted below.
The following Subgroup and Sensitivity Analyses will be performed for the primary outcome:
- Lesion complexity (overall); by operator grading and angiography (per core laboratory assessment)
- Support catheter use (overall)
- Support catheter use in complex lesions
Secondary Endpoints:
12. Angiographic procedural success, defined as final diameter stenosis less than or equal to 50% in at least one of the Lacrosse NSE ALPHA attempted lesions following completion of the interventional procedure, including adjunctive stenting (per core laboratory assessment).
13. Major adverse cardiac events (MACE) through hospital discharge (per CEC adjudication). MACE is defined as a composite of:
- All-cause death
- Myocardial infarction
- Clinically indicated target lesion revascularization
14. Stent thrombosis within the target vessel(s) through hospital discharge using ARC-2 definitions for definite and probable (per CEC adjudication).
15. Clinically significant arrhythmia (defined as those requiring intervention) through hospital discharge (per CEC adjudication).
16. Occurrence of Lacrosse NSE ALPHA balloon rupture (per device deficiency electronic case report form [eCRF]).
17. Change in minimum lumen diameter following use of Lacrosse NSE ALPHA catheter, measured by quantitative coronary angiography (per core laboratory assessment).
18. Device procedural success defined as for the primary endpoint, calculated and presented per target lesion.
19. Rate of device success and failure to cross lesions, including how this may relate to specific lesion characteristics (i.e., calcified and/or torturous).
20. Number of crossing attempts with the device, if any complications with crossing are observed.
21. Details regarding the use of support catheters and/or additional devices to enable lesion crossing after failure from use of device.
22. Physician a priori evaluation of crossing complexity.
Secondary endpoints 1 through 5 are defined at the subject level and will be presented as the proportion of subjects experiencing the above events. In cases where multiple events per subject may be observed (e.g., adverse events), the main analysis will be at the subject level, but total event counts will also be reported. Secondary endpoints 6 through 11 are defined at the lesion level and statistical analyses of these endpoints will account for within-subject correlation as some subjects will have more than one target lesion, using generalized estimating equations with an exchangeable correlation structure.
Follow-up Visits and Length of Follow-up The overall study duration, from screening the first patient to the final follow-up visit, data analysis, and final report, is expected to be approximately 27 months.


NSE Alpha New Enrollment PAS Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
6 month report 12/05/2024 11/27/2024 On Time
1 year report 06/06/2025 05/02/2025 On Time
18 month report 12/05/2025 12/05/2025 On Time
2 year report 06/06/2026 06/05/2026 On Time
3 year report 06/06/2027    
4 year report 06/05/2028    