Post-Approval Studies (PAS) Database

  • Print
  • Share
  • E-mail
-

The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

New Enrollment Study


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Ongoing
Application Number /
Requirement Number
P240003 / PAS002
Date Original Protocol Accepted 11/08/2024
Date Current Protocol Accepted  
Study Name New Enrollment Study
Device Name Minima Stent System and Minima Pro Stent System
Clinical Trial Number(s) NCT05086016 
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group No Control
Analysis Type Descriptive
Study Population Neonate: Birth < 29 days, Infant: 29 days < 2 yrs, Child: 2 yrs < 12 yrs
Detailed Study Protocol Parameters
Study Objectives Single arm, prospective, non-randomized, multi-center, open-label, post approval study
To collect post-market, real-world safety and effectiveness data of the Minima Stent and Minima Pro Stent when used in treatment of native and post-operative congenital vascular stenoses within the aortic and pulmonary circulations, in neonates, infants, and small children.
Study Population The study population will comprise patients treated with the Minima Stent System and the Minima Pro Stent System.
Sample Size Minimum 100 patients with minimum of 15 subjects enrolled with the 10mm Minima Pro Stent System

Number of sites: N/A
Sites location: N/A
Key Study Endpoints Safety Endpoints
Primary:
Freedom from procedure- or device-related SAEs resulting in an event listed below:
•Death
•Cardiac arrest and/or emergency ECMO cannulation
•Stroke
•Limb loss
•Vessel dissection of target lesion
•Device thrombosis/occlusion
•Cardiac perforation requiring percutaneous or open surgical intervention
•Persistent cardiac arrhythmia requiring a pacemaker
Secondary:
• Freedom from intraprocedural stent movement to non-intended location during the procedure.
• Freedom from postprocedural stent embolization or migration
• Freedom from stent fracture that led to reintervention
• Freedom from non-elective Minima Stent or Minima Pro Stent explant
• Freedom from procedure- or device-related SAE during re-dilation that results in the following:
o Death
o Cardiac arrest and/or emergency ECMO cannulation
o Stroke
o Limb loss
o Vessel dissection of target lesion
o Device thrombosis/occlusion
o Cardiac perforation requiring percutaneous or open surgical intervention
o Persistent cardiac arrhythmia requiring a pacemaker
Effectiveness Endpoints
Primary:
Acute device success as defined by:
•Stenosis relief, defined by stent outer diameter = 75% of the normal surrounding vessel immediately after deployment.
•Freedom from open surgical intervention required to treat Minima Stent or Minima Pro Stent dysfunction at implantation
Secondary:
•Peak-to-peak pressure gradient (ventricle to arterial or arterial to arterial) < 20 mmHg after stent placement, in treatment of Aortic Coarctation, when applicable.
•Successful stent re-dilation (when indicated) at re-catheterization, defined as angiographic improvement of stenosis to >50% of the normal surrounding vessel.
•Maintenance of stented vessel diameter = 50% of post implant diameter at the time of re-dilation (pre re-dilation).
Interim or Final Data Summary
Interim Results Safety Results: N/A
Effectiveness Results: N/A
Actual Number of Patients Enrolled 4
Actual Number of Sites Enrolled Following protocol approval, 15 sites were recruited to take part in the Minima Post-Approval Study. The first site completed activation activities on April 8, 2025 and there is only one site pending activation as of August 27, 2025. The last site is only pending IRB approval and is anticipating approval within the next week. This site, like the others, began site initiation activities in November of 2024 following Protocol Approval and experienced numerous delays within their site due to new processes and resourcing issues. The 14 other sites have completed IRB Approvals and are able to enroll patients.
Patient Follow-up Rate N/A


New Enrollment Study Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
6 month report 02/26/2025 01/28/2025 On Time
1 year report 08/28/2025 08/28/2025 On Time
18 month report 02/26/2026 03/02/2026 Overdue/Received
2 year report 08/28/2026 08/26/2026 On Time
30 month report 02/28/2027    
3 year report 08/28/2027    