U.S. Department of Health and Human Services

Post-Approval Studies (PAS) Database

  • Print
  • Share
  • E-mail
-

The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

Completion of WAVE Study


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Ongoing
Application Number /
Requirement Number
P240023 / PAS001
Date Original Protocol Accepted 02/12/2025
Date Current Protocol Accepted 02/12/2025
Study Name Completion of WAVE Study
Device Name WRAPSODY® Cell-Impermeable Endoprosthesis
Clinical Trial Number(s) NCT03644017 NCT04540302 NCT05062291 
General Study Protocol Parameters
Study Design Randomized Clinical Trial
Data Source New Data Collection
Comparison Group Concurrent Control
Analysis Type Analytical
Study Population Transit.Adolescent B(as adults): 18 yrs < 22 yrs, Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives This study is a prospective, randomized (AVF cohort only), controlled, multicenter study comparing the Merit WRAPSODY Endovascular Stent Graft to Percutaneous Transluminal Angioplasty for treatment of venous outflow circuit stenosis or occlusion in hemodialysis patients.

The objective of this study is to demonstrate the long-term safety and efficacy of the Merit WRAPSODY Endovascular Stent Graft for treatment of stenosis or occlusion within the dialysis access outflow circuit, including:
a) AVF Peripheral - the peripheral veins of subjects with an arteriovenous (AV) fistula, and
b) AVG Peripheral – at the venous anastomosis of subjects with a synthetic arteriovenous graft access

Cohort (a) will be compared to PTA. Cohort (b) will be compared to performance goals.
Study Population The study comprises two (2) independent cohorts: 1) Subjects with AVF for hemodialysis who have stenosis or occlusion of the peripheral venous outflow circuit, including the cephalic arch; and 2) subjects with AVG for hemodialysis who have stenosis or occlusion at the graft-vein anastomosis or juxta-anastomosis.
Sample Size Number of subjects: 246 AVF peripheral subjects; 112 AVG Anastomosis Subjects; 16 Central Cohort
Number of sites: 43 centers; 10 international
Sites location: United States, United Kingdom, and Brazil
Key Study Endpoints Two primary endpoints are proposed:

Primary Safety: Proportion of subjects without any localized or systemic safety events through 30 days that affect the access or venous outflow circuit and result in reintervention, hospitalization, or death

Primary Effectiveness: Target Lesion Primary Patency (TLPP) at 6 months
Secondary:

Key secondary endpoints include:
TLPP at months 12 and 24
Access Circuit Primary Patency (ACPP) at months 6, 12, and 24
Rates of procedure- and device- related adverse events
Rate of device observations and potential malfunctions or failures
Number of reinterventions at months 6, 12, and 24 months
Follow-up Visits and Length of Follow-up After the index procedure on Day 0, subjects shall be evaluated within the clinic at 30 days, then at months 6, 12 and 24. Telephone follow-up shall be completed at months 3, 9 and 18.
Interim or Final Data Summary
Interim Results Safety Results
AVF Cohort (WRAPSODY | PTA):
30 Day freedom from localized or systemic safety events: 96.6% (115/119) |
95.0% (115/121)
AVG Cohort:
30 Day freedom from localized or systemic safety events: 95.4% (104/109)

Effectiveness Results
AVF Cohort (WRAPSODY | PTA):
Effectiveness:
6 Month Target Lesion Primary Patency (TLPP): 89.6% (103/115) | 71/114
(62.3%)
12 Month TLPP: 69.4% (75/108) | 40/104 (38.5%)
6 Month Access Circuit Primary Patency (ACPP): 72.2% (83/115) | 65/114
(57.0%)
12 Month ACPP: 56.4% (62/110) | 31.1% (33/106)
6 Month Secondary Patency: 100% | 100%
12 Month Secondary Patency: 100% | 97%
Rates of Procedure-Related Adverse Events:
6 Months: 5.3% (6/114) | 17.5% (20/114)
12 Months: 5.7% (6/105) | 21.4% (21/98)
Rates of Device-Related Adverse Events:
6 Months: 2.6% (3/114) | 8.0% (9/112)
12 Months: 2.9% (3/105) | 9.5% (9/95)
Rate of device observations and potential malfunctions or failures: 2.5% (3/122)
| NA
Number of target lesion reinterventions at 6 Months: 0.18 ± 0.51 (115) | 0.47 ±
0.65 (114)
Number of target lesion reinterventions at 12 Months: 0.50 ± 0.92 (108) | 1.08 ±
1.16 (104)
Number of interventions to maintain access circuit patency at 6 Months: 0.48 ±
0.91 (115) | 0.78 ± 1.12 (114)
Number of interventions to maintain access circuit patency at 12 Months: 1.08 ±
1.63 (110) | 1.70 ± 1.93 (105)
AVG Cohort:
Effectiveness:
6 Month TLPP: 81.4% (79/97)
12 Month TLPP: 58.0% (51/88)
6 Month ACPP: 68.0% (68/100)
12 Month ACPP: 33.7% (31/92)
6 Month Secondary Patency: 95.1%
12 Month Secondary Patency: 88.4%
Rates of Procedure-Related Adverse Events:
6 Months: 14.7% (14/95)
12 Months: 17.5% (14/80)
Rates of Device-Related Adverse Events:
6 Months: 8.4% (8/95)
12 Months: 10.1% (8/79)
Rate of device observations and potential malfunctions or failures: 0.0% (0/112)
Number of target lesion reinterventions at 6 Months: 0.21 -/+ 0.46 (97)
Number of target lesion reinterventions at 12 Months: 0.65 -/+ 0.87 (88)
Number of interventions to maintain access circuit patency at 6 Months: 0.70 -/+ 1.41 (98)
Number of interventions to maintain access circuit patency at 12 Months: 1.89 -/+ 2.20 (90)
Actual Number of Patients Enrolled 374 patients; 295 in the US; 79 internationally (Brazil, Canada, and UK)
Actual Number of Sites Enrolled 43 sites; 33 US sites; 4 Brazilian sites; 1 Canadian site; 5 UK sites
Patient Follow-up Rate Completed Study:
Overall: 184/374 (49.2%)
AVF: 128/246 (52.0%)
AVG: 47/112 (42.0%)
Terminated early:
Overall: 154/374 (41.2%)
AVF: 92/246 (37.4%)
AVG: 56/112 (50.0%)
Active in the study:
Overall: 36/374 (9.6%)
AVF: 26/246 (10.6%)
AVG: 9/112 (8.0%)


Completion of WAVE Study Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
6 month report 06/19/2025 06/18/2025 On Time
final report 01/31/2026 01/29/2026 On Time