| General |
| Study Status |
Ongoing |
Application Number / Requirement Number |
P240028 / PAS001 |
| Date Original Protocol Accepted |
06/18/2025
|
| Date Current Protocol Accepted |
06/18/2025
|
| Study Name |
WiSE CRT System Post-Approval Study (PAS)
|
| Device Name |
WiSE® (Wireless Stimulation of the Endocardium Technology) CRT System
|
| Clinical Trial Number(s) |
NCT02922036
|
| General Study Protocol Parameters |
| Study Design |
Other Study Design
|
| Data Source |
Sponsor Registry
|
| Comparison Group |
No Control
|
| Analysis Type |
Descriptive
|
| Study Population |
Adult: At least 22 yrs
|
| Detailed Study Protocol Parameters |
| Study Objectives |
The WiSE-UP Registry is a prospective, non-randomized, observational, multi-center study tracking patients implanted with the WiSE System to understand clinical outcomes, CRT response, & long-term product performance. Patients enrolled in the study will be followed per the standard of care practices of their provider.
|
| Study Population |
Adults at least 22 years of age
|
| Sample Size |
Number of subjects: 320 Assumptions for sample size estimation: • Projected attrition rates of 10% annual mortality and 15% annual loss to follow-up. • Estimate a 1-month 15% event rate based on the observed rate during the SOLVE-CRT Pivotal Study • Estimation that at least 275 patients will complete 1-month follow-up. • Estimation that approximately 240 patients will remain evaluable at 1 year. Number of sites: 50 Sites location: US and OUS
|
| Key Study Endpoints |
The rate of serious adverse events (SAEs) at < 30 days post implant related to the WiSE System and/or implant procedure. Secondary Endpoints: The outcome measures for this study include the following evaluations: • Procedural Success • CRT Response Assessment • HF Clinical Composite Score • NYHA Classification • QRS Duration • Minnesota Living with Heart Failure Questionnaire (6 & 12 months only) • Events Assessment • Mortality • Patient Global Assessment • Device Longevity & Battery Performance • Left Ventricular End-Systolic Volume (LVESV)
|
| Follow-up Visits and Length of Follow-up |
Baseline visit, 1 month, 6 months, annually for 5 years
|
| Interim or Final Data Summary |
| Interim Results |
Safety Results: No Serious Adverse Events (SAEs) have been reported.
Effectiveness Results: Procedural success rate was 100% (17/17). At 1-month follow-up, 12/13 patients had Cardiac Resynchronization Therapy (CRT) data available: 3/12 patients were classified as “Responders”, 7/12 patients were classified as “Likely Responders”. A mean reduction in QRS duration of 14 ms was observed from baseline to 1 month (n=13).
|
| Actual Number of Patients Enrolled |
17
|
| Actual Number of Sites Enrolled |
5
|
| Patient Follow-up Rate |
13 patient had 1-month follow-up data available at the time of data cutoff.
|