The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

TOUCH CMC 1 New Enrollment Study


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Delayed
Application Number /
Requirement Number
P240020 / PAS002
Date Original Protocol Accepted 09/05/2025
Date Current Protocol Accepted 12/17/2025
Study Name TOUCH CMC 1 New Enrollment Study
Device Name TOUCH® CMC 1 Prosthesis
Clinical Trial Number(s) NCT01954433 
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group Objective Performance Criterion
Analysis Type Analytical
Study Population Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives Prospective, multi-center, single-arm observational post-approval study (PAS) with hypothesis testing.
The purpose of this study is to meet a PMA (P240020) condition of approval of the TOUCH® CMC 1 Prosthesis to evaluate the effectiveness and safety of the prosthesis in the United States (US) population. The device performance in newly enrolled US patients implanted by US surgeons will be compared to device performance observed within the outside the US (OUS) premarket cohort.
Study Population US Adult patients requiring surgical intervention for treating 1st CMC joint osteoarthritis that are indicated for the device, do not have any contraindications and who have failed non-conservative treatment.
Sample Size Minimum of 163 subjects
Up to 8 US sites, with no single site enrolls greater than 25% of subjects
Key Study Endpoints A composite clinical success (CSS) endpoint will be evaluated at 24 months/2 years post-implantation.

To be considered as a success, a subject must meet all of the following criteria:
Improvement of pain: A clinically meaningful improvement in pain during activities defined as a decrease in the Numerical Rating Scale (NRS) pain score of greater than or equal to 30% on a 10-point scale.
Maintenance or improvement in function: defined by key pinch strength, which is greater than or equal to 85% of the subject’s pre-operative key pinch strength.
Safety: success is defined as freedom from:
o Subsequent surgical interventions (SSI) (i.e., reoperation, revision, removal of any study component, or supplemental fixations) on the study carpometacarpal joint, or
Serious, device- or procedure-related adverse events
Secondary endpoints will be evaluated through 2 years.
Secondary endpoints will be evaluated through 2 years.
Pain at rest
Pain during activities
Key pinch (kg)
Kapandji index
Brief MHQ
Range of Motion (ROM):
Thumb Metacarpal Flexion
Thumb Metacarpal Radial Abduction
Thumb Metacarpal Palmar Abduction
Thumb length
PROMIS® (Patient-Reported Outcomes Measurement Information System) Upper Extremity Computer Adaptive Test (CAT)
PROMIS® Pain Interference CAT
EQ-5D-5L
EQ-5D-VAS
Return to work (days) in those working at baseline
Patient Satisfaction, including aesthetic
Radiographic endpoints:
Device migration
Osteolysis
Radiolucencies
Fracture
Individual components of the safety endpoint, including SSI(s) and adverse events

Follow-up Visits and Length of Follow-up Screening/Baseline (within 90 days of surgery)
Surgery
6 weeks -/+ 14 days
3 months -/+ 14 days
12 months -/+ 60 days
24 months -/+ 60 days
Patients are to be consented to 5-years for additional follow-up and evaluation, if necessary.
Interim or Final Data Summary
Interim Results Follow-up Rate: N/A
Safety Results: N/A
Actual Number of Patients Enrolled 7
Actual Number of Sites Enrolled 4
Patient Follow-up Rate 4.29% (7/163)


TOUCH CMC 1 New Enrollment Study Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
6 month report 01/08/2026 01/09/2026 On Time
1 year report 07/10/2026 06/22/2026 On Time
2 year report 07/10/2027    
3 year report 07/09/2028    
4 year report 07/09/2029    