The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

WATCHMAN FLX Pro OPTION PAS


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Ongoing
Application Number /
Requirement Number
P130013 S074/ PAS001
Date Original Protocol Accepted 12/15/2025
Date Current Protocol Accepted 12/15/2025
Study Name WATCHMAN FLX Pro OPTION PAS
Device Name WATCHMAN FLX Left Atrial Appendage (LAA) Closure Technology, WATCHMAN FLX Left Atrial Appendage Closure Device with Delivery System, WATCHMAN FLX™ Pro Left Atrial Appendage Closure (LAAC) Device with Delivery System
Clinical Trial Number(s) NCT02699957   NCT03795298  
General Study Protocol Parameters
Study Design Comprehensive/Linked/RegistryBased Surveillance
Data Source External Registry
Comparison Group No Control
Analysis Type Descriptive
Study Population Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives This is an observational, prospective, non-randomized, multicenter registry surveillance analysis designed to characterize clinical outcomes and to assess the real-world use of the commercial WATCHMAN FLX Pro LAA Closure Device with Delivery System. The analysis is designed to assess safety outcomes associated with the use of non-vitamin K oral anticoagulant (NOAC) alone as an alternative to an oral anticoagulant (OAC) and aspirin as the post-implant medication regimen for the WATCHMAN FLX Pro Left Atrial Appendage (LAA) Closure Device with Delivery System in a routine clinical setting. Data will be captured from the Left Atrial Appendage Occlusion (LAAO) Registry within the American College of Cardiology Foundation’s (ACCF) National Cardiovascular Data Registry (NCDR) conducted in the United States.
Study Population LAAO registry subjects who were successfully implanted with a WATCHMAN FLX Pro device, discharged 6 months prior to the NOAC alone labeling change implementation (16Jan2025 to 16Jul2025) or 6 months post labeling change implementation (17Jul2025 to 17Jan2026), and who meet inclusion/exclusion criteria.
Sample Size Number of subjects: N/A. See above study population
Assumptions for sample size estimation: N/A
Number of sites: Data from all sites participating in the LAAO Registry in the United States will be used for the analysis.
Sites location: US
Key Study Endpoints Co-Primary Endpoint: Composite Endpoint
The composite event rate is defined as all-cause death, stroke (ischemic and/or hemorrhagic), major bleeding, and systemic embolism. The composite event endpoint will be assessed in successfully implanted subjects between the date of hospital discharge and 45 days post-implant procedure and between hospital discharge and 6 months post-implant procedure.

Co-Primary Endpoint: Safety Endpoint
The occurrence of one of the following events between the time of the first implant attempt procedure and within seven days of that procedure or by hospital discharge, whichever is later:
• All-cause death,
• Ischemic stroke,
• Systemic embolism, or
• Device or procedure-related events requiring open cardiac
surgery or major endovascular intervention such as pseudoaneurysm repair, AV fistula repair, or other major
endovascular repair.

Percutaneous catheter drainage of pericardial effusions, snaring of an embolized device, thrombin injection to treat femoral pseudoaneurysm and nonsurgical treatments of access site complications are excluded from this endpoint and will be reported separately.

Co-Primary Endpoint: Effectiveness Endpoint
The occurrence of ischemic stroke or systemic embolism at 24 months post-implant procedure.

Additional Endpoints
Additional analyses will include the evaluation of the following events prior to and after the NOAC alone labeling implementation.

Clinical events will be assessed between hospital discharge and 45 days, 6 months, 12 months and 24 months post-implant procedure:
• All-cause death
• Cardiovascular death
• All stroke
• Ischemic stroke
• Hemorrhagic stroke
• Systemic thromboembolism
• Major bleeding
• Pericardial effusion
• Device embolization
• Device margin residual leak
• Device-related thrombus
Follow-up Visits and Length of Follow-up Subjects in all cohorts will be followed at hospital discharge, 45 days, 6 months, 12 months and 24 months post-implant procedure.
Interim or Final Data Summary
Interim Results Safety Results:
Implanted subjects
All-cause death:
- Concomitant cohort 0.08% (6/7679)
- Sequential cohort 0.00% (0/1490)
Cardiovascular death:
- Concomitant cohort 0.08% (6/7679)
- Sequential cohort 0.00% (0/1490)
Ischemic stroke:
- Concomitant cohort 0.17% (13/7679)
- Sequential cohort 0.00% (0/1490)
Hemorrhagic stroke:
- Concomitant cohort 0.01% (1/7679)
- Sequential cohort 0.00% (0/1490)
Systemic thromboembolism:
- Concomitant cohort 0.08% (6/7679)
- Sequential cohort 0.00% (0/1490)
Major Bleeding:
- Concomitant cohort 0.94% (72/7679)
- Sequential cohort 0.40% (6/1490)
Pericardial effusion:
- Concomitant cohort 0.20% (15/7679)
- Sequential cohort 0.01% (1/1490)
Device embolization:
- Concomitant cohort 0.00% (0/7679)
- Sequential cohort 0.00% (0/1490)
Device margin residual leak:
Complete seal was 99.3% in the concomitant cohort and 99.5% in the sequential cohort after the procedure, leading to an effective device closure (device margin residual leak <5mm) rate of 100% after the procedure in ITT concomitant and sequential patients with available imaging data.
In the Implanted data set, device related thrombus occurred in 0.02% (1/6519) of Concomitant Cohort subjects and 0.07% (1/1351) of Sequential Cohort subjects who had a TEE or CT at discharge.
Effectiveness Results:
Safety and effectiveness endpoints were reported as a composite endpoint; see the safety results section for study outcomes.
Actual Number of Patients Enrolled 9,169 Impanted subjects
- 7,679 subjects who were successfully implanted with a WATCHMAN FLX Pro device during a concomitant ablation procedure
- 1,490 subjects successfully implanted with a WATCHMAN FLX Pro device following a historical ablation procedure that occurred between 3-12 months prior to the WATCHMAN FLX Pro implant
Actual Number of Sites Enrolled N/A -LAOO Registry
Patient Follow-up Rate Subject follow-up is currently ongoing. Implanted 45-day follow-up rate for the Concomitant Cohort was 45.9% and the Sequential Cohort was 51.4%.


WATCHMAN FLX Pro OPTION PAS Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
6 month report 01/14/2026 01/09/2026 On Time
1 year report 07/16/2026 07/14/2026 On Time
2 year report 07/16/2027    
3 year report 07/15/2028    