The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

The GORE VIABAHN FORTEGRA Venous Stent New Enrollment Study


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Study Pending
Application Number /
Requirement Number
P250017 / PAS002
Date Original Protocol Accepted 04/07/2026
Date Current Protocol Accepted 04/07/2026
Study Name The GORE VIABAHN FORTEGRA Venous Stent New Enrollment Study
Device Name GORE® VIABAHN® FORTEGRA Venous Stent
Clinical Trial Number(s) NCT05409976 NCT05489588 
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group No Control
Analysis Type Descriptive
Study Population Transit.Adolescent B(as adults): 18 yrs < 22 yrs, Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives A prospective, multicenter, single-arm, post-market clinical study. To evaluate the Gore Viabahn Fortegra Venous Stent in real world use.
Study Population Adult patients (at least 18 years of age) intended to receive the Gore Viabahn Fortegra Venous Stent in endovascular treatment. Only patients who meet all the inclusion criteria and none of the exclusion criteria will be implanted. Significant Inclusion Criteria Preoperative
- Presence of lesion expected to necessitate treatment with the Gore Viabahn Fortegra Venous Stent. Intraoperative
- Patient has adequate inflow/outflow to/from the target lesion(s), per investigator/sub-investigator discretion.
Significant Exclusion Criteria Intraoperative Patient is contraindicated for treatment with the Gore Viabahn Fortegra Venous Stent according to the IFU. Intent to Treat population: All subjects who had a study device introduced into the vasculature. Per Protocol population: All subjects who had a study device introduced into the vasculature and were not determined to be ineligible for the study.
Sample Size Number of subjects: 150 subjects with the Gore Viabahn Fortegra device introduced into the vasculature, with a limit of 30 treated subjects per site Assumptions for sample size estimation: A sample size of 150 subjects with a study device introduced into the vasculature will be used to ensure broad representation of device usage. Number of sites: 20 sites up to a maximum of 35 sites will be enrolled Sites location: United States
Key Study Endpoints Primary Effectiveness
- Primary patency through 12 months Primary Safety
- Device Freedom from device- or index procedure-related SAEs through 30 days Secondary Endpoints
- Primary patency through 36 months
- Secondary patency through 36 months
- Clinically Freedom from clinically driven target lesion revascularization through 36 months
- Stent embolization through 12-month follow-up
- Device- or index procedure-related death through 30 days
- Clinically significant pulmonary embolism through 30 days
- Device- or index procedure-related vascular injury through 30 days requiring surgical or endovascular intervention
- Device- or index procedure-related major bleeding events through 30 days
- Change in EQ-5D-5L Measurement through 36-month follow-up compared to baseline prior to treatment
- Technical success
- Lesion success Additional assessment at 1, 6, 12, 24 and 36 months:
- Adverse events
- Concomitant medications
- rVCSS and Villalta (applicable only to subjects intended to be treated in the IVC or iliofemoral veins)
- Imaging of lesions /stents
Follow-up Visits and Length of Follow-up Dischage,1, 6-, 12-, 24-, and 36 months post treatment Subjects Lost to follow-up A subject is considered lost to follow-up after 2 consecutive follow-up visits, and 3 documented attempts have been made by the investigator or designee to contact the subject or next of kin. One of the three documented attempts must include a certified letter.


The GORE VIABAHN FORTEGRA Venous Stent New Enrollment Study Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
6 month report 06/19/2026 06/16/2026 On Time
1 year report 12/19/2026    
18 month report 06/19/2027    
2 year report 12/19/2027    
3 year report 12/18/2028    
4 year report 12/18/2029    


Contact Us

Mandated Studies Program
Food and Drug Administration
10903 New Hampshire Ave.
Silver Spring, MD 20993-0002
Email: MandatedStudiesPrograms@fda.hhs.gov

Additional Resources

-
-