The FDA has the authority to require sponsors to perform a post-approval study (or studies) at the time of approval of a premarket approval (PMA), humanitarian device exemption (HDE), or product development protocol (PDP) application. Post-approval studies can provide patients, health care professionals, the device industry, the FDA and other stakeholders information on the continued safety and effectiveness (or continued probable benefit, in the case of an HDE) of approved medical devices. This database allows you to search Post-Approval Study information by applicant or device information.

Learn more...


           

Navastar Thermocool


Suggest Enhancement / Report Issue | export reports to excelExport to Excel
General
Study Status Completed
Application Number /
Requirement Number
P030031 / PAS001
Date Original Protocol Accepted 12/19/2005
Date Current Protocol Accepted 12/19/2005
Study Name Navastar Thermocool
Device Name BIOSENSE WEBSTER NAVISTAR/CELSIUS THERMO COOL DIAGNOSTIC/ABLATION DEFLECTABLE TIP CATHETERS
General Study Protocol Parameters
Study Design Prospective Cohort Study
Data Source New Data Collection
Comparison Group Objective Performance Criterion
Analysis Type Analytical
Study Population Transit.Adolescent B(as adults): 18 yrs < 22 yrs, Adult: At least 22 yrs
Detailed Study Protocol Parameters
Study Objectives This was a multi-center, observational, one-armed prospective cohort study of patietns with atrial flutter treated with the device.
Study Population Subjects from several different centers with symptomatic typical atrial flutter who meet the inclusion/exclusion criteria. It is indicated for catheter-based cardiac electrophysiological mapping (stimulating and recording), and when used with the Stockert 70 generator, for the treatment of Type I atrial flutter in patients age 18 or older.
Sample Size 291 subjects at 21 sites
Key Study Endpoints The primary effectiveness endpoint was confirmation of complete bidirectional conduction block across the subeustachian isthmus at a minimum of 30 minutes following the last radio frequency application. Success was considered to be 88% effectiveness, with a one-sided 95% lower confidence bound of 80%. The primary safety endpoint was the percentage of subjects experiencing cardiovascular specific adverse events within seven days of the ablation procedure. Success was considered to be a CSAE no greater than 2.7%, with one-sided 95% confidence level no greater than 7%.
Follow-up Visits and Length of Follow-up One month, one year, and annually thereafter for a total follow-up of 5 years.
Interim or Final Data Summary
Actual Number of Patients Enrolled 291 patients
Actual Number of Sites Enrolled 21 sites
Patient Follow-up Rate 100%
Final Safety Findings Cardiac specific adverse events occurred in 4 of 267 subjects (two complete heart blocks, one pulmonary embolus, one myocardial infarction) in the safety cohort (1.5%). The upper one-side 95% confidence bound was 3.6%. All cardiac specific adverse events occurred in the primary group, in which the rate was 1.7% (upper 95% bound of 3.8%). Only one of the 4 was determined to be related to the device or the procedure. All 4 resolved. There were no deaths or unanticipated adverse events.
Study Strengths & Weaknesses The strength of the study was the excellent follow-up and prospective cohort study design.
Recommendations for Labeling Changes None


Navastar Thermocool Reporting Schedule

Reporting Schedule
Report
Date Due
FDA Receipt
Date
Applicant's Reporting Status
18 month report 06/19/2007 06/19/2007 On Time
Final PAS Report 03/31/2008 03/31/2008 On Time
2 year Report 11/05/2008 12/20/2007 Overdue/Received