| Class 2 Device Recall Clinac, Trilogy. Trilogy Tx and Novalis Tx Linear Accelerators |  |
Date Initiated by Firm | May 26, 2011 |
Date Posted | August 10, 2011 |
Recall Status1 |
Terminated 3 on July 13, 2012 |
Recall Number | Z-2978-2011 |
Recall Event ID |
59087 |
510(K)Number | K033343 K100890 |
Product Classification |
Linear Accelerator for Radiation Therapy System - Product Code IYE
|
Product | Clinac, Trilogy. Trilogy Tx and Novalis Tx Linear Accelerators.
Varian Medical Systems, Inc. Oncology Systems
Intended for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions in the body where radiation treatment is indicated. |
Code Information |
H14, H18, H27 and H29. Serial numbers: H140001 H140063 H140123 H140227 H140745 H140803 H140861 H140921 H180107 H180226 H180324 H180474 H180549 H140002 H140064 H140124 H140228 H140746 H140804 H140862 H140923 H180108 H180228 H180325 H180475 H180550 H140003 H140065 H140125 H140229 H140747 H140805 H140863 H140924 H180115 H180229 H180326 H180476 H180551 H140004 H140066 H140126 H140230 H140748 H140806 H140864 H140925 H180116 H180230 H180327 H180477 H180552 H140005 H140067 H140127 H140231 H140749 H140807 H140865 H140926 H180117 H180235 H180328 H180478 H180553 H140007 H140068 H140128 H140232 H140750 H140808 H140866 H140927 H180119 H180236 H180329 H180479 H180554 H140008 H140069 H140129 H140233 H140751 H140809 H140867 H140929 H180122 H180237 H180401 H180480 H180555 H140009 H140070 H140130 H140234 H140752 H140810 H140868 H140930 H180124 H180238 H180402 H180482 H180556 H140010 H140071 H140131 H140235 H140753 H140811 H140869 H140931 H180128 H180240 H180404 H180484 H180557 H140011 H140072 H140132 H140236 H140754 H140812 H140870 H140932 H180129 H180242 H180405 H180485 H180559 H140012 H140073 H140133 H140237 H140755 H140813 H140871 H140933 H180131 H180243 H180407 H180486 H180560 H140014 H140074 H140134 H140238 H140756 H140814 H140872 H140934 H180132 H180245 H180409 H180487 H180561 H140015 H140075 H140135 H140239 H140757 H140815 H140873 H140935 H180133 H180246 H180410 H180489 H180562 H140016 H140076 H140136 H140240 H140758 H140816 H140874 H140936 H180135 H180249 H180411 H180490 H180563 H140017 H140077 H140137 H140700 H140759 H140817 H140875 H140938 H180143 H180250 H180412 H180491 H180564 H140018 H140078 H140138 H140701 H140760 H140818 H140876 H140940 H180144 H180251 H180413 H180492 H180565 H140019 H140079 H140139 H140702 H140761 H140819 H140877 H140941 H180147 H180254 H180414 H180493 H180566 H140021 H140080 H140140 H140703 H140762 H140820 H140878 H140944 H180148 H180255 H180417 H180494 H180567 H140022 H140082 H140141 H140704 H140763 H140821 H140879 H140945 H180151 H180256 H180419 H180495 H180569 H140023 H140083 H140142 H140705 H140764 H140822 H140881 H140947 H180154 H180257 H180420 H180496 H180570 H140024 H140084 H140143 H140706 H140765 H140823 H140882 H180003 H180157 H180258 H180422 H180498 H180571 H140025 H140085 H140144 H140707 H140766 H140824 H140883 H180004 H180158 H180260 H180424 H180499 H180572 H140026 H140086 H140145 H140708 H140767 H140825 H140884 H180007 H180159 H180262 H180425 H180501 H180573 H140027 H140087 H140146 H140709 H140768 H140826 H140885 H180008 H180161 H180264 H180426 H180505 H180574 H140028 H140088 H140147 H140710 H140769 H140827 H140886 H180016 H180164 H180267 H180427 H180507 H180575 H140029 H140089 H140148 H140711 H140770 H140828 H140887 H180017 H180166 H180268 H180432 H180508 H180576 H140030 H140090 H140149 H140712 H140771 H140829 H140888 H180019 H180167 H180271 H180433 H180509 H180577 H140031 H140091 H140150 H140713 H140772 H140830 H140889 H180025 H180168 H180272 H180436 H180510 H180578 H140032 H140092 H140151 H140714 H140773 H140831 H140890 H180026 H180170 H180273 H180437 H180511 H180579 H140033 H140093 H140152 H140715 H140774 H140832 H140891 H180028 H180171 H180275 H180440 H180513 H180580 H140034 H140094 H140153 H140716 H140775 H140833 H140892 H180040 H180173 H180283 H180441 H180514 H180581 H140035 H140095 H140154 H140717 H140776 H140834 H140893 H180043 H180178 H180284 H180442 H180515 H180583 H140036 H140096 H140201 H140718 H140777 H140835 H140895 H180044 H180179 H180285 H180443 H180518 H180584 H140037 H140097 H140202 H140719 H140778 H140836 H140896 H180045 H180181 H180287 H180444 H180519 H180585 H140038 H140098 H140203 H140720 H140779 H140837 H140897 H180048 H180182 H180289 H180445 H180520 H180586 H140039 H140099 H140204 H140721 H140780 H140838 H140898 H180049 H180185 H180290 H180446 H180521 H180587 H140040 H140100 H140205 H140722 H140781 H140839 H140899 H180050 H180186 H180291 H180447 H180522 H180588 H140041 H140101 H140206 H140723 H140782 H140840 H140900 H180052 H180188 H180292 H180449 H180524 H180590 H140042 H140102 H140207 H140724 H140783 H140841 H140901 H180053 H180192 H180293 H180450 H180525 H180591 H140043 H140103 H140208 H140725 H140784 H140842 H140902 H180056 H180193 H180294 H180451 H180527 H180592 H140044 H140105 H140209 H140726 H140785 H140843 H140903 H180058 H180195 H180295 H180452 H180528 H180593 H140045 H140106 H140210 H140727 H140786 H140844 H140904 H180059 H180197 H180296 H180453 H180529 H180595 H140046 H140107 H140211 H140728 H140787 H140845 H140905 H180065 H180199 H180298 H180454 H180530 H180596 H140047 H140108 H140212 H140729 H140788 H140846 H140906 H180066 H180201 H180299 H180455 H180531 H180597 H140048 H140109 H140213 H140730 H140789 H140847 H140907 H180073 H180203 H180303 H180456 H180532 H180598 H140049 H140110 H140214 H140731 H140790 H140848 H140908 H180075 H180204 H180304 H180457 H180533 H180599 H140051 H140111 H140215 H140732 H140791 H140849 H140909 H180079 H180206 H180305 H180459 H180534 H180600 H140052 H140112 H140216 H140733 H140792 H140850 H140910 H180080 H180208 H180307 H180460 H180535 H180601 H140053 H140113 H140217 H140734 H140793 H140851 H140911 H180082 H180209 H180309 H180461 H180536 H180603 H140054 H140114 H140218 H140735 H140794 H140852 H140912 H180083 H180210 H180310 H180463 H180538 H180604 H140055 H140115 H140219 H140736 H140795 H140853 H140913 H180087 H180214 H180311 H180466 H180539 H180605 H140056 H140116 H140220 H140737 H140796 H140854 H140914 H180088 H180215 H180312 H180467 H180540 H180606 H140057 H140117 H140221 H140738 H140797 H140855 H140915 H180093 H180217 H180314 H180468 H180541 H180607 H140058 H140118 H140222 H140739 H140798 H140856 H140916 H180098 H180218 H180318 H180469 H180542 H180608 H140059 H140119 H140223 H140740 H140799 H140857 H140917 H180099 H180220 H180319 H180470 H180543 H180613 H140060 H140120 H140224 H140742 H140800 H140858 H140918 H180102 H180221 H180321 H180471 H180544 H180614 H140061 H140121 H140225 H140743 H140801 H140859 H140919 H180103 H180222 H180322 H180472 H180545 H180616 H140062 H140122 H140226 H140744 H140802 H140860 H140920 H180106 H180225 H180323 H180473 H180548 H180618 H180619 H180693 H180776 H180856 H181030 H181108 H181181 H270134 H270353 H270474 H270635 H270739 H270845 H180620 H180694 H180777 H180857 H181031 H181109 H181182 H270136 H270355 H270476 H270637 H270740 H270846 H180621 H180695 H180778 H180858 H181032 H181112 H181183 H270138 H270356 H270477 H270639 H270741 H270847 H180622 H180696 H180779 H180859 H181033 H181113 H181184 H270144 H270360 H270479 H270641 H270746 H270848 H180624 H180697 H180780 H180860 H181036 H181114 H181185 H270201 H270361 H270482 H270642 H270749 H270852 H180625 H180698 H180781 H180861 H181038 H181115 H181187 H270209 H270364 H270483 H270643 H270751 H270854 H180626 H180699 H180784 H180862 H181039 H181116 H181189 H270211 H270365 H270484 H270644 H270752 H270855 H180627 H180703 H180785 H180863 H181041 H181117 H181191 H270214 H270370 H270486 H270645 H270755 H270857 H180628 H180704 H180786 H180864 H181042 H181118 H181192 H270215 H270371 H270487 H270646 H270757 H270858 H180629 H180705 H180788 H180866 H181043 H181119 H181193 H270216 H270373 H270489 H270647 H270760 H270859 H180630 H180706 H180791 H180867 H181045 H181120 H181194 H270219 H270375 H270490 H270648 H270761 H270861 H180631 H180707 H180792 H180868 H181046 H181121 H181195 H270227 H270378 H270491 H270649 H270762 H270863 H180632 H180709 H180793 H180869 H181047 H181124 H181196 H270228 H270380 H270495 H270651 H270763 H270864 H180633 H180710 H180794 H180870 H181048 H181125 H181197 H270229 H270384 H270498 H270652 H270766 H270865 H180634 H180715 H180795 H180871 H181049 H181126 H181198 H270230 H270385 H270499 H270653 H270767 H270866 H180635 H180716 H180796 H180872 H181052 H181127 H181199 H270232 H270386 H270500 H270654 H270769 H270867 H180636 H180717 H180803 H180873 H181053 H181128 H181200 H270235 H270391 H270501 H270656 H270777 H270868 H180637 H180718 H180804 H180874 H181055 H181129 H181201 H270240 H270392 H270502 H270660 H270778 H270869 H180639 H180719 H180805 H180877 H181056 H181130 H181202 H270242 H270395 H270503 H270661 H270781 H270870 H180640 H180720 H180806 H180878 H181057 H181132 H181203 H270244 H270397 H270504 H270662 H270782 H270871 H180641 H180721 H180807 H180879 H181058 H181133 H181205 H270245 H270398 H270505 H270663 H270783 H270872 H180642 H180722 H180808 H180880 H181059 H181134 H181209 H270251 H270399 H270506 H270664 H270785 H270873 H180643 H180723 H180809 H180881 H181060 H181135 H181210 H270252 H270400 H270508 H270665 H270786 H270875 H180644 H180724 H180810 H180882 H181061 H181136 H181211 H270254 H270401 H270511 H270667 H270787 H270877 H180645 H180725 H180811 H180883 H181062 H181137 H181212 H270255 H270404 H270512 H270669 H270788 H270878 H180646 H180726 H180812 H180884 H181065 H181138 H182000 H270256 H270409 H270513 H270670 H270789 H270879 H180647 H180728 H180815 H180885 H181066 H181139 H182001 H270260 H270410 H270514 H270671 H270791 H270880 H180649 H180729 H180816 H180886 H181067 H181140 H182002 H270261 H270412 H270516 H270675 H270793 H270881 H180650 H180730 H180817 H180887 H181068 H181142 H182003 H270264 H270416 H270518 H270680 H270794 H270882 H180651 H180731 H180819 H180888 H181069 H181143 H182005 H270267 H270418 H270520 H270681 H270796 H270883 H180652 H180733 H180820 H180890 H181070 H181145 H182006 H270271 H270419 H270522 H270682 H270797 H270884 H180653 H180734 H180821 H180891 H181071 H181146 H182007 H270273 H270420 H270525 H270684 H270799 H270885 H180654 H180735 H180822 H180894 H181073 H181147 H182008 H270274 H270421 H270526 H270685 H270800 H270886 H180655 H180736 H180823 H180895 H181074 H181148 H182009 H270277 H270425 H270529 H270686 H270801 H270888 H180657 H180737 H180824 H180896 H181075 H181149 H182010 H270280 H270426 H270530 H270690 H270803 H270889 H180659 H180738 H180825 H180898 H181076 H181151 H182012 H270281 H270427 H270532 H270692 H270804 H270890 H180660 H180739 H180826 H180899 H181077 H181152 H182016 H270282 H270428 H270534 H270693 H270805 H270891 H180661 H180741 H180827 H180901 H181078 H181153 H182017 H270285 H270430 H270535 H270697 H270806 H270892 H180662 H180742 H180828 H180907 H181079 H181154 H270004 H270290 H270431 H270539 H270698 H270808 H270893 H180663 H180744 H180829 H181002 H181082 H181155 H270009 H270293 H270433 H270540 H270699 H270812 H270894 H180664 H180745 H180830 H181003 H181083 H181156 H270010 H270297 H270435 H270542 H270700 H270813 H270895 H180665 H180746 H180831 H181004 H181084 H181157 H270021 H270302 H270436 H270544 H270705 H270814 H270896 H180667 H180747 H180832 H181005 H181085 H181158 H270026 H270303 H270437 H270547 H270708 H270817 H270897 H180668 H180749 H180833 H181006 H181086 H181159 H270027 H270307 H270438 H270550 H270709 H270819 H270898 H180669 H180750 H180834 H181007 H181087 H181160 H270029 H270310 H270439 H270601 H270710 H270820 H270899 H180670 H180751 H180835 H181008 H181089 H181161 H270030 H270314 H270443 H270602 H270711 H270821 H271000 H180671 H180752 H180836 H181009 H181090 H181162 H270038 H270320 H270445 H270605 H270712 H270822 H271002 H180672 H180753 H180837 H181010 H181091 H181163 H270045 H270321 H270446 H270606 H270714 H270823 H271004 H180674 H180754 H180838 H181011 H181092 H181164 H270047 H270323 H270447 H270607 H270715 H270825 H271005 H180676 H180756 H180839 H181012 H181093 H181165 H270073 H270324 H270448 H270613 H270716 H270826 H271006 H180678 H180757 H180840 H181013 H181094 H181166 H270074 H270329 H270450 H270615 H270718 H270827 H271009 H180679 H180758 H180841 H181015 H181095 H181167 H270079 H270332 H270451 H270616 H270719 H270828 H271013 H180680 H180759 H180842 H181018 H181096 H181168 H270088 H270335 H270452 H270622 H270722 H270829 H271015 H180681 H180760 H180843 H181019 H181097 H181170 H270094 H270337 H270454 H270623 H270725 H270833 H271016 H180682 H180761 H180844 H181020 H181098 H181173 H270096 H270338 H270455 H270624 H270726 H270835 H271017 H180683 H180762 H180848 H181021 H181101 H181174 H270106 H270341 H270458 H270625 H270727 H270836 H271018 H180684 H180763 H180850 H181022 H181102 H181175 H270107 H270343 H270461 H270626 H270728 H270837 H271019 H180685 H180765 H180851 H181023 H181103 H181176 H270110 H270344 H270462 H270627 H270732 H270838 H271020 H180687 H180768 H180852 H181026 H181104 H181177 H270111 H270345 H270464 H270631 H270733 H270840 H271021 H180690 H180769 H180853 H181027 H181105 H181178 H270118 H270346 H270466 H270632 H270735 H270841 H271023 H180691 H180774 H180854 H181028 H181106 H181179 H270126 H270351 H270470 H270633 H270736 H270842 H271024 H180692 H180775 H180855 H181029 H181107 H181180 H270128 H270352 H270473 H270634 H270737 H270844 H271025 H271026 H271113 H271199 H271275 H271342 H271413 H271483 H271549 H271616 H271682 H271748 H271810 H271872 H271027 H271114 H271200 H271276 H271343 H271414 H271484 H271550 H271617 H271683 H271749 H271811 H271873 H271029 H271115 H271201 H271277 H271344 H271415 H271485 H271551 H271618 H271684 H271750 H271812 H271874 H271031 H271116 H271202 H271278 H271345 H271416 H271486 H271552 H271619 H271685 H271751 H271813 H271875 H271032 H271117 H271204 H271280 H271347 H271417 H271487 H271553 H271620 H271686 H271752 H271814 H271876 H271033 H271118 H271205 H271281 H271349 H271418 H271488 H271554 H271621 H271687 H271753 H271815 H271877 H271034 H271120 H271206 H271282 H271350 H271419 H271490 H271555 H271622 H271688 H271754 H271816 H271878 H271035 H271121 H271207 H271283 H271351 H271420 H271492 H271557 H271624 H271689 H271755 H271817 H271879 H271037 H271123 H271208 H271284 H271353 H271421 H271493 H271559 H271625 H271691 H271756 H271818 H271880 H271038 H271125 H271209 H271285 H271355 H271422 H271494 H271560 H271626 H271692 H271757 H271819 H271881 H271039 H271127 H271210 H271286 H271356 H271423 H271495 H271561 H271627 H271693 H271758 H271820 H271882 H271041 H271131 H271211 H271287 H271357 H271424 H271497 H271562 H271628 H271695 H271759 H271821 H271883 H271042 H271132 H271212 H271288 H271358 H271425 H271498 H271563 H271629 H271696 H271760 H271822 H271884 H271043 H271135 H271214 H271289 H271359 H271426 H271499 H271564 H271630 H271697 H271761 H271823 H271885 H271044 H271136 H271216 H271290 H271360 H271427 H271500 H271565 H271631 H271698 H271762 H271824 H271886 H271051 H271137 H271217 H271291 H271361 H271428 H271501 H271566 H271632 H271699 H271763 H271825 H271887 H271052 H271139 H271218 H271292 H271362 H271429 H271502 H271567 H271633 H271700 H271764 H271826 H271888 H271055 H271141 H271219 H271293 H271363 H271430 H271503 H271568 H271634 H271701 H271765 H271827 H271889 H271056 H271142 H271222 H271294 H271364 H271431 H271504 H271569 H271635 H271702 H271766 H271828 H271890 H271057 H271143 H271223 H271295 H271365 H271432 H271505 H271570 H271636 H271703 H271767 H271829 H271891 H271059 H271144 H271224 H271296 H271367 H271433 H271506 H271571 H271637 H271704 H271768 H271830 H271892 H271061 H271145 H271225 H271297 H271369 H271434 H271507 H271572 H271638 H271705 H271769 H271831 H271893 H271062 H271149 H271226 H271298 H271370 H271435 H271508 H271573 H271639 H271706 H271770 H271832 H271894 H271063 H271151 H271227 H271299 H271371 H271436 H271509 H271574 H271640 H271707 H271771 H271833 H271895 H271065 H271152 H271228 H271300 H271372 H271437 H271510 H271575 H271641 H271708 H271772 H271834 H271896 H271066 H271153 H271229 H271301 H271373 H271438 H271511 H271576 H271642 H271709 H271773 H271835 H271897 H271067 H271154 H271230 H271302 H271374 H271439 H271512 H271577 H271643 H271710 H271774 H271836 H271898 H271068 H271155 H271231 H271303 H271375 H271440 H271513 H271578 H271644 H271711 H271775 H271837 H271899 H271071 H271157 H271232 H271304 H271376 H271441 H271514 H271580 H271645 H271712 H271776 H271838 H271901 H271073 H271159 H271233 H271305 H271377 H271442 H271515 H271581 H271646 H271713 H271777 H271839 H271902 H271075 H271160 H271234 H271306 H271378 H271443 H271516 H271582 H271647 H271714 H271778 H271840 H271903 H271076 H271161 H271235 H271307 H271379 H271444 H271517 H271583 H271648 H271715 H271779 H271841 H271904 H271079 H271162 H271236 H271308 H271380 H271445 H271518 H271586 H271649 H271716 H271780 H271842 H271905 H271080 H271163 H271238 H271309 H271381 H271447 H271519 H271587 H271650 H271717 H271781 H271843 H271906 H271082 H271165 H271239 H271311 H271382 H271448 H271520 H271588 H271651 H271718 H271782 H271844 H271907 H271083 H271166 H271241 H271312 H271383 H271449 H271521 H271589 H271652 H271719 H271783 H271845 H271908 H271084 H271167 H271242 H271314 H271384 H271450 H271522 H271590 H271653 H271720 H271784 H271846 H271909 H271085 H271168 H271243 H271315 H271385 H271451 H271523 H271591 H271654 H271722 H271785 H271847 H271910 H271086 H271171 H271245 H271316 H271386 H271453 H271525 H271592 H271655 H271723 H271786 H271848 H271911 H271088 H271172 H271247 H271317 H271387 H271454 H271526 H271593 H271656 H271724 H271787 H271849 H271912 H271091 H271174 H271248 H271318 H271388 H271456 H271527 H271594 H271657 H271725 H271788 H271850 H271913 H271092 H271175 H271249 H271319 H271389 H271458 H271528 H271595 H271658 H271726 H271789 H271851 H271914 H271093 H271176 H271250 H271320 H271390 H271459 H271529 H271596 H271659 H271727 H271790 H271852 H271915 H271094 H271177 H271251 H271321 H271391 H271460 H271530 H271597 H271660 H271728 H271791 H271853 H271916 H271095 H271178 H271253 H271322 H271392 H271461 H271531 H271598 H271662 H271729 H271792 H271854 H271917 H271096 H271179 H271256 H271323 H271393 H271462 H271532 H271599 H271663 H271730 H271793 H271855 H271918 H271097 H271182 H271257 H271324 H271394 H271463 H271533 H271600 H271664 H271731 H271794 H271856 H271919 H271098 H271183 H271258 H271325 H271395 H271464 H271534 H271601 H271666 H271732 H271795 H271857 H271920 H271099 H271184 H271259 H271327 H271397 H271465 H271535 H271602 H271667 H271733 H271796 H271858 H271921 H271100 H271185 H271260 H271328 H271398 H271466 H271536 H271603 H271669 H271734 H271797 H271859 H271922 H271101 H271186 H271263 H271329 H271399 H271467 H271537 H271604 H271670 H271735 H271798 H271860 H271923 H271102 H271188 H271264 H271330 H271401 H271468 H271538 H271605 H271671 H271736 H271799 H271861 H271924 H271103 H271189 H271265 H271331 H271402 H271470 H271539 H271606 H271672 H271737 H271800 H271862 H271925 H271104 H271190 H271266 H271332 H271403 H271472 H271540 H271607 H271673 H271738 H271801 H271863 H271926 H271105 H271191 H271267 H271333 H271404 H271473 H271541 H271608 H271674 H271739 H271802 H271864 H271927 H271106 H271192 H271268 H271334 H271406 H271474 H271542 H271609 H271675 H271740 H271803 H271865 H271928 H271107 H271193 H271269 H271335 H271407 H271475 H271543 H271610 H271676 H271741 H271804 H271866 H271929 H271108 H271194 H271270 H271336 H271408 H271476 H271544 H271611 H271677 H271742 H271805 H271867 H271930 H271109 H271195 H271271 H271337 H271409 H271477 H271545 H271612 H271678 H271743 H271806 H271868 H271931 H271110 H271196 H271272 H271338 H271410 H271478 H271546 H271613 H271679 H271744 H271807 H271869 H271932 H271111 H271197 H271273 H271339 H271411 H271480 H271547 H271614 H271680 H271746 H271808 H271870 H271933 H271112 H271198 H271274 H271340 H271412 H271481 H271548 H271615 H271681 H271747 H271809 H271871 H271934 H271935 H271997 H272064 H272127 H272189 H272251 H272316 H272378 H272440 H272807 H272871 H272934 H272997 H271936 H271998 H272065 H272128 H272190 H272253 H272317 H272379 H272441 H272808 H272872 H272935 H272998 H271937 H271999 H272066 H272129 H272191 H272254 H272318 H272380 H272442 H272809 H272873 H272936 H272999 H271938 H272000 H272067 H272130 H272192 H272255 H272319 H272381 H272443 H272810 H272875 H272937 H273000 H271939 H272001 H272068 H272131 H272193 H272256 H272320 H272382 H272444 H272811 H272876 H272938 H273001 H271940 H272002 H272069 H272132 H272194 H272257 H272321 H272383 H272445 H272812 H272877 H272939 H273002 H271941 H272003 H272070 H272133 H272195 H272258 H272322 H272384 H272446 H272813 H272878 H272940 H273003 H271942 H272004 H272071 H272134 H272196 H272259 H272323 H272385 H272447 H272814 H272879 H272941 H273004 H271943 H272005 H272072 H272135 H272197 H272260 H272324 H272386 H272448 H272815 H272880 H272942 H273005 H271944 H272006 H272073 H272136 H272198 H272261 H272325 H272387 H272449 H272816 H272881 H272943 H273006 H271945 H272007 H272074 H272137 H272199 H272262 H272326 H272388 H272450 H272817 H272882 H272944 H273007 H271946 H272008 H272075 H272138 H272200 H272263 H272327 H272389 H272451 H272818 H272883 H272946 H273008 H271947 H272009 H272076 H272139 H272201 H272264 H272328 H272390 H272452 H272819 H272884 H272947 H273010 H271948 H272010 H272077 H272140 H272202 H272265 H272329 H272391 H272453 H272820 H272885 H272948 H273011 H271949 H272011 H272078 H272141 H272203 H272266 H272330 H272392 H272454 H272821 H272886 H272949 H273012 H271950 H272012 H272079 H272142 H272204 H272268 H272331 H272393 H272455 H272822 H272887 H272950 H273013 H271951 H272013 H272080 H272143 H272205 H272269 H272332 H272394 H272456 H272823 H272888 H272951 H273014 H271952 H272014 H272081 H272144 H272206 H272270 H272333 H272395 H272457 H272824 H272889 H272952 H273016 H271953 H272015 H272082 H272145 H272207 H272271 H272334 H272396 H272458 H272825 H272890 H272953 H273017 H271954 H272016 H272083 H272146 H272208 H272272 H272335 H272397 H272459 H272826 H272891 H272954 H273018 H271955 H272017 H272084 H272147 H272209 H272273 H272336 H272398 H272460 H272827 H272892 H272955 H273019 H271956 H272018 H272085 H272148 H272210 H272274 H272337 H272399 H272461 H272828 H272893 H272956 H273020 H271957 H272019 H272086 H272149 H272211 H272276 H272338 H272400 H272462 H272829 H272894 H272957 H273021 H271958 H272020 H272087 H272150 H272212 H272277 H272339 H272401 H272463 H272830 H272895 H272958 H273022 H271959 H272021 H272088 H272151 H272213 H272278 H272340 H272402 H272464 H272831 H272896 H272959 H273023 H271960 H272022 H272089 H272152 H272214 H272279 H272341 H272403 H272465 H272833 H272897 H272960 H273024 H271961 H272023 H272090 H272153 H272215 H272280 H272342 H272404 H272466 H272834 H272898 H272961 H273025 H271962 H272024 H272091 H272154 H272216 H272281 H272343 H272405 H272467 H272836 H272899 H272962 H273026 H271963 H272025 H272092 H272155 H272217 H272282 H272344 H272406 H272468 H272837 H272900 H272963 H273027 H271964 H272026 H272093 H272156 H272218 H272283 H272345 H272407 H272469 H272838 H272901 H272964 H273028 H271965 H272027 H272094 H272157 H272219 H272284 H272346 H272408 H272470 H272839 H272902 H272965 H273029 H271966 H272028 H272095 H272158 H272220 H272285 H272347 H272409 H272471 H272840 H272903 H272966 H273030 H271967 H272029 H272096 H272159 H272221 H272286 H272348 H272410 H272472 H272841 H272904 H272967 H273031 H271968 H272030 H272097 H272160 H272222 H272287 H272349 H272411 H272473 H272842 H272905 H272968 H273032 H271969 H272031 H272098 H272161 H272223 H272288 H272350 H272412 H272475 H272843 H272906 H272969 H273033 H271970 H272032 H272099 H272162 H272224 H272289 H272351 H272413 H272476 H272844 H272907 H272970 H273034 H271971 H272033 H272100 H272163 H272225 H272290 H272352 H272414 H272477 H272845 H272908 H272971 H273035 H271972 H272034 H272101 H272164 H272226 H272291 H272353 H272415 H272478 H272846 H272909 H272972 H273036 H271973 H272035 H272102 H272165 H272227 H272292 H272354 H272416 H272479 H272847 H272910 H272973 H273037 H271974 H272036 H272103 H272166 H272228 H272293 H272355 H272417 H272480 H272848 H272911 H272974 H273038 H271975 H272039 H272104 H272167 H272229 H272294 H272356 H272418 H272482 H272849 H272912 H272975 H273039 H271976 H272040 H272106 H272168 H272230 H272295 H272357 H272419 H272483 H272850 H272913 H272976 H273040 H271977 H272041 H272107 H272169 H272231 H272296 H272358 H272420 H272484 H272851 H272914 H272977 H273041 H271978 H272043 H272108 H272170 H272232 H272297 H272359 H272421 H272485 H272852 H272915 H272978 H273042 H271979 H272045 H272109 H272171 H272233 H272298 H272360 H272422 H272488 H272853 H272916 H272979 H273043 H271980 H272046 H272110 H272172 H272234 H272299 H272361 H272423 H272489 H272854 H272917 H272980 H273044 H271981 H272047 H272111 H272173 H272235 H272300 H272362 H272424 H272490 H272855 H272918 H272981 H273045 H271982 H272048 H272112 H272174 H272236 H272301 H272363 H272425 H272491 H272856 H272919 H272982 H273046 H271983 H272049 H272113 H272175 H272237 H272302 H272364 H272426 H272492 H272857 H272920 H272983 H273047 H271984 H272050 H272114 H272176 H272238 H272303 H272365 H272427 H272493 H272858 H272921 H272984 H273048 H271985 H272051 H272115 H272177 H272239 H272304 H272366 H272428 H272494 H272859 H272922 H272985 H273049 H271986 H272052 H272116 H272178 H272240 H272305 H272367 H272429 H272495 H272860 H272923 H272986 H273050 H271987 H272053 H272117 H272179 H272241 H272306 H272368 H272430 H272496 H272861 H272924 H272987 H273051 H271988 H272054 H272118 H272180 H272242 H272307 H272369 H272431 H272497 H272862 H272925 H272988 H273052 H271989 H272055 H272119 H272181 H272243 H272308 H272370 H272432 H272500 H272863 H272926 H272989 H273053 H271990 H272056 H272120 H272182 H272244 H272309 H272371 H272433 H272800 H272864 H272927 H272990 H273054 H271991 H272057 H272121 H272183 H272245 H272310 H272372 H272434 H272801 H272865 H272928 H272991 H273055 H271992 H272058 H272122 H272184 H272246 H272311 H272373 H272435 H272802 H272866 H272929 H272992 H273056 H271993 H272059 H272123 H272185 H272247 H272312 H272374 H272436 H272803 H272867 H272930 H272993 H273057 H271994 H272060 H272124 H272186 H272248 H272313 H272375 H272437 H272804 H272868 H272931 H272994 H273058 H271995 H272061 H272125 H272187 H272249 H272314 H272376 H272438 H272805 H272869 H272932 H272995 H273059 H271996 H272062 H272126 H272188 H272250 H272315 H272377 H272439 H272806 H272870 H272933 H272996 H273060 H273061 H273126 H273191 H290154 H290233 H290304 H290367 H290507 H291008 H291072 H291136 H291198 H291260 H273062 H273127 H273192 H290155 H290234 H290305 H290368 H290508 H291009 H291073 H291137 H291199 H291261 H273063 H273128 H273193 H290156 H290235 H290306 H290369 H290509 H291010 H291074 H291138 H291200 H291262 H273064 H273130 H273194 H290157 H290236 H290307 H290370 H290510 H291011 H291075 H291139 H291201 H291263 H273065 H273131 H273195 H290158 H290237 H290308 H290371 H290511 H291012 H291076 H291140 H291202 H291264 H273066 H273132 H273196 H290159 H290239 H290309 H290372 H290512 H291013 H291077 H291141 H291203 H291265 H273067 H273133 H273197 H290160 H290240 H290310 H290373 H290513 H291014 H291078 H291142 H291204 H291266 H273068 H273134 H273198 H290161 H290241 H290311 H290374 H290514 H291015 H291079 H291143 H291205 H291267 H273069 H273135 H273199 H290162 H290242 H290312 H290375 H290515 H291016 H291080 H291144 H291206 H291268 H273070 H273136 H273200 H290164 H290243 H290313 H290376 H290516 H291017 H291081 H291145 H291207 H291269 H273071 H273137 H273201 H290165 H290244 H290314 H290377 H290517 H291018 H291082 H291146 H291208 H291270 H273072 H273138 H273202 H290166 H290246 H290315 H290378 H290518 H291019 H291083 H291147 H291209 H291271 H273074 H273139 H273203 H290168 H290247 H290316 H290379 H290519 H291020 H291084 H291148 H291210 H291272 H273075 H273140 H273204 H290171 H290248 H290317 H290380 H290520 H291021 H291085 H291149 H291211 H291273 H273077 H273141 H273205 H290173 H290249 H290318 H290381 H290521 H291022 H291086 H291150 H291212 H292832 H273078 H273142 H273206 H290175 H290250 H290319 H290382 H290522 H291023 H291087 H291151 H291213 H293009 H273079 H273143 H273208 H290176 H290253 H290320 H290383 H290523 H291024 H291088 H291152 H291214 H293207 H273080 H273144 H273209 H290180 H290254 H290321 H290384 H290524 H291025 H291089 H291153 H291215 H293300 H273081 H273145 H273210 H290181 H290255 H290322 H290385 H290525 H291026 H291090 H291154 H291216 H293301 H273082 H273146 H273547 H290182 H290256 H290323 H290386 H290526 H291027 H291091 H291155 H291217 H293302 H273083 H273147 H273711 H290183 H290257 H290324 H290387 H290527 H291028 H291092 H291156 H291218 H293303 H273084 H273148 H273929 H290185 H290258 H290325 H290388 H290528 H291029 H291093 H291157 H291219 H293304 H273085 H273149 H290001 H290186 H290259 H290326 H290389 H290529 H291030 H291094 H291158 H291220 H293305 H273086 H273150 H290003 H290187 H290260 H290327 H290390 H290530 H291031 H291095 H291159 H291221 H293306 H273087 H273151 H290021 H290189 H290261 H290328 H290392 H290531 H291032 H291096 H291160 H291222 H293307 H273088 H273152 H290026 H290191 H290262 H290329 H290393 H290532 H291034 H291097 H291161 H291223 H293308 H273089 H273153 H290027 H290192 H290263 H290330 H290394 H290533 H291035 H291098 H291162 H291224 H293309 H273090 H273155 H290030 H290193 H290264 H290331 H290395 H290534 H291036 H291099 H291163 H291225 H293310 H273091 H273156 H290031 H290194 H290265 H290332 H290396 H290535 H291037 H291101 H291164 H291226 H293311 H273092 H273157 H290032 H290195 H290266 H290333 H290397 H290536 H291038 H291102 H291165 H291227 H293312 H273093 H273158 H290033 H290196 H290267 H290334 H290398 H290537 H291039 H291103 H291166 H291228 H293313 H273094 H273159 H290034 H290197 H290268 H290335 H290399 H290538 H291040 H291104 H291167 H291229 H293314 H273095 H273160 H290103 H290198 H290269 H290336 H290400 H290539 H291041 H291105 H291168 H291230 H293315 H273097 H273162 H290104 H290201 H290271 H290337 H290401 H290540 H291042 H291106 H291169 H291231 H293316 H273098 H273163 H290105 H290202 H290272 H290338 H290402 H290541 H291043 H291107 H291170 H291232 H293317 H273099 H273164 H290108 H290203 H290273 H290339 H290403 H290542 H291044 H291108 H291171 H291233 H293318 H273100 H273165 H290109 H290205 H290274 H290340 H290404 H290543 H291045 H291109 H291172 H291234 H293319 H273101 H273166 H290110 H290206 H290275 H290341 H290405 H290544 H291046 H291111 H291173 H291235 H293320 H273102 H273167 H290111 H290208 H290276 H290342 H290406 H290545 H291047 H291112 H291174 H291236 H293321 H273103 H273168 H290113 H290209 H290277 H290343 H290408 H290546 H291048 H291113 H291175 H291237 H293322 H273104 H273169 H290114 H290210 H290278 H290344 H290409 H290547 H291049 H291114 H291176 H291238 H293323 H273105 H273170 H290116 H290211 H290279 H290345 H290410 H290548 H291050 H291115 H291177 H291239 H293324 H273106 H273171 H290118 H290212 H290280 H290346 H290411 H290549 H291051 H291116 H291178 H291240 H293325 H273107 H273172 H290121 H290213 H290281 H290347 H290412 H290550 H291052 H291117 H291179 H291241 H293326 H273108 H273173 H290122 H290214 H290282 H290348 H290413 H290551 H291053 H291118 H291180 H291242 H293327 H273109 H273174 H290123 H290215 H290283 H290349 H290414 H290552 H291054 H291119 H291181 H291243 H293328 H273110 H273175 H290125 H290216 H290284 H290350 H290415 H290553 H291055 H291120 H291182 H291244 H293329 H273111 H273176 H290128 H290217 H290285 H290351 H290416 H290554 H291056 H291121 H291183 H291245 H293330 H273112 H273177 H290130 H290219 H290286 H290352 H290417 H290555 H291057 H291122 H291184 H291246 H293331 H273113 H273178 H290132 H290220 H290287 H290353 H290422 H290556 H291059 H291123 H291185 H291247 H293332 H273114 H273179 H290135 H290221 H290288 H290354 H290423 H290557 H291060 H291124 H291186 H291248 H293333 H273115 H273180 H290137 H290222 H290289 H290355 H290424 H290558 H291061 H291125 H291187 H291249 H293334 H273116 H273181 H290138 H290223 H290290 H290356 H290425 H290559 H291062 H291126 H291188 H291250 H293335 H273117 H273182 H290139 H290224 H290291 H290357 H290426 H290560 H291063 H291127 H291189 H291251 H293336 H273118 H273183 H290141 H290225 H290292 H290358 H290427 H290561 H291064 H291128 H291190 H291252 H293337 H273119 H273184 H290142 H290226 H290293 H290359 H290500 H290600 H291065 H291129 H291191 H291253 H293338 H273120 H273185 H290143 H290227 H290295 H290360 H290501 H291001 H291066 H291130 H291192 H291254 H293339 H273121 H273186 H290144 H290228 H290296 H290362 H290502 H291003 H291067 H291131 H291193 H291255 H293340 H273122 H273187 H290145 H290229 H290297 H290363 H290503 H291004 H291068 H291132 H291194 H291256 H293341 H273123 H273188 H290146 H290230 H290299 H290364 H290504 H291005 H291069 H291133 H291195 H291257 H293342 H273124 H273189 H290151 H290231 H290301 H290365 H290505 H291006 H291070 H291134 H291196 H291258 H293343 H273125 H273190 H290152 H290232 H290302 H290366 H290506 H291007 H291071 H291135 H291197 H291259 H293344 H293345 H293409 H293471 H293534 H293598 H293661 H293725 H293788 H293850 H293913 H293977 H294040 H294104 H293346 H293410 H293472 H293535 H293599 H293662 H293726 H293789 H293851 H293914 H293978 H294041 H294105 H293347 H293411 H293473 H293536 H293600 H293663 H293727 H293790 H293852 H293915 H293979 H294042 H294106 H293348 H293412 H293474 H293537 H293601 H293664 H293728 H293791 H293854 H293916 H293980 H294043 H294107 H293349 H293413 H293475 H293538 H293602 H293665 H293729 H293792 H293855 H293917 H293981 H294044 H294108 H293350 H293414 H293476 H293539 H293603 H293666 H293730 H293793 H293856 H293918 H293982 H294045 H294109 H293351 H293415 H293477 H293540 H293604 H293667 H293731 H293794 H293857 H293919 H293983 H294046 H294110 H293352 H293416 H293478 H293541 H293605 H293668 H293732 H293795 H293858 H293920 H293984 H294047 H294111 H293353 H293417 H293479 H293542 H293606 H293669 H293733 H293796 H293859 H293921 H293985 H294048 H294112 H293354 H293418 H293480 H293543 H293607 H293670 H293734 H293797 H293860 H293922 H293986 H294049 H294113 H293355 H293419 H293481 H293544 H293608 H293671 H293735 H293798 H293861 H293923 H293987 H294050 H294114 H293356 H293420 H293482 H293545 H293609 H293672 H293736 H293799 H293862 H293924 H293988 H294051 H294115 H293357 H293421 H293483 H293546 H293610 H293673 H293737 H293800 H293863 H293925 H293989 H294052 H294116 H293358 H293422 H293484 H293548 H293611 H293674 H293738 H293801 H293864 H293926 H293990 H294053 H294117 H293359 H293423 H293485 H293549 H293612 H293675 H293739 H293802 H293865 H293927 H293991 H294054 H294118 H293360 H293424 H293486 H293550 H293613 H293676 H293740 H293803 H293866 H293928 H293992 H294055 H294119 H293361 H293425 H293487 H293551 H293614 H293677 H293741 H293804 H293867 H293929 H293993 H294056 H294120 H293362 H293426 H293488 H293552 H293615 H293678 H293742 H293805 H293868 H293930 H293994 H294057 H294121 H293363 H293427 H293489 H293553 H293616 H293679 H293743 H293806 H293869 H293931 H293995 H294058 H294122 H293364 H293428 H293490 H293554 H293617 H293681 H293744 H293807 H293870 H293932 H293996 H294059 H294123 H293365 H293429 H293491 H293555 H293618 H293682 H293745 H293808 H293871 H293933 H293997 H294060 H294124 H293366 H293430 H293492 H293556 H293619 H293683 H293746 H293809 H293872 H293934 H293998 H294061 H294125 H293367 H293431 H293493 H293557 H293620 H293684 H293747 H293810 H293873 H293935 H293999 H294062 H294126 H293368 H293432 H293494 H293558 H293621 H293685 H293748 H293811 H293874 H293936 H294000 H294063 H294127 H293369 H293433 H293495 H293559 H293622 H293686 H293749 H293812 H293875 H293937 H294001 H294064 H294128 H293370 H293434 H293496 H293560 H293623 H293687 H293750 H293813 H293876 H293938 H294002 H294065 H294129 H293371 H293435 H293497 H293561 H293624 H293688 H293751 H293814 H293877 H293940 H294003 H294066 H294130 H293372 H293436 H293498 H293562 H293625 H293689 H293752 H293815 H293878 H293941 H294004 H294067 H294131 H293373 H293437 H293499 H293563 H293626 H293690 H293753 H293816 H293879 H293942 H294005 H294068 H294132 H293374 H293438 H293500 H293564 H293627 H293691 H293754 H293817 H293880 H293943 H294006 H294069 H294133 H293375 H293439 H293501 H293565 H293628 H293692 H293755 H293818 H293881 H293944 H294007 H294070 H294134 H293376 H293440 H293502 H293566 H293629 H293693 H293756 H293819 H293882 H293945 H294008 H294071 H294135 H293377 H293441 H293503 H293567 H293630 H293694 H293758 H293820 H293883 H293946 H294009 H294072 H294136 H293378 H293442 H293504 H293568 H293631 H293695 H293759 H293821 H293884 H293947 H294010 H294073 H294137 H293379 H293443 H293505 H293569 H293632 H293696 H293760 H293822 H293885 H293948 H294011 H294074 H294139 H293380 H293444 H293506 H293570 H293633 H293697 H293761 H293823 H293886 H293949 H294012 H294075 H294140 H293381 H293445 H293507 H293571 H293634 H293698 H293762 H293824 H293887 H293950 H294013 H294076 H294141 H293382 H293446 H293508 H293572 H293635 H293699 H293763 H293825 H293888 H293951 H294014 H294077 H294142 H293383 H293447 H293509 H293573 H293636 H293700 H293764 H293826 H293889 H293952 H294015 H294078 H294143 H293384 H293448 H293510 H293574 H293637 H293701 H293765 H293827 H293890 H293953 H294016 H294079 H294144 H293385 H293449 H293511 H293575 H293638 H293702 H293766 H293828 H293891 H293954 H294017 H294080 H294145 H293386 H293450 H293512 H293576 H293639 H293703 H293767 H293829 H293892 H293955 H294018 H294081 H294146 H293387 H293451 H293513 H293577 H293640 H293704 H293768 H293830 H293893 H293956 H294020 H294083 H294147 H293388 H293452 H293514 H293578 H293641 H293705 H293769 H293831 H293894 H293957 H294021 H294084 H294148 H293389 H293453 H293515 H293579 H293642 H293706 H293770 H293832 H293895 H293958 H294022 H294085 H294149 H293390 H293454 H293516 H293581 H293643 H293707 H293771 H293833 H293896 H293959 H294023 H294086 H294150 H293391 H293455 H293517 H293582 H293644 H293708 H293772 H293834 H293897 H293960 H294024 H294087 H294151 H293393 H293456 H293518 H293583 H293645 H293709 H293773 H293835 H293898 H293961 H294025 H294088 H294152 H293394 H293457 H293519 H293584 H293646 H293710 H293774 H293836 H293899 H293962 H294026 H294089 H294153 H293395 H293458 H293520 H293585 H293648 H293712 H293775 H293837 H293900 H293963 H294027 H294090 H294154 H293396 H293459 H293521 H293586 H293649 H293713 H293776 H293838 H293901 H293964 H294028 H294091 H294155 H293397 H293460 H293522 H293587 H293650 H293714 H293777 H293839 H293902 H293965 H294029 H294092 H294156 H293398 H293461 H293523 H293588 H293651 H293715 H293778 H293840 H293903 H293966 H294030 H294093 H294157 H293399 H293462 H293524 H293589 H293652 H293716 H293779 H293841 H293904 H293967 H294031 H294094 H294158 H293401 H293463 H293525 H293590 H293653 H293717 H293780 H293842 H293905 H293968 H294032 H294095 H294159 H293402 H293464 H293527 H293591 H293654 H293718 H293781 H293843 H293906 H293969 H294033 H294097 H294160 H293403 H293465 H293528 H293592 H293655 H293719 H293782 H293844 H293907 H293970 H294034 H294098 H294161 H293404 H293466 H293529 H293593 H293656 H293720 H293783 H293845 H293908 H293971 H294035 H294099 H294162 H293405 H293467 H293530 H293594 H293657 H293721 H293784 H293846 H293909 H293972 H294036 H294100 H294163 H293406 H293468 H293531 H293595 H293658 H293722 H293785 H293847 H293910 H293974 H294037 H294101 H294164 H293407 H293469 H293532 H293596 H293659 H293723 H293786 H293848 H293911 H293975 H294038 H294102 H294165 H293408 H293470 H293533 H293597 H293660 H293724 H293787 H293849 H293912 H293976 H294039 H294103 H294166 H294167 H294229 H294292 H294355 H294418 H294483 H294545 H294615 H294677 H294743 H294806 H294871 H294938 H294168 H294230 H294293 H294356 H294419 H294484 H294546 H294616 H294678 H294744 H294807 H294872 H294940 H294169 H294231 H294294 H294357 H294420 H294485 H294547 H294617 H294679 H294745 H294808 H294874 H294941 H294170 H294232 H294295 H294358 H294421 H294486 H294548 H294618 H294680 H294746 H294809 H294875 H294942 H294171 H294233 H294296 H294359 H294422 H294487 H294549 H294619 H294681 H294747 H294810 H294876 H294943 H294172 H294234 H294297 H294360 H294423 H294488 H294550 H294620 H294682 H294748 H294811 H294877 H294945 H294173 H294235 H294298 H294361 H294424 H294489 H294551 H294621 H294683 H294749 H294812 H294878 H294946 H294174 H294236 H294299 H294362 H294425 H294490 H294552 H294622 H294684 H294750 H294813 H294879 H294947 H294175 H294237 H294300 H294363 H294426 H294491 H294553 H294623 H294685 H294751 H294814 H294880 H294949 H294176 H294238 H294301 H294364 H294427 H294492 H294554 H294624 H294686 H294752 H294815 H294881 H294950 H294177 H294239 H294302 H294365 H294428 H294493 H294555 H294625 H294687 H294753 H294816 H294882 H294951 H294178 H294240 H294303 H294366 H294429 H294494 H294556 H294626 H294688 H294754 H294817 H294883 H294952 H294179 H294241 H294304 H294367 H294430 H294495 H294557 H294627 H294689 H294755 H294818 H294884 H294953 H294180 H294242 H294305 H294368 H294431 H294496 H294564 H294628 H294690 H294756 H294820 H294885 H294954 H294181 H294243 H294306 H294369 H294432 H294497 H294565 H294629 H294691 H294757 H294821 H294886 H294956 H294182 H294244 H294307 H294370 H294433 H294498 H294567 H294630 H294692 H294758 H294822 H294887 H294957 H294183 H294245 H294308 H294371 H294434 H294499 H294568 H294631 H294693 H294759 H294823 H294888 H294958 H294184 H294246 H294309 H294372 H294435 H294500 H294569 H294632 H294694 H294760 H294824 H294889 H294959 H294185 H294247 H294310 H294373 H294436 H294501 H294570 H294633 H294695 H294761 H294825 H294890 H294960 H294186 H294248 H294311 H294374 H294437 H294502 H294571 H294634 H294696 H294762 H294826 H294891 H294961 H294187 H294249 H294312 H294376 H294438 H294503 H294572 H294635 H294697 H294763 H294827 H294892 H294962 H294188 H294250 H294313 H294377 H294439 H294504 H294573 H294636 H294698 H294764 H294828 H294893 H294963 H294189 H294251 H294314 H294378 H294440 H294505 H294574 H294637 H294699 H294765 H294829 H294894 H294965 H294190 H294252 H294315 H294379 H294441 H294506 H294575 H294638 H294700 H294766 H294830 H294895 H294966 H294191 H294253 H294316 H294380 H294442 H294507 H294576 H294639 H294701 H294767 H294831 H294896 H294967 H294192 H294254 H294317 H294381 H294443 H294508 H294577 H294640 H294702 H294768 H294832 H294897 H294969 H294193 H294255 H294318 H294382 H294444 H294509 H294578 H294641 H294704 H294769 H294833 H294898 H294970 H294194 H294256 H294319 H294383 H294445 H294510 H294579 H294642 H294705 H294770 H294834 H294899 H294971 H294195 H294257 H294320 H294384 H294446 H294511 H294580 H294643 H294706 H294771 H294835 H294900 H294972 H294196 H294258 H294321 H294385 H294447 H294512 H294581 H294644 H294707 H294772 H294836 H294901 H294974 H294197 H294259 H294322 H294386 H294448 H294513 H294582 H294645 H294708 H294773 H294837 H294902 H294975 H294198 H294260 H294323 H294387 H294449 H294514 H294583 H294646 H294709 H294774 H294838 H294903 H294976 H294199 H294261 H294324 H294388 H294450 H294515 H294584 H294647 H294710 H294775 H294839 H294904 H294977 H294200 H294262 H294325 H294389 H294451 H294516 H294585 H294648 H294711 H294776 H294840 H294905 H294978 H294201 H294263 H294326 H294390 H294452 H294517 H294586 H294649 H294712 H294777 H294841 H294906 H294979 H294202 H294264 H294327 H294391 H294454 H294518 H294587 H294650 H294713 H294778 H294842 H294907 H294980 H294203 H294265 H294328 H294392 H294457 H294519 H294588 H294651 H294714 H294779 H294843 H294908 H294981 H294204 H294266 H294329 H294393 H294458 H294520 H294589 H294652 H294715 H294780 H294844 H294910 H294982 H294205 H294267 H294330 H294394 H294459 H294521 H294590 H294653 H294716 H294781 H294845 H294911 H294984 H294206 H294268 H294331 H294395 H294460 H294522 H294591 H294654 H294717 H294782 H294846 H294914 H294986 H294207 H294269 H294332 H294396 H294461 H294523 H294592 H294655 H294718 H294783 H294847 H294915 H294987 H294208 H294270 H294333 H294397 H294462 H294524 H294593 H294656 H294719 H294784 H294848 H294916 H294988 H294209 H294271 H294334 H294398 H294463 H294525 H294594 H294657 H294720 H294785 H294850 H294917 H294989 H294210 H294272 H294335 H294399 H294464 H294526 H294595 H294658 H294721 H294786 H294851 H294918 H294990 H294211 H294273 H294336 H294400 H294465 H294527 H294596 H294659 H294723 H294787 H294853 H294919 H294991 H294212 H294274 H294337 H294401 H294466 H294528 H294597 H294660 H294724 H294788 H294854 H294920 H294992 H294213 H294275 H294338 H294402 H294467 H294529 H294598 H294661 H294726 H294789 H294855 H294921 H294995 H294214 H294276 H294339 H294403 H294468 H294530 H294599 H294662 H294727 H294790 H294856 H294922 H294996 H294215 H294277 H294340 H294404 H294469 H294531 H294600 H294663 H294728 H294791 H294857 H294923 H294997 H294216 H294278 H294341 H294405 H294470 H294532 H294601 H294664 H294729 H294792 H294858 H294924 H294998 H294217 H294279 H294342 H294406 H294471 H294533 H294602 H294665 H294730 H294793 H294859 H294925 H294999 H294218 H294280 H294343 H294407 H294472 H294534 H294603 H294666 H294731 H294794 H294860 H294926 H295001 H294219 H294281 H294345 H294408 H294473 H294535 H294604 H294667 H294732 H294795 H294861 H294928 H295002 H294220 H294283 H294346 H294409 H294474 H294536 H294605 H294668 H294733 H294796 H294862 H294929 H295003 H294221 H294284 H294347 H294410 H294475 H294537 H294606 H294669 H294734 H294797 H294863 H294930 H295004 H294222 H294285 H294348 H294411 H294476 H294538 H294607 H294670 H294735 H294798 H294864 H294931 H295006 H294223 H294286 H294349 H294412 H294477 H294539 H294608 H294671 H294736 H294799 H294865 H294932 H295008 H294224 H294287 H294350 H294413 H294478 H294540 H294610 H294672 H294737 H294800 H294866 H294933 H295009 H294225 H294288 H294351 H294414 H294479 H294541 H294611 H294673 H294738 H294801 H294867 H294934 H295010 H294226 H294289 H294352 H294415 H294480 H294542 H294612 H294674 H294739 H294802 H294868 H294935 H295011 H294227 H294290 H294353 H294416 H294481 H294543 H294613 H294675 H294740 H294803 H294869 H294936 H295012 H294228 H294291 H294354 H294417 H294482 H294544 H294614 H294676 H294742 H294804 H294870 H294937 H295013 H295014 H295034 H295058 H295080 H295114 H295015 H295035 H295059 H295082 H295116 H295016 H295036 H295060 H295083 H295117 H295017 H295037 H295062 H295084 H295118 H295018 H295038 H295063 H295085 H295119 H295019 H295041 H295064 H295087 H295121 H295021 H295042 H295065 H295090 H295122 H295022 H295044 H295066 H295094 H295126 H295023 H295045 H295067 H295095 H295127 H295024 H295046 H295068 H295096 H295128 H295025 H295047 H295070 H295097 H295131 H295026 H295048 H295071 H295100 H295132 H295027 H295049 H295072 H295101 H295134 H295028 H295050 H295073 H295102 H295139 H295029 H295051 H295074 H295103 H295140 H295030 H295053 H295075 H295105 H295142 H295031 H295054 H295076 H295109 H295144 H295032 H295056 H295077 H295111 H295150 H295033 H295057 H295078 H295112 H295151. ***12-20-12, AMENDED INFO ***: H140971, H140942, H295370, H295371, H295271, H295266 |
Recalling Firm/ Manufacturer |
Varian Medical Systems, Inc. Oncology Systems 911 Hansen Way Palo Alto CA 94304-1028
|
For Additional Information Contact | Michael Pignataro 650-424-6471 |
Manufacturer Reason for Recall | On Varian Linear Accelerators where the customer may calibrate the collimator angle position readout in reverse, switching the 90 degree and 270 degree positions. |
FDA Determined Cause 2 | Device Design |
Action | Varian Medical Systems sent an "Urgent Medical Device Correction-Urgent Field Safety Notice" letter dated May 26, 2011 to all affected customers.
***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.***
The letters identifies the product, describes the problem and provides recommendations. Varian recommends maintenance only be performed by qualified personnel with appropriate training, always verify collimator calibration and anytime a collimator posthion readout calibration is performed, and always verify that the collimator angle is not calibrated in reverse by performing the appropriate checks.
For questions contact Varian at (888) 827-4265. |
Quantity in Commerce | 5685 |
Distribution | Worldwide Distribution: Nationwide (USA) Distribution and the countries of; Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Belorussia, Bolivia, Bosnia and Herzegoniva, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Rep., Denmark, D.R. Ecuador, Egypt, Estonia, Finland, France, French Poly., Georgia, Germany, Greece, Guam, Guatemala, H.K., Honduras, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Macau, Macedonia, Malaysia, Martinique, Mexico, Moldova, Monte Negro, Morocco, Nepal, Netherlands, N.Z., Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, P.R., Qatar, Reunion, Romania, Russia, S.A.,Serbia, Singapore, Slovakia, South Africa, S.K., Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad/Tobago, Tunisia, Turkey, Ukrain, U.A.E., U.K., Uruguay, Venezuela, Vietnam, V.I., and Zimbabwe. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = IYE 510(K)s with Product Code = IYE
|
|
|
|