| | Class 2 Device Recall Wound Management Tray |  |
| Date Initiated by Firm | December 08, 2011 |
| Date Posted | March 21, 2012 |
| Recall Status1 |
Terminated 3 on April 12, 2017 |
| Recall Number | Z-1264-2012 |
| Recall Event ID |
60753 |
| Product Classification |
Kit, wound dressing - Product Code MCY
|
| Product | Wound Management Tray Asoc. Agencia Salud, catalog# 900-1154 |
| Code Information |
lot# 111010025, 111020497, 111030548, 111041068, 111061631, 111102825 |
| FEI Number |
2648727
|
Recalling Firm/ Manufacturer |
Customed, Inc Calle Igualdad Final #7 Fajardo PR 00738
|
Manufacturer Reason for Recall | The firm has determined that there is a possibility that packaging seals may be compromised. The issue could result in an injury to the patient due to product contamination or loss of sterility condition. |
FDA Determined Cause 2 | Packaging |
| Action | Customed Inc sent an Urgent Medical Device Recall letter dated December 8, 2011 to distributors via e-mail; then by certified (return and receipt) mail. The letter identified the product, description of problem, and actions to be taken. The letter included the following instructions:
1. Product should be removed from sale (Quarantine them) and cease distribution. Products affected shall NOT be used.
2. Return the enclosed acknowledgement form to Customed, Inc. Recall Coordinator whether or not the consignee has affected products.
3. If products has been futher distributed, sub-recall products - It is extremely important to notify your customers of this action by including a copy of this leller along with your formal notification. Please, instruct your customer to follow your directions for credit/replacemenl if they purchased the product from you.
For qucstions. contact Cuslomed, Inc. Recall Coordinator at 787-622-5151 Ext. 7510. |
| Quantity in Commerce | 7,656 units |
| Distribution | USA Nationwide Distribution - including Florida and Puerto Rico. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|