| Date Initiated by Firm | October 12, 2010 |
| Date Posted | June 21, 2012 |
| Recall Status1 |
Terminated 3 on June 21, 2012 |
| Recall Number | Z-1849-2012 |
| Recall Event ID |
61741 |
| 510(K)Number | K092564 |
| Product Classification |
System, x-ray, mobile - Product Code IZL
|
| Product | Medtronic 0-arm Imaging System with Software Version 3.1.1, Catalog Numbers: BI-700-00027-120R and BI-700-00027-120. The 0-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging. |
| Code Information |
Serial Numbers: 125, 156, 234R. |
Recalling Firm/ Manufacturer |
Medtronic Navigation, Inc. 300 Foster St Littleton MA 01460-2017
|
| For Additional Information Contact | 978-698-6008 |
Manufacturer Reason for Recall | Software Version 3.1.1 does not meet navigational accuracy. |
FDA Determined Cause 2 | Software design |
| Action | Medtronic Navigation contacted customers by telephone on 10/12/2010 and with "Urgent Field Safety Notice" follow up letters dated 10/14 or 10/15/2010. Sites were visited by Medtronic Service Representative to complete navigational accuracy testing and either replace or correct the unit. |
| Quantity in Commerce | 3 units |
| Distribution | Nationwide Distribution including the states of CO, IN, and MN. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = IZL
|