Date Initiated by Firm | September 08, 2011 |
Date Posted | July 09, 2012 |
Recall Status1 |
Terminated 3 on January 28, 2013 |
Recall Number | Z-1977-2012 |
Recall Event ID |
62319 |
510(K)Number | K001387 |
Product Classification |
Glucose oxidase, glucose - Product Code CHL
|
Product | i-STAT1 Abaxis Analyzer; Abbott Point of Care, Princeton, NJ 08540
Model 200. The 37 degree tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2, and the i-Stat 200 analyzer as port of the i-Stat System, are intended for use in the in vitro quantitization of arterial, venous, or capillary whole blood 510 K940918 Model 300. The i-Stat Model 300 Portable Clinical Analyzer is intended to be used by trained medical professional for use with i-Stat test cartridges and MediSense blood glucose test strips. i-State cartridges comprise a variety of clinical chemistry tests and test panels. 510K 001387 |
Code Information |
Military Kit iStat1 04J48-201 215001 All lots |
Recalling Firm/ Manufacturer |
Abbott Point Of Care Inc. 400 College Rd E Princeton NJ 08540-6607
|
For Additional Information Contact | Mr. Larry W. Krasley 609-454-9272 |
Manufacturer Reason for Recall | Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.
|
FDA Determined Cause 2 | Device Design |
Action | Abbott Point Care sent an Urgent Product Recall letter dated September 2011 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to confirm the number of affected product at their facility that were using the Old Battery Carrier. Fill out the affected Buesiness Reply Card and submit their facility information as per the instructions on the card. Upon the receipt of the New Battery Carriers, replace the Old Battery Center with the New Battery Center. Customers were instructed to immediately dispose of all Old Battery Carriers per local regulations. If customers forward the affected product to another facility, they should provide a copy of the recall letter to them and advise them to replace the carriers if required.
For any questions call 800-366-8020, Option 1 or your Abbott Point of Care representative.
For questions regarding this recall call 609-454-9272. |
Quantity in Commerce | 04J48-201-1069 units; 215001-217 units |
Distribution | Worldwide Distribution - USA (nationwide) |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = CHL
|