| | Class 2 Device Recall Terumo Advanced Perfusion System 1 |  |
| Date Initiated by Firm | June 25, 2012 |
| Date Posted | July 09, 2012 |
| Recall Status1 |
Terminated 3 on March 01, 2013 |
| Recall Number | Z-1970-2012 |
| Recall Event ID |
62373 |
| 510(K)Number | K022947 |
| Product Classification |
Console, heart-lung machine, cardiopulmonary bypass - Product Code DTQ
|
| Product | System 1 Base 100/120V
The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. |
| Code Information |
catalog number: 801763 and serial numbers: 0011-0335, 0337-0610, 1001-1034, 1037-1049, 1100, 1103-1106,1109-1445, and 1447-1450. |
Recalling Firm/ Manufacturer |
Terumo Cardiovascular Systems Corporation 6200 Jackson Road Ann Arbor MI 48103-9586
|
| For Additional Information Contact | 734-741-6173 |
Manufacturer Reason for Recall | Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump malfunctions of the Terumo Advanced Perfusion System 1 that could result in the inability to use the pump and would then require replacement with back-up equipment. Description of malfunctions:
" Unplanned pump stop or pause
" Changes in pump speed
If it is necessary to replace a roller pump during cardiopu |
FDA Determined Cause 2 | Device Design |
| Action | Terumo CVS sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 4, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Customers were asked to complete and return a Customer Response Form to the firm via fax at 1-800-292-6551. Contact the firm at 1-800-521-2818 for questions regarding this recall. |
| Quantity in Commerce | 992 units |
| Distribution | Worldwide Distribution-USA (nationwide) and the countries of ARGENTINA, AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatamala, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = DTQ
|
|
|
|