| | Class 2 Device Recall LifeCare PCA infusion System with Hospira MedNet |  |
| Date Initiated by Firm | March 08, 2013 |
| Create Date | September 22, 2016 |
| Recall Status1 |
Terminated 3 on November 07, 2016 |
| Recall Number | Z-2873-2016 |
| Recall Event ID |
64230 |
| 510(K)Number | K042800 |
| Product Classification |
Pump, infusion, pca - Product Code MEA
|
| Product | LifeCare PCA Infusion System with Hospira MedNet Software, PCA Serial List Number 20709
Allows for clinician administration or self-administration of analgesic medications. |
| Code Information |
Serial Number 15553367, 16224548, 16224939, 16225091, 16225646, 16225851, 16226006, 16226095, 16226117, 16226159, 16226360, 16226383, 16226435, 16226460, 16226680, 16226707, 16226726, 16226730, 16226756, 16226763, 16226891, 16226895, 16226898, 16226910, 16226917, 16226918, 16226919, 16226920, 16226921, 16226922, 16226923, 16226924, 16226925, 16226926, 16226927, 16226928, 16226929, 16226930, 16226931, 16226932, 16226933, 16226934, 16226935, 16226936, 16226937, 16226938, 16226939, 16226940, 16226941, 16226943, 16226944, 16226945, 16226946, 16226947, 16226948, 16226949, 16226950, 16226951, 16226952, 16226953, 16226954, 16226955, 16226956, 16226957, 16226958, 16226959, 16226960, 16226961, 16226962, 16226963, 16226964, 16226965, 16226966, 16226967, 16226968, 16226969, 16226970, 16226971, 16226972, 16226973, 16226974, 16226975, 16226976, 16226977, 16226978, 16226979, 16226980, 16226981, 16226982, 16226983, 16226984, 16226985, 16226986, 16226987, 16226988, 16226989, 16226990, 16226991, 16226992, 16226993, 16226994, 16226995, 16226996, 16226997, 16226998, 16226999, 16227000, 16227001, 16227002, 16227003, 16227004, 16227005, 16227006, 16227007, 16227008, 16227009, 16227010, 16227011, 16227012, 16227013, 16227014, 16227015, 16227016, 16227017, 16227018, 16227019, 16227020, 16227021, 16227022, 16227023, 16227024, 16227026, 16227027, 16227028, 16227029, 16227030, 16227031, 16227032, 16227033, 16227035, 16227036, 16227038, 16227039, 16227040, 16227041, 16227042, 16227043, 16227044, 16227045, 16227046, 16227047, 16227048, 16227049, 16227050, 16227051, 16227052, 16227053, 16227054, 16227055, 16227056, 16227057, 16227058, 16227059, 16227061, 16227062, 16227063, 16227064, 16227065, 16227066, 16227067, 16227068, 16227069, 16227070, 16227071, 16227072, 16227073, 16227074, 16227075, 16227076, 16227077, 16227078, 16227079, 16227080, 16227082, 16227083, 16227084, 16227085, 16227086, 16227087, 16227088, 16227089, 16227090, 16227091, 16227092, 16227093, 16227094, 16227095, 16227096, 16227097, 16227098, 16227099, 16227100, 16227101, 16227102, 16227103, 16227104, 16227105, 16227106, 16227107, 16227108, 16227109, 16227110, 16227111, 16227112, 16227113, 16227114, 16227115, 16227116, 16227117, 16227119, 16227120, 16227122, 16227123, 16227124, 16227125, 16227126, 16227127, 16227128, 16227129, 16227130, 16227131, 16227132, 16227133, 16227134, 16227135, 16227136, 16227138, 16227139, 16227140, 16227141, 16227142, 16227143, 16227144, 16227145, 16227146, 16227147, 16227148, 16227149, 16227150, 16227151, 16227152, 16227153, 16227154, 16227155, 16227156, 16227157, 16227158, 16227159, 16227160, 16227161, 16227162, 16227163, 16227164, 16227165, 16227166, 16227167, 16227168, 16227169, 16227170, 16227171, 16227172, 16227173, 16227174, 16227175, 16227176, 16227177, 16227178, 16227179, 16227180, 16227181, 16227182, 16227183, 16227184, 16227185, 16227186, 16227187, 16227188, 16227189, 16227190, 16227191, 16227192, 16227193, 16227194, 16227195, 16227196, 16227197, 16227198, 16227199, 16227200, 16227201, 16227202, 16227203, 16227204, 16227205, 16227206, 16227207, 16227208, 16227209, 16227210, 16227211, 16227212, 16227213, 16227214, 16227215, 16227216, 16227217, 16227218, 16227219, 16227220, 16227221, 16227222, 16227223, 16227224, 16227225, 16227226, 16227227, 16227228, 16227229, 16227230, 16227231, 16227232, 16227233, 16227234, 16227235, 16227236, 16227237, 16227238, 16227239, 16227240, 16227241, 16227242, 16227243, 16227244, 16227245, 16227246, 16227247, 16227248, 16227249, 16227250, 16227251, 16227252, 16227253, 16227254, 16227255, 16227256, 16227257, 16227258, 16227259, 16227260, 16227261, 16227262, 16227263, 16227264, 16227265, 16227266, 16227267, 16227268, 16227269, 16227270, 16227271, 18577401, 18577402, 18577405, 18577409, 18577410, 18577411, 18577412, 18577414, 18577415, 18577419, 18577420, 18577421, 18577422, 18577425, 18577428, 18577430, 18577432, 18577433, 18577435, 18577437, 18577438, 18577439, 18577440, 18577443, 18577445, 18577450, 18577451, 18577453, 18577454, 18577457, 18577458, 18577462, 18577477, 18577479, 18577481, 18577482, 18577483, 18577484, 18577485, 18577487, 18577488, 18577489, 18577490, 18577491, 18577492, 18577493, 18577494, 18577495, 18577496, 18577497, 18577498, 18577499, 18577500, 18577501, 18577502, 18577503, 18577504, 18577505, 19350406, 19350407, 19350408, 19350409, 19350410, 19350411, 19350412, 19350413, 19350414, 19350415, 19350416, 19350417, 19350418, 19350419, 19350420, 19350421, 19350422, 19350423, 19350424, 19350425, 19350426, 19350427, 19350428, 19350429, 19350430, 19350431, 19350432, 19350433, 19350435, 19350436, 19350437, 19350438, 19350439, 19350440, 19350441, 19350442, 19350443, 19350444, 19350445, 19350446, 19350447, 19350448, 19350449, 19350450, 19350451, 19350452, 19350453, 19350454, 19350455, 19350456, 19350457, 19350459, 19350460, 19350461, 19350463, 19350464, 19350465, 19350466, 19350467, 19350468, 19350469, 19350470, 19350471, 19350472, 19350473, 19350474, 19350475, 19350476, 19350477, 19350478, 19350479, 19350480, 19350481, 19350482, 19350483, 19350484, 19350485, 19350486, 19350487, 19350488, 19350489, 19350490, 19350491, 19350492, 19350493, 19350494, 19350495, 19350496, 19350497, 19350498, 19350499, 19350500, 19350501, 19350502, 19350503, 19350504, 19350505, 19350506, 19350507, 19350508, 19350509, 19350510, 19350511, 19350512, 19350513, 19350514, 19350515, 19350516, 19350517, 19350518, 19350519, 19350520, 19350521, 19350522, 19350523, 19350524, 19350525, 19350526, 19350527, 19350528, 19350529, 19350530, 19350531, 19350532, 19350533, 19350534, 19350535, 19350536, 19350537, 19350538, 19350539, 19350540, 19350541, 19350542, 19350543, 19350544, 19350545, 19350546, 19350547, 19350548, 19350549, 19350550, 19350551, 19350552, 19350553, 19350554, 19350555, 19350556, 19350557, 19350559, 19350560, 19350561, 19350562, 19350563, 19350564, 19350565, 19350566, 19350567, 19350568, 19350569, 19350570, 19350571, 19350572, 19350573, 19350574, 19350575, 19350576, 19350577, 19350578, 19350579, 19350580, 19350581, 19350582, 19350583, 19350585, 19350587, 19350588, 19350589, 19350590, 19350591, 19350592, 19350593, 19350594, 19350595, 19350596, 19350597, 19350598, 19350599, 19350600, 19350601, 19350602, 19350603, 19350604, 19350605, 19350606, 19350607, 19350608, 19350609, 19350610, 19350611, 19350612, 19350614, 19350615, 19350616, 19350617, 19350618, 19350619, 19350620, 19350621, 19350622, 19350623, 19350624, 19350625, 19350626, 19350627, 19350628, 19350629, 19350630, 19350631, 19350632, 19350633, 19350634, 19350635, 19350636, 19350637, 19350638, 19350640, 19350641, 19350642, 19350643, 19350644, 19350645, 19350646, 19350647, 19350648, 19350649, 19350650, 19350651, 19350652, 19350653, 19350654, 19350655, 19350656, 19350657, 19350658, 19350659, 19350660, 19350661, 19350662, 19350663, 19350664, 19350665, 19350666, 19350667, 19350668, 19350669, 19350670, 19350671, 19350672, 19350673, 19350674, 19350675, 19350676, 19350677, 19350678, 19350679, 19350680, 19350681, 19350682, 19350683, 19350684, 19350685, 19350687, 19350688, 19350689, 19350690, 19350691, 19350692, 19350693, 19350694, 19350695, 19350696, 19350697, 19350698, 19350699, 19350700, 19350701, 19350702, 19350703, 19350704, 19350705, 19350706, 19350707, 19350708, 19350709, 19350710, 19350711, 19350712, 19350713, 19350714, 19350715, 19350716, 19350717, 19350718, 19350719, 19350720, 19350721, 19350722, 19350723, 19350724, 19350725, 19350726, 19350727, 19350728, 19350729, 19350730, 19350731, 19350732, 19350733, 19350734, 19350735, 19350736, 19350737, 19350738, 19350739, 19350740, 19350741, 19350742, 19350743, 19350744, 19350745, 19350746, 19350747, 19350748, 19350749, 19350750, 19350751, 19350752, 19350753, 19350754, 19350755, 19350756, 19350757, 19350758, 19350759, 19350760, 19350761, 19350762, 19350763, 19350764, 19350765, 19350766, 19350767, 19350768, 19350769, 19350770, 19350771, 19350772, 19350773, 19350774, 19350775, 19350776, 19350777, 19350778, 19350779, 19350780, 19350781, 19350782, 19350783, 19350784, 19350785, 19350786, 19350787, 19350788, 19350789, 19350790, 19350791, 19350792, 19350793, 19350794, 19350795, 19350796, 19350797, 19350798, 19350799, 19350800, 19350801, 19350802, 19350803, 19350804, 19350805, 19350806, 19350807, 19350808, 19350809, 19350810, 19350811, 19350812, 19350813, 19350814, 19350815, 19350816, 19350817, 19350818, 19350819, 19350820, 19350821, 19350822, 19350823, 19350824, 19350825, 19350826, 19350827, 19350828, 19350829, 19350830, 19350831, 19350832, 19350833, 19350834, 19350835, 19350836, 19350837, 19350838, 19350839, 19350840, 19350841, 19350842, 19350843, 19350844, 19350845, 19350846, 19350847, 19350848, 19350849, 19350850, 19350851, 19350852, 19350853, 19350854, 19350855, 19350856, 19350857, 19350858, 19350859, 19350860, 19350861, 19350862, 19350863, 19350864, 19350865, 19350866, 19350867, 19350868, 19350869, 19350870, 19350871, 19350872, 19350873, 19350874, 19350875, 19350876, 19350877, 19350878, 19350879, 19350880, 19350881, 19350882, 19350883, 19350884, 19350885, 19350886, 19350887, 19350888, 19350889, 19350890, 19350891, 19350892, 19350893, 19350894, 19350895, 19350896, 19350897, 19350898, 19350899, 19350900, 19350901, 19350902, 19350903, 19350904, 19350905, 19350906, 19350907, 19350908, 19350909, 19350910, 19350911, 19350912, 19350913, 19350914, 19350915, 19350917, 19350918, 19350919, 19350920, 19350921, 19350922, 19350923, 19350925, 19350926, 19350927, 19350928, 19350929, 19350930, 19350931, 19350932, 19350933, 19350934, 19350935, 19350936, 19350937, 19350938, 19350939, 19350940, 19350941, 19350942, 19350943, 19350944, 19350945, 19350946, 19350947, 19350948, 19350949, 19350950, 19350951, 19350952, 19350953, 19350954, 19350955, 19350956, 19350957, 19350958, 19350959, 19350960, 19350961, 19350962, 19350963, 19350965, 19350966, 19350967, 19350968, 19350969, 19350970, 19350971, 19350972, 19350973, 19350974, 19350975, 19350976, 19350977, 19350978, 19350979, 19350980, 19350981, 19350982, 19350983, 19350984, 19350985, 19350986, 19350987, 19350988, 19350989, 19350990, 19350991, 19350992, 19350993, 19350994, 19350995, 19350996, 19350997, 19350998, 19350999, 19351000, 19351001, 19351002, 19351003, 19351004, 19351005, 19351006, 19351007, 19351008, 19351009, 19351010, 19351011, 19351012, 19351013, 19351014, 19351015, 19351016, 19351017, 19351018, 19351019, 19351020, 19351021, 19351022, 19351023, 19351024, 19351025, 19351026, 19351027, 19351028, 19351029, 19351030, 19351031, 19351032, 19351033, 19351034, 19351035, 19351036, 19351037, 19351038, 19351039, 19351040, 19351041, 19351042, 19351043, 19351044, 19351045, 19351046, 19351047, 19351048, 19351049, 19351050, 19351051, 19351052, 19351053, 19351054, 19351055, 19351056, 19351057, 19351058, 19351059, 19351060, 19351061, 19351062, 19351063, 19351064, 19351065, 19351066, 19351067, 19351068, 19351069, 19351070, 19351071, 19351072, 19351073, 19351074, 19351075, 19351076, 19351077, 19351078, 19351079, 19351080, 19351081, 19351082, 19351083, 19351084, 19351085, 19351086, 19351087, 19351088, 19351089, 19351090, 19351091, 19351092, 19351093, 19351094, 19351098, 19351099, 19351100, 19351101, 19351102, 19351103, 19351104, 19351105, 19351106, 19351107, 19351108, 19351109, 19351110, 19351111, 19351112, 19351113, 19351114, 19351115, 19351116, 19351117, 19351118, 19351119, 19351120, 19351121, 19351122, 19351123, 19351124, 19351125, 19351126, 19351127, 19351128, 19351129, 19351130, 19351131, 19351132, 19351133, 19351134, 19351135, 19351136, 19351137, 19351138, 19351139, 19351140, 19351141, 19351142, 19351143, 19351144, 19351145, 19351146, 19351147, 19351148, 19351149, 19351150, 19351151, 19351152, 19351153, 19351154, 19351155, 19351156, 19351157, 19351158, 19351159, 19351160, 19351161, 19351162, 19351163, 19351164, 19351165, 19351166, 19351167, 19351168, 19351169, 19351170, 19351171, 19351172, 19351173, 19351174, 19351175, 19351176, 19351177, 19351178, 19351179, 19351180, 19351181, 19351182, 19351183, 19351184, 19351185, 19351186, 19351187, 19351188, 19351189, 19351190, 19351191, 19351192, 19351193, 19351194, 19351195, 19351196, 19351197, 19351198, 19351199, 19351200, 19351201, 19351202, 19351203, 19351204, 19351205, 19351206, 19351207, 19351208, 19351209, 19351210, 19351211, 19351212, 19351213, 19351214, 19351215, 19351216, 19351217, 19351218, 19351219, 19351220, 19351221, 19351222, 19351223, 19351224, 19351225, 19351226, 19351227, 19351228, 19351229, 19351230, 19351231, 19351232, 19351233, 19351234, 19351235, 19351236, 19351237, 19351238, 19351239, 19351240, 19351241, 19351242, 19351243, 19351244, 19351245, 19351246, 19351247, 19351248, 19351249, 19351250, 19351251, 19351252, 19351253, 19351254, 19351255, 19351256, 19351257, 19351258, 19351259, 19351260, 19351261, 19351262, 19351263, 19351264, 19351265, 19351266, 19351267, 19351268, 19351269, 19351270, 19351271, 19351272, 19351273, 19351274, 19351275, 19351276, 19351277, 19351278, 19351279, 19351280, 19351281, 19351282, 19351283, 19351284, 19351285, 19351286, 19351287, 19351288, 19351289, 19351290, 19351291, 19351292, 19351293, 19351294, 19351295, 19351296, 19351297, 19351298, 19351299, 19351300, 19351301, 19351302, 19351303, 19351304, 19351305, 19351306, 19351307, 19351308, 19351309, 19351310, 19351311, 19351312, 19351313, 19351314, 19351315, 19351316, 19351317, 19351318, 19351319, 19351320, 19351321, 19351322, 19351323, 19351324, 19351325, 19351326, 19351327, 19351329, 19351330, 19351331, 19351332, 19351333, 19351334, 19351335, 19351336, 19351337, 19351338, 19351339, 19351340, 19351341, 19351342, 19351343, 19351344, 19351345, 19351346, 19351347, 19351348, 19351349, 19351350, 19351351, 19351352, 19351353, 19351354, 19351355, 19351356, 19351357, 19351358, 19351359, 19351360, 19351361, 19351362, 19351363, 19351364, 19351365, 19351366, 19351367, 19351368, 19351369, 19351370, 19351371, 19351372, 19351373, 19351374, 19351375, 19351376, 19351377, 19351378, 19351379, 19351380, 19351381, 19351382, 19351383, 19351384, 19351385, 19351386, 19351387, 19351388, 19351389, 19351390, 19351391, 19351392, 19351393, 19351394, 19351395, 19351396, 19351397, 19351398, 19351399, 19351400, 19351401, 19351402, 19351403, 19351404, 19351405, 19351406, 19351407, 19351408, 19351409, 19351410, 19351411, 19351412, 19351414, 19351415, 19351416, 19351417, 19351419, 19351420, 19351421, 19351422, 19351423, 19351424, 19351425, 19351426, 19351427, 19351428, 19351429, 19351430, 19351431, 19351432, 19351433, 19351434, 19351435, 19351436, 19351437, 19351438, 19351439, 19351440, 19351441, 19351442, 19351443, 19351444, 19351445, 19351446, 19351447, 19351448, 19351449, 19351450, 19351451, 19351452, 19351453, 19351454, 19351455, 19351456, 19351457, 19351458, 19351459, 19351460, 19351461, 19351462, 19351463, 19351464, 19351465, 19351466, 19351467, 19351468, 19351469, 19351470, 19351471, 19351472, 19351473, 19351474, 19351475, 19351476, 19351477, 19351478, 19351479, 19351480, 19351481, 19351482, 19351483, 19351484, 19351485, 19351486, 19351487, 19351488, 19351489, 19351490, 19351491, 19351492, 19351493, 19351494, 19351495, 19351496, 19351497, 19351498, 19351499, 19351500, 19351501, 19351502, 19351503, 19351504, 19351505, 19351506, 19351507, 19351508, 19351509, 19351510, 19351511, 19351512, 19351513, 19351514, 19351515, 19351516, 19351517, 19351518, 19351519, 19351520, 19351521, 19351522, 19351523, 19351524, 19351525, 19351526, 19351527, 19351528, 19351529, 19351530, 19351531, 19351532, 19351533, 19351534, 19351535, 19351536, 19351537, 19351538, 19351539, 19351540, 19351541, 19351542, 19351543, 19351544, 19351545, 19351546, 19351547, 19351548, 19351549, 19351550, 19351551, 19351552, 19351553, 19351554, 19351555, 19351556, 19351557, 19351558, 19351559, 19351560, 19351561, 19351562, 19351564, 19351565, 19351566, 19351567, 19351568, 19351569, 19351570, 19351571, 19351572, 19351573, 19351574, 19351575, 19351576, 19351577, 19351578, 19351579, 19351580, 19351581, 19351582, 19351583, 19351584, 19351585, 19351586, 19351587, 19351588, 19351589, 19351590, 19351591, 19351592, 19351593, 19351594, 19351595, 19351596, 19351597, 19351598, 19351599, 19351600, 19351601, 19351602, 19351603, 19351604, 19351605, 19351606, 19351607, 19351608, 19351609, 19351610, 19351611, 19351612, 19351613, 19351614, 19351615, 19351616, 19351617, 19351618, 19351619, 19351620, 19351621, 19351622, 19351623, 19351624, 19351625, 19351626, 19351627, 19351628, 19351629, 19351630, 19351631, 19351632, 19351633, 19351634, 19351635, 19351636, 19351637, 19351638, 19351639, 19351640, 19351641, 19351642, 19351644, 19351645, 19351646, 19351647, 19351648, 19351649, 19351650, 19351651, 19351652, 19351653, 19351654, 19351655, 19351656, 19351657, 19351658, 19351659, 19351660, 19351661, 19351662, 19351663, 19351664, 19351665, 19351666, 19351667, 19351668, 19351669, 19351670, 19351671, 19351672, 19351673, 19351674, 19351675, 19351676, 19351677, 19351678, 19351679, 19351680, 19351681, 19351682, 19351683, 19351684, 19351685, 19351686, 19351687, 19351688, 19351689, 19351690, 19351691, 19351692, 19351693, 19351694, 19351695, 19351696, 19351697, 19351698, 19351699, 19351700, 19351701, 19351702, 19351703, 19351704, 19351705, 19351706, 19351707, 19351708, 19351709, 19351710, 19351711, 19351712, 19351713, 19351714, 19351716, 19351717, 19351718, 19351719, 19351720, 19351721, 19351722, 19351723, 19351724, 19351725, 19351726, 19351727, 19351728, 19351729, 19351730, 19351731, 19351732, 19351733, 19351734, 19351735, 19351736, 19351737, 19351738, 19351739, 19351740, 19351741, 19351742, 19351743, 19351744, 19351745, 19351746, 19351747, 19351748, 19351749, 19351750, 19351751, 19351752, 19351753, 19351755, 19351756, 19351757, 19351758, 19351759, 19351760, 19351761, 19351762, 19351763, 19351764, 19351765, 19351766, 19351767, 19351768, 19351769, 19351770, 19351771, 19351772, 19351773, 19351774, 19351775, 19351776, 19351777, 19351778, 19351779, 19351780, 19351781, 19351782, 19351783, 19351784, 19351785, 19351786, 19351787, 19351788, 19351789, 19351790, 19351791, 19351792, 19351793, 19351794, 19351795, 19351796, 19351797, 19351798, 19351799, 19351800, 19351801, 19351802, 19351803, 19351804, 19351805, 19351806, 19351807, 19351808, 19351809, 19351810, 19351811, 19351812, 19351813, 19351814, 19351815, 19351816, 19351817, 19351818, 19351819, 19351820, 19351821, 19351822, 19351823, 19351824, 19351825, 19351826, 19351827, 19351828, 19351829, 19351830, 19351831, 19351832, 19351833, 19351834, 19351835, 19351836, 19351837, 19351838, 19351839, 19351840, 19351841, 19351842, 19351843, 19351844, 19351845, 19351846, 19351847, 19351848, 19351849, 19351850, 19351851, 19351852, 19351853, 19351854, 19351855, 19351856, 19351857, 19351858, 19351859, 19351860, 19351861, 19351862, 19351863, 19351864, 19351865, 19351866, 19351867, 19351868, 19351870, 19351871, 19351872, 19351873, 19351874, 19351875, 19351876, 19351877, 19351878, 19351879, 19351880, 19351881, 19351882, 19351883, 19351884, 19351885, 19351886, 19351887, 19351888, 19351889, 19351890, 19351891, 19351892, 19351893, 19351894, 19351895, 19351896, 19351897, 19351899, 19351900, 19351901, 19351902, 19351903, 19351904, 19351905, 19351906, 19351907, 19351908, 19351909, 19351910, 19351911, 19351912, 19351913, 19351914, 19351915, 19351916, 19351917, 19351918, 19351919, 19351920, 19351921, 19351922, 19351923, 19351924, 19351925, 19351926, 19351927, 19351928, 19351929, 19351930, 19351931, 19351932, 19351933, 19351934, 19351935, 19351936, 19351937, 19351938, 19351939, 19351940, 19351941, 19351942, 19351943, 19351944, 19351945, 19351946, 19351947, 19351948, 19351949, 19351950, 19351951, 19351952, 19351953, 19351954, 19351955, 19351956, 19351957, 19351958, 19351959, 19351960, 19351961, 19351962, 19351963, 19351964, 19351965, 19351966, 19351967, 19351968, 19351969, 19351970, 19351971, 19351972, 19351973, 19351974, 19351975, 19351976, 19351977, 19351978, 19351979, 19351980, 19351981, 19351982, 19351983, 19351984, 19351985, 19351986, 19351987, 19351988, 19351989, 19351990, 19351991, 19351992, 19351993, 19351995, 19351996, 19351997, 19351998, 19351999, 19352000, 19352001, 19352002, 19352003, 19352004, 19352007, 19352008, 19352009, 19352011, 19352012, 19352014, 19352015, 19352018, 19352020, 19352022, 19352025, 19352026, 19352027, 19352028, 19352031, 19352034, 19352035, 19352036, 19352037, 19352039, 19352044, 19352048, 19352049, 19352050, 19352051, 19352053, 19352054, 19352055, 19352056, 19352057, 19352058, 19352060, 19352064, 19352065, 19352066, 19352067, 19352068, 19352070, 19352072, 19352074, 19352076, 19352080, 19352081, 19352082, 19352083, 19352084, 19352085, 19352086, 19352087, 19352088, 19352089, 19352090, 19352091, 19352092, 19352095, 19352096, 19352097, 19352098, 19352099, 19352100, 19352101, 19352106, 19352107, 19352108, 19352110, 19352111, 19352112, 19352113, 19352115, 19352116, 19352117, 19352118, 19352120, 19352121, 19352122, 19352123, 19352124, 19352125, 19352126, 19352127, 19352128, 19352129, 19352132, 19352135, 19352136, 19352137, 19352140, 19352141, 19352142, 19352143, 19352144, 19352145, 19352146, 19352147, 19352149, 19352150, 19352151, 19352152, 19352153, 19352154, 19352156, 19352157, 19352158, 19352159, 19352161, 19352162, 19352163, 19352164, 19352165, 19352166, 19352167, 19352168, 19352169, 19352170, 19352171, 19352172, 19352173, 19352174, 19352175, 19352176, 19352177, 19352178, 19352179, 19352180, 19352181, 19352182, 19352183, 19352184, 19352185, 19352186, 19352187, 19352188, 19352189, 19352190, 19352191, 19352192, 19352193, 19352194, 19352195, 19352196, 19352199, 19352200, 19352201, 19352202, 19352203, 19352204, 19352205, 19352206, 19352207, 19352209, 19352210, 19352211, 19352214, 19352215, 19352217, 19352218, 19352219, 19352220, 19352222, 19352225, 19352226, 19352230, 19352231, 19352232, 19352233, 19352234, 19352236, 19352238, 19352240, 19352241, 19352242, 19352243, 19352244, 19352246, 19352247, 19352248, 19352249, 19352251, 19352252, 19352253, 19352254, 19352257, 19352258, 19352259, 19352261, 19352262, 19352263, 19352267, 19352270, 19352276, 19352278, 19352279, 19352280, 19352281, 19352282, 19352283, 19352284, 19352285, 19352286, 19352287, 19352288, 19352289, 19352290, 19352291, 19352292, 19352293, 19352294, 19352295, 19352296, 19352297, 19352298, 19352299, 19352300, 19352302, 19352303, 19352304, 19352305, 19352306, 19352307, 19352308, 19352309, 19352310, 19352311, 19352312, 19352313, 19352314, 19352315, 19352316, 19352317, 19352318, 19352319, 19352320, 19352321, 19352322, 19352323, 19352324, 19352325, 19352326, 19352327, 19352328, 19352329, 19352330, 19352331, 19352332, 19352333, 19352334, 19352335, 19352336, 19352337, 19352338, 19352339, 19352340, 19352341, 19352343, 19352344, 19352345, 19352346, 19352348, 19352349, 19352350, 19352351, 19352352, 19352353, 19352354, 19352355, 19352356, 19352358, 19352359, 19352360, 19352361, 19352362, 19352363, 19352364, 19352365, 19352366, 19352367, 19352368, 19352369, 19352370, 19352371, 19352372, 19352373, 19352374, 19352375, 19352376, 19352377, 19352378, 19352379, 19352380, 19352381, 19352382, 19352383, 19352384, 19352385, 19352386, 19352387, 19352388, 19352389, 19352390, 19352391, 19352392, 19352393, 19352394, 19352395, 19352396, 19352397, 19352398, 19352399, 19352400, 19352401, 19352402, 19352403, 19352404, 19352405, 19352406, 19352407, 19352408, 19352409, 19352410, 19352411, 19352412, 19352413, 19352414, 19352415, 19352416, 19352417, 19352418, 19352419, 19352420, 19352421, 19352422, 19352423, 19352424, 19352425, 19352426, 19352427, 19352428, 19352429, 19352430, 19352431, 19352432, 19352433, 19352434, 19352435, 19352436, 19352437, 19352438, 19352439, 19352440, 19352441, 19352442, 19352443, 19352444, 19352445, 19352446, 19352447, 19352448, 19352449, 19352450, 19352451, 19352452, 19352453, 19352454, 19352455, 19352456, 19352457, 19352458, 19352459, 19352460, 19352461, 19352462, 19352463, 19352464, 19352465, 19352466, 19352467, 19352468, 19352469, 19352470, 19352471, 19352472, 19352473, 19352474, 19352475, 19352476, 19352477, 19352478, 19352479, 19352480, 19352481, 19352482, 19352483, 19352484, 19352485, 19352486, 19352487, 19352488, 19352489, 19352490, 19352491, 19352492, 19352493, 19352494, 19352495, 19352496, 19352497, 19352498, 19352499, 19352500, 19352501, 19352502, 19352503, 19352504, 19352505, 19352506, 19352507, 19352508, 19352509, 19352510, 19352511, 19352512, 19352513, 19352514, 19352515, 19352516, 19352517, 19352518, 19352519, 19352520, 19352521, 19352522, 19352523, 19352524, 19352525, 19352526, 19352527, 19352528, 19352529, 19352530, 19352531, 19352532, 19352533, 19352534, 19352535, 19352536, 19352537, 19352538, 19352539, 19352540, 19352541, 19352542, 19352543, 19352544, 19352545, 19352546, 19352547, 19352548, 19352549, 19352550, 19352551, 19352552, 19352553, 19352554, 19352555, 19352556, 19352557, 19352558, 19352559, 19352560, 19352561, 19352562, 19352563, 19352564, 19352566, 19352567, 19352568, 19352569, 19352570, 19352571, 19352572, 19352573, 19352574, 19352575, 19352576, 19352577, 19352578, 19352579, 19352580, 19352581, 19352582, 19352583, 19352584, 19352585, 19352586, 19352587, 19352588, 19352589, 19352590, 19352591, 19352592, 19352593, 19352594, 19352595, 19352596, 19352597, 19352598, 19352599, 19352600, 19352601, 19352602, 19352603, 19352604, 19352605, 19352606, 19352607, 19352608, 19352609, 19352610, 19352611, 19352612, 19352613, 19352614, 19352615, 19352616, 19352617, 19352618, 19352619, 19352620, 19352621, 19352622, 19352623, 19352624, 19352625, 19352626, 19352627, 19352628, 19352629, 19352630, 19352631, 19352632, 19352633, 19352634, 19352635, 19352636, 19352637, 19352638, 19352639, 19352640, 19352641, 19352642, 19352643, 19352644, 19352645, 19352646, 19352647, 19352648, 19352649, 19352650, 19352651, 19352652, 19352653, 19352654, 19352655, 19352656, 19352657, 19352658, 19352659, 19352660, 19352661, 19352662, 19352663, 19352664, 19352665, 19352666, 19352667, 19352668, 19352669, 19352670, 19352671, 19352672, 19352673, 19352674, 19352675, 19352676, 19352677, 19352678, 19352679, 19352680, 19352681, 19352682, 19352683, 19352684, 19352685, 19352686, 19352687, 19352688, 19352689, 19352690, 19352691, 19352692, 19352693, 19352694, 19352695, 19352696, 19352697, 19352698, 19352699, 19352700, 19352701, 19352702, 19352703, 19352704, 19352705, 19352706, 19352707, 19352708, 19352709, 19352710, 19352711, 19352712, 19352713, 19352714, 19352715, 19352716, 19352717, 19352718, 19352719, 19352720, 19352721, 19352722, 19352723, 19352724, 19352725, 19352726, 19352727, 19352728, 19352729, 19352730, 19352731, 19352732, 19352733, 19352734, 19352735, 19352736, 19352737, 19352738, 19352739, 19352740, 19352741, 19352742, 19352743, 19352744, 19352745, 19352746, 19352747, 19352748, 19352749, 19352750, 19352751, 19352752, 19352753, 19352754, 19352755, 19352756, 19352757, 19352758, 19352759, 19352760, 19352762, 19352763, 19352764, 19352765, 19352766, 19352767, 19352768, 19352769, 19352770, 19352771, 19352772, 19352773, 19352774, 19352775, 19352776, 19352777, 19352778, 19352779, 19352780, 19352781, 19352782, 19352783, 19352784, 19352785, 19352786, 19352787, 19352788, 19352789, 19352790, 19352791, 19352792, 19352793, 19352794, 19352795, 19352796, 19352797, 19352798, 19352799, 19352800, 19352801, 19352802, 19352803, 19352804, 19352805, 19352806, 19352807, 19352808, 19352809, 19352810, 19352811, 19352812, 19352813, 19352814, 19352815, 19352816, 19352817, 19352818, 19352819, 19352820, 19352821, 19352822, 19352823, 19352824, 19352825, 19352826, 19352827, 19352828, 19352829, 19352830, 19352831, 19352832, 19352833, 19352834, 19352835, 19352836, 19352837, 19352838, 19352839, 19352840, 19352841, 19352842, 19352843, 19352844, 19352845, 19352846, 19352847, 19352848, 19352849, 19352850, 19352851, 19352852, 19352853, 19352854, 19352855, 19352856, 19352857, 19352858, 19352859, 19352860, 19352861, 19352862, 19352863, 19352864, 19352865, 19352866, 19352867, 19352868, 19352869, 19352870, 19352871, 19352872, 19352873, 19352874, 19352875, 19352876, 19352877, 19352878, 19352879, 19352880, 19352881, 19352882, 19352883, 19352884, 19352885, 19352886, 19352887, 19352888, 19352889, 19352890, 19352891, 19352892, 19352893, 19352894, 19352895, 19352896, 19352897, 19352898, 19352899, 19352900, 19352901, 19352902, 19352903, 19352904, 19352905, 19352906, 19352907, 19352908, 19352909, 19352910, 19352911, 19352912, 19352913, 19352914, 19352915, 19352916, 19352917, 19352918, 19352919, 19352920, 19352921, 19352922, 19352923, 19352924, 19352925, 19352926, 19352927, 19352928, 19352929, 19352930, 19352931, 19352932, 19352933, 19352934, 19352935, 19352936, 19352937, 19352938, 19352939, 19352940, 19352941, 19352942, 19352943, 19352944, 19352945, 19352946, 19352947, 19352948, 19352949, 19352950, 19352951, 19352953, 19352954, 19352955, 19352956, 19352957, 19352958, 19352959, 19352960, 19352961, 19352962, 19352963, 19352964, 19352965, 19352966, 19352967, 19352968, 19352969, 19352970, 19352971, 19352972, 19352973, 19352974, 19352975, 19352976, 19352977, 19352978, 19352979, 19352980, 19352981, 19352982, 19352983, 19352984, 19352985, 19352986, 19352987, 19352988, 19352989, 19352990, 19352991, 19352992, 19352993, 19352994, 19352995, 19352996, 19352997, 19352998, 19352999, 19353000, 19353001, 19353002, 19353003, 19353004, 19353005, 19353006, 19353007, 19353008, 19353009, 19353010, 19353011, 19353012, 19353013, 19353014, 19353015, 19353016, 19353017, 19353018, 19353019, 19353020, 19353021, 19353022, 19353023, 19353024, 19353025, 19353026, 19353027, 19353028, 19353029, 19353030, 19353031, 19353032, 19353033, 19353034, 19353035, 19353036, 19353037, 19353038, 19353039, 19353040, 19353041, 19353042, 19353043, 19353044, 19353045, 19353046, 19353047, 19353048, 19353049, 19353050, 19353051, 19353052, 19353053, 19353054, 19353055, 19353056, 19353057, 19353058, 19353059, 19353060, 19353061, 19353062, 19353063, 19353064, 19353065, 19353066, 19353067, 19353068, 19353069, 19353070, 19353071, 19353072, 19353073, 19353074, 19353075, 19353076, 19353077, 19353078, 19353079, 19353080, 19353081, 19353082, 19353083, 19353084, 19353085, 19353086, 19353087, 19353088, 19353089, 19353090, 19353091, 19353092, 19353093, 19353094, 19353095, 19353096, 19353097, 19353098, 19353099, 19353100, 19353101, 19353102, 19353103, 19353104, 19353105, 19353106, 19353107, 19353108, 19353109, 19353110, 19353111, 19353112, 19353113, 19353114, 19353115, 19353116, 19353117, 19353118, 19353119, 19353120, 19353121, 19353122, 19353123, 19353124, 19353125, 19353126, 19353127, 19353128, 19353129, 19353130, 19353131, 19353132, 19353133, 19353134, 19353135, 19353136, 19353137, 19353138, 19353139, 19353140, 19353141, 19353142, 19353143, 19353144, 19353145, 19353146, 19353147, 19353148, 19353149, 19353150, 19353151, 19353152, 19353153, 19353154, 19353155, 19353156, 19353157, 19353158, 19353159, 19353160, 19353161, 19353162, 19353163, 19353164, 19353165, 19353166, 19353167, 19353168, 19353170, 19353171, 19353172, 19353173, 19353174, 19353175, 19353176, 19353177, 19353179, 19353180, 19353181, 19353182, 19353183, 19353184, 19353185, 19353186, 19353187, 19353188, 19353189, 19353190, 19353191, 19353192, 19353193, 19353194, 19353195, 19353197, 19353198, 19353199, 19353200, 19353201, 19353202, 19353203, 19353204, 19353205, 19353206, 19353207, 19353208, 19353209, 19353210, 19353211, 19353212, 19353213, 19353214, 19353215, 19353216, 19353217, 19353218, 19353219, 19353220, 19353221, 19353222, 19353223, 19353224, 19353225, 19353226, 19353227, 19353228, 19353229, 19353230, 19353231, 19353232, 19353233, 19353234, 19353235, 19353236, 19353237, 19353238, 19353239, 19353240, 19353241, 19353242, 19353243, 19353244, 19353245, 19353246, 19353247, 19353248, 19353249, 19353250, 19353251, 19353252, 19353253, 19353254, 19353255, 19353256, 19353257, 19353258, 19353259, 19353260, 19353261, 19353262, 19353263, 19353264, 19353265, 19353266, 19353267, 19353268, 19353269, 19353270, 19353271, 19353272, 19353273, 19353274, 19353275, 19353276, 19353277, 19353278, 19353279, 19353281, 19353282, 19353283, 19353284, 19353285, 19353286, 19353287, 19353288, 19353289, 19353290, 19353291, 19353292, 19353293, 19353295, 19353297, 19353298, 19353299, 19353300, 19353301, 19353302, 19353303, 19353304, 19353305, 19353306, 19353307, 19353308, 19353309, 19353310, 19353311, 19353312, 19353313, 19353314, 19353315, 19353316, 19353317, 19353318, 19353319, 19353320, 19353321, 19353322, 19353323, 19353324, 19353325, 19353326, 19353327, 19353328, 19353329, 19353330, 19353331, 19353332, 19353333, 19353334, 19353335, 19353336, 19353338, 19353339, 19353340, 19353341, 19353342, 19353343, 19353344, 19353345, 19353346, 19353347, 19353348, 19353349, 19353351, 19353352, 19353353, 19353354, 19353355, 19353356, 19353357, 19353358, 19353359, 19353360, 19353361, 19353362, 19353363, 19353364, 19353365, 19353366, 19353367, 19353368, 19353369, 19353370, 19353371, 19353372, 19353373, 19353374, 19353375, 19353376, 19353377, 19353378, 19353379, 19353380, 19353381, 19353382, 19353383, 19353384, 19353385, 19353386, 19353387, 19353388, 19353389, 19353390, 19353391, 19353392, 19353393, 19353394, 19353395, 19353396, 19353397, 19353398, 19353399, 19353400, 19353401, 19353402, 19353403, 19353404, 19353405, 19353406, 19353407, 19353408, 19353409, 19353410, 19353411, 19353412, 19353413, 19353414, 19353415, 19353416, 19353417, 19353418, 19353419, 19353420, 19353421, 19353422, 19353423, 19353424, 19353425, 19353426, 19353427, 19353428, 19353429, 19353430, 19353431, 19353432, 19353433, 19353434, 19353435, 19353436, 19353437, 19353438, 19353439, 19353440, 19353441, 19353442, 19353443, 19353444, 19353445, 19353446, 19353448, 19353449, 19353450, 19353451, 19353452, 19353453, 19353454, 19353455, 19353456, 19353457, 19353458, 19353459, 19353460, 19353461, 19353462, 19353463, 19353464, 19353465, 19353466, 19353467, 19353468, 19353469, 19353470, 19353471, 19353472, 19353473, 19353474, 19353475, 19353476, 19353477, 19353478, 19353479, 19353480, 19353481, 19353482, 19353483, 19353484, 19353485, 19353486, 19353487, 19353488, 19353489, 19353490, 19353491, 19353492, 19353493, 19353494, 19353495, 19353496, 19353497, 19353498, 19353499, 19353500, 19353501, 19353502, 19353503, 19353504, 19353505, 19353506, 19353507, 19353508, 19353509, 19353510, 19353511, 19353512, 19353513, 19353514, 19353515, 19353516, 19353517, 19353518, 19353519, 19353520, 19353521, 19353522, 19353523, 19353524, 19353525, 19353526, 19353527, 19353528, 19353529, 19353530, 19353531, 19353532, 19353533, 19353534, 19353535, 19353536, 19353537, 19353538, 19353539, 19353540, 19353541, 19353542, 19353543, 19353544, 19353545, 19353546, 19353547, 19353548, 19353549, 19353550, 19353551, 19353552, 19353553, 19353554, 19353555, 19353556, 19353557, 19353558, 19353559, 19353560, 19353561, 19353562, 19353563, 19353564, 19353565, 19353566, 19353567, 19353568, 19353569, 19353570, 19353571, 19353572, 19353573, 19353574, 19353575, 19353576, 19353577, 19353578, 19353579, 19353580, 19353581, 19353582, 19353583, 19353584, 19353585, 19353586, 19353587, 19353588, 19353589, 19353590, 19353591, 19353592, 19353593, 19353594, 19353595, 19353596, 19353597, 19353598, 19353599, 19353600, 19353601, 19353602, 19353603, 19353604, 19353605, 19353606, 19353607, 19353608, 19353609, 19353610, 19353611, 19353612, 19353613, 19353614, 19353615, 19353616, 19353617, 19353618, 19353619, 19353620, 19353621, 19353622, 19353623, 19353624, 19353625, 19353626, 19353627, 19353628, 19353629, 19353630, 19353631, 19353632, 19353633, 19353634, 19353635, 19353636, 19353637, 19353638, 19353639, 19353640, 19353641, 19353642, 19353643, 19353644, 19353645, 19353646, 19353647, 19353648, 19353649, 19353650, 19353651, 19353652, 19353653, 19353654, 19353655, 19353656, 19353657, 19353658, 19353659, 19353660, 19353661, 19353662, 19353663, 19353664, 19353665, 19353666, 19353667, 19353668, 19353669, 19353670, 19353671, 19353672, 19353673, 19353674, 19353675, 19353676, 19353677, 19353678, 19353679, 19353680, 19353681, 19353682, 19353683, 19353684, 19353685, 19353686, 19353687, 19353688, 19353689, 19353690, 19353691, 19353692, 19353693, 19353694, 19353695, 19353696, 19353697, 19353698, 19353699, 19353700, 19353701, 19353702, 19353703, 19353704, 19353705, 19353706, 19353707, 19353708, 19353709, 19353710, 19353711, 19353712, 19353713, 19353714, 19353715, 19353716, 19353717, 19353718, 19353719, 19353720, 19353721, 19353722, 19353723, 19353724, 19353725, 19353726, 19353727, 19353728, 19353729, 19353730, 19353731, 19353732, 19353733, 19353734, 19353735, 19353736, 19353737, 19353738, 19353739, 19353740, 19353741, 19353742, 19353743, 19353744, 19353745, 19353746, 19353747, 19353748, 19353749, 19353750, 19353751, 19353752, 19353753, 19353754, 19353755, 19353756, 19353757, 19353758, 19353759, 19353760, 19353761, 19353762, 19353763, 19353764, 19353765, 19353766, 19353767, 19353768, 19353769, 19353770, 19353771, 19353772, 19353773, 19353774, 19353775, 19353776, 19353777, 19353778, 19353779, 19353780, 19353781, 19353782, 19353783, 19353784, 19353785, 19353786, 19353787, 19353788, 19353789, 19353790, 19353792, 19353793, 19353794, 19353795, 19353796, 19353797, 19353798, 19353799, 19353800, 19353801, 19353802, 19353803, 19353804, 19353805, 19353806, 19353807, 19353808, 19353809, 19353811, 19353812, 19353813, 19353814, 19353815, 19353816, 19353817, 19353818, 19353819, 19353820, 19353821, 19353822, 19353823, 19353824, 19353825, 19353826, 19353827, 19353828, 19353829, 19353830, 19353831, 19353832, 19353833, 19353834, 19353835, 19353836, 19353837, 19353838, 19353839, 19353840, 19353841, 19353842, 19353843, 19353845, 19353846, 19353847, 19353848, 19353849, 19353850, 19353851, 19353852, 19353853, 19353854, 19353855, 19353856, 19353857, 19353858, 19353859, 19353861, 19353862, 19353863, 19353864, 19353865, 19353866, 19353867, 19353868, 19353869, 19353870, 19353871, 19353872, 19353873, 19353874, 19353876, 19353877, 19353878, 19353879, 19353880, 19353881, 19353882, 19353883, 19353884, 19353885, 19353886, 19353887, 19353888, 19353889, 19353890, 19353891, 19353892, 19353894, 19353895, 19353896, 19353897, 19353898, 19353899, 19353900, 19353901, 19353902, 19353903, 19353904, 19353905, 19353906, 19353907, 19353908, 19353909, 19353910, 19353911, 19353913, 19353915, 19353916, 19353917, 19353918, 19353919, 19353920, 19353921, 19353922, 19353923, 19353924, 19353925, 19353926, 19353927, 19353928, 19353930, 19353931, 19353932, 19353933, 19353934, 19353935, 19353936, 19353937, 19353938, 19353939, 19353940, 19353941, 19353942, 19353943, 19353944, 19353945, 19353948, 19353949, 19353950, 19353951, 19353952, 19353953, 19353954, 19353955, 19353956, 19353957, 19353958, 19353959, 19353960, 19353961, 19353962, 19353963, 19353964, 19353965, 19353966, 19353967, 19353968, 19353969, 19353970, 19353971, 19353972, 19353973, 19353974, 19353975, 19353976, 19353977, 19353979, 19353980, 19353981, 19353982, 19353983, 19353984, 19353985, 19353986, 19353987, 19353988, 19353989, 19353990, 19353991, 19353992, 19353993, 19353994, 19353995, 19353996, 19353997, 19353998, 19353999, 19354000, 19354001, 19354002, 19354003, 19354004, 19354005, 19354006, 19354007, 19354008, 19354009, 19354010, 19354011, 19354013, 19354014, 19354015, 19354016, 19354017, 19354018, 19354019, 19354020, 19354021, 19354022, 19354023, 19354024, 19354025, 19354026, 19354027, 19354028, 19354029, 19354030, 19354031, 19354032, 19354033, 19354034, 19354035, 19354036, 19354037, 19354038, 19354040, 19354041, 19354042, 19354043, 19354044, 19354045, 19354046, 19354048, 19354049, 19354050, 19354051, 19354052, 19354053, 19354054, 19354055, 19354056, 19354057, 19354058, 19354059, 19354060, 19354061, 19354062, 19354063, 19354064, 19354065, 19354066, 19354067, 19354068, 19354069, 19354070, 19354071, 19354072, 19354073, 19354075, 19354076, 19354077, 19354078, 19354079, 19354080, 19354081, 19354082, 19354083, 19354084, 19354085, 19354086, 19354087, 19354088, 19354089, 19354090, 19354091, 19354092, 19354093, 19354094, 19354095, 19354096, 19354097, 19354098, 19354099, 19354100, 19354101, 19354102, 19354103, 19354104, 19354106, 19354107, 19354108, 19354109, 19354110, 19354111, 19354112, 19354113, 19354114, 19354116, 19354117, 19354118, 19354120, 19354121, 19354122, 19354123, 19354124, 19354125, 19354126, 19354127, 19354128, 19354129, 19354130, 19354131, 19354132, 19354133, 19354135, 19354136, 19354137, 19354139, 19354140, 19354141, 19354142, 19354143, 19354144, 19354146, 19354147, 19354148, 19354149, 19354150, 19354151, 19354152, 19354153, 19354154, 19354155, 19354156, 19354157, 19354158, 19354159, 19354160, 19354161, 19354162, 19354163, 19354164, 19354165, 19354166, 19354167, 19354168, 19354169, 19354170, 19354171, 19354172, 19354173, 19354174, 19354175, 19354176, 19354177, 19354178, 19354179, 19354180, 19354181, 19354182, 19354183, 19354184, 19354185, 19354186, 19354187, 19354188, 19354189, 19354190, 19354191, 19354192, 19354193, 19354194, 19354196, 19354197, 19354198, 19354199, 19354200, 19354202, 19354203, 19354204, 19354205, 19354206, 19354207, 19354208, 19354209, 19354210, 19354211, 19354212, 19354213, 19354214, 19354215, 19354216, 19354217, 19354218, 19354219, 19354220, 19354221, 19354222, 19354223, 19354224, 19354225, 19354226, 19354227, 19354228, 19354229, 19354230, 19354231, 19354233, 19354234, 19354235, 19354236, 19354237, 19354238, 19354239, 19354240, 19354242, 19354243, 19354244, 19354245, 19354247, 19354248, 19354249, 19354250, 19354251, 19354252, 19354253, 19354254, 19354255, 19354256, 19354257, 19354258, 19354259, 19354260, 19354261, 19354262, 19354263, 19354264, 19354265, 19354267, 19354268, 19354269, 19354270, 19354271, 19354272, 19354273, 19354274, 19354275, 19354276, 19354277, 19354278, 19354279, 19354280, 19354281, 19354282, 19354283, 19354284, 19354285, 19354286, 19354287, 19354288, 19354289, 19354291, 19354292, 19354294, 19354295, 19354296, 19354297, 19354298, 19354299, 19354300, 19354301, 19354302, 19354304, 19354305, 19354306, 19354307, 19354308, 19354309, 19354310, 19354311, 19354313, 19354314, 19354315, 19354318, 19354319, 19354320, 19354321, 19354323, 19354324, 19354325, 19354326, 19354327, 19354330, 19354331, 19354332, 19354333, 19354334, 19354336, 19354337, 19354338, 19354339, 19354340, 19354341, 19354342, 19354343, 19354344, 19354345, 19354346, 19354347, 19354348, 19354349, 19354351, 19354353, 19354354, 19354355, 19354356, 19354357, 19354358, 19354360, 19354361, 19354362, 19354363, 19354365, 19354366, 19354367, 19354368, 19354369, 19354370, 19354371, 19354372, 19354373, 19354374, 19354375, 19354376, 19354378, 19354382, 19354384, 19354386, 19354387, 19354388, 19354392, 19354393, 19354394, 19354395, 19354396, 19354398, 19354401, 19354402, 19354403, 19354406, 19354407, 19354408, 19354409, 19354410, 19354412, 19354413, 19354415, 19354417, 19354419, 19354420, 19354421, 19354422, 19354423, 19354424, 19354425, 19354426, 19354427, 19354428, 19354429, 19354430, 19354431, 19354432, 19354433, 19354434, 19354436, 19354437, 19354438, 19354440, 19354441, 19354442, 19354443, 19354444, 19354445, 19354446, 19354447, 19354448, 19354449, 19354450, 19354451, 19354452, 19354453, 19354454, 19354455, 19354456, 19354457, 19354458, 19354459, 19354460, 19354461, 19354462, 19354463, 19354464, 19354465, 19354466, 19354468, 19354469, 19354470, 19354471, 19354472, 19354473, 19354474, 19354475, 19354476, 19354477, 19354478, 19354479, 19354480, 19354481, 19354482, 19354483, 19354484, 19354485, 19354486, 19354487, 19354488, 19354489, 19354490, 19354491, 19354492, 19354493, 19354494, 19354495, 19354496, 19354497, 19354498, 19354499, 19354500, 19354501, 19354502, 19354504, 19354505, 19354506, 19354507, 19354508, 19354509, 19354510, 19354511, 19354512, 19354513, 19354514, 19354515, 19354516, 19354517, 19354518, 19354520, 19354521, 19354522, 19354523, 19354524, 19354525, 19354526, 19354527, 19354528, 19354529, 19354530, 19354531, 19354532, 19354533, 19354534, 19354535, 19354536, 19354537, 19354538, 19354539, 19354540, 19354541, 19354542, 19354543, 19354544, 19354545, 19354546, 19354547, 19354548, 19354549, 19354550, 19354551, 19354552, 19354553, 19354555, 19354556, 19354557, 19354558, 19354559, 19354560, 19354561, 19354562, 19354563, 19354564, 19354565, 19354566, 19354567, 19354568, 19354569, 19354570, 19354571, 19354572, 19354573, 19354574, 19354575, 19354576, 19354578, 19354579, 19354580, 19354581, 19354582, 19354583, 19354584, 19354585, 19354587, 19354589, 19354590, 19354591, 19354592, 19354593, 19354594, 19354595, 19354596, 19354597, 19354598, 19354599, 19354600, 19354602, 19354603, 19354604, 19354605, 19354606, 19354607, 19354608, 19354609, 19354610, 19354611, 19354612, 19354613, 19354614, 19354615, 19354617, 19354618, 19354619, 19354620, 19354621, 19354622, 19354623, 19354624, 19354625, 19354626, 19354627, 19354628, 19354629, 19354630, 19354631, 19354632, 19354633, 19354634, 19354635, 19354636, 19354637, 19354638, 19354639, 19354640, 19354641, 19354642, 19354644, 19354646, 19354647, 19354648, 19354649, 19354650, 19354651, 19354652, 19354653, 19354654, 19354655, 19354656, 19354660, 19354661, 19354662, 19354664, 19354666, 19354667, 19354668, 19354669, 19354670, 19354671, 19354672, 19354673, 19354674, 19354675, 19354676, 19354677, 19354678, 19354679, 19354680, 19354681, 19354685, 19354686, 19354687, 19354688, 19354689, 19354692, 19354693, 19354694, 19354695, 19354696, 19354697, 19354698, 19354699, 19354700, 19354701, 19354702, 19354704, 19354705, 19354706, 19354707, 19354708, 19354709, 19354710, 19354712, 19354713, 19354714, 19354715, 19354716, 19354717, 19354718, 19354719, 19354720, 19354721, 19354722, 19354723, 19354724, 19354725, 19354726, 19354728, 19354729, 19354730, 19354731, 19354732, 19354733, 19354735, 19354736, 19354737, 19354738, 19354739, 19354740, 19354741, 19354742, 19354743, 19354744, 19354745, 19354746, 19354747, 19354748, 19354749, 19354750, 19354751, 19354752, 19354754, 19354755, 19354756, 19354757, 19354758, 19354759, 19354760, 19354763, 19354764, 19354765, 19354766, 19354767, 19354768, 19354769, 19354770, 19354771, 19354772, 19354773, 19354774, 19354776, 19354777, 19354779, 19354780, 19354781, 19354782, 19354783, 19354784, 19354785, 19354786, 19354787, 19354790, 19354791, 19354792, 19354793, 19354794, 19354795, 19354796, 19354797, 19354798, 19354800, 19354802, 19354803, 19354804, 19354805, 19354806, 19354807, 19354808, 19354809, 19354810, 19354811, 19354812, 19354813, 19354814, 19354815, 19354817, 19354818, 19354819, 19354820, 19354821, 19354822, 19354823, 19354824, 19354825, 19354826, 19354827, 19354828, 19354829, 19354830, 19354832, 19354834, 19354835, 19354836, 19354837, 19354838, 19354839, 19354841, 19354842, 19354843, 19354844, 19354845, 19354846, 19354847, 19354848, 19354849, 19354850, 19354851, 19354852, 19354853, 19354854, 19354855, 19354857, 19354858, 19354859, 19354860, 19354862, 19354865, 19354868, 19354869, 19354871, 19354872, 19354873, 19354874, 19354875, 19354877, 19354880, 19354881, 19354883, 19354884, 19354885, 19354886, 19354887, 19354888, 19354889, 19354890, 19354891, 19354892, 19354893, 19354894, 19354895, 19354896, 19354897, 19354898, 19354899, 19354900, 19354901, 19354902, 19354903, 19354904, 19354905, 19354906, 19354907, 19354908, 19354909, 19354911, 19354912, 19354913, 19354914, 19354915, 19354916, 19354917, 19354918, 19354919, 19354920, 19354921, 19354922, 19354923, 19354924, 19354925, 19354926, 19354927, 19354928, 19354929, 19354930, 19354931, 19354932, 19354933, 19354934, 19354935, 19354936, 19354937, 19354938, 19354939, 19354940, 19354941, 19354942, 19354943, 19354944, 19354945, 19354946, 19354947, 19354948, 19354949, 19354950, 19354951, 19354952, 19354953, 19354954, 19354956, 19354957, 19354958, 19354959, 19354960, 19354961, 19354962, 19354963, 19354964, 19354965, 19354966, 19354967, 19354968, 19354969, 19354970, 19354971, 19354972, 19354973, 19354974, 19354975, 19354976, 19354977, 19354978, 19354979, 19354980, 19354981, 19354982, 19354983, 19354984, 19354985, 19354986, 19354987, 19354988, 19354990, 19354991, 19354992, 19354993, 19354994, 19354995, 19354996, 19354997, 19354998, 19354999, 19355000, 19355001, 19355002, 19355003, 19355004, 19355005, 19355006, 19355008, 19355009, 19355010, 19355011, 19355012, 19355013, 19355014, 19355015, 19355016, 19355017, 19355018, 19355021, 19355022, 19355023, 19355024, 19355025, 19355026, 19355027, 19355028, 19355029, 19355030, 19355031, 19355032, 19355033, 19355034, 19355035, 19355036, 19355037, 19355038, 19355039, 19355040, 19355041, 19355042, 19355043, 19355044, 19355045, 19355046, 19355047, 19355048, 19355049, 19355050, 19355051, 19355052, 19355053, 19355054, 19355055, 19355056, 19355057, 19355058, 19355059, 19355060, 19355061, 19355062, 19355063, 19355064, 19355065, 19355066, 19355067, 19355068, 19355069, 19355070, 19355071, 19355072, 19355073, 19355074, 19355075, 19355076, 19355077, 19355078, 19355079, 19355080, 19355081, 19355082, 19355083, 19355084, 19355085, 19355086, 19355087, 19355088, 19355089, 19355090, 19355091, 19355093, 19355094, 19355095, 19355096, 19355097, 19355098, 19355099, 19355100, 19355101, 19355102, 19355103, 19355104, 19355105, 19355106, 19355107, 19355108, 19355109, 19355110, 19355111, 19355112, 19355113, 19355114, 19355115, 19355117, 19355118, 19355119, 19355120, 19355121, 19355122, 19355123, 19355124, 19355125, 19355126, 19355127, 19355128, 19355129, 19355130, 19355131, 19355132, 19355133, 19355134, 19355135, 19355136, 19355137, 19355138, 19355139, 19355140, 19355141, 19355142, 19355143, 19355144, 19355145, 19355146, 19355147, 19355148, 19355149, 19355150, 19355151, 19355152, 19355153, 19355154, 19355155, 19355156, 19355157, 19355158, 19355159, 19355160, 19355161, 19355162, 19355163, 19355164, 19355165, 19355166, 19355167, 19355168, 19355169, 19355170, 19355171, 19355172, 19355173, 19355174, 19355175, 19355176, 19355177, 19355178, 19355180, 19355181, 19355182, 19355183, 19355184, 19355185, 19355186, 19355187, 19355188, 19355189, 19355190, 19355191, 19355193, 19355194, 19355195, 19355196, 19355197, 19355198, 19355199, 19355200, 19355201, 19355202, 19355203, 19355204, 19355205, 19355206, 19355207, 19355208, 19355209, 19355211, 19355212, 19355213, 19355214, 19355215, 19355216, 19355217, 19355218, 19355219, 19355220, 19355221, 19355222, 19355223, 19355224, 19355225, 19355226, 19355227, 19355228, 19355229, 19355230, 19355231, 19355232, 19355233, 19355234, 19355235, 19355236, 19355237, 19355238, 19355239, 19355240, 19355241, 19355243, 19355244, 19355245, 19355246, 19355247, 19355248, 19355249, 19355250, 19355251, 19355252, 19355253, 19355254, 19355255, 19355256, 19355258, 19355259, 19355260, 19355261, 19355262, 19355263, 19355264, 19355265, 19355266, 19355267, 19355268, 19355269, 19355270, 19355271, 19355272, 19355273, 19355274, 19355275, 19355276, 19355277, 19355278, 19355279, 19355280, 19355281, 19355282, 19355283, 19355284, 19355285, 19355286, 19355287, 19355288, 19355289, 19355290, 19355291, 19355293, 19355294, 19355295, 19355296, 19355297, 19355298, 19355300, 19355301, 19355302, 19355303, 19355304, 19355305, 19355306, 19355307, 19355308, 19355309, 19355310, 19355311, 19355312, 19355313, 19355314, 19355315, 19355316, 19355317, 19355319, 19355320, 19355321, 19355323, 19355324, 19355325, 19355326, 19355327, 19355328, 19355329, 19355330, 19355331, 19355332, 19355333, 19355334, 19355335, 19355336, 19355337, 19355338, 19355339, 19355340, 19355341, 19355342, 19355343, 19355344, 19355345, 19355346, 19355347, 19355348, 19355350, 19355351, 19355352, 19355353, 19355354, 19355355, 19355356, 19355357, 19355358, 19355359, 19355360, 19355361, 19355362, 19355363, 19355364, 19355365, 19355366, 19355367, 19355368, 19355369, 19355370, 19355371, 19355372, 19355373, 19355374, 19355375, 19355376, 19355377, 19355378, 19355379, 19355380, 19355381, 19355382, 19355383, 19355384, 19355385, 19355386, 19355387, 19355388, 19355389, 19355390, 19355392, 19355393, 19355394, 19355395, 19355396, 19355397, 19355398, 19355399, 19355400, 19355401, 19355402, 19355403, 19355404, 19355405, 19355406, 19355407, 19355408, 19355409, 19355410, 19355412, 19355413, 19355414, 19355416, 19355417, 19355418, 19355421, 19355422, 19355424, 19355425, 19355426, 19355427, 19355428, 19355429, 19355430, 19355431, 19355432, 19355433, 19355435, 19355436, 19355437, 19355438, 19355440, 19355441, 19355442, 19355443, 19355444, 19355446, 19355447, 19355448, 19355449, 19355450, 19355451, 19355452, 19355453, 19355454, 19355455, 19355456, 19355457, 19355458, 19355459, 19355460, 19355461, 19355462, 19355463, 19355464, 19355465, 19355466, 19355467, 19355468, 19355469, 19355470, 19355471, 19355472, 19355473, 19355474, 19355475, 19355476, 19355477, 19355478, 19355479, 19355480, 19355481, 19355482, 19355483, 19355484, 19355485, 19355486, 19355487, 19355488, 19355489, 19355490, 19355491, 19355492, 19355493, 19355494, 19355495, 19355496, 19355497, 19355498, 19355499, 19355500, 19355501, 19355502, 19355503, 19355504, 19355505, 19355506, 19355507, 19355508, 19355509, 19355510, 19355511, 19355512, 19355513, 19355514, 19355516, 19355517, 19355518, 19355519, 19355520, 19355521, 19355522, 19355523, 19355524, 19355525, 19355526, 19355527, 19355529, 19355530, 19355531, 19355532, 19355533, 19355534, 19355535, 19355536, 19355537, 19355538, 19355539, 19355540, 19355541, 19355542, 19355543, 19355545, 19355546, 19355547, 19355548, 19355549, 19355550, 19355551, 19355552, 19355553, 19355554, 19355555, 19355556, 19355557, 19355558, 19355559, 19355560, 19355561, 19355562, 19355563, 19355564, 19355565, 19355566, 19355567, 19355568, 19355569, 19355570, 19355571, 19355572, 19355573, 19355574, 19355575, 19355576, 19355577, 19355578, 19355579, 19355580, 19355581, 19355582, 19355583, 19355584, 19355585, 19355587, 19355588, 19355589, 19355590, 19355591, 19355592, 19355593, 19355594, 19355595, 19355596, 19355597, 19355599, 19355600, 19355601, 19355603, 19355605, 19355606, 19355607, 19355608, 19355609, 19355610, 19355611, 19355612, 19355613, 19355614, 19355615, 19355616, 19355617, 19355618, 19355619, 19355620, 19355621, 19355622, 19355623, 19355624, 19355625, 19355626, 19355627, 19355628, 19355629, 19355630, 19355631, 19355632, 19355633, 19355634, 19355635, 19355636, 19355638, 19355640, 19355641, 19355643, 19355644, 19355645, 19355646, 19355647, 19355648, 19355650, 19355654, 19355655, 19355657, 19355658, 19355659, 19355660, 19355661, 19355662, 19355663, 19355664, 19355665, 19355666, 19355667, 19355668, 19355669, 19355670, 19355671, 19355672, 19355673, 19355674, 19355675, 19355676, 19355677, 19355678, 19355679, 19355680, 19355681, 19355682, 19355683, 19355684, 19355685, 19355686, 19355687, 19355688, 19355689, 19355691, 19355692, 19355693, 19355694, 19355695, 19355696, 19355697, 19355698, 19355699, 19355700, 19355702, 19355703, 19355704, 19355705, 19355706, 19355707, 19355708, 19355709, 19355710, 19355711, 19355712, 19355713, 19355714, 19355715, 19355716, 19355717, 19355718, 19355719, 19355720, 19355721, 19355722, 19355724, 19355725, 19355727, 19355728, 19355729, 19355730, 19355731, 19355732, 19355733, 19355734, 19355735, 19355736, 19355737, 19355739, 19355740, 19355741, 19355743, 19355744, 19355745, 19355748, 19355749, 19355750, 19355751, 19355752, 19355753, 19355754, 19355755, 19355756, 19355757, 19355758, 19355759, 19355761, 19355762, 19355763, 19355764, 19355765, 19355766, 19355767, 19355770, 19355771, 19355772, 19355773, 19355774, 19355775, 19355776, 19355777, 19355778, 19355779, 19355780, 19355782, 19355783, 19355784, 19355785, 19355786, 19355787, 19355789, 19355790, 19355791, 19355792, 19355793, 19355794, 19355795, 19355796, 19355797, 19355798, 19355799, 19355800, 19355801, 19355802, 19355803, 19355804, 19355805, 19355806, 19355808, 19355810, 19355811, 19355812, 19355814, 19355816, 19355817, 19355822, 19355828, 19355830, 19355831, 19355832, 19355834, 19355835, 19355836, 19355837, 19355838, 19355839, 19355840, 19355841, 19355842, 19355843, 19355844, 19355845, 19355846, 19355847, 19355848, 19355849, 19355850, 19355851, 19355852, 19355853, 19355854, 19355855, 19355856, 19355857, 19355858, 19355859, 19355860, 19355861, 19355862, 19355863, 19355864, 19355865, 19355866, 19355867, 19355868, 19355869, 19355870, 19355871, 19355872, 19355873, 19355874, 19355875, 19355876, 19355877, 19355878, 19355879, 19355880, 19355881, 19355882, 19355884, 19355885, 19355886, 19355887, 19355888, 19355889, 19355890, 19355891, 19355892, 19355893, 19355894, 19355895, 19355896, 19355897, 19355898, 19355899, 19355900, 19355901, 19355902, 19355903, 19355904, 19355905, 19355906, 19355907, 19355908, 19355909, 19355910, 19355911, 19355913, 19355914, 19355915, 19355916, 19355917, 19355918, 19355919, 19355920, 19355921, 19355922, 19355923, 19355924, 19355925, 19355926, 19355927, 19355928, 19355929, 19355930, 19355933, 19355934, 19355935, 19355936, 19355937, 19355938, 19355939, 19355940, 19355941, 19355942, 19355943, 19355944, 19355945, 19355946, 19355947, 19355948, 19355949, 19355950, 19355951, 19355953, 19355954, 19355955, 19355956, 19355957, 19355958, 19355959, 19355960, 19355961, 19355962, 19355963, 19355964, 19355965, 19355967, 19355968, 19355969, 19355970, 19355971, 19355972, 19355973, 19355974, 19355979, 19355981, 19355982, 19355987, 19355988, 19355991, 19355992, 19355994, 19355997, 19355998, 19355999, 19356001, 19356003, 19356005, 19356006, 19356008, 19356011, 19356012, 19356013, 19356014, 19356015, 19356016, 19356017, 19356018, 19356019, 19356020, 19356021, 19356022, 19356023, 19356024, 19356025, 19356026, 19356027, 19356028, 19356029, 19356030, 19356031, 19356032, 19356034, 19356035, 19356036, 19356037, 19356038, 19356039, 19356040, 19356041, 19356042, 19356043, 19356044, 19356045, 19356046, 19356047, 19356048, 19356049, 19356050, 19356051, 19356052, 19356053, 19356054, 19356055, 19356056, 19356057, 19356058, 19356059, 19356060, 19356061, 19356062, 19356065, 19356066, 19356067, 19356075, 19356077, 19356079, 19356087, 19356097, 19356098, 19356099, 19356100, 19356101, 19356102, 19356103, 19356104, 19356105, 19356106, 19356107, 19356108, 19356109, 19356110, 19356111, 19356112, 19356113, 19356114, 19356115, 19356117, 19356118, 19356119, 19356120, 19356121, 19356122, 19356123, 19356124, 19356125, 19356126, 19356127, 19356128, 19356129, 19356130, 19356131, 19356132, 19356133, 19356134, 19356135, 19356136, 19356137, 19356138, 19356139, 19356140, 19356141, 19356142, 19356143, 19356144, 19356145, 19356146, 19356147, 19356148, 19356149, 19356150, 19356151, 19356152, 19356153, 19356154, 19356155, 19356156, 19356157, 19356158, 19356159, 19356160, 19356161, 19356162, 19356163, 19356164, 19356165, 19356166, 19356167, 19356168, 19356169, 19356170, 19356171, 19356172, 19356173, 19356174, 19356175, 19356176, 19356177, 19356178, 19356179, 19356180, 19356181, 19356182, 19356183, 19356184, 19356185, 19356186, 19356187, 19356188, 19356189, 19356190, 19356191, 19356192, 19356194, 19356195, 19356196, 19356197, 19356198, 19356199, 19356200, 19356201, 19356202, 19356203, 19356204, 19356205, 19356206, 19356207, 19356208, 19356209, 19356210, 19356211, 19356212, 19356213, 19356214, 19356215, 19356216, 19356217, 19356218, 19356219, 19356220, 19356221, 19356222, 19356223, 19356224, 19356225, 19356226, 19356227, 19356228, 19356229, 19356230, 19356231, 19356232, 19356233, 19356234, 19356236, 19356237, 19356238, 19356239, 19356240, 19356241, 19356242, 19356243, 19356244, 19356245, 19356246, 19356248, 19356249, 19356250, 19356251, 19356252, 19356253, 19356254, 19356255, 19356256, 19356257, 19356258, 19356259, 19356260, 19356261, 19356262, 19356263, 19356264, 19356265, 19356266, 19356267, 19356268, 19356269, 19356270, 19356271, 19356272, 19356273, 19356274, 19356275, 19356276, 19356277, 19356278, 19356279, 19356280, 19356281, 19356282, 19356283, 19356284, 19356285, 19356286, 19356287, 19356288, 19356289, 19356290, 19356291, 19356292, 19356293, 19356294, 19356295, 19356296, 19356297, 19356298, 19356299, 19356300, 19356301, 19356302, 19356303, 19356304, 19356305, 19356306, 19356307, 19356308, 19356309, 19356310, 19356311, 19356312, 19356313, 19356314, 19356315, 19356316, 19356317, 19356318, 19356319, 19356320, 19356321, 19356322, 19356323, 19356324, 19356325, 19356326, 19356327, 19356328, 19356329, 19356330, 19356331, 19356332, 19356333, 19356334, 19356335, 19356336, 19356337, 19356338, 19356339, 19356340, 19356341, 19356342, 19356343, 19356344, 19356345, 19356347, 19356348, 19356349, 19356350, 19356351, 19356352, 19356353, 19356354, 19356355, 19356356, 19356357, 19356358, 19356359, 19356360, 19356361, 19356362, 19356363, 19356364, 19356365, 19356366, 19356367, 19356368, 19356369, 19356370, 19356371, 19356372, 19356373, 19356374, 19356375, 19356376, 19356377, 19356378, 19356379, 19356380, 19356381, 19356382, 19356383, 19356384, 19356385, 19356386, 19356387, 19356388, 19356389, 19356390, 19356391, 19356392, 19356393, 19356394, 19356395, 19356396, 19356397, 19356398, 19356400, 19356401, 19356402, 19356403, 19356405, 19356406, 19356407, 19356408, 19356409, 19356410, 19356411, 19356412, 19356413, 19356414, 19356415, 19356417, 19356418, 19356419, 19356420, 19356421, 19356422, 19356423, 19356424, 19356425, 19356426, 19356427, 19356428, 19356429, 19356430, 19356431, 19356432, 19356433, 19356434, 19356435, 19356436, 19356437, 19356438, 19356440, 19356441, 19356442, 19356443, 19356444, 19356445, 19356446, 19356447, 19356448, 19356449, 19356450, 19356451, 19356452, 19356453, 19356454, 19356455, 19356456, 19356457, 19356458, 19356459, 19356460, 19356461, 19356462, 19356463, 19356464, 19356465, 19356466, 19356467, 19356468, 19356469, 19356470, 19356471, 19356472, 19356473, 19356474, 19356475, 19356476, 19356477, 19356478, 19356479, 19356480, 19356481, 19356482, 19356483, 19356485, 19356486, 19356487, 19356488, 19356489, 19356490, 19356491, 19356492, 19356493, 19356494, 19356495, 19356496, 19356497, 19356498, 19356499, 19356500, 19356501, 19356502, 19356503, 19356504, 19356505, 19356506, 19356507, 19356508, 19356509, 19356510, 19356511, 19356512, 19356513, 19356514, 19356515, 19356516, 19356517, 19356518, 19356519, 19356520, 19356521, 19356522, 19356523, 19356525, 19356526, 19356527, 19356528, 19356529, 19356530, 19356531, 19356532, 19356533, 19356534, 19356535, 19356536, 19356537, 19356538, 19356539, 19356540, 19356541, 19356542, 19356543, 19356544, 19356545, 19356546, 19356547, 19356548, 19356549, 19356550, 19356551, 19356552, 19356553, 19356554, 19356555, 19356556, 19356557, 19356558, 19356559, 19356560, 19356561, 19356562, 19356563, 19356564, 19356565, 19356566, 19356567, 19356568, 19356570, 19356571, 19356572, 19356573, 19356574, 19356575, 19356576, 19356577, 19356578, 19356579, 19356580, 19356581, 19356582, 19356583, 19356584, 19356585, 19356586, 19356587, 19356588, 19356589, 19356590, 19356591, 19356592, 19356593, 19356594, 19356595, 19356596, 19356597, 19356598, 19356599, 19356600, 19356601, 19356602, 19356603, 19356604, 19356605, 19356606, 19356607, 19356608, 19356609, 19356610, 19356611, 19356612, 19356613, 19356614, 19356615, 19356616, 19356617, 19356618, 19356619, 19356620, 19356621, 19356622, 19356623, 19356624, 19356625, 19356626, 19356627, 19356628, 19356629, 19356630, 19356631, 19356632, 19356633, 19356634, 19356635, 19356636, 19356637, 19356638, 19356639, 19356641, 19356642, 19356644, 19356645, 19356646, 19356647, 19356648, 19356649, 19356651, 19356652, 19356653, 19356654, 19356655, 19356656, 19356657, 19356658, 19356659, 19356660, 19356661, 19356662, 19356663, 19356664, 19356665, 19356666, 19356667, 19356668, 19356669, 19356670, 19356671, 19356672, 19356673, 19356674, 19356676, 19356677, 19356678, 19356679, 19356680, 19356681, 19356683, 19356684, 19356686, 19356687, 19356688, 19356689, 19356690, 19356691, 19356692, 19356693, 19356694, 19356695, 19356696, 19356697, 19356698, 19356699, 19356702, 19356703, 19356705, 19356706, 19356708, 19356709, 19356710, 19356711, 19356713, 19356714, 19356715, 19356716, 19356717, 19356718, 19356719, 19356720, 19356721, 19356722, 19356723, 19356724, 19356725, 19356726, 19356727, 19356728, 19356729, 19356730, 19356731, 19356732, 19356733, 19356734, 19356735, 19356736, 19356737, 19356738, 19356739, 19356740, 19356741, 19356742, 19356743, 19356744, 19356745, 19356746, 19356748, 19356749, 19356750, 19356752, 19356753, 19356754, 19356755, 19356756, 19356757, 19356758, 19356759, 19356760, 19356761, 19356762, 19356763, 19356765, 19356767, 19356768, 19356770, 19356771, 19356772, 19356773, 19356774, 19356775, 19356776, 19356777, 19356778, 19356779, 19356780, 19356781, 19356782, 19356783, 19356784, 19356786, 19356787, 19356789, 19356790, 19356791, 19356792, 19356793, 19356794, 19356795, 19356796, 19356797, 19356798, 19356799, 19356800, 19356801, 19356802, 19356803, 19356804, 19356805, 19356806, 19356807, 19356808, 19356809, 19356810, 19356811, 19356812, 19356813, 19356814, 19356815, 19356816, 19356817, 19356818, 19356819, 19356820, 19356821, 19356822, 19356823, 19356824, 19356825, 19356826, 19356827, 19356828, 19356829, 19356830, 19356831, 19356832, 19356833, 19356834, 19356835, 19356836, 19356837, 19356838, 19356839, 19356840, 19356841, 19356842, 19356843, 19356844, 19356845, 19356846, 19356847, 19356849, 19356850, 19356851, 19356852, 19356853, 19356854, 19356855, 19356856, 19356857, 19356858, 19356859, 19356860, 19356861, 19356862, 19356863, 19356864, 19356865, 19356866, 19356867, 19356868, 19356869, 19356870, 19356871, 19356872, 19356873, 19356874, 19356875, 19356876, 19356877, 19356878, 19356879, 19356880, 19356881, 19356882, 19356883, 19356884, 19356885, 19356886, 19356888, 19356889, 19356890, 19356891, 19356892, 19356893, 19356894, 19356895, 19356896, 19356897, 19356899, 19356900, 19356901, 19356902, 19356903, 19356904, 19356905, 19356906, 19356907, 19356908, 19356910, 19356911, 19356912, 19356913, 19356914, 19356915, 19356918, 19356919, 19356920, 19356921, 19356922, 19356924, 19356925, 19356926, 19356927, 19356928, 19356929, 19356930, 19356931, 19356932, 19356933, 19356934, 19356935, 19356936, 19356937, 19356938, 19356939, 19356940, 19356941, 19356942, 19356943, 19356944, 19356945, 19356946, 19356947, 19356948, 19356949, 19356950, 19356951, 19356952, 19356953, 19356954, 19356955, 19356956, 19356957, 19356958, 19356959, 19356960, 19356961, 19356962, 19356963, 19356964, 19356965, 19356966, 19356967, 19356968, 19356969, 19356970, 19356971, 19356972, 19356973, 19356974, 19356975, 19356976, 19356977, 19356978, 19356979, 19356980, 19356981, 19356982, 19356983, 19356984, 19356985, 19356986, 19356987, 19356989, 19356990, 19356991, 19356992, 19356994, 19356995, 19356996, 19356998, 19356999, 19357000, 19357001, 19357002, 19357003, 19357004, 19357005, 19357006, 19357007, 19357008, 19357009, 19357010, 19357011, 19357012, 19357013, 19357014, 19357015, 19357016, 19357017, 19357018, 19357019, 19357020, 19357021, 19357022, 19357023, 19357024, 19357025, 19357026, 19357027, 19357028, 19357029, 19357030, 19357031, 19357032, 19357033, 19357034, 19357035, 19357036, 19357037, 19357038, 19357039, 19357040, 19357041, 19357043, 19357044, 19357045, 19357046, 19357047, 19357048, 19357050, 19357051, 19357052, 19357053, 19357054, 19357055, 19357056, 19357057, 19357058, 19357059, 19357060, 19357061, 19357062, 19357063, 19357064, 19357065, 19357066, 19357067, 19357068, 19357069, 19357070, 19357071, 19357072, 19357073, 19357074, 19357075, 19357076, 19357077, 19357078, 19357079, 19357080, 19357081, 19357082, 19357083, 19357084, 19357085, 19357086, 19357087, 19357088, 19357089, 19357090, 19357091, 19357092, 19357093, 19357094, 19357095, 19357096, 19357097, 19357098, 19357099, 19357100, 19357101, 19357102, 19357103, 19357104, 19357105, 19357106, 19357107, 19357108, 19357109, 19357110, 19357111, 19357112, 19357113, 19357114, 19357115, 19357116, 19357117, 19357118, 19357119, 19357120, 19357121, 19357122, 19357123, 19357124, 19357125, 19357126, 19357127, 19357128, 19357129, 19357130, 19357131, 19357132, 19357133, 19357134, 19357135, 19357136, 19357137, 19357138, 19357139, 19357140, 19357141, 19357142, 19357143, 19357144, 19357145, 19357146, 19357147, 19357148, 19357150, 19357151, 19357152, 19357153, 19357155, 19357156, 19357157, 19357158, 19357159, 19357160, 19357161, 19357162, 19357164, 19357165, 19357166, 19357167, 19357168, 19357169, 19357170, 19357171, 19357172, 19357173, 19357174, 19357175, 19357176, 19357177, 19357178, 19357179, 19357180, 19357181, 19357182, 19357183, 19357184, 19357185, 19357187, 19357188, 19357189, 19357190, 19357191, 19357192, 19357193, 19357194, 19357195, 19357196, 19357197, 19357198, 19357199, 19357200, 19357201, 19357202, 19357203, 19357204, 19357205, 19357206, 19357207, 19357208, 19357209, 19357210, 19357211, 19357212, 19357213, 19357214, 19357216, 19357217, 19357218, 19357219, 19357220, 19357221, 19357222, 19357223, 19357224, 19357225, 19357226, 19357227, 19357229, 19357230, 19357231, 19357233, 19357234, 19357235, 19357236, 19357238, 19357240, 19357241, 19357242, 19357243, 19357244, 19357246, 19357249, 19357250, 19357251, 19357252, 19357253, 19357254, 19357255, 19357256, 19357257, 19357258, 19357259, 19357260, 19357261, 19357262, 19357263, 19357264, 19357265, 19357266, 19357267, 19357268, 19357269, 19357270, 19357271, 19357272, 19357273, 19357274, 19357275, 19357276, 19357277, 19357278, 19357279, 19357281, 19357282, 19357284, 19357288, 19357289, 19357290, 19357291, 19357293, 19357294, 19357295, 19357296, 19357297, 19357299, 19357300, 19357301, 19357302, 19357303, 19357304, 19357305, 19357306, 19357307, 19357308, 19357309, 19357310, 19357311, 19357312, 19357313, 19357314, 19357315, 19357316, 19357317, 19357318, 19357319, 19357320, 19357322, 19357323, 19357324, 19357325, 19357326, 19357327, 19357328, 19357330, 19357331, 19357332, 19357333, 19357335, 19357336, 19357337, 19357338, 19357339, 19357340, 19357341, 19357342, 19357343, 19357344, 19357345, 19357346, 19357347, 19357348, 19357349, 19357350, 19357351, 19357352, 19357353, 19357354, 19357355, 19357356, 19357357, 19357358, 19357359, 19357360, 19357361, 19357362, 19357363, 19357364, 19357365, 19357366, 19357367, 19357368, 19357369, 19357370, 19357371, 19357372, 19357373, 19357374, 19357375, 19357376, 19357377, 19357378, 19357379, 19357380, 19357381, 19357382, 19357383, 19357384, 19357385, 19357386, 19357387, 19357388, 19357389, 19357390, 19357391, 19357392, 19357394, 19357395, 19357396, 19357397, 19357398, 19357399, 19357400, 19357401, 19357402, 19357403, 19357404, 19357405, 19357406, 19357407, 19357410, 19357411, 19357412, 19357413, 19357414, 19357415, 19357416, 19357419, 19357420, 19357422, 19357425, 19357426, 19357427, 19357429, 19357431, 19357433, 19357434, 19357435, 19357436, 19357437, 19357438, 19357439, 19357440, 19357441, 19357444, 19357445, 19357446, 19357447, 19357448, 19357449, 19357450, 19357451, 19357452, 19357453, 19357454, 19357455, 19357456, 19357457, 19357458, 19357459, 19357460, 19357461, 19357462, 19357463, 19357464, 19357465, 19357466, 19357467, 19357468, 19357469, 19357470, 19357471, 19357472, 19357473, 19357474, 19357475, 19357476, 19357477, 19357478, 19357479, 19357480, 19357481, 19357482, 19357483, 19357485, 19357486, 19357487, 19357488, 19357489, 19357490, 19357491, 19357493, 19357494, 19357495, 19357496, 19357497, 19357498, 19357499, 19357501, 19357502, 19357503, 19357504, 19357505, 19357506, 19357507, 19357508, 19357510, 19357511, 19357512, 19357513, 19357515, 19357516, 19357517, 19357518, 19357519, 19357520, 19357522, 19357523, 19357524, 19357525, 19357526, 19357529, 19357530, 19357531, 19357532, 19357533, 19357534, 19357535, 19357536, 19357537, 19357538, 19357539, 19357540, 19357541, 19357542, 19357543, 19357544, 19357545, 19357546, 19357547, 19357548, 19357549, 19357550, 19357551, 19357552, 19357553, 19357554, 19357555, 19357556, 19357557, 19357558, 19357559, 19357560, 19357561, 19357562, 19357563, 19357564, 19357565, 19357566, 19357567, 19357568, 19357569, 19357570, 19357571, 19357572, 19357573, 19357574, 19357575, 19357576, 19357577, 19357578, 19357579, 19357580, 19357582, 19357583, 19357584, 19357586, 19357587, 19357588, 19357589, 19357590, 19357591, 19357592, 19357594, 19357595, 19357596, 19357597, 19357599, 19357600, 19357601, 19357602, 19357603, 19357604, 19357605, 19357606, 19357607, 19357608, 19357609, 19357610, 19357611, 19357612, 19357613, 19357614, 19357615, 19357616, 19357617, 19357618, 19357619, 19357620, 19357621, 19357622, 19357623, 19357624, 19357626, 19357627, 19357628, 19357629, 19357630, 19357631, 19357632, 19357633, 19357634, 19357635, 19357636, 19357637, 19357638, 19357639, 19357641, 19357642, 19357643, 19357644, 19357645, 19357646, 19357647, 19357648, 19357649, 19357650, 19357651, 19357652, 19357654, 19357656, 19357658, 19357660, 19357661, 19357663, 19357664, 19357665, 19357667, 19357670, 19357676, 19357679, 19357680, 19357681, 19357682, 19357683, 19357684, 19357685, 19357686, 19357687, 19357688, 19357689, 19357690, 19357691, 19357692, 19357693, 19357694, 19357695, 19357696, 19357697, 19357698, 19357699, 19357700, 19357701, 19357702, 19357703, 19357704, 19357705, 19357706, 19357707, 19357708, 19357709, 19357710, 19357711, 19357712, 19357713, 19357714, 19357715, 19357716, 19357717, 19357718, 19357719, 19357720, 19357721, 19357722, 19357723, 19357724, 19357725, 19357726, 19357727, 19357728, 19357729, 19357730, 19357731, 19357732, 19357733, 19357734, 19357735, 19357736, 19357737, 19357738, 19357739, 19357740, 19357741, 19357742, 19357743, 19357744, 19357745, 19357746, 19357747, 19357748, 19357749, 19357750, 19357751, 19357752, 19357753, 19357754, 19357755, 19357757, 19357758, 19357759, 19357760, 19357761, 19357762, 19357763, 19357764, 19357765, 19357766, 19357767, 19357768, 19357769, 19357770, 19357771, 19357772, 19357773, 19357774, 19357775, 19357776, 19357777, 19357778, 19357779, 19357780, 19357781, 19357782, 19357783, 19357784, 19357785, 19357786, 19357787, 19357788, 19357789, 19357790, 19357791, 19357792, 19357793, 19357794, 19357795, 19357796, 19357797, 19357798, 19357799, 19357800, 19357801, 19357802, 19357803, 19357804, 19357805, 19357806, 19357807, 19357808, 19357809, 19357810, 19357811, 19357812, 19357813, 19357814, 19357815, 19357816, 19357817, 19357818, 19357819, 19357820, 19357821, 19357822, 19357823, 19357824, 19357825, 19357826, 19357827, 19357828, 19357829, 19357831, 19357832, 19357833, 19357834, 19357835, 19357836, 19357837, 19357838, 19357839, 19357840, 19357841, 19357842, 19357844, 19357845, 19357846, 19357847, 19357848, 19357849, 19357850, 19357851, 19357852, 19357853, 19357854, 19357855, 19357856, 19357857, 19357858, 19357859, 19357860, 19357861, 19357862, 19357863, 19357864, 19357865, 19357866, 19357867, 19357868, 19357869, 19357870, 19357871, 19357872, 19357873, 19357874, 19357875, 19357876, 19357877, 19357878, 19357879, 19357880, 19357881, 19357882, 19357883, 19357884, 19357885, 19357886, 19357887, 19357888, 19357889, 19357890, 19357891, 19357892, 19357893, 19357894, 19357895, 19357896, 19357897, 19357898, 19357899, 19357900, 19357901, 19357902, 19357903, 19357904, 19357905, 19357906, 19357907, 19357908, 19357909, 19357910, 19357911, 19357912, 19357913, 19357914, 19357915, 19357916, 19357917, 19357918, 19357919, 19357920, 19357921, 19357922, 19357923, 19357924, 19357925, 19357926, 19357927, 19357928, 19357929, 19357930, 19357931, 19357932, 19357933, 19357934, 19357935, 19357936, 19357938, 19357940, 19357941, 19357942, 19357943, 19357944, 19357948, 19357949, 19357951, 19357952, 19357955, 19357956, 19357957, 19357960, 19357961, 19357962, 19357963, 19357964, 19357965, 19357968, 19357970, 19357973, 19357975, 19357977, 19357978, 19357979, 19357980, 19357982, 19357983, 19357984, 19357985, 19357986, 19357987, 19357988, 19357989, 19357990, 19357991, 19357992, 19357993, 19357994, 19357995, 19358011, 19358015, 19358022, 19358032, 19358034, 19358035, 19358040, 19358046, 19358048, 19358049, 19358050, 19358059, 19358064, 19358075, 19358076, 19358077, 19358084, 19358086, 19358088, 19358089, 19358090, 19358092, 19358093, 19358094, 19358095, 19358096, 19358097, 19358098, 19358100, 19358102, 19358105, 19358107, 19358108, 19358109, 19358111, 19358114, 19358115, 19358117, 19358118, 19358131, 19358132, 19358134, 19358142, 19358145, 19358146, 19358147, 19358150, 19358152, 19358153, 19358155, 19358157, 19358158, 19358159, 19358160, 19358161, 19358163, 19358165, 19358167, 19358168, 19358169, 19358170, 19358171, 19358173, 19358174, 19358175, 19358177, 19358178, 19358180, 19358181, 19358182, 19358184, 19358185, 19358186, 19358187, 19358188, 19358189, 19358190, 19358191, 19358192, 19358193, 19358194, 19358195, 19358197, 19358198, 19358199, 19358200, 19358201, 19358202, 19358203, 19358204, 19358205, 19358208, 19358210, 19358211, 19358212, 19358213, 19358214, 19358215, 19358216, 19358218, 19358219, 19358220, 19358221, 19358222, 19358224, 19358225, 19358226, 19358228, 19358229, 19358230, 19358231, 19358232, 19358233, 19358234, 19358235, 19358236, 19358237, 19358238, 19358239, 19358240, 19358241, 19358242, 19358243, 19358244, 19358245, 19358246, 19358247, 19358248, 19358249, 19358250, 19358251, 19358252, 19358253, 19358254, 19358255, 19358256, 19358257, 19358258, 19358259, 19358260, 19358261, 19358262, 19358263, 19358264, 19358265, 19358266, 19358267, 19358268, 19358270, 19358271, 19358272, 19358273, 19358274, 19358275, 19358277, 19358278, 19358279, 19358280, 19358281, 19358282, 19358283, 19358285, 19358286, 19358287, 19358288, 19358289, 19358290, 19358291, 19358293, 19358295, 19358296, 19358297, 19358298, 19358299, 19358301, 19358302, 19358303, 19358304, 19358305, 19358306, 19358308, 19358309, 19358310, 19358311, 19358312, 19358313, 19358314, 19358315, 19358316, 19358317, 19358318, 19358319, 19358320, 19358321, 19358322, 19358323, 19358324, 19358326, 19358328, 19358329, 19358330, 19358332, 19358333, 19358334, 19358335, 19358336, 19358337, 19358338, 19358340, 19358341, 19358342, 19358343, 19358344, 19358345, 19358346, 19358347, 19358348, 19358349, 19358350, 19358351, 19358352, 19358353, 19358354, 19358355, 19358356, 19358357, 19358358, 19358359, 19358360, 19358361, 19358362, 19358363, 19358364, 19358365, 19358366, 19358367, 19358368, 19358369, 19358370, 19358371, 19358372, 19358373, 19358374, 19358375, 19358376, 19358378, 19358379, 19358380, 19358381, 19358382, 19358383, 19358384, 19358385, 19358386, 19358387, 19358389, 19358390, 19358391, 19358392, 19358393, 19358394, 19358395, 19358396, 19358397, 19358398, 19358399, 19358400, 19358401, 19358402, 19358403, 19358404, 19358405, 19358406, 19358407, 19358408, 19358409, 19358410, 19358411, 19358413, 19358414, 19358416, 19358417, 19358418, 19358419, 19358420, 19358421, 19358422, 19358424, 19358425, 19358426, 19358427, 19358428, 19358429, 19358430, 19358431, 19358433, 19358434, 19358435, 19358436, 19358437, 19358438, 19358439, 19358440, 19358441, 19358442, 19358443, 19358444, 19358447, 19358448, 19358449, 19358451, 19358452, 19358453, 19358454, 19358455, 19358456, 19358457, 19358458, 19358459, 19358460, 19358461, 19358462, 19358464, 19358465, 19358466, 19358467, 19358470, 19358471, 19358472, 19358473, 19358474, 19358475, 19358476, 19358477, 19358478, 19358479, 19358480, 19358481, 19358482, 19358483, 19358484, 19358485, 19358486, 19358487, 19358488, 19358490, 19358491, 19358494, 19358495, 19358496, 19358497, 19358498, 19358499, 19358501, 19358502, 19358506, 19358507, 19358516, 19358517, 19358520, 19358521, 19358522, 19358523, 19358524, 19358526, 19358527, 19358529, 19358530, 19358533, 19358534, 19358537, 19358540, 19358541, 19358542, 19358543, 19358544, 19358545, 19358546, 19358547, 19358548, 19358549, 19358550, 19358552, 19358553, 19358554, 19358555, 19358556, 19358557, 19358561, 19358568, 19358570, 19358572, 19358573, 19358574, 19358576, 19358577, 19358578, 19358579, 19358580, 19358581, 19358582, 19358583, 19358585, 19358587, 19358589, 19358591, 19358592, 19358593, 19358594, 19358595, 19358596, 19358598, 19358601, 19358606, 19358607, 19358610, 19358611, 19358612, 19358613, 19358615, 19358616, 19358617, 19358618, 19358619, 19358620, 19358622, 19358623, 19358624, 19358626, 19358627, 19358628, 19358629, 19358630, 19358632, 19358633, 19358634, 19358635, 19358636, 19358637, 19358638, 19358639, 19358642, 19358643, 19358644, 19358645, 19358647, 19358648, 19358649, 19358650, 19358651, 19358652, 19358653, 19358654, 19358655, 19358656, 19358657, 19358658, 19358659, 19358660, 19358661, 19358662, 19358663, 19358664, 19358665, 19358666, 19358667, 19358668, 19358669, 19358670, 19358671, 19358672, 19358673, 19358674, 19358677, 19358678, 19358679, 19358680, 19358681, 19358682, 19358683, 19358684, 19358685, 19358686, 19358687, 19358688, 19358689, 19358690, 19358691, 19358692, 19358693, 19358694, 19358695, 19358696, 19358697, 19358698, 19358699, 19358700, 19358701, 19358702, 19358703, 19358705, 19358706, 19358707, 19358708, 19358709, 19358710, 19358712, 19358713, 19358716, 19358717, 19358723, 19358724, 19358741, 19358742, 19358743, 19358744, 19358745, 19358746, 19358747, 19358748, 19358749, 19358751, 19358752, 19358753, 19358754, 19358755, 19358756, 19358757, 19358758, 19358759, 19358760, 19358761, 19358762, 19358763, 19358764, 19358765, 19358766, 19358767, 19358768, 19358769, 19358770, 19358771, 19358772, 19358773, 19358774, 19358775, 19358776, 19358777, 19358778, 19358779, 19358780, 19358781, 19358782, 19358783, 19358784, 19358785, 19358786, 19358787, 19358788, 19358789, 19358790, 19358791, 19358792, 19358794, 19358795, 19358796, 19358798, 19358799, 19358800, 19358801, 19358802, 19358803, 19358804, 19358805, 19358806, 19358807, 19358808, 19358809, 19358810, 19358811, 19358812, 19358813, 19358814, 19358815, 19358816, 19358818, 19358819, 19358820, 19358821, 19358822, 19358823, 19358824, 19358825, 19358826, 19358827, 19358828, 19358830, 19358831, 19358832, 19358834, 19358835, 19358837, 19358838, 19358839, 19358840, 19358841, 19358842, 19358843, 19358844, 19358845, 19358846, 19358848, 19358849, 19358850, 19358851, 19358852, 19358853, 19358854, 19358855, 19358856, 19358857, 19358858, 19358859, 19358860, 19358861, 19358862, 19358864, 19358865, 19358866, 19358868, 19358869, 19358870, 19358871, 19358872, 19358873, 19358874, 19358875, 19358878, 19358879, 19358882, 19358884, 19358885, 19358889, 19358890, 19358891, 19358892, 19358893, 19358894, 19358895, 19358896, 19358897, 19358898, 19358899, 19358900, 19358901, 19358902, 19358903, 19358904, 19358905, 19358906, 19358907, 19358908, 19358909, 19358910, 19358911, 19358912, 19358913, 19358915, 19358916, 19358917, 19358918, 19358919, 19358920, 19358921, 19358922, 19358923, 19358924, 19358925, 19358926, 19358927, 19358928, 19358929, 19358930, 19358931, 19358932, 19358933, 19358934, 19358935, 19358936, 19358937, 19358938, 19358940, 19358941, 19358942, 19358943, 19358944, 19358945, 19358946, 19358947, 19358948, 19358950, 19358951, 19358952, 19358953, 19358954, 19358955, 19358956, 19358957, 19358958, 19358959, 19358960, 19358961, 19358962, 19358963, 19358964, 19358965, 19358966, 19358967, 19358968, 19358969, 19358970, 19358971, 19358972, 19358973, 19358975, 19358976, 19358977, 19358978, 19358979, 19358980, 19358981, 19358982, 19358983, 19358984, 19358985, 19358986, 19358987, 19358988, 19358989, 19358990, 19358992, 19358993, 19358994, 19358996, 19358997, 19358998, 19358999, 19359000, 19359001, 19359002, 19359003, 19359004, 19359005, 19359006, 19359007, 19359008, 19359009, 19359011, 19359012, 19359013, 19359014, 19359015, 19359016, 19359017, 19359018, 19359019, 19359020, 19359021, 19359022, 19359024, 19359025, 19359026, 19359027, 19359031, 19359032, 19359033, 19359034, 19359035, 19359036, 19359037, 19359038, 19359039, 19359041, 19359042, 19359043, 19359044, 19359045, 19359046, 19359047, 19359048, 19359049, 19359050, 19359051, 19359052, 19359053, 19359054, 19359055, 19359056, 19359057, 19359058, 19359059, 19359060, 19359061, 19359062, 19359063, 19359064, 19359065, 19359066, 19359067, 19359068, 19359069, 19359071, 19359072, 19359073, 19359074, 19359075, 19359076, 19359077, 19359078, 19359079, 19359080, 19359082, 19359084, 19359085, 19359086, 19359087, 19359088, 19359089, 19359090, 19359091, 19359092, 19359093, 19359094, 19359095, 19359096, 19359097, 19359098, 19359103, 19359104, 19359106, 19359107, 19359108, 19359109, 19359110, 19359111, 19359112, 19359113, 19359114, 19359115, 19359116, 19359117, 19359118, 19359119, 19359124, 19359125, 19359126, 19359127, 19359128, 19359129, 19359131, 19359132, 19359133, 19359134, 19359144, 19359172, 19359173, 19359181, 19359185, 19359188, 19359192, 19359193, 19359199, 19359200, 19359204, 19359207, 19359218, 19359223, 19359226, 19359227, 19359228, 19359231, 19359232, 19359233, 19359234, 19359235, 19359236, 19359237, 19359238, 19359239, 19359240, 19359241, 19359242, 19359244, 19359245, 19359246, 19359247, 19359334, 19359335, 19359336, 19359337, 19359338, 19359339, 19359340, 19359341, 19359342, 19359343, 19359345, 19359348, 19359349, 19359350, 19359351, 19359352, 19359353, 15551627, 15551779, 15552083, 15552091, 15551891, 15551947, 15552083, 15552800, 15553633, 15554610, 15554979, 15555464, 15555936, 15556183, 15556650, 15557414, 15557591, 15557892, 15558169, 15558271, 15558352, 15558484, 15558557, 15558786, 15559063, 15559236, 15559265, 15559469, 15559997, 15562783, 15562963, 15563482, 15563793, 15564485, 15565014, 15565055, 15565056, 15565071, 15565075, 15565078, 15565137, 15565141, 15565642, 15566009, 15566208, 15566452, 15566624, 15566627, 15566631, 15551980, 15552147, 15552149, 15552192, 15552855, 15554391, 15554416, 15554816, 15554854, 15555017, 15555242, 15555393, 15555576, 15555763, 15555939, 15556169, 15556538, 15557751, 15557758, 15557765, 15557954, 15557989, 15558484, 15558546, 15559265, 15559398, 15559469, 15559620, 15559858, 15559933, 15559943, 15560287, 15560387, 15562919, 15562940, 15564349, 15564376, 15565055, 15565056, 15565071, 15566208, 15566605, 16212298, 16212825, 16213074, 16213187, 16213536, 16213548, 16213741, 16214352, 16214635, 16214711, 16214773, 16214786, 16214846, 16214956, 16215218, 16215274, 16215304, 16215338, 16215355, 16215405, 16215549, 16215671, 16215822, 16215835, 16215835, 16216274, 16216285, 16217357, 16217732, 16217733, 16217800, 16217859, 16217939, 16218021, 16218023, 16218267, 16218513, 16218826, 16219763, 16220183, 16220258, 16220303, 16221219, 16222006, 16222047, 16222086, 16222116, 16222164, 16222346, 16223131, 16223866, 16223925, 16224430, 16224535, 16224730, 16224743, 16224749, 16224781, 16224808, 16224905, 16224924, 16225234, 15560287, 15564349, 15566605, 15566624, 15566631, 16212543, 16213536, 16213587, 16214352, 16214718, 16214786, 16214846, 16214956, 16215460, 16215822, 16217359, 16218023, 16218226, 16218233, 16218559, 16219039, 16220150, 16220183, 16222346, 16225234, 16225666, 16225679, 16225763, 16225917, 16226859, 16226859, 16227272, 15554610, 16217859, 16225763, 19352316, 19352317, 19352318, 19352332, 19352373, 19352664, 19352914, 19352966, 19353053, 19353091, 19353092, 19353112, 19353120, 19353173, 19353173, 19353220, 19353220, 19353261, 19353261, 19353447, 19353646, 19353647, 19353669, 19353690, 19353690, 19353713, 19353746, 19353748, 19353791, 19353802, 19353819, 19353836, 19353851, 19353860, 19353862, 19353864, 19353875, 19353892, 19353900, 19353902, 19353917, 19353932, 19353933, 19353934, 19353995, 19354001, 19354009, 19354032, 19354055, 19354056, 19354072, 19354076, 19354085, 19354105, 19354139, 19354235, 19354266, 19354293, 19354300, 19354316, 19354317, 19354322, 19354328, 19354377, 19354379, 19354382, 19354384, 19354432, 19354434, 19354448, 19354460, 19354545, 19354556, 19354562, 19354572, 19354581, 19354593, 19354608, 19354621, 19354630, 19354665, 19354668, 19354671, 19354683, 19354684, 19354753, 19354761, 19354784, 19354790, 19354794, 19354804, 19354834, 19354865, 19354869, 19354872, 19354887, 19354905, 19354997, 19355011, 19355043, 19355043, 19355091, 19355198, 19355208, 19355212, 19355248, 19355264, 19355269, 19355277, 19355296, 19355302, 19355345, 19355451, 19355460, 19355471, 19355706, 19355731, 19355773, 19355782, 19355793, 19355931, 19355932, 19355960, 19356003, 19356013, 19356030, 19356033, 19356033, 19356115, 19356150, 19356173, 19356520, 19356520, 19356526, 19356636, 19356636, 19356665, 19356988, 19357188, 19352373, 19352664, 19353173, 19353220, 19353261, 19353791, 19353860, 19353875, 19353932, 19354076, 19354105, 19354293, 19354317, 19354377, 19354384, 19354683, 19355212, 19355451, 19355471, 19355793, 19356520, 16262527. |
Recalling Firm/ Manufacturer |
Hospira Inc. 275 N Field Dr Lake Forest IL 60045-2579
|
| For Additional Information Contact | Chris Eustace 224-212-2000 |
Manufacturer Reason for Recall | Screw rotation error occurring when the pump detects a mismatch between the number of steps given to the motor driving delivery and the number of steps the motor actually took. When this malfunction occurs, the pump will alarm and infusion stops. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Hospira will contract with Stericycle, Inc. in Indianapolis, IN to send letter and a reply form to Hospira, Inc. Receipt of the letters will be confirmed by UPS tracking. Customers will be requested to return a reply form to Stericycle indicating the number of pumps on hand and that they have received the letter and notified their users.
In addition, letters will be sent by email to ECRI, NRAC, Noblis and ISMP for notification of their members and posting on their websites. The letter will also be posted to the Hospira website.
Letters will be sent to Hospira customers of record who purchased the devices directly from Hospira. Customers will be directed to notify their consignees if they further distributed the product and confirm they have done so.
Telephone attempts will be made to contact customers that do not respond in order to confirm receipt of the letter. Final effectiveness will be done as the devices are upgraded with new plastic motors. |
| Quantity in Commerce | 8,794 |
| Distribution | United States (Nationwide) Distribution including Puerto Rico and American Virgin Islands. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MEA
|
|
|
|