• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall M/DN Intramedullary Fixation

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall M/DN Intramedullary Fixationsee related information
Date Initiated by FirmOctober 11, 2013
Date PostedFebruary 20, 2014
Recall Status1 Terminated 3 on April 14, 2017
Recall NumberZ-1009-2014
Recall Event ID 66500
510(K)NumberK965098 
Product Classification Rod, fixation, intramedullary and accessories - Product Code HSB
ProductINTERLOCKING IM SCREW SML, MED and LRG; M/DN TIBIAL I/M NAIL 6 MM to 15 MM. Some items starting with codes 0022530, 0022531, 002532, 0022533, 0022534
Code Information 60179289 60196075 60208629 60272379 60279637 60367538 60438316 60575719 60597680 60620340 60714150 60735642 60864016 60937504 61031036 61040654 61120277 61133164 61186040 61325064 61446262 61551621 60129673 60177185 60187385 60363817 60387525 60569948 60602446 60620341 60630482 60711126 60714151 60735643 60808002 60819266 60864018 60864019 60876889 60876890 60956565 61104687 61112613 61133165 61266235 61422229 61458906 61489975 60690743 60711143 60714152 60728336 60735644 60747675 60757046 60789824 60808003 60817590 60838120 60838121 60876892 60876893 60937495 60937496 61031038 61112614 61112651 61158629 61161384 60466661 60615939 60735645 60817572 60847807 61120278 61221431 61263174 61333468 60152966 60196077 60196079 60750994 60806391 60847809 60979630 61226089 61263175 61408290 61504079 60753364 60806396 60817586 60829585 60876926 60931129 61331079 60120568 60140390 60150071 60152967 60165627 60177186 60185172 60190843 60195440 60195442 60243126 60253766 60279638 60279639 60279640 60338431 60349085 60387526 60393389 60422609 60438317 60444815 60472686 60528779 60569949 60602447 60602448 60620342 60653343 60690744 60714153 60735646 60735647 60735648 60811023 60819429 60819430 60864026 60864029 60876928 60937499 60937501 60979623 60985453 61079807 61079808 61086366 61100812 61104688 61140198 61143244 61150035 61161407 61173560 61194688 61231196 61244342 61280079 61293775 61303278 61319424 61383085 61428464 61453389 61516672 61564740 61613173 61657631 61709056 61709069 61763484 60103523 60127531 60130218 60140391 60152968 60172506 60179290 60201128 60201130 60229987 60242009 60265708 60279641 60323303 60323305 60338432 60349086 60387527 60393390 60407650 60438318 60444317 60491698 60569950 60602207 60620343 60653345 60653346 60690745 60714154 60735649 60735650 60808004 60819270 60819273 60864032 60864034 60956555 60956558 60956566 60956567 61064852 61075702 61086367 61100813 61104689 61120279 61143245 61158631 61164107 61173035 61173562 61173564 61173565 61512411 61729530 61763487 60349309 60602449 60674160 60690747 60714155 60735652 60757049 60808006 60814361 60829589 60829590 60876895 60876897 60937502 60937503 60966581 61067793 61112652 61158633 61161356 61190345 60615916 60735653 60745777 60806394 60819279 60876929 60926469 61150072 61150073 61202901 61224266 61231239 61238805 61349139 61356914 61383087 61428466 61489976 61496668 61516673 60109571 60367539 60407652 60597682 60653348 60735654 60814200 60847818 60857999 61120979 61140199 61226092 61263176 61267707 61434739 62326518 60117045 60135211 60145198 60157582 60177187 60185173 60186867 60208631 60242010 60253767 60253768 60279642 60279643 60279646 60338433 60338434 60349087 60349088 60461528 60461529 60487153 60517097 60517098 60537305 60537306 60550321 60569954 60572975 60575720 60602450 60620364 60653349 60674184 60735655 60735657 60811024 60819285 60819286 60819287 60876931 60876934 60937497 60937498 60979626 60985463 61059954 61096522 61100814 61104691 61131925 61256012 61263177 61267708 61475183 61484475 61547943 61555829 61564741 61597005 61664932 61729533 60097908 60099562 60107380 60114655 60115159 60117047 60123301 60124858 60130219 60135212 60135213 60142917 60145199 60145201 60152969 60154189 60157583 60165630 60167143 60172507 60177188 60177190 60186868 60186869 60189042 60193856 60196080 60196081 60196082 60201133 60201134 60218842 60218843 60218844 60229988 60239641 60242011 60242012 60259288 60259290 60265709 60265710 60265711 60279652 60279654 60279655 60279656 60309136 60309138 60323307 60323308 60323309 60338435 60338436 60338437 60338438 60349089 60349090 60349091 60349092 60380449 60407654 60407655 60407657 60407658 60429161 60438319 60438320 60444318 60452043 60452044 60461532 60479128 60491712 60505965 60505966 60513231 60513232 60537308 60538409 60538410 60550323 60550326 60569957 60569959 60575721 60602453 60602457 60602458 60602459 60602460 60602461 60620366 60620368 60653351 60653352 60653353 60653354 60674187 60690749 60714157 60714158 60714159 60714160 60735658 60735659 60735660 60735661 60808007 60808008 60808009 60814245 60814262 60814263 60814264 60819314 60819315 60876937 60876939 60876940 60876942 60876943 60876945 60876947 60876949 60937505 60956560 60956561 60956562 60956564 60958484 60958485 60958486 61024749 61024750 61030971 61030972 61030975 61040670 61040676 61040679 61040680 61064853 61071423 61071424 61075703 61075704 61079809 61084122 61086368 61089038 61131926 61147784 61158637 61158639 61161359 61186042 61190342 61199889 61221433 61221435 61267709 61271266 61280080 61286468 61303280 61306507 61358789 61383089 61388905 61414386 61421224 61428467 61434740 61434741 61446318 61446319 61453390 61464939 61464940 61464941 61509880 61512412 61516674 61516675 61621141 61657633 61676193 61676194 61677324 61677325 61729536 61729537 61729539 61729541 61753006 61763494 62262060 60099564 60103534 60140392 60142919 60150074 60152970 60165632 60165634 60172509 60189043 60208633 60208635 60208637 60229989 60242013 60265712 60265713 60279666 60309142 60323313 60349094 60349310 60367541 60380451 60391592 60391593 60429162 60444816 60461540 60487154 60513234 60513236 60531976 60531977 60537309 60537310 60550327 60550328 60550329 60569964 60569965 60575722 60575723 60575724 60620375 60653360 60653361 60711163 60711166 60714163 60714164 60735664 60757053 60808012 60808013 60808014 60814575 60814576 60876951 60876952 60876953 60876954 60876955 60876956 60876958 60931196 60931197 60966628 60979631 60985474 61023558 61029738 61031019 61086370 61090244 61100816 61104692 61120281 61120980 61125214 61158644 61161360 61178963 61186044 61190333 61303282 61383090 61421226 61421227 61446320 61489977 61538687 61547944 61613175 60272380 60615915 60808015 60814290 60873118 61173572 61199892 61202909 61202911 61319442 61383091 61395917 60103527 60127533 60137566 60154190 60175709 60186870 60210701 60243128 60323314 60358123 60380455 60407660 60438322 60472687 60513238 60537311 60550330 60602463 60620379 60653362 60757056 60817574 60873117 60876960 60876961 61024751 61031018 61100818 61100819 61100820 61249861 61267710 61277976 61421228 61564744 61729555 60608665 60620388 60714167 60747679 60757060 60806411 60847844 60847846 60931187 60931188 61120282 61140203 61158647 61211627 61226093 61242316 61263178 61303286 61484476 61547945 61657674 60145202 60165640 60181943 60190845 60208183 60218847 60243125 60245720 60253771 60253772 60253773 60253774 60253775 60265715 60265717 60265719 60265720 60265721 60531984 60531986 60569966 60569968 60602483 60620389 60653366 60690754 60714168 60735667 60735668 60806413 60806415 60806416 60811025 60819458 60876963 60876964 60876965 60876966 60876967 60876969 61003990 61003993 61009887 61086372 61120284 61153536 61173585 61190869 61190870 61199895 61199896 61202924 61215206 61215210 61231270 61288929 61288931 61307063 61307066 61329362 61329363 61367090 61371404 61379424 61383094 61395919 61395920 61489980 61489981 61504083 61565268 60097909 60097910 60101715 60107381 60107382 60107383 60107384 60112820 60114656 60115160 60115161 60117048 60118881 60119085 60123305 60123306 60123307 60124860 60129675 60129677 60132559 60132560 60132561 60135215 60135216 60135217 60140395 60142921 60142922 60142923 60145203 60145206 60150076 60152972 60154191 60154192 60154193 60154194 60155334 60155335 60157584 60165642 60165645 60165646 60167145 60167146 60167148 60172512 60175712 60175716 60175718 60177192 60177193 60185174 60186871 60186872 60186873 60186874 60187386 60189044 60189045 60193848 60196084 60196087 60196089 60202490 60208194 60208639 60210692 60218848 60218849 60218850 60218851 60218852 60218853 60218854 60229990 60229991 60229992 60229993 60229994 60229995 60242014 60242015 60242016 60243120 60243122 60245721 60245722 60253776 60253777 60259293 60259294 60259295 60259296 60265722 60281575 60281576 60281577 60281579 60281580 60281581 60281582 60281583 60281584 60281585 60288491 60288492 60288493 60288494 60288495 60288497 60288498 60310566 60310567 60310568 60310569 60310570 60310571 60310572 60310573 60310575 60310576 60310577 60323344 60323345 60323346 60323347 60323348 60323349 60323350 60338522 60338523 60338524 60338526 60338528 60338529 60338530 60338531 60358125 60358126 60367546 60367548 60367549 60367550 60367552 60367554 60380463 60380466 60385465 60385466 60393396 60393397 60393398 60393399 60393401 60399618 60407665 60407666 60407667 60410615 60422631 60422633 60425158 60425159 60438324 60438326 60438327 60438328 60438329 60444319 60444817 60444818 60452045 60452046 60452047 60472690 60472691 60472692 60479130 60479131 60479132 60491721 60491722 60505971 60505972 60513239 60513240 60513241 60513242 60513243 60531988 60531989 60537312 60537314 60537340 60537341 60550333 60550334 60550335 60550337 60550339 60550340 60550342 60550344 60572976 60575725 60602484 60602486 60602489 60602491 60602493 60602495 60602496 60615930 60615931 60620390 60620391 60620392 60620393 60653368 60653369 60653370 60653371 60653372 60653374 60653375 60653376 60674198 60674201 60674204 60674205 60674208 60674209 60690755 60690756 60690757 60690758 60690759 60714169 60714174 60714175 60714176 60714177 60735669 60735670 60735671 60735673 60757061 60757062 60757063 60757064 60757065 60757066 60757068 60819295 60819301 60819302 60819303 60819304 60819305 60819306 60819307 60819308 60819310 60876970 60876971 60876972 60876973 60876974 60876975 60876976 60876977 60876978 60876979 60915888 60915890 60915891 60915892 60915893 60915894 60915895 60915896 60915897 60931138 60931139 60931140 60931141 60931143 60931144 60958479 60958480 60966537 60966538 60966539 60966540 60966541 60974167 60974168 60974169 60974170 60974171 60979614 60979615 60979616 60985473 61020125 61020126 61020127 61020128 61020130 61024660 61024661 61031012 61031013 61031014 61041605 61041606 61041607 61041608 61059949 61071428 61075705 61075706 61075707 61075708 61075709 61079813 61086375 61089040 61096528 61096529 61096530 61096531 61096532 61104694 61104695 61112654 61120286 61125216 61125218 61131927 61131928 61133872 61140204 61140206 61140207 61140208 61140209 61143247 61143248 61143249 61147786 61148360 61148361 61164109 61186047 61186050 61186051 61186052 61186053 61211628 61211629 61211630 61211631 61242319 61242320 61242321 61249863 61263179 61267711 61271268 61293776 61303302 61306508 61306509 61387093 61388907 61388910 61396177 61428468 61428469 61434742 61439807 61439809 61439810 61446322 61446323 61446324 61446325 61464946 61464948 61489982 61496683 61496684 61504084 61509883 61512415 61542363 61542638 61551622 61551623 61551624 61555830 61597012 61597014 61693599 61693602 61693603 61727562 61727564 61727565 61727566 61727567 61727568 61727571 61729561 61729562 61753009 61753011 61753012 61763527 61763529 61763531 62262071 62262075 62285850 60099565 60101720 60103535 60105502 60120571 60124862 60135219 60135220 60140396 60142926 60152973 60154199 60154202 60157585 60165647 60165648 60167150 60172514 60177194 60185175 60186876 60189046 60193849 60196091 60196093 60208196 60218856 60218857 60218858 60229997 60229998 60229999 60230001 60239642 60242195 60259301 60259306 60293426 60293427 60293428 60293429 60293430 60293431 60293432 60293433 60310583 60310584 60323353 60323354 60338543 60338544 60338546 60349312 60358128 60358129 60367561 60367563 60367566 60380471 60380473 60380475 60387529 60393404 60399620 60407670 60407671 60422636 60429164 60438330 60438331 60438332 60444821 60452049 60461546 60461550 60461552 60461553 60497841 60497842 60497843 60497845 60513244 60513245 60531996 60532000 60532003 60537346 60550346 60550350 60550351 60550354 60569971 60569979 60569987 60569988 60569989 60575727 60575728 60575729 60602508 60602509 60602510 60620398 60620399 60620401 60620402 60653382 60653383 60653384 60653386 60653389 60653392 60674214 60690762 60690763 60714188 60714189 60714191 60714192 60714193 60735677 60735678 60757075 60757077 60757078 60814581 60814582 60814583 60814584 60814585 60814586 60859220 60859221 60859222 60859223 60859224 60859225 60876990 60876991 60876992 60876993 60876995 60876996 60876997 60876998 60876999 60979608 60979609 60979610 61003955 61003957 61003963 61003968 61009890 61029742 61029743 61029745 61041726 61041732 61041734 61041735 61075714 61075715 61084127 61090245 61104697 61112656 61112657 61120287 61120288 61125220 61131929 61133878 61148362 61148363 61175291 61178948 61178949 61190329 61211632 61221443 61221445 61226094 61242324 61242325 61242326 61263182 61266234 61267714 61271262 61271263 61286475 61293778 61303303 61303304 61339508 61358791 61358792 61379426 61379428 61414389 61421230 61421231 61421232 61439811 61464951 61484478 61489986 61504090 61504092 61512416 61516684 61538692 61564756 61613183 61613184 61621166 61652458 61664946 61664947 61729563 61729564 60230002 60597278 60757079 60806418 60811005 60908910 61112658 61140210 61215269 61258937 61263183 61277977 61504093 60115162 60118882 60127536 60132565 60152974 60165649 60186877 60196095 60208202 60230003 60265726 60310585 60323355 60338547 60358130 60380478 60393405 60407674 60422637 60438333 60452051 60472696 60513246 60550355 60602512 60653398 60711125 60714195 60819457 60858001 60877001 60926467 60926468 61059943 61112659 61120289 61140211 61158651 61211635 61226096 61242328 61263184 61280081 61340070 61420295 61446326 61530538 61555831 61664949 62306310 60152975 60177196 60193850 60218859 60218860 60253781 60253783 60253784 60253785 60253786 60253788 60513248 60532007 60532008 60532009 60569995 60575730 60602523 60620404 60690766 60714201 60735680 60789987 60806420 60811009 60811026 60873120 60877013 60877016 60877017 60877018 60877019 60877020 60956539 61066624 61066625 61084129 61153537 61173614 61190875 61190876 61215218 61231291 61238868 61275395 61275396 61288944 61307068 61307069 61319467 61319468 61371406 61383097 61464953 61484480 61484481 60097911 60097912 60097913 60097914 60107385 60107386 60107387 60109572 60109573 60115163 60115164 60115165 60117050 60117051 60118883 60119086 60123310 60124863 60124864 60127537 60129679 60129681 60130221 60130222 60132566 60132567 60132568 60135221 60135222 60135223 60140397 60142927 60142928 60142929 60142930 60145208 60145210 60150079 60152976 60154205 60154209 60154211 60154214 60155337 60155338 60157586 60159608 60159609 60165650 60165651 60165652 60165653 60167155 60167156 60167157 60172517 60175721 60175724 60175725 60175728 60177197 60177198 60181944 60186878 60186879 60186880 60186881 60187387 60187389 60189047 60193851 60193852 60196121 60196122 60196124 60196126 60196127 60201137 60201138 60202491 60202492 60218861 60218862 60218863 60218864 60218865 60230004 60230005 60230007 60230008 60230009 60230010 60230011 60230012 60239234 60242018 60242019 60242020 60242021 60243129 60245723 60245724 60253789 60259311 60259312 60259314 60259322 60259325 60265727 60265728 60265730 60265731 60265732 60265733 60281588 60281589 60281590 60281591 60281592 60281593 60281594 60281595 60281596 60281597 60288508 60288509 60288510 60288511 60310588 60310589 60310590 60310591 60310592 60310593 60310594 60310595 60310596 60310597 60310598 60310599 60323356 60323357 60323358 60323359 60323360 60323361 60323362 60323363 60323364 60338549 60338550 60338551 60338553 60338557 60338559 60358133 60358135 60358136 60358138 60367573 60367574 60367575 60367576 60380482 60380484 60380485 60380486 60385474 60385479 60393410 60393411 60393412 60393413 60393415 60393418 60393419 60399626 60407676 60407678 60407679 60407680 60410617 60410618 60410619 60422639 60422640 60422641 60425162 60425163 60438335 60438336 60438337 60438338 60438339 60444320 60444822 60452053 60452055 60452056 60452057 60452058 60461554 60461555 60472697 60472699 60479136 60479137 60479139 60484263 60484264 60484265 60491730 60491731 60491732 60505980 60505982 60505983 60513261 60513262 60513264 60513265 60513266 60513267 60532011 60537351 60538411 60538412 60550356 60550357 60550358 60550359 60550361 60550362 60550364 60550366 60569996 60569997 60572977 60575731 60602526 60602529 60602530 60602532 60602534 60602535 60602538 60615923 60615924 60615926 60615927 60620405 60620406 60620407 60620408 60653404 60653408 60653409 60653411 60653412 60653413 60674218 60674219 60674221 60674222 60674228 60674229 60691437 60691438 60691439 60691441 60714202 60714203 60714204 60714205 60714206 60714207 60714209 60714210 60714211 60714212 60735681 60735682 60735683 60735684 60735685 60735686 60735687 60735689 60735690 60757084 60757085 60757086 60757087 60757088 60757089 60808018 60808019 60808021 60808022 60819316 60819317 60819318 60819321 60819323 60819324 60819325 60873116 60877024 60877025 60877027 60877028 60877029 60877030 60877031 60877032 60877865 60877866 60877868 60877869 60925017 60925018 60925019 60925020 60925021 60925023 60925024 60925025 60925026 60937493 60950663 60950664 60950665 60966542 60966543 60966544 60966545 60966546 60966548 60966549 60966550 60966551 60966571 60974158 60974159 60974161 60979596 60979597 60979598 60979599 60985478 61024740 61024743 61024744 61040697 61040701 61040704 61059934 61059935 61071450 61071453 61075717 61075720 61084130 61086377 61086378 61089041 61091740 61096539 61096540 61096542 61096546 61096559 61100823 61100824 61100825 61104701 61104702 61109132 61109133 61109134 61117873 61117874 61117876 61131930 61133880 61143250 61143251 61143252 61143254 61147788 61158655 61158656 61161388 61171486 61178957 61178958 61186056 61186057 61186058 61186059 61211638 61242329 61242332 61271270 61271271 61286472 61306511 61306512 61306513 61358795 61358796 61358797 61383098 61383099 61387095 61388912 61388913 61388916 61388917 61396180 61396181 61414390 61421235 61421240 61434743 61439814 61446328 61446329 61446331 61446332 61464955 61464956 61464957 61464958 61489988 61496693 61496694 61496695 61512418 61516686 61516687 61523817 61523818 61530541 61542365 61542639 61542640 61551626 61551628 61559559 61597017 61597018 61621168 61664955 61664956 61677339 61677347 61677413 61677414 61693597 61709129 61709138 61709142 61709144 61729570 61729572 61729573 61729575 61729577 61729578 61763555 61763558 62285855 60090461 60097924 60099567 60099568 60103536 60127538 60130223 60132570 60135225 60135226 60140398 60142931 60152977 60154216 60154217 60155339 60165655 60165656 60167162 60172520 60177199 60186883 60189048 60195443 60195444 60208208 60210693 60218866 60218867 60230013 60230014 60253791 60253792 60265736 60265737 60288514 60288515 60288516 60310610 60310611 60310612 60323366 60323367 60323368 60338568 60338569 60338572 60358140 60358141 60367583 60367584 60380487 60385484 60393423 60399628 60407683 60407684 60422645 60422646 60429166 60431518 60451201 60461559 60461563 60497848 60497849 60497851 60497852 60513269 60513272 60532015 60532016 60532017 60537353 60550371 60550373 60550374 60570001 60570002 60570003 60575732 60575733 60575734 60575735 60602570 60602571 60620414 60620415 60620417 60630487 60653418 60653419 60653420 60653421 60653422 60691444 60691445 60691446 60714217 60714218 60714219 60714220 60735693 60735694 60757082 60757096 60757097 60817592 60817594 60817595 60817596 60859227 60859228 60877057 60877061 60877062 60877063 60877064 60877065 60877066 60877067 60925645 60925646 60925647 60925648 60925649 61009892 61024745 61024746 61024747 61024748 61029734 61029735 61038477 61074472 61075722 61090254 61104703 61104704 61112666 61112667 61120290 61120291 61143255 61148364 61158658 61190341 61211643 61215270 61221449 61249866 61263186 61267718 61271265 61280084 61303305 61306517 61358800 61379430 61383102 61414391 61421242 61421243 61446333 61475187 61484483 61489989 61504096 61516688 61516689 61547946 61567420 61613185 61621171 61657690 61665007 61665010 61677623 61729582 60266019 60615910 60789893 60813552 60847848 60858006 61263188 61333472 61479004 61597020 60152978 60175754 60201140 60202493 60210694 60210695 60338573 60358143 60367585 60380490 60407685 60422647 60438344 60472703 60513277 60575736 60602573 60620419 60653423 60691447 60714221 60735695 60757099 60819313 60869320 60877317 60877318 60943957 60943958 61003953 61120292 61125225 61140212 61158660 61211646 61226097 61242335 61266231 61267725 61340071 61428471 61551629 61597021 61665019 60064860 60181945 60190846 60242201 60253793 60288517 60323369 60338574 60385490 60461564 60461565 60513279 60532023 60532025 60570005 60572978 60575737 60620424 60735697 60757102 60811027 60819282 60819283 60819284 60877322 61009691 61009692 61067801 61091741 61120294 61133885 61231198 61242336 61367028 61383103 61475188 61489990 61729583 61752997 60097915 60097916 60097917 60097918 60101717 60107388 60107389 60109574 60109575 60112821 60114658 60115166 60115167 60117053 60117054 60118885 60120573 60123312 60123313 60124865 60124866 60127539 60129682 60132571 60135227 60135228 60142932 60142933 60145212 60145213 60150080 60152979 60154218 60154219 60154220 60154221 60155340 60157587 60165657 60165658 60165659 60167167 60167168 60167169 60169751 60172522 60175756 60177201 60177202 60186884 60186885 60187390 60187391 60187392 60189055 60193853 60195445 60195446 60201142 60202494 60202495 60208213 60208641 60208642 60218868 60218869 60218871 60218872 60230015 60230016 60230017 60230018 60230019 60239235 60242024 60242025 60242202 60242203 60242204 60245725 60245726 60253794 60253796 60253797 60253798 60265738 60265739 60265740 60265741 60281601 60281603 60281604 60288520 60288521 60304023 60304024 60304026 60304027 60304028 60304029 60310613 60310615 60323370 60323371 60323372 60323373 60323374 60323375 60338575 60338577 60338578 60338579 60338580 60338581 60358144 60358145 60358146 60367588 60367590 60367591 60367592 60380491 60380492 60380493 60385491 60385495 60393424 60393425 60393426 60393427 60393429 60399630 60407686 60407687 60407688 60410621 60422648 60422649 60425165 60425166 60431519 60431521 60431522 60444321 60444825 60452060 60452061 60452062 60455030 60461566 60461567 60479140 60479141 60479142 60484266 60484267 60484268 60491734 60491735 60505989 60505990 60505993 60505994 60505995 60505997 60505998 60532026 60537355 60537356 60550378 60550379 60550380 60550382 60550383 60550384 60550385 60550386 60550388 60575738 60602575 60602576 60602577 60602581 60602584 60615934 60615935 60620429 60620431 60620433 60630488 60653426 60653427 60653428 60653430 60653431 60653432 60653433 60653434 60674230 60674231 60674232 60674233 60674234 60691459 60691460 60691461 60691462 60714233 60714235 60714236 60714238 60714240 60714241 60735698 60735699 60735701 60735702 60735704 60735707 60757103 60757104 60757105 60808023 60808024 60808025 60808027 60819326 60819327 60819328 60819329 60819330 60819332 60819333 60819334 60819335 60877324 60877326 60877328 60877329 60877330 60877331 60877333 60877334 60877335 60877337 60931145 60931146 60931147 60931148 60931149 60931150 60931151 60931152 60931153 60950616 60950617 60950618 60950619 60950620 60966617 60966622 60966623 60966624 60966625 60966626 60974172 60974173 60974174 60974176 60979604 60979605 60979606 60985454 60985455 60985456 61031031 61040691 61040692 61040694 61040695 61071457 61075723 61075724 61084132 61084133 61086380 61091743 61096564 61096565 61096566 61100826 61100827 61100828 61104705 61104706 61109135 61109136 61112668 61112669 61112670 61112671 61120295 61125226 61131932 61133886 61133887 61140214 61140215 61143256 61143257 61143259 61143260 61161391 61178950 61178951 61178952 61186060 61211648 61242337 61242339 61249869 61267734 61293779 61303306 61303308 61303312 61306518 61306520 61383104 61388918 61388922 61396182 61421244 61439816 61446334 61446335 61446336 61464959 61464961 61484485 61512419 61516690 61516691 61516692 61530545 61538694 61542641 61542642 61547947 61561728 61567421 61584890 61597024 61597026 61597028 61657691 61665037 61677633 61677638 61677639 61677640 61677648 61677650 61693738 61693739 61709155 61729584 61729585 61729587 61729590 61729591 61729592 62282803 62282805 60090463 60099569 60099570 60103537 60124867 60130224 60135229 60135230 60140399 60150081 60152980 60154223 60165661 60165662 60167171 60175757 60177203 60189049 60193854 60196130 60196135 60208215 60208645 60210697 60230020 60230021 60242027 60242028 60253799 60253801 60265743 60265744 60288523 60288524 60304033 60304034 60323377 60323378 60323379 60338589 60338591 60363828 60363829 60363830 60367593 60367594 60380494 60380495 60387530 60399633 60399634 60407690 60410622 60422651 60425167 60438348 60451203 60461568 60461572 60461573 60497854 60497863 60506008 60506010 60532030 60537357 60537359 60550389 60550390 60550393 60550394 60570006 60570007 60572979 60575741 60575742 60575743 60602591 60620438 60620440 60630491 60653438 60653439 60653440 60653441 60691464 60691465 60714246 60714247 60714248 60714249 60735709 60735710 60735711 60757109 60757110 60757111 60814569 60814571 60814573 60832656 60832657 60877352 60877353 60877354 60877356 60877358 60877360 60877363 60931183 60931184 60931185 60931186 60950649 60950650 60950652 60950653 61009873 61009875 61009876 61059941 61059942 61064824 61071462 61075725 61084134 61096567 61096568 61112672 61120296 61120297 61125228 61133892 61143261 61143262 61186062 61186063 61211649 61221452 61221453 61242341 61249872 61258952 61263189 61267736 61280085 61303314 61306523 61358801 61379432 61388923 61388924 61414393 61421246 61421247 61446337 61475189 61489991 61504097 61516693 61564762 61613186 61621175 61709158 61729594 62298609 62298610 60152326 60468389 60811010 60847850 60956536 60958478 61395927 61578717 60208217 60265745 60323380 60363832 60380496 60452063 60513280 60550396 60597687 60630492 60653442 60691466 60747673 60757112 60819455 60860818 60860819 60926466 61125229 61175301 61226099 61277979 61428474 61555834 60367595 60387531 60461575 60461576 60506011 60506012 60550397 60550398 60597689 60620442 60757113 60811011 60819281 60877365 60877368 60877369 61009715 61009716 61067804 61079817 61091744 61186064 61194473 61231202 61339511 61358802 61379433 61383106 61414394 61453393 61489992 61512420 61729595 60107390 60109576 60112822 60114659 60115168 60117056 60120574 60123315 60124868 60129684 60135231 60135232 60142934 60145214 60145215 60154224 60154225 60154226 60155341 60165663 60165664 60167172 60167175 60175761 60177204 60179291 60186886 60187393 60189050 60195448 60202496 60202497 60208646 60218873 60218874 60230022 60230023 60230024 60230025 60242029 60242030 60245727 60253802 60253803 60253804 60265746 60265747 60281610 60281611 60288592 60310616 60310617 60310618 60323381 60323382 60323383 60323384 60338594 60338595 60338596 60358150 60367598 60367601 60380497 60380498 60385510 60387532 60393433 60393434 60393435 60399637 60407691 60407692 60407693 60407694 60422653 60429169 60438349 60438350 60438351 60444322 60452065 60452066 60461577 60461578 60479144 60479146 60484272 60491739 60491740 60506014 60506015 60506016 60506017 60506018 60506019 60537361 60537362 60538416 60550399 60550400 60550401 60550402 60550403 60572980 60602594 60602596 60602599 60615933 60620445 60620446 60630493 60654007 60654008 60654009 60654010 60654011 60654016 60654018 60654020 60674235 60674236 60691519 60691531 60714252 60714256 60714257 60714258 60714260 60735713 60735714 60735715 60735716 60735717 60757114 60808030 60808031 60808032 60808033 60819336 60819337 60877374 60877379 60877381 60877382 60877385 60877387 60877388 60931176 60931177 60931178 60931179 60931180 60931181 60974162 60974164 60974166 60985443 60985444 60985446 60985447 60985449 60985450 61004037 61004038 61024752 61031009 61031010 61040718 61040719 61059938 61059939 61096573 61096574 61104708 61109137 61109138 61112673 61112675 61120299 61125232 61131933 61140217 61143281 61148365 61158663 61186065 61194485 61215272 61242343 61263190 61271264 61303315 61306524 61306541 61388926 61388927 61396184 61439817 61446338 61446339 61453394 61464964 61464966 61464967 61464968 61504098 61509885 61516694 61542367 61665047 61677652 61729652 61729657 61753004 61753005 62262080 62282806 60090464 60099571 60103538 60127540 60130225 60135233 60135234 60140401 60150083 60152981 60154227 60155342 60165667 60165668 60167177 60172523 60177205 60186887 60189051 60196138 60201144 60208222 60210699 60218876 60218877 60230026 60230027 60242031 60242032 60253806 60265748 60265749 60310620 60310621 60310622 60323387 60338600 60338601 60338602 60358152 60363833 60367607 60367608 60380499 60380501 60387533 60393437 60399638 60407696 60422656 60422657 60438356 60451124 60451205 60461602 60491742 60506027 60506028 60506032 60506033 60506037 60537365 60537367 60537368 60550404 60550405 60550406 60550409 60570009 60572981 60597691 60597692 60602604 60602606 60602609 60654027 60654031 60654032 60691535 60691536 60714262 60714263 60714264 60714265 60735719 60735720 60757117 60757119 60817597 60817598 60817601 60873122 60877689 60877690 60877691 60877692 60877693 60877695 60877696 60877697 60877698 60931170 60931171 60931172 60931174 60950621 60950624 60950625 61004034 61004035 61004036 61009877 61009878 61059953 61062526 61062529 61066628 61075726 61075727 61075728 61084135 61100830 61104709 61109139 61125233 61125234 61133896 61143283 61158664 61171499 61190331 61211653 61215273 61221454 61221455 61226102 61242345 61242346 61263192 61267742 61280089 61286473 61303333 61387098 61388928 61388929 61388930 61434744 61446340 61464969 61475190 61484488 61512421 61516695 61564765 61613189 61621177 61728675 61728676 62137925 62298611 62319594 62319595 60393247 60711123 60789894 60817573 60908911 61220850 61408987 60140402 60155343 60208224 60208647 60620452 60679617 60735721 60806470 60860821 60926472 61140218 61249877 61256015 61277980 60288596 60461614 60461616 60532034 60597693 60602620 60757120 60853635 61009729 61067805 61091747 61131934 61133898 61231203 61280090 61293780 61325066 61332796 61379435 61414396 61475192 61530547 61547949 61621178 60097919 60101718 60109577 60117057 60120575 60123318 60135235 60140403 60142935 60145216 60154228 60154229 60159610 60167180 60177206 60185176 60189052 60202498 60218878 60218879 60230028 60230029 60242033 60242034 60253807 60265751 60288598 60288599 60310623 60310624 60323388 60323389 60338604 60338606 60358153 60358154 60367610 60380505 60385521 60387535 60399639 60407697 60422665 60431523 60438357 60444323 60444827 60452068 60461618 60479148 60484288 60484290 60491744 60506038 60506040 60506041 60506042 60506043 60550412 60550413 60550414 60550415 60602623 60602626 60615917 60615918 60630495 60674238 60691538 60714266 60714267 60714268 60735723 60757122 60808034 60819346 60819348 60819349 60819350 60877701 60877702 60877704 60877705 60966601 60966602 60979594 60979595 60985465 60985466 61040720 61059930 61059931 61075729 61084136 61091748 61096575 61104710 61109142 61112677 61120302 61131935 61143284 61158666 61164130 61190344 61211657 61221456 61226104 61242348 61249879 61267743 61280091 61286474 61306558 61421249 61434745 61439819 61446341 61464970 61475193 61504099 61516696 61547951 61564766 61613190 61621179 61665050 61728677 61728678 61729659 61763550 60090465 60099572 60103539 60117039 60129687 60135236 60135237 60145217 60150089 60152954 60154231 60154232 60165669 60165670 60167198 60175764 60177207 60186888 60187394 60196139 60208648 60208649 60208650 60218884 60218885 60242206 60242207 60245728 60253808 60253809 60265752 60288628 60288635 60310716 60310717 60310718 60323395 60323396 60338609 60338610 60358155 60358156 60367611 60380506 60387536 60393438 60407698 60422667 60438359 60451125 60452069 60461619 60461621 60461622 60491746 60497865 60497867 60532041 60532045 60537369 60537371 60550417 60570012 60570013 60570014 60570015 60575745 60602628 60602629 60602632 60602634 60620454 60620455 60620456 60620457 60654042 60654043 60654044 60654045 60691540 60691541 60714270 60714271 60714272 60714273 60735786 60735787 60735788 60757133 60757147 60757148 60757149 60866149 60866151 60866152 60866156 60866158 60866160 60867936 60867937 60867938 60867939 60867940 60877712 60877713 60979612 60979613 60991171 60991172 60991173 60991176 61040644 61040645 61040646 61040647 61059932 61059933 61071463 61071464 61075730 61091749 61112678 61112679 61112680 61112681 61120301 61120303 61133901 61140220 61143285 61178961 61190334 61190340 61211659 61221457 61242349 61249881 61263193 61267744 61280092 61303336 61388931 61396185 61414398 61421250 61453395 61475195 61484489 61504101 61516697 61516698 61555836 61613192 61621180 61709160 60393248 60757134 60807411 61226105 61331080 61383109 61467827 61631409 60132574 60338611 60468392 60620458 60691543 60747666 60814271 60889690 61226106 61325067 60253811 60513286 60513287 60550420 60550421 60573738 60620459 60819462 60858007 60877717 61009865 61091751 61125235 61131936 61186068 61256018 61263194 61280094 61293782 61319684 61325068 61358804 61383110 61414401 61489993 61530550 61719709 60101719 60117058 60127542 60135238 60140404 60154233 60157588 60167199 60175772 60186889 60196147 60196149 60242208 60259330 60265753 60288636 60288637 60288639 60323397 60338612 60338613 60338614 60363835 60380507 60385528 60387537 60410623 60422669 60429171 60431524 60431525 60444324 60444828 60461627 60461628 60506052 60506054 60506056 60506057 60537373 60550422 60550423 60575746 60602635 60620460 60620461 60654047 60691544 60714274 60735789 60757136 60757137 60808036 60808037 60819451 60819452 60819453 60877718 60877719 60877720 60931195 60985540 60985541 61040681 61040682 61040684 61109144 61112682 61120304 61120981 61131937 61190343 61211661 61267746 61271273 61286469 61306562 61420296 61420297 61446342 61464971 61484490 61523821 61547953 61650561 61729661 61729663 61763514 61763515 62262082 60099573 60099574 60099585 60124870 60129688 60135240 60140405 60145218 60145219 60154234 60154235 60154237 60157589 60165672 60167201 60167202 60172525 60177208 60186890 60187396 60196178 60196181 60201147 60201149 60218886 60218887 60230030 60242035 60242036 60259331 60265754 60265755 60288640 60288641 60288648 60288649 60310719 60310720 60323398 60323399 60338615 60338619 60358157 60367613 60380508 60380509 60385531 60387538 60393439 60399641 60407699 60410624 60422670 60425169 60438360 60438361 60438362 60461630 60461634 60461635 60497870 60497871 60497872 60497873 60497874 60497876 60513295 60532046 60532048 60537375 60537376 60550424 60550425 60550428 60561124 60570017 60570018 60570019 60572982 60575749 60575750 60575751 60602637 60602638 60620463 60620464 60630497 60654049 60654054 60691549 60691550 60691551 60714276 60714277 60714278 60714280 60735791 60735792 60735793 60735794 60757140 60757142 60814546 60814548 60814550 60814551 60873121 60877855 60877857 60877858 60877859 60889697 60889698 60915898 60915899 60915901 60915902 60937494 60950658 60950660 60950661 60950662 60991178 60991179 60991180 60991181 60991182 60991183 60991184 61040648 61040649 61040650 61059950 61059951 61064861 61075731 61104711 61109145 61112683 61112684 61112685 61120305 61125236 61131938 61140221 61143286 61164132 61186069 61221459 61226108 61242350 61249891 61267747 61280104 61303339 61306563 61387099 61388932 61388933 61396186 61421251 61453396 61475196 61484492 61489994 61504102 61516699 61516701 61547955 61555837 61564767 61613193 61621181 61665052 61729664 62262083 62262084 62285833 62285836 60393249 60660430 60789892 60813556 60873119 60877728 61395933 61630708 60152327 60367615 60460689 60664487 60714282 60807420 60871494 60877729 60877730 61629206 61752998 60807428 60832663 60876551 60943956 61059940 61120306 61221460 61267749 61383111 61635830 60620465 60730668 60814291 60847853 60858008 61031035 61037802 61091752 61120307 61131939 61133905 61293783 61349674 61367029 61387101 61399251 61472026 61475197 61763512 60097920 60142936 60152955 60157590 60167203 60185177 60196186 60230031 60242037 60253815 60288656 60288657 60288658 60338620 60367616 60380510 60387541 60410625 60444325 60444829 60484292 60484293 60491747 60506058 60506059 60506060 60506061 60691554 60714286 60735795 60757143 60814295 60814296 60814298 60873115 60877732 60877733 60877734 60877735 60877736 60931189 60931190 61041702 61041703 61075732 61091753 61104714 61109146 61133907 61211664 61221461 61226110 61263196 61280110 61428475 61446343 61484493 61502717 61584898 61657697 61729667 61763521 62036744 62285866 60095057 60099575 60101721 60101722 60132576 60135241 60135242 60140406 60142937 60145220 60150090 60154238 60154240 60155345 60165673 60165674 60167204 60172526 60177209 60177210 60186891 60189053 60196204 60196205 60201154 60201156 60230034 60230035 60239236 60242038 60242039 60253816 60253817 60265756 60265758 60288659 60288670 60310721 60310722 60310723 60323401 60323402 60323403 60338621 60338623 60338625 60358158 60358159 60367617 60380511 60380512 60385546 60393440 60393441 60393442 60399642 60407701 60410626 60425171 60431526 60431528 60451127 60461636 60461637 60461638 60479149 60479150 60497879 60497880 60497881 60497882 60506062 60506064 60537380 60550429 60550430 60550431 60550432 60550433 60570022 60572983 60575752 60602639 60602641 60602642 60620467 60620468 60620469 60630499 60654111 60654112 60654113 60654114 60654115 60674239 60691557 60691558 60714289 60714291 60714293 60735797 60735798 60735799 60814558 60814560 60814562 60814563 60814564 60814566 60832665 60832666 60832669 60877738 60877739 60877740 60877741 60877742 60931104 60931111 60931113 60931116 60931117 60931118 60931121 60991187 60991188 60991189 60991190 60991191 61041609 61041610 61041611 61041612 61066634 61066635 61075733 61075734 61084137 61084138 61084139 61109147 61112686 61112687 61112688 61120308 61120309 61125237 61131940 61178956 61186070 61186072 61211665 61211666 61221463 61293784 61303342 61306567 61379441 61383112 61387102 61387103 61388934 61421252 61421253 61421254 61453397 61484494 61489995 61504103 61516702 61523823 61542644 61547956 61555838 61564769 61597042 61621182 61677670 61677671 61677672 61693746 61711264 61711265 62262086 62286546 60783556 60813557 60847854 61306359 61388936 61403066 61439820 61578719 60129689 60151183 60370096 60606754 60745776 60897960 61341491 61641814 61709163 60807439 60832670 60877743 61003946 61091754 61120310 61133909 61293785 61367030 61578721 60615932 60620553 60757150 60807441 60860822 60860823 60877744 60931157 61120311 61120982 61133911 61186073 61323017 61349675 61395935 61414402 61453399 61501013 60145221 60152956 60169753 60177211 60230038 60242040 60288673 60288688 60338626 60422674 60438363 60444326 60444830 60484294 60491748 60506067 60506069 60561129 60620554 60654116 60691614 60714294 60735801 60757151 60819258 60819260 60819261 60819263 60877745 60877746 60877747 60877748 61059936 61091758 61109148 61120312 61186074 61194491 61211668 61242351 61263197 61280111 61306568 61439821 61464972 61523824 61564770 61763522 61780107 60090466 60097927 60099576 60103540 60112823 60115169 60117041 60123324 60129690 60132577 60135243 60135244 60142938 60145222 60150092 60154241 60154242 60165675 60165676 60167205 60167206 60175773 60177212 60186892 60187397 60195450 60208651 60208652 60218888 60218889 60230041 60242041 60242042 60242209 60253819 60253820 60265759 60265760 60288701 60288705 60288710 60288712 60310724 60310725 60310726 60310727 60323404 60323405 60338627 60338629 60338630 60358160 60358161 60358162 60367619 60380513 60387543 60393443 60399643 60407702 60407703 60425172 60431530 60431531 60455031 60461639 60461640 60487164 60491749 60497883 60497884 60497885 60506071 60506072 60506073 60537390 60537391 60537393 60550434 60550435 60550437 60550438 60570026 60570028 60570030 60575755 60575757 60602645 60602646 60602647 60602648 60620556 60620557 60620558 60654117 60654118 60654119 60654120 60654121 60674241 60691615 60691623 60714296 60714303 60714304 60714306 60735803 60735805 60757154 60757155 60757156 60814489 60814490 60814491 60814492 60814493 60814494 60814495 60814497 60814498 60886477 60886478 60886480 60886481 60886483 60886488 60886490 60886492 60886493 60931131 60931132 60931133 60931135 60931136 60966633 60979633 60991219 60991223 60991225 60991226 60991228 60991229 60991230 60991231 61040725 61040726 61040727 61059952 61075735 61075736 61084140 61084141 61084142 61120313 61120314 61120315 61120983 61140222 61143287 61148366 61171500 61186075 61211671 61211672 61293786 61303344 61306574 61383113 61387105 61388937 61420298 61420301 61420302 61420303 61453398 61464973 61484495 61484496 61504104 61516703 61516704 61542645 61547957 61665056 61677674 61693740 61693741 61693742 61693743 61719710 61719712 61719713 62285824 62285827 62285831 60393250 60783557 60873144 60877752 60877753 60877754 60877755 60877756 60877757 60877758 60877759 60877760 60877761 60877762 60711122 60789896 60817570 60847858 61091761 61112689 61277982 61358806 61595053 61629207 61753001 60807444 60832674 60877763 61019955 61091763 61133914 61275341 61578723 60789994 60860827 60897958 60897959 61057573 61057575 61091764 61133916 61186076 61266236 61280113 61293787 61306581 61383114 61395936 61428476 61472028 61496696 61629208 60159611 60189054 60253821 60288725 60323406 60338631 60438364 60444327 60506078 60506079 60602650 60615941 60679618 60714307 60757159 60757160 60814288 60858003 60860831 60860832 60860833 60860834 60860835 60860836 60860837 60877764 60877765 61112690 61158669 61215275 61226112 61242352 61271275 61293788 61306582 61428477 61509886 61530551 61652153 61652484 60101723 60117042 60135245 60140407 60142939 60145223 60150094 60154244 60154245 60155346 60165677 60167208 60172528 60177213 60186893 60186894 60187398 60196254 60196256 60196257 60208654 60208655 60218890 60230042 60242043 60242044 60253822 60265761 60288727 60288728 60288729 60288730 60310728 60310729 60323407 60323408 60323410 60338632 60338633 60338634 60358163 60367620 60367621 60380515 60380516 60380517 60385550 60387544 60393444 60407704 60407706 60438365 60451129 60452071 60452072 60461643 60461644 60487165 60497886 60497887 60506080 60506082 60506083 60506084 60537396 60537397 60537399 60550439 60550441 60550443 60550444 60572984 60575862 60602651 60602652 60602653 60620559 60620560 60654124 60654125 60654126 60654127 60654128 60691633 60691635 60714309 60714310 60714312 60735807 60735808 60757163 60757165 60807445 60807446 60807455 60814540 60814542 60814543 60869325 60873145 60873146 60873147 60873148 60873149 60873152 60873155 60873156 60873159 60877767 60877772 60877773 60877774 60991233 60991239 60991241 60991242 60991243 60991244 61040685 61040687 61040688 61040689 61104715 61109149 61112691 61112692 61112693 61112694 61120316 61125239 61131942 61140223 61148367 61271269 61286462 61303346 61306584 61421255 61421256 61421257 61446344 61484497 61504105 61516706 61547959 61584903 61665057 61677676 61677677 61693744 61693745 61709164 60376354 60629551 60956537 61331086 61383115 61434746 60152328 60725143 60789995 60817578 60886501 61332831 61597045 60748954 60776479 60813561 60832675 60966580 61153535 61226113 61597049 60606755 60615921 60620562 60877777 60915933 60915934 61064862 61120319 61120324 61133917 61306585 61325069 61367032 61400452 61420294 61443612 61453403 61484498 61613194 60154247 60155347 60155348 60155349 60157591 60444328 60620563 60735809 60789996 60819288 60877779 60877780 61064863 61104717 61112696 61131943 61133919 61263199 61280114 61293789 61306586 61331088 61367033 61383116 61414404 61439822 61458907 61530559 61613196 60088614 60099577 60103541 60135246 60140408 60142940 60154248 60154250 60155350 60165678 60167209 60172529 60175775 60186895 60196260 60208661 60218891 60230044 60242045 60253824 60253825 60288733 60288737 60323411 60323412 60338637 60358164 60367622 60380518 60387545 60393445 60407708 60422679 60438367 60451130 60461645 60461646 60479152 60479153 60487166 60497888 60497889 60497890 60506087 60506088 60537404 60550445 60550446 60550447 60570035 60572986 60575864 60602662 60602663 60620565 60620566 60620567 60654132 60691643 60691644 60714317 60714318 60735810 60735811 60757170 60757171 60814552 60814553 60814554 60814555 60814556 60877783 60877784 60877785 60877787 60877788 60877790 60877792 60926488 60926490 60926492 60926493 60926494 60926498 60991246 60991248 60991250 60991255 60991257 61059955 61071466 61075737 61075738 61109150 61112711 61120320 61120321 61125240 61547960 61709167 60376355 60813563 60877793 61394803 61403068 61434747 60665677 60711124 60789897 60814293 60873143 60877794 61621040 61745931 60741222 60789874 60832676 60877796 61077765 61147789 61211674 61319685 60811013 60853648 60909995 61064864 61075739 61120984 61186077 61266238 61293790 61306587 61358808 61399254 61420304 61454197 61458908 61502721 60288743 60323413 60367623 60597696 60620568 60757173 60847863 60873140 60877797 61037803 61040724 61067808 61140225 61186078 61211675 61263201 61267752 61331090 61358809 61383117 61428480 61458909 61489998 60152329 60373540 60538419 60811014 60875774 60877809 60152958 60208658 60757177 60797059 60956559 60958481 61733621 60750996 60789881 60813565 60832679 60902685 61091766 61148368 61191334 61242353 61629209 60620572 60811015 60814183 60814186 60877811 60877812 61091767 61120326 61131944 61186079 61367034 61410609 61427108 61446265 61471936 61496666 60620574 60757179 60789985 60858004 60877813 60926470 61064865 61091768 61120327 61131945 61263202 61267754 61325070 61348472 61400451 61422230 61458910 61501010 61584906 60165681 60196264 60208659 60242212 60288748 60310731 60323415 60358166 60367625 60422689 60461651 60461653 60497895 60532059 60532060 60532061 60532062 60532063 60570045 60570048 60575882 60575883 60654136 60691646 60714328 60735815 60757181 60757182 60814363 60814369 60814577 60814578 60869323 60877814 60877815 60877816 60877818 60937507 60937508 60966584 61037813 61037814 61109151 61125241 61271277 61453406 61538696 60784487 60813566 60847864 60847865 60877819 61140226 61191335 61221464 60265763 60310732 60363836 60387548 60513298 60538420 60570049 60597698 60691647 60714335 60735816 60757183 60807469 60814355 60832680 60853691 60853692 60869324 60877820 60915922 60925701 60925704 60966582 60966583 61104718 61120328 61161390 61267755 61275343 61286459 61453407 61484499 60757184 60807464 60877821 60966532 61104724 60528781 60536324 60537405 60570050 60615912 60620580 60654138 60679616 60691648 60711129 60714336 60730285 60757185 60807467 60814229 60847866 60847867 60847868 60847869 60847870 60877823 60877824 60979627 60985461 61109152 61120330 61226114 61266232 368071 61117743 61183889 61183892 61117744 61140736 61224424 61290842 61117744R 368073 60231602 60933755 61140737 61298469 61383009 00368073R 60231602R 367945 60933756 61113660 61146938 61249079 61276093 61328158 61113660R 61276093R 368074 60913737 60933757 60988458 60988459 61249083 61276097 60988459R 61249083R 368075 60933758 61140738 61254904 61254905 61276104 61344232 61614752 61620478 60933758R 61344232R 368076 60893366 61183908 60893366R 368077 61075886 61117745 61140739 61290843 61075886R 368078 369083 61198494 61244779 61277983 61298470 61338796 368079 60922793 61117746 61183911 61327768 60922793R 61327768R 368080 369378 60783531 60784483 61079844 61117747 61276098 61332779 00368080R 60784483R 61079844R 61117747R 368081 369084 61049723 61120898 61146939 61276106 61344233 61120898R 61276106R 368082 60848468 60984602 61249082 61332780 61475203 60848468R 61249082R 368083 60753359 61143359 61320049 60753359R 61143359R 368084 60893370 61320050 61415317 00368084R 60893370R 368085 369085 61075887 61277984 61317591 61338797 61434768 00368085R 61079845 61187615 61276094 61327769 00367918R 61079845R 368086 60708946 61049725 61049727 61267606 61306362 61332781 61049725R 368087 61092441 61171124 61276107 61332782 61351683 00369086R 61332782R 61351683R 368089 60848469 61064766 61194137 61213187 61276113 61351684 61064766R 61194137R 368090 61113661 61117748 61344234 00368090R 368091 60893371 61079850 61117749 61377332 00368091R 60893371R 368092 60630217 60893372 61140741 61146941 61382376 367919 60241629 60373379 61069557 61276088 61320071 61395126 61590997 00367919R 61069557R 367920 369854 60159586 60426219 60704475 61031702 61276095 61327770 61348468 61351685 60426219R 60704475R 368093 60173915 60598578 61049676 61049678 61120899 61276099 61327771 61351686 61382377 00368093R 367357 60345797 60665766 61049703 61096698 61276108 61317592 61344235 61413337 61344235R 368095 60708940 61020653 61048028 61276114 61324293 61377333 61020653R 368096 61040187 61048032 61280050 61310741 61344236 00368096R 61048032R 61280050R 368097 60753360 61162315 61172716 61188444 61282983 61327772 61338798 60753360R 61162315R 368098 60665009 60858232 61187616 61188446 60665009R 60858232R 60753361 60893373 61344237 60422979 60893374 61290846 61344238 60422979R 367190 60127928 60159587 60197144 60197145 60241631 60592014 60894639 60961020 61069558 61069831 61217318 61281842 61303771 61340887 61370226 61516764 367301 60150753 60266241 60388938 60515407 60806261 60894578 60894580 61203583 61217319 61303772 61329272 61329273 61338803 61338804 61340888 61377335 61382380 60385447R 367229 367927 60119148 60173917 60278563 60360171 60377212 60426221 60487379 60645075 60703527 60874471 60913735 61217320 61240217 61240218 61269486 61281843 61329274 61347427 61356879 61390160 60360171R 60377212R 368101 368399 369444 60108248 60161914 60252754 60342198 60345807 60426222 60733117 60797050 61079859 61276109 61315091 61410000 61445329 61475204 60108248R 368102 368400 60386465 60670029 60837645 61064767 61140743 61276115 61306361 61360838 61475205 60386465R 60837645R 368103 60138508 60718441 61117750 61172719 61307933 61344239 00368103R 60138508R 61172719R 368104 61170167 61172720 61185471 61332783 61458949 61170167R 61185471R 61332783R 368105 369093 60858220 61327773 00368105R 60858220R 368106 60138509 60255270 61069559 61327774 60255270R 61069559R 367784 60255303 60684924 60893375 61064494 61069560 61332784 60255303R 60684924R 61064494R 367931 368107 369095 60119149 60123924 60140130 60197146 60208539 60231174 60252755 60266242 60293989 60310135 60695331 60758102 60931748 60931749 61143361 61162313 61172721 61187618 61187619 61240219 61269487 61340889 61382387 61779974 60123924R 60231174R 60310135R 61269487R 367885 60098447 60119150 60123925 60123927 60138510 60138511 60140131 60173918 60173921 60197147 60219085 60252756 60252757 60278565 60367707 60380225 60393456 60399552 60416149 60444869 60444870 60515408 60536751 60601732 60654173 60671985 60671986 60715110 60715112 60735586 60758103 60925120 60925121 60925122 60931251 60931252 60931254 61031618 61031620 61117751 61147688 61186091 61217322 61217323 61269488 61294510 61329277 61329279 61361876 61379587 61382388 61433463 61508361 61516765 61780038 60278565R 60367707R 60536751R 60735586R 368108 60093668 60104809 60123930 60140133 60154659 60173922 60173923 60197148 60197150 60241634 60241635 60252758 60258782 60293990 60293991 60380227 60399553 60407627 60416150 60487381 60528776 60536752 60592023 60621840 60654174 60654175 60671987 60695332 60695336 60695338 60758104 60789438 60874472 60874473 60913746 60913747 61069562 61146942 61171132 61172724 61203587 61217324 61217325 61234897 61234898 61240253 61347442 61377336 61413338 61516766 61618130 61690443 61735139 60098448R 60104809R 60140133R 60695338R 60140134 60173924 60173925 60197151 60241636 60252759 60258783 60278567 60293992 60345810 60345813 60401283 60407628 60416151 60515409 60561213 60583666 60691728 60789439 60913738 60913739 61069563 61171133 61194139 61217326 61224434 61303777 61317598 61346031 61428531 61453180 61508362 61719083 60583666R 61224434R 367191 367847 369097 60135458 60258785 60528777 60621846 60691729 60745758 60789440 60844413 60848470 61113663 61276117 61307934 61607678 61113663R 368233 369098 60119151 60703528 60833761 61338799 61397501 00368233R 60703528R 368234 60848472 61324292 61327775 61338800 60848472R 368235 60848473 61184356 61324291 00368235R 368236 60899321 60931705 61117752 61140745 61221370 00368236R 60899321R 60126575 61092445 61224436 61263724 61290847 61377337 61420487 61684788 60126575R 61377337R 368172 60080830 60093679 60119152 60123932 60154660 60156398 60173926 60197152 60241637 60252760 60266245 60266246 60278568 60487383 60536753 60601734 60621847 60645053 60654176 60671988 60695339 60695340 60894717 61003746 61020652 61092446 61120902 61147692 61217327 61269489 61307912 61329285 61340891 61361878 61565521 60080830R 61340891R 60093680 60098449 60123933 60123934 60123935 60131277 60138513 60138514 60140135 60140136 60173927 60173928 60175568 60176988 60186991 60186992 60186993 60186994 60186995 60219088 60231175 60231176 60241638 60241639 60241640 60258786 60260229 60266248 60266249 60310136 60310137 60310138 60317281 60317282 60333374 60333375 60337473 60338208 60338209 60416153 60444872 60445146 60487384 60515410 60536754 60550459 60573198 60592024 60621848 60630444 60645063 60645064 60654178 60695341 60695355 60735587 60735588 60789442 60894596 60894598 60894601 60894602 60894615 60985627 60985628 60985629 60991117 60991118 60991119 61031646 61113664 61113665 61113666 61120903 61186093 61198495 61203592 61240256 61263732 61303778 61307913 61329291 61361880 61382392 61433465 61508363 61516767 61618133 61722806 61735140 61775172 60241640R 60654179R 61217328R 60080831 60080832 60093681 60098450 60119154 60123936 60123938 60138515 60138516 60140139 60154661 60154662 60156401 60156402 60161915 60161916 60173929 60173930 60197153 60197154 60208541 60208542 60219089 60219090 60231177 60241641 60241642 60266250 60269628 60278569 60278570 60310139 60310140 60325571 60333376 60337475 60338211 60338212 60401284 60401285 60416154 60437825 60444873 60444874 60445147 60445148 60515413 60515415 60536755 60583670 60583671 60630447 60645049 60645050 60654180 60671990 60671991 60695361 60715114 60715115 60735589 60758107 60758108 60789443 60913701 60937734 60937735 60937736 60937737 60937742 60973749 60973750 60973751 60985630 60985631 60991120 60991121 61054830 61113667 61113668 61113669 61117753 61117754 61128202 61146947 61162308 61172725 61172726 61198498 61224438 61269491 61277992 61329292 61329293 61356883 61361881 61377338 61382394 61382396 61490756 61502930 61612585 61672778 61693769 61693771 61708239 61722807 60278569R 368238 60080833 60080834 60089206 60093682 60119155 60127611 60127613 60127937 60130253 60131279 60138518 60140141 60173931 60173932 60175569 60197156 60219091 60219092 60219093 60241643 60241644 60260257 60278571 60278572 60338213 60367708 60380228 60399554 60407629 60407630 60416155 60416156 60444876 60445149 60550461 60654181 60671992 60715116 60735590 60813986 60874479 60913740 60913741 60913742 60973753 60973754 61052177 61052178 61052179 61052180 61140746 61147697 61194142 61203609 61217329 61234902 61298472 61303781 61338817 61423278 61423286 61428532 61501014 61501015 61607679 61719011 61735141 60173931R 61147696R 368239 369101 60119156 60123942 60127617 60127938 60130254 60131280 60131281 60138519 60247128 60247129 60260258 60310141 60338214 60429239 60437826 60443909 60444877 60444878 60711450 60837647 61040180 61040182 61048030 61143362 61186089 61187622 61194143 61458951 61612586 61722808 60131280R 368241 60156403 60197157 60833763 61140748 61179438 61183893 00368241R 368242 369180 60156404 60750988 61162318 61194144 61327776 61338801 61475206 60156404R 60750988R 61162318R 368403 60718442 60858223 61322926 61389804 61322926R 61140752 61281250 61344240 00368404R 61140752R 61281250R 368405 60252761 61260570 61360839 61415321 60252761R 60080835 60119157 60140148 60173933 60186999 60197158 60219095 60231178 60241645 60258787 60266251 60278573 60310143 60310144 60407632 60416157 60515416 60630449 60654182 60695369 60735591 60735593 60735594 60789445 60894705 60894707 60894710 60894715 61117755 61117756 61125121 61147714 61240261 61268007 61276089 61294511 61307915 61329299 61338819 61360840 60080835R 60173933R 60219095R 60278573R 60630449R 60735594R 367888 60127628 60127631 60127939 60130251 60130252 60131282 60131283 60136408 60138520 60138521 60138522 60140143 60140147 60173934 60173935 60175570 60187002 60187007 60197159 60208543 60208544 60208545 60219096 60219097 60219098 60219099 60241646 60241647 60252762 60258788 60278575 60293999 60294000 60294001 60310145 60317283 60333377 60333378 60338215 60338216 60387684 60407633 60416158 60437827 60444880 60444881 60444882 60445151 60445152 60515417 60515418 60515419 60536757 60573202 60601742 60621858 60630450 60654183 60654184 60671993 60671994 60671995 60695377 60715118 60735595 60735596 60758109 60758110 60789446 60789447 60894641 60894658 60931255 60931258 60931259 60931261 60973735 60973736 61003749 61113670 61113671 61113672 61117757 61120905 61128205 61143363 61172728 61172729 61203613 61217332 61224444 61240263 61281845 61294512 61298473 61303782 61303784 61377340 61379593 61382399 61382401 61388021 61428533 61428534 61428535 61433466 61516768 61590999 61597119 61676934 61719076 61719078 61719079 61719080 61719081 61735142 61775174 60127631R 60973736R 61172729R 61379593R 368245 60080839 60080840 60080841 60089207 60093685 60093686 60104814 60123945 60127940 60138523 60138524 60140149 60150757 60154664 60156405 60156406 60156407 60165704 60173936 60173937 60175571 60176989 60187009 60187011 60197160 60197161 60208546 60208547 60208548 60219101 60219102 60231179 60231180 60231181 60231182 60241648 60241649 60247130 60252763 60258789 60260260 60266253 60266254 60269629 60278576 60310147 60310148 60310149 60317284 60325574 60325575 60333379 60333380 60337476 60337477 60337478 60338217 60338218 60367709 60380229 60380230 60393462 60393464 60393466 60401287 60407634 60416159 60423166 60426223 60437828 60437829 60437830 60445089 60445090 60445154 60445155 60445156 60445157 60515421 60536758 60536759 60550462 60573204 60573205 60583675 60583676 60601744 60601745 60645037 60645043 60654185 60671997 60671998 60671999 60672001 60695379 60695381 60695383 60715121 60735598 60735599 60789449 60789450 60874463 60874464 60874475 60874477 60913708 60913710 60913711 60931715 60931718 60931719 60937727 60973752 60991127 61052196 61052197 61052198 61052199 61052201 61069565 61069566 61100916 61113673 61113674 61113675 61120912 61120915 61128207 61128208 61146956 61146957 61146958 61147731 61162309 61162310 61172733 61203615 61203617 61209927 61224448 61234906 61240267 61277996 61277997 61281848 61298475 61361886 61382406 61382407 61388023 61428537 61428538 61441066 61502931 61502933 61523838 61530388 61674475 61676938 61676939 61676940 61676942 61676943 61676944 61693757 61693758 61693759 61708240 61735143 61735144 61747578 60241649R 368392 60080842 60080843 60093688 60098451 60098453 60123946 60123947 60127633 60130257 60131285 60131286 60138525 60138526 60140151 60140153 60154665 60156408 60165705 60173938 60173939 60187013 60187015 60197162 60208549 60208550 60208551 60219103 60219104 60219105 60231183 60231184 60241650 60252764 60258817 60260261 60310150 60310151 60310152 60317285 60333381 60337479 60337480 60338219 60338220 60342202 60399555 60401288 60401289 60407637 60437831 60437832 60445092 60445158 60445159 60515422 60515423 60573206 60583678 60592026 60601746 60601748 60621863 60630451 60645025 60645026 60645028 60654186 60672002 60672003 60672004 60672006 60695399 60695404 60695407 60695410 60758112 60758113 60789451 60813993 60813994 60874474 60925123 60925124 60925125 60931706 60931707 60937724 60937725 61052191 61052192 61052193 61113678 61117758 61128209 61128210 61128211 61143366 61143367 61171138 61194155 61224449 61237553 61263728 61268002 61276110 61329310 61338826 61379600 61388024 61441067 61591002 61674435 61676999 61677003 61708235 60219103R 60231184R 60621863R 60758112R 368339 60127941 60127943 60130247 60130248 60138527 60140155 60140156 60173940 60187017 60187019 60197163 60231185 60241651 60252765 60269630 60294002 60294003 60310153 60338221 60338222 60423168 60426224 60443911 60443912 60445095 60445096 60583679 60630452 60695414 60695416 60695418 60695423 60837648 60837649 61040183 61040184 61048031 61151164 61171139 61179440 61183895 61224450 61263727 61294515 61327547 61338827 61361890 61382409 61641824 61722814 60241651R 60338222R 60695414R 368408 60140158 60173941 60197164 60252766 60342203 60342204 60582870 60703529 61172743 61172744 00368408R 60140158R 368409 60123949 60173942 60747428 61277985 61298479 00368409R 60173942R 61277985R 368410 60154666 60837650 60858225 61187623 61315092 61344241 60154666R 61187623R 61315092R 368411 61095371 61120916 61187625 61281251 61327778 61338802 61095371R 61120916R 61187625R 368412 61140759 61194156 61290174 61327779 61428539 61512877 61194156R 367195 60104816 60119160 60127944 60187023 60197165 60208552 60247131 60252767 60266255 60282781 60445160 60668363 60735601 60735603 60789452 60789453 60894623 60894625 60894627 60894629 61187570 61240268 61281849 61307919 61340895 61382412 61388026 61490757 61747583 60208552R 60266255R 60735603R 365806 367300 367889 368247 60080846 60080847 60080848 60082909 60093689 60093690 60123950 60123951 60131289 60131291 60131292 60138528 60138529 60140161 60140162 60165706 60173943 60173944 60187025 60187026 60197166 60208553 60219106 60219107 60219108 60231186 60231187 60231188 60231189 60241652 60266260 60282782 60294004 60294005 60294006 60310205 60310208 60317286 60337481 60338223 60393478 60401290 60437833 60443914 60445102 60445103 60445162 60515427 60536762 60550466 60573208 60583680 60601753 60601755 60621865 60645055 60645056 60645057 60672007 60672008 60695446 60715122 60735604 60735605 60758114 60758115 60789457 60894587 60894589 60894590 60894591 60973740 60973741 60973742 60985618 60985620 60991122 61092447 61100925 61113679 61128212 61128214 61147745 61162312 61171140 61172753 61172754 61172755 61198499 61234909 61240271 61281851 61298481 61409982 61409983 61413340 61433467 61433468 61437644 61516769 61578052 61618136 61747584 61775175 61775176 60140162R 60219108R 61281851R 367890 368248 60080849 60080850 60080851 60080852 60082910 60089208 60093691 60093692 60098457 60098458 60123952 60127640 60131293 60138530 60140163 60140164 60154667 60154668 60156409 60156410 60161917 60173946 60176990 60187029 60187030 60197167 60197168 60208554 60208555 60208556 60219109 60219110 60219111 60219112 60231190 60241653 60241654 60247132 60252769 60258926 60260263 60266261 60269632 60282783 60282785 60294007 60294008 60294009 60310210 60310212 60310213 60317287 60317288 60325577 60333382 60333383 60337483 60338224 60380232 60393479 60423173 60423174 60437834 60437835 60437836 60437837 60443918 60443919 60445105 60445107 60445108 60445163 60515429 60515431 60550468 60550470 60550471 60557533 60573209 60573210 60583681 60583683 60583684 60601758 60601760 60621866 60621867 60630453 60645058 60645059 60654191 60654192 60665768 60672009 60672010 60672011 60691730 60695449 60695451 60715123 60715124 60735607 60758117 60758118 60789459 60789460 60789461 60789462 60789463 60913730 60913731 60913732 60913733 60925115 60925117 60925118 60931709 60931711 60931712 60973745 60973746 60985632 60985633 61052182 61052184 61052185 61052186 61069567 61113680 61113681 61113682 61117760 61120918 61120919 61128215 61128216 61146959 61146960 61162305 61171145 61172758 61209929 61209931 61217339 61217340 61224462 61240274 61269498 61269499 61276102 61278004 61356889 61361894 61382414 61413342 61413343 61428540 61428541 61434770 61468144 61480003 61480007 61490759 61490762 61618137 61677071 61677072 61693765 61708243 61708248 61722816 61735186 61735188 60310212R 60338224R 60789459R 60789462R 60913731R 61128215R 61217340R 368261 60080853 60080855 60082911 60123953 60123954 60127641 60127642 60127643 60127945 60127946 60127947 60131294 60131295 60131296 60138531 60138532 60138533 60140165 60140166 60175572 60187033 60197169 60197170 60208557 60208558 60219113 60219114 60249934 60252771 60258927 60258928 60260264 60282786 60294010 60294011 60310215 60317289 60333384 60333385 60337485 60337486 60338225 60399556 60407644 60407646 60423176 60437838 60445110 60445111 60445112 60445165 60487386 60515432 60515433 60536765 60550476 60573211 60583685 60583686 60601761 60645046 60645047 60645048 60654193 60672012 60672013 60672014 60695453 60695455 60715125 60735608 60758119 60789464 60789468 60789469 60789470 60813999 60913727 60913728 60913729 60925113 60925114 60937746 60985634 60985635 60991123 60991124 61069568 61092449 61100992 61113683 61113684 61120920 61162311 61170171 61170172 61171146 61172761 61187580 61209932 61240278 61269502 61269503 61298484 61303797 61307925 61329316 61329317 61329318 61379606 61379607 61382416 61441060 61512878 61565524 61674468 61677105 61693755 61722819 61722833 61735207 61747588 60445110R 368413 60082912 60123955 60123956 60127645 60130258 60135461 60140167 60140168 60140169 60173947 60175573 60187035 60187037 60208559 60219115 60219116 60249935 60310217 60310218 60342205 60380235 60429254 60443920 60443921 60445116 60445118 60445120 60550480 60582873 60583687 60621870 60672015 60695458 60715127 60758120 60813941 60837651 60844414 61003741 61003742 61020650 61040175 61128218 61171147 61172762 61186086 61186087 61198505 61263730 61285985 61290178 61307936 61329319 61360842 61502935 365787 60123957 60165707 60187038 60260266 60443923 60443924 60445121 60582871 61186081 61186082 61186083 60123957R 60187038R 60445121R 368415 60150761 60187040 60426969 60733116 61263726 61285984 61332785 61428542 00368415R 60150761R 60187040R 60426969R 60733116R 61285984R 368327 369104 60067825 61172763 61277986 61329351 61475208 00368327R 60067825R 61172763R 61277986R 368416 61100993 61183909 61277987 61317694 00368416R 61100993R 61183909R 61317694R 60897034 61281252 61327780 61453184 60897034R 367849 368249 60140170 60187043 60187044 60252776 60294012 60443927 60592027 60630454 60695460 60725704 60894719 60894720 61092451 61194160 61224466 61240280 61329321 61329322 61340899 61382418 61667268 61735218 61747648 60140170R 60252776R 60725704R 60894720R 365809 367196 367892 368417 60080856 60093695 60127949 60131297 60138534 60140171 60140172 60175574 60187046 60187047 60197171 60208561 60208562 60219117 60260267 60282788 60310220 60317290 60380238 60443928 60443931 60445124 60445125 60445126 60445167 60515434 60583688 60592028 60621871 60654195 60672016 60715128 60735609 60758121 60789471 60789472 60894684 60894685 60894688 61003743 61003745 61020647 61031661 61172765 61172766 61224468 61240281 61356892 61356894 61409989 61437645 61509900 61591003 61667269 61722836 61735219 00367196R 60672016R 365792 367798 60072087 60080858 60080859 60082914 60093697 60093698 60119161 60123958 60131298 60131299 60140173 60154669 60154670 60156411 60173948 60175575 60187049 60187050 60197172 60197173 60208563 60219118 60219119 60219120 60241655 60241656 60252778 60260268 60266263 60266264 60282790 60310222 60310223 60317291 60338226 60380239 60393480 60407647 60429255 60437840 60443933 60443934 60445128 60445129 60445168 60536766 60550483 60550484 60573212 60573213 60582874 60583689 60601769 60621872 60621873 60645022 60654197 60672017 60691731 60695462 60695464 60735610 60735611 60758123 60789474 60874470 60913704 60913705 60913707 60937728 60937729 60973747 60973748 61052173 61052174 61052175 61052176 61128220 61146964 61147749 61147750 61147751 61162303 61187590 61203634 61217341 61224471 61240284 61240285 61269507 61278010 61278011 61278013 61329324 61382421 61388029 61428543 61441061 61441062 61480008 61502936 61523844 61563318 61591004 61667270 61693767 61693768 61708255 61747591 60131299R 60208563R 60310222R 60310223R 60443934R 60735611R 60789474R 365810 367893 368418 370731 60080860 60080861 60082915 60093699 60127649 60127950 60131301 60140174 60140175 60156412 60165708 60173949 60175576 60176991 60187051 60197174 60197175 60208564 60208565 60219121 60219122 60219123 60252780 60260269 60282791 60282792 60310225 60317292 60367711 60367712 60380240 60393481 60423185 60423187 60437841 60437842 60443935 60443936 60445131 60445133 60445169 60536767 60550485 60573214 60583690 60621874 60621875 60645032 60654198 60654199 60654200 60672019 60695468 60715130 60758124 60758125 60814007 60814008 60925126 60937745 60991131 60991132 61069570 61100996 61113686 61117764 61117765 61128221 61171150 61186085 61224472 61224473 61249733 61253802 61253803 61263733 61303799 61307927 61329326 61341148 61341149 61346036 61382422 61388030 61614584 61677112 61677113 61693760 61693761 61708237 60140175R 60175576R 60621874R 60654198R 61303799R 367197 60080862 60093703 60123960 60127658 60130256 60131302 60131303 60138537 60140176 60175577 60197176 60197177 60231191 60240130 60240131 60260270 60266265 60380241 60423188 60437843 60437844 60443937 60445135 60515435 60583691 60592029 60622428 60645054 60695480 60715131 60758126 60758127 60837652 60837653 60844415 61064769 61143368 61183922 61194162 61224475 61240287 61253804 61303801 61327548 61327549 61368154 61382423 61388031 61490764 60197176R 60437843R 61183922R 365789 368420 60108249 60131304 60159588 60208566 60393482 60426242 60445136 60715132 60837654 61183919 61307937 61388033 60131304R 60393482R 60426242R 60445136R 367894 368421 60140177 60837655 61194163 61285988 61327781 61388034 61468145 00367894R 61194163R 61388034R 368422 60759282 60837656 61277988 61307938 61324290 61441063 00368422R 60759282R 60837656R 61277988R 61324290R 368423 60933759 61120921 61217342 61327782 61415326 00368423R 61120921R 61217342R 61327782R 61415326R 368424 61120922 61151165 61320258 61375279 61415327 61120922R 61151165R 367895 369106 60266266 60282793 60282794 61162319 61186090 61194164 61224476 61249734 61307928 61329327 61340903 61382424 61453186 61747595 60282793R 61162319R 61249734R 365811 367302 367850 368253 60093704 60093705 60130259 60165709 60197179 60208567 60241657 60260271 60282795 60601773 60630455 60894689 60894690 61128222 61143369 61172771 61194165 61203635 61224477 61269511 61298486 61347483 61356897 61368160 61373301 61377348 61390163 61409977 61409978 60093705R 60208567R 367231 368254 60072090 60080863 60093706 60093707 60140178 60173950 60175578 60187052 60187053 60208568 60208569 60241658 60266269 60282797 60282798 60437845 60443938 60445138 60445139 60536768 60573215 60592032 60601774 60601775 60645073 60672020 60695523 60715133 60758128 60789477 60913743 60913744 60913745 61117766 61140760 61172772 61172774 61187596 61194167 61194168 61203640 61209937 61209940 61373304 61377349 61390164 61409971 61409973 61409974 61530391 61597133 61612597 61747596 60175578R 60208569R 60443938R 60445139R 60601775R 60695523R 367721 369827 60080864 60080865 60104818 60127954 60131306 60138538 60140179 60173951 60208575 60208576 60247133 60266273 60266274 60282799 60325582 60325583 60423189 60437846 60443939 60445140 60445141 60550487 60582875 60582876 60622430 60695525 60814009 60814010 60913722 60913723 60937748 61049705 61113689 61128223 61146965 61187597 61224484 61249102 61263735 61269514 61281879 61379614 61422223 61422224 61434771 61468146 61490765 61509901 61747597 60437846R 60814010R 61382425R 368425 368426 60080866 60123962 60135471 60165710 60173952 60219124 60266272 60282801 60429256 60443941 60445142 60445143 60583692 60592034 60622431 60645062 60695526 60779104 60779105 61183912 61187607 61194212 61249739 61281881 61294516 61327551 61368167 61382426 61468147 60266272R 60645062R 60695526R 60779104R 61468147R 368427 60072092 60208577 60443942 60445144 61327783 61434772 60208577R 60443942R 60445144R 368428 369828 60072093 60156413 60670030 60837657 60848481 61458952 60072093R 60156413R 60670030R 60837657R 368429 60072094 60738224 61244786 61263725 61307940 61327784 60072094R 60738224R 61244786R 61263725R 61327784R 368430 60730286 61101499 61140761 61290181 61415332 00368430R 61140761R 368333 369109 60378682 60423632 61214445 61276091 61307941 61344242 60423632R 61214445R 61276091R 367851 368256 368432 60127955 60260272 60401120 60708941 61069571 61187608 61194221 61203641 61217348 61224487 61249740 61329329 61347491 61347492 61373307 61458954 60127955R 60260272R 60401120R 368257 60127956 60140180 60333386 60515436 60573217 60582877 60711452 60821410 60931684 61187609 61194222 61203642 61249741 61269516 61281885 61320163 61394191 61434773 61475209 61523845 60127956R 368258 60093708 60138540 60208578 60297195 60550215 60567045 60733118 60806380 61155637 61187610 61194224 61224491 61249744 61317732 61329330 61346040 61368170 61379616 61388035 61490766 60550215R 61194224R 368393 60053581 60154671 60348770 60472301 60747418 60821413 61162316 61194225 61224492 61269517 61320164 61327553 61382427 60053581R 60154671R 60348770R 60472301R 60821413R 61194225R 61382427R 368433 60154672 60531299 60729914 60973738 60973739 60154672R 60531299R 368434 60154673 60748950 60750986 61277989 61327785 60748950R 60750986R 368394 60072097 61170168 61277990 61307942 61332786 61344243 00368394R 60072097R 61170168R 61277990R 61307942R 368262 369532 60933761 61069574 61187611 61276092 61382428 61621125 00367787R 60933761R 61069574R 61187611R 365814 368259 368435 369110 60161918 60708942 60933762 61217351 61224493 61249745 61281886 61356910 61368171 61377351 61441064 61475210 61523848 00368435R 61224493R 61249745R 367852 368260 60074254 60378683 60602192 60685552 60789437 60806382 61136722 61194234 61217352 61224494 61249746 61281887 61356912 61368172 61377352 60074254R 60602192R 60789437R 61194234R 368263 60154674 60378684 60426971 60617709 61079863 61184357 61187612 61276111 61332787 61351688 61428544 00368263R 60426971R 60617709R 61079863R 61184357R 61428544R 368436 60067835 60154675 61095372 61120924 61276121 61315093 61327786 61382429 61458956 00368436R 60154675R 61095372R 61315093R 61327786R 60067836 60541593 60067836R 60541593R 368438 60665010 60848484 61267999 61327787 60665010R 60848484R 368439 60173953 60848485 61343281 60173953R 60848485R 368328 369112 60138541 60588707 60670031 60933763 61276103 61277885 61351689 61389807 00368328R 60138541R 60588707R 60670031R 61276103R 368395 60708943 61079864 61187613 61341150 61490767 61641826 00368395R 60708943R 61079864R 61187613R 61341150R 368329 60154676 61040188 61048033 61276123 61341151 60154676R 61040188R 61048033R 368440 60138542 60985622 60985623 60985624 60985625 00368440R 60138542R 60985622R 365799 368441 60689046 60753362 60848488 61188445 00368441R 60689046R 60848488R 368396 369114 60067837 60753363 60848489 61188447 61490768 00368396R 00369114R 60067837R 60848489R 369115 60689048 60933764 61143370 61276112 61351690 60689048R 61143370R 61276112R 61351690R 368442 60984603 61003748 61040177 61040178 61040179 61048029 00368442R 61003748R 61040179R 368331 60973733 60976652 61382431 61621127 60973733R 61382431R 
Recalling Firm/
Manufacturer
Zimmer, Inc.
345 E Main St
Warsaw IN 46580-2746
For Additional Information Contact
800-613-6131
Manufacturer Reason
for Recall
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
FDA Determined
Cause 2
Packaging
ActionDistributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.
Quantity in Commerce7,044,680 total
DistributionWorldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = HSB
-
-