|
Class 2 Device Recall StandAlone Resuscitation Unit, Bag and Mask |
|
Date Initiated by Firm |
December 12, 2013 |
Date Posted |
February 03, 2014 |
Recall Status1 |
Terminated 3 on June 01, 2015 |
Recall Number |
Z-0914-2014 |
Recall Event ID |
67225 |
510(K)Number |
K070247
|
Product Classification |
Ventilator, emergency, manual (resuscitator) - Product Code BTM
|
Product |
Stand-Alone Resuscitation Unit, Bag and Mask, part number M1226444, labeled in part ***GE Healthcare 8880 Gorman Road Laurel, MD 20723***www.gehealthcare.com***.
Provides the basic equipment required for pulmonary resuscitation of infants. |
Code Information |
Serial Numbers: QABS50026 QABR50066 QABR50068 QABR50133 QABS50023 QABR50102 QABR50103 QABR50064 QABR50065 QABR50067 QABS50070 QABS50069 QABR50275 QABS50027 QABR50274 |
Recalling Firm/ Manufacturer |
GE Healthcare 8880 Gorman Rd Laurel MD 20723-5800
|
Manufacturer Reason for Recall |
Medical device adjustment knob may be loose and affect the accuracy of blended Air/Oxygen gas mixture to patients.
|
FDA Determined Cause 2 |
Process control |
Action |
GE Healthcare notified end users with a letter flagged as Urgent Medical Device Correction letter on 12/12/2013. The notification advised that GE Healthcare has become aware of a potential safety issue associated with the loosening of the blender knob shaft on all Resuscitation units. The firm advised that accounts ensure that all potential users in their facility are made aware of this safety notification and the recommended actions described in the notification:
1. Perform blender accuracy check as outlined in Section 2.5-2.5.1
of the Service Manual to verify that the O2 concentration set at
21% is within 16-26% O2 and the concentration at 100% setting
is within the recommended range (95-105% O2).
2. If the blender accuracy test results do not meet specifications,
remove the unit from patient use and contact a GE Healthcare
Service Representative at the number listed below.
3. If the blender accuracy check passes in Step 1 above, you may
continue to use your resuscitation system. The following precautions
are recommended to ensure blender accuracy is maintained between
system checks.
¿ Use an independent oxygen analyzer, as determined by your hospital
policy, to check blended Air/O2 accuracy.
¿ When delivering Oxygen, use a pulse oximeter to monitor your patient.
Customers were requested to contact a GE Healthcare Service Representative at 1-800-345-2700 to arrange for correction if the device is found to be defective. |
Quantity in Commerce |
15 units |
Distribution |
Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Uruguay, Venezuela, Albania, Algeria, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Canada, Chile, China, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Kazakhstan, Korea, Kuwait, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Arab Emirates, and United Kingdom. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = BTM and Original Applicant = OHMEDA MEDICAL
|
|
|
|