|
Class 2 Device Recall PCA 3 |
|
Date Initiated by Firm |
March 08, 2013 |
Create Date |
July 08, 2016 |
Recall Status1 |
Terminated 3 on November 07, 2016 |
Recall Number |
Z-2140-2016 |
Recall Event ID |
68318 |
510(K)Number |
K043256
|
Product Classification |
Pump, infusion, pca - Product Code MEA
|
Product |
The PCA 3 system cam be used in a wide range of clinical settings, including but not limited to: general floor, labor/delivery,post-partum, burn unit, medical surgical, operating room, oncology, critical care units, post-anesthesia care unit (PACU), and pediatrics. The PCA 3 infusion pump allows clinicians to administer or patients to self-administer analgesia safely and effectively within clinician programmed limits. The epidural route can be used to provide anesthesia or analgesia. |
Code Information |
List No. 12384; Serial Numbers: 12920041 ; 12920048 ; 12920052 ; 12920056 ; 12920066 ; 12920069 ; 12920079 ; 12920082 ; 12920101 ; 12920102 ; 12920104 ; 12920105 ; 12920106 ; 12920107 ; 12920117 ; 12920121 ; 12920122 ; 12920152 ; 12920158 ; 12920159 ; 12920160 ; 12920174 ; 12920246 ; 12920248 ; 12920251 ; 12920252 ; 12920255 ; 12920257 ; 12920264 ; 12920265 ; 12920268 ; 12920270 ; 12920275 ; 12920276 ; 12920277 ; 12920278 ; 12920282 ; 12920287 ; 12920289 ; 12920291 ; 12920294 ; 12920304 ; 12920305 ; 12920311 ; 12920312 ; 12920320 ; 12920328 ; 12920330 ; 12920332 ; 12920340 ; 12920343 ; 12920346 ; 12920347 ; 12920352 ; 12920353 ; 12920357 ; 12920358 ; 12920359 ; 12920360 ; 12920362 ; 12920364 ; 12920369 ; 12920371 ; 12920372 ; 12920377 ; 12920381 ; 12920382 ; 12920383 ; 12920384 ; 12920387 ; 12920388 ; 12920390 ; 12920392 ; 12920398 ; 12920401 ; 12920403 ; 12920404 ; 12920405 ; 12920410 ; 12920412 ; 12920413 ; 12920415 ; 12920416 ; 12920420 ; 12920424 ; 12920425 ; 12920428 ; 12920429 ; 12920432 ; 12920433 ; 12920434 ; 12920435 ; 12920436 ; 12920441 ; 12920444 ; 12920446 ; 12920448 ; 12920449 ; 12920453 ; 12920455 ; 12920457 ; 12920458 ; 12920460 ; 12920461 ; 12920462 ; 12920466 ; 12920467 ; 12920468 ; 12920469 ; 12920472 ; 12920473 ; 12920476 ; 12920477 ; 12920478 ; 12920480 ; 12920481 ; 12920482 ; 12920484 ; 12920485 ; 12920486 ; 12920487 ; 12920488 ; 12920489 ; 12920490 ; 12920492 ; 12920496 ; 12920497 ; 12920498 ; 12920499 ; 12920501 ; 12920505 ; 12920506 ; 12920507 ; 12920508 ; 12920509 ; 12920510 ; 12920514 ; 12920516 ; 12920517 ; 12920518 ; 12920519 ; 12920520 ; 12920521 ; 12920522 ; 12920523 ; 12920524 ; 12920525 ; 12920526 ; 12920530 ; 12920533 ; 12920534 ; 12920535 ; 12920537 ; 12920540 ; 12920541 ; 12920542 ; 12920544 ; 12920545 ; 12920546 ; 12920548 ; 12920555 ; 12920558 ; 12920559 ; 12920560 ; 12920562 ; 12920565 ; 12920566 ; 12920567 ; 12920568 ; 12920569 ; 12920570 ; 12920571 ; 12920574 ; 12920576 ; 12920577 ; 12920578 ; 12920581 ; 12920582 ; 12920585 ; 12920586 ; 12920587 ; 12920589 ; 12920590 ; 12920591 ; 12920592 ; 12920593 ; 12920596 ; 12920598 ; 12920599 ; 12920600 ; 12920601 ; 12920602 ; 12920603 ; 12920604 ; 12920605 ; 12920606 ; 12920607 ; 12920608 ; 12920610 ; 12920611 ; 12920612 ; 12920614 ; 12920615 ; 12920620 ; 12920621 ; 12920623 ; 12920626 ; 12920627 ; 12920629 ; 12920632 ; 12920633 ; 12920634 ; 12920635 ; 12920639 ; 12920641 ; 12920644 ; 12920646 ; 12920647 ; 12920648 ; 12920649 ; 12920651 ; 12920654 ; 12920657 ; 12920659 ; 12920660 ; 12920663 ; 12920664 ; 12920665 ; 12920666 ; 12920668 ; 12920669 ; 12920670 ; 12920671 ; 12920674 ; 12920675 ; 12920677 ; 12920679 ; 12920680 ; 12920681 ; 12920682 ; 12920686 ; 12920688 ; 12920690 ; 12920691 ; 12920692 ; 12920696 ; 12920699 ; 12920702 ; 12920703 ; 12920706 ; 12920708 ; 12920710 ; 12920714 ; 12920715 ; 12920716 ; 12920717 ; 12920720 ; 12920724 ; 12920726 ; 12920727 ; 12920728 ; 12920729 ; 12920731 ; 12920733 ; 12920734 ; 12920736 ; 12920743 ; 12920744 ; 12920746 ; 12920747 ; 12920749 ; 12920750 ; 12920756 ; 12920759 ; 12920761 ; 12920762 ; 12920763 ; 12920766 ; 12920767 ; 12920770 ; 12920771 ; 12920773 ; 12920774 ; 12920778 ; 12920779 ; 12920782 ; 12920783 ; 12920788 ; 12920792 ; 12920793 ; 12920795 ; 12920796 ; 12920797 ; 12920798 ; 12920800 ; 12920802 ; 12920808 ; 12920809 ; 12920811 ; 12920813 ; 12920815 ; 12920819 ; 12920820 ; 12920821 ; 12920826 ; 12920827 ; 12920828 ; 12920834 ; 12920836 ; 12920837 ; 12920839 ; 12920840 ; 12920841 ; 12920844 ; 12920848 ; 12920850 ; 12920851 ; 12920852 ; 12920853 ; 12920856 ; 12920857 ; 12920858 ; 12920859 ; 12920861 ; 12920863 ; 12920865 ; 12920866 ; 12920868 ; 12920871 ; 12920872 ; 12920873 ; 12920874 ; 12920875 ; 12920879 ; 12920880 ; 12920881 ; 12920884 ; 12920885 ; 12920887 ; 12920888 ; 12920890 ; 12920891 ; 12920894 ; 12920895 ; 12920896 ; 12920900 ; 12920902 ; 12920903 ; 12920904 ; 12920908 ; 12920909 ; 12920910 ; 12920911 ; 12920916 ; 12920917 ; 12920918 ; 12920924 ; 12920925 ; 12920926 ; 12920928 ; 12920931 ; 12920932 ; 12920933 ; 12920935 ; 12920936 ; 12920938 ; 12920940 ; 12920941 ; 12920945 ; 12920946 ; 12920947 ; 12920948 ; 12920950 ; 12920951 ; 12920954 ; 12920956 ; 12920957 ; 12920958 ; 12920960 ; 12920961 ; 12920963 ; 12920966 ; 12920968 ; 12920969 ; 12920970 ; 12920971 ; 12920973 ; 12920974 ; 12920978 ; 12920979 ; 12920980 ; 12920982 ; 12920983 ; 12920984 ; 12920986 ; 12920990 ; 12920991 ; 12920997 ; 12921003 ; 12921005 ; 12921006 ; 12921008 ; 12921010 ; 12921011 ; 12921012 ; 12921017 ; 12921018 ; 12921020 ; 12921021 ; 12921023 ; 12921024 ; 12921025 ; 12921026 ; 12921030 ; 12921032 ; 12921040 ; 12921042 ; 12921044 ; 12921050 ; 12921053 ; 12921054 ; 12921055 ; 12921056 ; 12921057 ; 12921060 ; 12921063 ; 12921064 ; 12921065 ; 12921066 ; 12921067 ; 12921069 ; 12921070 ; 12921072 ; 12921074 ; 12921076 ; 12921082 ; 12921083 ; 12921084 ; 12921086 ; 12921090 ; 12921091 ; 12921093 ; 12921095 ; 12921096 ; 12921097 ; 12921098 ; 12921099 ; 12921100 ; 12921102 ; 12921106 ; 12921107 ; 12921109 ; 12921110 ; 12921111 ; 12921115 ; 12921116 ; 12921117 ; 12921118 ; 12921119 ; 12921121 ; 12921123 ; 12921124 ; 12921125 ; 12921126 ; 12921128 ; 12921130 ; 12921135 ; 12921137 ; 12921142 ; 12921144 ; 12921145 ; 12921146 ; 12921147 ; 12921151 ; 12921155 ; 12921158 ; 12921159 ; 12921160 ; 12921161 ; 12921163 ; 12921168 ; 12921172 ; 12921173 ; 12921174 ; 12921175 ; 12921177 ; 12921179 ; 12921180 ; 12921181 ; 12921183 ; 12921184 ; 12921185 ; 12921186 ; 12921187 ; 12921188 ; 12921189 ; 12921192 ; 12921193 ; 12921194 ; 12921195 ; 12921197 ; 12921199 ; 12921202 ; 12921204 ; 12921205 ; 12921207 ; 12921212 ; 12921217 ; 12921219 ; 12921221 ; 12921223 ; 12921226 ; 12921227 ; 12921230 ; 12921232 ; 12921234 ; 12921235 ; 12921236 ; 12921238 ; 12921239 ; 12921241 ; 12921242 ; 12921243 ; 12921247 ; 12921249 ; 12921250 ; 12921254 ; 12921255 ; 12921256 ; 12921259 ; 12921260 ; 12921262 ; 12921263 ; 12921264 ; 12921265 ; 12921266 ; 12921267 ; 12921269 ; 12921270 ; 12921271 ; 12921272 ; 12921273 ; 12921274 ; 12921276 ; 12921280 ; 12921281 ; 12921282 ; 12921288 ; 12921290 ; 12921294 ; 12921296 ; 12921297 ; 12921298 ; 12921299 ; 12921300 ; 12921302 ; 12921304 ; 12921305 ; 12921306 ; 12921307 ; 12921309 ; 12921311 ; 12921312 ; 12921314 ; 12921315 ; 12921316 ; 12921317 ; 12921319 ; 12921324 ; 12921325 ; 12921326 ; 12921328 ; 12921330 ; 12921331 ; 12921332 ; 12921335 ; 12921338 ; 12921340 ; 12921342 ; 12921343 ; 12921344 ; 12921347 ; 12921348 ; 12921349 ; 12921351 ; 12921352 ; 12921353 ; 12921354 ; 12921356 ; 12921359 ; 12921360 ; 12921361 ; 12921363 ; 12921366 ; 12921368 ; 12921370 ; 12921372 ; 12921373 ; 12921374 ; 12921376 ; 12921377 ; 12921378 ; 12921379 ; 12921380 ; 12921381 ; 12921385 ; 12921386 ; 12921387 ; 12921389 ; 12921390 ; 12921391 ; 12921392 ; 12921394 ; 12921395 ; 12921396 ; 12921399 ; 12921400 ; 12921403 ; 12921404 ; 12921405 ; 12921406 ; 12921408 ; 12921410 ; 12921413 ; 12921414 ; 12921415 ; 12921417 ; 12921418 ; 12921419 ; 12921420 ; 12921421 ; 12921422 ; 12921423 ; 12921425 ; 12921426 ; 12921427 ; 12921428 ; 12921432 ; 12921439 ; 12921441 ; 12921442 ; 12921444 ; 12921445 ; 12921446 ; 12921447 ; 12921448 ; 12921449 ; 12921450 ; 12921451 ; 12921459 ; 12921462 ; 12921463 ; 12921469 ; 12921475 ; 12921477 ; 12921478 ; 12921479 ; 12921480 ; 12921482 ; 12921483 ; 12921485 ; 12921486 ; 12921487 ; 12921488 ; 12921489 ; 12921490 ; 12921491 ; 12921493 ; 12921495 ; 12921497 ; 12921498 ; 12921499 ; 12921500 ; 12921501 ; 12921502 ; 12921504 ; 12921507 ; 12921508 ; 12921509 ; 12921510 ; 12921512 ; 12921514 ; 12921517 ; 12921519 ; 12921521 ; 12921525 ; 12921527 ; 12921530 ; 12921531 ; 12921532 ; 12921533 ; 12921534 ; 12921535 ; 12921538 ; 12921539 ; 12921540 ; 12921541 ; 12921542 ; 12921544 ; 12921545 ; 12921546 ; 12921547 ; 12921548 ; 12921550 ; 12921551 ; 12921552 ; 12921554 ; 12921555 ; 12921556 ; 12921558 ; 12921560 ; 12921561 ; 12921562 ; 12921563 ; 12921564 ; 12921565 ; 12921569 ; 12921570 ; 12921576 ; 12921577 ; 12921578 ; 12921580 ; 12921583 ; 12921584 ; 12921585 ; 12921587 ; 12921588 ; 12921591 ; 12921592 ; 12921597 ; 12921603 ; 12921606 ; 12921607 ; 12921613 ; 12921615 ; 12921616 ; 12921618 ; 12921621 ; 12921626 ; 12921636 ; 12921637 ; 12921638 ; 12921639 ; 12921640 ; 12921641 ; 12921643 ; 12921644 ; 12921646 ; 12921647 ; 12921648 ; 12921650 ; 12921652 ; 12921654 ; 12921655 ; 12921656 ; 12921657 ; 12921658 ; 12921661 ; 12921663 ; 12921665 ; 12921669 ; 12921671 ; 12921672 ; 12921676 ; 12921680 ; 12921686 ; 12921689 ; 12921690 ; 12921693 ; 12921694 ; 12921695 ; 12921700 ; 12921701 ; 12921702 ; 12921703 ; 12921704 ; 12921705 ; 12921706 ; 12921710 ; 12921712 ; 12921716 ; 12921717 ; 12921719 ; 12921720 ; 12921721 ; 12921723 ; 12921724 ; 12921728 ; 12921730 ; 12921733 ; 12921734 ; 12921738 ; 12921748 ; 12921750 ; 12921756 ; 12921757 ; 12921761 ; 12921769 ; 12921770 ; 12921774 ; 12921776 ; 12921777 ; 12921778 ; 12921784 ; 12921785 ; 12921786 ; 12921787 ; 12921788 ; 12921789 ; 12921791 ; 12921794 ; 12921797 ; 12921800 ; 12921803 ; 12921805 ; 12921806 ; 12921807 ; 12921812 ; 12921814 ; 12921817 ; 12921818 ; 12921819 ; 12921824 ; 12921825 ; 12921827 ; 12921828 ; 12921829 ; 12921830 ; 12921831 ; 12921833 ; 12921834 ; 12921839 ; 12921840 ; 12921841 ; 12921846 ; 12921850 ; 12921851 ; 12921854 ; 12921855 ; 12921857 ; 12921858 ; 12921860 ; 12921861 ; 12921862 ; 12921863 ; 12921864 ; 12921866 ; 12921867 ; 12921868 ; 12921871 ; 12921872 ; 12921876 ; 12921877 ; 12921878 ; 12921881 ; 12921882 ; 12921883 ; 12921886 ; 12921889 ; 12921892 ; 12921895 ; 12921896 ; 12921897 ; 12921898 ; 12921902 ; 12921908 ; 12921912 ; 12921913 ; 12921919 ; 12921924 ; 12921925 ; 12921926 ; 12921929 ; 12921931 ; 12921932 ; 12921938 ; 12921941 ; 12921942 ; 12921944 ; 12921947 ; 12921948 ; 12921950 ; 12921951 ; 12921952 ; 12921953 ; 12921956 ; 12921957 ; 12921958 ; 12921963 ; 12921964 ; 12921965 ; 12921967 ; 12921968 ; 12921972 ; 12921975 ; 12921976 ; 12921977 ; 12921978 ; 12921983 ; 12921984 ; 12921985 ; 12921986 ; 12921987 ; 12921990 ; 12921995 ; 12921998 ; 12922001 ; 12922020 ; 12922024 ; 12922030 ; 12922034 ; 12922036 ; 12922041 ; 12922043 ; 12922044 ; 12922045 ; 12922047 ; 12922051 ; 12922052 ; 12922054 ; 12922055 ; 12922056 ; 12922058 ; 12922059 ; 12922060 ; 12922061 ; 12922062 ; 12922065 ; 12922066 ; 12922067 ; 12922069 ; 12922071 ; 12922072 ; 12922074 ; 12922075 ; 12922076 ; 12922077 ; 12922078 ; 12922079 ; 12922081 ; 12922083 ; 12922084 ; 12922085 ; 12922087 ; 12922088 ; 12922092 ; 12922093 ; 12922094 ; 12922096 ; 12922097 ; 12922098 ; 12922099 ; 12922100 ; 12922101 ; 12922103 ; 12922105 ; 12922106 ; 12922107 ; 12922108 ; 12922110 ; 12922111 ; 12922113 ; 12922114 ; 12922116 ; 12922117 ; 12922119 ; 12922120 ; 12922121 ; 12922122 ; 12922123 ; 12922126 ; 12922127 ; 12922128 ; 12922129 ; 12922131 ; 12922132 ; 12922133 ; 12922135 ; 12922139 ; 12922141 ; 12922142 ; 12922143 ; 12922144 ; 12922145 ; 12922148 ; 12922150 ; 12922153 ; 12922155 ; 12922156 ; 12922157 ; 12922158 ; 12922161 ; 12922162 ; 12922163 ; 12922164 ; 12922165 ; 12922167 ; 12922168 ; 12922170 ; 12922171 ; 12922172 ; 12922174 ; 12922175 ; 12922176 ; 12922177 ; 12922178 ; 12922179 ; 12922180 ; 12922181 ; 12922182 ; 12922183 ; 12922184 ; 12922185 ; 12922186 ; 12922187 ; 12922188 ; 12922189 ; 12922190 ; 12922191 ; 12922192 ; 12922193 ; 12922194 ; 12922195 ; 12922196 ; 12922198 ; 12922199 ; 12922200 ; 12922201 ; 12922202 ; 12922203 ; 12922204 ; 12922205 ; 12922206 ; 12922207 ; 12922208 ; 12922209 ; 12922210 ; 12922211 ; 12922212 ; 12922213 ; 12922215 ; 12922216 ; 12922219 ; 12922220 ; 12922221 ; 12922222 ; 12922224 ; 12922225 ; 12922226 ; 12922227 ; 12922229 ; 12922230 ; 12922232 ; 12922233 ; 12922235 ; 12922236 ; 12922240 ; 12922241 ; 12922242 ; 12922244 ; 12922250 ; 12922251 ; 12922252 ; 12922253 ; 12922254 ; 12922256 ; 12922257 ; 12922258 ; 12922260 ; 12922261 ; 12922262 ; 12922263 ; 12922264 ; 12922266 ; 12922267 ; 12922269 ; 12922270 ; 12922272 ; 12922273 ; 12922274 ; 12922277 ; 12922278 ; 12922279 ; 12922280 ; 12922281 ; 12922283 ; 12922284 ; 12922285 ; 12922286 ; 12922288 ; 12922289 ; 12922290 ; 12922292 ; 12922293 ; 12922294 ; 12922295 ; 12922297 ; 12922298 ; 12922299 ; 12922300 ; 12922301 ; 12922302 ; 12922303 ; 12922304 ; 12922306 ; 12922307 ; 12922309 ; 12922310 ; 12922312 ; 12922313 ; 12922314 ; 12922315 ; 12922316 ; 12922317 ; 12922318 ; 12922319 ; 12922320 ; 12922321 ; 12922322 ; 12922324 ; 12922325 ; 12922326 ; 12922327 ; 12922331 ; 12922333 ; 12922334 ; 12922336 ; 12922338 ; 12922340 ; 12922341 ; 12922342 ; 12922343 ; 12922344 ; 12922345 ; 12922346 ; 12922348 ; 12922350 ; 12922351 ; 12922352 ; 12922353 ; 12922356 ; 12922357 ; 12922358 ; 12922359 ; 12922360 ; 12922361 ; 12922363 ; 12922366 ; 12922367 ; 12922368 ; 12922369 ; 12922370 ; 12922371 ; 12922373 ; 12922376 ; 12922378 ; 12922380 ; 12922382 ; 12922383 ; 12922384 ; 12922385 ; 12922387 ; 12922388 ; 12922389 ; 12922393 ; 12922395 ; 12922398 ; 12922404 ; 12922405 ; 12922406 ; 12922412 ; 12922420 ; 12922421 ; 12922422 ; 12922424 ; 12922426 ; 12922427 ; 12922429 ; 12922430 ; 12922433 ; 12922434 ; 12922436 ; 12922437 ; 12922439 ; 12922440 ; 12922442 ; 12922444 ; 12922450 ; 12922451 ; 12922452 ; 12922456 ; 12922461 ; 12922463 ; 12922464 ; 12922465 ; 12922466 ; 12922468 ; 12922470 ; 12922471 ; 12922472 ; 12922474 ; 12922475 ; 12922477 ; 12922478 ; 12922481 ; 12922482 ; 12922483 ; 12922485 ; 12922486 ; 12922487 ; 12922489 ; 12922494 ; 12922495 ; 12922496 ; 12922497 ; 12922498 ; 12922501 ; 12922507 ; 12922509 ; 12922510 ; 12922514 ; 12922515 ; 12922520 ; 12922523 ; 12922526 ; 12922527 ; 12922528 ; 12922532 ; 12922534 ; 12922536 ; 12922537 ; 12922539 ; 12922540 ; 12922542 ; 12922543 ; 12922546 ; 12922547 ; 12922548 ; 12922549 ; 12922551 ; 12922552 ; 12922554 ; 12922555 ; 12922556 ; 12922558 ; 12922559 ; 12922560 ; 12922561 ; 12922562 ; 12922563 ; 12922564 ; 12922567 ; 12922569 ; 12922571 ; 12922572 ; 12922573 ; 12922575 ; 12922576 ; 12922580 ; 12922581 ; 12922582 ; 12922584 ; 12922585 ; 12922589 ; 12922592 ; 12922593 ; 12922596 ; 12922597 ; 12922602 ; 12922607 ; 12922608 ; 12922609 ; 12922610 ; 12922612 ; 12922614 ; 12922616 ; 12922617 ; 12922619 ; 12922620 ; 12922622 ; 12922623 ; 12922625 ; 12922631 ; 12922635 ; 12922636 ; 12922639 ; 12922640 ; 12922641 ; 12922645 ; 12922646 ; 12922647 ; 12922648 ; 12922649 ; 12922650 ; 12922654 ; 12922658 ; 12922659 ; 12922660 ; 12922663 ; 12922664 ; 12922665 ; 12922668 ; 12922669 ; 12922675 ; 12922677 ; 12922678 ; 12922679 ; 12922680 ; 12922683 ; 12922686 ; 12922689 ; 12922690 ; 12922691 ; 12922693 ; 12922695 ; 12922696 ; 12922697 ; 12922698 ; 12922699 ; 12922700 ; 12922701 ; 12922702 ; 12922703 ; 12922704 ; 12922705 ; 12922708 ; 12922710 ; 12922711 ; 12922712 ; 12922713 ; 12922714 ; 12922715 ; 12922715 ; 12922715 ; 12922716 ; 12922717 ; 12922718 ; 12922719 ; 12922720 ; 12922721 ; 12922722 ; 12922724 ; 12922725 ; 12922726 ; 12922727 ; 12922731 ; 12922732 ; 12922736 ; 12922737 ; 12922738 ; 12922741 ; 12922742 ; 12922743 ; 12922744 ; 12922745 ; 12922749 ; 12922750 ; 12922751 ; 12922754 ; 12922755 ; 12922756 ; 12922757 ; 12922758 ; 12922759 ; 12922761 ; 12922763 ; 12922765 ; 12922766 ; 12922767 ; 12922771 ; 12922773 ; 12922776 ; 12922780 ; 12922782 ; 12922783 ; 12922786 ; 12922788 ; 12922791 ; 12922795 ; 12922796 ; 12922798 ; 12922799 ; 12922800 ; 12922801 ; 12922802 ; 12922803 ; 12922805 ; 12922806 ; 12922808 ; 12922809 ; 12922810 ; 12922811 ; 12922813 ; 12922815 ; 12922816 ; 12922817 ; 12922818 ; 12922820 ; 12922821 ; 12922823 ; 12922824 ; 12922825 ; 12922826 ; 12922827 ; 12922828 ; 12922829 ; 12922830 ; 12922832 ; 12922833 ; 12922834 ; 12922835 ; 12922837 ; 12922838 ; 12922841 ; 12922842 ; 12922843 ; 12922846 ; 12922847 ; 12922848 ; 12922850 ; 12922852 ; 12922854 ; 12922861 ; 12922863 ; 12922864 ; 12922867 ; 12922872 ; 12922873 ; 12922884 ; 12922885 ; 12922887 ; 12922889 ; 12922891 ; 12922892 ; 12922893 ; 12922896 ; 12922897 ; 12922910 ; 12922915 ; 12922918 ; 12922919 ; 12922921 ; 12922922 ; 12922926 ; 12922927 ; 12922930 ; 12922931 ; 12922933 ; 12922937 ; 12922939 ; 12922945 ; 12922947 ; 12922949 ; 12922950 ; 12922951 ; 12922952 ; 12922953 ; 12922955 ; 12922956 ; 12922957 ; 12922958 ; 12922960 ; 12922962 ; 12922964 ; 12922965 ; 12922966 ; 12922967 ; 12922968 ; 12922971 ; 12922972 ; 12922973 ; 12922975 ; 12922976 ; 12922977 ; 12922978 ; 12922979 ; 12922980 ; 12922981 ; 12922982 ; 12922984 ; 12922985 ; 12922986 ; 12922987 ; 12922988 ; 12922989 ; 12922990 ; 12922991 ; 12922993 ; 12922994 ; 12922995 ; 12922998 ; 12923001 ; 12923002 ; 12923005 ; 12923006 ; 12923007 ; 12923008 ; 12923012 ; 12923015 ; 12923017 ; 12923020 ; 12923021 ; 12923023 ; 12923027 ; 12923030 ; 12923031 ; 12923032 ; 12923033 ; 12923035 ; 12923036 ; 12923040 ; 12923042 ; 12923049 ; 12923051 ; 12923055 ; 12923057 ; 12923058 ; 12923059 ; 12923062 ; 12923063 ; 12923064 ; 12923065 ; 12923067 ; 12923068 ; 12923069 ; 12923072 ; 12923073 ; 12923074 ; 12923075 ; 12923076 ; 12923077 ; 12923079 ; 12923080 ; 12923081 ; 12923082 ; 12923083 ; 12923084 ; 12923085 ; 12923087 ; 12923090 ; 12923091 ; 12923092 ; 12923095 ; 12923096 ; 12923097 ; 12923098 ; 12923101 ; 12923102 ; 12923103 ; 12923105 ; 12923106 ; 12923108 ; 12923109 ; 12923110 ; 12923111 ; 12923114 ; 12923115 ; 12923116 ; 12923117 ; 12923118 ; 12923120 ; 12923121 ; 12923122 ; 12923124 ; 12923125 ; 12923126 ; 12923127 ; 12923128 ; 12923129 ; 12923132 ; 12923133 ; 12923134 ; 12923136 ; 12923137 ; 12923138 ; 12923139 ; 12923140 ; 12923142 ; 12923143 ; 12923144 ; 12923146 ; 12923147 ; 12923148 ; 12923149 ; 12923150 ; 12923152 ; 12923153 ; 12923155 ; 12923156 ; 12923157 ; 12923158 ; 12923159 ; 12923160 ; 12923161 ; 12923162 ; 12923163 ; 12923164 ; 12923167 ; 12923168 ; 12923169 ; 12923171 ; 12923172 ; 12923173 ; 12923174 ; 12923177 ; 12923178 ; 12923179 ; 12923181 ; 12923183 ; 12923184 ; 12923186 ; 12923187 ; 12923189 ; 12923190 ; 12923191 ; 12923192 ; 12923193 ; 12923194 ; 12923195 ; 12923197 ; 12923199 ; 12923201 ; 12923202 ; 12923204 ; 12923205 ; 12923206 ; 12923207 ; 12923208 ; 12923211 ; 12923212 ; 12923216 ; 12923219 ; 12923221 ; 12923224 ; 12923228 ; 12923230 ; 12923235 ; 12923237 ; 12923241 ; 12923244 ; 12923247 ; 12923251 ; 12923256 ; 12923258 ; 12923259 ; 12923260 ; 12923262 ; 12923264 ; 12923265 ; 12923269 ; 12923271 ; 12923272 ; 12923274 ; 12923275 ; 12923277 ; 12923278 ; 12923279 ; 12923282 ; 12923284 ; 12923285 ; 12923286 ; 12923287 ; 12923288 ; 12923289 ; 12923292 ; 12923293 ; 12923295 ; 12923297 ; 12923299 ; 12923300 ; 12923301 ; 12923302 ; 12923303 ; 12923304 ; 12923305 ; 12923308 ; 12923311 ; 12923312 ; 12923314 ; 12923315 ; 12923316 ; 12923322 ; 12923323 ; 12923332 ; 12923333 ; 12923336 ; 12923340 ; 12923342 ; 12923344 ; 12923350 ; 12923353 ; 12923355 ; 12923356 ; 12923357 ; 12923360 ; 12923373 ; 12923378 ; 12923379 ; 12923382 ; 12923383 ; 12923385 ; 12923388 ; 12923389 ; 12923396 ; 12923402 ; 12923405 ; 12923406 ; 12923408 ; 12923414 ; 12923415 ; 12923416 ; 12923417 ; 12923418 ; 12923420 ; 12923421 ; 12923423 ; 12923425 ; 12923426 ; 12923427 ; 12923428 ; 12923431 ; 12923433 ; 12923435 ; 12923437 ; 12923438 ; 12923440 ; 12923441 ; 12923446 ; 12923447 ; 12923448 ; 12923451 ; 12923452 ; 12923453 ; 12923454 ; 12923458 ; 12923460 ; 12923461 ; 12923465 ; 12923466 ; 12923467 ; 12923468 ; 12923470 ; 12923472 ; 12923474 ; 12923475 ; 12923477 ; 12923478 ; 12923480 ; 12923481 ; 12923483 ; 12923484 ; 12923496 ; 12923502 ; 12923506 ; 12923507 ; 12923508 ; 12923511 ; 12923518 ; 12923533 ; 12923548 ; 12923552 ; 12923565 ; 12923571 ; 12923576 ; 12923579 ; 12923584 ; 12923589 ; 12923593 ; 12923594 ; 12923599 ; 12923606 ; 12923609 ; 12923613 ; 12923614 ; 12923615 ; 12923617 ; 12923618 ; 12923621 ; 12923622 ; 12923623 ; 12923625 ; 12923627 ; 12923628 ; 12923630 ; 12923632 ; 12923635 ; 12923637 ; 12923640 ; 12923645 ; 12923648 ; 12923653 ; 12923657 ; 12923658 ; 12923659 ; 12923660 ; 12923661 ; 12923663 ; 12923664 ; 12923665 ; 12923666 ; 12923667 ; 12923669 ; 12923671 ; 12923673 ; 12923676 ; 12923677 ; 12923680 ; 12923681 ; 12923684 ; 12923685 ; 12923687 ; 12923689 ; 12923690 ; 12923691 ; 12923692 ; 12923695 ; 12923697 ; 12923699 ; 12923700 ; 12923703 ; 12923704 ; 12923705 ; 12923710 ; 12923711 ; 12923712 ; 12923714 ; 12923715 ; 12923716 ; 12923718 ; 12923719 ; 12923720 ; 12923721 ; 12923722 ; 12923725 ; 12923726 ; 12923728 ; 12923730 ; 12923731 ; 12923732 ; 12923734 ; 12923735 ; 12923736 ; 12923737 ; 12923738 ; 12923741 ; 12923742 ; 12923746 ; 12923747 ; 12923748 ; 12923749 ; 12923750 ; 12923751 ; 12923753 ; 12923754 ; 12923757 ; 12923760 ; 12923774 ; 12923777 ; 12923779 ; 12923784 ; 12923786 ; 12923792 ; 12923794 ; 12923795 ; 12923796 ; 12923799 ; 12923800 ; 12923802 ; 12923803 ; 12923805 ; 12923811 ; 12923815 ; 12923816 ; 12923818 ; 12923820 ; 12923822 ; 12923823 ; 12923824 ; 12923825 ; 12923826 ; 12923829 ; 12923830 ; 12923834 ; 12923836 ; 12923838 ; 12923839 ; 12923840 ; 12923841 ; 12923842 ; 12923843 ; 12923845 ; 12923847 ; 12923848 ; 12923850 ; 12923851 ; 12923853 ; 12923854 ; 12923855 ; 12923856 ; 12923857 ; 12923858 ; 12923859 ; 12923860 ; 12923861 ; 12923862 ; 12923863 ; 12923864 ; 12923865 ; 12923866 ; 12923869 ; 12923870 ; 12923871 ; 12923872 ; 12923873 ; 12923874 ; 12923876 ; 12923877 ; 12923878 ; 12923879 ; 12923880 ; 12923881 ; 12923882 ; 12923883 ; 12923884 ; 12923885 ; 12923887 ; 12923888 ; 12923889 ; 12923890 ; 12923891 ; 12923892 ; 12923893 ; 12923894 ; 12923895 ; 12923896 ; 12923897 ; 12923899 ; 12923900 ; 12923901 ; 12923902 ; 12923904 ; 12923905 ; 12923906 ; 12923907 ; 12923908 ; 12923909 ; 12923910 ; 12923911 ; 12923912 ; 12923913 ; 12923914 ; 12923915 ; 12923916 ; 12923917 ; 12923918 ; 12923919 ; 12923920 ; 12923921 ; 12923922 ; 12923924 ; 12923925 ; 12923926 ; 12923927 ; 12923928 ; 12923929 ; 12923930 ; 12923931 ; 12923932 ; 12923933 ; 12923934 ; 12923935 ; 12923936 ; 12923937 ; 12923938 ; 12923939 ; 12923940 ; 12923941 ; 12923942 ; 12923943 ; 12923944 ; 12923945 ; 12923946 ; 12923949 ; 12923950 ; 12923951 ; 12923952 ; 12923954 ; 12923955 ; 12923956 ; 12923957 ; 12923958 ; 12923959 ; 12923960 ; 12923962 ; 12923963 ; 12923964 ; 12923966 ; 12923967 ; 12923968 ; 12923969 ; 12923970 ; 12923971 ; 12923972 ; 12923973 ; 12923974 ; 12923975 ; 12923977 ; 12923978 ; 12923979 ; 12923980 ; 12923981 ; 12923982 ; 12923983 ; 12923984 ; 12923985 ; 12923986 ; 12923987 ; 12923988 ; 12923989 ; 12923990 ; 12923992 ; 12923993 ; 12923994 ; 12923995 ; 12923996 ; 12923997 ; 12923998 ; 12923999 ; 12924000 ; 12924001 ; 12924002 ; 12924003 ; 12924004 ; 12924005 ; 12924006 ; 12924007 ; 12924008 ; 12924010 ; 12924011 ; 12924012 ; 12924013 ; 12924014 ; 12924015 ; 12924017 ; 12924019 ; 12924020 ; 12924021 ; 12924022 ; 12924023 ; 12924024 ; 12924025 ; 12924026 ; 12924027 ; 12924028 ; 12924029 ; 12924032 ; 12924033 ; 12924034 ; 12924035 ; 12924036 ; 12924037 ; 12924038 ; 12924039 ; 12924040 ; 12924041 ; 12924042 ; 12924043 ; 12924045 ; 12924048 ; 12924052 ; 12924053 ; 12924058 ; 12924059 ; 12924060 ; 12924062 ; 12924063 ; 12924065 ; 12924066 ; 12924067 ; 12924068 ; 12924069 ; 12924070 ; 12924071 ; 12924072 ; 12924073 ; 12924074 ; 12924075 ; 12924076 ; 12924077 ; 12924078 ; 12924080 ; 12924081 ; 12924082 ; 12924083 ; 12924085 ; 12924086 ; 12924088 ; 12924089 ; 12924091 ; 12924092 ; 12924093 ; 12924094 ; 12924095 ; 12924096 ; 12924097 ; 12924098 ; 12924099 ; 12924100 ; 12924101 ; 12924102 ; 12924103 ; 12924104 ; 12924105 ; 12924106 ; 12924107 ; 12924108 ; 12924111 ; 12924112 ; 12924113 ; 12924114 ; 12924115 ; 12924116 ; 12924117 ; 12924118 ; 12924119 ; 12924121 ; 12924122 ; 12924123 ; 12924125 ; 12924127 ; 12924128 ; 12924129 ; 12924130 ; 12924131 ; 12924132 ; 12924133 ; 12924134 ; 12924135 ; 12924136 ; 12924137 ; 12924138 ; 12924139 ; 12924140 ; 12924141 ; 12924142 ; 12924145 ; 12924146 ; 12924147 ; 12924148 ; 12924149 ; 12924150 ; 12924151 ; 12924152 ; 12924153 ; 12924154 ; 12924155 ; 12924156 ; 12924157 ; 12924158 ; 12924159 ; 12924162 ; 12924163 ; 12924165 ; 12924166 ; 12924167 ; 12924168 ; 12924171 ; 12924174 ; 12924175 ; 12924176 ; 12924177 ; 12924178 ; 12924179 ; 12924180 ; 12924181 ; 12924182 ; 12924183 ; 12924185 ; 12924186 ; 12924187 ; 12924189 ; 12924192 ; 12924193 ; 12924194 ; 12924195 ; 12924196 ; 12924197 ; 12924198 ; 12924199 ; 12924200 ; 12924201 ; 12924202 ; 12924203 ; 12924204 ; 12924205 ; 12924206 ; 12924207 ; 12924209 ; 12924210 ; 12924211 ; 12924212 ; 12924213 ; 12924214 ; 12924216 ; 12924217 ; 12924218 ; 12924219 ; 12924221 ; 12924222 ; 12924223 ; 12924224 ; 12924225 ; 12924226 ; 12924227 ; 12924228 ; 12924229 ; 12924230 ; 12924231 ; 12924232 ; 12924233 ; 12924235 ; 12924237 ; 12924238 ; 12924239 ; 12924241 ; 12924242 ; 12924243 ; 12924244 ; 12924245 ; 12924247 ; 12924248 ; 12924249 ; 12924250 ; 12924251 ; 12924252 ; 12924253 ; 12924255 ; 12924257 ; 12924258 ; 12924259 ; 12924260 ; 12924261 ; 12924262 ; 12924263 ; 12924264 ; 12924265 ; 12924266 ; 12924267 ; 12924269 ; 12924270 ; 12924271 ; 12924272 ; 12924273 ; 12924274 ; 12924275 ; 12924277 ; 12924278 ; 12924279 ; 12924281 ; 12924282 ; 12924283 ; 12924284 ; 12924286 ; 12924287 ; 12924288 ; 12924290 ; 12924291 ; 12924293 ; 12924295 ; 12924296 ; 12924298 ; 12924299 ; 12924301 ; 12924302 ; 12924303 ; 12924307 ; 12924308 ; 12924309 ; 12924311 ; 12924313 ; 12924314 ; 12924315 ; 12924317 ; 12924321 ; 12924322 ; 12924324 ; 12924326 ; 12924328 ; 12924329 ; 12924332 ; 12924335 ; 12924337 ; 12924344 ; 12924347 ; 12924348 ; 12924349 ; 12924352 ; 12924353 ; 12924355 ; 12924357 ; 12924358 ; 12924359 ; 12924363 ; 12924364 ; 12924367 ; 12924370 ; 12924371 ; 12924372 ; 12924373 ; 12924377 ; 12924378 ; 12924379 ; 12924380 ; 12924384 ; 12924385 ; 12924386 ; 12924388 ; 12924389 ; 12924391 ; 12924392 ; 12924397 ; 12924398 ; 12924399 ; 12924403 ; 12924404 ; 12924407 ; 12924408 ; 12924409 ; 12924411 ; 12924417 ; 12924421 ; 12924428 ; 12924429 ; 12924430 ; 12924431 ; 12924432 ; 12924433 ; 12924435 ; 12924437 ; 12924443 ; 12924446 ; 12924449 ; 12924452 ; 12924459 ; 12924464 ; 12924467 ; 12924469 ; 12924473 ; 12924480 ; 12924481 ; 12924483 ; 12924489 ; 12924490 ; 12924498 ; 12924501 ; 12924504 ; 12924513 ; 12924515 ; 12924518 ; 12924520 ; 12924521 ; 12924525 ; 12924529 ; 12924530 ; 12924531 ; 12924536 ; 12924537 ; 12924538 ; 12924541 ; 12924542 ; 12924543 ; 12924546 ; 12924547 ; 12924550 ; 12924554 ; 12924556 ; 12924560 ; 12924562 ; 12924564 ; 12924565 ; 12924566 ; 12924567 ; 12924572 ; 12924575 ; 12924576 ; 12924578 ; 12924579 ; 12924580 ; 12924581 ; 12924584 ; 12924585 ; 12924586 ; 12924587 ; 12924588 ; 12924589 ; 12924590 ; 12924592 ; 12924593 ; 12924594 ; 12924595 ; 12924596 ; 12924599 ; 12924600 ; 12924603 ; 12924605 ; 12924606 ; 12924609 ; 12924610 ; 12924611 ; 12924613 ; 12924614 ; 12924615 ; 12924617 ; 12924618 ; 12924619 ; 12924620 ; 12924621 ; 12924625 ; 12924627 ; 12924628 ; 12924629 ; 12924630 ; 12924632 ; 12924635 ; 12924636 ; 12924637 ; 12924638 ; 12924640 ; 12924642 ; 12924643 ; 12924644 ; 12924645 ; 12924646 ; 12924648 ; 12924649 ; 12924650 ; 12924651 ; 12924652 ; 12924653 ; 12924654 ; 12924656 ; 12924657 ; 12924658 ; 12924659 ; 12924664 ; 12924667 ; 12924668 ; 12924670 ; 12924671 ; 12924673 ; 12924674 ; 12924675 ; 12924676 ; 12924677 ; 12924678 ; 12924679 ; 12924680 ; 12924681 ; 12924682 ; 12924684 ; 12924685 ; 12924686 ; 12924688 ; 12924689 ; 12924691 ; 12924693 ; 12924694 ; 12924695 ; 12924697 ; 12924698 ; 12924699 ; 12924700 ; 12924703 ; 12924705 ; 12924709 ; 12924711 ; 12924713 ; 12924715 ; 12924716 ; 12924719 ; 12924723 ; 12924724 ; 12924727 ; 12924728 ; 12924729 ; 12924733 ; 12924734 ; 12924735 ; 12924736 ; 12924740 ; 12924742 ; 12924743 ; 12924747 ; 12924748 ; 12924749 ; 12924751 ; 12924752 ; 12924754 ; 12924756 ; 12924757 ; 12924759 ; 12924760 ; 12924761 ; 12924764 ; 12924766 ; 12924767 ; 12924768 ; 12924769 ; 12924770 ; 12924771 ; 12924772 ; 12924773 ; 12924774 ; 12924775 ; 12924776 ; 12924777 ; 12924778 ; 12924779 ; 12924780 ; 12924781 ; 12924782 ; 12924784 ; 12924785 ; 12924786 ; 12924787 ; 12924788 ; 12924790 ; 12924791 ; 12924792 ; 12924793 ; 12924794 ; 12924795 ; 12924796 ; 12924798 ; 12924799 ; 12924800 ; 12924801 ; 12924802 ; 12924803 ; 12924804 ; 12924805 ; 12924806 ; 12924807 ; 12924809 ; 12924811 ; 12924812 ; 12924813 ; 12924814 ; 12924815 ; 12924816 ; 12924817 ; 12924818 ; 12924819 ; 12924821 ; 12924822 ; 12924824 ; 12924825 ; 12924826 ; 12924827 ; 12924828 ; 12924832 ; 12924834 ; 12924835 ; 12924836 ; 12924837 ; 12924842 ; 12924844 ; 12924845 ; 12924846 ; 12924847 ; 12924848 ; 12924852 ; 12924853 ; 12924858 ; 12924859 ; 12924861 ; 12924863 ; 12924864 ; 12924865 ; 12924868 ; 12924869 ; 12924871 ; 12924873 ; 12924875 ; 12924877 ; 12924878 ; 12924881 ; 12924882 ; 12924886 ; 12924888 ; 12924891 ; 12924892 ; 12924893 ; 12924894 ; 12924895 ; 12924896 ; 12924899 ; 12924902 ; 12924906 ; 12924907 ; 12924908 ; 12924911 ; 12924917 ; 12924918 ; 12924919 ; 12924920 ; 12924925 ; 12924926 ; 12924928 ; 12924929 ; 12924930 ; 12924933 ; 12924936 ; 12924937 ; 12924938 ; 12924939 ; 12924940 ; 12924950 ; 12924952 ; 12924953 ; 12924954 ; 12924964 ; 12924965 ; 12924966 ; 12924971 ; 12924974 ; 12924976 ; 12924977 ; 12924978 ; 12924979 ; 12924981 ; 12924983 ; 12924984 ; 12924985 ; 12924986 ; 12924987 ; 12924988 ; 12924990 ; 12924993 ; 12924994 ; 12924995 ; 12924997 ; 12925000 ; 12925005 ; 12925007 ; 12925008 ; 12925009 ; 12925010 ; 12925012 ; 12925014 ; 12925017 ; 12925018 ; 12925025 ; 12925026 ; 12925029 ; 12925031 ; 12925032 ; 12925034 ; 12925035 ; 12925036 ; 12925037 ; 12925041 ; 12925043 ; 12925045 ; 12925046 ; 12925047 ; 12925048 ; 12925049 ; 12925050 ; 12925051 ; 12925052 ; 12925053 ; 12925054 ; 12925056 ; 12925058 ; 12925059 ; 12925060 ; 12925062 ; 12925063 ; 12925064 ; 12925065 ; 12925066 ; 12925067 ; 12925068 ; 12925069 ; 12925070 ; 12925071 ; 12925072 ; 12925073 ; 12925074 ; 12925076 ; 12925077 ; 12925078 ; 12925079 ; 12925080 ; 12925082 ; 12925084 ; 12925085 ; 12925086 ; 12925089 ; 12925091 ; 12925092 ; 12925095 ; 12925097 ; 12925098 ; 12925100 ; 12925106 ; 12925108 ; 12925109 ; 12925111 ; 12925112 ; 12925121 ; 12925122 ; 12925125 ; 12925127 ; 12925129 ; 12925134 ; 12925137 ; 12925139 ; 12925141 ; 12925142 ; 12925143 ; 12925144 ; 12925145 ; 12925146 ; 12925147 ; 12925148 ; 12925150 ; 12925152 ; 12925154 ; 12925155 ; 12925156 ; 12925157 ; 12925158 ; 12925159 ; 12925160 ; 12925161 ; 12925162 ; 12925163 ; 12925164 ; 12925165 ; 12925167 ; 12925168 ; 12925169 ; 12925170 ; 12925171 ; 12925173 ; 12925174 ; 12925175 ; 12925176 ; 12925177 ; 12925178 ; 12925179 ; 12925181 ; 12925183 ; 12925184 ; 12925185 ; 12925187 ; 12925188 ; 12925190 ; 12925191 ; 12925192 ; 12925193 ; 12925194 ; 12925195 ; 12925196 ; 12925197 ; 12925198 ; 12925199 ; 12925200 ; 12925201 ; 12925202 ; 12925203 ; 12925204 ; 12925205 ; 12925206 ; 12925207 ; 12925208 ; 12925209 ; 12925210 ; 12925211 ; 12925212 ; 12925213 ; 12925214 ; 12925215 ; 12925217 ; 12925218 ; 12925219 ; 12925220 ; 12925221 ; 12925222 ; 12925223 ; 12925224 ; 12925225 ; 12925226 ; 12925227 ; 12925228 ; 12925229 ; 12925230 ; 12925231 ; 12925232 ; 12925234 ; 12925235 ; 12925236 ; 12925237 ; 12925238 ; 12925239 ; 12925240 ; 12925241 ; 12925242 ; 12925243 ; 12925244 ; 12925245 ; 12925246 ; 12925247 ; 12925248 ; 12925249 ; 12925250 ; 12925251 ; 12925252 ; 12925253 ; 12925254 ; 12925255 ; 12925256 ; 12925258 ; 12925259 ; 12925260 ; 12925262 ; 12925263 ; 12925264 ; 12925265 ; 12925266 ; 12925268 ; 12925269 ; 12925270 ; 12925272 ; 12925274 ; 12925276 ; 12925277 ; 12925280 ; 12925282 ; 12925286 ; 12925296 ; 12925297 ; 12925300 ; 12925302 ; 12925304 ; 12925305 ; 12925313 ; 12925314 ; 12925315 ; 12925319 ; 12925321 ; 12925322 ; 12925324 ; 12925326 ; 12925327 ; 12925328 ; 12925329 ; 12925331 ; 12925333 ; 12925334 ; 12925335 ; 12925336 ; 12925340 ; 12925341 ; 12925342 ; 12925343 ; 12925344 ; 12925346 ; 12925347 ; 12925350 ; 12925352 ; 12925353 ; 12925354 ; 12925355 ; 12925356 ; 12925357 ; 12925358 ; 12925359 ; 12925360 ; 12925361 ; 12925362 ; 12925363 ; 12925364 ; 12925365 ; 12925366 ; 12925367 ; 12925369 ; 12925370 ; 12925371 ; 12925372 ; 12925373 ; 12925374 ; 12925375 ; 12925376 ; 12925377 ; 12925378 ; 12925379 ; 12925380 ; 12925381 ; 12925382 ; 12925383 ; 12925384 ; 12925385 ; 12925388 ; 12925389 ; 12925390 ; 12925392 ; 12925393 ; 12925394 ; 12925395 ; 12925396 ; 12925397 ; 12925398 ; 12925401 ; 12925402 ; 12925403 ; 12925405 ; 12925409 ; 12925410 ; 12925411 ; 12925412 ; 12925413 ; 12925414 ; 12925416 ; 12925419 ; 12925420 ; 12925422 ; 12925423 ; 12925424 ; 12925425 ; 12925426 ; 12925427 ; 12925429 ; 12925431 ; 12925437 ; 12925441 ; 12925445 ; 12925446 ; 12925447 ; 12925451 ; 12925456 ; 12925457 ; 12925458 ; 12925460 ; 12925462 ; 12925464 ; 12925465 ; 12925466 ; 12925466 ; 12925466 ; 12925467 ; 12925468 ; 12925473 ; 12925474 ; 12925478 ; 12925480 ; 12925482 ; 12925483 ; 12925485 ; 12925491 ; 12925493 ; 12925494 ; 12925497 ; 12925501 ; 12925505 ; 12925506 ; 12925508 ; 12925510 ; 12925512 ; 12925515 ; 12925516 ; 12925518 ; 12925519 ; 12925520 ; 12925523 ; 12925524 ; 12925533 ; 12925535 ; 12925536 ; 12925538 ; 12925540 ; 12925542 ; 12925544 ; 12925548 ; 12925549 ; 12925550 ; 12925551 ; 12925554 ; 12925555 ; 12925557 ; 12925563 ; 12925564 ; 12925566 ; 12925567 ; 12925575 ; 12925575 ; 12925575 ; 12925577 ; 12925578 ; 12925579 ; 12925580 ; 12925581 ; 12925583 ; 12925589 ; 12925590 ; 12925591 ; 12925595 ; 12925600 ; 12925601 ; 12925603 ; 12925605 ; 12925607 ; 12925608 ; 12925610 ; 12925611 ; 12925612 ; 12925613 ; 12925614 ; 12925617 ; 12925618 ; 12925620 ; 12925621 ; 12925624 ; 12925626 ; 12925627 ; 12925628 ; 12925630 ; 12925631 ; 12925633 ; 12925635 ; 12925637 ; 12925638 ; 12925639 ; 12925641 ; 12925642 ; 12925643 ; 12925644 ; 12925645 ; 12925647 ; 12925648 ; 12925649 ; 12925651 ; 12925652 ; 12925653 ; 12925654 ; 12925655 ; 12925656 ; 12925657 ; 12925658 ; 12925660 ; 12925661 ; 12925662 ; 12925663 ; 12925664 ; 12925665 ; 12925666 ; 12925669 ; 12925671 ; 12925672 ; 12925673 ; 12925677 ; 12925678 ; 12925679 ; 12925680 ; 12925681 ; 12925682 ; 12925683 ; 12925684 ; 12925685 ; 12925686 ; 12925687 ; 12925688 ; 12925689 ; 12925690 ; 12925691 ; 12925692 ; 12925693 ; 12925694 ; 12925695 ; 12925696 ; 12925697 ; 12925698 ; 12925699 ; 12925700 ; 12925701 ; 12925702 ; 12925703 ; 12925704 ; 12925705 ; 12925708 ; 12925710 ; 12925711 ; 12925716 ; 12925717 ; 12925721 ; 12925723 ; 12925728 ; 12925729 ; 12925731 ; 12925733 ; 12925735 ; 12925737 ; 12925738 ; 12925741 ; 12925742 ; 12925744 ; 12925745 ; 12925746 ; 12925747 ; 12925748 ; 12925750 ; 12925752 ; 12925753 ; 12925754 ; 12925755 ; 12925756 ; 12925758 ; 12925759 ; 12925760 ; 12925761 ; 12925761 ; 12925761 ; 12925762 ; 12925764 ; 12925765 ; 12925766 ; 12925767 ; 12925768 ; 12925770 ; 12925771 ; 12925772 ; 12925773 ; 12925773 ; 12925773 ; 12925774 ; 12925775 ; 12925776 ; 12925779 ; 12925779 ; 12925779 ; 12925780 ; 12925781 ; 12925783 ; 12925784 ; 12925787 ; 12925788 ; 12925789 ; 12925792 ; 12925793 ; 12925794 ; 12925795 ; 12925798 ; 12925799 ; 12925802 ; 12925803 ; 12925804 ; 12925807 ; 12925808 ; 12925810 ; 12925811 ; 12925812 ; 12925813 ; 12925814 ; 12925817 ; 12925820 ; 12925821 ; 12925824 ; 12925825 ; 12925827 ; 12925828 ; 12925830 ; 12925831 ; 12925832 ; 12925833 ; 12925834 ; 12925837 ; 12925839 ; 12925840 ; 12925844 ; 12925845 ; 12925848 ; 12925851 ; 12925854 ; 12925857 ; 12925858 ; 12925859 ; 12925862 ; 12925863 ; 12925864 ; 12925866 ; 12925867 ; 12925868 ; 12925869 ; 12925872 ; 12925876 ; 12925880 ; 12925881 ; 12925887 ; 12925888 ; 12925889 ; 12925891 ; 12925892 ; 12925893 ; 12925896 ; 12925897 ; 12925898 ; 12925899 ; 12925900 ; 12925901 ; 12925902 ; 12925906 ; 12925907 ; 12925910 ; 12925911 ; 12925915 ; 12925917 ; 12925920 ; 12925921 ; 12925922 ; 12925923 ; 12925924 ; 12925926 ; 12925928 ; 12925929 ; 12925930 ; 12925931 ; 12925932 ; 12925933 ; 12925936 ; 12925937 ; 12925939 ; 12925940 ; 12925941 ; 12925942 ; 12925943 ; 12925944 ; 12925950 ; 12925951 ; 12925952 ; 12925953 ; 12925954 ; 12925955 ; 12925958 ; 12925960 ; 12925961 ; 12925963 ; 12925964 ; 12925966 ; 12925968 ; 12925972 ; 12925973 ; 12925976 ; 12925977 ; 12925979 ; 12925981 ; 12925983 ; 12925984 ; 12925985 ; 12925986 ; 12925987 ; 12925989 ; 12925991 ; 12925993 ; 12925994 ; 12925995 ; 12925996 ; 12925999 ; 12926002 ; 12926003 ; 12926004 ; 12926006 ; 12926008 ; 12926009 ; 12926010 ; 12926011 ; 12926012 ; 12926015 ; 12926017 ; 12926019 ; 12926020 ; 12926021 ; 12926022 ; 12926024 ; 12926026 ; 12926027 ; 12926029 ; 12926030 ; 12926031 ; 12926033 ; 12926035 ; 12926036 ; 12926037 ; 12926038 ; 12926039 ; 12926040 ; 12926041 ; 12926043 ; 12926044 ; 12926045 ; 12926047 ; 12926048 ; 12926049 ; 12926065 ; 12926066 ; 12926067 ; 12926068 ; 12926069 ; 12926070 ; 12926072 ; 12926074 ; 12926075 ; 12926076 ; 12926077 ; 12926078 ; 12926079 ; 12926080 ; 12926081 ; 12926082 ; 12926083 ; 12926084 ; 12926085 ; 12926086 ; 12926087 ; 12926088 ; 12926089 ; 12926090 ; 12926091 ; 12926092 ; 12926093 ; 12926094 ; 12926095 ; 12926096 ; 12926097 ; 12926098 ; 12926099 ; 12926100 ; 12926101 ; 12926103 ; 12926104 ; 12926108 ; 12926111 ; 12926112 ; 12926113 ; 12926116 ; 12926121 ; 12926123 ; 12926124 ; 12926127 ; 12926129 ; 12926130 ; 12926132 ; 12926133 ; 12926135 ; 12926136 ; 12926137 ; 12926138 ; 12926139 ; 12926140 ; 12926141 ; 12926142 ; 12926143 ; 12926144 ; 12926145 ; 12926146 ; 12926149 ; 12926150 ; 12926153 ; 12926154 ; 12926155 ; 12926157 ; 12926158 ; 12926159 ; 12926160 ; 12926161 ; 12926162 ; 12926164 ; 12926165 ; 12926167 ; 12926169 ; 12926174 ; 12926176 ; 12926177 ; 12926183 ; 12926184 ; 12926185 ; 12926188 ; 12926191 ; 12926194 ; 12926195 ; 12926196 ; 12926197 ; 12926199 ; 12926201 ; 12926202 ; 12926203 ; 12926205 ; 12926207 ; 12926209 ; 12926210 ; 12926211 ; 12926212 ; 12926213 ; 12926214 ; 12926216 ; 12926218 ; 12926220 ; 12926222 ; 12926224 ; 12926225 ; 12926226 ; 12926227 ; 12926232 ; 12926233 ; 12926235 ; 12926237 ; 12926240 ; 12926241 ; 12926243 ; 12926244 ; 12926245 ; 12926246 ; 12926247 ; 12926248 ; 12926249 ; 12926251 ; 12926252 ; 12926254 ; 12926255 ; 12926257 ; 12926259 ; 12926261 ; 12926262 ; 12926263 ; 12926271 ; 12926273 ; 12926277 ; 12926278 ; 12926279 ; 12926280 ; 12926281 ; 12926282 ; 12926286 ; 12926289 ; 12926295 ; 12926296 ; 12926299 ; 12926301 ; 12926305 ; 12926306 ; 12926311 ; 12926312 ; 12926313 ; 12926315 ; 12926318 ; 12926323 ; 12926326 ; 12926330 ; 12926333 ; 12926336 ; 12926343 ; 12926344 ; 12926347 ; 12926348 ; 12926351 ; 12926354 ; 12926355 ; 12926358 ; 12926362 ; 12926365 ; 12926366 ; 12926369 ; 12926370 ; 12926373 ; 12926374 ; 12926375 ; 12926377 ; 12926380 ; 12926381 ; 12926383 ; 12926384 ; 12926388 ; 12926392 ; 12926394 ; 12926395 ; 12926396 ; 12926398 ; 12926399 ; 12926400 ; 12926401 ; 12926402 ; 12926403 ; 12926407 ; 12926408 ; 12926410 ; 12926411 ; 12926419 ; 12926421 ; 12926422 ; 12926426 ; 12926428 ; 12926430 ; 12926431 ; 12926434 ; 12926435 ; 12926437 ; 12926438 ; 12926439 ; 12926441 ; 12926448 ; 12926451 ; 12926452 ; 12926453 ; 12926456 ; 12926457 ; 12926458 ; 12926459 ; 12926460 ; 12926461 ; 12926463 ; 12926464 ; 12926465 ; 12926469 ; 12926470 ; 12926473 ; 12926474 ; 12926475 ; 12926477 ; 12926477 ; 12926477 ; 12926478 ; 12926479 ; 12926481 ; 12926481 ; 12926481 ; 12926482 ; 12926483 ; 12926483 ; 12926483 ; 12926486 ; 12926489 ; 12926489 ; 12926489 ; 12926490 ; 12926491 ; 12926491 ; 12926491 ; 12926492 ; 12926492 ; 12926492 ; 12926493 ; 12926493 ; 12926493 ; 12926494 ; 12926495 ; 12926495 ; 12926495 ; 12926496 ; 12926497 ; 12926500 ; 12926502 ; 12926504 ; 12926504 ; 12926504 ; 12926506 ; 12926507 ; 12926508 ; 12926510 ; 12926511 ; 12926511 ; 12926511 ; 12926512 ; 12926513 ; 12926514 ; 12926515 ; 12926515 ; 12926515 ; 12926516 ; 12926516 ; 12926516 ; 12926518 ; 12926518 ; 12926518 ; 12926519 ; 12926520 ; 12926520 ; 12926520 ; 12926521 ; 12926522 ; 12926523 ; 12926523 ; 12926523 ; 12926525 ; 12926526 ; 12926527 ; 12926527 ; 12926527 ; 12926528 ; 12926528 ; 12926528 ; 12926529 ; 12926530 ; 12926531 ; 12926532 ; 12926532 ; 12926532 ; 12926534 ; 12926534 ; 12926534 ; 12926536 ; 12926536 ; 12926536 ; 12926537 ; 12926537 ; 12926537 ; 12926538 ; 12926538 ; 12926538 ; 12926540 ; 12926540 ; 12926540 ; 12926541 ; 12926541 ; 12926541 ; 12926542 ; 12926543 ; 12926544 ; 12926545 ; 12926546 ; 12926548 ; 12926548 ; 12926548 ; 12926549 ; 12926550 ; 12926551 ; 12926552 ; 12926553 ; 12926555 ; 12926557 ; 12926558 ; 12926559 ; 12926560 ; 12926561 ; 12926562 ; 12926563 ; 12926564 ; 12926565 ; 12926569 ; 12926570 ; 12926572 ; 12926573 ; 12926574 ; 12926575 ; 12926576 ; 12926578 ; 12926580 ; 12926581 ; 12926582 ; 12926584 ; 12926585 ; 12926587 ; 12926588 ; 12926589 ; 12926590 ; 12926591 ; 12926592 ; 12926593 ; 12926594 ; 12926595 ; 12926596 ; 12926597 ; 12926598 ; 12926600 ; 12926602 ; 12926603 ; 12926604 ; 12926606 ; 12926608 ; 12926613 ; 12926615 ; 12926622 ; 12926624 ; 12926630 ; 12926638 ; 12926639 ; 12926642 ; 12926645 ; 12926646 ; 12926647 ; 12926652 ; 12926656 ; 12926659 ; 12926660 ; 12926661 ; 12926662 ; 12926664 ; 12926669 ; 12926670 ; 12926672 ; 12926673 ; 12926677 ; 12926679 ; 12926680 ; 12926683 ; 12926684 ; 12926687 ; 12926688 ; 12926689 ; 12926690 ; 12926692 ; 12926694 ; 12926696 ; 12926697 ; 12926703 ; 12926706 ; 12926709 ; 12926710 ; 12926711 ; 12926712 ; 12926713 ; 12926715 ; 12926716 ; 12926718 ; 12926722 ; 12926723 ; 12926725 ; 12926726 ; 12926727 ; 12926728 ; 12926729 ; 12926730 ; 12926732 ; 12926733 ; 12926734 ; 12926735 ; 12926736 ; 12926737 ; 12926738 ; 12926739 ; 12926740 ; 12926741 ; 12926742 ; 12926743 ; 12926744 ; 12926745 ; 12926746 ; 12926748 ; 12926749 ; 12926750 ; 12926751 ; 12926752 ; 12926753 ; 12926754 ; 12926755 ; 12926757 ; 12926759 ; 12926761 ; 12926762 ; 12926763 ; 12926764 ; 12926766 ; 12926767 ; 12926768 ; 12926769 ; 12926770 ; 12926771 ; 12926772 ; 12926775 ; 12926776 ; 12926777 ; 12926781 ; 12926782 ; 12926784 ; 12926785 ; 12926787 ; 12926788 ; 12926789 ; 12926791 ; 12926792 ; 12926793 ; 12926794 ; 12926795 ; 12926796 ; 12926800 ; 12926809 ; 12926813 ; 12926815 ; 12926816 ; 12926819 ; 12926824 ; 12926830 ; 12926832 ; 12926837 ; 12926840 ; 12926849 ; 12926853 ; 12926854 ; 12926855 ; 12926858 ; 12926859 ; 12926862 ; 12926863 ; 12926864 ; 12926866 ; 12926872 ; 12926873 ; 12926877 ; 12926879 ; 12926880 ; 12926883 ; 12926885 ; 12926887 ; 12926890 ; 12926891 ; 12926898 ; 12926899 ; 12926900 ; 12926907 ; 12926910 ; 12926911 ; 12926915 ; 12926916 ; 12926925 ; 12926928 ; 12926929 ; 12926932 ; 12926938 ; 12926941 ; 12926946 ; 12926948 ; 12926949 ; 12926953 ; 12926954 ; 12926956 ; 12926958 ; 12926962 ; 12926963 ; 12926965 ; 12926966 ; 12926967 ; 12926969 ; 12926970 ; 12926971 ; 12926972 ; 12926973 ; 12926974 ; 12926975 ; 12926976 ; 12926978 ; 12926981 ; 12926982 ; 12926983 ; 12926984 ; 12926985 ; 12926986 ; 12926987 ; 12926988 ; 12926989 ; 12926990 ; 12926991 ; 12926993 ; 12926994 ; 12926995 ; 12926996 ; 12926997 ; 12926998 ; 12926999 ; 12927000 ; 12927001 ; 12927002 ; 12927003 ; 12927004 ; 12927005 ; 12927008 ; 12927009 ; 12927010 ; 12927012 ; 12927013 ; 12927014 ; 12927015 ; 12927016 ; 12927017 ; 12927018 ; 12927019 ; 12927020 ; 12927021 ; 12927022 ; 12927023 ; 12927024 ; 12927025 ; 12927026 ; 12927027 ; 12927028 ; 12927029 ; 12927030 ; 12927031 ; 12927032 ; 12927033 ; 12927034 ; 12927035 ; 12927036 ; 12927037 ; 12927041 ; 12927042 ; 12927043 ; 12927044 ; 12927045 ; 12927046 ; 12927047 ; 12927049 ; 12927051 ; 12927052 ; 12927053 ; 12927055 ; 12927056 ; 12927057 ; 12927058 ; 12927059 ; 12927061 ; 12927062 ; 12927063 ; 12927064 ; 12927065 ; 12927067 ; 12927069 ; 12927070 ; 12927071 ; 12927072 ; 12927073 ; 12927074 ; 12927075 ; 12927077 ; 12927078 ; 12927079 ; 12927080 ; 12927081 ; 12927082 ; 12927083 ; 12927084 ; 12927085 ; 12927086 ; 12927087 ; 12927089 ; 12927090 ; 12927091 ; 12927092 ; 12927093 ; 12927095 ; 12927097 ; 12927099 ; 12927100 ; 12927101 ; 12927102 ; 12927103 ; 12927104 ; 12927106 ; 12927107 ; 12927108 ; 12927109 ; 12927110 ; 12927111 ; 12927112 ; 12927113 ; 12927114 ; 12927115 ; 12927117 ; 12927118 ; 12927119 ; 12927121 ; 12927123 ; 12927125 ; 12927126 ; 12927127 ; 12927130 ; 12927131 ; 12927132 ; 12927134 ; 12927136 ; 12927137 ; 12927138 ; 12927141 ; 12927142 ; 12927143 ; 12927144 ; 12927145 ; 12927146 ; 12927147 ; 12927148 ; 12927149 ; 12927150 ; 12927151 ; 12927153 ; 12927155 ; 12927156 ; 12927157 ; 12927158 ; 12927160 ; 12927161 ; 12927162 ; 12927164 ; 12927165 ; 12927166 ; 12927167 ; 12927169 ; 12927170 ; 12927171 ; 12927172 ; 12927173 ; 12927174 ; 12927175 ; 12927176 ; 12927177 ; 12927178 ; 12927179 ; 12927180 ; 12927181 ; 12927182 ; 12927184 ; 12927185 ; 12927186 ; 12927187 ; 12927188 ; 12927189 ; 12927190 ; 12927191 ; 12927192 ; 12927194 ; 12927195 ; 12927196 ; 12927197 ; 12927198 ; 12927200 ; 12927201 ; 12927202 ; 12927203 ; 12927204 ; 12927205 ; 12927206 ; 12927207 ; 12927208 ; 12927210 ; 12927211 ; 12927212 ; 12927213 ; 12927214 ; 12927215 ; 12927219 ; 12927223 ; 12927224 ; 12927226 ; 12927227 ; 12927228 ; 12927229 ; 12927230 ; 12927231 ; 12927232 ; 12927233 ; 12927234 ; 12927235 ; 12927237 ; 12927238 ; 12927239 ; 12927240 ; 12927242 ; 12927244 ; 12927245 ; 12927246 ; 12927248 ; 12927251 ; 12927252 ; 12927253 ; 12927254 ; 12927262 ; 12927263 ; 12927264 ; 12927265 ; 12927266 ; 12927268 ; 12927269 ; 12927270 ; 12927271 ; 12927272 ; 12927273 ; 12927274 ; 12927277 ; 12927278 ; 12927279 ; 12927280 ; 12927281 ; 12927282 ; 12927283 ; 12927285 ; 12927286 ; 12927288 ; 12927289 ; 12927290 ; 12927291 ; 12927292 ; 12927293 ; 12927295 ; 12927296 ; 12927297 ; 12927298 ; 12927299 ; 12927301 ; 12927302 ; 12927303 ; 12927306 ; 12927307 ; 12927308 ; 12927309 ; 12927311 ; 12927312 ; 12927313 ; 12927314 ; 12927315 ; 12927316 ; 12927317 ; 12927318 ; 12927320 ; 12927321 ; 12927322 ; 12927323 ; 12927324 ; 12927326 ; 12927327 ; 12927328 ; 12927342 ; 12927344 ; 12927346 ; 12927347 ; 12927348 ; 12927349 ; 12927351 ; 12927352 ; 12927354 ; 12927356 ; 12927357 ; 12927358 ; 12927359 ; 12927360 ; 12927361 ; 12927363 ; 12927364 ; 12927367 ; 12927368 ; 12927369 ; 12927370 ; 12927371 ; 12927372 ; 12927373 ; 12927375 ; 12927376 ; 12927377 ; 12927378 ; 12927380 ; 12927381 ; 12927382 ; 12927383 ; 12927384 ; 12927385 ; 12927386 ; 12927387 ; 12927388 ; 12927389 ; 12927390 ; 12927391 ; 12927392 ; 12927395 ; 12927396 ; 12927397 ; 12927398 ; 12927399 ; 12927400 ; 12927401 ; 12927402 ; 12927403 ; 12927404 ; 12927405 ; 12927407 ; 12927408 ; 12927409 ; 12927410 ; 12927411 ; 12927412 ; 12927413 ; 12927414 ; 12927415 ; 12927416 ; 12927417 ; 12927418 ; 12927420 ; 12927421 ; 12927422 ; 12927423 ; 12927424 ; 12927425 ; 12927426 ; 12927427 ; 12927428 ; 12927429 ; 12927430 ; 12927431 ; 12927432 ; 12927433 ; 12927434 ; 12927435 ; 12927437 ; 12927439 ; 12927440 ; 12927441 ; 12927445 ; 12927447 ; 12927450 ; 12927451 ; 12927452 ; 12927454 ; 12927455 ; 12927456 ; 12927458 ; 12927465 ; 12927466 ; 12927467 ; 12927468 ; 12927469 ; 12927470 ; 12927471 ; 12927472 ; 12927476 ; 12927479 ; 12927480 ; 12927490 ; 12927494 ; 12927507 ; 12927508 ; 12927514 ; 12927515 ; 12927516 ; 12927517 ; 12927520 ; 12927524 ; 12927525 ; 12927526 ; 12927527 ; 12927529 ; 12927531 ; 12927533 ; 12927534 ; 12927535 ; 12927536 ; 12927537 ; 12927538 ; 12927541 ; 12927544 ; 12927548 ; 12927549 ; 12927550 ; 12927553 ; 12927554 ; 12927556 ; 12927557 ; 12927558 ; 12927559 ; 12927560 ; 12927561 ; 12927562 ; 12927563 ; 12927564 ; 12927565 ; 12927566 ; 12927567 ; 12927570 ; 12927571 ; 12927572 ; 12927573 ; 12927574 ; 12927575 ; 12927576 ; 12927577 ; 12927578 ; 12927579 ; 12927580 ; 12927581 ; 12927582 ; 12927583 ; 12927585 ; 12927586 ; 12927587 ; 12927589 ; 12927590 ; 12927591 ; 12927596 ; 12927603 ; 12927605 ; 12927606 ; 12927613 ; 12927615 ; 12927617 ; 12927619 ; 12927620 ; 12927623 ; 12927626 ; 12927650 ; 12927674 ; 12927675 ; 12927677 ; 12927678 ; 12927679 ; 12927680 ; 12927681 ; 12927682 ; 12927683 ; 12927684 ; 12927685 ; 12927687 ; 12927688 ; 12927689 ; 12927690 ; 12927691 ; 12927692 ; 12927693 ; 12927694 ; 12927696 ; 12927698 ; 12927699 ; 12927700 ; 12927702 ; 12927704 ; 12927706 ; 12927707 ; 12927708 ; 12927709 ; 12927710 ; 12927711 ; 12927712 ; 12927713 ; 12927714 ; 12927715 ; 12927716 ; 12927717 ; 12927718 ; 12927719 ; 12927720 ; 12927721 ; 12927722 ; 12927723 ; 12927724 ; 12927725 ; 12927726 ; 12927727 ; 12927728 ; 12927729 ; 12927730 ; 12927731 ; 12927732 ; 12927733 ; 12927734 ; 12927736 ; 12927737 ; 12927738 ; 12927739 ; 12927740 ; 12927741 ; 12927742 ; 12927743 ; 12927744 ; 12927745 ; 12927748 ; 12927749 ; 12927752 ; 12927753 ; 12927754 ; 12927756 ; 12927757 ; 12927758 ; 12927762 ; 12927763 ; 12927764 ; 12927765 ; 12927766 ; 12927766 ; 12927766 ; 12927767 ; 12927769 ; 12927770 ; 12927771 ; 12927780 ; 12927787 ; 12927789 ; 12927793 ; 12927805 ; 12927829 ; 12927856 ; 12927857 ; 12927867 ; 12927874 ; 12927879 ; 12927881 ; 12927889 ; 12927898 ; 12927900 ; 12927902 ; 12927903 ; 12927906 ; 12927908 ; 12927909 ; 12927912 ; 12927917 ; 12927918 ; 12927921 ; 12927923 ; 12927925 ; 12927926 ; 12927931 ; 12927932 ; 12927933 ; 12927934 ; 12927935 ; 12927936 ; 12927937 ; 12927938 ; 12927939 ; 12927941 ; 12927944 ; 12927945 ; 12927946 ; 12927947 ; 12927948 ; 12927951 ; 12927952 ; 12927955 ; 12927956 ; 12927957 ; 12927958 ; 12927962 ; 12927964 ; 12927967 ; 12927968 ; 12927970 ; 12927971 ; 12927972 ; 12927973 ; 12927974 ; 12927978 ; 12927979 ; 12927980 ; 12927981 ; 12927983 ; 12927984 ; 12927985 ; 12927986 ; 12927987 ; 12927988 ; 12927989 ; 12927990 ; 12927991 ; 12927993 ; 12927994 ; 12927995 ; 12927997 ; 12927998 ; 12927999 ; 12928000 ; 12928001 ; 12928002 ; 12928003 ; 12928004 ; 12928005 ; 12928006 ; 12928007 ; 12928009 ; 12928010 ; 12928011 ; 12928012 ; 12928013 ; 12928014 ; 12928015 ; 12928016 ; 12928017 ; 12928018 ; 12928019 ; 12928020 ; 12928024 ; 12928025 ; 12928026 ; 12928027 ; 12928028 ; 12928029 ; 12928030 ; 12928031 ; 12928032 ; 12928033 ; 12928034 ; 12928035 ; 12928036 ; 12928037 ; 12928038 ; 12928039 ; 12928040 ; 12928041 ; 12928042 ; 12928043 ; 12928044 ; 12928045 ; 12928046 ; 12928047 ; 12928048 ; 12928051 ; 12928052 ; 12928053 ; 12928055 ; 12928056 ; 12928058 ; 12928059 ; 12928060 ; 12928061 ; 12928062 ; 12928063 ; 12928064 ; 12928065 ; 12928066 ; 12928067 ; 12928068 ; 12928069 ; 12928070 ; 12928071 ; 12928072 ; 12928073 ; 12928074 ; 12928076 ; 12928077 ; 12928080 ; 12928082 ; 12928084 ; 12928085 ; 12928086 ; 12928087 ; 12928090 ; 12928091 ; 12928092 ; 12928094 ; 12928095 ; 12928096 ; 12928097 ; 12928098 ; 12928099 ; 12928102 ; 12928103 ; 12928104 ; 12928105 ; 12928106 ; 12928107 ; 12928109 ; 12928110 ; 12928111 ; 12928112 ; 12928113 ; 12928115 ; 12928121 ; 12928123 ; 12928130 ; 12928135 ; 12928137 ; 12928138 ; 12928142 ; 12928147 ; 12928151 ; 12928154 ; 12928156 ; 12928159 ; 12928160 ; 12928161 ; 12928162 ; 12928163 ; 12928164 ; 12928166 ; 12928167 ; 12928168 ; 12928175 ; 12928178 ; 12928179 ; 12928180 ; 12928182 ; 12928183 ; 12928185 ; 12928188 ; 12928191 ; 12928194 ; 12928197 ; 12928198 ; 12928200 ; 12928201 ; 12928206 ; 12928208 ; 12928209 ; 12928210 ; 12928211 ; 12928212 ; 12928215 ; 12928216 ; 12928218 ; 12928221 ; 12928222 ; 12928224 ; 12928225 ; 12928226 ; 12928227 ; 12928229 ; 12928230 ; 12928231 ; 12928234 ; 12928236 ; 12928241 ; 12928250 ; 12928258 ; 12928260 ; 12928261 ; 12928262 ; 12928264 ; 12928265 ; 12928267 ; 12928268 ; 12928269 ; 12928270 ; 12928271 ; 12928273 ; 12928274 ; 12928275 ; 12928276 ; 12928277 ; 12928278 ; 12928281 ; 12928282 ; 12928285 ; 12928286 ; 12928290 ; 12928291 ; 12928296 ; 12928298 ; 12928300 ; 12928303 ; 12928304 ; 12928305 ; 12928306 ; 12928307 ; 12928308 ; 12928309 ; 12928310 ; 12928311 ; 12928312 ; 12928313 ; 12928314 ; 12928316 ; 12928317 ; 12928318 ; 12928319 ; 12928320 ; 12928322 ; 12928323 ; 12928324 ; 12928325 ; 12928326 ; 12928327 ; 12928329 ; 12928331 ; 12928332 ; 12928333 ; 12928334 ; 12928337 ; 12928338 ; 12928339 ; 12928340 ; 12928341 ; 12928342 ; 12928344 ; 12928350 ; 12928351 ; 12928353 ; 12928354 ; 12928355 ; 12928356 ; 12928357 ; 12928358 ; 12928361 ; 12928362 ; 12928363 ; 12928364 ; 12928366 ; 12928367 ; 12928368 ; 12928369 ; 12928371 ; 12928372 ; 12928378 ; 12928379 ; 12928382 ; 12928384 ; 12928385 ; 12928386 ; 12928388 ; 12928390 ; 12928391 ; 12928392 ; 12928394 ; 12928396 ; 12928398 ; 12928399 ; 12928402 ; 12928404 ; 12928407 ; 12928408 ; 12928409 ; 12928410 ; 12928411 ; 12928412 ; 12928414 ; 12928415 ; 12928416 ; 12928418 ; 12928419 ; 12928420 ; 12928421 ; 12928422 ; 12928423 ; 12928424 ; 12928427 ; 12928428 ; 12928429 ; 12928430 ; 12928432 ; 12928437 ; 12928443 ; 12928445 ; 12928446 ; 12928447 ; 12928450 ; 12928453 ; 12928456 ; 12928459 ; 12928461 ; 12928462 ; 12928463 ; 12928464 ; 12928465 ; 12928466 ; 12928467 ; 12928468 ; 12928469 ; 12928470 ; 12928471 ; 12928472 ; 12928473 ; 12928474 ; 12928475 ; 12928476 ; 12928477 ; 12928479 ; 12928480 ; 12928482 ; 12928483 ; 12928486 ; 12928487 ; 12928489 ; 12928493 ; 12928496 ; 12928499 ; 12928500 ; 12928507 ; 12928512 ; 12928513 ; 12928514 ; 12928515 ; 12928516 ; 12928517 ; 12928518 ; 12928519 ; 12928520 ; 12928521 ; 12928522 ; 12928523 ; 12928524 ; 12928525 ; 12928526 ; 12928527 ; 12928528 ; 12928529 ; 12928530 ; 12928531 ; 12928532 ; 12928533 ; 12928534 ; 12928535 ; 12928536 ; 12928537 ; 12928538 ; 12928539 ; 12928540 ; 12928541 ; 12928542 ; 12928543 ; 12928544 ; 12928545 ; 12928546 ; 12928548 ; 12928549 ; 12928550 ; 12928551 ; 12928552 ; 12928553 ; 12928554 ; 12928555 ; 12928556 ; 12928557 ; 12928558 ; 12928560 ; 12928561 ; 12928562 ; 12928563 ; 12928564 ; 12928565 ; 12928566 ; 12928567 ; 12928568 ; 12928569 ; 12928570 ; 12928571 ; 12928572 ; 12928573 ; 12928574 ; 12928575 ; 12928576 ; 12928577 ; 12928578 ; 12928579 ; 12928580 ; 12928581 ; 12928582 ; 12928583 ; 12928584 ; 12928585 ; 12928586 ; 12928587 ; 12928589 ; 12928590 ; 12928591 ; 12928592 ; 12928593 ; 12928594 ; 12928595 ; 12928596 ; 12928598 ; 12928602 ; 12928603 ; 12928606 ; 12928607 ; 12928608 ; 12928609 ; 12928610 ; 12928611 ; 12928612 ; 12928613 ; 12928614 ; 12928616 ; 12928617 ; 12928618 ; 12928619 ; 12928620 ; 12928621 ; 12928622 ; 12928623 ; 12928624 ; 12928625 ; 12928626 ; 12928627 ; 12928628 ; 12928629 ; 12928630 ; 12928631 ; 12928632 ; 12928633 ; 12928634 ; 12928636 ; 12928637 ; 12928640 ; 12928641 ; 12928642 ; 12928643 ; 12928644 ; 12928645 ; 12928646 ; 12928647 ; 12928648 ; 12928649 ; 12928651 ; 12928652 ; 12928653 ; 12928654 ; 12928655 ; 12928656 ; 12928657 ; 12928659 ; 12928660 ; 12928661 ; 12928662 ; 12928663 ; 12928664 ; 12928666 ; 12928670 ; 12928673 ; 12928674 ; 12928675 ; 12928676 ; 12928677 ; 12928682 ; 12928684 ; 12928685 ; 12928686 ; 12928687 ; 12928688 ; 12928689 ; 12928690 ; 12928693 ; 12928694 ; 12928695 ; 12928696 ; 12928698 ; 12928700 ; 12928703 ; 12928706 ; 12928708 ; 12928710 ; 12928712 ; 12928713 ; 12928717 ; 12928718 ; 12928722 ; 12928723 ; 12928725 ; 12928726 ; 12928731 ; 12928732 ; 12928734 ; 12928735 ; 12928740 ; 12928741 ; 12928742 ; 12928744 ; 12928745 ; 12928746 ; 12928749 ; 12928750 ; 12928751 ; 12928752 ; 12928753 ; 12928754 ; 12928755 ; 12928757 ; 12928758 ; 12928759 ; 12928761 ; 12928762 ; 12928763 ; 12928764 ; 12928766 ; 12928767 ; 12928768 ; 12928769 ; 12928770 ; 12928771 ; 12928774 ; 12928776 ; 12928777 ; 12928778 ; 12928779 ; 12928781 ; 12928782 ; 12928787 ; 12928788 ; 12928789 ; 12928790 ; 12928791 ; 12928792 ; 12928793 ; 12928794 ; 12928796 ; 12928797 ; 12928798 ; 12928799 ; 12928800 ; 12928801 ; 12928802 ; 12928804 ; 12928805 ; 12928806 ; 12928807 ; 12928809 ; 12928810 ; 12928811 ; 12928812 ; 12928813 ; 12928816 ; 12928819 ; 12928820 ; 12928822 ; 12928823 ; 12928824 ; 12928825 ; 12928826 ; 12928828 ; 12928832 ; 12928833 ; 12928834 ; 12928835 ; 12928836 ; 12928838 ; 12928839 ; 12928840 ; 12928841 ; 12928843 ; 12928844 ; 12928845 ; 12928846 ; 12928847 ; 12928848 ; 12928849 ; 12928851 ; 12928852 ; 12928853 ; 12928854 ; 12928855 ; 12928856 ; 12928857 ; 12928858 ; 12928859 ; 12928860 ; 12928861 ; 12928864 ; 12928865 ; 12928866 ; 12928867 ; 12928868 ; 12928870 ; 12928871 ; 12928872 ; 12928873 ; 12928874 ; 12928875 ; 12928876 ; 12928877 ; 12928878 ; 12928879 ; 12928880 ; 12928882 ; 12928883 ; 12928884 ; 12928885 ; 12928886 ; 12928887 ; 12928889 ; 12928890 ; 12928893 ; 12928894 ; 12928895 ; 12928896 ; 12928897 ; 12928898 ; 12928899 ; 12928900 ; 12928901 ; 12928902 ; 12928903 ; 12928904 ; 12928905 ; 12928906 ; 12928907 ; 12928908 ; 12928909 ; 12928911 ; 12928913 ; 12928914 ; 12928915 ; 12928917 ; 12928918 ; 12928919 ; 12928920 ; 12928921 ; 12928922 ; 12928923 ; 12928924 ; 12928925 ; 12928926 ; 12928928 ; 12928929 ; 12928930 ; 12928931 ; 12928932 ; 12928933 ; 12928934 ; 12928935 ; 12928936 ; 12928937 ; 12928938 ; 12928939 ; 12928940 ; 12928941 ; 12928942 ; 12928943 ; 12928944 ; 12928945 ; 12928946 ; 12928947 ; 12928948 ; 12928950 ; 12928951 ; 12928952 ; 12928953 ; 12928954 ; 12928955 ; 12928957 ; 12928958 ; 12928959 ; 12928960 ; 12928961 ; 12928962 ; 12928963 ; 12928964 ; 12928965 ; 12928967 ; 12928968 ; 12928969 ; 12928970 ; 12928971 ; 12928972 ; 12928973 ; 12928974 ; 12928975 ; 12928976 ; 12928977 ; 12928978 ; 12928979 ; 12928980 ; 12928981 ; 12928983 ; 12928984 ; 12928985 ; 12928987 ; 12928988 ; 12928989 ; 12928991 ; 12928992 ; 12928993 ; 12928994 ; 12928995 ; 12928997 ; 12928998 ; 12928999 ; 12929000 ; 12929001 ; 12929002 ; 12929003 ; 12929004 ; 12929005 ; 12929006 ; 12929007 ; 12929008 ; 12929009 ; 12929012 ; 12929013 ; 12929014 ; 12929015 ; 12929016 ; 12929017 ; 12929018 ; 12929019 ; 12929020 ; 12929021 ; 12929022 ; 12929023 ; 12929024 ; 12929025 ; 12929026 ; 12929027 ; 12929028 ; 12929029 ; 12929030 ; 12929031 ; 12929033 ; 12929034 ; 12929036 ; 12929037 ; 12929038 ; 12929039 ; 12929040 ; 12929041 ; 12929042 ; 12929043 ; 12929044 ; 12929045 ; 12929047 ; 12929048 ; 12929049 ; 12929050 ; 12929051 ; 12929052 ; 12929053 ; 12929054 ; 12929055 ; 12929056 ; 12929057 ; 12929058 ; 12929059 ; 12929060 ; 12929061 ; 12929062 ; 12929063 ; 12929064 ; 12929066 ; 12929067 ; 12929068 ; 12929069 ; 12929070 ; 12929072 ; 12929075 ; 12929076 ; 12929078 ; 12929079 ; 12929081 ; 12929083 ; 12929084 ; 12929087 ; 12929089 ; 12929090 ; 12929091 ; 12929092 ; 12929093 ; 12929094 ; 12929095 ; 12929097 ; 12929099 ; 12929103 ; 12929105 ; 12929106 ; 12929107 ; 12929109 ; 12929111 ; 12929116 ; 12929117 ; 12929119 ; 12929120 ; 12929121 ; 12929122 ; 12929123 ; 12929124 ; 12929125 ; 12929126 ; 12929128 ; 12929131 ; 12929132 ; 12929135 ; 12929138 ; 12929139 ; 12929140 ; 12929141 ; 12929142 ; 12929143 ; 12929145 ; 12929146 ; 12929147 ; 12929148 ; 12929149 ; 12929151 ; 12929156 ; 12929160 ; 12929161 ; 12929162 ; 12929163 ; 12929164 ; 12929165 ; 12929166 ; 12929167 ; 12929168 ; 12929169 ; 12929170 ; 12929171 ; 12929172 ; 12929175 ; 12929176 ; 12929177 ; 12929178 ; 12929179 ; 12929180 ; 12929181 ; 12929182 ; 12929183 ; 12929184 ; 12929185 ; 12929188 ; 12929190 ; 12929191 ; 12929192 ; 12929194 ; 12929195 ; 12929196 ; 12929197 ; 12929198 ; 12929201 ; 12929202 ; 12929205 ; 12929206 ; 12929208 ; 12929209 ; 12929210 ; 12929212 ; 12929213 ; 12929214 ; 12929216 ; 12929217 ; 12929220 ; 12929221 ; 12929222 ; 12929223 ; 12929224 ; 12929225 ; 12929227 ; 12929228 ; 12929229 ; 12929230 ; 12929231 ; 12929232 ; 12929233 ; 12929234 ; 12929236 ; 12929240 ; 12929241 ; 12929242 ; 12929244 ; 12929247 ; 12929249 ; 12929252 ; 12929253 ; 12929255 ; 12929259 ; 12929260 ; 12929263 ; 12929264 ; 12929270 ; 12929271 ; 12929272 ; 12929274 ; 12929276 ; 12929277 ; 12929283 ; 12929284 ; 12929285 ; 12929286 ; 12929288 ; 12929289 ; 12929290 ; 12929291 ; 12929292 ; 12929295 ; 12929298 ; 12929299 ; 12929300 ; 12929301 ; 12929302 ; 12929303 ; 12929306 ; 12929307 ; 12929309 ; 12929310 ; 12929311 ; 12929312 ; 12929313 ; 12929314 ; 12929315 ; 12929316 ; 12929317 ; 12929318 ; 12929319 ; 12929321 ; 12929323 ; 12929324 ; 12929325 ; 12929327 ; 12929329 ; 12929331 ; 12929332 ; 12929333 ; 12929334 ; 12929335 ; 12929336 ; 12929337 ; 12929338 ; 12929339 ; 12929340 ; 12929341 ; 12929342 ; 12929343 ; 12929344 ; 12929345 ; 12929349 ; 12929350 ; 12929351 ; 12929352 ; 12929353 ; 12929354 ; 12929355 ; 12929356 ; 12929357 ; 12929358 ; 12929359 ; 12929360 ; 12929361 ; 12929362 ; 12929363 ; 12929364 ; 12929365 ; 12929366 ; 12929369 ; 12929370 ; 12929371 ; 12929373 ; 12929374 ; 12929375 ; 12929376 ; 12929377 ; 12929378 ; 12929380 ; 12929381 ; 12929382 ; 12929385 ; 12929386 ; 12929387 ; 12929388 ; 12929389 ; 12929390 ; 12929391 ; 12929392 ; 12929393 ; 12929394 ; 12929395 ; 12929396 ; 12929400 ; 12929401 ; 12929402 ; 12929403 ; 12929404 ; 12929405 ; 12929406 ; 12929407 ; 12929408 ; 12929409 ; 12929410 ; 12929414 ; 12929420 ; 12929421 ; 12929422 ; 12929423 ; 12929426 ; 12929427 ; 12929428 ; 12929429 ; 12929430 ; 12929431 ; 12929432 ; 12929433 ; 12929434 ; 12929435 ; 12929436 ; 12929437 ; 12929438 ; 12929439 ; 12929441 ; 12929442 ; 12929443 ; 12929444 ; 12929446 ; 12929447 ; 12929448 ; 12929450 ; 12929451 ; 12929452 ; 12929453 ; 12929454 ; 12929455 ; 12929456 ; 12929457 ; 12929458 ; 12929459 ; 12929460 ; 12929461 ; 12929462 ; 12929463 ; 12929464 ; 12929465 ; 12929466 ; 12929467 ; 12929468 ; 12929470 ; 12929471 ; 12929472 ; 12929473 ; 12929474 ; 12929475 ; 12929476 ; 12929477 ; 12929478 ; 12929487 ; 12929488 ; 12929489 ; 12929490 ; 12929492 ; 12929493 ; 12929494 ; 12929495 ; 12929496 ; 12929497 ; 12929499 ; 12929500 ; 12929501 ; 12929502 ; 12929503 ; 12929504 ; 12929505 ; 12929506 ; 12929507 ; 12929508 ; 12929509 ; 12929510 ; 12929511 ; 12929512 ; 12929513 ; 12929515 ; 12929517 ; 12929518 ; 12929521 ; 12929522 ; 12929523 ; 12929524 ; 12929525 ; 12929526 ; 12929529 ; 12929531 ; 12929533 ; 12929534 ; 12929535 ; 12929537 ; 12929538 ; 12929539 ; 12929540 ; 12929541 ; 12929542 ; 12929543 ; 12929544 ; 12929545 ; 12929546 ; 12929547 ; 12929548 ; 12929549 ; 12929550 ; 12929551 ; 12929554 ; 12929555 ; 12929556 ; 12929557 ; 12929558 ; 12929559 ; 12929560 ; 12929561 ; 12929563 ; 12929564 ; 12929565 ; 12929566 ; 12929568 ; 12929569 ; 12929570 ; 12929571 ; 12929572 ; 12929575 ; 12929576 ; 12929577 ; 12929579 ; 12929582 ; 12929583 ; 12929584 ; 12929586 ; 12929587 ; 12929588 ; 12929590 ; 12929591 ; 12929592 ; 12929593 ; 12929594 ; 12929595 ; 12929596 ; 12929597 ; 12929599 ; 12929600 ; 12929601 ; 12929603 ; 12929604 ; 12929605 ; 12929606 ; 12929607 ; 12929608 ; 12929609 ; 12929610 ; 12929611 ; 12929612 ; 12929613 ; 12929614 ; 12929615 ; 12929616 ; 12929618 ; 12929619 ; 12929620 ; 12929621 ; 12929622 ; 12929623 ; 12929624 ; 12929625 ; 12929626 ; 12929627 ; 12929628 ; 12929630 ; 12929631 ; 12929632 ; 12929633 ; 12929634 ; 12929635 ; 12929636 ; 12929637 ; 12929638 ; 12929640 ; 12929641 ; 12929642 ; 12929643 ; 12929644 ; 12929645 ; 12929646 ; 12929647 ; 12929648 ; 12929649 ; 12929650 ; 12929651 ; 12929652 ; 12929653 ; 12929654 ; 12929655 ; 12929656 ; 12929657 ; 12929658 ; 12929659 ; 12929660 ; 12929661 ; 12929662 ; 12929663 ; 12929664 ; 12929665 ; 12929666 ; 12929667 ; 12929668 ; 12929670 ; 12929671 ; 12929672 ; 12929673 ; 12929674 ; 12929675 ; 12929676 ; 12929677 ; 12929678 ; 12929680 ; 12929681 ; 12929682 ; 12929683 ; 12929684 ; 12929685 ; 12929686 ; 12929687 ; 12929688 ; 12929689 ; 12929690 ; 12929691 ; 12929692 ; 12929693 ; 12929694 ; 12929695 ; 12929696 ; 12929697 ; 12929698 ; 12929699 ; 12929700 ; 12929701 ; 12929702 ; 12929704 ; 12929705 ; 12929706 ; 12929707 ; 12929709 ; 12929711 ; 12929712 ; 12929713 ; 12929714 ; 12929715 ; 12929716 ; 12929717 ; 12929718 ; 12929719 ; 12929721 ; 12929722 ; 12929723 ; 12929724 ; 12929725 ; 12929726 ; 12929728 ; 12929729 ; 12929731 ; 12929733 ; 12929734 ; 12929735 ; 12929736 ; 12929738 ; 12929739 ; 12929740 ; 12929741 ; 12929742 ; 12929743 ; 12929744 ; 12929745 ; 12929746 ; 12929747 ; 12929748 ; 12929750 ; 12929751 ; 12929752 ; 12929753 ; 12929755 ; 12929756 ; 12929757 ; 12929758 ; 12929759 ; 12929760 ; 12929761 ; 12929763 ; 12929764 ; 12929765 ; 12929766 ; 12929768 ; 12929769 ; 12929770 ; 12929771 ; 12929772 ; 12929774 ; 12929775 ; 12929776 ; 12929777 ; 12929778 ; 12929780 ; 12929781 ; 12929782 ; 12929784 ; 12929785 ; 12929786 ; 12929788 ; 12929790 ; 12929791 ; 12929792 ; 12929793 ; 12929794 ; 12929795 ; 12929796 ; 12929797 ; 12929798 ; 12929799 ; 12929800 ; 12929801 ; 12929802 ; 12929803 ; 12929804 ; 12929805 ; 12929808 ; 12929809 ; 12929810 ; 12929811 ; 12929812 ; 12929813 ; 12929814 ; 12929815 ; 12929816 ; 12929817 ; 12929818 ; 12929819 ; 12929820 ; 12929821 ; 12929822 ; 12929823 ; 12929824 ; 12929825 ; 12929827 ; 12929828 ; 12929830 ; 12929831 ; 12929832 ; 12929833 ; 12929834 ; 12929835 ; 12929837 ; 12929840 ; 12929844 ; 12929845 ; 12929850 ; 12929853 ; 12929854 ; 12929855 ; 12929857 ; 12929858 ; 12929860 ; 12929861 ; 12929862 ; 12929863 ; 12929864 ; 12929865 ; 12929866 ; 12929867 ; 12929869 ; 12929871 ; 12929872 ; 12929873 ; 12929874 ; 12929876 ; 12929877 ; 12929878 ; 12929879 ; 12929882 ; 12929883 ; 12929888 ; 12929891 ; 12929892 ; 12929893 ; 12929894 ; 12929895 ; 12929896 ; 12929897 ; 12929898 ; 12929899 ; 12929900 ; 12929901 ; 12929902 ; 12929903 ; 12929904 ; 12929905 ; 12929906 ; 12929907 ; 12929909 ; 12929912 ; 12929913 ; 12929914 ; 12929915 ; 12929916 ; 12929918 ; 12929919 ; 12929920 ; 12929921 ; 12929923 ; 12929925 ; 12929926 ; 12929927 ; 12929928 ; 12929931 ; 12929932 ; 12929933 ; 12929934 ; 12929935 ; 12929936 ; 12929937 ; 12929938 ; 12929939 ; 12929940 ; 12929942 ; 12929943 ; 12929944 ; 12929945 ; 12929946 ; 12929947 ; 12929948 ; 12929949 ; 12929950 ; 12929951 ; 12929953 ; 12929954 ; 12929955 ; 12929956 ; 12929957 ; 12929958 ; 12929959 ; 12929960 ; 12929961 ; 12929962 ; 12929963 ; 12929964 ; 12929965 ; 12929967 ; 12929968 ; 12929969 ; 12929973 ; 12929976 ; 12929977 ; 12929978 ; 12929980 ; 12929983 ; 12929987 ; 12929989 ; 12929990 ; 12929994 ; 12929995 ; 12929996 ; 12929997 ; 12929998 ; 12929999 ; 12930000 ; 12930001 ; 12930003 ; 12930004 ; 12930008 ; 12930009 ; 12930011 ; 12930012 ; 12930014 ; 12930015 ; 12930016 ; 12930017 ; 12930018 ; 12930019 ; 12930020 ; 12930021 ; 12930022 ; 12930023 ; 12930025 ; 12930026 ; 12930030 ; 12930031 ; 12930036 ; 12930037 ; 12930038 ; 12930039 ; 12930040 ; 12930042 ; 12930043 ; 12930044 ; 12930045 ; 12930046 ; 12930047 ; 12930048 ; 12930049 ; 12930050 ; 12930053 ; 12930054 ; 12930055 ; 12930056 ; 12930057 ; 12930058 ; 12930060 ; 12930062 ; 12930063 ; 12930064 ; 12930066 ; 12930068 ; 12930069 ; 12930070 ; 12930071 ; 12930072 ; 12930074 ; 12930075 ; 12930076 ; 12930077 ; 12930078 ; 12930080 ; 12930081 ; 12930083 ; 12930084 ; 12930085 ; 12930086 ; 12930087 ; 12930088 ; 12930089 ; 12930091 ; 12930092 ; 12930093 ; 12930094 ; 12930095 ; 12930096 ; 12930097 ; 12930098 ; 12930099 ; 12930101 ; 12930102 ; 12930103 ; 12930104 ; 12930105 ; 12930106 ; 12930107 ; 12930108 ; 12930109 ; 12930110 ; 12930111 ; 12930112 ; 12930113 ; 12930115 ; 12930118 ; 12930119 ; 12930120 ; 12930123 ; 12930128 ; 12930129 ; 12930130 ; 12930131 ; 12930132 ; 12930134 ; 12930136 ; 12930138 ; 12930139 ; 12930140 ; 12930141 ; 12930143 ; 12930144 ; 12930145 ; 12930146 ; 12930148 ; 12930149 ; 12930150 ; 12930152 ; 12930153 ; 12930155 ; 12930156 ; 12930157 ; 12930158 ; 12930160 ; 12930161 ; 12930162 ; 12930163 ; 12930164 ; 12930165 ; 12930167 ; 12930168 ; 12930169 ; 12930170 ; 12930172 ; 12930173 ; 12930174 ; 12930175 ; 12930176 ; 12930177 ; 12930178 ; 12930180 ; 12930182 ; 12930183 ; 12930184 ; 12930185 ; 12930187 ; 12930190 ; 12930191 ; 12930192 ; 12930193 ; 12930194 ; 12930195 ; 12930196 ; 12930197 ; 12930199 ; 12930200 ; 12930201 ; 12930202 ; 12930203 ; 12930204 ; 12930205 ; 12930206 ; 12930207 ; 12930208 ; 12930209 ; 12930210 ; 12930211 ; 12930212 ; 12930213 ; 12930214 ; 12930215 ; 12930216 ; 12930217 ; 12930218 ; 12930219 ; 12930220 ; 12930221 ; 12930222 ; 12930223 ; 12930224 ; 12930225 ; 12930226 ; 12930227 ; 12930228 ; 12930229 ; 12930231 ; 12930232 ; 12930235 ; 12930236 ; 12930237 ; 12930238 ; 12930239 ; 12930240 ; 12930241 ; 12930242 ; 12930243 ; 12930244 ; 12930245 ; 12930246 ; 12930247 ; 12930248 ; 12930249 ; 12930250 ; 12930251 ; 12930252 ; 12930253 ; 12930254 ; 12930255 ; 12930256 ; 12930257 ; 12930258 ; 12930260 ; 12930261 ; 12930262 ; 12930263 ; 12930264 ; 12930265 ; 12930266 ; 12930267 ; 12930268 ; 12930269 ; 12930270 ; 12930271 ; 12930272 ; 12930273 ; 12930274 ; 12930275 ; 12930276 ; 12930277 ; 12930278 ; 12930279 ; 12930280 ; 12930281 ; 12930282 ; 12930283 ; 12930284 ; 12930285 ; 12930286 ; 12930287 ; 12930288 ; 12930289 ; 12930290 ; 12930291 ; 12930292 ; 12930293 ; 12930294 ; 12930295 ; 12930296 ; 12930298 ; 12930299 ; 12930300 ; 12930301 ; 12930302 ; 12930303 ; 12930304 ; 12930305 ; 12930306 ; 12930307 ; 12930308 ; 12930309 ; 12930310 ; 12930311 ; 12930312 ; 12930313 ; 12930314 ; 12930315 ; 12930316 ; 12930317 ; 12930318 ; 12930320 ; 12930321 ; 12930322 ; 12930323 ; 12930324 ; 12930325 ; 12930326 ; 12930328 ; 12930329 ; 12930331 ; 12930332 ; 12930333 ; 12930334 ; 12930335 ; 12930336 ; 12930337 ; 12930338 ; 12930339 ; 12930340 ; 12930341 ; 12930342 ; 12930343 ; 12930344 ; 12930345 ; 12930346 ; 12930347 ; 12930348 ; 12930349 ; 12930350 ; 12930352 ; 12930353 ; 12930354 ; 12930355 ; 12930356 ; 12930357 ; 12930358 ; 12930359 ; 12930360 ; 12930361 ; 12930362 ; 12930363 ; 12930364 ; 12930365 ; 12930366 ; 12930367 ; 12930368 ; 12930369 ; 12930370 ; 12930371 ; 12930372 ; 12930373 ; 12930374 ; 12930375 ; 12930376 ; 12930377 ; 12930378 ; 12930379 ; 12930380 ; 12930381 ; 12930382 ; 12930383 ; 12930384 ; 12930385 ; 12930386 ; 12930387 ; 12930388 ; 12930389 ; 12930390 ; 12930391 ; 12930392 ; 12930393 ; 12930394 ; 12930396 ; 12930398 ; 12930399 ; 12930401 ; 12930402 ; 12930403 ; 12930404 ; 12930405 ; 12930406 ; 12930407 ; 12930408 ; 12930409 ; 12930410 ; 12930411 ; 12930412 ; 12930413 ; 12930414 ; 12930416 ; 12930417 ; 12930421 ; 12930422 ; 12930424 ; 12930425 ; 12930426 ; 12930428 ; 12930429 ; 12930430 ; 12930431 ; 12930432 ; 12930433 ; 12930434 ; 12930435 ; 12930436 ; 12930437 ; 12930438 ; 12930439 ; 12930440 ; 12930441 ; 12930442 ; 12930443 ; 12930444 ; 12930445 ; 12930446 ; 12930447 ; 12930448 ; 12930449 ; 12930450 ; 12930451 ; 12930452 ; 12930453 ; 12930454 ; 12930455 ; 12930456 ; 12930458 ; 12930460 ; 12930462 ; 12930463 ; 12930464 ; 12930465 ; 12930466 ; 12930467 ; 12930468 ; 12930469 ; 12930470 ; 12930471 ; 12930472 ; 12930473 ; 12930474 ; 12930475 ; 12930476 ; 12930477 ; 12930478 ; 12930479 ; 12930480 ; 12930481 ; 12930482 ; 12930483 ; 12930485 ; 12930486 ; 12930488 ; 12930489 ; 12930490 ; 12930491 ; 12930492 ; 12930495 ; 12930496 ; 12930497 ; 12930498 ; 12930499 ; 12930501 ; 12930502 ; 12930503 ; 12930506 ; 12930507 ; 12930508 ; 12930509 ; 12930510 ; 12930511 ; 12930512 ; 12930513 ; 12930514 ; 12930515 ; 12930517 ; 12930518 ; 12930519 ; 12930520 ; 12930523 ; 12930524 ; 12930525 ; 12930526 ; 12930527 ; 12930528 ; 12930530 ; 12930531 ; 12930532 ; 12930534 ; 12930537 ; 12930538 ; 12930539 ; 12930540 ; 12930541 ; 12930542 ; 12930543 ; 12930545 ; 12930546 ; 12930547 ; 12930548 ; 12930550 ; 12930552 ; 12930553 ; 12930554 ; 12930555 ; 12930556 ; 12930557 ; 12930561 ; 12930564 ; 12930565 ; 12930566 ; 12930567 ; 12930568 ; 12930569 ; 12930570 ; 12930571 ; 12930572 ; 12930573 ; 12930575 ; 12930576 ; 12930577 ; 12930578 ; 12930579 ; 12930581 ; 12930582 ; 12930583 ; 12930584 ; 12930585 ; 12930588 ; 12930589 ; 12930590 ; 12930592 ; 12930593 ; 12930594 ; 12930596 ; 12930597 ; 12930598 ; 12930599 ; 12930600 ; 12930601 ; 12930602 ; 12930603 ; 12930604 ; 12930605 ; 12930606 ; 12930607 ; 12930610 ; 12930611 ; 12930613 ; 12930614 ; 12930615 ; 12930616 ; 12930617 ; 12930618 ; 12930620 ; 12930621 ; 12930622 ; 12930623 ; 12930624 ; 12930625 ; 12930626 ; 12930627 ; 12930628 ; 12930632 ; 12930635 ; 12930637 ; 12930638 ; 12930639 ; 12930640 ; 12930642 ; 12930644 ; 12930647 ; 12930648 ; 12930649 ; 12930653 ; 12930657 ; 12930658 ; 12930659 ; 12930660 ; 12930661 ; 12930668 ; 12930669 ; 12930670 ; 12930671 ; 12930672 ; 12930673 ; 12930674 ; 12930676 ; 12930678 ; 12930682 ; 12930683 ; 12930689 ; 12930691 ; 12930692 ; 12930694 ; 12930696 ; 12930697 ; 12930704 ; 12930711 ; 12930712 ; 12930714 ; 12930720 ; 12930721 ; 12930722 ; 12930723 ; 12930725 ; 12930726 ; 12930727 ; 12930728 ; 12930732 ; 12930749 ; 12930758 ; 12930761 ; 12930762 ; 12930766 ; 12930767 ; 12930768 ; 12930769 ; 12930770 ; 12930771 ; 12930772 ; 12930773 ; 12930774 ; 12930775 ; 12930776 ; 12930778 ; 12930779 ; 12930780 ; 12930781 ; 12930782 ; 12930783 ; 12930784 ; 12930785 ; 12930786 ; 12930787 ; 12930788 ; 12930790 ; 12930791 ; 12930792 ; 12930793 ; 12930794 ; 12930795 ; 12930796 ; 12930797 ; 12930798 ; 12930799 ; 12930801 ; 12930803 ; 12930804 ; 12930805 ; 12930806 ; 12930807 ; 12930808 ; 12930809 ; 12930810 ; 12930811 ; 12930813 ; 12930815 ; 12930816 ; 12930817 ; 12930818 ; 12930821 ; 12930822 ; 12930823 ; 12930824 ; 12930825 ; 12930826 ; 12930827 ; 12930828 ; 12930829 ; 12930830 ; 12930831 ; 12930832 ; 12930834 ; 12930841 ; 12930845 ; 12930847 ; 12930848 ; 12930850 ; 12930851 ; 12930852 ; 12930853 ; 12930854 ; 12930857 ; 12930859 ; 12930862 ; 12930866 ; 12930868 ; 12930869 ; 12930871 ; 12930872 ; 12930873 ; 12930875 ; 12930878 ; 12930879 ; 12930880 ; 12930881 ; 12930883 ; 12930886 ; 12930887 ; 12930890 ; 12930891 ; 12930892 ; 12930893 ; 12930894 ; 12930895 ; 12930896 ; 12930898 ; 12930899 ; 12930900 ; 12930902 ; 12930904 ; 12930908 ; 12930909 ; 12930910 ; 12930911 ; 12930912 ; 12930913 ; 12930914 ; 12930915 ; 12930916 ; 12930917 ; 12930918 ; 12930919 ; 12930920 ; 12930921 ; 12930922 ; 12930923 ; 12930924 ; 12930926 ; 12930927 ; 12930928 ; 12930929 ; 12930930 ; 12930931 ; 12930932 ; 12930933 ; 12930935 ; 12930936 ; 12930937 ; 12930938 ; 12930940 ; 12930941 ; 12930942 ; 12930943 ; 12930944 ; 12930945 ; 12930946 ; 12930947 ; 12930948 ; 12930949 ; 12930951 ; 12930952 ; 12930953 ; 12930955 ; 12930956 ; 12930958 ; 12930960 ; 12930961 ; 12930963 ; 12930965 ; 12930968 ; 12930969 ; 12930971 ; 12930972 ; 12930974 ; 12930975 ; 12930976 ; 12930977 ; 12930978 ; 12930979 ; 12930980 ; 12930981 ; 12930982 ; 12930983 ; 12930985 ; 12930986 ; 12930987 ; 12930989 ; 12930990 ; 12930991 ; 12930992 ; 12930994 ; 12930995 ; 12930996 ; 12930997 ; 12930998 ; 12931000 ; 12931001 ; 12931002 ; 12931003 ; 12931004 ; 12931005 ; 12931006 ; 12931007 ; 12931008 ; 12931009 ; 12931011 ; 12931012 ; 12931013 ; 12931014 ; 12931015 ; 12931016 ; 12931017 ; 12931018 ; 12931019 ; 12931020 ; 12931021 ; 12931022 ; 12931023 ; 12931024 ; 12931026 ; 12931027 ; 12931028 ; 12931030 ; 12931031 ; 12931032 ; 12931033 ; 12931034 ; 12931035 ; 12931036 ; 12931037 ; 12931038 ; 12931039 ; 12931040 ; 12931041 ; 12931042 ; 12931043 ; 12931044 ; 12931045 ; 12931046 ; 12931047 ; 12931048 ; 12931049 ; 12931050 ; 12931051 ; 12931052 ; 12931053 ; 12931054 ; 12931055 ; 12931056 ; 12931057 ; 12931058 ; 12931059 ; 12931060 ; 12931061 ; 12931062 ; 12931063 ; 12931064 ; 12931065 ; 12931066 ; 12931067 ; 12931068 ; 12931069 ; 12931070 ; 12931071 ; 12931072 ; 12931073 ; 12931074 ; 12931075 ; 12931076 ; 12931077 ; 12931078 ; 12931079 ; 12931080 ; 12931081 ; 12931082 ; 12931083 ; 12931084 ; 12931085 ; 12931086 ; 12931087 ; 12931089 ; 12931090 ; 12931091 ; 12931092 ; 12931093 ; 12931094 ; 12931095 ; 12931096 ; 12931098 ; 12931099 ; 12931100 ; 12931101 ; 12931103 ; 12931104 ; 12931105 ; 12931106 ; 12931107 ; 12931108 ; 12931109 ; 12931110 ; 12931111 ; 12931112 ; 12931113 ; 12931114 ; 12931115 ; 12931116 ; 12931117 ; 12931119 ; 12931120 ; 12931121 ; 12931122 ; 12931124 ; 12931125 ; 12931126 ; 12931127 ; 12931128 ; 12931129 ; 12931130 ; 12931131 ; 12931132 ; 12931133 ; 12931134 ; 12931135 ; 12931136 ; 12931137 ; 12931138 ; 12931139 ; 12931140 ; 12931141 ; 12931142 ; 12931143 ; 12931144 ; 12931145 ; 12931146 ; 12931147 ; 12931148 ; 12931149 ; 12931150 ; 12931151 ; 12931152 ; 12931153 ; 12931154 ; 12931155 ; 12931156 ; 12931157 ; 12931158 ; 12931159 ; 12931160 ; 12931161 ; 12931162 ; 12931163 ; 12931164 ; 12931165 ; 12931166 ; 12931167 ; 12931168 ; 12931169 ; 12931170 ; 12931171 ; 12931172 ; 12931173 ; 12931174 ; 12931175 ; 12931176 ; 12931177 ; 12931178 ; 12931179 ; 12931180 ; 12931181 ; 12931182 ; 12931183 ; 12931184 ; 12931185 ; 12931186 ; 12931187 ; 12931188 ; 12931189 ; 12931190 ; 12931191 ; 12931192 ; 12931193 ; 12931194 ; 12931195 ; 12931196 ; 12931197 ; 12931198 ; 12931199 ; 12931200 ; 12931201 ; 12931202 ; 12931203 ; 12931204 ; 12931205 ; 12931206 ; 12931207 ; 12931208 ; 12931210 ; 12931211 ; 12931212 ; 12931213 ; 12931214 ; 12931215 ; 12931216 ; 12931217 ; 12931218 ; 12931219 ; 12931220 ; 12931221 ; 12931222 ; 12931223 ; 12931224 ; 12931225 ; 12931226 ; 12931227 ; 12931228 ; 12931229 ; 12931230 ; 12931231 ; 12931232 ; 12931233 ; 12931234 ; 12931235 ; 12931236 ; 12931237 ; 12931238 ; 12931239 ; 12931240 ; 12931241 ; 12931242 ; 12931243 ; 12931244 ; 12931245 ; 12931246 ; 12931248 ; 12931249 ; 12931250 ; 12931251 ; 12931252 ; 12931253 ; 12931254 ; 12931255 ; 12931256 ; 12931257 ; 12931258 ; 12931259 ; 12931260 ; 12931261 ; 12931262 ; 12931263 ; 12931264 ; 12931265 ; 12931266 ; 12931267 ; 12931268 ; 12931269 ; 12931270 ; 12931271 ; 12931272 ; 12931273 ; 12931274 ; 12931275 ; 12931276 ; 12931277 ; 12931278 ; 12931279 ; 12931280 ; 12931281 ; 12931282 ; 12931283 ; 12931284 ; 12931285 ; 12931286 ; 12931287 ; 12931288 ; 12931289 ; 12931290 ; 12931291 ; 12931292 ; 12931293 ; 12931294 ; 12931295 ; 12931296 ; 12931297 ; 12931298 ; 12931299 ; 12931300 ; 12931301 ; 12931302 ; 12931303 ; 12931304 ; 12931305 ; 12931306 ; 12931307 ; 12931308 ; 12931309 ; 12931310 ; 12931311 ; 12931312 ; 12931313 ; 12931314 ; 12931315 ; 12931316 ; 12931317 ; 12931318 ; 12931319 ; 12931320 ; 12931322 ; 12931323 ; 12931324 ; 12931325 ; 12931326 ; 12931327 ; 12931328 ; 12931329 ; 12931330 ; 12931331 ; 12931332 ; 12931333 ; 12931334 ; 12931335 ; 12931336 ; 12931337 ; 12931338 ; 12931339 ; 12931340 ; 12931341 ; 12931342 ; 12931343 ; 12931344 ; 12931345 ; 12931346 ; 12931347 ; 12931348 ; 12931349 ; 12931350 ; 12931351 ; 12931352 ; 12931353 ; 12931354 ; 12931355 ; 12931356 ; 12931357 ; 12931358 ; 12931359 ; 12931360 ; 12931361 ; 12931362 ; 12931363 ; 12931364 ; 12931365 ; 12931366 ; 12931367 ; 12931368 ; 12931369 ; 12931370 ; 12931371 ; 12931372 ; 12931373 ; 12931374 ; 12931375 ; 12931376 ; 12931377 ; 12931379 ; 12931380 ; 12931381 ; 12931382 ; 12931383 ; 12931384 ; 12931385 ; 12931386 ; 12931387 ; 12931388 ; 12931389 ; 12931390 ; 12931391 ; 12931392 ; 12931393 ; 12931394 ; 12931395 ; 12931399 ; 12931400 ; 12931401 ; 12931402 ; 12931403 ; 12931404 ; 12931405 ; 12931406 ; 12931407 ; 12931408 ; 12931410 ; 12931411 ; 12931412 ; 12931413 ; 12931414 ; 12931415 ; 12931416 ; 12931418 ; 12931419 ; 12931420 ; 12931421 ; 12931423 ; 12931424 ; 12931425 ; 12931426 ; 12931427 ; 12931428 ; 12931429 ; 12931430 ; 12931431 ; 12931432 ; 12931433 ; 12931434 ; 12931435 ; 12931436 ; 12931437 ; 12931438 ; 12931439 ; 12931441 ; 12931442 ; 12931443 ; 12931445 ; 12931446 ; 12931447 ; 12931448 ; 12931449 ; 12931450 ; 12931451 ; 12931453 ; 12931455 ; 12931456 ; 12931457 ; 12931458 ; 12931459 ; 12931460 ; 12931461 ; 12931462 ; 12931463 ; 12931464 ; 12931465 ; 12931466 ; 12931467 ; 12931468 ; 12931470 ; 12931471 ; 12931472 ; 12931473 ; 12931474 ; 12931475 ; 12931476 ; 12931477 ; 12931478 ; 12931479 ; 12931480 ; 12931481 ; 12931482 ; 12931483 ; 12931484 ; 12931485 ; 12931486 ; 12931487 ; 12931488 ; 12931489 ; 12931490 ; 12931491 ; 12931492 ; 12931495 ; 12931496 ; 12931498 ; 12931499 ; 12931500 ; 12931501 ; 12931502 ; 12931503 ; 12931504 ; 12931505 ; 12931506 ; 12931507 ; 12931508 ; 12931509 ; 12931510 ; 12931511 ; 12931512 ; 12931513 ; 12931514 ; 12931515 ; 12931516 ; 12931517 ; 12931518 ; 12931520 ; 12931521 ; 12931522 ; 12931523 ; 12931524 ; 12931525 ; 12931526 ; 12931527 ; 12931528 ; 12931529 ; 12931530 ; 12931532 ; 12931533 ; 12931534 ; 12931535 ; 12931536 ; 12931537 ; 12931538 ; 12931539 ; 12931540 ; 12931541 ; 12931542 ; 12931543 ; 12931544 ; 12931545 ; 12931546 ; 12931547 ; 12931548 ; 12931549 ; 12931550 ; 12931551 ; 12931552 ; 12931553 ; 12931555 ; 12931558 ; 12931559 ; 12931560 ; 12931561 ; 12931562 ; 12931563 ; 12931564 ; 12931565 ; 12931566 ; 12931567 ; 12931568 ; 12931569 ; 12931570 ; 12931571 ; 12931572 ; 12931573 ; 12931574 ; 12931575 ; 12931576 ; 12931577 ; 12931578 ; 12931579 ; 12931580 ; 12931581 ; 12931582 ; 12931583 ; 12931584 ; 12931585 ; 12931586 ; 12931587 ; 12931588 ; 12931589 ; 12931590 ; 12931591 ; 12931592 ; 12931593 ; 12931594 ; 12931595 ; 12931596 ; 12931597 ; 12931598 ; 12931599 ; 12931600 ; 12931601 ; 12931602 ; 12931603 ; 12931604 ; 12931605 ; 12931606 ; 12931607 ; 12931608 ; 12931609 ; 12931610 ; 12931611 ; 12931612 ; 12931613 ; 12931614 ; 12931615 ; 12931616 ; 12931617 ; 12931618 ; 12931619 ; 12931620 ; 12931621 ; 12931622 ; 12931623 ; 12931624 ; 12931625 ; 12931626 ; 12931627 ; 12931628 ; 12931629 ; 12931630 ; 12931631 ; 12931632 ; 12931633 ; 12931634 ; 12931635 ; 12931636 ; 12931637 ; 12931638 ; 12931639 ; 12931640 ; 12931641 ; 12931642 ; 12931643 ; 12931644 ; 12931645 ; 12931646 ; 12931647 ; 12931648 ; 12931649 ; 12931650 ; 12931651 ; 12931652 ; 12931653 ; 12931654 ; 12931655 ; 12931656 ; 12931657 ; 12931658 ; 12931659 ; 12931660 ; 12931661 ; 12931662 ; 12931663 ; 12931664 ; 12931665 ; 12931666 ; 12931667 ; 12931668 ; 12931669 ; 12931670 ; 12931671 ; 12931672 ; 12931673 ; 12931676 ; 12931677 ; 12931678 ; 12931681 ; 12931682 ; 12931683 ; 12931684 ; 12931685 ; 12931686 ; 12931687 ; 12931688 ; 12931689 ; 12931690 ; 12931691 ; 12931692 ; 12931693 ; 12931694 ; 12931695 ; 12931696 ; 12931697 ; 12931699 ; 12931701 ; 12931703 ; 12931705 ; 12931707 ; 12931710 ; 12931712 ; 12931713 ; 12931714 ; 12931715 ; 12931717 ; 12931718 ; 12931719 ; 12931720 ; 12931721 ; 12931722 ; 12931723 ; 12931724 ; 12931725 ; 12931726 ; 12931727 ; 12931728 ; 12931729 ; 12931730 ; 12931731 ; 12931735 ; 12931736 ; 12931737 ; 12931738 ; 12931739 ; 12931740 ; 12931744 ; 12931745 ; 12931746 ; 12931747 ; 12931748 ; 12931750 ; 12931751 ; 12931752 ; 12931753 ; 12931754 ; 12931755 ; 12931757 ; 12931758 ; 12931760 ; 12931761 ; 12931763 ; 12931764 ; 12931768 ; 12931770 ; 12931771 ; 12931772 ; 12931773 ; 12931774 ; 12931775 ; 12931776 ; 12931777 ; 12931778 ; 12931779 ; 12931780 ; 12931781 ; 12931782 ; 12931785 ; 12931786 ; 12931787 ; 12931788 ; 12931789 ; 12931790 ; 12931791 ; 12931792 ; 12931793 ; 12931794 ; 12931795 ; 12931797 ; 12931798 ; 12931799 ; 12931800 ; 12931801 ; 12931802 ; 12931803 ; 12931804 ; 12931805 ; 12931806 ; 12931807 ; 12931808 ; 12931809 ; 12931811 ; 12931814 ; 12931815 ; 12931816 ; 12931817 ; 12931818 ; 12931819 ; 12931821 ; 12931822 ; 12931823 ; 12931824 ; 12931825 ; 12931826 ; 12931827 ; 12931828 ; 12931829 ; 12931830 ; 12931831 ; 12931832 ; 12931833 ; 12931834 ; 12931835 ; 12931836 ; 12931837 ; 12931838 ; 12931839 ; 12931840 ; 12931841 ; 12931842 ; 12931843 ; 12931844 ; 12931845 ; 12931846 ; 12931847 ; 12931848 ; 12931849 ; 12931850 ; 12931851 ; 12931853 ; 12931854 ; 12931855 ; 12931856 ; 12931857 ; 12931858 ; 12931859 ; 12931860 ; 12931861 ; 12931862 ; 12931863 ; 12931864 ; 12931865 ; 12931866 ; 12931867 ; 12931869 ; 12931870 ; 12931871 ; 12931872 ; 12931873 ; 12931875 ; 12931877 ; 12931878 ; 12931879 ; 12931880 ; 12931881 ; 12931882 ; 12931883 ; 12931884 ; 12931885 ; 12931886 ; 12931889 ; 12931891 ; 12931892 ; 12931894 ; 12931895 ; 12931896 ; 12931898 ; 12931901 ; 12931902 ; 12931903 ; 12931905 ; 12931906 ; 12931907 ; 12931909 ; 12931910 ; 12931911 ; 12931913 ; 12931914 ; 12931915 ; 12931916 ; 12931917 ; 12931918 ; 12931919 ; 12931920 ; 12931921 ; 12931922 ; 12931923 ; 12931924 ; 12931926 ; 12931927 ; 12931928 ; 12931930 ; 12931931 ; 12931932 ; 12931933 ; 12931934 ; 12931935 ; 12931937 ; 12931938 ; 12931940 ; 12931941 ; 12931942 ; 12931943 ; 12931944 ; 12931945 ; 12931946 ; 12931947 ; 12931948 ; 12931949 ; 12931950 ; 12931953 ; 12931955 ; 12931956 ; 12931957 ; 12931958 ; 12931960 ; 12931961 ; 12931962 ; 12931963 ; 12931965 ; 12931966 ; 12931967 ; 12931968 ; 12931969 ; 12931970 ; 12931972 ; 12931973 ; 12931974 ; 12931975 ; 12931976 ; 12931977 ; 12931978 ; 12931979 ; 12931980 ; 12931981 ; 12931982 ; 12931983 ; 12931985 ; 12931986 ; 12931987 ; 12931990 ; 12931991 ; 12931992 ; 12931993 ; 12931994 ; 12931995 ; 12931996 ; 12931997 ; 12931998 ; 12931999 ; 12932000 ; 12932001 ; 12932002 ; 12932003 ; 12932004 ; 12932005 ; 12932006 ; 12932007 ; 12932009 ; 12932010 ; 12932011 ; 12932012 ; 12932015 ; 12932016 ; 12932018 ; 12932020 ; 12932021 ; 12932022 ; 12932023 ; 12932024 ; 12932025 ; 12932026 ; 12932028 ; 12932029 ; 12932030 ; 12932031 ; 12932032 ; 12932033 ; 12932034 ; 12932035 ; 12932036 ; 12932038 ; 12932039 ; 12932040 ; 12932041 ; 12932043 ; 12932044 ; 12932045 ; 12932046 ; 12932047 ; 12932049 ; 12932050 ; 12932051 ; 12932052 ; 12932054 ; 12932055 ; 12932056 ; 12932057 ; 12932058 ; 12932060 ; 12932062 ; 12932063 ; 12932064 ; 12932067 ; 12932068 ; 12932069 ; 12932070 ; 12932072 ; 12932073 ; 12932074 ; 12932075 ; 12932076 ; 12932077 ; 12932078 ; 12932079 ; 12932080 ; 12932081 ; 12932082 ; 12932083 ; 12932084 ; 12932085 ; 12932086 ; 12932088 ; 12932089 ; 12932090 ; 12932091 ; 12932092 ; 12932094 ; 12932095 ; 12932096 ; 12932097 ; 12932098 ; 12932099 ; 12932100 ; 12932102 ; 12932103 ; 12932105 ; 12932106 ; 12932107 ; 12932108 ; 12932111 ; 12932112 ; 12932113 ; 12932114 ; 12932115 ; 12932116 ; 12932118 ; 12932119 ; 12932121 ; 12932122 ; 12932123 ; 12932124 ; 12932125 ; 12932126 ; 12932127 ; 12932128 ; 12932129 ; 12932130 ; 12932131 ; 12932132 ; 12932133 ; 12932135 ; 12932137 ; 12932138 ; 12932139 ; 12932140 ; 12932141 ; 12932142 ; 12932143 ; 12932144 ; 12932145 ; 12932146 ; 12932147 ; 12932148 ; 12932149 ; 12932150 ; 12932151 ; 12932152 ; 12932153 ; 12932154 ; 12932155 ; 12932156 ; 12932157 ; 12932158 ; 12932159 ; 12932161 ; 12932162 ; 12932163 ; 12932164 ; 12932165 ; 12932166 ; 12932167 ; 12932168 ; 12932169 ; 12932170 ; 12932172 ; 12932173 ; 12932174 ; 12932175 ; 12932176 ; 12932177 ; 12932178 ; 12932179 ; 12932180 ; 12932181 ; 12932182 ; 12932183 ; 12932184 ; 12932185 ; 12932186 ; 12932189 ; 12932191 ; 12932192 ; 12932193 ; 12932194 ; 12932196 ; 12932197 ; 12932198 ; 12932200 ; 12932201 ; 12932203 ; 12932206 ; 12932207 ; 12932208 ; 12932209 ; 12932210 ; 12932211 ; 12932212 ; 12932213 ; 12932214 ; 12932215 ; 12932216 ; 12932217 ; 12932218 ; 12932219 ; 12932220 ; 12932221 ; 12932222 ; 12932223 ; 12932225 ; 12932227 ; 12932228 ; 12932229 ; 12932230 ; 12932231 ; 12932232 ; 12932234 ; 12932235 ; 12932236 ; 12932237 ; 12932238 ; 12932239 ; 12932241 ; 12932243 ; 12932244 ; 12932245 ; 12932246 ; 12932248 ; 12932249 ; 12932250 ; 12932251 ; 12932252 ; 12932253 ; 12932255 ; 12932257 ; 12932258 ; 12932259 ; 12932260 ; 12932261 ; 12932262 ; 12932263 ; 12932264 ; 12932265 ; 12932268 ; 12932271 ; 12932272 ; 12932273 ; 12932274 ; 12932275 ; 12932276 ; 12932278 ; 12932280 ; 12932281 ; 12932282 ; 12932283 ; 12932284 ; 12932285 ; 12932286 ; 12932287 ; 12932288 ; 12932289 ; 12932290 ; 12932291 ; 12932292 ; 12932293 ; 12932294 ; 12932295 ; 12932296 ; 12932297 ; 12932298 ; 12932299 ; 12932300 ; 12932301 ; 12932302 ; 12932303 ; 12932304 ; 12932305 ; 12932307 ; 12932308 ; 12932309 ; 12932311 ; 12932312 ; 12932313 ; 12932314 ; 12932315 ; 12932316 ; 12932317 ; 12932318 ; 12932319 ; 12932320 ; 12932321 ; 12932323 ; 12932325 ; 12932326 ; 12932327 ; 12932328 ; 12932329 ; 12932330 ; 12932331 ; 12932332 ; 12932333 ; 12932334 ; 12932335 ; 12932336 ; 12932337 ; 12932338 ; 12932339 ; 12932340 ; 12932341 ; 12932342 ; 12932343 ; 12932344 ; 12932345 ; 12932346 ; 12932347 ; 12932348 ; 12932349 ; 12932350 ; 12932351 ; 12932352 ; 12932353 ; 12932354 ; 12932355 ; 12932356 ; 12932357 ; 12932358 ; 12932359 ; 12932360 ; 12932361 ; 12932362 ; 12932363 ; 12932364 ; 12932365 ; 12932366 ; 12932367 ; 12932368 ; 12932369 ; 12932370 ; 12932371 ; 12932372 ; 12932373 ; 12932374 ; 12932376 ; 12932377 ; 12932378 ; 12932379 ; 12932380 ; 12932381 ; 12932382 ; 12932383 ; 12932384 ; 12932385 ; 12932386 ; 12932387 ; 12932388 ; 12932389 ; 12932390 ; 12932391 ; 12932392 ; 12932393 ; 12932394 ; 12932395 ; 12932396 ; 12932397 ; 12932398 ; 12932399 ; 12932400 ; 12932401 ; 12932402 ; 12932403 ; 12932404 ; 12932405 ; 12932406 ; 12932407 ; 12932408 ; 12932409 ; 12932410 ; 12932411 ; 12932412 ; 12932413 ; 12932414 ; 12932415 ; 12932416 ; 12932417 ; 12932418 ; 12932420 ; 12932421 ; 12932422 ; 12932423 ; 12932424 ; 12932425 ; 12932426 ; 12932427 ; 12932428 ; 12932429 ; 12932430 ; 12932431 ; 12932432 ; 12932433 ; 12932434 ; 12932435 ; 12932436 ; 12932437 ; 12932438 ; 12932439 ; 12932440 ; 12932441 ; 12932442 ; 12932443 ; 12932444 ; 12932448 ; 12932449 ; 12932450 ; 12932451 ; 12932452 ; 12932454 ; 12932459 ; 12932460 ; 12932461 ; 12932462 ; 12932463 ; 12932464 ; 12932465 ; 12932466 ; 12932467 ; 12932468 ; 12932469 ; 12932470 ; 12932471 ; 12932472 ; 12932473 ; 12932474 ; 12932475 ; 12932476 ; 12932477 ; 12932478 ; 12932479 ; 12932480 ; 12932481 ; 12932482 ; 12932483 ; 12932484 ; 12932485 ; 12932486 ; 12932487 ; 12932488 ; 12932489 ; 12932490 ; 12932491 ; 12932492 ; 12932493 ; 12932494 ; 12932495 ; 12932496 ; 12932498 ; 12932499 ; 12932500 ; 12932501 ; 12932502 ; 12932505 ; 12932506 ; 12932508 ; 12932509 ; 12932510 ; 12932511 ; 12932512 ; 12932513 ; 12932514 ; 12932515 ; 12932516 ; 12932517 ; 12932518 ; 12932519 ; 12932520 ; 12932521 ; 12932522 ; 12932523 ; 12932524 ; 12932525 ; 12932526 ; 12932527 ; 12932528 ; 12932529 ; 12932530 ; 12932531 ; 12932532 ; 12932533 ; 12932534 ; 12932535 ; 12932536 ; 12932537 ; 12932538 ; 12932539 ; 12932540 ; 12932541 ; 12932542 ; 12932543 ; 12932544 ; 12932546 ; 12932547 ; 12932548 ; 12932549 ; 12932550 ; 12932551 ; 12932552 ; 12932553 ; 12932555 ; 12932556 ; 12932557 ; 12932558 ; 12932559 ; 12932560 ; 12932562 ; 12932563 ; 12932564 ; 12932567 ; 12932568 ; 12932569 ; 12932570 ; 12932571 ; 12932572 ; 12932573 ; 12932574 ; 12932575 ; 12932576 ; 12932577 ; 12932579 ; 12932580 ; 12932581 ; 12932583 ; 12932584 ; 12932585 ; 12932586 ; 12932587 ; 12932588 ; 12932589 ; 12932590 ; 12932591 ; 12932592 ; 12932594 ; 12932596 ; 12932597 ; 12932598 ; 12932599 ; 12932600 ; 12932601 ; 12932602 ; 12932603 ; 12932604 ; 12932605 ; 12932606 ; 12932607 ; 12932610 ; 12932612 ; 12932613 ; 12932614 ; 12932615 ; 12932616 ; 12932617 ; 12932618 ; 12932619 ; 12932620 ; 12932621 ; 12932622 ; 12932623 ; 12932624 ; 12932625 ; 12932626 ; 12932627 ; 12932628 ; 12932629 ; 12932630 ; 12932631 ; 12932632 ; 12932633 ; 12932634 ; 12932635 ; 12932636 ; 12932637 ; 12932638 ; 12932639 ; 12932640 ; 12932641 ; 12932642 ; 12932643 ; 12932644 ; 12932645 ; 12932646 ; 12932647 ; 12932648 ; 12932649 ; 12932650 ; 12932651 ; 12932652 ; 12932653 ; 12932654 ; 12932655 ; 12932656 ; 12932657 ; 12932658 ; 12932659 ; 12932660 ; 12932661 ; 12932662 ; 12932663 ; 12932664 ; 12932665 ; 12932666 ; 12932667 ; 12932669 ; 12932670 ; 12932671 ; 12932672 ; 12932673 ; 12932674 ; 12932675 ; 12932676 ; 12932677 ; 12932678 ; 12932679 ; 12932680 ; 12932681 ; 12932682 ; 12932683 ; 12932684 ; 12932685 ; 12932686 ; 12932687 ; 12932688 ; 12932689 ; 12932690 ; 12932691 ; 12932692 ; 12932693 ; 12932694 ; 12932695 ; 12932696 ; 12932697 ; 12932698 ; 12932699 ; 12932700 ; 12932701 ; 12932702 ; 12932703 ; 12932704 ; 12932706 ; 12932707 ; 12932708 ; 12932709 ; 12932710 ; 12932711 ; 12932712 ; 12932713 ; 12932714 ; 12932715 ; 12932716 ; 12932717 ; 12932718 |
Recalling Firm/ Manufacturer |
Hospira Inc. 275 N Field Dr Lake Forest IL 60045-2579
|
For Additional Information Contact |
Hospira Global Complaint Management 800-441-4100
|
Manufacturer Reason for Recall |
Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.
|
FDA Determined Cause 2 |
Component design/selection |
Action |
Hospira sent an IMPORTANT DEVICE INFORMATION letter dated March 8, 2014 to all affected customers.. The letter advised customers to: 1) Perform the Performance Verification Test (PVT) Occlusion Test as defined in the PCA Technical Service Manual (TSM); 2) If the device does not pass the test, remove it from clinical service and contact Hospira at 800-441-4100 (Monday - Friday, 8:00 AM - 5:00 PM CST) to report the issue; and 3) Perform the appropriate troubleshooting and repair activities defined by your facility, which may include returning the device to Hospira for further diagnosing and servicing.
The letter advised the Hospira is in the process of establishing a useful life for the half nut and to determine when it will require replacement. A new Technical Service Manual will be issued in late 2014 that adds a requirement for annual PVT Occlusion testing. The System Operating Manual will include new instructions to reduce the excessive wear on the half nut resulting from incorrectly using the vial information.
Customers who further distributed the devices are requested to notify their customers of the recall and ask them to call Stericycle at 888-965-5801 (Monday - Friday, 8:00 AM - 5:00 PM EST) to receive a reply form.
Questions about this recall should be direct to Hospira at 800-441-4100 (Monday - Friday, 8:00 AM - 5:00 PM CST). |
Quantity in Commerce |
8,216 infusion pump systems |
Distribution |
Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY) + District of Columbia + Puerto Rico + Virgin Islands; *** FOREIGN: |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MEA and Original Applicant = HOSPIRA, INC.
|
|
|
|