Class 2 Device Recall Boston Scientific COGNIS CRTD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Boston Scientific COGNIS CRTDsee related information
Date Initiated by FirmSeptember 17, 2014
Date PostedOctober 18, 2014
Recall Status1 Terminated 3 on February 20, 2015
Recall NumberZ-0087-2015
Recall Event ID 69291
PMA NumberP010012S165 
Product Classification Implantable pulse generator, pacemaker (non-CRT) - Product Code LWP
ProductBoston Scientific COGNIS CRT-D, Cardiac Resynchronization Therapy Defibrillator, models N118, N119. The following models are not available in the US: models N106, N107, P106, P107. Sterilized using ethylene oxide. Product Usage: Boston Scientific cardiac resynchronization therapy defibrillators (CRT-Ds) are indicated for patients with moderate to severe heart failure who remain symptomatic despite stable, optimal heart failure drug therapy and have left ventricular dysfunction and QRS duration > or equal to 120 ms. This family of pulse generators provide a variety of therapies including: -Ventricular tachyarrhythmia therapy, which is used to treat rhythms associated with sudden cardiac death such as VT and VF; -Cardiac Resynchronization Therapy (CRT), which treats heart failure by resynchronizing ventricular contractions through biventricular electrical stimulation; -Bradycardia pacing, including adaptive rate pacing, to detect and treat bradyarrhythmias and to provide cardiac rate support after defibrillation therapy.
Code Information Model N106   - Serial numbers: 222276, 222277, 222279, 222280, 222282, 222283, 222286, 222289, 222290, 222291, 222292, 222293, 222295, 222311, 222312, 222314, 222315, 222316, 222318, 222320, 222321, 222324, 222325, 222335, 222338, 222340, 222342, 222343, 222346, 222361, 222363, 222366, 222372, 222374, 222409.   Model N107   - Serial numbers: 100278, 100279, 100280, 100281, 100282, 100283, 100284, 100285, 100286, 100287, 100288, 100289, 100290, 100291, 100293, 100294, 100295, 100297, 100298, 100299, 100300, 100301, 100303, 100304, 100305, 100306, 100307, 100308, 100309, 100311, 100312, 100314, 100315, 100317, 100318, 100319, 100320, 100321, 100322, 100324, 100325, 100327, 100328, 100329, 100330, 100331, 100333, 100334, 100335, 100336, 100408, 100409, 100410, 100411, 100413, 100414, 100417, 100418, 100419, 100420, 100421, 100422, 100423, 100425, 100426, 100427, 100429, 100430, 100431, 100432, 100433, 100434, 100435, 100436, 100437, 100438, 100439, 100441, 100442, 100445, 100446, 100447, 100448, 100449, 100450, 100451, 100453, 100454, 100455, 100456, 100458, 100459, 100460, 100461, 100462, 100463, 100464, 100466, 100467, 100468, 100469, 100470, 100471, 100472, 100474, 100475, 100476, 100477, 100478, 100479, 100480, 100481, 100482, 100483, 100484, 100485, 100486, 100488, 100489, 100492, 100494, 100495, 100496, 100497, 100498, 100499, 100500, 100501, 100503, 100505, 100506, 100507, 100509, 100518, 100525, 100527, 100531, 100534, 100535, 100536, 100538, 100539, 100540, 100541, 100542, 100543, 100544, 100546, 100547, 100548, 100549, 100550, 100551, 100554, 100555, 100556, 100557, 100558, 100559, 100562, 100563, 100564, 100565, 100566, 100567, 100854, 100855, 100856, 100857, 100858, 100860, 100862, 100863, 100864, 100865, 100866, 100867, 100868, 100869, 100870, 100871, 100872, 100873.   Model N108   - Serial numbers: 666667, 666668, 666669, 666670, 666671, 666672, 666673, 666675, 666676, 666677, 666678, 666679, 666680, 666681, 666682, 666683, 666684, 666685, 666686, 666687, 666688, 666689, 666690, 666691, 666693, 666694, 666695, 666696, 666697, 666699, 666700, 666702, 666703, 666704, 666705, 666706, 666707, 666708, 666710, 666711, 666713, 666714, 666715, 666716, 666717, 666718, 666719, 666720, 666721.   Model N118   - Serial numbers: , 565136, 565138, 565140, 565146, 565149, 565151, 565152, 565154, 565155, 565157, 565158, 565160, 565163, 565164, 565165, 565167, 565168, 565170, 565173, 565176, 565177, 565178, 565179, 565181, 565182, 565186, 565187, 565188, 565189, 565190, 565191, 565192, 565193, 565194, 565197, 565198, 565199, 565200, 565202, 565204, 565209, 565210, 565319, 565320, 565321, 565326, 565328, 565329, 565330, 565331, 565332, 565334, 565335, 565336, 565337, 565338, 565339, 565342, 565343, 565344, 565345, 565356, 565359, 565363, 565364, 565365, 565367, 565368, 565371, 565374, 565375, 565376, 565378, 565379, 565383, 565384, 565386, 565388, 565390, 565394, 565396, 565397, 565398, 565401, 565402, 565404, 565406, 565407, 565408, 565409, 565410, 565412, 565415, 565416, 565418, 565421, 565427, 565429, 565433, 565434, 565437, 565440, 565441, 565442, 565444, 565446, 565447, 565448, 565449, 565450, 565452, 565454, 565455, 565457, 565460, 565467, 565469, 565471, 565475, 565476, 565478, 565480, 565482, 565483, 565485, 565488, 565489, 565490, 565492, 565494, 565495, 565497, 565501, 565502, 565503, 565508, 565510, 565511, 565513, 565517, 565519, 565523, 565524, 565525, 565529, 565531, 565532, 565535, 565537, 565540, 565542, 565543, 565544, 565545, 565550, 565551, 565555, 565556, 565558, 565562, 565563, 565566, 565567, 565568, 565569, 565570, 565571, 565572, 565573, 565575, 565577, 565579, 565580, 565581, 565583, 565584, 565587, 565589, 565590, 565592, 565594, 565595, 565597, 565598, 565603, 565605, 565606, 565607, 565608, 565610, 565611, 565615, 565622, 565627, 565628, 565629, 565630, 565631, 565632, 565633, 565634, 565637, 565638, 565640, 565643, 565644, 565645, 565646, 565648, 565649, 565650, 565653, 565654, 565656.565659, 565662, 565664, 565669, 565670, 565671, 565676, 565677, 565678, 565683, 565684, 565685, 565687, 565688, 565689, 565690, 565692, 565693, 565695, 565697, 565700, 565701, 565702, 565708, 565710, 565711, 565712, 565713, 565717, 565720, 565724, 565735, 565737, 565738, 565741, 565743, 565746, 565747, 565748, 565749, 565750, 565751, 565752, 565753, 565755, 565757, 565759, 565760, 565762, 565764, 565765, 565767, 565768, 565769, 565771, 565772, 565774, 565775, 565777, 565779, 565780, 565782, 565783, 565784, 565785, 565788, 565789, 565793, 565794, 565797, 565798, 565800, 565801, 565802, 565807, 565808, 565809, 565811, 565812, 565815, 565816, 565817, 565818, 565882, 565883, 565884, 565887, 565888, 565890, 565895, 565897, 565898, 565899, 568886, 568887, 568895, 568901, 568904, 568909, 568910, 568911, 568912, 568915, 568916, 568921, 568923, 568926, 568931, 568939, 568942, 568943, 568944, 568945, 568949, 568950, 568952, 568954, 568958, 568962, 568970, 568971, 568972, 568975, 568976, 568977, 568978, 568979, 568981, 568984, 568985, 568986, 568987, 568992, 568994, 568995, 568996, 568997, 569000, 569001, 569004, 569007, 569010, 569011, 569012, 569014, 569015, 569016, 569017, 569018, 569021, 569022, 569027, 569031, 569032, 569033, 569035, 569038, 569040, 569041, 569042, 569046, 569052, 569057, 569059, 569060, 569061, 569062, 569064, 569065, 569066, 569067, 569068, 569069, 569070, 569077, 569080, 569084, 569086, 569091, 569093, 569094, 569097, 569099, 569103, 569106, 569110, 569116, 569119, 569120, 569122, 569123, 569127, 569129, 569130, 569131, 569134, 569138, 569139, 569141, 569142, 569143, 569144, 569145, 569147, 569148, 569149, 569151, 569154, 569156, 569157, 569160, 569161, 569164, 569165, 569166, 569168, 569169, 569170, 569177, 569178, 569180, 569183, 569184, 569186, 569187, 569188, 569193, 569196, 569197, 569198, 569199, 569200, 569202, 569203, 569206, 569211, 569212, 569215, 569216, 569217, 569218, 569219, 569220, 569225, 569226, 569228, 569230, 569231, 569234, 569235, 569236, 569238, 569240, 569243, 569244, 569245, 569246, 569247, 569250, 569253, 569254, 569256, 569258, 569259, 569260, 569261, 569264, 569265, 569269, 569271, 569277, 569282, 569285, 569286, 569287, 569289, 569290, 569292, 569295, 569296, 569297, 569298, 569307, 569311, 569314, 569319, 569327, 569328, 569329, 569337, 569339, 569340, 569341, 569342, 569343, 569346, 569347, 569352, 569354, 569357, 569358, 569360, 569362, 569364, 569365, 569366, 569368, 569371, 569372, 569376, 569378, 569381, 569383, 569384, 569387, 569390, 569393, 569394, 569395, 569402, 569404, 569405, 569406, 569411, 569412, 569413, 569415, 569416, 569418, 569421, 569422, 569425, 569426, 569427, 569434, 569438, 569442, 569443, 569445, 569449, 569450, 569452, 569453, 569456, 569457, 569459, 569460, 569461, 569462, 569465, 569469, 569471, 569472, 569473, 569474, 569475, 569478, 569480, 569482, 569486, 569487, 569489, 569490, 569491, 569493, 569494, 569495, 569496, 569497, 569502, 569503, 569504, 569505, 569506, 569508, 569509, 569513, 569514, 569516, 569517, 569518, 569519, 569521, 569522, 569523, 569524, 569526, 569527, 569531, 569532, 569533, 569534, 569537, 569541, 569543, 569544, 569546, 569548, 569549, 569552, 569554, 569555, 569563, 569564, 569568, 569576, 569577, 569578, 569582, 569585, 569590, 569591, 569595, 569596, 569599, 569602, 569606, 569611, 569613, 569614, 569618, 569624, 569627, 569637, 569639, 569641, 569647, 569650, 569653, 569655, 569656, 569657, 569659, 569663, 569664, 569666, 569668, 569673, 569674, 569675, 569678, 569679, 569682, 569684, 569685, 569686, 569687, 569689, 569691, 569692, 569694, 569697, 569699, 569700, 569701, 569709, 569711, 569712, 569713, 569715, 569719, 569720, 569721, 569724, 569728, 569731, 569733, 569734, 569735, 569736, 569737, 569739, 569741, 569747, 569748, 569749, 569750, 569751, 569752, 569753, 569756, 569759, 569760, 569762, 569763, 569765, 569774, 569778, 569781, 569783, 569784, 569787, 569789, 569792, 569793, 569794, 569795, 569797, 569799, 569801.569802, 569803, 569804, 569805, 569807, 569812, 569813, 569815, 569816, 569817, 569819, 569823, 569824, 569826, 569829, 569830, 569832, 569833, 569834, 569836, 569837, 569838, 569839, 569841, 569843, 569845, 569846, 569855, 569856, 569857, 569858, 569860, 569864, 569865, 569868, 569870, 569871, 569872, 569873, 569877, 569878, 569879, 569880, 569884, 569887, 569889, 569890, 569892, 569894, 569895, 569896, 569897, 569900, 569901, 569902, 569904, 569909, 569912, 569917, 569920, 569921, 569927, 569930, 569933, 569935, 569939, 569942, 569943, 569945, 569946, 569949, 569950, 569951, 569954, 569955, 569957, 569959, 569961, 569962, 569963, 569965, 569967, 569968, 569969, 569972, 569973, 569974, 569976, 569978, 569980, 569983, 569984, 569986, 569989, 569991, 569992, 569993, 569995, 569997, 569999, 570001, 570006, 570007, 570009, 570012, 570013, 570014, 570017, 570018, 570019, 570020, 570024, 570026, 570027, 570029, 570036, 570037, 570041, 570042, 570043, 570048, 570050, 570051, 570053, 570054, 570066, 570077, 570078, 570079, 570080, 570082, 570084, 570085, 570090, 570093, 570097, 570098, 570100, 570102, 570104, 570105, 570107, 570109, 570110, 570112, 570114, 570116, 570117, 570118, 570120, 570121, 570123, 570124, 570125, 570128, 570131, 570135, 570140, 577912, 577914, 577917, 577919, 577923, 577929, 577931, 577935, 577943, 577944, 577945, 577946, 577949, 577950, 577952, 577953, 577954, 577956, 577959, 577961, 577963, 577965, 577966, 577967, 577968, 577969, 577973, 577974, 577977, 577978, 577983, 577986, 577990, 577993, 577997, 577998, 578000, 578002, 578004, 578005, 578006, 578007, 578010, 578011, 578013, 578015, 578021, 578024, 578027, 578028, 578030, 578032, 578033, 578036, 578037, 578039, 578040, 578043, 578045, 578048, 578049, 578053, 578054, 578058, 578061, 578062, 578063, 578064, 578065, 578066, 578073, 578075, 578076, 578079, 578086, 578087, 578089, 578092, 578101, 578103, 578105, 578106, 578108, 578111, 578112, 578114, 578119, 578120, 578121, 578122, 578123, 578124, 578125, 578126, 578127, 578129, 578135, 578137, 578142, 578146, 578150, 578151, 578153, 578154, 578159, 578163, 578164, 578166, 578167, 578169, 578171, 578173, 578177, 578178, 578182, 578183, 578186, 581905, 581911, 581913, 581914, 581916, 581917, 581919, 581920, 581921, 581924, 581925, 581926, 581927, 581928, 581929, 581930, 581932, 581934, 581935, 581936, 581939, 581940, 581941, 581942, 581944, 581946, 581947, 581948, 581950, 581956, 581959, 581960, 581961, 581963, 581964, 581966, 581968, 581972, 581973, 581974, 581976, 581978, 581984, 581985, 581987, 581989, 581990, 581992, 581993, 581995, 581998, 581999, 582001, 582002, 582004, 582005, 582008, 582009, 582010, 582012, 582015, 582016, 582018, 582020, 582022, 582023, 582025, 582026, 582027, 582029, 582035, 582036, 582037, 582038, 582040, 582041, 582042, 582043, 582044, 582045, 582046, 582047, 582048, 582049, 582050, 582052, 582053, 582054, 582055, 582056, 582057, 582059, 582062, 582064, 582066, 582067, 582071, 582074, 582077, 582079, 582080, 582081, 582082, 582083, 582084, 582086, 582087, 582088, 582090, 582092, 582093, 582094, 582096, 582104, 582108, 582109, 582110, 582112, 582113, 582114, 582115, 582117, 582118, 582119, 582121, 582125, 582126, 582128, 582130, 582132, 582133, 582135, 582136, 582137, 582139, 582140, 582142, 582143, 582144, 582145, 582148, 582151, 582153, 582155, 582156, 582157, 582158, 582161, 582165, 582166, 582169, 582171, 582173, 582174, 582175, 582178, 582179, 582180, 582182, 582183, 582184, 582185, 582187, 582188, 582189, 582191, 582192, 582193, 582195, 582198, 582200, 582201, 582202, 582203, 582204, 582206, 582208, 582209, 582210, 582214, 582215, 582216, 582219, 582221, 582224, 582226, 582227, 582230, 582234, 582235, 582236, 582240, 582242, 582243, 582245, 582246, 582247, 582249, 582250, 582252, 582255, 582257, 582258, 582259, 582260, 582261, 582263, 582264, 582267, 582268, 582269, 582270, 582271, 582275, 582276, 582277, 582279, 582280, 582281, 582283, 582284, 582285, 582286, 582288, 582290.582292, 582297, 582299, 582301, 582303, 582304, 582305, 582306, 582308, 582309, 582310, 582311, 582312, 582313, 582314, 582315, 582316, 582318, 582319, 582320, 582321, 582323, 582325, 582327, 582331, 582332, 582333, 582335, 582337, 582340, 582341, 582342, 582343, 582344, 582345, 582346, 582347, 582350, 582351, 582353, 582355, 582356, 582358, 582360, 582361, 582362, 582363, 582364, 582366, 582368, 582369, 582371, 582373, 582374, 582375, 582376, 582377, 582378, 582380, 582381, 582383, 582385, 582387, 582388, 582389, 582392, 582394, 582397, 582399, 582401, 582404, 582408, 582410, 582413, 582419, 582420, 582421, 582422, 582423, 582425, 582430, 582431, 582434, 582436, 582437, 582444, 582445, 582446, 582447, 582448, 582449, 582450, 582452, 582455, 582458, 582459, 582460, 582462, 582464, 582466, 582468, 582469, 582470, 582472, 582473, 582475, 582476, 582477, 582478, 582480, 582482, 582483, 582485, 582486, 582487, 582490, 582491, 582492, 582493, 582494, 582496, 582501, 582503, 582504, 582505, 582510, 582511, 582512, 582513, 582514, 582515, 582518, 582520, 582522, 582523, 582529, 582530, 582531, 582533, 582534, 582536, 582540, 582541, 582544, 582545, 582547, 582548, 582549, 582550, 582551, 582554, 582555, 582556, 582557, 582559, 582560, 582561, 582562, 582563, 582564, 582565, 582566, 582570, 582571, 582572, 582573, 582574, 582575, 582576, 582577, 582578, 582581, 582582, 582583, 582585, 582586, 582587, 582588, 582589, 582590, 582592, 582594, 582597, 582598, 582600, 582601, 582603, 582606, 582608, 582609, 582614, 582616, 582620, 582623, 582628, 582629, 582630, 582632, 582633, 582634, 582635, 582637, 582640, 582644, 582646, 582648, 582649, 582650, 582651, 582652, 582653, 582654, 582659, 582660, 582661, 582662, 582665, 582666, 582670, 582671, 582674, 582675, 582676, 582678, 582679, 582682, 582684, 582686, 582688, 582689, 582690, 582691, 582692, 582693, 582697, 582698, 582699, 582701, 582703, 582704, 582705, 582706, 582708, 582713, 582714, 582716, 582719, 582726, 582727, 582730, 582731, 582732, 582735, 582736, 582739, 582741, 582744, 582745, 582747, 582749, 582750, 582751, 582752, 582753, 582754, 582756, 582757, 582758, 582761, 582765, 582767, 582771, 582772, 582773, 582775, 582778, 582779, 582780, 582781, 582782, 582783, 582785, 582786, 582787, 582788, 582790, 582791, 582792, 582793, 582794, 582797, 582799, 582801, 582803, 582805, 582807, 582809, 582810, 582813, 582814, 582818, 582820, 582822, 582823, 582824, 582827, 582830, 582831, 582835, 582836, 582837, 582839, 582841, 582842, 582844, 582845, 582847, 582849, 582850, 582851, 582853, 582855, 582856, 582857, 582863, 582864, 582865, 582869, 582870, 582873, 582874, 582876, 582881, 582882, 582884, 582885, 582889, 582890, 582891, 582892, 582893, 582894, 582896, 582899, 582902, 582905, 582907, 582908, 582910, 582912, 582913, 582914, 582915, 582916, 582917, 582919, 582923, 582924, 582925, 582927, 582929, 582930, 582931, 582933, 582934, 582936, 582937, 582938, 582940, 582941, 582942, 582943, 582944, 582948, 582951, 582952, 582953, 582954, 582955, 582956, 582958, 582959, 582960, 582962, 582965, 582966, 582970, 582971, 582972, 582975, 582976, 582977, 582978, 582979, 582981, 582983, 582984, 582986, 582988, 582989, 582990, 582991, 582996, 582997, 582999, 583000, 583002, 583005, 583006, 583008, 583009, 583013, 583015, 583017, 583018, 583023, 583024, 583027, 583028, 583029, 583031, 583032, 583033, 583034, 583035, 583036, 583038, 583039, 583043, 583045, 583046, 583049, 583050, 583051, 583052, 583053, 583054, 583055, 583056, 583060, 583061, 583062, 583063, 583064, 583065, 583066, 583069, 583071, 583072, 583073, 583076, 583078, 583079, 583080, 583082, 583083, 583084, 583085, 583088, 583089, 583090, 583092, 583094, 583095, 583096, 583098, 583100, 583101, 583102, 583103, 583105, 583108, 583109, 583110, 583111, 583116, 583118, 583119, 583121, 583123, 583124, 583125, 583126, 583131, 583132, 583134, 583135, 583136, 583137, 583139, 583141, 583146, 583147, 583150, 583151, 583153, 583156, 583157, 583158, 583166.583167, 583169, 583170, 583172, 583173, 583175, 583177, 583179, 583181, 583183, 583185, 583186, 583189, 583192, 583194, 583195, 583196, 583197, 583198, 583201, 583204, 583205, 583206, 583208, 583209, 583211, 583213, 583214, 583216, 583217, 583218, 583219, 583221, 583222, 583223, 583225, 583226, 583227, 583229, 583232, 583235, 583236, 583237, 583239, 583240, 583241, 583244, 583245, 583246, 583249, 583250, 583251, 583255, 583256, 583258, 583260, 583261, 583262, 583263, 583264, 583267, 583268, 583271, 583274, 583276, 583279, 583280, 583283, 583284, 583285, 583286, 583287, 583289, 583294, 583296, 583298, 583299, 583300, 583302, 583306, 583307, 583308, 583309, 583311, 583313, 583314, 583317, 583318, 583320, 583322, 583323, 583324, 583326, 583328, 583329, 583330, 583331, 583333, 583334, 583335, 583336, 583337, 583338, 583341, 583342, 583344, 583345, 583349, 583350, 583351, 583353, 583355, 583357, 583360, 583362, 583364, 583365, 583366, 583370, 583371, 583373, 583374, 583375, 583379, 583380, 583385, 583386, 583387, 583389, 583390, 583392, 583393.   Model N119   - Serial numbers: 048289, 048292, 048294, 048295, 048296, 048297, 048298, 048299, 048300, 048302, 048303, 048304, 048306, 048307, 048308, 048309, 048310, 048311, 048312, 048313, 048314, 048315, 048318, 048319, 048323, 048325, 048326, 048330, 048332, 048336, 048337, 048338, 048341, 048344, 048408, 048410, 048412, 048413, 048414, 048415, 048416, 048418, 048419, 048420, 048422, 048423, 048425, 048426, 048428, 048430, 048434, 048436, 048440, 048442, 048444, 048447, 048534, 048535, 048537, 048538, 048539, 048540, 048541, 048542, 048543, 048545, 048546, 048547, 048548, 048550, 048551, 048552, 048553, 048554, 048555, 048558, 048560, 048561, 048563, 048564, 048565, 048567, 048570, 048571, 048673, 048676, 048679, 048680, 048681, 048686, 048689, 048691, 048693, 048695, 048698, 048699, 048700, 048701, 048704, 048705, 048707, 048708, 048709, 048710, 048711, 048712, 048713, 048715, 048716, 048720, 048724, 048725, 048726, 048793, 048794, 048812, 048815, 048817, 048818, 048819, 048820, 048821, 048822, 048823, 048824, 048826, 048828, 048831, 048832, 048833, 048836, 048838, 048839, 048840, 048841, 048842, 048848, 048875, 048876, 048877, 048879, 048880, 048881, 048882, 048883, 048884, 048885, 048886, 048887, 048888, 048889, 049042, 049043, 049044, 049046, 049049, 049054, 049056, 049059, 049061, 049063, 049064, 049066, 049067, 049070, 049072, 049073, 049074, 049075, 049079, 049080, 049081, 049082, 049084, 049086, 049087, 049088, 049089, 049090, 049091, 049094, 049097, 049098, 049099, 049100, 049101, 049105, 049106, 049107, 049109, 049110, 049111, 049280, 049281, 049282, 049283, 049285, 049286, 049287, 049294, 049295, 049297, 049299, 049320, 049321, 049322, 049323, 049324, 049326, 049328, 049329, 049330, 049333, 049376, 049380, 049384, 049387, 049412, 049414, 049415, 049417, 049420, 049423, 049424, 049427, 049429, 049431, 049432, 049435, 049437, 049438, 049461, 049462, 049468, 049469, 049470, 049472, 049473, 049474, 049476, 049477, 049480, 049481, 049482, 049484, 049485, 049486, 049487, 049491, 049492, 049494, 049495, 049496, 049497, 049501, 049502, 049503, 049504, 049505, 049507, 049508, 049509, 049511, 049512, 049515, 049517, 049519, 049520, 049521, 049522, 049524, 049525, 049527, 049528, 049783, 049784, 049786, 049787, 049788, 049790, 049793, 049794, 049796, 049798, 049802, 049803, 049805, 049806, 049808, 049810, 049811, 049873, 049874, 049875, 049877, 049879, 049881, 049882, 049883, 049884, 049886, 049907, 049908, 049911, 049912, 049913, 049915, 049916, 049918, 049919, 049920, 049966, 049969, 049970, 049972, 049973, 049976, 049978, 049980, 049981, 049982, 049983, 049984, 049985, 049986, 049987, 049988, 049990, 050009, 050010, 050011, 050013, 050014, 050016, 050018, 050019, 050020, 050021, 050022, 050023, 050026, 050028, 050030, 050032, 050033, 050074, 050075, 050080, 050083, 050084, 050085, 050088, 050092, 050093, 050095, 050096, 050098, 050099, 050103, 050164, 050166, 050167, 050168, 050249, 050253, 050256, 050257, 050262, 050263, 050267, 050270, 050271, 050272, 050275, 050277, 050279, 050282, 050283, 050286, 050287, 050291, 050292, 050296, 050298, 050299, 050301, 050304, 050305, 050307, 050308, 050329, 050331, 050332, 050333, 050337, 050339, 050341, 050343, 050344, 050345, 050347, 050348, 050349, 050350, 050353, 050357, 050358, 050399, 050400, 050401, 050402, 050404, 050407, 050409, 050412, 050415, 050416, 050418, 050423, 050426, 050427, 050457, 050458, 050459, 050460, 050465, 050466, 050469, 050470, 050471, 050472, 050474, 050479, 050480, 050482, 050484, 050501, 050502, 050504, 050505, 050506, 050507, 050509, 050514, 050515, 050516, 050517, 050518, 050519, 050520, 050521, 050522, 050523, 050524, 050526, 050527, 050528, 050529, 050530, 050531, 050532, 050533, 050536, 050538, 050540, 050542, 050543, 050546, 050547, 050548, 050549, 050551, 050552, 050556, 050557, 050762, 050765, 050767, 050768, 050769, 050772, 050773, 050775, 050776, 050778, 050781, 050783, 050785, 050787, 050788, 050790, 050796, 050799, 050801, 050809, 050813, 050816, 050818, 050820, 050821, 050822, 050823, 050824, 050827, 050828, 050833, 050835, 050836, 050838, 050839, 050840, 050841, 050842, 050843, 050844, 050846, 050848, 050849, 050851, 050852, 050853, 050855, 050856, 050865, 050866, 050867, 050891, 050892, 050893, 050894, 050899, 050901, 050902, 050903, 050904, 050905, 050907, 050908, 050909, 050911, 050914, 050939, 050941, 050942, 050944, 050945, 050950, 050952, 050953, 051099, 051101, 051102, 051103, 051107, 051124, 051125, 051127, 051129, 051132, 051133, 051134, 051137, 051138, 051139, 051141, 051143, 051144, 051145, 051146, 051147, 051150, 051153, 051155, 051156, 051157, 051158, 051160, 051162, 051163, 051165, 051166, 051170, 051171, 051172, 051175, 051176, 051177, 051180, 051181, 051183, 051185, 051186, 051188, 051189, 051191, 051194, 051197, 051198, 051423, 051427, 051428, 051431, 051435, 051438, 051442, 051444, 051445, 051446, 051448, 051451, 051452, 051454, 051455, 051457, 051459, 051461, 051462, 051464, 051466, 051467, 051468, 051469, 051473, 051476, 051477, 051478, 051479, 051480, 051482, 051484, 051487, 051488, 051490, 051493, 051494, 051496, 051497, 051499, 051501, 051502, 051503, 051505, 051506, 051510, 051511, 051513, 051515, 051519, 051521, 051522, 051523, 051525, 051526, 051527, 051528, 051529, 051531, 051533, 051535, 051536, 051538, 051539, 051542, 051544, 051755, 051756, 051761, 051762, 051763, 051764, 051768, 051769, 051770, 051771, 051772, 051773, 051775, 051777, 051781, 051782, 051783, 051785, 051787, 051790, 051791, 051794, 051795, 051797, 051798, 051800, 051801, 051802, 051806, 051813, 051822, 051823, 051825, 051826, 051827, 051829, 051831, 051832, 051954, 051955, 051956, 051957, 051958, 051959, 051961, 051964, 051966, 051967, 051968, 051969, 051973, 051974, 051975, 051976, 051977, 051979, 051981, 051982, 051983, 051984, 051985, 051987, 051988, 051989, 051990, 051992, 052157, 052158, 052161, 052162, 052163, 052164, 052165, 052166, 052167, 052169, 052170, 052171, 052173, 052176, 052177, 052179, 052181, 052184, 052185, 052188, 052192, 052195, 052198, 052199, 052200, 052201, 052209, 052212, 052214, 052215, 052216, 052221, 052223, 052224, 052225, 052227, 052233, 052242, 052243, 052244, 052245, 052246, 052248, 052249, 052253, 052254, 052256, 052258, 052261, 052262, 052263, 052264, 052266, 052289, 052290, 052292, 052293, 052295, 052297, 052298, 052299, 052300, 052302, 052304, 052347, 052348, 052349, 052350, 052351, 052353, 052355, 052356, 052358, 052359, 052360, 052361, 052362, 052363, 052527, 052529, 052531, 052532, 052533, 052536, 052537, 052538, 052540, 052541, 052542, 052543, 052545, 052546, 052767, 052771, 052777, 052786, 052787, 052789, 052790, 052791, 052792, 052793, 052796, 052797, 052801, 052802, 052803, 052804, 052805, 052949, 052951, 052952, 052955, 052956, 052959, 052962, 052965, 052966, 052968, 052972, 052973, 052974, 052978, 052980, 052982, 052985, 052986, 052987, 052988, 052989, 052990, 052992, 052995, 052997, 053001, 053004, 053005, 053007, 053008, 053009, 053010, 053011, 053012, 053013, 053014, 053015, 053018, 053020, 053022, 053024, 053328, 053329, 053331, 053332, 053335, 053336, 053338, 053344, 053345, 053346, 053348, 053349, 053350, 053351, 053353, 053354, 053355, 053357, 053361, 053362, 053363, 053364, 053365, 053411, 053412, 053413, 053414, 053415, 053416, 053417, 053422, 053426, 053427, 053429, 053430, 053431, 053433, 053434, 053435, 053436, 053437, 053439, 053442, 053444, 053445, 053446, 053447, 053449, 053450, 053452, 053453, 053455, 053457, 053459, 053461, 053462, 053465, 053469, 053499, 053500, 053502, 053503, 053506, 053507, 053508, 053509, 053510, 053512, 053513, 053514, 053516, 053517, 053540, 053543, 053544, 053545, 053546, 053547, 053585, 053587, 053590, 053591, 053592, 053595, 053597, 053598, 053603, 053604, 053605, 053609, 053612, 053614, 053616, 053617, 053619, 053621, 053622, 053624, 053625, 053626, 053627, 053630, 053843, 053844, 053847, 053848, 053849, 053850, 053852, 053855, 053856, 053857, 053858, 053860, 053861, 053862, 053863, 053866, 053867, 053869, 053870, 053872, 053873, 053874, 053875, 053877, 053879, 053881, 053882, 053883, 053885, 053887, 053888, 053889, 053890, 053892, 053897, 053898, 053899, 053902, 053903, 053904, 053906, 053907, 053908, 053910, 053914, 053915, 053917, 053918, 053922, 053923, 053925, 053926, 053928, 053929, 053930, 053932, 053934, 053936, 053937, 053939, 053940, 053941, 053943, 053944, 053947, 053948, 053949, 053950, 053952, 053956, 053960, 055591, 055593, 055595, 055597, 055598, 055599, 055601, 055602, 055603, 055604, 055605, 055607, 055608, 055609, 055610, 055611, 055613, 055614, 055615, 055618, 055619, 055622, 055625, 055626, 055631, 055635, 055640, 055641, 055642, 055643, 055644, 055645, 055647, 055648, 055649, 055650, 055651, 055655, 055657, 055659, 055661, 055664, 055667, 055668, 055745, 055747, 055751, 055753, 055759, 055761, 055762, 055766, 055769, 055770, 055772, 055773, 055774, 055776, 055777, 055778, 055779, 055780, 055783, 055786, 055788, 055789, 055790, 055791, 055792, 055793, 055815, 055816, 055819, 055821, 055822, 055823, 055824, 055825, 055826, 055831, 055832, 055834, 055835, 055836, 055837, 055838, 055839, 055841, 055842, 055844, 055846, 055853, 055969, 055970, 055971, 055972, 055974, 055975, 055976, 055977, 055978, 055981, 055982, 055984, 055986, 055987, 056388, 056389, 056391, 056393, 056394, 056395, 056397, 056398, 056399, 056402, 056408, 056410, 056411, 056412, 056413, 056416, 056457, 056460, 056462, 056463, 056466, 056469, 056471, 056473, 056476, 056477, 056478, 056479, 056480, 056482, 056483, 056484, 056485, 056486, 056487, 056489, 056491, 056493, 056494, 056495, 056497, 056498, 056500, 056501, 056504, 056505, 056509, 056510, 056511, 056513, 056560, 056562, 056563, 056564, 056565, 056566, 056568, 056570, 056572, 056575, 056576, 056598, 056602, 056603, 056606, 056607, 056609, 056611, 056613, 056614, 056616, 056638, 056640, 056807, 056808, 056809, 056810, 056812, 056815, 056816, 056817, 056818, 056820, 056821, 056823, 056824, 056827, 056829, 056830, 056831, 056832, 056835, 056837, 056840, 056842, 056844, 056852, 056854, 056855, 056856, 056858, 056859, 056862, 056864, 056865, 056866, 056868, 056870, 056871, 056872, 056875, 056876, 056877, 056878, 056879, 057098, 057101, 057102, 057103, 057106, 057107, 057108, 057110, 057112, 057113, 057115, 057118, 057120, 057121, 057122, 057124, 057125, 057127, 057129, 057133, 057137, 057139, 057161, 057162, 057163, 057164, 057165, 057166, 057167, 057169, 057171, 057173, 057174, 057176, 057177, 057179, 057180, 057181, 057182, 057183, 057184, 057185, 057188, 057207, 057208, 057209, 057211, 057215, 057216, 057241, 057243, 057244, 057245, 057248, 057250, 057251, 057254, 057258, 057259, 057260, 057304, 057312, 057314, 057315, 177091, 178618, 178621, 178628, 178630, 178631, 178635, 178637, 178638, 178639, 178642, 178645, 178646, 178647, 178650, 178677, 178678, 178681, 178685, 178687, 178689, 178690, 178692, 178694, 178695, 178896, 178897, 178899, 178900, 178906, 178907, 178908, 178909, 178911, 178912, 178913, 178914, 178915, 178917, 178920, 178921, 178922, 178923, 178924, 178926, 178927, 178931, 178932, 178933, 178976, 178980, 178981, 178985, 178987, 178988, 178989, 178991, 178993, 178994, 179016, 179017, 179018, 179020, 179021, 179022, 179023, 179024, 179025, 179186, 179188, 179189, 179192, 179195, 179197, 179198, 179199, 179200, 179202, 179203, 179204, 179225, 179228, 179229, 179233, 179259, 179260, 179264, 179267, 179272, 179273, 179296, 179298, 179299, 179300, 179301, 179303, 179304, 179305, 179307, 179309, 179311, 179312, 179313, 179549, 179552, 179555, 179557, 179559, 179560, 179561, 179562, 179564, 179566, 179571, 179579, 179580, 179581, 179582, 179587, 179609, 179610, 179611, 179614, 179618, 179619, 179620, 179621, 179625, 179648, 179649, 179650, 179654, 179656, 179661, 179665, 179667, 179768, 179769, 179771, 179772, 179775, 179776, 179777, 179781, 179782, 179790, 179794, 179795, 179802, 179804, 179806, 179809, 179812, 179817, 179820, 179822, 179824, 179825, 179826, 179829, 179832, 179833, 179839, 179841, 179842, 179847, 179850, 179851, 179852, 179853, 179855, 180196, 180197, 180198, 180200, 180221, 180222, 180224, 180227, 180229, 180233, 180235, 180236, 180237, 180239, 180241, 180245, 180247, 180249, 180251, 180254, 180255, 180257, 180259, 180308, 180311, 180314, 180318, 180319, 180320, 180322, 180363, 180366, 180368, 180370, 180372, 180376, 180377, 180379, 180384, 180385, 180387, 180388, 180391, 180395, 180396, 180399, 180401, 180443, 180446, 180448, 180449, 180450, 180452, 180455, 180456, 180457, 180498, 180499, 180500, 180501, 180502, 180503, 180504, 180505, 180506, 180548, 180549, 180550, 180554, 180556, 180559, 180560, 180561, 180562, 180783, 180784, 180786, 180787, 180790, 180791, 180792, 180793, 180799, 180800, 180802, 180803, 180804, 180805, 180806, 180807, 180811, 180853, 180858, 180860, 180861, 180862, 180863, 180864, 180867, 180869, 180870, 180871, 180872, 180974, 180975, 180976, 180977, 180980, 180981, 180984, 180986, 180988, 180989, 180990, 180991, 180994, 180995, 180998, 181003, 181004, 181006, 181007, 181008, 181010, 181012, 181013, 181016, 181018, 181020, 181022, 181025, 181026, 181028, 181030, 181032, 181214, 181215, 181216, 181217, 181220, 181222, 181225, 181226, 181230, 181231, 181334, 181336, 181337, 181338, 181340, 181341, 181342, 181344, 181345, 181347, 181348, 181349, 181350, 181351, 181352, 181353, 181357, 181358, 181362, 181464, 181465, 181470, 181471, 181472, 181473, 181475, 181476, 181477, 181478, 181479, 181481, 181483, 181484, 181490, 181492, 181494, 181495, 181496, 181497, 181498, 181500, 181501, 181503, 181505, 181506, 181507, 181508, 181510, 181513, 181519, 181803, 181805, 181806, 181807, 181808, 181809, 181813, 181816, 181918, 181921, 181923, 181924, 181926, 181929, 181933, 181934, 181935, 181936, 181937, 181938, 181939, 181940, 181941, 181943, 181945, 181947, 181948, 181949, 181950, 181951, 181952, 181953, 181954, 181957, 181960, 181961, 181962, 181964, 181965, 181966, 181968, 181971, 181974, 181977, 181981, 181984, 181986, 181987, 181989, 181997, 182003, 182005, 182010, 182011, 182012, 182014, 182015, 182016, 182017, 182020, 182021, 182022, 182024, 182028, 182029, 182032, 182034, 182035, 182039, 182040, 182042, 182046, 182047, 182048, 182049, 182050, 182051, 182053, 182061, 182064, 182065, 182067, 182248, 182394, 182395, 182396, 182397, 182398, 182400, 182403, 182404, 182405, 182407, 182409, 182410, 182413, 182420, 182449, 182451, 182452, 182453, 182454, 182455, 182456, 182458, 182460, 182461, 182462, 182466, 182468, 182470, 182494, 182495, 182496, 182497, 182499, 182500, 182504, 182505, 182506, 182508, 182510, 182512, 182514, 182539, 182540, 182541, 182542, 182546, 182547, 182549, 182551, 182648, 182649, 182650, 182651, 182653, 182654, 182655, 182656, 182657, 182659, 182660, 182664, 182669, 182671, 182705, 182706, 182707, 182708, 182709, 182710, 182711, 182714, 182716, 182720, 182726, 182727, 182729, 182730, 182732, 182775, 182778, 182779, 182780, 182782, 182787, 182789, 182790, 182795, 182796, 182797, 182800, 182801, 182803, 182804, 182806, 182808, 182812, 182813, 182814, 182815, 182820, 182821, 182823, 182864, 182868, 182869, 182870, 182871, 182873, 182874, 182875, 182877, 182878, 182879, 182882, 182883, 182884, 182885, 182890, 182892, 182894, 182896, 182898, 182910, 182911, 182914, 182916, 182918, 182920, 182921, 182992, 183064, 183065, 183066, 183069, 183070, 183269, 183270, 183271, 183272, 183273, 183276, 183277, 183278, 183279, 183281, 183282, 183288, 183289, 183292, 183294, 183295, 183296, 183297, 183300, 183301, 183302, 183303, 183304, 183305, 183306, 183307, 183308, 183373, 183374, 183375, 183376, 183377, 183383, 183386, 183388, 183392, 183393, 183394, 183397, 183399, 183402, 183403, 183404, 183407, 183410, 183411, 183413, 183415, 183416, 183420, 183422, 183423, 183424, 183426, 183429, 183431, 183432, 183433, 183434, 183435, 183437, 183439, 183440, 183443, 183445, 183446, 183448, 183510, 183511, 183512, 183513, 183515, 183517, 183518, 183539, 183542, 183543, 183544, 183545, 183547, 183548, 183554, 183581, 183582, 183584, 183585, 184009, 184013, 184014, 184015, 184017, 184018, 184022, 184023, 184024, 184025, 184027, 184028, 184029, 184030, 184031, 184033, 184035, 184036, 184038, 184040, 184041, 184043, 184046, 184089, 184090, 184091, 184092, 184094, 184095, 184096, 184097, 184098, 184120, 184122, 184125, 184127, 184128, 184130, 184135, 184136, 184137, 184823, 184824, 184825, 184830, 184833, 184837, 184838, 184839, 184841, 184842, 184843, 184848, 184856, 184858, 184861, 184862, 184864, 184866, 184867, 184868, 184869, 184870, 184873, 184874, 184875, 184876, 184880, 184881, 184882, 184883, 184884, 184885, 184886, 184887, 184889, 184890, 184891, 184893, 184894, 184895, 184897, 184898, 184899, 184901, 184902, 184965, 184966, 184967, 184968, 184970, 184973, 184974, 184976, 184978, 184979, 184980, 184981, 184982, 184983, 184984, 184987, 184988, 184989, 184991, 184993, 185195, 185196, 185197, 185198, 185199, 185202, 185204, 185207, 185251, 185255, 185256, 185257, 185259, 185260, 185261, 185262, 185264, 185266, 185268, 185269, 185270, 185274, 185276, 185277, 185280, 185282, 185283, 185284, 185287, 185288, 185290, 185291, 185292, 185294, 185298, 185302, 185303, 185304, 185426, 185427, 185429, 185432, 185436, 185437, 185438, 185439, 185440, 185441, 185446, 185447, 185448, 185449, 185454, 185543, 185549, 185550, 185551, 185552, 185553, 185554, 185556, 185559, 185560, 185561, 185562, 185563, 185566, 185567, 185570, 185571, 185572, 185573, 185574, 185575, 185579, 185581, 185584, 185585, 185588, 185591, 185593, 185594, 185603, 185604, 185605, 185608, 185609, 185610, 185611, 186085, 186087, 186095, 186096, 186103, 186104, 186105, 186108, 186109, 186110, 186111, 186135, 186138, 186139, 186140, 186141, 186143, 186144, 186145, 186146, 186148, 186149, 186150, 186152, 186153, 186155, 186156, 186159, 186165, 186167, 186169, 186171, 186172, 186173, 186216, 186220, 186221, 186225, 186226, 186228, 186250, 186251, 186253, 186254, 186255, 186256, 186259, 186263, 186305, 186306, 186308, 186311, 186312, 186314, 186536, 186538, 186541, 186543, 186544, 186605, 186610, 186611, 186613, 186615, 186616, 186617, 186618, 186619, 186621, 186622, 186624, 186625, 186627, 186628, 186629, 186630, 186633, 186635, 186636, 186637, 186638, 186639, 186640, 186641, 186645, 186652, 186653, 186654, 186655, 186658, 186660, 186662, 186667, 186668, 186670, 186672, 186675, 186676, 186681, 186682, 186683, 186684, 186685, 186686, 186688, 186690, 186691, 186692, 186697, 186699, 186700, 186704, 186705, 186706, 186707, 186708, 186720, 186721, 186722, 186724, 186726, 186728, 186734, 186735, 186737, 186741, 186742, 186745, 186746, 186747, 186748, 186749, 186754, 186756, 186757, 186760, 186761, 186762, 186763, 186765, 186766, 186767, 186768, 186770, 186771, 186773, 186774, 186997, 187000, 187001, 187003, 187004, 187009, 187010, 187011, 187012, 187013, 187016, 187017, 187018, 187020, 187024, 187026, 187027, 187028, 187029, 187033, 187034, 187036, 187037, 187038, 187039, 187044, 187045, 187048, 187049, 187050, 187052, 187057, 187058, 187059, 187118, 187120, 187121, 187124, 187126, 187128, 187133, 187134, 187136, 187137, 187140, 187142, 187145, 187175, 187176, 187180, 187182, 187184, 187187, 187189, 187526, 187527, 187528, 187531, 187532, 187539, 187540, 187582, 187583, 187584, 187585, 187586, 187587, 187588, 187591, 187592, 187593, 187594, 187596, 187599, 187600, 187603, 187604, 187605, 187606, 187607, 187608, 187610, 187611, 187612, 187613, 187614, 187619, 187621, 187622, 187623, 187726, 187728, 187730, 187731, 187733, 187735, 187742, 187743, 187747, 187748, 187749, 187750, 187757, 187758, 187762, 187764, 187765, 187889, 187891, 187896, 187899, 187900, 187901, 187902, 187903, 187906, 187908, 187909, 187912, 187915, 187918, 187919, 187921, 187923, 187930, 187932, 187934, 187938, 187939, 187940, 187941, 187944, 187947, 187948, 187949, 187951, 187953, 187954, 187959, 187964, 187965, 187967, 187968, 187969, 187970, 187971, 187973, 187974, 187978, 187981, 187984, 187986, 187988, 188009, 188013, 188015, 188016, 188017, 188201, 188204, 188205, 188206, 188207, 188208, 188212, 188214, 188215, 188217, 188219, 188220, 188221, 188222, 188224, 188227, 188228, 188229, 188230, 188232, 188235, 188236, 188239, 188241, 188303, 188304, 188305, 188306, 188308, 188311, 188313, 188314, 188315, 188316, 188582, 188584, 188585, 188586, 188587, 188588, 188589, 188590, 188591, 188594, 188595, 188597, 188598, 188599, 188601, 188602, 188603, 188609, 188610, 188611, 188612, 188613, 188619, 188620, 188622, 188624, 188627, 188629, 188634, 188635, 188701, 188702, 188703, 188704, 188705, 188706, 188712, 188713, 188714, 188717, 188720, 188725, 188726, 188727, 188729, 188730, 188732, 188733, 188735, 188736, 188738, 188739, 188740, 188743, 188747, 188748, 188750, 188751, 188752, 188753, 188754, 188755, 188756, 188757, 188758, 188761, 188763, 188764, 188765, 188766, 188767, 189109, 189110, 189111, 189112, 189114, 189115, 189116, 189117, 189119, 189120, 189121, 189122, 189123, 189126, 189127, 189173, 189177, 189178, 189182, 189183, 189185, 189186, 189189, 189190, 189191, 189194, 189195, 189196, 189197, 189200, 189201, 189205, 189206, 189210, 189212, 189213, 189214, 189215, 189216, 189218, 189220, 189222, 189223, 189224, 189225, 189226, 189227, 189228, 189231, 189234, 189236, 189411, 189412, 189413, 189414, 189415, 189416, 189417, 189418, 189422, 189425, 189448, 189449, 189450, 189451, 189455, 189456, 189457, 189459, 189462, 189463, 189465, 189466, 189467, 189468, 189469, 189470, 189472, 189473, 189475, 189476, 189478, 189479, 189482, 189483, 189484, 189485, 189487, 189488, 189489, 189492, 189494, 189496, 189498, 189500, 189501, 189503, 189504, 189506, 189507, 189540, 189542, 189544, 189546, 189549, 189584, 189585, 189587, 189591, 189592, 189593, 189595, 189596, 189597, 189808, 189812, 189813, 189814, 189815, 189816, 189817, 189819, 189822, 189823, 189824, 189825, 189826, 189828, 189830, 189831, 189834, 189837, 189839, 189840, 189841, 189842, 189845, 189846, 189847, 189848, 189849, 189850, 189851, 189853, 189854, 189856, 189857, 189861, 189865, 189866, 189867, 189869, 189870, 189872, 189873, 189874, 189875, 189876, 189878, 189879, 189881, 189882, 189883, 189884, 189887, 189888, 189891, 189893, 189895, 189896, 190108, 190109, 190112, 190114, 190116, 190117, 190119, 190120, 190121, 190123, 190127, 190128, 190129, 190132, 190134, 190136, 190138, 190139, 190143, 190145, 190149, 190151, 190152, 190153, 190154, 190156, 190157, 190159, 190161, 190163, 190165, 190167, 190168, 190170, 190172, 190174, 190177, 190178, 190182, 190214, 190215, 190216, 190217, 190218, 190219, 190221, 190225, 190226, 190228, 190229, 190230, 190231, 190232, 190358, 190359, 190360, 190361, 190363, 190365, 190367, 190370, 190375, 190378, 190380, 190381, 190382, 190386, 190387, 190388, 190390, 190393, 190394, 190395, 190400, 190404, 190405, 190406, 190444, 190445, 190446, 190450, 190452, 190453, 190454, 190455, 190456, 190458, 190459, 190460, 190483, 190484, 190485, 190486, 190549, 190550, 190552, 190553, 190554, 190557, 190558, 190560, 190561, 190564, 190565, 190566, 190568, 190570, 190572, 190575, 190577, 190581, 190582, 190583, 190585, 190588, 190589, 190590, 190591, 190592, 190623, 190624, 190625, 190627, 190628, 190630, 190723, 190726, 190728, 190729, 190730, 190731, 190732, 190734, 190736, 190854, 190861, 190862, 190868, 190872, 190873, 190874, 190877, 190878, 190879, 190880, 190881, 190884, 190888, 190889, 190895, 190896, 190900, 190905, 190907, 190908, 190909, 190912, 190913, 190914, 190915, 190916, 190917, 190918, 190920, 190921, 190923, 190924, 190925, 190927, 190929, 190930, 190933, 190934, 190935, 190936, 191086, 191087, 191088, 191089, 191090, 191092, 191095, 191099, 191100, 191101, 191102, 191103, 191104, 191105, 191107, 191115, 191118, 191121, 191122, 191123, 191125, 191126, 191127, 191128, 191132, 191134, 191136, 191140, 191141, 191158, 191160, 191162, 191163, 191164, 191167, 191168, 191169, 191171, 191172, 191173, 191175, 191176, 191179, 191180, 191181, 191184, 191185, 191186, 191206, 191208, 191209, 191210, 191212, 191213, 191214, 191216, 191218, 191219, 191221, 191223, 191224, 191228, 191229, 191232, 191233, 191234, 191235, 191357, 191358, 191360, 191363, 191366, 191367, 191370, 191371, 191397, 191400, 191402, 191403, 191405, 191406, 191407, 191408, 191409, 191410, 191707, 191711, 191713, 191714, 191716, 191721, 191722, 191723, 191725, 191727, 191728, 191730, 191732, 191733, 191734, 191736, 191737, 191739, 191740, 191742, 191743, 191744, 191747, 191749, 191750, 191752, 191754, 191821, 191823, 191827, 191828, 191829, 191830, 191831, 191833, 191837, 191841, 191843, 191844, 191845, 191846, 191848, 191849, 191850, 191851, 191852, 191853, 191857, 191858, 191859, 191860, 191897, 191898, 191899, 191901, 191902, 191903, 191904, 191907, 191908, 191910, 191912, 191915, 191917, 191919, 191920, 191921, 191925, 191926, 191927, 191929, 191930, 191931, 191932, 191933, 191934, 191937, 191942, 191943, 191944, 191945, 191947, 191949, 191950, 191958, 191959, 191960, 191961, 192084, 192085, 192086, 192089, 192090, 192091, 192092, 192093, 192094, 192095, 192096, 192097, 192098, 192099, 192100, 192104, 192107, 192108, 192109, 192110, 192111, 192138, 192139, 192140, 192141, 192142, 192144, 192145, 192149, 192151, 192153, 192155, 192156, 192157, 192158, 192161, 192162, 192164, 192203, 192204, 192205, 192207, 192209, 192211, 192212, 192213, 192214, 192216, 192217, 192222, 192223, 192224, 192225, 192227, 192230, 192231, 192312, 192313, 192315, 192317, 192319, 192321, 192322, 192324, 192326, 192327, 192332, 192334, 192336, 192338, 192340, 192341, 192343, 192344, 192346, 192348, 192349, 192351, 192352, 192354, 192355, 192356, 192360, 192361, 192366, 192368, 192370, 192372, 192374, 192375, 192376, 192617, 192618, 192619, 192620, 192624, 192626, 192627, 192628, 192629, 192633, 192637, 192640, 192643, 192644, 192645, 192646, 192708, 192710, 192711, 192714, 192715, 192716, 192719, 192720, 192721, 192724, 192726, 192727, 192729, 192730, 192732, 192734, 192735, 192736, 192738, 192739, 192740, 192741, 192743, 192744, 192746, 192747, 192748, 192753, 192754, 192756, 192758, 192759, 192761, 192762, 192764, 192765, 192767, 192809, 192810, 192811, 192813, 192814, 192815, 192816, 192818, 192819, 192820, 192821, 192824, 192825, 192828, 192833, 192835, 192838, 192839, 192841, 192844, 192846, 192848, 192850, 192851, 192858, 192860, 192862, 192864, 192865, 192867, 193168, 193169, 193170, 193171, 193172, 193174, 193175, 193177, 193179, 193181, 193184, 193187, 193188, 193190, 193191, 193192, 193194, 193195, 193259, 193260, 193261, 193262, 193263, 193265, 193269, 193273, 193277, 193598, 193599, 193600, 193601, 193602, 193604, 193606, 193607, 193612, 193613, 193615, 193617, 193619, 193620, 193621, 193622, 193623, 193626, 193628, 193629, 193630, 193631, 193632, 193633, 193634, 193636, 193637, 193639, 193642, 193643, 193644, 193647, 193648, 193649, 193651, 193652, 193694, 193695, 193698, 193701, 193702, 193703, 193704, 193706, 193707, 193708, 193709, 193710, 193714, 193716, 193717, 193718, 193721, 193723, 193724, 193727, 193728, 193730, 193731, 193732, 193734, 193735, 193736, 193818, 193819, 193820, 193821, 193822, 193823, 193826, 193827, 193828, 193831, 193832, 193834, 193835, 193839, 193840, 193843, 193844, 193845, 193846, 193847, 193848, 193849, 193852, 193854, 193856, 193858, 193859, 193861, 193864, 193865, 193866, 193867, 193869, 193870, 193871, 194054, 194055, 194060, 194061, 194062, 194063, 194064, 194065, 194066, 194067, 194068, 194069, 194070, 194071, 194072, 194073, 194074, 194077, 194079, 194081, 194087, 194089, 194090, 194091, 194094, 194095, 194098, 194099, 194100, 194101, 194102, 194106, 194107, 194108, 194112, 194115, 194116, 194117, 194118, 194119, 194122, 194123, 194126, 194129, 194130, 194132, 194133, 194134, 194136, 194139, 194140, 194141, 194194, 194195, 194196, 194199, 194202, 194203, 194242, 194245, 194246, 194247, 194248, 194249, 194250, 194252, 194254, 194255, 194258, 194259, 194260, 194262, 194263, 194264, 194265, 194266, 194267, 194268, 194429, 194430, 194431, 194432, 194433, 194434, 194435, 194437, 194438, 194821, 194823, 194824, 194825, 194826, 194828, 194829, 194830, 194832, 194833, 194836, 194837, 194839, 194840, 194842, 194848, 194851, 194854, 194857, 194858, 194860, 194861, 194862, 194865, 194866, 194867, 194868, 194872, 194873, 194876, 194878, 194940, 194941, 194942, 194944, 194946, 194947, 194949, 194950, 194951, 194954, 194958, 194960, 194962, 194965, 194967, 194970, 194972, 194973, 194974, 194975, 194976, 194977, 194978, 194979, 194980, 194981, 194983, 194986, 194988, 194989, 194993, 194994, 194995, 194996, 194998, 195060, 195062, 195063, 195064, 195065, 195067, 195068, 195071, 195073, 195074, 195075, 195076, 195078, 195079, 195081, 195082, 195083, 195085, 195086, 195087, 195129, 195130, 195132, 195133, 195137, 195138, 195140, 195142, 195152, 195153, 195155, 195158, 195159, 195160, 195161, 195163, 195165, 195166, 195167, 195168, 195169, 195171, 195174, 195175, 195177, 195178, 195179, 195180, 195182, 195184, 195185, 195354, 195355, 195357, 195358, 195359, 195360, 195361, 195362, 195363, 195364, 195367, 195429, 195431, 195433, 195434, 195435, 195440, 195441, 195442, 195443, 195445, 195447, 195448, 195630, 195631, 195633, 195635, 195637, 195639, 195640, 195642, 195643, 195644, 195645, 195646, 195647, 195649, 195650, 195652, 195653, 195655, 195658, 195659, 195660, 195662, 195665, 195666, 195667, 195668, 196106, 196108, 196111, 196112, 196113, 196114, 196118, 196119, 196121, 196123, 196125, 196127, 196128, 196129, 196131, 196132, 196133, 196134, 196135, 196152, 196155, 196156, 196157, 196159, 196162, 196164, 196165, 196166, 196168, 196171, 196173, 196177, 196178, 196179, 196180, 196181, 196182, 196183, 196184, 196185, 196298, 196300, 196301, 196302, 196308, 196309, 196310, 196311, 196313, 196315, 196316, 196318, 196322, 196324, 196325, 196326, 196327, 196328, 196330, 196332, 196333, 196334, 196336, 196338, 196340, 196342, 196343, 196344, 196350, 196353, 196359, 196360, 196361, 196362, 196364, 196369, 196381, 196385, 196389, 196391, 196392, 196394, 196395, 196396, 196397, 196399, 196401, 196407, 196408, 196409, 196411, 196414, 196415, 196417, 196421, 196422, 196423, 196526, 196527, 196532, 196533, 196534, 196535, 196538, 196543, 196544, 196545, 196768, 196774, 196778, 196779, 196827, 196828, 196832, 196833, 196835, 196838, 196840, 196918, 196920, 196922, 196923, 196925, 196928, 196930, 196937, 196938, 196939, 196940, 196942, 196946, 196948, 196949, 196950, 196955, 196958, 196959, 196960, 196961, 196962, 196963, 196964, 196966, 196967, 196970, 196971, 196974, 196975, 196976, 196980, 196981, 196982, 196983, 196984, 196986, 196987, 196990, 196991, 196992, 196993, 196994, 196995, 196996, 197107, 197108, 197110, 197111, 197114, 197115, 197119, 197120, 197133, 197135, 197237, 197239, 197242, 197247, 197249, 197250, 197254, 197256, 197337, 197340, 197342, 197343, 197345, 197346, 197348, 197350, 197351, 197512, 197514, 197515, 197516, 197518, 197521, 197523, 197524, 197525, 197527, 197528, 197529, 197530, 197532, 197534, 197535, 197536, 197537, 197539, 197540, 197541, 197543, 197544, 197546, 197548, 197549, 197551, 197553, 197554, 197555, 197557, 197559, 197560, 197561, 197568, 197572, 197573, 197580, 197581, 197582, 197583, 197585, 197587, 197588, 197589, 197590, 197591, 197592, 197593, 197594, 197595, 197597, 197599, 197600, 197602, 197604, 197606, 197611, 197612, 197613, 197614, 197615, 197616, 198018, 198019, 198020, 198021, 198022, 198025, 198028, 198029, 198030, 198032, 198037, 198039, 198041, 198044, 198045, 198046, 198048, 198049, 198050, 198051, 198052, 198054, 198057, 198059, 198061, 198062, 198064, 198066, 198068, 198069, 198070, 198071, 198072, 198073, 198075, 198082, 198083, 198084, 198091, 198172, 198173, 198174, 198176, 198177, 198178, 198182, 198183, 198184, 198188, 198189, 198191, 198192, 198194, 198195, 198196, 198560, 198561, 198562, 198564, 198566, 198567, 198568, 198571, 198572, 198573, 198576, 198577, 198579, 198580, 198581, 198585, 198586, 198587, 198591, 198606, 198608, 198609, 198610, 198611, 198613, 198614, 198615, 198618, 198619, 198620, 198623, 198624, 198625, 198626, 198627, 198629, 198631, 198633, 198636, 198641, 198642, 198643, 198644, 198645, 198648, 198649, 198650, 198652, 198653, 198655, 198656, 198660, 198661, 198663, 198665, 198846, 198848, 198850, 198852, 198853, 198854, 198855, 198856, 198857, 198859, 198861, 198864, 198867, 198868, 198869, 198871, 198872, 198873, 198874, 198875, 198877, 198878, 198879, 198880, 198882, 198883, 198884, 198885, 198889, 198890, 198891, 198892, 198894, 198895, 198899, 198901, 198903, 198904, 199086, 199087, 199088, 199089, 199090, 199091, 199196, 199197, 199200, 199201, 199203, 199205, 199206, 199208, 199209, 199212, 199215, 199216, 199219, 199220, 199221, 199223, 199224, 199228, 199229, 199231, 199232, 199235, 199240, 199243, 199244, 199246, 199248, 199249, 199251, 199255, 199256, 199258, 199259, 199260, 199261, 199262, 199263, 199264, 199265, 199326, 199328, 199329, 199331, 199332, 199333, 199334, 199338, 199340, 199345, 199449, 199454, 199455, 199456, 199457, 199460, 199482, 199483, 199485, 199489, 199491, 199492, 199494, 199500, 199504, 199505, 199506, 199507, 199508, 199509, 199510, 199511, 199513, 199514, 199516, 199517, 199518, 199519, 199521, 199522, 199523, 199551, 199554, 199555, 199556, 199557, 199559, 199720, 199721, 199722, 199723, 199728, 199730, 199731, 199736, 199737, 199742, 199743, 199744, 199745, 200128, 200129, 200130, 200131, 200132, 200135, 200136, 200137, 200138, 200141, 200142, 200143, 200145, 200147, 200148, 200150, 200153, 200154, 200155, 200156, 200159, 200160, 200161, 200162, 200165, 200166, 200167, 200168, 200169, 200174, 200284, 200285, 200287, 200289, 200291, 200292, 200294, 200298, 200299, 200302, 200303, 200304, 200307, 200309, 200315, 200316, 200319, 200322, 200324, 200326, 200329, 200330, 200334, 200436, 200440, 200441, 200442, 200443, 200444, 200445, 200446, 200450, 200452, 200453, 200456, 200457, 200459, 200460, 200462, 200463, 200464, 200465, 200526, 200527, 200528, 200530, 200531, 200532, 200536, 200537, 200538, 200539, 200540, 200617, 200622, 200624, 200628, 200629, 200630, 200631, 200632, 200637, 200643, 200645, 200646, 200647, 200650, 200655, 200717, 200719, 200721, 200725, 200781, 200784, 200785, 200786, 200787, 200790, 200791, 200792, 200795, 200796, 200797, 200798, 200800, 200801, 200805, 200806, 200808, 200809, 200812, 200813, 200815, 200817, 200820, 200821, 200823, 200824, 200825, 200826, 200827, 200828, 200829, 200830, 200831, 200834, 200835, 200836, 200837, 200839, 200845, 200846, 200847, 200849, 200850, 200966, 200967, 200969, 200972, 200974, 200977, 200980, 201048, 201049, 201051, 201053, 201054, 201055, 201057, 201058, 201060, 201061, 201063, 201066, 201067, 201071, 201074, 201076, 201077, 201078, 201079, 201081, 201082, 201083, 201084, 201092, 201098, 201099, 201100, 201301, 201302, 201304, 201305, 201306, 201308, 201309, 201310, 201311, 201312, 201314, 201316, 201319, 201324, 201327, 201330, 201334, 201335, 201336, 201338, 201342, 201347, 201349, 201411, 201412, 201413, 201414, 201419, 201420, 201421, 201423, 201425, 201429, 201430, 201432, 201433, 201434, 201435, 201436, 201438, 201439, 201441, 201442, 201443, 201444, 201448, 201450, 201772, 201775, 201779, 201784, 201785, 201787, 201789, 201790, 201791, 201794, 201796, 201798, 201799, 201800, 201863, 201864, 201865, 201866, 201870, 201871, 201872, 201873, 201874, 201876, 201877, 201878, 201879, 201880, 201881, 201883, 201885, 201887, 201889, 201890, 201891, 201896, 201897, 201899, 201902, 201903, 201905, 201906, 201907, 201909, 201951, 201952, 201957, 201958, 201959, 202262, 202269, 202275, 202276, 202277, 202278, 202280, 202282, 202286, 202287, 202288, 202291, 202294, 202297, 202300, 202301, 202302, 202303, 202304, 202307, 202310, 202351, 202354, 202355, 202356, 202359, 202360, 202361, 202368, 202369, 202370, 202371, 202374, 202375, 202379, 202581, 202582, 202583, 202584, 202585, 202586, 202588, 202590, 202596, 202600, 202603, 202604, 202605, 202607, 202608, 202611, 202613, 202615, 202616, 202618, 202621, 202622, 202623, 202624, 202626, 202627, 202628, 202629, 202631, 202633, 202634, 202636, 202639, 202642, 202644, 202645, 202648, 203226, 208006, 968903, 968908, 968910, 968912, 968914, 968915, 968917, 968918, 968919, 968920, 968923, 968924, 968925, 968926, 968927, 968929, 968930, 968931, 968932, 968986, 968987, 968988, 968990, 968991, 968995, 968997, 968998, 968999, 969001, 969004, 969005, 969006, 969008, 969010, 969011, 969012, 969033, 969035, 969037, 969038, 969039, 969041, 969042, 969043, 969044, 969045, 969046, 969047, 969048, 969050, 969051, 969052, 969055, 969056, 969057, 969058, 969059, 969062, 969103, 969104, 969106, 969107, 969109, 969111, 969112, 969113, 969115, 969116, 969117, 969118, 969119, 969120, 969121, 969122, 969123, 969127, 969128, 969130, 969131, 969132, 969173, 969174, 969175, 969176, 969177, 969178, 969179, 969182, 970613, 970614, 970615, 970616, 970617, 970618, 970619, 970620, 970621, 970622, 970625, 970626, 970628, 970630, 970703, 970704, 970707, 970708, 970709, 970710, 970711, 970712, 970713, 970714, 970717, 970719, 970720, 970721, 970733, 970734, 970736, 970737, 970738, 970739, 970740, 970741, 970742, 970743, 970745, 970746, 970747, 970748, 970749, 970750, 970751, 970752, 970753, 970758, 970759, 970761, 970763, 970766, 970767, 970768, 970769, 970770, 970771, 970772, 970773, 970776, 970777, 970781, 970799, 970800, 970801, 970805, 970806, 970809, 970810, 970812, 970813, 970814, 970816, 970819, 970820, 970822, 970824, 970825, 970827, 970849, 970850, 970851, 970853, 970854, 970855, 970856, 970857, 970858, 970859, 970860, 970862, 970863, 970866, 970867, 970870, 970872, 970873, 970876, 970877, 970878, 970879, 970881, 970882, 970884, 970885, 970886, 970888, 970889, 970890, 970892, 970893, 970894, 970895, 970896, 970897, 970898, 970900, 970901, 970902, 970903, 970904, 970905, 970906, 970968, 970970, 970972, 970974, 970976, 970977, 970979, 971013, 971014, 971015, 971017, 971020, 971021, 971023, 971029, 971030, 971033, 971034, 971035, 971036, 971037, 971038, 971039, 971040, 971041, 971042, 971044, 971045, 971046, 971047, 971048, 971049, 971050, 971051, 971054, 971055, 971056, 971057, 971218, 971225, 971226, 971227, 971228, 971229, 971230, 971231, 971233, 971236, 971237, 971238, 971239, 971241, 971242, 971243, 971245, 971247, 971248, 971249, 971250, 971252, 971253, 971254, 971257, 971259, 971262, 971410, 971411, 971413, 971415, 971416, 971417, 971420, 971422, 971424, 971427, 971428, 971429, 971430, 971434, 971435, 971436, 971439, 971440, 971441, 971442, 971443, 971444, 971445, 971453, 971454, 971455, 971456, 971457, 971458, 971459, 971464, 971467, 971468, 971469, 971470, 971471, 971473, 971474, 971476, 971477, 971478, 971479, 971480, 971481, 971482, 971483, 971484, 971485, 971486, 971487, 971488, 971490, 971492, 971494, 971497, 971508, 971509, 971510, 971512, 971513, 971517, 972125, 972126, 972127, 972128, 972129, 972130, 972132, 972133, 972134, 972136, 972139, 972140, 972141, 972143, 972145, 972146, 972148, 972149, 972151, 972152, 972153, 972154, 972156, 972157, 972160, 972161, 972162, 972164, 972165, 972166, 972167, 972168, 972170, 972171, 972172, 972175, 972177, 972413, 972415, 972417, 972419, 972423, 972424, 972425, 972426, 972512, 972514, 972516, 972517, 972519, 972521, 972524, 972525, 972527, 972528, 972529, 972530, 972531, 972532, 972533, 972534, 972536, 972537, 972538, 972539, 972540, 972542, 972543, 972544, 972547, 972548, 972549, 972550, 972551, 972552, 972553, 972555, 972557, 972558, 972562, 972564, 972566, 972568, 972569, 972592, 972595, 972597, 972599, 972600, 972602, 972603, 972606, 972607, 972608, 972611, 972615, 972618, 972619, 972620, 972621, 972622, 972623, 972625, 972626, 972627, 972628, 972629, 972631, 972632, 972633, 972634, 972636, 972637, 972638, 972639, 972640, 972641, 972642, 972643, 972644, 972647, 972651, 972653, 972656, 972878, 972880, 972882, 972883, 972884, 972885, 972886, 972889, 972893, 972894, 972898, 972899, 972901, 972902, 972904, 972905, 972906, 972907, 972909, 972910, 972915, 972916, 972917, 972918, 972919, 972920, 972921, 972923, 972924, 972925, 972926, 972927, 972928, 972930, 972931, 972933, 972934, 972935, 972937, 972940, 972941, 972943, 972944, 972945, 972947, 972950, 972951, 972953, 972954, 972957, 972958, 972959, 972962, 972963, 972964, 972965, 972966, 973267, 973269, 973270, 973271, 973274, 973275, 973278, 973280, 973281, 973324, 973326, 973327, 973328, 973329, 973330, 973331, 973332, 973333, 973334, 973335, 973336, 973338, 973339, 973340, 973341, 973342, 973343, 973346, 973347, 973349, 973350, 973351, 973452, 973453, 973455, 973458, 973459, 973460, 973461, 973462, 973463, 973464, 973465, 973466, 973467, 973468, 973471, 973472, 973473, 973475, 973477, 973478, 973480, 973482, 973483, 973484, 973486, 973487, 973488, 973489, 973490, 973492, 973493, 973495, 973497, 973498, 973500, 973501, 973503, 973505, 973508, 973509, 973511, 973512, 973513, 973515, 973516, 973517, 973519, 973521, 973524, 973525, 973526, 973528, 973529, 973531, 973532, 973533, 973536, 973538, 973545, 973546, 973588, 973589, 973590, 973593, 973594, 973595, 973596, 973597, 973600, 973602, 973603, 973604, 973606, 973707, 973708, 973709, 973710, 973711, 973712, 973713, 973714, 973715, 973717, 973718, 973719, 973720, 973721, 973722, 973725, 973726, 973728, 973729, 973730, 973732, 973733, 973735, 973736, 973737, 973738, 973740, 973741, 973744, 973745, 973746, 973747, 973749, 973750, 973751, 973752, 973753, 973754, 973755, 973756, 973757, 973758, 973760, 973761, 973762, 973763, 973764, 973765, 973766, 974233, 974235, 974236, 974237, 974239, 974240, 974241, 974245, 974247, 974248, 974251, 974253, 974254, 974255, 974256, 974258, 974259, 974261, 974262, 974264, 974265, 974266, 974268, 974269, 974271, 974272, 974273, 974274, 974275, 974276, 974278, 974280, 974282, 974283, 974284, 974285, 974286, 974287, 974288, 974290, 974291, 974292, 974293, 974294, 974296, 974297, 974298, 974300, 974302, 974303, 974305, 974306, 974307, 974308, 974311, 974372, 974375, 974376, 974377, 974379, 974380, 974382, 974386, 974390, 974391, 974392, 974393, 974394, 974396, 974397, 974400, 974401, 974422, 974425, 974426, 974428, 974429, 974433, 974434, 974435, 974518, 974519, 974523, 974524, 974526, 974527, 974528, 974529, 974530, 974531, 974532, 974536, 974558, 974559, 974560, 974562, 974564, 974565, 974566, 974567, 974569, 974573, 974574, 974576, 974578, 974579, 974580, 974582, 974584, 974585, 974586, 974589, 974590, 974591, 974592, 974594, 974595, 974596, 974597, 974599, 974601, 974623, 974624, 974626, 974627, 974628, 974630, 974631, 974633, 974634, 974635, 974637, 974638, 974639, 974640, 974938, 974939, 975003, 975004, 975005, 975010, 975011, 975013, 975015, 975016, 975017, 975018, 975019, 975022, 975023, 975028, 975029, 975031, 975032, 975033, 975034, 975035, 975038, 975041, 975042, 975043, 975044, 975046, 975047, 975048, 975051, 975052, 975053, 975054, 975055, 975056, 975057, 975059, 975061, 975124, 975128, 975129, 975131, 975134, 975135, 975136, 975137, 975138, 975159, 975162, 975163, 975164, 975165, 975166, 975173, 975194, 975196, 975197, 975198, 975199, 975200, 975201, 975202, 975203, 975204, 975206, 975208, 975209, 975210, 975212, 975215, 975216, 975217, 975218, 975460, 975461, 975462, 975463, 975464, 975465, 975466, 975471, 975472, 975473, 975477, 975479, 975480, 975481, 975482, 975483, 975484, 975485, 975486, 975487, 975489, 975490, 975493, 975594, 975596, 975597, 975599, 975601, 975604, 975605, 975607, 975608, 975609, 975610, 975611, 975613, 975614, 975616, 975619, 975621, 975623, 975624, 975625, 975626, 975627, 975628, 975629, 975630, 975631, 975632, 975635, 975638, 975640, 975641, 975642, 975665, 975666, 975668, 975673, 975674, 975677, 975678, 975681, 975683, 975684, 975687, 975689, 975690, 975691, 975694, 975695, 975697, 975699, 975703, 975705, 975706, 975707, 975708, 975729, 975730, 975731, 975733, 975734, 975735, 975736, 975739, 975742, 975745, 975746, 975747, 975749, 975750, 975751, 975752, 975753, 975755, 975757, 975758, 975811, 975812, 975813, 975814, 975815, 975817, 975819, 975820, 975821, 975823, 975826, 975827, 975828, 975829, 975830, 975831, 975833, 975834, 975835, 975837, 975838, 975860, 975861, 975862, 975867, 975868, 975869, 975870, 975872, 975873, 975874, 975875, 975876, 975877, 975878, 975879, 975882, 975884, 975886, 975889, 975891, 975893, 975894, 975895, 975896, 975897, 975899, 975901, 975902, 975965, 975966, 975967, 975968, 975969, 975970, 975971, 975973, 975974, 975976, 975978, 976449, 976450, 976452, 976453, 976454, 976455, 976459, 976460, 976461, 976462, 976463, 976464, 976465, 976466, 976467, 976470, 976473, 976475, 976476, 976477, 976478, 976490, 976491, 976492, 976493, 976494, 976496, 976499, 976500, 976501, 976502, 976503, 976505, 976507, 976508, 976509, 976510, 976511, 976512, 976513, 976518, 976521, 976523, 976524, 976525, 976526, 976527, 976528, 976529, 976530, 976533, 976534, 976535, 976536, 976537, 976538, 976539, 976540, 976541, 976542, 976544, 976545, 976547, 976551, 976552, 976556, 976557, 976558, 976569, 976570, 976574, 976575, 976576, 976577, 976579, 976581, 976840, 976841, 976843, 976844, 976845, 976847, 976848, 976849, 976851, 976852, 976854, 976857, 976858, 976859, 976860, 976862, 976863, 976868, 976873, 976894, 976896, 976897, 976898, 976899, 976901, 976904, 976907, 976908, 976909, 976912, 976915, 976916, 976917, 976918, 976920, 976921, 976922, 976923, 976925, 976926, 976927, 976928, 976930, 976933, 977014, 977015, 977017, 977018, 977021, 977022, 977023, 977027, 977028, 977031, 977032, 977033, 977036, 977039, 977040, 977041, 977042, 977044, 977046, 977048, 977049, 977050, 977051, 977052, 977053, 977115, 977116, 977117, 977118, 977120, 977121, 977122, 977123, 977124, 977125, 977126, 977129, 977131, 977133, 977134, 977136, 977137, 977138, 977140, 977141, 977142, 977143, 977145, 977146, 977147, 977148, 977149, 977151, 977152, 977155, 977156, 977157, 977158, 977159, 977160, 977162, 977165, 977166, 977167, 977168, 977171, 977172, 977174, 977176, 977177, 977178, 977179, 977180, 977181, 977182, 977183, 977184, 977185, 977187, 977188, 977190, 977192, 977193, 977195, 977198, 977199, 977201, 977202, 977203, 977727, 977728, 977729, 977730, 977731, 977732, 977733, 977734, 977736, 977737, 977740, 977741, 977743, 977744, 977745, 977747, 977748, 977749, 977750, 977751, 977752, 977753, 977754, 977756, 977757, 977759, 977760, 977761, 977762, 977763, 977765, 977766, 977767, 977769, 977770, 977771, 977772, 977773, 977774, 977775, 977776, 977777, 977778, 977779, 977780, 977785, 977786, 977787, 977788, 977789, 977790, 977791, 977792, 977793, 977794, 977795, 977796, 977797, 977798, 977799, 977800, 977801, 977802, 977803, 977804, 977807, 977809, 977810, 977811, 977812, 977815, 977816, 977819, 977820, 977824, 977826, 977828, 977830, 977831, 977832, 977836, 977838, 977839, 977841, 977844, 977845, 977846, 977849, 977850, 977894, 977895, 977896, 977897, 977900, 977902, 977903, 977904, 977906, 977929, 977930, 977931, 977932, 977933, 977934, 977935, 977937, 977939, 977940, 977942, 977943, 977944, 977945, 977946, 977947, 977948, 977950, 977951, 977953, 977954, 977955, 977956, 977957, 977958, 978019, 978020, 978023, 978024, 978027, 978029, 978030, 978031, 978032, 978194, 978195, 978198, 978199, 978202, 978206, 978389, 978391, 978393, 978395, 978396, 978397, 978398, 978401, 978404, 978407, 978409, 978410, 978412, 978413, 978414, 978415, 978416, 978417, 978419, 978420, 978422, 978423, 978424, 978427, 978428, 978429, 978430, 978431, 978433, 978435, 978436, 978438, 978440, 978441, 978443, 978445, 978446, 978447, 978448, 978449, 978451, 978453, 978454, 978456, 978457, 978458, 978459, 978460, 978462, 978463, 978465, 978468, 978469, 978470, 978471, 978472, 978473, 978474, 978475, 978476, 978477, 978479, 978480, 978481, 978484, 978486, 978487, 978488, 978489, 978490, 978491, 978492, 978493, 978495, 978499, 978500, 978501, 978503, 978505, 978508, 978549, 978551, 978553, 978555, 978557, 978560, 978561, 978563, 978564, 978566, 978567, 978568, 978569, 978571, 978573, 978574, 978575, 978576, 978577, 978578, 978580, 978581, 978582, 978583, 978584, 978585, 978586, 978590, 978592, 978593, 978594, 978595, 978596, 978597, 978598, 978600, 978602, 978606, 978607, 978608, 978609, 978610, 978612, 978616, 978617, 978618, 978621, 978623, 978624, 978625, 978626, 978627, 978689, 978690, 978692, 978693, 978694, 978695, 978696, 978698, 978700, 978701, 978702, 978705, 978707, 978709, 978711, 978712, 978714, 978715, 978716, 978717, 978718, 978963, 978964, 978965, 978967, 978969, 978970, 978971, 978972, 978973, 978974, 978976, 978978, 979019, 979020, 979023, 979024, 979025, 979026, 979027, 979032, 979036, 979037, 979038, 979039, 979040, 979041, 979042, 979043, 979045, 979046, 979047, 979048, 979049, 979051, 979054, 979058, 979099, 979101, 979102, 979104, 979105, 979107, 979389, 979391, 979392, 979394, 979395, 979396, 979398, 979399, 979400, 979401, 979406, 979407, 979409, 979411, 979412, 979415, 979417, 979418, 979519, 979520, 979521, 979522, 979523, 979525, 979526, 979528, 979530, 979532, 979536, 979537, 979599, 979601, 979602, 979603, 979604, 979605, 979606, 979607, 979608, 979611, 979613, 979615, 979616, 979617, 979619, 979621, 979622, 979623, 979624, 979625, 979626, 979627, 979628, 979692, 979693, 979694, 979695, 979697, 979698, 979699, 979700, 979702, 979703, 979706, 979709, 979710, 979711, 979712, 979713, 979714, 979716, 979717, 979718, 979741, 979742, 979745, 979746, 979747, 979748, 979750, 979753, 979754, 979756, 979757, 979759, 979761, 979762, 979764, 979766, 979767, 979769, 979770, 979772, 979774, 979775, 979776, 979778, 979779, 979780, 979781, 979804, 979809, 979810, 979811, 979812, 979813, 979817, 979818, 979835, 979836, 979839, 979840, 979841, 979842, 979846, 979847, 979850, 979851, 979853, 979856, 979857, 979858, 979860, 979861, 979862, 979863, 979864, 979865, 979866, 979868, 979869, 979871, 979872, 979873, 979874, 979876, 979878, 979879, 979880, 979881, 979882, 979883, 980105, 980107, 980108, 980109, 980110, 980111, 980114, 980115, 980116, 980119, 980120, 980121, 980122, 980123, 980126, 980127, 980129, 980520, 980522, 980523, 980524, 980525, 980527, 980529, 980530, 980532, 980533, 980534, 980536, 980539, 980540, 980542, 980544, 980545, 980549, 980550, 980551, 980553, 980557, 980558, 980560, 980561, 980562, 980564, 980565, 980566, 980567, 980568, 980569, 980571, 980572, 980573, 980574, 980575, 980577, 980578, 980650, 980651, 980652, 980653, 980657, 980658, 980660, 980662, 980663, 980665, 980666, 980667, 980675, 980676, 980677, 980678, 980679, 980695, 980697, 980699, 980701, 980703, 980704, 980705, 980706, 980707, 980709, 980713, 980714, 980855, 980857, 980858, 980859, 980862, 980863, 980865, 980866, 980867, 980868, 980872, 980873, 980874, 980875, 980876, 980877, 980878, 980880, 980881, 980882, 980883, 980886, 980889, 980890, 980891, 980892, 980893, 980894, 980895, 980896, 980897, 980898, 980900, 980901, 980902, 981340, 981341, 981342, 981343, 981344, 981345, 981347, 981348, 981349, 981351, 981352, 981353, 981354, 981355, 981405, 981406, 981407, 981408, 981410, 981411, 981412, 981414, 981415, 981417, 981419, 981420, 981421, 981470, 981471, 981473, 981475, 981476, 981477, 981478, 981479, 981480, 981481, 981482, 981483, 981484, 981487, 981488, 981490, 981492, 981493, 981494, 981495, 981496, 981498, 981499, 981500, 981501, 981502, 981503, 981831, 981832, 981833, 981834, 981836, 981838, 981839, 981840, 981901, 981903, 981904, 981905, 981906, 981907, 981908, 981909, 981910, 981911, 981913, 981914, 981915, 981917, 981920, 981921, 981922, 981924, 981925, 981927, 981930, 981931, 981932, 981933, 981935, 981938, 981940, 981941, 981942, 981943, 981944, 981947, 981948, 981950, 981951, 981952, 981955, 981958, 981959, 981961, 981963, 981964, 981966, 981967, 981968, 981971, 981972, 981975, 981977, 981979, 981980, 981981, 981983, 981985, 981986, 981989, 981994, 981996, 981997, 981998, 982338, 982340, 982341, 982342, 982343, 982344, 982345, 982346, 982351, 982354, 982589, 982590, 982592, 982594, 982595, 982596, 982597, 982598, 982599, 982602, 982603, 982604, 982605, 982607, 982609, 982610, 982612, 982613, 982614, 982615, 982618, 982619, 982621, 982623, 982624, 982626, 982627, 982628, 982629, 982633, 982636, 982637, 982640, 982641, 982642, 982643, 982647, 982648, 982649, 982650, 982651, 982652, 982653, 982654, 982657, 982658, 982659, 982661, 982662, 982663, 982665, 982666, 982667, 982668, 982670, 982671, 982673, 982678, 982681, 982683, 982685, 982686, 982687, 982688, 982692, 982694, 982695, 982696, 982697, 982698, 982699, 982701, 982702, 982703, 982704, 982705, 982706, 982707, 982708, 982709, 982710, 982711, 982712, 982713, 982714, 982715, 982716, 982767, 982769, 982770, 982772, 982773, 982774, 982776, 982779, 982780, 982782, 982785, 982786, 982787, 982790, 982791, 982792, 982793, 982794, 982796, 982798, 982799, 982800, 982801, 982802, 982803, 982806, 982809, 982810, 982812, 982813, 982814, 982816, 982858, 982861, 982864, 982865, 982866, 982867, 982868, 982869, 982871, 982872, 982874, 982877, 982880, 982884, 982885, 982886, 982977, 982978, 982979, 982982, 982983, 982985, 982986, 982992, 982994, 982995, 982996, 983435, 983436, 983437, 983438, 983439, 983440, 983442, 983443, 983444, 983445, 983447, 983448, 983450, 983451, 983452, 983455, 983456, 983457, 983463, 983464, 983465, 983467, 983468, 983470, 983471, 983472, 983473, 983475, 983477, 983478, 983479, 983480, 983482, 983483, 983486, 983487, 983489, 983492, 983494, 983495, 983496, 983499, 983500, 983501, 983503, 983504, 983505, 983506, 983507, 983508, 983510, 983511, 983512, 983513, 983517, 983518, 983520, 983521, 983522, 983524, 983595, 983596, 983600, 983601, 983603, 983604, 983605, 983606, 983607, 983608, 983609, 983661, 983662, 983663, 983664, 983665, 983666, 983667, 983668, 983670, 983671, 983672, 983673, 983674, 983676, 983678, 983679, 983680, 983681, 983682, 983684, 983686, 983687, 983689, 983712, 983713, 983716, 983718, 983720, 983721, 983723, 983724, 983725, 983726, 983728, 983730, 983731, 983732, 983733, 983734, 983735, 983736, 983737, 983738, 983739, 983740, 983741, 983742, 983744, 983805, 983808, 983810, 983812, 983814, 983816, 983817, 983818, 983819, 983820, 983821, 983822, 983823, 983826, 983828, 983829, 983831, 983833, 983834, 983835, 983836, 983837, 983838, 983839, 983840, 983841, 983842, 983843, 983844, 983946, 983948, 983949, 983950, 983951, 983953, 983954, 983955, 983956, 983957, 983958, 983959, 983960, 983961, 983962, 983965, 983966, 983967, 983968, 983971, 983972, 983973, 983974, 983977, 983980, 983981, 983982, 983983, 983984, 983985, 983986, 983987, 983988, 983989, 983991, 983992, 983993, 983994, 983995, 983997, 983998, 984003, 984004, 984005, 984006, 984008, 984009, 984012, 984013, 984014, 984016, 984018, 984020, 984021, 984022, 984023, 984026, 984027, 984028, 984030, 984031, 984033, 984035, 984036, 984037, 984039, 984040, 984041, 984067, 984068, 984069, 984074, 984075, 984077, 984078, 984079, 984082, 984083, 984366, 984367, 984370, 984371, 984372, 984373, 984374, 984376, 984379, 984380, 984381, 984382, 984383, 984385, 984387, 984391, 984392, 984394, 984396, 984397, 984398, 984400, 984401, 984403, 984404, 984405, 984406, 984407, 984408, 984410, 984412, 984413, 984415, 984416, 984417, 984418, 984419, 984420, 984423, 984424, 984608, 984609, 984610, 984611, 984612, 984613, 984615, 984616, 984617, 984618, 984620, 984621.   Model P106   - Serial numbers: 001589, 001590, 001591, 001592, 001594, 001596, 001597, 001598, 001599, 001600, 001601, 001602, 001603, 001605, 001606, 001607, 001608, 001609, 001610, 001611, 001612, 001613, 001615, 001616, 001617, 001618, 001619, 001621, 001622, 001623, 001624, 001625, 001626, 001627, 001628, 001629, 001632, 001633, 001635, 001636, 001638, 001639, 001641, 001642, 001643, 001644, 001645, 001646, 001647, 001648, 001650, 001652, 001653, 001654, 001655, 001656, 001657, 001658, 001659, 001660, 001661, 001662, 001663, 001664, 001665, 001666, 001667, 001668, 001669, 001670, 001671, 001672, 001673, 001674, 001675, 001676, 001678, 001679, 001681, 001682, 001684, 001685, 001686, 001687, 001689, 001690, 001691, 001692, 001693, 001694, 001695, 001696, 001697, 001698, 001699, 001700, 001701, 001702, 001703, 001704, 001705, 001706, 001707, 001708, 001709, 001710, 001711, 001712, 001713, 001714, 001715, 001716, 001717, 001718, 001719, 001721, 001722, 001723, 001724, 001726, 001727, 001729, 001730, 001733, 001735, 001736, 001739, 001740, 001741, 001742, 001743, 001745, 001746, 001747, 001748, 001749, 001750, 001751, 001752, 001753, 001754, 001755, 001756, 001757, 001758, 001759, 001760, 001761, 001762, 001763, 001765, 001766, 001767, 001768, 001769, 001770, 001771, 001772, 001773, 001774, 001775, 001777, 001779, 001781, 001782, 001783, 001785, 001786, 001788, 001790, 001791, 001792, 001795, 001796, 001797, 001798, 001799, 001800, 001801, 001802, 001803, 001804, 001805, 001806, 001807, 001808, 001809, 001810, 001812, 001813, 001814, 001815, 001816, 001818, 001819, 001823, 001934, 001935, 001937, 001938, 001940, 001941, 001942, 001943, 001944, 001945, 001946, 001947, 001949, 001950, 001951, 001952, 001955, 001956, 001957, 001958, 001959, 001961, 001962, 001963, 001964, 001965, 001966, 001967, 001968, 001969, 001970, 001972, 001973, 001974, 001975, 001976, 001977, 001978, 001980, 001981, 001982, 001985, 001986, 001987, 001988, 001989, 001990, 001991, 001992, 001994, 001996, 001997, 001998, 001999, 002000, 002001, 002003, 002004, 002005, 002006, 002007, 002009, 002010, 002011, 002012, 002013, 002014, 002015, 002016, 002017, 002019, 002020, 002022, 002024, 002026, 002027, 002028, 002029, 002031, 002032, 002033, 002034, 002035, 002036, 002038, 002040, 002041, 002042, 002043, 002044, 002045, 002046, 002048, 002049, 002050, 002051, 002052, 002053, 002054, 002055, 002060, 002061, 002063, 002064, 002065, 002066, 002067, 002068, 002069, 002071, 002073, 002074, 002075, 002077, 002078, 002080, 002081, 002085, 002087, 002088, 002089, 002090, 002091, 002092, 002094, 002095, 002096, 002098, 002101, 002102, 002103, 002104, 002105, 002106, 002107, 002108, 002109, 002110, 002111, 002112, 002113, 002114, 002115, 002116, 002118, 002119, 002120, 002121, 002124, 002125, 002126, 002127, 002128, 002129, 002130, 002131, 002132, 002133, 002134, 002136, 002137, 002140, 002141, 002142, 002143, 002144, 002145, 002146, 002148, 002151, 002152, 002154, 002158, 002159, 002161, 002162, 002163, 002164, 002165, 002166, 002171, 002172, 002174, 002176, 002178, 002179, 002180, 002181, 002182, 002183, 002184, 002186, 002187, 002188, 002189, 002190, 002191, 002192, 002193, 002194, 002196, 002198, 002199, 002200, 002201, 002202, 002203, 002564, 002565, 002566, 002569, 002570, 002571, 002572, 002573, 002575, 002577, 002580, 002582, 002583, 002584, 002585, 002586, 002587, 002588, 002589, 002590, 002591, 002592, 002593, 002594, 002595, 002596, 002597, 002598, 002599, 002601, 002602, 002603, 002604, 002605, 002606, 002607, 002608, 002609, 002610, 002611, 002612, 002613, 002614, 002615, 002617, 002618, 002619, 002620, 002621, 002622, 002623, 002624, 002625, 002626, 002627, 002628, 002629, 002630, 002631, 002632, 002633, 002634, 002635, 002636, 002637, 002639, 002640, 002641, 002642, 002643, 002644, 002645, 002646, 002647, 002648, 002649, 002650, 002651, 002652, 002653, 002654, 002655, 002656, 002657, 002658, 002659, 002660, 002661, 002662, 002664, 002665, 002666, 002667, 002668, 002669, 002670, 002671, 002672, 002673, 002674, 002675, 002677, 002678, 002679, 002680, 002681, 002684, 002685, 002686, 002687, 002688, 002689, 002690, 002691, 002692, 002693, 002694, 002695, 002696, 002697, 002698, 002699, 002701, 002702, 002703, 002704, 002705, 002706, 002708, 002709, 002710, 002711, 002712, 002713, 002714, 002716, 002718, 002719, 002720, 002721, 002723, 002724, 002725, 002727, 002728, 002729, 002730, 002732, 002733, 002734, 002735, 002737, 002739, 002741, 002743, 002744, 002746, 002747, 002748, 002749, 002751, 002753, 002755, 002756, 002758, 002759, 002760, 002761, 002762, 002763, 002764, 002765, 002766, 002767, 002769, 002770, 002772, 002774, 002775, 002776, 002777, 002779, 002780, 002781, 002783, 002784, 002785, 002786, 002787, 002788.   Model P107   - Serial numbers: 008276, 009310, 009311, 009312, 009313, 009315, 009317, 009320, 009323, 009324, 009325, 009327, 009328, 009329, 009330, 009331, 009332, 009333, 009334, 009335, 009337, 009338, 009339, 009340, 009342, 009343, 009344, 009345, 009347, 009349, 009350, 009351, 009352, 009353, 009354, 009355, 009357, 009358, 009359, 009360, 009361, 009362, 009363, 009364, 009365, 009366, 009367, 009368, 009369, 009490, 009491, 009492, 009493, 009494, 009495, 009496, 009497, 009498, 009499, 009500, 009501, 009502, 009503, 009504, 009505, 009506, 009507, 009508, 009509, 009510, 009512, 009514, 009516, 009517, 009518, 009519, 009540, 009541, 009542, 009543, 009545, 009546, 009547, 009548, 009549, 009550, 009551, 009553, 009554, 009555, 009556, 009557, 009558, 009559, 009560, 009561, 009562, 009564, 009565, 009566, 009567, 009568, 009569, 009570, 009571, 009572, 009573, 009574, 009576, 009577, 009578, 009579, 009580, 009581, 009582, 009583, 009584, 009585, 009586, 009587, 009588, 009589, 009590, 009591, 009592, 009593, 009594, 009595, 009596, 009597, 009598, 009599, 009600, 009601, 009602, 009603, 009604, 009605, 009606, 009607, 009608, 009609, 009781, 009782, 009785, 009786, 009787, 009788, 009790, 009791, 009792, 009794, 009795, 009796, 009797, 009798, 009799, 009800, 009801, 009803, 009804, 009805, 009806, 009807, 009809, 009810, 009811, 009812, 009813, 009814, 009815, 009817, 009818, 009819, 009821, 009823, 009824, 009825, 009826, 009827, 009828, 009829, 009830, 009891, 009892, 009893, 009894, 009895, 009896, 009897, 009898, 009899, 009900, 009901, 009902, 009903, 009904, 009905, 009906, 009907, 009908, 009909, 009910, 009911, 009914, 009915, 009916, 009917, 009918, 009919, 009920, 009921, 009922, 009923, 009924, 009925, 009926, 009927, 009928, 009929, 009930, 009932, 009933, 009934, 009935, 009936, 009937, 009939, 009940, 009941, 009942, 009943, 009944, 009945, 009946, 009947, 009948, 009949, 009950, 009991, 009992, 009993, 009994, 009995, 009996, 009997, 009998, 009999, 010000, 010001, 010002, 010003, 010005, 010006, 010007, 010008, 010009, 010010, 010011, 010012, 010013, 010014, 010015, 010056, 010057, 010058, 010059, 010060, 010061, 010062, 010064, 010065, 010066, 010067, 010068, 010069, 010070, 010071, 010074, 010075, 010077, 010078, 010079, 010080, 010081, 010082, 010083, 010084, 010085, 010086, 010087, 010089, 010090, 010091, 010092, 010093, 010094, 010096, 010097, 010098, 010099, 010100, 010101, 010102, 010103, 010104, 010105, 010106, 010108, 010109, 010110, 010111, 010112, 010113, 010114, 010115, 010116, 010117, 010118, 010119, 010120, 010122, 010123, 010124, 010125, 010126, 010127, 010128, 010129, 010131, 010132, 010133, 010134, 010135, 010136, 010137, 010138, 010139, 010140, 010141, 010142, 010143, 010144, 010145, 010146, 010147, 010148, 010149, 010150, 010151, 010152, 010153, 010154, 010155, 010156, 010157, 010158, 010159, 010160, 010161, 010162, 010163, 010164, 010165, 010166, 010167, 010168, 010169, 010170, 010171, 010172, 010173, 010174, 010175, 010176, 010177, 010178, 010179, 010180, 010181, 010182, 010183, 010184, 010185, 010186, 010187, 010188, 010189, 010190, 010191, 010192, 010193, 010194, 010195, 010272, 010273, 010274, 010275, 010276, 010278, 010279, 010281, 010282, 010283, 010284, 010285, 010286, 010287, 010288, 010289, 010290, 010291, 010292, 010293, 010294, 010295, 010296, 010297, 010299, 010300, 010301, 010302, 010303, 010305, 010306, 010307, 010308, 010309, 010310, 010311, 010312, 010313, 010314, 010315, 010316, 010317, 010318, 010319, 010320, 010321, 010322, 010323, 010324, 010325, 010326, 010327, 010328, 010329, 010330, 010331, 010389, 010390, 010391, 010392, 010394, 010395, 010396, 010397, 010398, 010399, 010400, 010401, 010402, 010403, 010405, 010406, 010407, 010408, 010409, 010410, 010411, 010412, 010413, 010414, 010415, 010417, 010418, 010663, 010664, 010665, 010666, 010667, 010668, 010669, 010671, 010672, 010696, 010697, 010698, 010699, 010700, 010703, 010704, 010705, 010706, 010707, 010708, 010710, 010711, 010712, 010713, 010714, 010715, 010735, 010736, 010737, 010739, 010740, 010741, 010742, 010764, 010765, 010766, 010767, 010768, 010769, 010770, 010771, 010772, 010773, 010774, 010776, 010777, 010779, 010780, 010781, 010782, 010783, 010810, 010811, 010812, 010813, 010814, 010815, 010816, 010817, 010818, 010834, 010835, 010836, 010837, 010838, 010839, 010840, 010841, 010898, 010899, 010900, 010901, 010903, 010904, 010905, 010906, 010907, 010908, 010909, 010910, 010911, 010912, 010914, 010915, 010916, 010917, 010918, 010919, 010920, 010921, 010922, 010923, 010924, 010925, 010926, 010927, 011011, 011012, 011013, 011014, 011015, 011016, 011017, 011018, 011020, 011021, 011022, 011023, 011024, 011026, 011027, 011028, 011050, 011051, 011052, 011053, 011054, 011055, 011056, 011057, 011058, 011059, 011060, 011061, 011062, 011063, 011064, 011065, 011067, 011068, 011069, 011070, 011071, 011073, 011074, 011075, 011076, 011077, 011078, 011079, 011080, 011081, 011082, 011083, 011084, 011086, 011087, 011088, 011089, 011090, 011091, 011093, 011094, 011095, 011096, 011099, 011100, 011101, 011102, 011103, 011106, 011107, 011108, 011109, 011110, 011111, 011112, 011113, 011114, 011115, 011116, 011117, 011118, 011119, 011120, 011122, 011123, 011124, 011125, 011126, 011127, 011128, 011129, 011131, 011132, 011133, 011134, 011137, 011138, 011139, 011140, 011141, 011142, 011143, 011145, 011146, 011147, 011148, 011150, 011151, 011152, 011154, 011158, 011159, 011160, 011161, 011162, 011163, 011164, 011167, 011168, 011169, 011170, 011173, 011174, 011175, 011176, 011177, 011180, 011182, 011183, 011184, 011185, 011186, 011187, 011188, 011189, 011190, 011193, 011194, 011195, 011196, 011197, 011239, 011240, 011241, 011242, 011243, 011245, 011246, 011248, 011249, 011250, 011251, 011252, 011253, 011254, 011255, 011256, 011257, 011258, 011259, 011260, 011261, 011262, 011263, 011264, 011266, 011267, 011268, 011269, 011270, 011271, 011272, 011273, 011498, 011499, 011500, 011501, 011502, 011503, 011504, 011506, 011507, 011508, 011509, 011510, 011511, 011512, 011513, 011514, 011515, 011517, 011518, 011521, 011522, 011523, 011524, 011525, 011526, 011527, 011528, 011530, 011531, 011532, 014351, 014352, 014353, 014354, 014355, 014356, 014357, 014358, 014359, 014360, 014361, 014362, 014363, 014364, 014365, 014366, 014367, 014368, 014369, 014370, 014372, 014373, 014374, 014375, 014377, 014378, 014380, 014381, 014382, 014383, 014384, 014385, 014386, 014387, 014388, 014389, 014390, 014391, 014392, 014393, 014394, 014395, 014396, 014398, 014399, 014400, 014441, 014442, 014443, 014445, 014446, 014447, 014448, 014449, 014450, 014451, 014452, 014453, 014454, 014456, 014457, 014458, 014459, 014460, 014461, 014462, 014463, 014464, 014465, 014546, 014547, 014548, 014549, 014550, 014551, 014552, 014553, 014554, 014555, 014556, 014557, 014558, 014560, 014561, 014562, 014563, 014564, 014565, 014566, 014567, 014568, 014569, 014570, 014571, 014572, 014573, 014574, 014575, 014576, 014577, 014578, 014579, 014580, 014581, 014583, 014584, 014585, 014586, 014587, 014588, 014589, 014590, 014631, 014632, 014633, 014634, 014635, 014636, 014637, 014638, 014639, 014641, 014643, 014644, 014645, 014646, 014647, 014648, 014649, 014651, 014652, 014653, 014654, 014655, 014657, 014659, 014660, 014661, 014662, 014664, 014665, 014666, 014667, 014668, 014669, 014670, 014671, 014672, 014673, 014674, 014675, 014676, 014678, 014679, 014680, 014684, 014685, 014707, 014708, 014709, 014710, 014711, 014712, 014713, 014715, 014716, 014717, 014718, 014719, 014720, 014721, 014722, 014723, 014724, 014725, 014729, 014730, 014731, 014733, 014739, 014740, 014741, 014742, 014746, 014747, 014749, 014751, 014752, 014753, 014754, 014755, 014756, 014757, 014758, 014759, 014760, 014761, 014762, 014764, 014765, 014965, 014966, 014967, 014968, 014969, 014971, 014972, 014973, 014974, 014977, 014978, 014979, 014980, 014981, 014982, 014983, 014984, 014985, 014986, 014987, 014988, 014989, 014991, 014992, 014993, 014994, 014995, 014996, 014997, 014998, 014999, 015000, 015001, 015002, 015003, 015004, 015006, 015007, 015008, 015009, 015010, 015011, 015012, 015013, 015014, 015035, 015036, 015037, 015039, 015041, 015042, 015043, 015045, 015047, 015048, 015049, 015050, 015051, 015052, 015053, 015054, 015055, 015056, 015057, 015058, 015059, 015060, 015061, 015062, 015063, 015064, 015065, 015066, 015069, 015070, 015071, 015072, 015074, 015075, 015076, 015077, 015078, 015079, 015080, 015081, 015082, 015083, 015084, 015085, 015086, 015087, 015088, 015089, 015090, 015091, 015092, 015094, 015095, 015096, 015097, 015098, 015099, 015100, 015101, 015102, 015103, 015104, 015145, 015146, 015147, 015148, 015149, 015150, 015151, 015152, 015153, 015155, 015156, 015157, 015158, 015160, 015161, 015162, 015163, 015164, 015165, 015166, 015167, 015168, 015169, 015170, 015171, 015172, 015174, 015176, 015177, 015178, 015179, 015180, 015181, 015182, 015183, 015184, 015185, 015186, 015187, 015188, 015270, 015272, 015273, 015274, 015275, 015276, 015277, 015279, 015280, 015281, 015282, 015283, 015284, 015305, 015307, 015309, 015310, 015311, 015313, 015315, 015316, 015319, 015320, 015322, 015323, 015325, 015326, 015327, 015328, 015329, 015330, 015331, 015332, 015334, 015335, 015337, 015338, 015840, 015841, 015842, 015844, 015845, 015846, 015847, 015848, 015849, 015850, 015851, 015852, 015853, 015854, 015855, 015856, 015857, 015858, 015860, 015861, 015862, 015863, 015864, 015865, 015866, 015867, 015868, 015869, 015870, 015871, 015872, 015873, 015874, 015875, 015876, 015877, 015879, 015880, 015881, 015882, 015883, 015886, 015887, 015888, 015889, 015890, 015891, 015892, 015894, 015895, 015896, 015897, 015898, 015899, 015900, 015901, 015903, 015904, 015905, 015906, 015907, 015908, 015909, 015910, 015911, 015912, 015913, 015915, 015916, 015918, 015919, 015921, 015922, 015923, 015924, 015926, 015927, 015928, 015929, 015930, 015931, 015932, 015933, 015935, 015936, 015937, 015938, 015939, 015940, 015942, 015944, 015945, 015948, 015950, 015951, 015952, 015953, 015954, 015956, 015957, 015958, 015959, 015960, 015961, 015962, 015963, 015964, 015965, 015968, 015969, 015970, 015972, 015973, 015974, 015976, 015977, 015979, 016001, 016004, 016005, 016007, 016008, 016009, 016011, 016012, 016014, 016015, 016016, 016017, 016018, 016019, 016020, 016021, 016022, 016023, 016026, 016028, 016029, 016031, 016032, 016033, 016034, 016035, 016037, 016038, 016039, 016040, 016041, 016042, 016043, 016044, 016150, 016151, 016152, 016153, 016154, 016155, 016156, 016157, 016158, 016159, 016260, 016261, 016262, 016263, 016264, 016265, 016266, 016267, 016268, 016269, 016270, 016271, 016272, 016273, 016274, 016275, 016276, 016277, 016278, 016279, 016280, 016281, 016282, 016284, 016426, 016427, 016428, 016429, 016430, 016431, 016432, 016433, 016434, 016435, 016436, 016437, 016438, 016439, 016440, 016441, 016442, 016443, 016444, 016445, 016447, 016448, 016449, 016450, 016451, 016453, 016454, 016455, 016456, 016457, 016458, 016459, 016565, 016566, 016567, 016568, 016570, 016571, 016572, 016573, 016574, 016575, 016576, 016577, 016578, 016579, 016580, 016581, 016582, 016583, 016584, 016585, 016586, 016587, 016588, 016589, 016590, 016591, 016593, 016594, 016655, 016656, 016657, 016659, 016660, 016661, 016662, 016663, 016664, 016666, 016667, 016668, 016669, 016670, 016671, 016672, 016673, 016674, 016676, 016677, 016679, 016681, 016682, 016745, 016746, 016747, 016748, 016749, 016750, 016751, 016752, 016753, 016754, 016757, 016758, 016759, 016760, 016761, 016762, 016763, 016764, 016765, 016766, 016767, 016769, 016770, 016771, 016772, 016773, 016774, 016855, 016856, 016857, 016858, 016859, 016860, 016861, 016862, 016863, 016864, 016865, 016866, 016867, 016868, 016869, 016870, 016872, 016873, 016874, 016875, 016876, 016877, 016879, 016880, 016881, 016882, 016883, 016885, 016887, 016888, 016889, 016891, 016892, 016893, 016894, 016895, 016896, 016897, 016898, 016899, 016900, 016901, 016902, 016903, 016904, 016945, 016946, 016947, 016948, 016949, 016950, 016951, 016952, 016953, 016954, 016955, 016956, 016957, 016958, 016959, 016960, 016961, 016962, 016963, 016964, 016965, 016966, 016967, 016968, 016969, 016970, 016971, 016972, 016973, 016974, 016975, 016976, 016977, 016978, 016979, 016980, 016981, 016982, 016983, 016984, 016985, 016986, 016987, 016988, 016989, 016993, 016994, 016995, 016997, 016998, 016999, 017000, 017001, 017002, 017005, 017006, 017007, 017008, 017009, 017010, 017011, 017012, 017013, 017014, 017055, 017056, 017057, 017058, 017059, 017060, 017061, 017062, 017063, 017064, 017067, 017068, 017069, 017150, 017151, 017152, 017153, 017154, 017155, 017156, 017157, 017158, 017159, 017200, 017201, 017202, 017203, 017205, 017208, 017209, 017210, 017212, 017213, 017214, 017216, 017219, 017220, 017221, 017222, 017223, 017224, 017225, 017226, 017227, 017228, 017229, 017230, 017232, 017234, 017235, 017236, 017237, 017238, 017239, 017240, 017241, 017243, 017244, 017245, 017246, 017247, 017248, 017249, 017250, 017251, 017253, 017254, 017255, 017256, 017257, 017258, 017259, 017336, 017337, 017338, 017339, 017341, 017342, 017343, 017344, 017346, 017349, 017350, 017351, 017352, 017354, 017355, 017356, 017357, 017358, 017359, 017360, 017361, 017363, 017364, 017365, 017366, 017367, 017368, 017369, 017370, 017372, 017373, 017374, 017375, 017376, 017377, 017379, 017380, 017381, 017382, 017383, 017384, 017385, 017426, 017427, 017428, 017429, 017430, 017431, 017432, 017433, 017434, 017435, 017476, 017477, 017478, 017481, 017482, 017483, 017484, 017485, 017486, 017487, 017488, 017489, 017490, 017491, 017492, 017493, 017494, 017495, 017698, 017699, 017700, 017701, 017702, 017703, 017704, 017705, 017706, 017707, 017708, 017709, 017710, 017711, 017712, 017713, 017714, 017715, 017717, 017718, 017719, 017720, 017721, 017722, 017725, 017726, 017727, 017728, 017729, 017730, 017731, 017733, 017734, 017735, 017738, 017739, 017740, 017741, 017742, 017743, 017744, 017745, 017746, 017747, 017748, 017749, 017750, 017751, 017752, 017753, 017754, 017755, 017836, 017837, 017838, 017839, 017840, 017841, 017842, 017843, 017844, 017845, 017846, 017847, 017848, 017849, 017850, 017851, 017852, 017854, 017855, 017917, 017918, 017919, 017920, 017921, 017922, 017924, 017925, 017926, 017927, 017928, 017929, 017930, 017931, 017932, 017933, 017934, 017935, 017936, 017937, 017938, 017939, 017940, 017942, 017943, 017944, 017945, 017948, 017949, 017950, 017951, 017953, 017954, 017955, 017956, 017957, 017958, 017960, 017961, 017962, 017963, 017964, 017965, 018127, 018128, 018129, 018130, 018131, 018132, 018133, 018135, 018136, 018137, 018138, 018139, 018140, 018141, 018142, 018144, 018146, 018147, 018148, 018149, 018150, 018151, 018152, 018153, 018154, 018156, 018217, 018218, 018220, 018221, 018222, 018223, 018224, 018225, 018228, 018229, 018230, 018231, 018233, 018234, 018235, 018236, 018277, 018278, 018279, 018280, 018282, 018283, 018284, 018285, 018286, 018287, 018288, 018289, 018290, 018291, 018292, 018293, 018294, 018295, 018296, 018298, 018299, 018300, 018301, 018302, 018303, 018304, 018305, 018306, 018347, 018348, 018349, 018350, 018351, 018352, 018353, 018354, 018355, 018356, 018358, 018361, 018363, 018364, 018365, 018366, 018567, 018568, 018569, 018570, 018571, 018572, 018573, 018574, 018575, 018576, 018577, 018578, 018579, 018580, 018581, 018582, 018583, 018584, 018585, 018586, 018588, 018589, 018590, 018591, 018592, 018593, 018594, 018596, 018658, 018659, 018660, 018661, 018662, 018663, 018664, 018665, 018666, 018668, 018669, 018670, 018671, 018672, 018673, 018674, 018675, 018676, 018778, 018779, 018781, 018785, 018786, 018788, 018789, 018790, 018791, 018832, 018833, 018834, 018835, 018836, 018837, 018838, 018839, 018840, 018841, 018842, 018845, 018847, 018848, 018849, 018850, 018852, 018854, 018856, 018858, 018859, 018860, 018864, 018865, 018866, 018967, 018968, 018969, 018971, 018972, 018973, 018975, 018976, 018979, 018980, 018981, 018982, 018983, 018984, 018986, 018988, 018989, 018990, 018992, 018993, 018994, 018995, 018996, 027174, 027175, 027176, 027177, 027178, 027179, 027180, 027181, 027182, 027183, 027184, 027185, 027186, 027187, 027189, 027190, 027191, 027192, 027193, 027214, 027215, 027216, 027217, 027218, 027219, 027220, 027221, 027222, 027223, 027224, 027225, 027227, 027228, 027229, 027230, 027231, 027233, 027234, 027235, 027236, 027237, 027238, 027239, 027240, 027241, 027242, 027243, 027244, 027245, 027246, 027247, 027248, 027249, 027250, 027251, 027252, 027253, 027314, 027315, 027316, 027317, 027318, 027319, 027320, 027321, 027322, 027323, 027324, 027325, 027326, 027327, 027328, 027329, 027330, 027331, 027332, 027333, 027334, 027335, 027336, 027337, 027338, 027339, 027340, 027341, 027342, 027343, 027344, 027345, 027346, 027347, 027348, 027349, 027350, 027351, 027352, 027353, 027354, 027355, 027356, 027357, 027358, 027359, 027361, 027362, 027384, 027385, 027386, 027387, 027388, 027389, 027390, 027391, 027392, 027393, 027394, 027395, 027396, 027397, 027398, 027399, 027401, 027402, 027403, 027404, 027405, 027406, 027407, 027409, 027410, 027411, 027412, 027413, 027415, 027416, 027417, 027418, 027419, 027420, 027422, 027423, 027424, 027425, 027426, 027427, 027428, 027430, 027431, 027432, 027434, 027436, 027437, 027439, 027440, 027441, 027442, 027443, 027444, 027445, 027446, 027447, 027448, 027449, 027450, 027451, 027452, 027453, 027454, 027455, 027456, 027457, 027458, 027459, 027460, 027461, 027462, 027463, 027464, 027465, 027466, 027467, 027468, 027469, 027470, 027471, 027473, 027474, 027475, 027476, 027477, 027478, 027480, 027481, 027482, 027483, 027484, 027485, 027486, 027487, 027488, 027489, 027490, 027491, 027493, 027554, 027555, 027556, 027557, 027559, 027560, 027561, 027562, 027563, 027564, 027568, 028010, 028011, 028012, 028013, 028014, 028015, 028016, 028017, 028018, 028019, 028020, 028021, 028022, 028023, 028024, 028125, 028126, 028127, 028128, 028129, 028130, 028131, 028132, 028133, 028134, 028135, 028136, 028137, 028139, 028140, 028141, 028142, 028143, 028144, 028145, 028146, 028147, 028148, 028149, 028150, 028151, 028152, 028153, 028154.   Model P108   - Serial numbers: 100355, 100356, 100357, 100358, 100359, 100360, 100361, 100364, 100365, 100366, 100367, 100369, 100370, 100371, 100372, 100373, 100374, 100375, 100376, 100377, 100378, 100379, 100381, 100382, 100383, 100384, 100386, 100387, 100388, 100389, 100390, 100391, 100392, 100393, 100394, 100397, 100398, 100399, 100400, 100402, 100403, 100404, 100405, 100406, 100407, 100409, 100410, 100411, 100412, 100413, 100414, 100415, 100416, 100417, 100418.100419, 100420, 100421, 100422, 100423, 100424, 100425, 100426, 100427, 100428, 100429, 100430, 100431, 100432, 100433, 100434, 100435, 100436, 100437, 100438, 100440, 100441, 100442, 100443, 100444, 100445, 100446, 100447, 100448, 100449, 100450, 100451, 100452, 100453, 100454, 100455, 100457, 100459, 100460, 100461, 100462, 100463, 100465, 100466, 100467, 100469, 100470, 100471, 100472, 100473, 100475, 100476, 100477, 100478, 100479, 100480, 100481, 100483, 100484, 100485, 100486, 100487, 100488, 100489, 100490, 100491, 100492, 100493, 100494, 100495, 100496, 100498, 100499, 100500, 100501, 100502, 100503, 100504, 100506, 100508, 100509, 100510, 100511, 100512, 100514, 100515, 100517, 100518, 100519, 100520, 100521, 100522, 100523, 100524, 100526, 100527, 100528, 100529, 100530, 100531, 100532, 100533, 100534, 100535, 100536, 100537, 100538, 100539, 100540, 100541, 100542, 100543, 100544, 100545, 100546, 100547, 100548, 100551, 100552, 100553, 100554, 100555, 100557, 100558, 100559, 100560, 100561, 100562, 100563, 100564, 100565, 100566, 100567, 100568, 100570, 100571, 100572, 100573, 100574, 100575, 100576, 100577, 100578, 100579, 100580, 100581, 100582, 100584, 100585, 100587, 100588, 100589, 100590, 100591, 100592, 100593, 100594, 100595, 100596, 100597, 100598, 100600, 100601, 100602, 100603, 100604, 100605, 100607, 100608, 100609, 100610, 100611, 100612, 100613, 100614, 100615, 100616, 100617, 100618, 100619, 100620, 100621, 100622, 100623, 100624, 100626, 100627, 100629, 100630, 100631, 100632, 100633, 100634, 100635, 100636, 100637, 100638, 100639, 100640, 100642, 100643, 100644, 100646, 100647, 100648, 100649, 100650, 100651, 100652, 100653, 100654, 100655, 100656, 100657, 100658, 100659, 100660, 100661, 100662, 100663, 100664, 100665, 100666, 100668, 100669, 100670, 100671, 100672, 100673, 100674, 100676, 100677, 100678, 100679, 100680, 100681, 100682, 100683, 100685, 100686, 100687, 100688, 100690, 100692, 100693, 100694, 100695, 100696, 100698, 100699, 100700, 100701, 100703, 100704.  
FEI Number 2124215
Recalling Firm/
Manufacturer
Boston Scientific CRM Corp
4100 Hamline Ave N
Saint Paul MN 55112-5700
For Additional Information ContactUS Technical Services
800-227-3422
Manufacturer Reason
for Recall
Boston Scientific identified a subset of COGNISTM CRT-Ds and TELIGENTM ICDs that had experienced an increased rate of premature battery depletion due to compromised performance of a low voltage (LV) capacitor. Boston Scientific updated their August 2013 Physician communication to include additional devices that may experience premature battery depletion due to compromised performance of a low volt
FDA Determined
Cause 2
Component design/selection
ActionBoston Scientific sent an Important Medical Device Information letter to physicians beginning September 17, 2014. The letter provided a summary of the issue, described clinical implications and expected device performance, and provided patient monitoring recommendations. Boston Scientific is introducing upgraded program software (Model 2868, version 3.04) that enhances the effectiveness of the Safety Architecture tools. Questions should be addressed to Boston Scientific Representative or Technical Services 1-800-227-3422 or tech.services@bsci.com.
Quantity in Commerce31,144 total
DistributionWorldwide Distribution - US Nationwide including American Samoa, Guam, Puerto Rico, and U.S. - Virgin Islands. OUS countries to include: Argentina, Aruba, Australia, Austria, Bahamas, Barbados, Belgium, Bermuda, Brazil, Canada, Cayman Islands, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Germany, Greece, Guadeloupe, Haiti, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Korea, Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Monaco, Mozambique, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, North Yemen, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan R O C, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.