| Class 1 Device Recall The Plum A Infusion System |  |
Date Initiated by Firm | July 02, 2014 |
Date Posted | February 23, 2015 |
Recall Status1 |
Terminated 3 on May 24, 2017 |
Recall Number | Z-1073-2015 |
Recall Event ID |
69336 |
510(K)Number | K042081 |
Product Classification |
Pump, infusion - Product Code FRN
|
Product | The Plum A+ Infusion System is designed to meet the fluid delivery requirements of today's evolving healthcare environments. It is a cassette based multi-function infusion system. The Plum A+ allows two lines in and one line out. The pump can be used for standard, piggyback, or concurrent delivery. Therapy modes include: Standard Infusions, Multistep Programming, Loading Dose, and Dose Calculation. The Plum A+ is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types. Each system includes a pumping module and an assortment of disposable IV sets, optional accessories, and an operator's manual. |
Code Information |
1) List Number: 20678: Serial Numbers: 15867001 , 15938721 , 15900791 , 15900971 , 15901051 , 15901101 , 15901301 , 15902021 , 15886121 , 15886191 , 15886201 , 15886331 , 15887441 , 15887451 , 15887501 , 15887541 , 15887551 , 15887561 , 15887571 , 15887591 , 15887601 , 15887621 , 15887721 , 15887781 , 15887801 , 15887811 , 15887841 , 15887851 , 15887861 , 15909721 , 15910521 , 15910551 , 15911281 , 15911441 , 15911501 , 15911631 , 15922951 , 15926291 , 15926731 , 15926961 , 15927281 , 15927621 , 15927731 , 15927781 , 15928041 , 15928061 , 15928131 , 15928171 , 15928221 , 15928231 , 15928291 , 15928951 , 15929051 , 15929151 , 15929351 , 15929531 , 15929901 , 15930141 , 15930191 , 15930341 , 15930351 , 15930641 , 15930691 , 15930711 , 15931171 , 15933301 , 15933501 , 15933821 , 15933941 , 15933991 , 15934001 , 15934171 , 15934261 , 15934681 , 15934711 , 15934841 , 15934891 , 15935011 , 15935071 , 15935141 , 15935191 , 15935221 , 15935261 , 15935361 , 15935381 , 15935411 , 15935431 , 15935451 , 15935471 , 15935521 , 15935541 , 15935551 , 15935561 , 15935601 , 15935891 , 15936011 , 15936181 , 15936401 , 15937241 , 15937311 , 15937381 , 15937451 , 15937461 , 15938101 , 15938601 , 15938621 , 15938731 , 15938741 , 15938751 , 15938761 , 15938871 , 15941441 , 15943391 , 15944341 , 15944421 , 15944721 , 15944751 , 15944861 , 15944881 , 15945021 , 15945091 , 15945121 , 15945381 , 15945531 , 15945921 , 15947671 , 15958991 , 15541871 , 15861851 , 15862021 , 15864801 , 15871841 , 15892651 , 15892671 , 15896621 , 15903101 , 15904551 , 15904961 , 15905391 , 15905491 , 15907531 , 15907591 , 15907751 , 15907791 , 15907801 , 15908091 , 15908471 , 15909671 , 15910131 , 15910371 , 15911701 , 15914421 , 15917421 , 15917721 , 15918431 , 15918731 , 15918951 , 15919281 , 15920761 , 15922001 , 15922761 , 15922801 , 15923281 , 15923551 , 15924971 , 15924991 , 15925081 , 15925701 , 15925731 , 15925741 , 15925791 , 15926141 , 15926191 , 15926271 , 15926281 , 15926301 , 15928071 , 15929601 , 15929621 , 15930171 , 15930631 , 15932561 , 15933451 , 15933561 , 15933721 , 15933811 , 15933831 , 15934781 , 15935321 , 15935761 , 15936121 , 15936361 , 15936561 , 15936601 , 15936951 , 15937031 , 15937041 , 15937061 , 15937191 , 15938121 , 15938471 , 15938481 , 15943001 , 15944541 , 15952631 , 15954621 , 15956511 , 15957721 , 15958091 , 15958161 , 15958241 , 15958511 , 15958691 , 15958941 , 15960891 , 15960971 , 19240471 , 19240541 , 19240771 , 19240801 , 19240861 , 19241171 , 19241441 , 15945821 , 15946021 , 15951911 , 19241901 , 19242221 , 19250221 , 19414401 , 19414451 , 19414521 , 19414571 , 19414581 , 19414601 , 19414631 , 19414641 , 19414651 , 19414661 , 19414691 , 19414711 , 19414721 , 19414751 , 19414781 , 19414811 , 19414821 , 19414841 , 19414881 , 19414901 , 19414921 , 19414981 , 19415011 , 19415041 , 19415061 , 19415071 , 19415081 , 19415091 , 19415101 , 19415111 , 19415131 , 19415191 , 19415241 , 19415251 , 19415261 , 19415321 , 19415351 , 19415361 , 19415391 , 19415421 , 19415461 , 19415491 , 19415511 , 19415521 , 19415551 , 19415581 , 19415701 , 19415731 , 19415741 , 19415751 , 19415801 , 19415831 , 19415851 , 19415891 , 19415951 , 19416331 , 19416341 , 19416361 , 19416371 , 19416381 , 19416401 , 19416411 , 19416441 , 19416451 , 19416461 , 19416471 , 19416491 , 19416521 , 19416551 , 19416571 , 19416591 , 19416611 , 19416621 , 19416661 , 19416681 , 19416751 , 19416771 , 19416781 , 19416801 , 19416811 , 19416821 , 19416831 , 19416911 , 19416941 , 19416961 , 19417061 , 19417181 , 19617581 , 15546971 , 15946081 , 15946181 , 15948941 , 15954341 , 19241451 , 19241471 , 19241731 , 19241801 , 19241891 , 19241921 , 19241971 , 19242061 , 19242081 , 19242161 , 19242191 , 19242441 , 19242541 , 19242621 , 19242651 , 19242691 , 19244991 , 19245031 , 19248621 , 19250231 , 19250351 , 19414501 , 19414761 , 19415861 , 19417901 , 19418771 , 19419631 , 19420311 , 19420471 , 19422911 , 19615331 , 19617421 , 19617661 , 19582951 , 19612341 , 19645601 , 19646941 , 19647301 , 19647311 , 19647741 , 19647821 , 18648161 , 18648201 , 18648541 , 18649471 , 18652351 , 18654881 , 18656291 , 18666591 , 18666791 , 18666821 , 19577461 , 19645631 , 19416561 , 19423351 , 19758231 , 19758241 , 19758251 , 19758261 , 19758271 , 19758291 , 19758301 , 19758311 , 19758321 , 19758351 , 19758361 , 19758371 , 19758391 , 19758401 , 19758411 , 19758421 , 19758431 , 19758441 , 19758451 , 19758461 , 19758471 , 19758481 , 19758501 , 19758521 , 19758531 , 19758551 , 19758591 , 19760451 , 19760481 , 19760521 , 19760591 , 19760671 , 19760711 , 19760771 , 19760811 , 19760851 , 19760921 , 19760951 , 19760961 , 19761011 , 19761031 , 19761041 , 19761061 , 19761071 , 19761091 , 19761111 , 19761141 , 19761151 , 19761191 , 19761231 , 19761251 , 19761351 , 19761381 , 19761581 , 19761651 , 19761931 , 19761961 , 19761991 , 19762001 , 19762111 , 19762241 , 19762271 , 19762311 , 19762361 , 19762381 , 19762401 , 19762481 , 19762521 , 19762541 , 19762571 , 19762581 , 19762631 , 19762651 , 19762711 , 19762721 , 19762731 , 19762771 , 19762821 , 19762841 , 19762891 , 19762971 , 19763041 , 19763111 , 19763141 , 19763181 , 19763251 , 19763291 , 19763301 , 19763341 , 19763351 , 19763491 , 19763511 , 19763641 , 19763721 , 19763751 , 19763761 , 19763781 , 19763801 , 19763841 , 19763871 , 19763911 , 19763971 , 19763981 , 19763991 , 19764041 , 19764061 , 19764171 , 19764181 , 19764191 , 19764201 , 19764421 , 19764461 , 19867171 , 19867311 , 19867471 , 19867931 , 19868541 , 19879621 , 18665541 , 19603901 , 19758491 , 19760721 , 19761361 , 19763461 , 19763921 , 19764071 , 19764091 , 19764151 , 19764251 , 19764301 , 19764361 , 19764371 , 19764391 , 19764401 , 19764451 , 19764691 , 19764701 , 15541171 , 15542831 , 15542861 , 15542911 , 15542921 , 15543011 , 15543031 , 15543281 , 15543521 , 15543571 , 15543591 , 15543621 , 15543741 , 15543751 , 15544571 , 15544701 , 15867971 , 15872871 , 15876321 , 15877001 , 15877441 , 15877841 , 15893591 , 15893631 , 15893671 , 15893851 , 15893861 , 15893891 , 15893951 , 15894061 , 15894091 , 15894181 , 15894201 , 15894221 , 15894271 , 15894281 , 15894751 , 15894791 , 15894961 , 15895091 , 15895181 , 15895201 , 15895341 , 15895401 , 15895471 , 15895481 , 15542891 , 15543631 , 15863781 , 15868271 , 15871071 , 15874161 , 15874321 , 15874381 , 15881751 , 15891981 , 15892221 , 15893761 , 15893871 , 15894011 , 15894101 , 15895071 , 15895291 ; 2) List Number: 20679; Serial Numbers: 15446876 , 15446886 , 15447262 , 15447455 , 15447612 , 15447652 , 15448232 , 15448296 , 15448362 , 15448426 , 15445226 , 15446166 , 15446197 , 15448292 , 15450782 , 15451773 ; 3) List Number: 20792; Serial Numbers: 17712308 , 17713271 , 17713907 , 17714143 , 17714392 , 17714490 , 17714586 , 17715518 , 17715534 , 17715716 , 17715896 , 17715918 , 17710912 , 16094467 , 16095171 , 16095742 , 16095903 , 16106872 , 16108272 , 16109155 , 16109175 , 16109210 , 16111862 , 16114766 , 16119246 , 16072646 , 17685345 , 17711612 , 16091311 , 16085563 , 16089622 , 16089677 , 16089915 , 16089917 , 16089971 , 16091539 , 16091753 , 16063744 , 16076969 , 16092882 , 16093457 , 16093477 , 16094461 , 16094516 , 16094526 , 16094543 , 16094546 , 16094547 , 16094560 , 16094574 , 16094584 , 16094643 , 16094655 , 16094656 , 16094657 , 16094662 , 16094719 , 16094722 , 16094852 , 16094869 , 16094894 , 16094910 , 16094923 , 16094953 , 16094955 , 16094957 , 16094967 , 16094968 , 16094975 , 16094990 , 16095005 , 16095015 , 16095016 , 16095027 , 16095033 , 16095036 , 16095042 , 16095044 , 16095050 , 16095052 , 16095056 , 16095058 , 16095059 , 16095063 , 16095069 , 16095072 , 16095073 , 16095083 , 16095084 , 16095086 , 16095088 , 16095106 , 16095107 , 16095109 , 16095110 , 16095126 , 16095133 , 16095142 , 16095145 , 16095173 , 16095174 , 16095180 , 16095183 , 16095185 , 16095193 , 16095198 , 16095199 , 16095201 , 16095204 , 16095211 , 16095213 , 16095218 , 16095221 , 16095224 , 16095344 , 16095350 , 16095618 , 16095758 , 16095777 , 16095787 , 16095791 , 16095793 , 16095813 , 16095825 , 16095829 , 16095830 , 16095832 , 16095836 , 16095846 , 16095853 , 16095874 , 16095876 , 16095877 , 16095878 , 16095894 , 16095898 , 16095907 , 16095913 , 16095940 , 16095945 , 16095962 , 16095986 , 16095998 , 16096008 , 16096014 , 16096038 , 16096039 , 16096170 , 16096227 , 16096236 , 16099106 , 16100030 , 16100100 , 16100102 , 16100213 , 16100231 , 16100245 , 16100251 , 16100292 , 16100337 , 16100363 , 16100504 , 16100551 , 16100577 , 16101218 , 16101445 , 16101523 , 16101543 , 16101566 , 16101574 , 16101577 , 16101581 , 16101586 , 16101587 , 16101588 , 16101589 , 16101591 , 16101594 , 16101596 , 16101600 , 16101607 , 16101608 , 16101612 , 16101628 , 16101629 , 16101630 , 16101631 , 16101635 , 16101639 , 16101667 , 16101668 , 16101677 , 16101682 , 16101685 , 16101696 , 16101697 , 16101704 , 16101705 , 16101706 , 16101709 , 16101716 , 16101719 , 16101721 , 16101746 , 16101748 , 16101759 , 16101761 , 16101764 , 16101768 , 16101773 , 16101774 , 16101775 , 16101781 , 16101788 , 16101796 , 16101853 , 16101858 , 16101886 , 16101892 , 16101911 , 16101922 , 16102581 , 16102583 , 16102592 , 16102604 , 16102605 , 16102612 , 16102613 , 16102616 , 16102618 , 16102621 , 16102625 , 16102626 , 16102636 , 16102638 , 16102641 , 16102651 , 16102653 , 16102660 , 16102664 , 16102667 , 16102668 , 16102679 , 16102682 , 16102692 , 16102699 , 16102704 , 16102706 , 16102708 , 16102714 , 16102722 , 16102728 , 16102729 , 16102733 , 16102739 , 16102742 , 16102744 , 16102745 , 16102764 , 16102770 , 16102791 , 16102801 , 16102802 , 16102803 , 16102804 , 16102805 , 16102806 , 16102807 , 16102812 , 16102815 , 16102819 , 16102840 , 16102844 , 16102846 , 16102859 , 16102882 , 16102906 , 16102913 , 16102934 , 16102946 , 16102970 , 16102972 , 16102984 , 16102997 , 16102999 , 16103010 , 16103016 , 16103021 , 16103075 , 16103087 , 16103088 , 16103097 , 16103098 , 16103107 , 16103130 , 16103139 , 16103140 , 16103146 , 16103169 , 16103172 , 16103187 , 16103445 , 16103721 , 16103722 , 16103827 , 16103829 , 16103879 , 16104110 , 16104137 , 16104164 , 16104179 , 16104181 , 16104185 , 16104195 , 16104200 , 16104208 , 16104210 , 16104216 , 16104220 , 16104229 , 16104324 , 16104384 , 16104387 , 16104394 , 16104396 , 16104398 , 16104401 , 16104402 , 16104403 , 16104404 , 16104417 , 16104423 , 16104432 , 16104462 , 16104840 , 16104843 , 16104900 , 16104921 , 16104954 , 16104978 , 16104985 , 16104993 , 16105025 , 16105027 , 16105141 , 16105186 , 16105231 , 16105237 , 16105316 , 16105328 , 16105388 , 16105418 , 16105495 , 16105519 , 16105604 , 16105730 , 16105987 , 16106018 , 16106087 , 16106202 , 16106401 , 16106410 , 16106513 , 16106542 , 16106545 , 16106596 , 16106686 , 16106774 , 16106875 , 16106908 , 16106935 , 16106939 , 16107109 , 16107127 , 16107129 , 16107268 , 16107354 , 16107386 , 16107417 , 16107426 , 16107468 , 16107475 , 16107476 , 16107478 , 16107481 , 16107500 , 16107522 , 16107523 , 16107525 , 16107533 , 16107542 , 16107555 , 16107562 , 16107574 , 16107603 , 16107607 , 16107732 , 16107864 , 16107869 , 16107911 , 16108044 , 16108105 , 16108155 , 16108904 , 16108924 , 16108928 , 16108936 , 16109006 , 16109015 , 16109085 , 16109098 , 16109113 , 16109131 , 16109134 , 16109150 , 16109177 , 16109179 , 16109191 , 16109192 , 16109196 , 16109199 , 16109200 , 16109202 , 16109203 , 16109205 , 16109215 , 16109220 , 16109223 , 16109224 , 16109225 , 16109228 , 16109229 , 16109231 , 16109232 , 16109237 , 16109255 , 16109256 , 16109260 , 16109263 , 16109269 , 16109271 , 16109276 , 16109285 , 16109286 , 16109288 , 16109293 , 16109296 , 16109308 , 16109309 , 16109310 , 16109312 , 16109314 , 16109315 , 16109319 , 16109322 , 16109335 , 16109345 , 16109350 , 16109353 , 16109542 , 16109582 , 16110012 , 16110318 , 16110361 , 16110562 , 16110565 , 16110864 , 16111056 , 16111103 , 16111325 , 16111326 , 16111360 , 16111426 , 16111438 , 16111442 , 16111477 , 16111561 , 16111589 , 16111632 , 16111638 , 16111658 , 16111661 , 16111664 , 16111666 , 16111669 , 16111679 , 16111681 , 16111696 , 16111742 , 16111746 , 16111772 , 16111788 , 16111798 , 16111816 , 16111827 , 16111833 , 16111895 , 16111896 , 16111906 , 16111908 , 16111909 , 16111917 , 16111924 , 16111925 , 16111930 , 16111935 , 16111940 , 16111979 , 16111980 , 16111981 , 16111982 , 16112006 , 16112007 , 16112010 , 16112012 , 16112013 , 16112014 , 16112016 , 16112017 , 16112018 , 16112023 , 16112025 , 16112027 , 16112029 , 16112030 , 16112032 , 16112047 , 16112048 , 16112057 , 16112059 , 16112078 , 16112087 , 16112089 , 16112091 , 16112092 , 16112094 , 16112095 , 16112100 , 16112111 , 16112117 , 16112118 , 16112124 , 16112132 , 16112138 , 16112140 , 16112142 , 16112147 , 16112149 , 16112156 , 16112158 , 16112159 , 16112161 , 16112166 , 16112169 , 16112174 , 16112176 , 16112632 , 16112926 , 16112965 , 16113668 , 16113790 , 16113838 , 16113934 , 16113979 , 16114003 , 16114065 , 16114195 , 16114229 , 16114247 , 16114324 , 16114491 , 16114532 , 16114562 , 16114605 , 16114614 , 16114627 , 16114648 , 16114677 , 16114697 , 16114746 , 16114756 , 16114758 , 16114773 , 16114779 , 16114784 , 16114788 , 16114792 , 16114799 , 16114800 , 16114804 , 16114814 , 16114821 , 16114822 , 16114831 , 16114839 , 16114850 , 16114853 , 16114876 , 16114878 , 16114894 , 16114895 , 16114901 , 16115157 , 16115475 , 16115641 , 16115681 , 16115691 , 16115776 , 16115824 , 16115870 , 16115929 , 16115975 , 16116145 , 16116189 , 16116419 , 16116425 , 16116428 , 16116466 , 16116487 , 16116512 , 16116521 , 16116542 , 16116560 , 16116592 , 16116639 , 16116644 , 16116684 , 16116707 , 16116786 , 16116788 , 16117540 , 16117603 , 16117678 , 16117783 , 16117826 , 16119315 , 16119354 , 16119394 , 16119505 , 16119540 , 16061264 , 16062593 , 16062866 , 16062867 , 16062911 , 16062917 , 16063364 , 16063416 , 16063492 , 16063550 , 16063656 , 16063884 , 16064002 , 16064014 , 16064026 , 16064104 , 16064106 , 16064110 , 16064122 , 16064123 , 16064330 , 16064336 , 16064430 , 16064493 , 16064739 , 16065745 , 16066056 , 16066063 , 16066076 , 16066406 , 16066409 , 16066475 , 16067218 , 16067801 , 16067853 , 16068248 , 16068543 , 16068584 , 16068617 , 16068637 , 16068707 , 16068718 , 16068724 , 16068827 , 16068957 , 16068968 , 16068983 , 16068984 , 16069002 , 16069248 , 16069757 , 16069765 , 16069906 , 16070176 , 16070787 , 16070928 , 16070999 , 16072230 , 16072668 , 16073961 , 16074179 , 16074772 , 16076031 , 16076057 , 16076601 , 16076632 , 16076957 , 16077006 , 16077023 , 16077144 , 16077233 , 16077292 , 16077313 , 16077347 , 16077364 , 16077556 , 16078216 , 16078458 , 16078577 , 16078972 , 16080175 , 16080599 , 16080683 , 16080844 , 16083530 , 16084924 , 16085715 , 16086023 , 16086416 , 16086460 , 16087306 , 16087366 , 16087373 , 16087454 , 16087540 , 16087569 , 16087987 , 16088130 , 16088275 , 16088305 , 16088742 , 16089357 , 16089607 , 16089620 , 16089630 , 16089633 , 16089637 , 16089641 , 16090767 , 16091106 , 16091623 , 16092297 , 16092747 , 16092876 , 16092977 , 16093056 , 16093120 , 16093126 , 16093129 , 16093141 , 16093143 , 16093153 , 16093156 , 16093159 , 16093178 , 16093195 , 16093196 , 16093202 , 16093226 , 16093238 , 16093242 , 16093247 , 16093253 , 16093255 , 16093259 , 16093280 , 16093283 , 16093285 , 16093300 , 16093307 , 16093322 , 16093323 , 16093329 , 16093513 , 16093574 , 16093585 , 16093679 , 16093791 , 16093809 , 16093814 , 16093828 , 16093843 , 16093857 , 16093869 , 16093870 , 16093878 , 16093880 , 16093893 , 16093898 , 16093924 , 16093990 , 16094028 , 16094087 , 16094127 , 16094251 , 16094388 , 16094392 , 16094470 , 16094472 , 16094484 , 16094485 , 16094486 , 16094498 , 16094499 , 16094502 , 16094525 , 16094541 , 16094552 , 16094561 , 16094566 , 16094567 , 16094572 , 16094608 , 16094647 , 16094660 , 16094661 , 16094669 , 16094672 , 16094677 , 16094680 , 16094694 , 16094717 , 16094725 , 16094731 , 16094732 , 16094734 , 16094753 , 16094766 , 16094769 , 16094774 , 16094778 , 16094782 , 16094787 , 16094876 , 16094914 , 16094917 , 16094945 , 16094954 , 16094963 , 16094965 , 16094966 , 16094973 , 16094976 , 16094978 , 16094983 , 16094984 , 16094996 , 16095002 , 16095007 , 16095010 , 16095023 , 16095030 , 16095031 , 16095035 , 16095048 , 16095051 , 16095053 , 16095054 , 16095061 , 16095064 , 16095081 , 16095093 , 16095100 , 16095108 , 16095113 , 16095116 , 16095118 , 16095119 , 16095123 , 16095148 , 16095154 , 16095156 , 16095162 , 16095165 , 16095166 , 16095176 , 16095191 , 16095282 , 16095323 , 16095324 , 16095625 , 16095727 , 16095736 , 16095740 , 16095745 , 16095759 , 16095773 , 16095782 , 16095790 , 16095807 , 16095808 , 16095811 , 16095819 , 16095820 , 16095822 , 16095835 , 16095842 , 16095844 , 16095848 , 16095851 , 16095852 , 16095863 , 16095864 , 16095866 , 16095870 , 16095875 , 16095879 , 16095882 , 16095883 , 16095891 , 16095897 , 16095902 , 16095921 , 16095924 , 16095931 , 16095934 , 16095951 , 16095954 , 16095956 , 16095957 , 16095960 , 16095975 , 16095976 , 16095977 , 16095981 , 16095982 , 16095984 , 16095989 , 16095999 , 16096020 , 16096041 , 16096047 , 16096049 , 16096052 , 16096061 , 16096095 , 16096112 , 16096113 , 16096120 , 16096140 , 16096158 , 16096209 , 16096225 , 16096244 , 16096266 , 16096417 , 16096508 , 16096598 , 16096963 , 16097242 , 16097254 , 16097273 , 16097342 , 16097370 , 16097379 , 16097384 , 16097418 , 16097467 , 16097468 , 16097483 , 16097504 , 16097631 , 16097720 , 16097743 , 16097777 , 16097784 , 16097797 , 16097890 , 16097919 , 16097922 , 16097942 , 16097944 , 16097946 , 16097970 , 16097971 , 16097972 , 16097974 , 16098007 , 16098022 , 16098028 , 16098034 , 16098035 , 16098045 , 16098051 , 16098055 , 16098095 , 16098139 , 16098154 , 16098161 , 16098207 , 16098219 , 16098231 , 16098232 , 16098249 , 16098253 , 16098264 , 16098296 , 16098312 , 16098398 , 16098419 , 16098439 , 16098522 , 16098552 , 16098554 , 16098558 , 16098563 , 16098678 , 16098724 , 16098874 , 16099114 , 16099259 , 16099675 , 16099689 , 16099935 , 16099981 , 16100694 , 16101290 , 16101717 , 16101808 , 16103577 , 16103850 , 16104042 , 16104096 , 16104145 , 16104159 , 16104160 , 16104161 , 16104182 , 16104184 , 16104186 , 16104201 , 16104202 , 16104203 , 16104206 , 16104209 , 16104212 , 16104214 , 16104222 , 16104228 , 16104251 , 16104274 , 16104466 , 16104480 , 16104509 , 16104518 , 16104647 , 16104688 , 16105676 , 16106450 , 16106532 , 16106789 , 16106835 , 16106942 , 16107110 , 16107128 , 16107362 , 16107430 , 16107482 , 16107483 , 16107504 , 16107655 , 16107702 , 16107887 , 16107900 , 16107930 , 16107939 , 16107986 , 16108059 , 16108156 , 16108485 , 16109159 , 16109165 , 16109235 , 16109279 , 16109786 , 16110841 , 16110848 , 16110899 , 16111074 , 16111267 , 16111702 , 16111919 , 16112031 , 16112401 , 16112445 , 16113747 , 16113784 , 16113793 , 16113852 , 16113900 , 16113984 , 16113990 , 16114001 , 16114020 , 16114022 , 16114277 , 16114628 , 16114723 , 16114757 , 16114813 , 16114914 , 16116649 , 16117414 , 16117622 , 16118028 , 16118161 , 16118316 , 16118402 , 16118487 , 16118522 , 16118570 , 16118578 , 16118678 , 16118683 , 16118710 , 16118714 , 16118759 , 16118769 , 16118771 , 16118794 , 16118862 , 16118874 , 16118958 , 16119007 , 16119013 , 16119018 , 16119021 , 16119023 , 16119028 , 16119045 , 16119076 , 16119082 , 16119095 , 16119159 , 16119173 , 16119176 , 16119181 , 16119187 , 16119206 , 16119212 , 16119265 , 16122888 , 16122893 , 16122894 , 16122898 , 16122900 , 16122903 , 16122911 , 16122923 , 16122927 , 16122928 , 16122947 , 16122952 , 16122953 , 16122955 , 16122957 , 16122961 , 16122964 , 16122968 , 16122969 , 16122971 , 16122973 , 16122974 , 16122984 , 16122987 , 16122989 , 16122999 , 16123002 , 16123003 , 16123009 , 16123010 , 16123025 , 16123028 , 16123032 , 16123034 , 16123035 , 16123037 , 16123044 , 16123360 , 16128020 , 16102615 , 16112157 , 16117462 , 16119391 , 16119526 , 16119532 , 16119597 , 16119792 , 16119867 , 16119889 , 16119898 , 16119993 , 16120022 , 16120028 , 16120045 , 16120061 , 16120086 , 16120145 , 16120147 , 16120153 , 16120154 , 16120170 , 16120182 , 16120204 , 16120205 , 16120224 , 16120256 , 16120329 , 16120337 , 16120394 , 16120889 , 16121467 , 16122673 , 16122889 , 16122906 , 16123030 , 16123038 , 16123043 , 16124689 , 16125324 , 16125540 , 16126882 , 16126904 , 16127058 , 16127426 , 16128234 , 16128410 , 16129394 , 16129484 , 17667642 , 17668222 , 17667548 , 17667894 , 17668067 , 17668293 , 17668341 , 17668509 , 17668120 , 17668374 , 17668453 , 17668814 , 17669162 , 17669192 , 17669437 , 17669718 , 17669748 , 17669773 , 17669787 , 17669828 , 17669861 , 17669867 , 17669901 , 17669902 , 17670019 , 17670188 , 17671466 , 17671467 , 17671469 , 17671472 , 17671473 , 17671479 , 17671480 , 17671483 , 17671484 , 17671485 , 17671488 , 17671490 , 17671492 , 17671493 , 17671494 , 17671495 , 17671497 , 17671499 , 17671500 , 17671501 , 17671502 , 17671503 , 17671505 , 17671507 , 17671509 , 17671510 , 17671511 , 17671512 , 17671515 , 17671518 , 17671519 , 17671522 , 17671528 , 17671529 , 17671532 , 17671534 , 17671535 , 17671539 , 17671540 , 17671543 , 17671544 , 17671546 , 17671547 , 17671548 , 17671549 , 17671551 , 17671553 , 17671555 , 17671558 , 17671562 , 17671563 , 17671564 , 17671565 , 17671567 , 17671568 , 17671570 , 17671572 , 17671573 , 17671577 , 17671578 , 17671580 , 17671583 , 17671587 , 17671588 , 17671594 , 17671604 , 17671605 , 17671612 , 17671614 , 17671617 , 17671620 , 17671621 , 17671623 , 17671628 , 17671629 , 17671640 , 17671643 , 17671646 , 17671650 , 17671651 , 17671664 , 17671671 , 17671674 , 17671675 , 17671678 , 17671686 , 17671688 , 17671690 , 17671692 , 17671704 , 17671707 , 17671708 , 17671711 , 17671712 , 17671713 , 17671714 , 17671715 , 17671716 , 17671717 , 17671718 , 17671719 , 17671720 , 17671721 , 17671735 , 17671738 , 17671739 , 17671740 , 17671741 , 17671743 , 17671745 , 17671747 , 17671748 , 17671749 , 17671756 , 17671758 , 17671760 , 17671763 , 17671765 , 17671766 , 17671770 , 17671774 , 17671776 , 17671779 , 17671782 , 17671784 , 17671788 , 17671804 , 17671810 , 17671815 , 17671819 , 17671821 , 17671822 , 17671824 , 17671825 , 17671826 , 17671827 , 17671828 , 17671831 , 17671836 , 17671843 , 17671845 , 17671847 , 17671849 , 17671852 , 17671853 , 17671856 , 17671857 , 17671859 , 17671860 , 17671861 , 17671863 , 17671867 , 17671876 , 17671879 , 17671888 , 17671892 , 17671913 , 17671946 , 17671947 , 17671952 , 17671954 , 17671955 , 17671956 , 17671958 , 17671961 , 17671962 , 17671963 , 17671964 , 17671965 , 17671969 , 17671973 , 17671977 , 17671980 , 17671983 , 17671989 , 17671991 , 17671993 , 17671995 , 17671998 , 17671999 , 17672004 , 17672005 , 17672009 , 17672012 , 17672013 , 17672016 , 17672021 , 17672022 , 17672026 , 17672030 , 17672031 , 17672035 , 17672046 , 17672059 , 17672061 , 17672062 , 17672063 , 17672064 , 17672065 , 17672066 , 17672068 , 17672069 , 17672070 , 17672071 , 17672072 , 17672073 , 17672074 , 17672076 , 17672081 , 17672084 , 17672085 , 17672086 , 17672087 , 17672089 , 17672098 , 17672101 , 17672112 , 17672113 , 17672119 , 17672120 , 17672121 , 17672122 , 17672125 , 17672126 , 17672128 , 17672130 , 17672134 , 17672139 , 17672150 , 17672151 , 17672154 , 17672156 , 17672159 , 17672163 , 17672165 , 17672166 , 17672169 , 17672170 , 17672174 , 17672176 , 17672180 , 17672183 , 17672187 , 17672188 , 17672191 , 17672192 , 17672197 , 17672200 , 17672201 , 17672203 , 17672206 , 17672207 , 17672210 , 17672211 , 17672212 , 17672215 , 17672220 , 17672221 , 17672223 , 17672224 , 17672226 , 17672227 , 17672228 , 17672232 , 17672234 , 17672237 , 17672238 , 17672239 , 17672240 , 17672241 , 17672242 , 17672243 , 17672244 , 17672245 , 17672246 , 17672255 , 17672262 , 17672263 , 17672264 , 17672267 , 17672268 , 17672270 , 17672278 , 17672288 , 17672290 , 17672292 , 17672294 , 17672298 , 17672304 , 17672310 , 17672313 , 17672421 , 17672450 , 17672452 , 17672455 , 17672456 , 17672460 , 17672464 , 17672466 , 17672468 , 17672511 , 17672513 , 17672519 , 17672521 , 17672525 , 17672529 , 17672533 , 17672537 , 17672538 , 17672540 , 17672541 , 17672546 , 17672548 , 17672553 , 17672555 , 17672556 , 17672559 , 17672560 , 17672562 , 17672563 , 17672568 , 17672569 , 17672571 , 17672574 , 17672593 , 17672599 , 17672600 , 17672607 , 17672613 , 17672614 , 17672615 , 17672616 , 17672617 , 17672618 , 17672620 , 17672623 , 17672625 , 17672626 , 17672630 , 17672631 , 17672632 , 17672637 , 17672638 , 17672639 , 17672644 , 17672645 , 17672646 , 17672647 , 17672649 , 17672652 , 17672653 , 17672655 , 17672656 , 17672658 , 17672659 , 17672664 , 17672666 , 17672671 , 17672672 , 17672676 , 17672677 , 17672681 , 17672686 , 17672688 , 17672689 , 17672690 , 17672691 , 17672699 , 17672700 , 17672709 , 17674336 , 17676423 , 17684084 , 17684126 , 17684139 , 17684675 , 17684830 , 17684859 , 17684929 , 17686270 , 17686938 , 17687789 , 17687794 , 17687803 , 17687829 , 17687840 , 17688279 , 17689506 , 17689916 , 17691022 , 17692207 , 17692218 , 17692447 , 17692467 , 17692468 , 17692472 , 17692495 , 17692499 , 17692506 , 17692615 , 17692738 , 17692741 , 17692750 , 17692751 , 17692757 , 17692779 , 17692781 , 17692791 , 17692800 , 17692808 , 17692810 , 17692820 , 17692823 , 17692832 , 17692834 , 17692837 , 17692839 , 17692843 , 17692844 , 17692845 , 17692846 , 17693034 , 17693040 , 17693450 , 17693467 , 17693492 , 17693508 , 17693519 , 17693538 , 17693563 , 17693591 , 17693596 , 17693621 , 17693652 , 17693686 , 17693695 , 17693840 , 17693843 , 17693896 , 17694032 , 17694040 , 17694053 , 17694066 , 17694067 , 17694079 , 17694087 , 17694092 , 17694097 , 17694112 , 17694204 , 17694210 , 17694455 , 17694597 , 17694603 , 17694610 , 17694616 , 17694619 , 17694629 , 17694659 , 17694663 , 17694666 , 17694671 , 17694673 , 17694675 , 17694694 , 17694695 , 17694700 , 17694702 , 17694746 , 17694753 , 17694762 , 17694980 , 17695057 , 17695059 , 17695064 , 17695067 , 17695075 , 17695080 , 17695092 , 17695460 , 17695484 , 17695487 , 17695494 , 17695498 , 17695504 , 17695505 , 17695507 , 17695509 , 17695513 , 17695519 , 17695521 , 17695526 , 17695528 , 17695534 , 17695574 , 17695585 , 17695663 , 17695675 , 17695677 , 17695682 , 17695686 , 17695690 , 17695696 , 17695699 , 17695701 , 17695703 , 17695706 , 17695714 , 17695723 , 17696054 , 17696056 , 17696062 , 17696088 , 17696101 , 17696120 , 17696121 , 17696122 , 17696130 , 17696146 , 17696159 , 17696161 , 17696171 , 17696173 , 17696180 , 17696185 , 17696186 , 17696187 , 17696190 , 17696197 , 17696198 , 17696199 , 17696200 , 17696201 , 17696202 , 17696203 , 17696204 , 17696208 , 17696210 , 17696224 , 17696279 , 17696286 , 17696297 , 17696298 , 17696305 , 17696306 , 17696307 , 17696308 , 17696311 , 17696314 , 17696555 , 17696943 , 17697064 , 17697065 , 17697067 , 17697068 , 17697072 , 17697075 , 17697083 , 17697685 , 17697692 , 17667676 , 17668631 , 17669083 , 17669275 , 17669840 , 17669843 , 17671487 , 17671520 , 17671552 , 17671581 , 17671595 , 17671600 , 17671726 , 17671730 , 17671813 , 17671833 , 17671838 , 17671846 , 17671851 , 17671869 , 17671904 , 17671960 , 17672043 , 17672092 , 17672103 , 17672108 , 17672209 , 17672230 , 17672260 , 17672281 , 17672359 , 17672379 , 17672388 , 17672430 , 17672440 , 17672457 , 17672467 , 17672473 , 17672516 , 17672536 , 17672551 , 17672554 , 17672674 , 17672675 , 17672727 , 17673198 , 17673364 , 17673381 , 17673792 , 17673902 , 17674164 , 17674418 , 17674544 , 17674716 , 17674894 , 17674950 , 17675333 , 17675403 , 17675518 , 17675541 , 17675694 , 17676084 , 17676095 , 17676100 , 17676146 , 17676244 , 17676561 , 17676598 , 17676740 , 17676928 , 17678633 , 17679250 , 17679435 , 17679682 , 17679692 , 17679752 , 17680513 , 17682117 , 17682129 , 17682150 , 17682200 , 17682610 , 17682805 , 17682839 , 17682915 , 17683029 , 17683041 , 17683043 , 17683115 , 17684736 , 17684907 , 17685425 , 17685562 , 17686544 , 17686617 , 17687165 , 17687881 , 17687975 , 17688281 , 17689145 , 17690424 , 17692487 , 17693497 , 17693525 , 17693547 , 17693630 , 17694069 , 17694413 , 17694446 , 17694472 , 17694747 , 17695333 , 17695347 , 17695371 , 17695722 , 17696116 , 17696184 , 17696207 , 17696287 , 17696434 , 17696594 , 17697082 , 17700689 , 16074381 , 16093785 , 16106145 , 16123840 , 16123912 , 16124048 , 16124832 , 16124964 , 17708186 , 17708190 , 17708231 , 17708233 , 17708236 , 17708240 , 17708247 , 17708250 , 17708252 , 17708255 , 17708259 , 17708265 , 17708649 , 17708732 , 17708738 , 17708742 , 17708748 , 17708749 , 17708771 , 17708773 , 17708784 , 17708793 , 17708834 , 17708842 , 17708843 , 17708846 , 17708848 , 17708849 , 17708852 , 17708853 , 17708854 , 17708855 , 17708856 , 17708857 , 17708858 , 17708859 , 17708860 , 17708861 , 17708862 , 17708866 , 17708870 , 17708871 , 17708873 , 17708874 , 17708876 , 17708878 , 17708880 , 17708882 , 17708884 , 17708885 , 17708888 , 17708889 , 17708890 , 17708891 , 17708892 , 17708893 , 17708895 , 17708896 , 17708897 , 17708898 , 17708900 , 17708901 , 17708902 , 17708904 , 17708906 , 17708907 , 17708908 , 17708911 , 17708914 , 17708915 , 17708916 , 17708919 , 17708920 , 17708921 , 17708922 , 17708925 , 17708927 , 17708930 , 17708932 , 17708933 , 17708934 , 17708936 , 17708939 , 17708951 , 17708957 , 17708958 , 17708959 , 17708963 , 17708964 , 17708965 , 17708966 , 17708967 , 17708968 , 17708969 , 17708970 , 17708972 , 17708975 , 17708976 , 17708980 , 17708982 , 17708984 , 17708985 , 17708986 , 17708988 , 17708989 , 17708990 , 17708991 , 17708993 , 17708995 , 17708996 , 17709005 , 17709010 , 17709011 , 17709012 , 17709014 , 17709015 , 17709017 , 17709019 , 17709023 , 17709024 , 17709026 , 17709028 , 17709029 , 17709117 , 17709125 , 17709128 , 17709129 , 17709130 , 17709131 , 17709132 , 17709133 , 17709134 , 17709137 , 17709141 , 17709145 , 17709146 , 17709147 , 17709148 , 17709152 , 17709156 , 17709157 , 17709158 , 17709159 , 17709160 , 17709161 , 17709163 , 17709164 , 17709169 , 17709170 , 17709172 , 17709176 , 17709182 , 17709183 , 17709184 , 17709185 , 17709198 , 17709199 , 17709202 , 17709203 , 17709204 , 17709205 , 17709206 , 17709207 , 17709208 , 17709211 , 17709213 , 17709343 , 17709345 , 17709347 , 17709349 , 17709352 , 17709357 , 17709362 , 17709366 , 17709393 , 17709397 , 17709398 , 17709502 , 17709507 , 17709509 , 17709511 , 17709515 , 17709520 , 17709523 , 17709525 , 17709526 , 17709527 , 17709528 , 17709529 , 17709530 , 17709531 , 17709533 , 17709534 , 17709537 , 17709538 , 17709539 , 17709540 , 17709541 , 17709542 , 17709543 , 17709544 , 17709545 , 17709546 , 17709547 , 17709548 , 17709551 , 17709552 , 17709553 , 17709554 , 17709555 , 17709556 , 17709557 , 17709558 , 17709559 , 17709560 , 17709561 , 17709563 , 17709564 , 17709565 , 17709567 , 17709568 , 17709569 , 17709570 , 17709571 , 17709574 , 17709575 , 17709576 , 17709577 , 17709578 , 17709579 , 17709580 , 17709581 , 17709582 , 17709583 , 17709585 , 17709586 , 17709587 , 17709588 , 17709589 , 17709590 , 17709591 , 17709592 , 17709593 , 17709594 , 17709595 , 17709596 , 17709597 , 17709598 , 17709599 , 17709600 , 17709601 , 17709602 , 17709603 , 17709604 , 17709606 , 17709607 , 17709609 , 17709610 , 17709613 , 17709615 , 17709616 , 17709696 , 17709708 , 17709709 , 17709712 , 17709715 , 17709718 , 17709721 , 17709726 , 17709729 , 17709730 , 17709732 , 17709736 , 17709738 , 17709739 , 17709744 , 17709745 , 17709749 , 17709762 , 17709763 , 17709764 , 17709765 , 17709767 , 17709768 , 17709770 , 17709772 , 17709773 , 17709774 , 17709776 , 17709777 , 17709778 , 17709779 , 17709780 , 17709781 , 17709782 , 17709783 , 17709784 , 17709785 , 17709786 , 17709787 , 17709788 , 17709789 , 17709790 , 17709793 , 17709794 , 17709796 , 17709797 , 17709798 , 17709799 , 17709800 , 17709801 , 17709805 , 17709806 , 17709807 , 17709809 , 17709810 , 17709812 , 17709813 , 17709814 , 17709816 , 17709817 , 17709818 , 17709819 , 17709821 , 17709822 , 17709823 , 17709824 , 17709825 , 17709829 , 17709830 , 17709831 , 17709832 , 17709834 , 17709835 , 17709836 , 17709840 , 17709841 , 17709843 , 17709844 , 17709847 , 17709855 , 17709856 , 17709858 , 17709859 , 17709860 , 17709861 , 17709864 , 17709866 , 17709867 , 17709868 , 17709869 , 17709870 , 17709872 , 17709873 , 17709875 , 17709876 , 17709878 , 17709879 , 17709881 , 17709882 , 17709883 , 17709885 , 17709886 , 17709887 , 17709889 , 17709890 , 17709891 , 17709892 , 17709893 , 17709894 , 17709895 , 17709896 , 17709897 , 17709899 , 17709900 , 17709901 , 17709902 , 17709903 , 17709904 , 17709905 , 17709907 , 17709908 , 17709909 , 17709910 , 17709911 , 17709912 , 17709913 , 17709914 , 17709915 , 17709918 , 17709920 , 17709921 , 17709924 , 17709925 , 17709927 , 17709928 , 17709929 , 17709941 , 17709950 , 17709956 , 17709960 , 17709966 , 17709968 , 17709974 , 17710005 , 17710011 , 17710012 , 17710014 , 17710015 , 17710017 , 17710019 , 17710026 , 17710028 , 17710034 , 17710035 , 17710037 , 17710038 , 17710039 , 17710040 , 17710042 , 17710043 , 17710044 , 17710045 , 17710046 , 17710047 , 17710048 , 17710049 , 17710050 , 17710052 , 17710055 , 17710056 , 17710057 , 17710058 , 17710059 , 17710062 , 17710063 , 17710064 , 17710065 , 17710066 , 17710067 , 17710068 , 17710069 , 17710071 , 17710073 , 17710074 , 17710075 , 17710077 , 17710078 , 17710079 , 17710080 , 17710082 , 17710083 , 17710085 , 17710086 , 17710088 , 17710089 , 17710090 , 17710091 , 17710092 , 17710094 , 17710095 , 17710096 , 17710097 , 17710098 , 17710099 , 17710102 , 17710103 , 17710107 , 17710108 , 17710110 , 17710111 , 17710112 , 17710113 , 17710114 , 17710115 , 17710116 , 17710117 , 17710119 , 17710120 , 17710121 , 17710122 , 17710123 , 17710124 , 17710125 , 17710128 , 17710129 , 17710130 , 17710131 , 17710133 , 17710136 , 17710138 , 17710140 , 17710141 , 17710142 , 17710146 , 17710148 , 17710152 , 17710153 , 17710155 , 17710156 , 17710157 , 17710159 , 17710160 , 17710161 , 17710162 , 17710172 , 17710173 , 17710199 , 17710200 , 17710208 , 17710216 , 17710218 , 17710221 , 17710224 , 17710225 , 17710239 , 17710245 , 17710250 , 17710253 , 17710256 , 17710257 , 17710266 , 17710270 , 17710275 , 17710276 , 17710279 , 17710285 , 17710288 , 17710291 , 17710292 , 17710294 , 17710297 , 17710301 , 17710302 , 17710304 , 17710325 , 17710334 , 17710344 , 17710345 , 17710346 , 17710349 , 17710351 , 17710352 , 17710353 , 17710358 , 17710360 , 17710367 , 17710368 , 17710369 , 17710372 , 17710374 , 17710375 , 17710376 , 17710378 , 17710379 , 17710380 , 17710382 , 17710383 , 17710384 , 17710386 , 17710388 , 17710389 , 17710390 , 17710391 , 17710392 , 17710393 , 17710395 , 17710397 , 17710398 , 17710399 , 17710400 , 17710401 , 17710402 , 17710403 , 17710404 , 17710405 , 17710407 , 17710420 , 17710421 , 17710425 , 17710430 , 17710433 , 17710434 , 17710437 , 17710439 , 17710440 , 17710443 , 17710444 , 17710446 , 17710452 , 17710457 , 17710471 , 17710472 , 17710475 , 17710477 , 17710479 , 17710480 , 17710481 , 17710482 , 17710483 , 17710484 , 17710485 , 17710486 , 17710487 , 17710488 , 17710489 , 17710493 , 17710494 , 17710496 , 17710497 , 17710499 , 17710500 , 17710505 , 17710506 , 17710508 , 17710509 , 17710511 , 17710517 , 17710518 , 17710519 , 17710520 , 17710521 , 17710522 , 17710524 , 17710526 , 17710527 , 17710528 , 17710529 , 17710530 , 17710531 , 17710533 , 17710535 , 17710537 , 17710538 , 17710539 , 17710541 , 17710545 , 17710546 , 17710547 , 17710548 , 17710549 , 17710551 , 17710552 , 17710553 , 17710554 , 17710556 , 17710557 , 17710558 , 17710559 , 17710562 , 17710564 , 17710565 , 17710566 , 17710568 , 17710570 , 17710576 , 17710577 , 17710580 , 17710581 , 17710583 , 17710587 , 17710588 , 17710590 , 17710591 , 17710592 , 17710597 , 17710598 , 17710601 , 17710602 , 17710608 , 17710609 , 17710610 , 17710611 , 17710612 , 17710613 , 17710614 , 17710615 , 17710616 , 17710617 , 17710618 , 17710619 , 17710620 , 17710621 , 17710622 , 17710623 , 17710624 , 17710625 , 17710629 , 17710630 , 17710633 , 17710634 , 17710638 , 17710640 , 17710641 , 17710643 , 17710655 , 17710661 , 17710667 , 17710668 , 17710669 , 17710674 , 17710675 , 17710676 , 17710678 , 17710683 , 17710684 , 17710686 , 17710687 , 17710689 , 17710693 , 17710694 , 17710695 , 17710696 , 17710697 , 17710700 , 17710701 , 17710704 , 17710706 , 17710707 , 17710708 , 17710709 , 17710711 , 17710718 , 17710719 , 17710723 , 17710724 , 17710730 , 17710731 , 17710734 , 17710735 , 17710741 , 17710743 , 17710744 , 17710746 , 17710748 , 17710750 , 17710761 , 17710778 , 17710786 , 17710796 , 17710797 , 17710798 , 17710801 , 17710802 , 17710804 , 17710805 , 17710806 , 17710807 , 17710809 , 17710810 , 17710816 , 17710821 , 17710836 , 17710840 , 17710842 , 17710844 , 17710847 , 17710848 , 17710851 , 17710852 , 17710853 , 17710854 , 17710856 , 17710857 , 17710877 , 17710883 , 17710888 , 17710895 , 17710897 , 17710922 , 17710928 , 17710937 , 17710938 , 17710956 , 17710958 , 17710959 , 17710961 , 17710963 , 17710967 , 17710968 , 17710969 , 17710970 , 17710972 , 17710975 , 17710976 , 17710977 , 17710978 , 17710979 , 17710980 , 17710981 , 17710982 , 17710983 , 17710985 , 17710987 , 17710989 , 17710990 , 17710991 , 17710992 , 17710993 , 17710994 , 17710995 , 17710996 , 17710997 , 17711000 , 17711001 , 17711002 , 17711003 , 17711004 , 17711005 , 17711006 , 17711007 , 17711008 , 17711009 , 17711011 , 17711012 , 17711019 , 17711022 , 17711023 , 17711027 , 17711028 , 17711029 , 17711030 , 17711032 , 17711035 , 17711037 , 17711038 , 17711039 , 17711040 , 17711044 , 17711045 , 17711047 , 17711048 , 17711050 , 17711051 , 17711053 , 17711054 , 17711055 , 17711056 , 17711059 , 17711062 , 17711063 , 17711064 , 17711065 , 17711066 , 17711068 , 17711073 , 17711074 , 17711075 , 17711077 , 17711089 , 17711090 , 17711093 , 17711095 , 17711102 , 17711103 , 17711116 , 17711118 , 17711120 , 17711125 , 17711126 , 17711127 , 17711128 , 17711131 , 17711132 , 17711134 , 17711135 , 17711136 , 17711138 , 17711140 , 17711141 , 17711143 , 17711144 , 17711145 , 17711163 , 17711164 , 17711166 , 17711168 , 17711174 , 17711188 , 17711190 , 17711191 , 17711193 , 17711194 , 17711198 , 17711199 , 17711201 , 17711202 , 17711203 , 17711204 , 17711206 , 17711213 , 17711214 , 17711215 , 17711220 , 17711223 , 17711230 , 17711234 , 17711242 , 17711244 , 17711246 , 17711287 , 17711297 , 17711308 , 17711316 , 17711323 , 17711332 , 17711345 , 17711352 , 17711353 , 17711417 , 17711418 , 17711419 , 17711420 , 17711421 , 17711423 , 17711424 , 17711425 , 17711426 , 17711427 , 17711430 , 17711432 , 17711433 , 17711435 , 17711436 , 17711437 , 17711438 , 17711440 , 17711441 , 17711444 , 17711445 , 17711449 , 17711450 , 17711451 , 17711455 , 17711461 , 17711464 , 17711465 , 17711466 , 17711470 , 17711472 , 17711473 , 17711474 , 17711475 , 17711477 , 17711478 , 17711479 , 17711482 , 17711483 , 17711484 , 17711485 , 17711491 , 17711497 , 17711498 , 17711499 , 17711500 , 17711501 , 17711503 , 17711504 , 17711507 , 17711509 , 17711510 , 17711511 , 17711513 , 17711515 , 17711516 , 17711517 , 17711520 , 17711521 , 17711523 , 17711524 , 17711526 , 17711527 , 17711529 , 17711530 , 17711536 , 17711537 , 17711543 , 17711545 , 17711551 , 17711554 , 17711555 , 17711572 , 17711578 , 17711581 , 17711601 , 17711603 , 17711617 , 17711626 , 17711654 , 17711669 , 17711704 , 17711716 , 17711719 , 17711721 , 17711727 , 17711757 , 17711764 , 17711765 , 17711779 , 17711780 , 17711788 , 17711790 , 17711801 , 17711805 , 17711811 , 17711817 , 17711829 , 17711830 , 17711835 , 17711861 , 17711863 , 17711864 , 17711865 , 17711866 , 17711868 , 17711878 , 17711883 , 17711900 , 17711921 , 17711947 , 17711948 , 17711952 , 17711954 , 17711963 , 17711971 , 17711974 , 17711975 , 17711981 , 17711983 , 17711987 , 17711990 , 17711994 , 17712000 , 17712002 , 17712004 , 17712006 , 17712007 , 17712008 , 17712011 , 17712013 , 17712027 , 17712028 , 17712032 , 17712066 , 17712074 , 17712076 , 17712077 , 17712078 , 17712082 , 17712085 , 17712087 , 17712088 , 17712090 , 17712091 , 17712092 , 17712098 , 17712101 , 17712106 , 17712109 , 17712112 , 17712113 , 17712114 , 17712115 , 17712117 , 17712118 , 17712119 , 17712124 , 17712125 , 17712126 , 17712131 , 17712132 , 17712133 , 17712134 , 17712136 , 17712138 , 17712139 , 17712140 , 17712141 , 17712142 , 17712143 , 17712144 , 17712145 , 17712146 , 17712152 , 17712153 , 17712155 , 17712156 , 17712157 , 17712158 , 17712159 , 17712160 , 17712163 , 17712167 , 17712168 , 17712170 , 17712171 , 17712173 , 17712175 , 17712176 , 17712177 , 17712178 , 17712179 , 17712180 , 17712181 , 17712182 , 17712183 , 17712185 , 17712191 , 17712194 , 17712195 , 17712196 , 17712197 , 17712198 , 17712199 , 17712200 , 17712201 , 17712202 , 17712203 , 17712204 , 17712205 , 17712206 , 17712208 , 17712209 , 17712210 , 17712211 , 17712230 , 17712233 , 17712281 , 17712282 , 17712292 , 17712313 , 17712315 , 17712357 , 17712364 , 17712381 , 17712384 , 17712434 , 17712437 , 17712438 , 17712440 , 17712441 , 17712443 , 17712446 , 17712447 , 17712457 , 17712458 , 17712459 , 17712466 , 17712471 , 17712472 , 17712474 , 17712481 , 17712482 , 17712483 , 17712485 , 17712486 , 17712493 , 17712494 , 17712500 , 17712507 , 17712508 , 17712509 , 17712511 , 17712512 , 17712513 , 17712514 , 17712515 , 17712519 , 17712522 , 17712545 , 17712548 , 17712579 , 17712598 , 17712601 , 17712610 , 17712615 , 17712632 , 17712633 , 17712637 , 17712638 , 17712640 , 17712641 , 17712642 , 17712643 , 17712645 , 17712646 , 17712648 , 17712649 , 17712650 , 17712651 , 17712652 , 17712654 , 17712655 , 17712656 , 17712657 , 17712660 , 17712661 , 17712662 , 17712663 , 17712664 , 17712665 , 17712666 , 17712667 , 17712668 , 17712669 , 17712670 , 17712671 , 17712675 , 17712681 , 17712682 , 17712686 , 17712687 , 17712689 , 17712692 , 17712693 , 17712694 , 17712695 , 17712697 , 17712715 , 17712732 , 17712745 , 17712758 , 17712795 , 17712818 , 17712826 , 17712830 , 17712856 , 17712857 , 17712858 , 17712867 , 17712869 , 17712870 , 17712871 , 17712872 , 17712873 , 17712875 , 17712876 , 17712878 , 17712879 , 17712880 , 17712882 , 17712883 , 17712884 , 17712888 , 17712891 , 17712892 , 17712894 , 17712895 , 17712896 , 17712897 , 17712900 , 17712901 , 17712902 , 17712903 , 17712904 , 17712906 , 17712908 , 17712909 , 17712910 , 17712911 , 17712914 , 17712917 , 17712918 , 17712919 , 17712920 , 17712921 , 17712922 , 17712923 , 17712925 , 17712929 , 17712931 , 17712932 , 17712933 , 17712934 , 17712935 , 17712936 , 17712937 , 17712938 , 17712939 , 17712940 , 17712942 , 17712943 , 17712944 , 17712946 , 17712947 , 17712948 , 17712949 , 17712956 , 17712972 , 17712974 , 17713013 , 17713014 , 17713026 , 17713043 , 17713051 , 17713064 , 17713069 , 17713099 , 17713143 , 17713145 , 17713147 , 17713209 , 17713210 , 17713212 , 17713214 , 17713216 , 17713218 , 17713224 , 17713226 , 17713230 , 17713237 , 17713238 , 17713242 , 17713246 , 17713247 , 17713248 , 17713250 , 17713251 , 17713255 , 17713256 , 17713262 , 17713274 , 17713283 , 17713285 , 17713287 , 17713289 , 17713293 , 17713295 , 17713296 , 17713297 , 17713300 , 17713305 , 17713306 , 17713307 , 17713310 , 17713311 , 17713312 , 17713315 , 17713319 , 17713340 , 17713345 , 17713346 , 17713360 , 17713361 , 17713363 , 17713366 , 17713367 , 17713370 , 17713389 , 17713415 , 17713439 , 17713440 , 17713442 , 17713444 , 17713447 , 17713458 , 17713463 , 17713465 , 17713466 , 17713467 , 17713472 , 17713480 , 17713494 , 17713512 , 17713513 , 17713514 , 17713519 , 17713520 , 17713522 , 17713523 , 17713524 , 17713529 , 17713536 , 17713538 , 17713539 , 17713542 , 17713543 , 17713547 , 17713555 , 17713557 , 17713562 , 17713565 , 17713568 , 17713573 , 17713575 , 17713577 , 17713579 , 17713583 , 17713584 , 17713585 , 17713586 , 17713594 , 17713598 , 17713600 , 17713601 , 17713603 , 17713605 , 17713607 , 17713678 , 17713681 , 17713683 , 17713684 , 17713685 , 17713686 , 17713687 , 17713688 , 17713689 , 17713691 , 17713692 , 17713693 , 17713694 , 17713695 , 17713696 , 17713697 , 17713698 , 17713700 , 17713701 , 17713704 , 17713705 , 17713706 , 17713707 , 17713708 , 17713709 , 17713710 , 17713711 , 17713712 , 17713714 , 17713716 , 17713719 , 17713720 , 17713721 , 17713722 , 17713723 , 17713724 , 17713726 , 17713727 , 17713728 , 17713731 , 17713732 , 17713733 , 17713734 , 17713735 , 17713737 , 17713738 , 17713739 , 17713740 , 17713742 , 17713743 , 17713744 , 17713745 , 17713746 , 17713747 , 17713759 , 17713768 , 17713769 , 17713775 , 17713776 , 17713777 , 17713779 , 17713781 , 17713787 , 17713795 , 17713845 , 17713847 , 17713851 , 17713853 , 17713859 , 17713862 , 17713863 , 17713864 , 17713866 , 17713868 , 17713869 , 17713870 , 17713888 , 17713896 , 17713897 , 17713905 , 17713909 , 17713913 , 17713915 , 17713916 , 17713918 , 17713919 , 17713920 , 17713931 , 17713933 , 17713934 , 17713937 , 17713940 , 17713946 , 17713948 , 17713949 , 17713950 , 17713951 , 17713952 , 17713954 , 17713961 , 17713962 , 17713965 , 17713974 , 17713980 , 17713982 , 17713987 , 17713989 , 17713991 , 17713992 , 17713994 , 17713999 , 17714012 , 17714013 , 17714015 , 17714018 , 17714019 , 17714020 , 17714031 , 17714032 , 17714035 , 17714036 , 17714039 , 17714048 , 17714062 , 17714063 , 17714066 , 17714070 , 17714076 , 17714077 , 17714078 , 17714080 , 17714106 , 17714115 , 17714117 , 17714119 , 17714122 , 17714127 , 17714128 , 17714130 , 17714134 , 17714137 , 17714139 , 17714142 , 17714145 , 17714150 , 17714154 , 17714157 , 17714180 , 17714185 , 17714187 , 17714189 , 17714190 , 17714195 , 17714205 , 17714219 , 17714221 , 17714225 , 17714237 , 17714238 , 17714249 , 17714250 , 17714251 , 17714261 , 17714263 , 17714269 , 17714273 , 17714278 , 17714284 , 17714286 , 17714288 , 17714289 , 17714293 , 17714295 , 17714298 , 17714303 , 17714307 , 17714313 , 17714321 , 17714324 , 17714328 , 17714332 , 17714343 , 17714344 , 17714345 , 17714355 , 17714357 , 17714360 , 17714363 , 17714365 , 17714366 , 17714373 , 17714374 , 17714375 , 17714378 , 17714379 , 17714381 , 17714382 , 17714383 , 17714384 , 17714386 , 17714391 , 17714396 , 17714399 , 17714402 , 17714408 , 17714411 , 17714418 , 17714425 , 17714428 , 17714435 , 17714437 , 17714440 , 17714441 , 17714443 , 17714445 , 17714448 , 17714449 , 17714450 , 17714456 , 17714460 , 17714462 , 17714463 , 17714465 , 17714466 , 17714471 , 17714482 , 17714485 , 17714489 , 17714495 , 17714509 , 17714510 , 17714513 , 17714519 , 17714520 , 17714558 , 17714569 , 17714571 , 17714576 , 17714588 , 17714589 , 17714591 , 17714604 , 17714606 , 17714609 , 17714614 , 17714620 , 17714624 , 17714627 , 17714632 , 17714644 , 17714647 , 17714661 , 17714707 , 17714777 , 17714869 , 17714887 , 17714889 , 17714891 , 17714892 , 17714893 , 17714900 , 17714905 , 17714924 , 17714933 , 17714938 , 17714946 , 17714956 , 17714957 , 17714960 , 17714961 , 17714963 , 17714964 , 17714970 , 17714973 , 17714978 , 17714980 , 17714983 , 17714990 , 17714992 , 17714993 , 17714994 , 17714995 , 17714998 , 17714999 , 17715000 , 17715001 , 17715002 , 17715006 , 17715007 , 17715009 , 17715011 , 17715014 , 17715019 , 17715021 , 17715023 , 17715024 , 17715026 , 17715030 , 17715035 , 17715037 , 17715051 , 17715052 , 17715060 , 17715062 , 17715063 , 17715065 , 17715072 , 17715077 , 17715171 , 17715188 , 17715418 , 17715421 , 17715427 , 17715428 , 17715432 , 17715434 , 17715437 , 17715446 , 17715465 , 17715469 , 17715474 , 17715485 , 17715487 , 17715488 , 17715489 , 17715491 , 17715492 , 17715493 , 17715495 , 17715496 , 17715497 , 17715498 , 17715499 , 17715503 , 17715504 , 17715509 , 17715511 , 17715514 , 17715517 , 17715520 , 17715522 , 17715537 , 17715540 , 17715547 , 17715551 , 17715555 , 17715557 , 17715558 , 17715565 , 17715566 , 17715572 , 17715573 , 17715574 , 17715576 , 17715577 , 17715580 , 17715586 , 17715601 , 17715621 , 17715632 , 17715634 , 17715639 , 17715644 , 17715645 , 17715646 , 17715653 , 17715654 , 17715658 , 17715662 , 17715664 , 17715666 , 17715668 , 17715671 , 17715675 , 17715677 , 17715678 , 17715681 , 17715684 , 17715687 , 17715688 , 17715693 , 17715699 , 17715701 , 17715702 , 17715704 , 17715705 , 17715711 , 17715717 , 17715722 , 17715724 , 17715727 , 17715728 , 17715738 , 17715739 , 17715747 , 17715750 , 17715763 , 17715765 , 17715770 , 17715772 , 17715773 , 17715780 , 17715784 , 17715785 , 17715788 , 17715789 , 17715790 , 17715797 , 17715801 , 17715802 , 17715803 , 17715807 , 17715808 , 17715812 , 17715822 , 17715824 , 17715832 , 17715833 , 17715835 , 17715836 , 17715841 , 17715851 , 17715852 , 17715857 , 17715859 , 17715864 , 17715874 , 17715875 , 17715887 , 17715888 , 17715899 , 17715903 , 17715909 , 17715920 , 17715922 , 17716554 , 17716556 , 17716566 , 17716567 , 17716588 , 17716595 , 17716601 , 17716610 , 17716626 , 17716854 , 17716868 , 17718774 , 17726380 , 17726508 , 17729909 , 17669398 , 17671560 , 17671606 , 17676916 , 17677515 , 17679500 , 17685126 , 17687602 , 17693534 , 17696172 , 17697080 , 17709124 , 17709165 , 17709549 , 17709737 , 17709769 , 17709802 , 17709815 , 17709851 , 17709862 , 17709940 , 17709943 , 17710030 , 17710053 , 17710135 , 17710143 , 17710431 , 17710492 , 17710604 , 17710699 , 17710715 , 17710803 , 17710829 , 17710845 , 17710861 , 17710868 , 17710926 , 17710933 , 17710946 , 17711024 , 17711052 , 17711082 , 17711092 , 17711094 , 17711137 , 17711156 , 17711189 , 17711233 , 17711471 , 17711519 , 17711613 , 17711722 , 17711819 , 17711932 , 17712014 , 17712089 , 17712121 , 17712259 , 17712340 , 17712492 , 17712495 , 17712505 , 17712644 , 17712881 , 17712898 , 17713012 , 17713249 , 17713349 , 17713390 , 17713434 , 17713521 , 17713893 , 17713900 , 17714007 , 17714075 , 17714096 , 17714105 , 17714140 , 17714335 , 17714352 , 17714409 , 17714415 , 17714424 , 17714426 , 17714454 , 17714499 , 17714528 , 17714538 , 17714539 , 17714554 , 17714903 , 17714907 , 17714922 , 17714965 , 17715010 , 17715039 , 17715460 , 17715476 , 17715521 , 17715597 , 17715607 , 17715642 , 17715650 , 17715676 , 17715730 , 17715736 , 17715766 , 17715820 , 17715821 , 17715823 , 17715889 , 17716552 , 16065951 , 16066466 , 16076615 , 16076979 , 16077000 , 16077160 , 16077166 , 16077172 , 16077173 , 16077174 , 16077177 , 16077187 , 16077188 , 16077191 , 16078190 , 16078307 , 16078315 , 16078316 , 16078537 , 16078542 , 16078547 , 16078548 , 16078549 , 16078550 , 16078556 , 16078558 , 16078599 , 16078904 , 16078906 , 16079051 , 16079208 , 16079238 , 16079299 , 16079358 , 16079401 , 16079403 , 16079404 , 16079407 , 16079408 , 16079409 , 16079410 , 16079412 , 16079413 , 16079414 , 16079415 , 16079417 , 16079418 , 16079419 , 16079421 , 16079426 , 16079428 , 16079432 , 16079494 , 16079667 , 16079673 , 16079675 , 16079683 , 16079704 , 16079708 , 16079737 , 16079750 , 16079754 , 16079755 , 16079761 , 16079762 , 16079764 , 16079766 , 16079769 , 16080454 , 16080475 , 16080704 , 16080829 , 16081348 , 16081355 , 16081356 , 16081358 , 16081360 , 16081365 , 16081371 , 16081374 , 16081376 , 16081378 , 16082275 , 16082405 , 16082461 , 16082492 , 16082536 , 16083296 , 16083387 , 16083398 , 16083526 , 16083541 , 16083547 , 16083994 , 16084010 , 16084015 , 16084052 , 16084076 , 16084081 , 16084084 , 16084108 , 16084112 , 16084118 , 16084127 , 16084130 , 16084133 , 16084141 , 16084168 , 16084184 , 16084187 , 16061037 , 16061394 , 16061451 , 16062665 , 16063264 , 16063290 , 16065853 , 16065891 , 16066282 , 16066645 , 16066665 , 16067356 , 16067986 , 16068174 , 16068272 , 16068798 , 16068817 , 16068882 , 16069007 , 16069688 , 16070828 , 16072878 , 16073611 , 16073744 , 16077200 , 16078106 , 16078159 , 16078370 , 16078543 , 16079353 , 16079416 , 16079658 , 16079711 , 16079751 , 16079763 , 16079765 , 16079773 |
Recalling Firm/ Manufacturer |
Hospira Inc. 275 N Field Dr Lake Forest IL 60045-2579
|
For Additional Information Contact | Hospira Global Complaint Management 800-441-4100 |
Manufacturer Reason for Recall | One lot of alarm assemblies used in Plum A+ and Plum A+3 infusion pumps may fail to sound at all volume levels. |
FDA Determined Cause 2 | Component change control |
Action | Hospira sent an Urgent Medical Device Correction letter dated July 2, 2014, to all affected customers. The letter notified customers that Hospira will be contacting them regarding the completion of an audible alarm test and will replace any alarm assemblies that fail to audible alarm test. Additionally, the letter included instructions for customers to: 1) inform potential users of this notification; 2) complete and return the attached reply form; and, 3) if the products were further distributed, notify those accounts of this recall and ask them to contact Stericycle at 888-912-7350 (Monday - Friday, 8:00 AM - 5:00 PM ET) to receive a reply form.
A second URGENT MEDICAL DEVICE CORRECTION letter dated July 16, 2014 was sent to all customers of record because the initial letters may have contained errors in the attached data files listing the potentially affected devices at the customer's' facility. The instructions to the customer were unchanged from the original letter.
For questions regarding this recall call 800-441-4100. |
Quantity in Commerce | 4,656 infusion pumps |
Distribution | Worldwide Distribution - USA including AK, AL, AR, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and Brazil, Canada, Colombia, Ecuador, France, Israel, Italy, Mexico, Oman, Peru, Saudi Arabia, Spain, Turkey, United Arab Emirates, United Kingdom, and Venezuela. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = FRN
|
|
|
|