• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Laksell Gamma Knife Perfexion

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Laksell Gamma Knife Perfexionsee related information
Date Initiated by FirmOctober 06, 2014
Date PostedOctober 30, 2014
Recall Status1 Terminated 3 on September 15, 2021
Recall NumberZ-0152-2015
Recall Event ID 69425
510(K)NumberK120811 
Product Classification System, radiation therapy, radionuclide - Product Code IWB
ProductLaksell Gamma Knife Perfexion
Code Information Serial No. 6052, 6144, 6088, 6122, 6158, 6005, 6028, 6097, 6153, 6064, 6081, 6103, 6120, 6143, 6145, 6213, 6215, 6076, 6070, 6136, 6002, 6090, 6131, 6063, 6160, 6077, 6018, 6031, 6040, 6105, 6121, 6013, 6026, 6038, 6050, 6089, 6042, 6055, 6058, 6071, 6080, 6087, 6091, 6096, 9098, 6106, 6108, 6109, 6116, 6118, 6125, 6132, 6133, 6142, 6148, 6149, 6161, 6162, 6175, 6181, 6183, 6198, 6205, 6206, 6163, 6169, 6033, 6041, 6046, 6068, 6113, 6019, 6067, 6110, 6104, 6043, 6044, 6062, 6085, 6086, 6101, 6111, 6123, 6126, 6157, 6177, 6178, 6024, 6069, 6094, 6007, 6075, 6182, 6065, 6159, 6199, 6003, 6017, 6027, 6039, 6112, 6146, 6194, 6004, 6006, 6008, 6009, 6010, 6011, 6012, 6014, 6015, 6016, 6020, 6021, 6022, 6023, 6025, 6029, 6030, 6032, 6034, 6035, 6036, 6037, 6045, 6047, 6048, 6053, 6054, 6056, 6057, 6060, 6061, 6066, 6072, 6073, 6074, 6078, 6079, 6082, 6083, 6084, 6092, 6093, 6095, 6099, 6100, 6102, 6107, 6114, 6115, 6117, 6119, 6124, 6127, 6128, 6129, 6130, 6134, 6135, 6139, 6141, 6150, 6151, 6165, 6167, 6171, 6172, 6173, 6176, 6179, 6180, 6184, 6185, 6191, 6192, 6200, 6203, 6207, 6211
FEI Number 1037831
Recalling Firm/
Manufacturer
Elekta, Inc.
400 Perimeter Center Ter Ne
Ste 50
Atlanta GA 30346-1227
For Additional Information ContactLinda Wetsel
770-300-9725
Manufacturer Reason
for Recall
Elekta has become aware that the latches may be locked before they have been fully turned resulting in a poorly locked frame.
FDA Determined
Cause 2
Device Design
ActionElekta sent an" Important Field Safety Notice" distributed on October 6, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. This Important Field Safety Notice describes the problem and possible work arounds for customers to follow until a permanent solution can be developed. They are: " Always assure that the plastic lever is only operated when at a right angle with the frame adapter. " Do not force the plastic lever in place if it meets significant resistance when turned. " Ensure that the plastic lever is completely flush with the frame adapter and that no angulations are present. Corrective Action #2 - Elekta will check and replace as necessary affected frame adapters in the field. This will be released at the end of October 2014, and Elekta will have six months to complete the checks and replacements for those products in the field. If you have any questions please contact your local Elekta office, or call (770) 300-9725.
Quantity in Commerce181
DistributionWorldwide Distribution - USA (nationwide) including the states of AK, AZ, AR, CA, CO, CT, FL, GA, HI, IL, KY, LA, MA, MI, MN, MS, MO, MT, NJ, NM, NY, NC, OH, OR, PA, RI, SC, TX, TN, UT, VA, WA and WV, and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Mexico, Morocco, Netherland, Norway, Poland, Portugal, Russia, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey and United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = IWB
-
-