| Class 2 Device Recall TSX301C Aquilion (Vision edition) | |
Date Initiated by Firm | November 05, 2014 |
Date Posted | November 26, 2014 |
Recall Status1 |
Terminated 3 on October 09, 2015 |
Recall Number | Z-0471-2015 |
Recall Event ID |
69748 |
510(K)Number | K132222 |
Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
Product | Cardiac Function Analysis Software
Product Data No. MPDCT0301EAD
CSCF-003A
The software is applicable to the following CT systems:
Aquilion ONE TSX-301C
X-ray CT system with a 320-row 0.5 mm detector is provided with a 7.5-MHU large-capacity X-ray tube unit, and permits scanning over large areas at the high-speed of 0.35 s per rotation. The system features are; 160-mm-wide area detector, station, gantry and patient couch operation controls, image reconstruction, high-speed volume scan, high-quality images, selectable image slice thickness, dual-monitor system, improved image diagnosis functions, high-speed volume data workflow, exposure reduction, ECG-gated scan and reconstruction, and SUREFluoro.
The CT cardiac function analysis software is used to analyze cardiac function using data obtained by performing cardiac MPR processing for ECG-gated scan data according to the cardiac axis. Aquilion ONE is a dynamic volume CT system that supports whole-body scanning. This 320-detector row system generates 640 slices per rotation using the coneXact reconstruction algorithm. In addition, the high-speed rotation mechanism and the fast reconstruction unit of the system allow the rapid image acquisition to further improve throughput in CT examinations. |
Code Information |
Item CFA Software Serial Number: Description TSX-301C/2D 2BA1362093 Aquilion One Vision |
Recalling Firm/ Manufacturer |
Toshiba American Medical Systems Inc 2441 Michelle Dr P.O. Box 2068 Tustin CA 92780-7047
|
For Additional Information Contact | 714-730-5000 |
Manufacturer Reason for Recall | Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire Heart displayed as analysis results of the CFA and in a Left-Ventricular Volume Curve generated based on some of those parameters. |
FDA Determined Cause 2 | Software design |
Action | Toshiba sent an Urgent Medical Device Correction letter dated November 5, 2014, to all affected customers. Customers were asked to refrain from using the function parameters for the entire heart and the left-ventricular volume curve until corrective actions have been taken.
Customers were instructed to share this information with all users and reviewing radiologist as well as clinical engineering or Biomedical group at facility.
Please complete and return the attached form and fax it to the toll free number at the top of the form 877-349-3054. This form can also be sent via email to raffairs@tams.com.
Revised software to prevent occurrence of this problem will be installed on the system. When the new software becomes available the Toshiba service representative will contact for an appointment to install it on the system.
Customers with questions were instructed to call 800-421-1968 or contact their local Toshiba Representative at 800-521-1968. |
Quantity in Commerce | 1 unit within the U.S. only |
Distribution | US Nationwide Distribution to CA, MA, MD |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JAK
|
|
|
|