| | Class 2 Device Recall ADVIA XPT Chemistry System |  |
| Date Initiated by Firm | February 11, 2015 |
| Date Posted | April 07, 2015 |
| Recall Status1 |
Terminated 3 on March 02, 2017 |
| Recall Number | Z-1399-2015 |
| Recall Event ID |
70591 |
| 510(K)Number | K990346 |
| Product Classification |
Electrode, ion specific, sodium - Product Code JGS
|
| Product | ADVIA Chemistry XPT System, Siemens Material Number 10723034, Software Version 1.0.2, Siemens Material Number 11219493, IVD.
The ADVIA XPT Chemistry System is an automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes at a throughput rate of both 1800 photometric tests per hour and 600 electrolyte (ISE) tests per hour. |
| Code Information |
Software Version 1.0.2; Lot Numbers: CA1275000080008, CA1275000090009, CA1275000060006, CA1275000050005, CA1275000070007, CA1275000130013, CA1275000100010, CA1275000120012, CA1275000170017, CA1275000260026, CA1275000270027, CA1275000280028, CA1275000240024, CA1275000200020, CA1275000220022, CA1275000410041 |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 511 Benedict Ave Tarrytown NY 10591-5005
|
| For Additional Information Contact | 914-631-8000 |
Manufacturer Reason for Recall | Siemens Healthcare Diagnostics has confirmed an issue with ADVIA Chemistry XPT software version 1.0.2. The system may cause samples to remain in an Inprocess state. Test results
on a sample that is held Inprocess will not transmit to the LIS. Manual intervention is necessary to complete the processing of the samples that are held Inprocess. |
FDA Determined Cause 2 | Software design |
| Action | Siemens Healthcare Diagnostics sent Urgent Medical Device Correction Letters (dated 2/11/2015) via FedEx to the domestic consignees informing users of the software issue with ADVIA Chemistry XPT Software Version 1.0.2. The letters included work instructions to handle the bug. Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter within 30 days. For questions about in process samples, the event message 02 981 8728, or need assistance with mitigation solutions, please contact your Siemens Customer Care Center or your local Siemens Technical Support Representative. Urgent Field Safety Notice Letters were sent to the foreign consignees (distribution is determined at the country level). Field service personnel were notified of the Urgent Medical Device Correction Letter. |
| Quantity in Commerce | 16 units: 3 Domestic, 13 Foreign |
| Distribution | Distributed in the states of AZ, CA, and WA, and the countries of Germany, Italy, Spain, and UK. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = JGS
|
|
|
|