• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Triglycerides

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Triglyceridessee related information
Date Initiated by FirmMay 07, 2015
Date PostedJune 27, 2015
Recall Status1 Terminated 3 on March 24, 2017
Recall NumberZ-1929-2015
Recall Event ID 71182
510(K)NumberK921661 
Product Classification Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
ProductTriglycerides GPO. In vitro test for the quantitative determination of triglycerides in human serum and plasma on the COBAS Integra 800 and 400 plus systems, cobas c111, Roche/Hitachi MODULAR systems, and cobas c systems.
Code Information Catalog Numbers: 04657594190; 20767107322; 05171407190; 20767107322; 11730711216; 11876023216; 11876040216; 11488872216.
FEI Number 1823260
Recalling Firm/
Manufacturer
Roche Diagnostics Operations, Inc.
9115 Hague Rd
Indianapolis IN 46256-1025
For Additional Information ContactTodd Siesky
317-521-3911 Ext. 249
Manufacturer Reason
for Recall
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
FDA Determined
Cause 2
Under Investigation by firm
ActionRoche Diagnostics initiated a voluntary recall and the Urgent Medical Device Correction notices were sent via UPS Ground on 05/07/2015 to customers. Customers were notified of the interference and the method sheets and package inserts will be updated with the new interference claims. The firm plans to monitor any non-responding accounts and address those customers with follow-up attempts to distribute recall notices. The firm plans to conduct level B effectiveness checks via fax and phone. Until the new method sheets are available, the Urgent Medical Device Correction serves as the updated labeling. Roche is not requesting return of the affected product. Customers may contact Todd Siesky Corporate Communications Director, Phone: 317.521.3911, Fax: 317.521.4696, E-mail: todd.siesky@roche.com. For technical questions, contact the Customer Support Center at 1-800-428-2336
Quantity in Commerce6255
DistributionUS (nationwide) including PR.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = JJE
-
-