Date Initiated by Firm |
July 10, 2015 |
Date Posted |
September 03, 2015 |
Recall Status1 |
Terminated 3 on April 12, 2017 |
Recall Number |
Z-2672-2015 |
Recall Event ID |
71721 |
Product Classification |
Cesarean section kit - Product Code OHM
|
Product |
CESAREAN PACK 3/CS Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. . |
Code Information |
Code:900403 Lots: 141015707 exp. 10/31/15 141216329 exp. 12/31/15 150116529 exp. 1/31/16 150116530 exp. 1/31/16 150116531 exp. 1/31/16 150116532 exp. 1/31/16 150317799 exp. 3/3/16 150317800 exp. 3/31/16 150317801 exp. 3/31/16 150317802 exp. 4/30/16 150418395 exp. 4/30/16 150418394 exp. 4/30/16 150418396 exp. 4/30/16 150518580 exp. 5/31/16 150518581 exp. 5/31/16 |
Recalling Firm/ Manufacturer |
Customed, Inc Calle Igualdad Final #7 Fajardo PR 00738
|
Manufacturer Reason for Recall |
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
|
FDA Determined Cause 2 |
Packaging |
Action |
Customed issued an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" notification letter dated July 10, 2015 to all affected customers. The letter identified the product, problem, and actions to be taken by the customer. |
Quantity in Commerce |
507 |
Distribution |
Distributed Only in Puerto Rico. |
Total Product Life Cycle |
TPLC Device Report
|
|