  Class 2 Device Recall SKIN STAPLE REMOVER KIT see related information
Date Initiated by Firm July 10, 2015
Date Posted September 03, 2015
Recall Status1 Terminated 3 on April 12, 2017
Recall Number Z-2703-2015
Recall Event ID 71721
Product Classification Staple, removable (skin) - Product Code GDT
Product SKIN STAPLE REMOVER KIT 80/CS
Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.
.
Code Information Code:900782 Lots: 150116966 exp. 1/31/16 150317662 exp. 3/31/16 150116966 exp. 1/31/16 150518842 exp. 5/31/16 
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
Manufacturer Reason
for Recall
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
FDA Determined
Cause 2
Packaging
Action Customed issued an "URGENT VOLUNTARY MEDICAL DEVICE RECALL" notification letter dated July 10, 2015 to all affected customers. The letter identified the product, problem, and actions to be taken by the customer.
Quantity in Commerce 2180
Distribution Distributed Only in Puerto Rico.
Total Product Life Cycle TPLC Device Report