| | Class 2 Device Recall INTEGRIS CV |  |
| Date Initiated by Firm | November 03, 2015 |
| Date Posted | January 30, 2016 |
| Recall Status1 |
Terminated 3 on September 15, 2020 |
| Recall Number | Z-0729-2016 |
| Recall Event ID |
72599 |
| 510(K)Number | K002016 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | INTEGRIS CV; Model Number: 722030
The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS Allura 12 & 15 system is intended for peripheral, abdominal, cerebral diagnostic and interventional angiography, neuro applications, cardiac applications and non-vascular interventions |
| Code Information |
All serial numbers. |
| FEI Number |
1218950
|
Recalling Firm/ Manufacturer |
Philips Electronics North America Corporation 3000 Minuteman Rd Andover MA 01810-1032
|
| For Additional Information Contact | 800-722-9377 |
Manufacturer Reason for Recall | The use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images. |
FDA Determined Cause 2 | Device Design |
| Action | The firm, Philips, issued "URGENT-Field Safety Notices" dated 10/26/2015 and 11/19/2015 to its customers. The notices described the product, problem and actions to be taken. The customers were instructed to ensure the audible signal (X-ray buzzer) is always enabled, to activate only ONE fluoroscopy pedal at time, and to configure the footswitch to the default Philips configuration. Customers were also instructed to review the information with all members of your staff who need to be aware; retain a copy with the equipment IFU and complete and return the Confirmation Respond form to Mail Stop 4302, 3000 Minuteman Road, Andover, MA 01810 or email: Debra.Gibbs@Philips.com. Philips is tracking the delivery of these notices via certified return receipt.
If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377. |
| Quantity in Commerce | 13297 in total |
| Distribution | Worldwide Distribution: US (Nationwide) including Puerto Rico and countries of: Albania, Algeria, US Virgin Islands, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Democratic Congo, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = OWB
|
|
|
|