| Date Initiated by Firm | December 23, 2015 |
| Create Date | February 26, 2016 |
| Recall Status1 |
Terminated 3 on October 25, 2016 |
| Recall Number | Z-0919-2016 |
| Recall Event ID |
73024 |
| 510(K)Number | K142297 |
| Product Classification |
Gastrointestinal Tubes with Enteral Specific Connectors - Product Code PIF
|
| Product | EndoVive 3s Low Profile Balloon Kits
Part Number:
M00549310
(XMD P/N 70-0050-122)
The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The EndoVive 3s Low Profile Balloon is indicated for use in adult and pediatric patients who require long-term feeding, are unable to tolerate oral feeding, are at low risk for aspiration or require gastric decompression and/or medication delivery directly into the stomach. |
| Code Information |
Part Number: M00549310 (XMD P/N 70-0050-122) Lot Numbers: 930138 |
Recalling Firm/ Manufacturer |
Xeridiem Mediem Medical Devices Inc 4700 S Overland Dr Tucson AZ 85714-3430
|
| For Additional Information Contact | 520-882-7794 |
Manufacturer Reason for Recall | A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric contents to leak out of the device in some instances. |
FDA Determined Cause 2 | Device Design |
| Action | Xeridiem sent a customer notification letter dated December 23, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer.
Customer was provided the following instructions:
1. Segregate all devices from the above table that in Boston Scientific Corporation's inventory and restrict them from being distributed to end users.
2. Determine the list of end users who have received devices from the above list, notify them of the quality issue, and have these devices returned to Boston Scientific Corporation. Provide Xeridiem Medical Devices with this list of end users.
For any questions the customers is instructed to contact the firm.
For questions regarding this recall call 520-882-7794. |
| Quantity in Commerce | 40 |
| Distribution | Nationwide distribution to MA only. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|