| | Class 2 Device Recall GE DatexOhmeda Aisys CS2 Anesthesia System |  |
| Date Initiated by Firm | February 17, 2016 |
| Create Date | April 13, 2016 |
| Recall Status1 |
Terminated 3 on September 04, 2018 |
| Recall Number | Z-1381-2016 |
| Recall Event ID |
73383 |
| 510(K)Number | K132530 |
| Product Classification |
Gas-machine, anesthesia - Product Code BSZ
|
| Product | GE Healthcare, Aisys CS2 anesthesia.. |
| Code Information |
APWU00723 APWU00724 APWU01057 APWU01058 APWU01059 APWU02682 APWU02683 APWU02684 APWU02685 APWU02686 APWU02687 APWU02688 APWU02689 APWU02690 APWU02691 APWU02692 APWU02693 APWU02694 APWU02695 APWU01240 APWU01241 APWU01242 APWU01243 APWU01244 APWU01785 APWU01827 APWU01988 APWU01989 APWU01990 APWU01991 APWU01992 APWU01993 APWU01994 APWU01995 APWU01996 APWU01997 APWU02708 APWT02996 APWT02384 APWT02254 APWT02255 APWT02256 APWT02257 APWT02258 APWT03126 APWT03127 APWT02326 APWU02707 APWT00391 APWT00392 APWT00393 APWT00394 APWT00395 APWU01576 APWU01577 APWU01578 APWU00363 APWU00364 APWU00365 APWU00366 APWT00898 APWT00899 APWT02417 APWV00144 APWU00707 APWU00708 APWU00709 APWU00710 APWU00711 APWU02287 APWU00190 APWU02166 APWU02167 APWU02168 APWU02169 APWU02170 APWU02171 APWU02172 APWU02316 APWU02317 APWU02318 APWU02319 APWU02320 APWU02321 APWU02322 APWT02043 APWT02044 APWT02045 APWT02046 APWT02047 APWT02048 APWT02049 APWT02050 APWT02051 APWT02052 APWU00712 APWT03125 APWT02250 APWT02251 APWU00505 APWU00506 APWU00507 APWT02328 APWT02329 APWT02330 APWT02331 APWT02332 APWU00474 APWU00475 APWU00476 APWU00477 APWU00478 APWU00479 APWT00589 APWT00590 APWT00591 APWT00592 APWT00593 APWT00603 APWT00977 APWT00978 APWU01486 APWU01487 APWU01488 APWU01489 APWU01490 APWU01491 APWU01492 APWU01493 APWU01494 APWU01495 APWT00561 APWT00562 APWU00611 APWU00612 APWU00613 APWU00614 APWU00145 APWU00146 APWU02163 APWU02164 APWU02165 APWU02051 APWU02052 APWU01825 APWU01909 APWU01599 APWU01600 APWU01601 APWU01602 APWU02017 APWU02018 APWU02019 APWU02020 APWT02407 APWT02408 APWT02409 APWU02576 APWT02385 APWU02237 APWU02648 APWU00297 APWU01054 APWU01055 APWU01056 APWU01608 APWU01609 APWU01610 APWU01611 APWU01612 APWU01613 APWU01614 APWU01615 APWU02066 APWU02067 APWT02750 APWT02746 APWT02747 APWT02748 APWT02749 APWT02751 APWT02752 APWT02851 APWT02852 APWT02853 APWT02854 APWT02855 APWT02856 APWT02857 APWT02858 APWU00576 APWU00579 APWU00580 APWU00581 APWU00582 APWU00583 APWU00584 APWU00585 APWU00586 APWU00587 APWU00588 APWU00589 APWU00590 APWU00591 APWU00592 APWU00593 APWU00594 APWU00595 APWU00596 APWU00597 APWU00598 APWU00599 APWU00600 APWU00601 APWU00602 APWU00603 APWU00728 APWU00729 APWU00730 APWU00751 APWU00752 APWU00753 APWU00754 APWU00755 5404700 APWU00564 APWU00565 APWU00566 APWU00604 APWU00605 APWU00606 APWU02070 APWT00553 APWT00554 APWT00555 APWT00556 APWT00557 APWT00558 APWT00559 APWT00560 APWU00567 APWU01595 APWU02326 APWU01359 APWU01949 APWT02705 APWT02066 APWT02067 APWT02068 APWT02069 APWT02070 APWT02071 APWT02072 APWT02073 APWT02074 APWT02075 APWT02076 APWT02077 APWT02078 APWT02079 APWT02080 APWT02081 APWT02082 APWT02083 APWT02084 APWT02085 APWT02086 APWT02087 APWT02088 APWT02089 APWT02090 APWT02091 APWT02092 APWT02093 APWT02094 APWT02095 APWT02096 APWT02097 APWT02098 APWT02099 APWT02100 APWT02101 APWT02149 APWT02150 APWU02421 APWU02422 APWU02423 APWU02424 APWT02418 APWU02110 APWU02111 APWU00713 APWU01410 APWU01411 APWU01412 APWU00176 APWU00177 APWU00178 APWU02327 APWU02328 APWU00767 APWU00768 APWU00769 APWT02696 APWT02697 APWU01131 APWU01132 APWU01133 APWU00141 APWU00142 APWU02223 APWU02224 APWT02639 APWT02640 APWT02641 APWT02642 APWT02643 APWT02644 APWT02645 APWT02646 APWT03004 APWT03005 APWT03006 APWT03007 APWT03008 APWT03009 APWT03010 APWT03011 APWT03012 APWU02108 APWU00293 APWT02305 APWT02308 APWT02595 APWT02596 APWT02597 APWT02598 APWT02599 APWT02600 APWT02601 APWT02602 APWT02603 APWT02604 APWT02605 APWT02606 APWT02607 APWT02608 APWT02609 APWT02610 APWT02611 APWT02612 APWT02613 APWT02614 APWT02615 APWT02616 APWT02617 APWT02618 APWT02619 APWT02620 APWT02621 APWT02622 APWT02623 APWT02624 APWT02625 APWT02626 APWT02627 APWT02628 APWT02629 APWT03029 APWS00103 APWT02663 APWU01340 APWU01341 APWU01342 APWU01343 APWU01344 APWT02245 APWU00260 APWU01616 APWU01617 APWU01618 APWU01619 APWU01620 APWU01621 APWU01622 APWU01623 APWT00594 APWT00796 APWT00797 APWT00798 APWT00799 APWT00800 APWT00801 5400770 APWU01998 APWT00238 APWT00242 APWT00791 APWT02997 APWT02998 APWT02999 APWT03000 APWT03001 APWT03002 APWT03003 APWU02485 APWU02505 APWU02506 APWT00246 APWU01699 APWU01700 APWU01701 APWU01702 APWU01703 APWU01704 APWU01705 APWU01706 APWU01707 APWU01708 APWU01709 APWU01710 APWU02635 APWU02636 APWU02637 APWU02638 APWU02639 APWU02640 APWU02641 APWU02642 APWU02643 APWU02644 APWU02645 APWU02696 APWU02697 APWU02698 APWU02699 APWU02700 APWU02701 APWT00532 APWT02333 APWT02334 APWT02335 APWT02336 APWT02337 APWT02338 APWT02339 APWT02340 APWT02341 APWT00362 APWT00363 APWT03056 APWT03057 APWT00245 APWT00790 APWT02194 APWT00236 APWT00240 APWT00244 APWT02191 APWT02848 APWT02849 APWU01652 APWU01653 APWU01654 APWU01655 APWU01897 APWU02379 APWU02381 APWU01096 APWU01097 APWU01098 APWU01099 APWU01100 APWU01101 APWU01102 APWU01103 APWU01104 APWU01105 APWU01106 APWU01107 APWU01108 APWU01109 APWU01496 APWU01497 APWU01498 APWU01499 APWU01500 APWU01501 APWU01502 APWU01503 APWU01504 APWU01505 APWU01736 APWU01737 APWT00320 APWT00341 APWT00342 APWT00343 APWT00344 APWT00375 APWT00376 APWU01579 APWU01580 APWU01581 APWU01582 APWU00864 APWU00865 APWU00866 APWU00867 APWU00868 APWU00869 APWU00870 APWU00871 APWU01149 APWU01150 APWV00102 APWV00103 APWT00239 APWU00607 APWU00608 APWU00609 APWU00610 APWU02176 APWU01910 APWU00913 APWU01265 APWU01266 APWT02053 APWU01911 APWU01912 APWU01913 APWU01914 APWU00168 APWU00169 APWU00170 APWU00171 APWU02723 APWT00891 APWT00892 APWT02312 APWT02317 APWT02318 APWT02320 APWT02321 APWU01922 APWU01923 APWU01929 APWU01930 APWU01931 APWU01932 APWU01933 APWU01934 APWU01935 APWU01936 APWU01937 APWU01938 APWT02700 APWT02701 APWT02702 APWT02703 APWU02392 APWU02393 APWU02394 APWU02395 APWU02400 APWU02414 APWU02415 APWU02416 APWU02417 APWU02653 APWU02654 APWU02655 APWU02656 APWU02657 APWU02658 APWU02659 APWU02660 APWU02661 APWU02662 APWU02663 APWU02664 APWU02665 APWU02666 APWU02667 APWU02668 APWU02669 APWU02670 APWU02671 APWU02672 APWU01422 APWU00764 APWU00765 APWU00766 APWU00889 APWU02249 APWU02250 APWU02632 APWU02061 APWT00202 APWT00203 APWT00204 APWT03065 APWT03066 APWT03067 APWT03068 APWT03069 APWT03070 APWT03071 APWT03072 APWT03073 APWT03074 APWT03075 APWT03076 APWT03077 APWT03078 APWU01443 APWU01444 APWU01445 APWU01446 APWU01447 APWU01448 APWU01449 APWU01450 APWU01451 APWU01452 APWU01774 APWU01775 APWT02859 APWT02860 APWT02861 APWT02862 APWT02863 APWT02864 APWT02865 APWT02866 APWT02867 APWT02868 APWT03030 APWT03031 APWT03032 APWT03033 APWT03034 APWT03035 APWT03036 APWT03037 APWT03038 APWT03039 APWT03040 APWT03041 APWT03042 APWT03043 APWT03044 APWT03045 APWT03046 APWT03047 APWT03048 APWT03049 APWT03050 APWT03051 APWT03052 APWT03053 APWT03054 APWT03055 APWT00256 APWT00257 APWT02568 APWU01095 APWU00725 APWU00726 APWU00727 APWU01245 APWT02298 APWT02299 APWT02300 APWT02301 APWT02302 APWT02303 APWT02304 APWU00241 APWU00242 APWU00243 APWU00244 APWU00245 APWT02539 APWU01788 APWU01789 APWU01790 APWU01791 APWU01246 APWU01247 APWU01248 APWU02577 APWU02578 APWU02579 APWU02580 APWU02581 APWU02582 APWU02583 APWU00486 APWU00487 APWU00575 APWU01267 APWT03059 APWT03060 APWT03061 APWT03062 APWT03063 APWT03064 APWU02724 APWU02725 APWU02726 APWU02727 APWU02728 APWU02729 APWU02730 APWU02731 APWU02649 APWU02650 APWT03174 APWT03175 APWU01383 APWU01384 APWT03079 APWT03080 APWT03081 APWT03082 APWT03083 APWT03084 APWT03085 APWT03086 APWT03087 APWT03088 APWT03089 APWT03090 APWT03091 APWT03092 APWT03093 APWT03094 APWT03095 APWT03096 APWT03097 APWT03098 APWT03099 APWT03100 APWT03101 APWT03102 APWT03103 APWT03104 APWT03105 APWT03106 APWT00144 APWT00145 APWT00146 APWT00147 APWT00148 APWT00149 APWT00150 APWT00151 APWT00152 APWT00153 APWT00154 APWT00155 APWT02200 APWT00575 APWT00576 APWT00577 APWT00578 APWT00611 APWT00612 APWT00613 APWT00614 APWT00615 APWT00616 APWT00617 APWT00618 APWT00619 APWT00620 APWT00621 APWT00622 APWT00623 APWT00624 APWT00625 APWT00626 APWT00627 APWT00628 APWT00629 APWT00630 APWT00903 APWT00904 APWT00905 APWT00906 APWT00907 APWT00908 APWT00909 APWU01632 APWU01633 APWU01634 APWU02702 APWU02703 APWU02704 APWU02705 APWU02706 APWU01665 APWU01666 APWU01667 APWU01668 APWU01669 APWU01670 APWU01671 APWU01672 APWU01673 APWU01674 APWU01675 APWU01684 APWU01685 APWU01686 APWU01687 APWU01688 APWU01689 APWU01690 APWU01691 APWU01692 APWU01693 APWU01694 APWU02197 APWT00267 APWT00268 APWT00269 APWT00270 APWT00271 APWT00272 APWT00273 APWT00274 APWT00275 APWT00305 APWT00306 APWT00307 APWT00308 APWT00309 APWT00310 APWT00311 APWT00312 APWT00313 APWT00314 APWT00315 APWT00316 APWT00317 APWT00318 APWT00319 APWU01540 APWU01541 APWU01542 APWU01543 APWU01544 APWU01545 APWU01546 APWU01547 APWU01548 APWU01549 APWU01550 APWU01551 APWU01552 APWU01553 APWU01554 APWU01555 APWU01556 APWU01557 APWU01558 APWU01559 APWU01560 APWU01561 APWU01562 APWU01563 APWU01564 APWU01565 APWU01566 APWU01567 APWU01568 APWU01569 APWU01570 APWU01641 APWT02813 APWT02814 APWT00450 APWT00451 APWT00452 APWT00453 APWT00454 APWT00455 APWT00456 APWT00457 APWT00458 APWT00459 APWT02869 APWT02870 APWT02901 APWT02902 APWT02903 APWT02904 APWT02905 APWT02906 APWT02907 APWT02908 APWT02909 APWT02910 APWT02911 APWT02912 APWT02913 APWT02914 APWT02915 APWT02916 APWT02917 APWT02918 APWT02919 APWT02920 APWT02921 APWT02922 APWT02923 APWT02924 APWU02647 APWU01114 APWU00937 APWU00938 APWU02053 APWU02054 APWU02055 APWU02056 APWU02057 APWU02058 APWU00540 APWU00541 APWU00542 APWU00543 APWU00621 APWU00622 APWU00623 APWU00624 APWU00625 APWU00626 APWU00627 APWU00628 APWU00629 APWU00630 APWU00631 APWU00632 APWU00633 APWU00634 APWU00635 APWU00636 APWU00637 APWU00638 APWU00639 APWU00640 APWU00641 APWU00642 APWU00643 APWU00644 APWU00645 APWU00646 APWU00647 APWU00648 APWU00649 APWU00650 APWU00651 APWU00652 APWU00653 APWU00654 APWU00655 APWU00656 APWU00657 APWU00658 APWU00659 APWU00660 APWU00661 APWU00662 APWU00663 APWU00664 APWU00665 APWU00666 APWU00667 APWU00668 APWU00669 APWU00670 APWU00671 APWU00672 APWU00673 APWU00674 APWU00675 APWU00676 APWU00677 APWU00678 APWT03013 APWT03014 APWT03015 APWT03016 APWT03017 APWT03018 APWT03019 APWT03020 APWT03021 APWT03022 APWT03023 APWT03024 APWT03025 APWT03026 APWT03027 APWT03028 APWT02246 APWT02247 APWV00138 APWV00139 APWU00573 APWU02068 APWU00250 APWU00252 APWU01345 APWT01007 APWT01008 APWS00175 APWS00176 APWS00177 APWS00178 APWS00179 APWS00180 APWS00181 APWS00182 APWS00183 APWS00184 APWS00185 APWS00186 APWS00187 APWS00188 APWU00274 APWU00275 APWU00276 APWU00277 APWU00278 APWU00279 APWU00280 APWU00281 APWU00282 APWU00283 APWU00284 APWU00285 APWU00286 APWU00287 APWU00288 APWU00289 APWU00290 APWU00291 APWU00292 APWU00106 APWU00107 APWU00108 APWU00109 APWU01360 APWU01361 APWU01362 APWU01363 APWU01364 APWU01365 APWU01366 APWU01367 APWU01368 APWU01369 APWU01370 APWU01088 APWU01089 APWU01090 APWU01346 APWU01347 APWU00114 APWU00485 APWU01426 APWU02589 APWU02590 APWU02591 APWU02592 APWU02593 APWU02594 APWU02595 APWU02596 APWU02597 APWU02598 APWU02297 APWU02298 APWU02299 APWU02300 APWU02301 APWU02302 APWU02303 APWU02304 APWT02419 APWU01794 APWU01795 APWU01796 APWU01826 APWU01348 APWU01349 APWU01350 APWU01351 APWU01352 APWU01353 APWU01354 APWU01355 APWU01356 APWU01357 APWU01358 APWU01898 APWU01899 APWU01900 APWU01901 APWU01902 APWU01903 APWU01296 APWU01297 APWU01298 APWU01299 APWU01300 APWU01301 APWU01302 APWU01303 APWU01304 APWU01305 APWU01306 APWU01307 APWU01308 APWU01309 APWU01310 APWU01311 APWU01312 APWU01313 APWU01314 APWU01315 APWU01316 APWU01317 APWU01318 APWU01319 APWU01320 APWU01321 APWU01322 APWU01323 APWU01324 APWU01325 APWU01326 APWU01327 APWU01328 APWU01329 APWU01330 APWU01331 APWU01332 APWU01333 APWU01334 APWU01335 APWU01336 APWU01337 APWU01338 APWT02540 APWT02541 APWT02542 APWT02543 APWT00588 APWT02664 APWT02666 APWT02667 APWT02669 APWT02670 APWT02671 APWT02672 APWT00247 APWT00793 APWT02665 APWT02668 APWU02175 APWU00208 APWU00529 APWU00530 APWU00531 APWU00532 APWU00533 APWT02892 APWU01083 APWT00969 APWT00970 APWT00971 APWT00972 APWT00973 APWT00974 APWT00975 APWT00976 APWT02535 APWT00794 APWT00985 APWT00986 APWT00987 APWT00988 APWT00989 APWT00990 APWV00114 APWV00115 APWV00116 APWV00117 APWV00118 APWV00119 APWV00120 APWV00121 APWV00122 APWT00961 APWT00962 APWT00963 APWT00964 APWT00965 APWT00966 APWT00967 APWU01155 APWU01156 APWU01196 APWU01197 APWU01198 APWU01199 APWU01200 APWU01201 APWU01202 APWU01203 APWU01204 APWU01205 APWT01037 APWU00756 APWU00757 APWU00758 APWU00759 APWU00760 APWU00761 APWU00762 APWU00763 APWU01161 APWU01162 APWU01163 APWU01249 APWU02115 APWU02116 APWU02117 APWU02118 APWU02119 APWU02120 APWU02121 APWU02122 APWU02123 APWU02124 APWU02125 APWU02126 APWU02127 APWU02128 APWU02129 APWU02130 APWU02131 APWU02132 APWU02133 APWU02134 APWU02135 APWU02136 APWU02137 APWU02138 APWU02139 APWU02140 APWU02141 APWU02142 APWU02143 APWU01681 APWU01682 APWU01683 APWT00604 APWT00605 APWT00606 APWT00607 APWU01924 APWU01925 APWU01926 APWU01927 APWU02104 APWU01168 APWU01169 APWU01170 APWU01171 APWU01172 APWU01173 APWU01174 APWU01175 APWU01176 APWU01177 APWU01178 APWU01179 APWU01180 APWU01181 APWU01182 APWU01183 APWU01184 APWU01185 APWU01186 APWU01187 APWU01188 APWU01189 APWU01190 APWU00534 APWU00535 APWU00536 APWU00537 APWU01939 APWU01940 APWU01941 APWU01942 APWU01943 APWU01781 APWT02884 APWT02885 APWT02886 APWT02887 APWT02888 APWT02889 APWT02890 APWT02891 APWU01782 APWU01783 APWU02105 APWU02106 APWU02107 APWU01571 APWU01572 APWU01573 APWU01574 APWU01575 APWU01584 APWU01585 APWU01586 APWU01587 APWU01885 APWU01886 APWU02177 APWU02178 APWU02179 APWU02180 APWT02011 APWU00121 APWU01676 APWU02389 APWU02390 APWU02391 APWU01828 APWU01829 APWT00250 APWU00509 APWU00510 APWU00511 APWT02517 APWT02518 APWT02519 APWT02520 APWT02521 APWU00568 APWU00569 APWU00570 APWU00571 APWU00572 APWT00787 APWT00788 APWT02504 APWT02505 APWU00577 APWU00578 APWT02437 APWU00512 APWT02253 APWT02192 APWT02778 APWT02779 APWT02780 APWT02781 APWT02782 APWT02783 APWT02784 APWT02785 APWT02786 APWT02787 APWT02788 APWT02789 APWT03128 APWT03129 APWT00243 APWU01157 APWU01158 APWU01159 APWU01160 APWU01418 APWT00253 APWT00254 APWT02722 APWT02723 APWT02724 APWT02410 APWT02411 APWT02412 APWT02413 APWT02414 APWT02415 APWU01797 APWU01798 APWT03130 APWT03131 APWT03132 APWT03133 APWT03134 APWT03135 APWU02709 APWU02710 APWU02711 APWU02712 APWU02713 APWU02714 APWU02715 APWU02716 APWU01638 APWU01639 APWU01640 APWT02406 APWT02704 APWU01520 APWU01521 APWU01522 APWU01523 APWU00574 APWU01191 APWU01192 APWU01529 APWU01999 APWU02000 APWU02001 APWU02002 APWU02003 APWU02004 APWU02005 APWU02006 APWU02007 APWU02008 APWU02009 APWU02010 APWU02011 APWU02012 APWU02013 APWU02014 APWU02015 APWU02584 APWT00817 APWT00818 APWT00819 APWT00820
APWT02189 APWU01624 APWU01625 APWU01626 APWU01627 APWU01628 APWU01629 APWU01630 APWU01631 APWU02064 APWU02065 APWU02633 APWU02634 APWT00821 APWT00822 APWT00823 APWT00824 APWT00825 APWT00826 APWT00827 APWT00828 APWT00829 APWT00830 APWT00831 APWT00832 APWT00833 APWT00834 APWT00835 APWT00836 APWU01421 APWU00149 APWU00150 APWU00151 APWU00152 APWU00153 APWU00154 APWU00155 APWU00156 APWU00157 APWT00472 APWT00473 APWT00597 APWT00598 APWT02708 APWT02709 APWT02710 APWT02711 APWT02712 APWT02995 APWU00990 APWU00991 APWU00992 APWU00993 APWU00994 APWU00995 APWU00996 APWU00997 APWU00998 APWU00999 APWU02305 APWU02306 APWT01038 APWU00139 APWU00714 APWU00715 APWU00716 APWU00717 APWU00718 APWU01193 APWU01194 APWU01195 APWT03136 APWT03137 APWT03138 APWT03139 APWT03140 APWT03141 APWT03142 APWT03143 APWT03144 APWT03145 APWT03146 APWT03147 APWT03148 APWT03149 APWT03150 APWU00294 APWU01413 APWU01414 APWU01415 APWT02462 APWT02463 APWT02464 APWT02465 APWT02466 APWT02467 APWT02468 APWT02259 APWT02260 APWT02261 APWT02262 APWT02263 APWT02264 APWT02265 APWT02266 APWT02267 APWT02268 APWT02269 APWT02270 APWT02271 APWT02272 APWT02273 APWT02274 APWT02275 APWT02285 APWT02286 APWT02287 APWT02288 APWT02289 APWT02290 APWT02306 APWT02307 APWT02309 APWT02310 APWT02311 APWT02343 APWT02344 APWT02345 APWT02347 APWT02348 APWT02349 APWT02350 APWT02351 APWT02352 APWT02353 APWT02354 APWT02355 APWT02356 APWT02357 APWT02358 APWT02359 APWT02360 APWT02361 APWT02362 APWT02363 APWT02364 APWT02365 APWT02366 APWT02367 APWT02368 APWT02369 APWT02370 APWT02371 APWT02372 APWT02373 APWT02374 APWT02375 APWT02376 APWT02377 APWT02378 APWT02379 APWT02380 APWT02381 APWT02382 APWT02420 APWT02421 APWT02422 APWT02423 APWT02424 APWT02425 APWT02426 APWT02427 APWT02428 APWT02429 APWT02430 APWT02438 APWT02439 APWT02440 APWT02441 APWT02442 APWT02443 APWT02444 APWT02445 APWT02447 APWT02448 APWT02449 APWT02450 APWT02451 APWT02452 APWT02453 APWT02454 APWT02455 APWT02456 APWT02457 APWT02458 APWT02459 APWT02460 APWT02461 APWT02576 APWT02577 APWT02578 APWT02579 APWT02580 APWT02581 APWT02582 APWT02583 APWT02584 APWT02585 APWT02586 APWT02587 APWT02588 APWT02589 APWT02590 APWT02591 APWT02592 APWT02593 APWT02594 APWT02674 APWT02675 APWT02676 APWT02677 APWT02727 APWT02947 APWT02948 APWT02949 APWT02950 APWT02951 APWT02952 APWT02953 APWT02954 APWT02955 APWT02956 APWU00508 APWT02383 APWU02396 APWU02397 APWU02398 APWU02399 APWU02457 APWU02458 APWU02459 APWU02425 APWU02426 APWU02427 APWT00601 APWT00602 APWT01009 APWT01010 APWT01011 APWT03058 APWU01153 APWU01154 APWU01915 APWU01916 APWU01917 APWU00100 APWU00101 APWU00102 APWU00103 APWU00104 APWU02329 APWU02604 APWU02605 APWU02606 APWU02607 APWU02608 APWU02609 APWU02610 APWT02446 APWU02428 APWU02429 APWU02430 APWU02431 APWU02514 APWU02515 APWU02516 APWU02517 APWU02518 APWU02519 APWU02520 APWU02521 APWU00973 APWU00975 APWU00976 APWU00980 APWU02611 APWU02612 APWU02613 APWU02614 APWU02615 APWU02616 APWU02617 APWU02618 APWU02619 APWU02620 APWU02621 APWU02622 APWU02623 APWU02624 APWU02625 APWU02626 APWU02627 APWU02628 APWU02629 APWU02630 APWU02631 APWU02680 APWU02681 APWU00480 APWU00481 APWU00482 APWU00483 APWU00484 APWU01950 APWU01951 APWU00352 APWT02514 APWT02515 APWT02516 APWU02112 5400321-1 5400321-10 5400321-2 5400321-3 5400321-4 5400321-5 5400321-6 5400321-7 5400321-8 5400321-9 APWU00972 APWU00974 APWU00977 APWU00978 APWU00979 APWU00981 APWU01835 APWU01836 APWU01837 APWU01842 APWU01843 APWT00984 APWU00251 APWT00583 APWT00584 APWT02203 APWT02204 APWT02205 APWT02206 APWT02207 APWT02208 APWU00912 APWU02201 APWU01377 APWU01786 APWT02715 APWT02716 APWU00513 APWU00514 APWU00558 APWT02017 APWT02018 APWU00143 APWU00144 APWU01635 APWU01636 APWT00531 APWU00246 5408943-1 5408943-2 5408943-3 5408943-4 5408943-5 APWS00106 APWS00107 APWS00108 APWS00109 APWS00110 APWS00111 APWS00112 APWS00113 APWS00114 APWS00115 APWU01838 APWU01839 APWU01840 APWU01841 APWU01844 APWU02109 APWU02374 APWU02375 APWU02585 APWU02586 APWU02587 APWU02588 APWV00171 APWV00172 APWV00190 APWV00191 APWV00192 APWV00193 APWV00194 APWV00195 APWV00196 APWV00197 APWV00198 APWV00199 APWV00200 APWV00201 APWV00202 APWV00203 APWV00204 APWV00205 APWV00207 APWV00208 APWV00209 APWV00210 APWV00218 APWV00219 APWV00220 APWV00221 APWV00222 APWV00223 APWV00224 APWV00234 APWV00235 APWV00236 APWV00237 APWV00238 APWV00239 APWV00240 APWV00241 APWV00242 APWV00243 APWV00244 APWV00245 APWV00246 APWV00247 APWV00248 APWV00249 APWV00250 APWV00252 APWV00253 APWV00255 APWV00256 APWV00259 APWV00260 APWV00261 APWV00264 APWV00265 APWV00284 APWV00285 APWV00286 APWV00295 APWV00296 APWV00297 APWV00298 APWV00299 APWV00300 APWV00301 APWV00302 APWV00303 APWV00304 APWV00305 APWV00306 APWV00307 APWV00308 APWV00314 APWV00315 APWV00316 APWV00317 APWV00318 APWV00319 APWV00322 APWV00323 APWV00325 APWV00327 APWV00328 APWV00329 APWV00331 APWV00332 APWV00333 APWV00334 APWV00335 APWV00336 APWV00337 APWV00338 APWV00339 APWV00340 APWV00342 APWV00343 APWV00344 APWV00353 APWV00355 APWU01000 APWU01001 APWU01002 APWU01003 APWU01004 APWU01005 APWU01006 APWU01007 APWU01008 APWU01009 APWU01010 APWU01011 APWU01012 APWU01013 APWU01014 APWU01015 APWU01016 APWU01017 APWU01018 APWU01019 APWU01020 APWU01021 APWS00101 APWS00102 APWS00104 APWS00105 APWT02850 APWT02196 APWT02197 5405278 5402641-1 5402641-2 5402641-3 5402641-4 5404412-1 5404412-2 5404412-3 5404412-4 5404412-5 APWT02190 APWU00325 APWU00402 APWU00403 APWU01371 APWU01372 APWU01373 APWU01374 APWU01375 APWU01376 APWU01918 APWU01919 APWU01920 APWU01921 APWU01642 APWU01643 APWU01644 APWU01645 APWU01646 APWU01647 APWU01648 APWU01649 APWU02200 APWT00237 APWT00241 APWT00792 APWT02122 APWT02123 APWT02124 APWT02125 APWT02126 APWT02127 APWT02193 APWT02195 APWT00585 APWT00586 APWU00722 APWT02102 APWU00130 APWU00131 APWU00132 APWU00133 APWU00134 APWU00135 APWU01435 APWU01436 APWU01437 APWU01438 APWU01439 APWU01440 APWU01441 APWU01442 APWU00551 APWU00552 APWU00553 APWU00554 APWU00555 APWU00556 APWU00557 APWU00825 APWU01135 APWU01136 APWU01137 APWU01138 APWU01139 APWU01140 APWU01141 APWU01142 APWU02323 APWU02324 APWS00236 APWS00415 APWS00416 APWS00417 APWS00418 APWS00419 APWS00420 APWS00421 APWS00452 APWT02297 APWU02030 APWU02031 APWU02072 APWU02073 APWU02074 APWU02075 APWU02076 APWU02077 APWU02078 APWU02079 APWU02080 APWU02092 APWU02093 APWU02094 APWU02095 APWU02096 APWU02097 APWU02113 APWU02159 APWU02160 APWU02161 APWU02162 APWU02246 APWU02288 APWU02290 APWU02332 APWU02410 APWU02411 APWU02412 APWU02413 APWU02501 APWU02651 APWU02652 APWU02717 APWV00134 APWV00135 APWV00137 APWV00159 APWU01592 APWU01779 APWU01780 APWU01952 APWU01953 APWU01954 APWU01955 APWU01956 APWU01957 APWU01958 APWU01792 APWU01793 APWT02248 APWT02839 APWU01679 APWU00789 APWU01378 APWU01379 APWU01380 APWU01381 APWU01382 APWT00692 APWT00693 APWT00694 APWT00697 APWT00698 APWT00702 APWT00703 APWT00704 APWT00705 APWT00706 APWT00707 APWT00710 APWT00711 APWT00712 APWT00713 APWT00715 APWT00719 APWT00720 APWT00721 APWT02803 APWS00502 APWS00503 APWU00806 APWU00807 APWU00808 APWU00809 APWU00810 APWU00811 APWU00812 APWU00813 APWU00814 APWU00815 APWU00816 APWU00817 APWU01400 APWU01401 APWU01402 APWU01805 APWU01745 APWU01746 APWU01747 APWT01015 APWT01016 APWS00189 APWS00190 APWS00191 APWS00192 APWS00193 APWS00194 APWS00195 APWS00196 APWS00197 APWS00198 APWS00199 APWS00200 APWS00201 APWS00202 APWS00203 APWS00204 APWS00205 APWS00206 APWS00207 APWS00208 APWS00209 APWS00210 APWS00211 APWS00212 APWS00213 APWS00214 APWS00215 APWU01799 APWS00237 APWT00695 APWT00696 APWT00700 APWT00701 APWT00708 APWT00709 APWT00714 APWT00716 APWT00718 APWT00779 APWT00780 APWT00781 APWT00782 APWT00783 APWT00784 APWT02107 APWT02108 APWT02109 APWT02110 APWT02111 APWT02112 APWT02113 APWT02114 APWT02199 APWT02820 APWU00115 APWU00116 APWU00117 APWS00373 APWS00374 APWS00375 APWS00376 APWS00377 APWU00884 APWU00122 APWU00123 APWU00124 APWU00125 APWT02404 APWT02531 APWT02532 APWT02533 APWT02534 APWT00785 APWT00786 APWU01419 APWS00117 APWS00453 APWS00461 APWU00925 APWU00926 APWS00458 APWS00460 APWT00885 APWT02798 APWT02405 APWU00180 APWU00172 APWU00173 APWU00174 APWT00844 APWS00232 APWU00488 APWU00489 APWU00490 APWU00491 APWU00492 APWU00493 APWU00494 APWU00495 APWU00496 APWU00313 APWT02662 APWT02403 APWT02714 APWT02826 APWT02818 APWT02819 APWT00896 APWT00897 APWT00421 APWT00422 APWT00423 APWT00424 APWT00425 APWT00426 APWT00427 APWT00428 APWT00429 APWT00811 APWU00989 APWT00771 APWT02673 APWS00455 APWU00799 APWU00800 APWT00124 APWT00125 APWT00126 APWT00127 APWT00128 APWT00129 APWT02399 APWT02400 APWT02401 APWT02402 APWS00454 APWS00457 APWS00459 APWS00456 APWS00485 APWT00369 APWT00812 APWT00813 APWT00814 APWT02001 APWT02002 APWT02003 APWT02004 APWT02005 APWT02006 APWT02007 APWT02008 APWT01017 APWU00914 APWS00407 APWS00408 APWS00409 APWS00410 APWS00411 APWS00412 APWS00413 APWT00579 APWT02831 APWT02832 APWT02833 APWT02834 APWU00898 APWT00770 APWT00769 APWU00826 APWT02063 APWT02064 APWT02065 APWS00167 APWS00168 APWS00169 APWS00170 APWS00171 APWS00172 APWS00173 APWS00174 APWT00699 APWT00717 APWS00233 APWT03117 APWU00822 APWU00823 APWU00824 APWU00872 APWU00873 APWU00874 APWT02680 APWT02836 APWT02837 APWT02838 APWT01013 APWT01014 APWU01807 APWU01808 APWT02828 APWS00378 APWU00915 APWS00379 APWS00380 APWU00836 APWT00110 APWT00111 APWT00112 APWT00113 APWT00114 APWT00115 APWT00116 APWT00117 APWT00118 APWU01111 APWT00364 APWT00365 APWT00366 APWT00367 APWT00368 APWT02827 APWU00370 APWU00371 APWS00433 APWS00434 APWS00435 APWS00436 APWS00437 APWU00307 APWS00430 APWS00431 APWS00432 APWU01110 APWU01112 APWT00435 APWT00749 APWT02234 APWT02235 APWT02236 APWS00531 APWT00134 APWS00422 APWU01887 APWU01888 APWU01889 APWS00155 APWS00156 APWS00157 APWS00216 APWS00217 APWS00251 APWS00252 APWS00256 APWS00257 APWS00258 APWS00259 APWS00260 APWS00446 APWS00552 APWT00183 APWT00199 APWT00200 APWT00201 APWT00251 APWT00252 APWT00255 APWT00742 APWT00743 APWT00927 APWT00928 APWT02033 APWT02034 APWT02035 APWT02036 APWT02037 APWT02038 APWT02039 APWT02040 APWT02041 APWT02042 APWT02506 APWT02507 APWT02508 APWT02509 APWT02510 APWT02511 APWT02513 APWT02530 APWT02879 APWT02880 APWT02935 APWU00317 APWU00318 APWU00783 APWU00784 APWU00785 APWU01117 APWU01118 APWU01119 APWU01120 APWU01121 APWU01122 APWT00845 APWS00138 APWS00140 APWS00218 APWS00219 APWS00220 APWS00221 APWS00222 APWS00223 APWS00266 APWS00267 APWS00316 APWS00317 APWS00318 APWS00319 APWS00320 APWS00321 APWS00322 APWS00543 APWT00213 APWT00214 APWT00215 APWT00216 APWT00217 APWT00218 APWT00219 APWT00220 APWT00221 APWT00222 APWT00223 APWT00503 APWT00504 APWT00505 APWT00506 APWT00507 APWT00508 APWT00509 APWT00510 APWT00511 APWT00512 APWT00530 APWT00595 APWT00837 APWT00838 APWT00839 APWT00840 APWT00841 APWT00842 APWT00843 APWT00900 APWT00901 APWT02055 APWT02141 APWT02142 APWT02143 APWT02144 APWT02145 APWT02146 APWT02147 APWT02148 APWT02681 APWU00111 APWU00112 APWU00412 APWU00464 APWU00465 APWU00466 APWU00467 APWU00468 APWU00469 APWU00470 APWU00471 APWU00472 APWU00473
APWU00886 APWU00887 APWU00888 APWU01904 APWT00258 APWT00259 APWT00260 APWT00261 APWT00262 APWT02103 APWU00319 APWU00320 APWU00321 APWU00322 APWU00323 APWU01290 APWU01291 APWU02153 APWU01981 APWU02233 APWU02234 APWU02235 APWU02247 APWU01787 APWU00545 APWU01982 APWU01983 APWU01984 APWU01985 APWU01986 APWU01987 APWU00353 APWU00827 APWT02009 APWT02957 APWT02958 APWT02959 APWT02960 APWT02961 APWT02979 APWT00460 APWT00547 APWT00548 APWT00549 APWT00550 APWT00551 APWT00552 APWT00855 APWT02313 APWT02314 APWT02315 APWT02316 APWT02319 APWT02322 APWT02821 APWT02822 APWT02823 APWT02824 APWT02825 APWU00840 APWU00841 APWU00842 APWU00843 APWU00844 APWU00845 APWU00846 APWU00847 APWT00438 APWT00884 APWU00247 APWT00940 APWU00936 APWS00381 APWS00382 APWS00383 APWT02284 APWT00539 APWT00540 APWT00541 APWT00346 APWT00347 APWT00348 APWT00349 APWT00350 APWT00351 APWT00352 APWU00211 APWU00212 APWU00213 APWU00214 APWU01761 APWU01762 APWU01763 APWU01764 APWU01765 APWU01766 APWU01767 APWU01768 APWU01769 APWU01770 APWU01771 APWU01772 APWU01773 APWT02946 APWT00722 APWT00723 APWT00724 APWT00725 APWT00726 APWT00727 APWT00728 APWT00729 APWT00730 APWT00731 APWT00732 APWT00733 APWT00439 APWT00440 APWT02162 APWU00891 APWU00892 APWT00377 APWU00343 APWU00933 APWU00934 APWU00935 APWS00505 APWS00506 APWS00507 APWS00508 APWS00509 APWS00544 APWS00545 APWS00546 APWS00547 APWS00548 APWS00549 APWS00550 APWT02387 APWT02388 APWT00475 APWT00476 APWT02342 APWU01220 APWU01221 APWU01222 APWU01223 APWU01224 APWU01225 APWU01226 APWU01227 APWU01228 APWU01229 APWU01230 APWU01231 APWU01232 APWU01233 APWU01234 APWU01235 APWU01236 APWU01237 APWU01238 APWU01239 APWU01255 APWU01256 APWU01257 APWU01258 APWU01259 APWU01260 APWU01261 APWU01262 APWU01263 APWU01264 APWU01268 APWU01269 APWU01270 APWU01271 APWU01272 APWU01273 APWU01274 APWU01275 APWU01276 APWU01277 APWU01278 APWU01279 APWU01280 APWU01281 APWU01282 APWU01283 APWU01284 APWU01285 APWU01286 APWU01287 APWU01288 APWU01289 APWT02416 APWU01206 APWU01207 APWU01208 APWU01209 APWU01210 APWU01211 APWU01212 APWU01213 APWU01214 APWU01215 APWU01216 APWU01217 APWU01218 APWU01219 APWU00191 APWU00192 APWU00193 APWU00194 APWU00195 APWU00196 APWU00197 APWU00198 APWT00141 APWT00142 APWT00143 APWT00542 APWT00543 APWT00544 APWT00545 APWT00546 APWU01458 APWU01459 APWU01460 APWU00931 APWU00932 APWS00482 APWU00391 APWU00450 APWU00181 APWU00615 APWU00616 APWU00617 APWU00618 APWU00233 APWU00234 APWU00235 APWU00236 APWU00237 APWU00238 APWU01818 APWU01819 APWU01820 APWU01821 APWU01822 APWU01823 APWU01824 APWT00991 APWT00992 APWT00993 APWT00994 APWT00995 APWU01928 APWS00484 APWU00930 APWU01851 APWU01852 APWU01853 APWU01854 APWU01855 APWU01856 APWU01857 APWU01858 APWU01859 APWU01860 APWU01861 APWU01862 APWU01863 APWU01864 APWU01865 APWU01866 APWU01867 APWU01868 APWU01869 APWU01870 APWU01871 APWU01872 APWU01873 APWU01874 APWU01875 APWU00939 APWU00940 APWU00941 APWU01029 APWU01030 APWU01031 APWU01032 APWU01033 APWU01034 APWU01035 APWU01036 APWU01037 APWU01038 APWU01039 APWU01040 APWU01041 APWU01042 APWU01043 APWU01044 APWU01045 APWU01046 APWU01047 APWU01048 APWU01049 APWU01050 APWU01051 APWU01387 APWU01388 APWU01389 APWU01390 APWU01391 APWU01392 APWU01393 APWU01394 APWU01395 APWU01396 APWU01397 APWU00679 APWU00680 APWU00681 APWU00682 APWU00683 APWU00684 APWU00685 APWU00686 APWU00687 APWU00688 APWU00689 APWU00690 APWU00691 APWU00692 APWU00693 APWU00694 APWU00695 APWU00696 APWU00697 APWU00698 APWU00699 APWU00700 APWU00701 APWU00702 APWU00703 APWU00704 APWU00705 APWU01661 APWU01662 APWU01663 APWU01664 APWT00913 APWT00914 APWT00156 APWT00157 APWU00119 APWU00120 APWT00734 APWT00735 APWT00736 APWT00737 APWT00738 APWT00739 APWT00740 APWT00741 APWT00140 APWT02984 APWT00931 APWT00674 APWT00675 APWT00676 APWT00677 APWT00678 APWT00679 APWT00680 APWT00681 APWT00682 APWT00683 APWT00684 APWU00199 APWU00200 APWU00201 APWU00202 APWU00203 APWU00204 APWU00205 APWU00206 APWU00207 APWT02962 APWT02963 APWT02964 APWT02965 APWT02966 APWT02967 APWT02968 APWT02969 APWT02970 APWT02971 APWT02972 APWT02973 APWT02974 APWT02975 APWT02976 APWU00421 APWU00422 APWU00423 APWU00424 APWU00425 APWU00426 APWU00427 APWU00428 APWU00429 APWU00430 APWU00431 APWU00432 APWU00433 APWU00434 APWU00435 APWU00436 APWU00437 APWU00438 APWU00439 APWU00440 APWU00441 APWU00442 APWU00443 APWU00444 APWU00445 APWU00446 APWU00447 APWT00448 APWU00367 APWU00368 APWU00369 APWU01461 APWU01462 APWU01463 APWU01596 APWU01597 APWU01512 APWT00533 APWT00534 APWT00535 APWT00536 APWT00537 APWT00538 APWT00949 APWT00178 APWT00179 APWT00180 APWT00181 APWT00182 APWT02647 APWT02648 APWT02649 APWT02650 APWT02651 APWT02652 APWT02653 APWT02654 APWT02655 APWT02656 APWS00514 APWT00184 APWT00185 APWT00186 APWT00187 APWU00209 APWT02019 APWT02020 APWT02021 APWT02022 APWT02023 APWT02024 APWT02025 APWT02026 APWT02027 APWT02028 APWT02029 APWT02030 APWU00374 APWU00375 APWU00376 APWU00377 APWU00378 5404025 5407086-1 5407086-2 5407086-3 5408547-1 5408547-2 5409649-1 5409649-2 5409858-1 5409858-2 5409858-3 5410348-1 5410348-2 5410348-3 5410348-4 5410348-5 APWU01850 APWU02229 APWU02230 APWU02231 APWU02388 APWU02511 APWU02512 APWU02573 APWU02574 APWU02575 APWU02599 APWU02600 APWU02601 APWU02602 APWU02603 APWT00353 APWT00354 APWT00355 APWT00356 APWT00357 APWT00358 APWU00908 APWU00909 APWU00910 APWU00911 APWS00438 APWU00405 APWU00406 APWU00407 APWU00408 APWU01711 APWU00344 APWU00345 APWU01511 APWU01534 APWU01678 APWU01729 APWU01730 APWU01731 APWU01749 APWU01750 APWU01778 APWU01881 APWU02198 APWU02199 APWU02236 APWU02307 APWU02308 APWU02315 APWU02347 APWU02362 APWU02363 APWU02364 APWU02365 APWU02366 APWU02367 APWU02368 APWU02369 APWU02370 APWU02371 APWU02406 APWT02128 APWT02129 APWT02130 APWT02131 APWU00126 APWU00848 APWU00849 APWU00850 APWU01776 APWU00127 APWS00414 APWU01892 APWU01893 APWU01894 APWU01895 APWU01896 APWT00748 APWT02159 APWT02202 APWT02660 APWT02713 APWU01533 APWU02037 APWU02038 APWU02380 APWU02478 APWU02479 APWU02480 APWU02562 APWU02563 APWU01148 APWT01034 APWT00599 APWT00600 APWT02657 APWT02470 APWT00445 APWT00968 APWU01429 APWU01430 APWT02536 APWT02537 APWT02538 5404357 APWU02039 APWU02045 APWU02181 APWU02182 APWU02183 APWU02184 APWU02185 APWU02186 APWU02225 APWU02226 APWU02227 APWU02232 APWU02244 APWU02245 APWU02291 APWU02325 APWU02331 APWU02333 APWU02334 APWU02348 APWU02349 APWU02351 APWU02372 APWU02373 APWU02376 APWU02377 APWU02385 APWU02386 APWU02419 APWU02420 APWU02433 APWU02434 APWU02482 APWU02483 APWU02507 APWU02522 APWU02526 APWU02527 APWU02528 APWU02529 APWU02530 APWU02531 APWU02532 APWU02533 APWU02534 APWU02535 APWU02536 APWU02537 APWU02538 APWU02539 APWU02540 APWU02548 APWU02557 APWU02558 APWU02559 APWU02560 APWU02561 APWU02646 APWV00123 APWV00124 APWV00125 APWV00147 APWV00148 APWT02658 APWT02659 APWT02771 APWU00312 APWT00446 APWT00932 APWT02151 APWS00489 APWT00673 APWU01423 APWU01485 APWU01967 APWU01968 APWU01969 APWU01970 APWU01971 APWU01972 APWU01973 APWU01974 APWU01975 APWU01976 APWU01977 APWU01978 APWU01979 APWT01029 APWU01403 APWT00943 APWT00944 APWU01151 APWT00998 APWT00999 APWU00921 APWU00922 APWU00923 APWU00969 APWU00970 APWU00971 APWS00429 APWT01035 APWT01036 APWS00530 APWS00504 APWT00447 APWT02016 APWU01883 APWU01884 APWT00941 APWT00942 APWS00342 APWS00501 APWU02387 APWT02661 APWU00349 APWT02943 APWT02944 APWT02945 APWU01876 APWU01877 APWU01878 APWT02012 APWT02013 APWU00828 APWU00829 APWU01589 APWU01591 APWT00929 APWT00930 APWU01804 APWS00466 APWT00915 APWU00409 APWU00410 APWT00130 APWT00131 APWT00132 APWT02168 APWT02187 APWT00513 APWU00796 APWT00120 APWT00121 APWU00924 APWT02552 APWU00893 APWU00894 APWU01084 APWU01738 APWU01739 APWU01740 APWU01741 APWT03107 APWT00123 APWS00139 APWT02791 APWT02134 APWT02135 APWT02136 APWT02137 APWT02138 APWU00326 APWS00328 APWT02527 APWT02528 APWT02529 APWU00448 APWU00801 APWU00802 APWU00803 APWU00804 APWU00805 APWS00166 APWT00574 APWT03113 APWT02161 APWT02167 APWU01696 APWU01697 APWT02525 APWT02526 APWU00347 APWU00348 APWT02842 APWT00916 APWU00113 APWT02201 APWT02323 APWU01593 APWU01594 APWS00464 APWS00478 APWS00479 APWS00480 APWS00481 APWT02548 APWT02549 APWT02550 APWT02551 APWS00439 APWS00440 APWU01085 APWU01086 APWS00477 APWU01023 APWU01025 APWU01432 APWU01433 APWT02718 APWS00146 APWU01431 APWT00133 APWU01091 APWU01092 APWS00147 APWT00502 APWT00122 APWU00411 APWT00635 APWT00659 APWU01677 APWT01000 APWS00490 APWU01947 APWU01948 APWU01959 APWU01960 APWU01961 APWU01962 APWU01963 APWU01964 APWU01965 APWU01966 APWT02790 APWT00805 APWU00163 APWU00781 APWU00782 APWU00787 APWS00149 APWT00232 APWT00233 APWT01022 APWU01590 APWU01604 APWU01607 APWU02021 APWU02022 APWU02023 APWU02024 APWU02025 APWU02026 APWU02027 APWU02028 APWU02029 APWU02035 APWU02040 APWU02041 APWU02046 APWU02047 APWU02060 APWU02062 APWU02063 APWU02098 APWU02099 APWU02100 APWU02151 APWU02202 APWU02203 APWU02204 APWU02205 APWU02206 APWU02207 APWU02208 APWU02209 APWU02210 APWU02211 APWU02212 APWU02213 APWU02214 APWU02215 APWU02241 APWU02481 APWU02508 APWU02509 APWU02510 APWU02553 APWU02554 APWU02555 APWU02556 APWV00109 APWV00110 APWV00111 APWV00112 APWV00113 APWV00126 APWV00127 APWV00128 APWV00161 APWS00329 APWT00580 APWU00838 APWU00839 APWT00378 APWT02932 APWU01024 APWT02188 APWT02160 APWS00491 APWS00492 APWS00493 APWS00494 APWS00495 APWS00496 APWS00497 APWS00498 APWS00499 APWS00500 APWT00887 APWS00158 APWS00159 APWS00160 APWS00161 APWS00162 APWU00794 APWT00634 APWT00636 APWT00637 APWT00638 APWT00639 APWT00640 APWT00641 APWT00642 APWT00643 APWT00644 APWT00645 APWT00646 APWT00647 APWT00648 APWT00649 APWT00650 APWT00651 APWT00652 APWT00653 APWT00654 APWT00655 APWT00656 APWT00657 APWT00658 APWT00660 APWT00661 APWT00662 APWT00663 APWT00664 APWT00665 APWT00666 APWT00667 APWU00821 APWT03109 APWU00162 APWS00165 APWU00875 APWU00876 APWU00877 APWU00878 APWU00879 APWS00253 APWT02934 APWU00449 APWU00401 APWS00330 APWS00331 APWS00332 APWS00333 APWT02707 APWU01800 APWS00462 APWS00463 APWS00234 APWS00235 APWS00324 APWS00325 APWS00326 APWS00327 APWT02717 APWU01806 APWT02105 APWS00262 APWS00263 APWT03111 APWT03112 APWT00668 APWU01116 APWT00745 APWT00746 APWT00747 APWU01660 APWT02573 APWT02574 APWT02575 APWS00246 APWT00768 APWT02139 APWT02140 APWT02198 APWU00372 APWU00373 APWU00458 APWS00247 APWS00250 APWS00248 APWS00249 APWU00504 APWU00400 APWS00486 APWT03108 APWT02106 APWS00148 APWS00261 APWT00449 APWU01022 APWS00272 APWS00273 APWS00274 APWT02524 APWU01603 APWT02799 APWT02801 APWU01742 APWU01743 APWU01744 APWT02683 APWT02941 APWT02942 APWT00776 APWT00777 APWT00778 APWT02324 APWT02325 APWT00849 APWT00850 APWT00851 APWT00852 APWS00309 APWS00310 APWS00311 APWS00312 APWU00502 APWS00343 APWS00344 APWS00345 APWS00346 APWS00347 APWS00348 APWS00349 APWS00350 APWS00351 APWS00352 APWS00353 APWS00354 APWS00355 APWS00356 APWS00357 APWS00358 APWS00334 APWS00264 APWS00265 APWT02678 APWT02679 APWU01605 APWS00335 APWS00336 APWS00337 APWS00338 APWT00501 APWS00313 APWS00314 APWT00888 APWT00889 APWT02933 APWU00818 APWT00587 APWS00315 APWT00804 APWT02841 APWU00503 APWU01606 APWT00744 APWT02684 APWT02685 APWT02706 APWU01152 APWT02835 APWS00126 APWT02686 APWT02687 APWT02688 APWT02689 APWS00467 APWS00468 APWS00469 APWS00470 APWS00471 APWS00472 APWS00473 APWS00474 APWS00475 APWS00476 APWU00916 APWU00917 APWU00918 APWU00919 APWU00920 APWT00487 APWT00492 APWT00690 APWT00691 APWT00461 APWT00462 APWT00463 APWT00464 APWT00465 APWT00466 APWT00467 APWT00468 APWT00469 APWT00470 APWT00471 APWT00496 APWS00131 APWT00491 APWT02987 APWT02988 APWS00133 APWS00135 APWT00490 APWS00136 APWS00163 APWS00164 APWS00307 APWT00882 APWT00101 APWT00102 APWT00103 APWT00104 APWT00105 APWT00106 APWT00107 APWT00108 APWT00109 APWS00127 APWS00128 APWS00130 APWT02058 APWS00448 APWU00546 APWS00308 APWT00494 APWU00549 APWU00550 APWT01018 APWT01019 APWT01020 APWT01021 APWS00129 APWS00137 APWS00389 APWS00390 APWS00391 APWS00392 APWS00393 APWS00394 APWS00395 APWS00396 APWS00397 APWS00398 APWS00399 APWS00400 APWS00401 APWS00402 APWS00403 APWS00404 APWS00405 APWS00406 APWT00488 APWT00489 APWT00493 APWT00495 APWT02209 APWT02210 APWT02211 APWT02212 APWT02213 APWT02214 APWT02215 APWT02216 APWT02217 APWT02218 APWS00515 APWS00516 APWS00517 APWS00518 APWS00519 APWS00520 APWT02346 APWT00856 APWS00465 APWU00927 APWU00928 APWU00929 APWS00134 APWS00447 APWT02062 APWT02431 APWT02432 APWT02433 APWT02434 APWT02435 APWT02436 APWU01134 APWT00997 APWT02986 APWT03151 APWT03152 APWT03153 APWT03154 APWT03155 APWT03156 APWT03157 APWT00497 APWU01416 APWT02169 APWT02170 APWT02171 APWT02172 APWT02173 APWT02174 APWT02175 APWT02176 APWT02177 APWT02178 APWT02179 APWT02180 APWT02181 APWT02182 APWT02183 APWS00132 APWU00415 APWT02296 APWU01060 APWU02196 APWU02251 APWT00137 APWT02553 APWT02554 APWT02555 APWT02556 APWT02557 APWT02558 APWT02559 APWT02560 APWT02561 APWT02562 APWU00790 APWS00154 APWT00205 APWT00206 APWT00207 APWT00208 APWT00209 APWT00210 APWT00211 APWT00212 APWT00950 APWT00951 APWT00952 APWT00953 APWT00954 APWT00955 APWT00956 APWT00957 APWT00958 APWT00959 APWT00960 APWU00771 APWU00772 APWU00773 APWU00774 APWU00775 APWU00797 APWU01890 APWU01891 APWU00986 APWT02566 APWT02567 APWT02231 APWT02232 APWT02233 APWU00516 APWU00518 APWU00521 APWU00522 APWU00523 APWU00525 APWT00934 APWT00935 APWS00362 APWT00933 APWU00361 APWU00362 APWU01398 APWU01399 APWU02252 APWU02253 APWU02254 APWU02524 APWU02525 APWT02990 APWT02991 APWT02992 APWT02993 APWT02994 APWT00474 APWU00982 APWT00436 APWT00437 APWT00563 APWU01166 APWU01167 APWT02735 APWT02736 APWT02737 APWU00515 APWU00517 APWU00519 APWU00520 APWU00524 APWU00526 APWT01033 APWU01814 APWU01815 APWT00119 APWT02244 APWT02937 APWT02938 APWU00987 APWS00141 APWU01845 APWT02761 APWT02762 APWT02763 APWT02764 APWT02765 APWT02766 APWT02767 APWT02805 APWT02989 APWT03158 APWU00359 APWU00399 APWU00903 APWU00904 APWU01292 APWU01777 APWT02804 APWU00820 APWU01123 APWU01124 APWU01125 APWU01126 APWU01127 APWU01128 APWU00261 APWU00262 APWU00263 APWU00264 APWU01598 APWU02150 APWU02402 APWU02484 APWS00275 APWT00188 APWT00189 APWU01424 APWU01052 APWT00231 APWU01339 APWU01698 APWU00776 APWU00777 APWU00779 APWU00780 APWU00778 APWT02881 APWT02882 APWT02883 APWU00416 APWU00417 APWU00418 APWT00816 APWT02792 APWT02793 APWT02794 APWT02795 APWU00265 APWU00266 APWU00267 APWU00268 APWU00269 APWU00270 APWT02845 APWT00945 APWU00346 APWU00527 APWU00528 APWT00902 APWT02522 APWT02523 APWU00719 APWU00310 APWT02015 APWT02738 APWT03123 APWU00222 APWU00223 APWU00224 APWU00225 APWU00226 APWU00227 APWU00414 APWT00920 APWT00926 APWT02014 APWU00404 APWU00984 APWT02733 APWT02734 APWT00758 APWU00539 APWU00158 APWU00159 APWT00795 APWU00538 APWT02152 APWT02292 APWT02293 APWT02294 APWT02295 APWT00917 APWT00918 APWT00919 APWT00921 APWT00922 APWT00923 APWT00924 APWT00925 APWT01001 APWT01002 APWT02242 APWU00137 APWT00789 APWT00399 APWT00400 APWU00302 APWU00305 APWU00306 APWU01832 APWU02032 APWU02048 APWU02049 APWU02050 APWU02071 APWU02101 APWU02102 APWU02103 APWU02146 APWU02147 APWU02148 APWU02149 APWU02157 APWU02158 APWU02216 APWU02217 APWU02218 APWU02219 APWU02220 APWU02221 APWU02222 APWU02248 APWU02345 APWU02346 APWU02405 APWU02407 APWU02408 APWU02409 APWU02418 APWU02432 APWU02456 APWU02513 APWV00104 APWV00105 APWV00106 APWV00107 APWV00108 APWV00129 APWV00130 APWV00131 APWV00132 APWV00133 APWV00136 APWV00145 APWV00146 APWT02728 APWT02729 APWT02730 APWT02731 APWT02732 APWU00720 APWU00721 APWT00854 APWU00273 APWT00566 APWU01583 APWU00544 APWU01026 APWU01732 APWT00569 APWU00351 APWU00413 APWU01531 APWU01532 APWU01404 APWU01405 APWU01406 APWT00568 APWT02032 APWT00571 APWU01784 APWU01830 APWU01879 APWU01831 APWU01945 APWU01944 APWT00570 APWU01880 APWU01801 APWU01802 APWU01803 APWU01464 APWU01465 APWU01466 APWU01946 APWT00564 APWT02802 APWU00324 APWT00757 APWT02389 APWU00304 APWU00559 APWU00560 APWU00561 APWT02104 APWU00160 APWU00161 APWT03114 APWT02184 APWT02185 APWT02186 APWU00210 APWU00619 APWU00620 APWT00853 APWU01386 APWU00308 APWU00309 APWT03122 APWU00240 APWT00567 APWT00572 APWT03121 APWT00573 APWT02243 APWT00802 APWT00803 APWT02240 APWT02241 APWU01053 APWT00565 APWU00248 APWU00138 APWT02565 APWU00706 APWU00164 APWU00165 APWU00166 APWU00167 APWT02940 APWS00513 APWU01833 APWS00276 APWU02144 APWU02145 APWU00890 APWU01846 APWU01847 APWU01848 APWU01849 APWT00396 APWT00481 APWT00482 APWT00483 APWT00484 APWT00485 APWT00486 APWT00883 APWT02719 APWT02720 APWT02721 APWU00327 APWU00328 APWU00329 APWU00330 APWU00331 APWU00332 APWU00333 APWU00334 APWU00335 APWU00336 APWU00337 APWU00338 APWU00339 APWU00340 APWU00341 APWU00342 APWU01407 APWU01408 APWU01409 APWU01656 APWU01657 APWU01658 APWU01659 APWU02033 APWU02034 APWT00895 APWT02390 APWT02391 APWT02563 APWT02564 APWT02772 APWT02773 APWT02774 APWT02775 APWT02800 APWU00147 APWU00148 APWU01113 APWU01524 APWU01530 APWT02776 APWT02777 APWT02796 APWT02797 APWT02977 APWT02978 APWU02173 APWU02174 APWU02187 APWU02188 APWU02189 APWU02190 APWU02191 APWU02192 APWU02193 APWU02194 APWU02195 APWT02291 APWT00750 APWT00751 APWS00363 APWS00364 APWS00365 APWS00366 APWS00367 APWS00368 APWS00369 APWS00370 APWS00371 APWS00372 APWT00477 APWT00478 APWT00479 APWT00480 APWT02980 APWT02981 APWT02982 APWT02983 APWU00392 APWU00393 APWU00394 APWU00395 APWU00396 APWU00419 APWU00420 APWU00451 APWU00452 APWS00145 APWU00314 APWT01023 APWT00224 APWT00225 APWT00226 APWT00227 APWT00228 APWT00229 APWT00230 APWT00432 APWT00498 APWT00499 APWT00500 APWU02036 APWU02465 APWU02486 APWU02487 APWU02488 APWU02489 APWU02490 APWU02491 APWU02492 APWU02493 APWU02494 APWU02495 APWU02496 APWU02497 APWU02498 APWU02499 APWU02500 APWU00880 APWT00433 APWT00434 APWU00792 APWU00793 APWT00263 APWT00264 APWT00265 APWT00266 APWU00459 APWU00460 APWU00461 APWU00462 APWU00463 APWS00144 APWS00385 APWS00386 APWS00387 APWS00388 APWT02569 APWT02570 APWT02571 APWT02572 APWU00851 APWT03124 APWU00852 APWU00853 APWU00854 APWU00855 APWU00856 APWU00857 APWU00858 APWU00859 APWU00860 APWU00861 APWU00862 APWU00863 APWU00239 APWU00453 APWS00116 APWU00562 APWU00563 APWU02154 APWU02155 APWU02156 APWU02242 APWU02243 APWU02452 APWU02453 APWU02454 APWU02455 APWU02564 APWU02565 APWU02566 APWU02567 APWU02568 APWU02569 APWU02570 APWU02571 APWU02572 APWV00143 APWS00441 APWS00442 APWS00443 APWS00444 APWS00445 APWS00425 APWS00426 APWS00427 APWS00428 APWT02163 APWT02164 APWT02165 APWT02166 APWU00905 APWU00906 APWU00907 APWT00397 APWT00398 APWT02279 APWU00175 APWS00238 APWU01712 APWU01713 APWU01714 APWU01715 APWU01716 APWU01717 APWU01718 APWU01719 APWU01720 APWU01721 APWU01722 APWU01723 APWU01724 APWU01725 APWU01726 APWU01727 APWU01728 APWU01751 APWU01752 APWU01753 APWU01754 APWU01755 APWU01756 APWU01757 APWU01758 APWU01759 APWU01760 APWU01908 APWT02815 APWU00983 APWU00988 APWU01427 APWU01428 APWT00946 APWT00947 APWT00948 APWU01385 APWT02816 APWU00833 APWU00834 APWU00835 APWS00449 APWS00450 APWS00451 APWT00248 APWT00249 APWT00359 APWT00360 APWT00361 APWT02281 APWT02282 APWT00370 APWT00371 APWT00372 APWT00373 APWT00374 APWT00379 APWT00380 APWT02939 APWS00254 APWS00255 APWS00153 APWT02396 APWU00296 APWU02435 APWU02436 APWU02437 APWU02438 APWU02439 APWU02440 APWU02441 APWU02442 APWU02443 APWU02444 APWU02445 APWU02549 APWU02550 APWU02551 APWU02552 APWU01434 APWT02846 APWU00118 APWU00788 APWU01417 APWU01748 APWU01882 APWU02523 5409303 APWU01980 APWU02042 APWU02043 APWU02044 APWU02289 APWU02378 APWV00140 APWV00141 APWS00323 APWT02682 APWU01695 APWU02228 APWU02401 APWT00810 APWU00770 APWT00596 APWU00791 APWU00786 APWU01733 APWU01734 APWU01735 APWT00996 APWU00136 APWT00806 APWT00807 APWT00808 APWT00809 APWS00423 APWS00424 APWU02081 APWU02082 APWU02083 APWU02084 APWU02085 APWU02086 APWU02087 APWU02088 APWU02089 APWU02090 APWU02091 APWU02114 APWU02310 APWU02311 APWU02312 APWU02313 APWU02314 APWU02352 APWU02353 APWU02354 APWU02355 APWU02356 APWU02357 APWU02358 APWU02359 APWU02360 APWU02361 APWT01003 APWT01004 APWT01005 APWT01006 APWT00135 APWT00136 APWU01513 APWU01514 APWU00182 APWU00183 APWU00184 APWU00185 APWU00186 APWU00187 APWU00188 APWU00189 APWS00268 APWU02460 APWU02461 APWU02462 APWU02463 APWU02464 APWU02468 APWU02469 APWU02470 APWU02471 APWU02472 APWU02473 APWU02474 APWU02475 APWU02476 APWU02477 APWV00154 APWT00234 APWT00235 APWT00861 APWT00862 APWT00863 APWT00864 APWT00865 APWT00866 APWT00867 APWT00868 APWT00869 APWT00870 APWT00871 APWT00872 APWT00873 APWT00874 APWT00875 APWT00876 APWT00877 APWT00878 APWT00879 APWT00880 APWT00881 APWT03120 APWT02237 APWT02238 APWT02239 APWU00795 APWS00285 APWS00299 APWT02059 APWS00279 APWS00281 APWT02060 APWS00291 APWT00608 APWT00609 APWS00296 APWS00294 APWS00304 APWS00305 APWS00306 APWS00282 APWS00284 APWS00289 APWS00292 APWS00295 APWU00271 APWU00272 APWT00939 APWS00303 APWT00938 APWS00287 APWT02061 APWT00633 APWS00280 APWS00288 APWS00290 APWS00278 APWS00300 APWS00293 APWS00301 APWS00302 APWT00936 APWT00937 APWT02219 APWT02220 APWT02221 APWT02222 APWT02223 APWT02224 APWT02225 APWT02226 APWT02227 APWT02228 APWT02229 APWT02230 APWT02698 APWT02699 APWT02806 APWT02807 APWT02808 APWT02809 APWT02810 APWT02811 APWT02812 APWT02871 APWT02872 APWT02873 APWT02874 APWT02875 APWT02876 APWT02877 APWT02878 APWT02925 APWT02926 APWT02927 APWT02928 APWT02929 APWT02930 APWU00128 APWU00129 APWU00253 APWU00254 APWU00255 APWU00256 APWU00257 APWU00258 APWU00259 APWU00881 APWU00882 APWU00883 APWU00895 APWU00896 APWU00897 APWU00899 APWU00900 APWU00901 APWU00902 APWU00942 APWU00943 APWU00944 APWU00945 APWU00946 APWU00947 APWU00948 APWU00949 APWU00950 APWU00951 APWU00952 APWU00953 APWU00954 APWU00955 APWU00956 APWU00957 APWU00958 APWU00959 APWU00960 APWU00961 APWU00962 APWU00963 APWU00964 APWU00965 APWU00966 APWU00967 APWU00968 APWU01027 APWU01028 APWU01061 APWU01062 APWU01063 APWU01064 APWU01065 APWU01066 APWU01067 APWU01068 APWU01069 APWU01070 APWU01071 APWU01072 APWU01073 APWU01074 APWU01075 APWU01076 APWU01077 APWU01078 APWU01079 APWU01080 APWU01081 APWU01082 APWU01145 APWU01146 APWU01147 APWU01420 APWU01425 APWU01483 APWU01484 APWU01515 APWU01516 APWU01517 APWU01518 APWU01519 APWU01816 APWU01817 APWU01834 APWU02016 APWS00277 APWT00610 APWS00283 APWS00286 APWS00298 APWS00297 APWU00249 APWT02768 APWU02152 APWU02309 APWU02502 APWU02503 APWU02504 APWU02541 APWU02542 APWU02543 APWU02544 APWU02545 APWU02546 APWU02547 APWT02985 APWU00140 APWU01637 APWT00345 APWT02847 APWU01143 APWU01144 APWU01588 APWT02115 APWT02116 APWT02117 APWT02118 APWT02119 APWT02120 APWT02121 APWT02054 APWT02469 APWT02397 APWT02398 APWT03118 APWT03119 APWT00139 APWS00152 APWU02069 APWT00430 APWS00245 APWU00498 APWU00499 APWU00500 APWU00501 APWU00819 APWT02936 APWT02753 APWT02754 APWT02755 APWT02756 APWT02757 APWT02758 APWT02759 APWT02760 APWT00431 APWU00179 APWT00759 APWT00760 APWT00761 APWT00762 APWT00763 APWT00764 APWT00765 APWT00766 APWT00767 APWU00455 APWU00456 APWU00457 APWU00497 APWT00190 APWT00191 APWT00192 APWT00193 APWT00194 APWT00195 APWT00196 APWT00197 APWT00198 APWU01467 APWU01468 APWU01469 APWU01470 APWU01471 APWU01472 APWU01473 APWU01474 APWU01475 APWU01476 APWU01477 APWU01478 APWU01479 APWU01480 APWU01481 APWU01482 APWT02000 APWU00311 APWU01506 APWU01507 APWT00138 APWT00815 APWU00454 APWU00885 APWS00118 APWS00119 APWS00120 APWS00121 APWS00142 APWS00143 APWS00269 APWS00270 APWS00271 APWS00339 APWS00340 APWS00341 APWS00483 APWS00536 APWS00537 APWS00538 APWT00581 APWT00582 APWT00890 APWT02276 APWT02327 APWT02725 APWT02726 APWT03163 APWT03164 APWT03165 APWU00110 APWU00379 APWU00380 APWU00381 APWU00382 APWU00383 APWU00384 APWU00385 APWU00386 APWU00387 APWU00388 APWU00389 APWU00390 APWU00798 APWU00837 APWU01087 APWU01164 APWU01165 APWU02350 APWU02403 APWU02404 APWU01250 APWU01251 APWU01252 APWU01253 APWT02252 APWU01525 APWU01526 APWU01527 APWU01528 APWT00158 APWT00159 APWT00160 APWT00161 APWT00162 APWT00163 APWT00164 APWT00165 APWT00166 APWT00167 APWT00168 APWT00169 APWT00170 APWT00171 APWT00172 APWT00173 APWT00174 APWT00175 APWT00176 APWT00177 APWT02010 APWT02056 APWT02057 APWU00315 APWU00316 APWU00220 APWU00221 APWT00886 APWT02395 APWS00384 APWU01536 APWU01537 APWU01538 APWU01539 APWT02392 APWT02393 APWT02394 APWT00689 APWU02292 APWU02293 APWU02294 APWU02295 APWU02296 APWU02446 APWU02447 APWU02448 APWU02449 APWU02450 APWU02451 APWU02466 APWU02467 APWT02843 APWT02844 APWT00752 APWT00753 APWT00754 APWT00755 APWT00756 APWU00547 APWU00548 APWT00685 APWT00686 APWT00687 APWT00688 APWT00846 APWT00847 APWT00848 APWT00893 APWT00894 APWT01030 APWT01031 APWT01032 APWT02546 APWT02547 APWT02829 APWT02830 APWT00276 APWT00277 APWT00278 APWT00279 APWT00280 APWT00281 APWT00282 APWT00283 APWT00284 APWT00285 APWT00286 APWT00287 APWT00288 APWT00289 APWT00290 APWT00291 APWT00292 APWT00293 APWT00294 APWT00295 APWT00296 APWT00297 APWT00298 APWT00299 APWT00300 APWT00301 APWT00302 APWT00303 APWT00304 APWT02153 APWT02154 APWT02155 APWT02156 APWT02157 APWT02158 APWU02255 APWU02256 APWU02257 APWU02258 APWU02259 APWU02260 APWU02261 APWU02262 APWU02263 APWU02264 APWU02265 APWU02266 APWU02267 APWU02268 APWU02269 APWU02270 APWU02271 APWU02272 APWU02273 APWU02274 APWU02275 APWU02276 APWU02277 APWU02278 APWU02279 APWU02280 APWU02281 APWU02282 APWU02283 APWU02284 APWU02285 APWU02286 APWU02382 APWU02383 APWU02384 APWV00149 APWU00105 APWU01650 APWU01651 APWU01809 APWU01810 APWU01811 APWU01812 APWU01813 APWU00228 APWU00229 APWU00230 APWU00231 APWU00232 APWT00857 APWT00858 APWT00859 APWT00860 5403155-1 5403155-2 5403155-3 APWT03116 APWU00831 APWU00832 APWU02059 APWU02238 APWU02239 APWU02240 APWU02335 APWU02336 APWU02337 APWU02338 APWU02339 APWU02340 APWU02341 APWU02342 APWU02343 APWU02344 APWT01012 APWT00381 APWT00382 APWT00383 APWT00384 APWT00385 APWT00386 APWT00387
APWT00388 APWT00389 APWT00390 APWU01093 APWU01094 APWS00239 APWS00240 APWS00241 APWS00242 APWS00243 APWS00244 APWT01039 APWT01040 APWT01041 APWT01042 APWT01043 APWT01044 APWT01045 APWT01046 APWT01047 APWU00354 APWU00355 APWU00356 APWU00357 APWU00358 APWS00510 APWS00511 APWS00512 APWT00910 APWT00911 APWT00912 APWT02278 APWU01535 APWU01905 APWU01906 APWU01907 APWS00527 APWS00528 APWS00529 APWS00487 APWS00488 APWT00670 APWS00123 APWS00124 APWU00295 APWU00298 APWU00299 APWU00300 APWU00301 APWU00360 APWT02740 APWT02741 APWT02742 APWT02743 APWT02744 APWT02745 APWS00224 APWS00225 APWS00226 APWS00227 APWS00228 APWS00229 APWS00230 APWS00231 APWT03115 APWT00514 APWT00515 APWT00516 APWT00517 APWT00518 APWT00519 APWT00520 APWT00521 APWT00522 APWT00523 APWT00524 APWT00525 APWT00526 APWT00527 APWT00528 APWT00529 APWT02544 APWT02545 APWS00122 APWS00125 APWS00359 APWU01293 APWU01294 APWU01295 APWT02132 APWT02133 APWT02896 APWT02897 APWT02898 APWT02899 APWT02900 APWT02931 APWT00671 APWU00303 APWS00532 APWS00533 APWS00534 APWS00535 APWT00772 APWT00773 APWT00774 APWT00775 APWT02031 APWT02283 APWT02817 APWS00539 APWS00540 APWS00541 APWS00542 APWU01115 APWS00521 APWS00522 APWS00523 APWS00524 APWS00525 APWS00526 APWU01129 APWU01130 APWS00361 APWT00441 APWT00442 APWT00443 APWT00444 APWT02893 APWT02894 APWT02895 APWT02769 APWT02770 APWT02277 APWS00360 APWT00672 APWU00830 APWT02280 APWU00215 APWU00216 APWU00217 APWU00218 APWU00219 APWS00551 APWV00142 APWT02690 APWT02691 APWT02692 APWT02693 APWT02694 APWT02695 5382484-1 5382484-2 11341993 5361392-3 5361392-4 5361393-2 5361394-1 11355650 5346182-1 5346182-2 5346182-3 5346182-4 5346182-5 5346182-6 5346182-7 5346182-8 11343381 5388065-1 5388065-2 5388065-3 5388065-4 5388065-5 5388065-6 5388065-7 5362262-1 5362262-2 11323900-1 11323900-2 11323900-3 11323900-4 11323900-5 11323900-6 11323900-7 11323900-8 5375033-1 5375033-2 5375033-3 11351727 6504249-1 5383809-1 11351570 5383223-1 5383223-2 5383223-3 5350316-1 5350316-2 5350316-3 5350316-4 5350316-5 5350316-6 5350316-7 5350316-8 5350316-9 11355191 5367002-1 5376729-1 5376729-2 5376729-3 5388059-1 5388059-2 5377707-1 5330734-1 5330734-2 5375042-1 5375042-2 5375042-3 5375042-4 5378701-1 5378701-2 5378701-3 5373220-1 5373220-2 5373220-3 5373220-4 5373220-5 5373220-6 5378963-1 5394512-1 5372155-1 5372155-2 5372155-3 6504252-1 5367773-1 5367773-2 5367773-3 5367773-4 5353419-3 5353419-4 5353420-2 5353421-1 5371113-1 5371113-2 5371113-3 5371113-4 5371113-5 5365353-4 5365353-5 5365354-3 5365355-2 5365356-1 6505037-1 5353403-1 5353403-2 5353403-3 5353403-4 5353403-5 5353403-6 5379292-1 5379292-2 5379292-3 5379292-4 5379292-5 5350814-1 5371747-1 ANAK00930 ANAP00606 ANAP00609 11342168 5384379-1 5384379-2 5384379-3 5384379-4 5384379-5 5384379-6 5384379-7 5384379-8 11341897 5378034-1 5378034-2 5367586-1 5367586-2 5390285-1 5376732-1 5376732-2 5376732-3 5379286-1 5379286-2 5379286-3 5350424-1 5350424-2 5350424-3 5350424-4 5390390-1 5390390-2 5370659-1 5368095-1 5368095-2 5368095-3 5368095-4 5368095-5 5330740-1 5330740-2 6511813-1 5371854-1 5371854-2 5373320-1 5373320-2 5371101-1 5350819-1 5384342-1 5384342-2 5384342-3 5384342-4 5384402-1 11354283 11342583 5394438-1 5371738-1 5371738-2 5371738-3 11339305 11344906 11322132 11322362 11322879 11326559 11328760 11331736 11331825 11332655 11335458 11325950-1 11325950-2 11325950-3 11331187-1 11331187-2 20080855-1 20080855-2 20080855-3 6513362 11325451-1 11325451-2 ANAP01382 11349669 11358733 5396821-1 5403850 5404058 5399439 5409592 5410409 5412734 5404083-1 5404083-10 5404083-2 5404083-3 5404083-4 5404083-5 5404083-6 5404083-7 5404083-8 5404083-9 6515107-1 6515107-10 6515107-11 6515107-12 6515107-13 6515107-14 6515107-15 6515107-16 6515107-17 6515107-18 6515107-19 6515107-2 6515107-20 6515107-21 6515107-22 6515107-23 6515107-24 6515107-25 6515107-26 6515107-27 6515107-28 6515107-29 6515107-3 6515107-30 6515107-4 6515107-5 6515107-6 6515107-7 6515107-8 6515107-9 5402663 11350346 11350773-1 11350773-2 6504250-1 6504251-2 5378257-1 5378257-2 5378257-3 6513271-1 6513271-2 |
| FEI Number |
2126677
|
Recalling Firm/ Manufacturer |
GE Medical Systems, LLC
3000 N Grandview Blvd
Waukesha
WI
53188-1615
|
| For Additional Information Contact | GE Healthcare Service 800-437-1171 |
Manufacturer Reason for Recall | GE Healthcare has recently become aware of a potential safety issue involving a momentary, self-correcting anesthetic agent bolus when using 21% oxygen on all Aisys CS2 and upgraded Aisys anesthesia devices. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Consignees were sent on 2/17/2016 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# 34075 dated February 17, 2016.The letter was addressed to Chief of Anesthesia, Director of Biomedical / Clinical Engineering, and Health Care Administrator / Risk Manager. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions contact GE
Healthcare Service at 1-800-437-1171 or your local Service Representative. |
| Quantity in Commerce | 5496 |
| Distribution | US: AL, AZ, AR, CA, CO, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NE, NV, NJ, NY, NC, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI.
OUS: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BERMUDA, BOLIVIA, BRAZIL, CANDA, CHILE, CHINA, COLOMBIA, CROATIA, CR¿CHE REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY,GREECE, HELSINKI, HONG KONG, HUNGARIA, ICELAND, INDIA, IRAQ, IRELAND, ITA LY, JAPAN, JORDAN, KAZAKHSTAN, KOREA, KUWAIT, MALAYSIA, MEXICO, MOLDOVA, MOROCCO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NORWAY, PERU, POLAND, PORTUGAL QATAR, REUNION, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM,URUGUAY, VIETNAM. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = BSZ
|
|
|
|