• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Stryker

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Strykersee related information
Date Initiated by FirmOctober 10, 2013
Date PostedApril 18, 2016
Recall Status1 Terminated 3 on April 21, 2016
Recall NumberZ-1507-2016
Recall Event ID 73505
510(K)NumberK130009 
Product Classification Screw, fixation, bone - Product Code HWC
ProductVariAx Compression Plating System
Code Information catalog numbers 629503S, 629504S, 629505S, 629506S, 629507S, 629508S, 629509S, 629510S, 629527S, 629537S, 629543S, 629544S, 629545S, 629546S, 629547S, 629548S, 629559S, 629560S, 629561S, 629562S, 629564S, 629566S, 629568S, 629570S
FEI Number 3003070421
Recalling Firm/
Manufacturer
Stryker Howmedica Osteonics Corp.
325 Corporate Dr
Mahwah NJ 07430-2006
For Additional Information ContactMr. Michael Van Ryn
201-831-5000
Manufacturer Reason
for Recall
Stryker received two reports indicating that during screw insertion under a high lateral angle a bone screw went through a oblong hole.
FDA Determined
Cause 2
Device Design
ActionStryker Orthopaedics sent notification letters/product accountability forms on October 10, 2013, via email. Customers are provided the product issue, hazards and risk mitigation along with actions needed. Customers should receive the affected products and complete the BRF and return to Stryker. Questions can be directed to Stryker. For further questions, please call (201) 831-5000.
Quantity in Commerce118 units
DistributionNationwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = HWC
-
-