|
Class 2 Device Recall Zimmer MagnaFx Cannulated Screw Fixation System |
 |
Date Initiated by Firm |
May 04, 2016 |
Date Posted |
May 16, 2016 |
Recall Status1 |
Terminated 3 on July 05, 2017 |
Recall Number |
Z-1679-2016 |
Recall Event ID |
73981 |
510(K)Number |
K142442
|
Product Classification |
Screw, fixation, bone - Product Code HWC
|
Product |
Mini MagnaFx Cannulated Screw Fixation System Small Cannulated Bone Screw, Fully Threaded, 5.0 mm Diameter, 20 mm Length 00114205020 Mini Magna-Fx Cannulated Screw Fixation System Small Cannulated Bone Screw, Fully Threaded, 5.0 mm Diameter, 36 mm Length 00114205036 Mini MagnaFx Cannulated Screw Fixation System Small Cannulated Bone Screw, Partially Threaded, 5.0 mm Diameter, 24 mm Length 00114205124 Mini MagnaFx Cannulated Screw Fixation System Small Cannulated Bone Screw, Partially Threaded, 5.0 mm Diameter, 28 mm Length 00114205128 Mini MagnaFx Cannulated Screw Fixation System Small Cannulated Bone Screw, Partially Threaded, 5.0 mm Diameter, 30 mm Length 00114205130 Mini MagnaFx Cannulated Screw Fixation System Small Cannulated Bone Screw, Partially Threaded, 5.0 mm Diameter, 38 mm Length 00114205138 Mini MagnaFx Cannulated Screw Fixation System Small Cannulated Bone Screw, Partially Threaded, 5.0 mm Diameter, 48 mm Length 00114205148 Mini MagnaFx Cannulated Screw Fixation System Small Cannulated Bone Screw, Partially Threaded, 5.0 mm Diameter, 65 mm Length 00114205165 MagnaFx Cannulated Screw Fixation System Cannulated Bone Screw, Fully Threaded, 7.0 mm Diameter, 55 mm Length 00114605599 MagnaFx Cannulated Screw Fixation System Cannulated Bone Screw, Fully Threaded, 7.0 mm Diameter, 60 mm Length 00114606099 MagnaFx Cannulated Screw Fixation System Cannulated Bone Screw, 32 mm Thread Length, 7.0 mm Diameter, 65 mm Length 00114606532 MagnaFx Cannulated Screw Fixation System Cannulated Bone Screw, Fully Threaded, 7.0 mm Diameter, 65 mm Length 00114606599 Mini MagnaFx Cannulated Screw Fixation System Small Cannulated Bone Screw, Fully Threaded, 5.0 mm Diameter, 12 mm Length 00114205012
|
Code Information |
LOTS: 63290148 63295349 63290143 63295363 63290146 63296605 63295371 63295372 63308565 63308562 63303825 63259852 63295341 |
Recalling Firm/ Manufacturer |
Zimmer Manufacturing B.V. Turpeaux Industrial Park Rd 1 km 123.4 Mercedita PR 00715
|
Manufacturer Reason for Recall |
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
|
FDA Determined Cause 2 |
Packaging change control |
Action |
All distributors will be notified via electronic mail. Distributors with affected lots will also be notified via courier. Hospitals risk managers and surgeons will be notified via recall notification letter, dated May 4, 2016. |
Quantity in Commerce |
751 |
Distribution |
Worldwide distribution. US nationwide, Korea, Japan, Argentina, Brazil, Singapore, Taiwan, China, El Salvador, Australia, India, Hong Kong, Chile, Germany, Italy, France, Spain, UK, Sweden, Canada. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = HWC and Original Applicant = ZIMMER, INC.
|
|
|
|