| | Class 2 Device Recall Plum A Infusion Pump, Single Channel |  |
| Date Initiated by Firm | February 06, 2013 |
| Date Posted | September 06, 2016 |
| Recall Status1 |
Terminated 3 on November 14, 2016 |
| Recall Number | Z-2729-2016 |
| Recall Event ID |
64261 |
| 510(K)Number | K081412 |
| Product Classification |
Pump, infusion - Product Code FRN
|
| Product | Plum A+3 Infusion Pump. List Number 12618. The Plum A+ Infusion Systems are a multi-purpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self-contained, microprocessor based infusion devices. |
| Code Information |
list number 12618. serial numbers: 14040001, 14040002, 14040003, 14040004, 14040005, 14040006, 14040008, 14040009, 14040010, 14040011, 14040012, 14040014, 14040015, 14040016, 14040017, 14040018, 14040019, 14040020, 14040021, 14040022, 14040023, 14040024, 14040025, 14040026, 14040027, 14040027, 14040027, 14040028, 14040028, 14040028, 14040029, 14040030, 14040031, 14040033, 14040034, 14040034, 14040034, 14040035, 14040036, 14040037, 14040038, 14040039, 14040040, 14040041, 14040042, 14040043, 14040044, 14040045, 14040047, 14040048, 14040049, 14040050, 14040051, 14040052, 14040053, 14040054, 14040056, 14040059, 14040061, 14040062, 14040063, 14040064, 14040065, 14040066, 14040067, 14040068, 14040069, 14040070, 14040071, 14040072, 14040073, 14040074, 14040075, 14040076, 14040077, 14040078, 14040079, 14040080, 14040081, 14040082, 14040083, 14040084, 14040085, 14040086, 14040087, 14040088, 14040089, 14040090, 14040091, 14040092, 14040093, 14040094, 14040095, 14040096, 14040097, 14040098, 14040099, 14040100, 14040101, 14040103, 14040104, 14040105, 14040106, 14040107, 14040108, 14040109, 14040110, 14040111, 14040112, 14040113, 14040115, 14040116, 14040117, 14040118, 14040119, 14040120, 14040121, 14040122, 14040123, 14040124, 14040125, 14040128, 14040129, 14040130, 14040131, 14040132, 14040133, 14040134, 14040135, 14040136, 14040138, 14040139, 14040140, 14040141, 14040142, 14040143, 14040144, 14040145, 14040146, 14040148, 14040149, 14040150, 14040152, 14040153, 14040154, 14040155, 14040156, 14040156, 14040156, 14040157, 14040158, 14040159, 14040160, 14040161, 14040162, 14040163, 14040164, 14040165, 14040166, 14040167, 14040168, 14040169, 14040170, 14040171, 14040172, 14040173, 14040174, 14040175, 14040176, 14040178, 14040179, 14040180, 14040181, 14040181, 14040181, 14040182, 14040183, 14040184, 14040185, 14040186, 14040187, 14040189, 14040190, 14040191, 14040192, 14040193, 14040194, 14040196, 14040198, 14040199, 14040200, 14040201, 14040202, 14040203, 14040205, 14040206, 14040207, 14040208, 14040209, 14040210, 14040211, 14040212, 14040213, 14040214, 14040215, 14040216, 14040217, 14040219, 14040220, 14040221, 14040223, 14040224, 14040225, 14040226, 14040227, 14040228, 14040229, 14040230, 14040231, 14040232, 14040233, 14040234, 14040235, 14040236, 14040238, 14040239, 14040240, 14040242, 14040243, 14040244, 14040245, 14040246, 14040247, 14040248, 14040249, 14040250, 14040251, 14040252, 14040253, 14040254, 14040255, 14040256, 14040257, 14040258, 14040260, 14040261, 14040262, 14040263, 14040264, 14040265, 14040266, 14040267, 14040269, 14040270, 14040271, 14040272, 14040274, 14040275, 14040276, 14040277, 14040278, 14040279, 14040280, 14040282, 14040283, 14040284, 14040285, 14040286, 14040287, 14040288, 14040289, 14040290, 14040291, 14040292, 14040293, 14040294, 14040295, 14040296, 14040297, 14040298, 14040299, 14040300, 14040301, 14040302, 14040303, 14040304, 14040305, 14040306, 14040307, 14040308, 14040309, 14040310, 14040310, 14040310, 14040311, 14040312, 14040313, 14040314, 14040315, 14040316, 14040317, 14040318, 14040319, 14040320, 14040321, 14040323, 14040324, 14040326, 14040327, 14040328, 14040329, 14040330, 14040331, 14040332, 14040333, 14040334, 14040335, 14040336, 14040337, 14040338, 14040339, 14040340, 14040341, 14040342, 14040343, 14040344, 14040345, 14040347, 14040348, 14040349, 14040350, 14040351, 14040352, 14040353, 14040354, 14040357, 14040359, 14040360, 14040361, 14040362, 14040363, 14040364, 14040365, 14040366, 14040367, 14040368, 14040369, 14040371, 14040372, 14040373, 14040374, 14040375, 14040376, 14040377, 14040379, 14040380, 14040382, 14040383, 14040384, 14040386, 14040387, 14040388, 14040389, 14040390, 14040391, 14040393, 14040394, 14040396, 14040397, 14040398, 14040399, 14040400, 14040401, 14040403, 14040404, 14040405, 14040406, 14040407, 14040408, 14040409, 14040411, 14040414, 14040415, 14040416, 14040417, 14040418, 14040419, 14040420, 14040421, 14040422, 14040423, 14040424, 14040425, 14040426, 14040427, 14040429, 14040430, 14040431, 14040432, 14040433, 14040434, 14040435, 14040437, 14040438, 14040440, 14040441, 14040442, 14040443, 14040444, 14040445, 14040446, 14040447, 14040448, 14040449, 14040453, 14040454, 14040455, 14040457, 14040458, 14040459, 14040460, 14040461, 14040462, 14040463, 14040464, 14040465, 14040466, 14040467, 14040468, 14040469, 14040470, 14040471, 14040472, 14040473, 14040474, 14040475, 14040476, 14040477, 14040478, 14040479, 14040480, 14040481, 14040482, 14040483, 14040484, 14040485, 14040486, 14040487, 14040488, 14040490, 14040491, 14040495, 14040496, 14040497, 14040498, 14040499, 14040500, 14040501, 14040502, 14040503, 14040504, 14040505, 14040506, 14040507, 14040508, 14040509, 14040510, 14040512, 14040513, 14040514, 14040515, 14040516, 14040517, 14040518, 14040519, 14040520, 14040522, 14040523, 14040524, 14040525, 14040526, 14040527, 14040528, 14040529, 14040530, 14040531, 14040532, 14040533, 14040534, 14040535, 14040536, 14040538, 14040539, 14040540, 14040541, 14040542, 14040543, 14040544, 14040545, 14040546, 14040547, 14040548, 14040549, 14040550, 14040551, 14040552, 14040553, 14040554, 14040556, 14040557, 14040558, 14040559, 14040560, 14040561, 14040562, 14040564, 14040565, 14040566, 14040567, 14040568, 14040569, 14040570, 14040571, 14040572, 14040573, 14040574, 14040575, 14040576, 14040577, 14040578, 14040579, 14040580, 14040581, 14040582, 14040583, 14040584, 14040585, 14040586, 14040587, 14040588, 14040589, 14040590, 14040591, 14040592, 14040593, 14040594, 14040595, 14040596, 14040597, 14040598, 14040599, 14040600, 14040601, 14040602, 14040603, 14040604, 14040605, 14040606, 14040607, 14040608, 14040609, 14040610, 14040611, 14040612, 14040613, 14040614, 14040615, 14040616, 14040617, 14040618, 14040620, 14040621, 14040623, 14040624, 14040625, 14040626, 14040627, 14040628, 14040629, 14040630, 14040631, 14040632, 14040633, 14040634, 14040635, 14040636, 14040637, 14040638, 14040639, 14040640, 14040641, 14040643, 14040644, 14040645, 14040646, 14040647, 14040648, 14040649, 14040650, 14040651, 14040652, 14040653, 14040654, 14040655, 14040656, 14040657, 14040658, 14040659, 14040660, 14040661, 14040662, 14040663, 14040664, 14040665, 14040666, 14040667, 14040668, 14040669, 14040670, 14040671, 14040672, 14040673, 14040674, 14040675, 14040676, 14040677, 14040678, 14040679, 14040680, 14040681, 14040682, 14040683, 14040684, 14040685, 14040686, 14040687, 14040688, 14040689, 14040690, 14040691, 14040692, 14040693, 14040694, 14040695, 14040696, 14040697, 14040698, 14040699, 14040700, 14040701, 14040702, 14040703, 14040704, 14040705, 14040706, 14040707, 14040708, 14040709, 14040710, 14040711, 14040712, 14040713, 14040714, 14040715, 14040716, 14040717, 14040718, 14040719, 14040720, 14040721, 14040722, 14040723, 14040724, 14040725, 14040726, 14040727, 14040728, 14040729, 14040730, 14040731, 14040732, 14040733, 14040734, 14040735, 14040736, 14040737, 14040738, 14040739, 14040740, 14040741, 14040742, 14040743, 14040744, 14040745, 14040746, 14040747, 14040748, 14040749, 14040750, 14040751, 14040752, 14040753, 14040755, 14040756, 14040757, 14040758, 14040759, 14040760, 14040762, 14040763, 14040764, 14040765, 14040766, 14040767, 14040768, 14040769, 14040771, 14040772, 14040773, 14040774, 14040775, 14040776, 14040777, 14040778, 14040779, 14040780, 14040781, 14040782, 14040783, 14040784, 14040785, 14040786, 14040787, 14040788, 14040790, 14040791, 14040792, 14040794, 14040797, 14040798, 14040803, 14040805, 14040806, 14040808, 14040810, 14040811, 14040812, 14040814, 14040816, 14040819, 14040820, 14040821, 14040828, 14040830, 14040831, 14040832, 14040834, 14040835, 14040839, 14040841, 14040842, 14040845, 14040847, 14040848, 14040849, 14040852, 14040853, 14040854, 14040855, 14040856, 14040857, 14040858, 14040859, 14040860, 14040861, 14040862, 14040863, 14040864, 14040865, 14040866, 14040868, 14040869, 14040870, 14040871, 14040873, 14040874, 14040875, 14040876, 14040877, 14040878, 14040879, 14040880, 14040881, 14040882, 14040883, 14040884, 14040885, 14040886, 14040887, 14040888, 14040889, 14040890, 14040891, 14040892, 14040893, 14040894, 14040895, 14040896, 14040897, 14040898, 14040899, 14040901, 14040902, 14040903, 14040904, 14040905, 14040906, 14040907, 14040908, 14040909, 14040910, 14040911, 14040913, 14040914, 14040915, 14040916, 14040917, 14040918, 14040919, 14040920, 14040921, 14040923, 14040924, 14040925, 14040926, 14040927, 14040928, 14040929, 14040930, 14040931, 14040933, 14040934, 14040935, 14040936, 14040937, 14040938, 14040939, 14040940, 14040941, 14040942, 14040943, 14040944, 14040945, 14040946, 14040947, 14040948, 14040949, 14040950, 14040951, 14040952, 14040953, 14040955, 14040956, 14040957, 14040958, 14040959, 14040960, 14040961, 14040963, 14040964, 14040965, 14040966, 14040968, 14040969, 14040970, 14040971, 14040972, 14040973, 14040974, 14040975, 14040976, 14040977, 14040978, 14040979, 14040981, 14040982, 14040983, 14040984, 14040985, 14040986, 14040987, 14040988, 14040989, 14040990, 14040991, 14040992, 14040993, 14040994, 14040995, 14040996, 14040997, 14040998, 14040999, 14041000, 14041001, 14041002, 14041003, 14041004, 14041005, 14041006, 14041007, 14041008, 14041009, 14041010, 14041011, 14041012, 14041013, 14041014, 14041015, 14041016, 14041017, 14041018, 14041019, 14041020, 14041022, 14041023, 14041024, 14041025, 14041026, 14041027, 14041028, 14041029, 14041030, 14041031, 14041032, 14041033, 14041034, 14041035, 14041036, 14041037, 14041038, 14041039, 14041040, 14041041, 14041042, 14041043, 14041044, 14041045, 14041046, 14041047, 14041048, 14041049, 14041050, 14041051, 14041052, 14041053, 14041054, 14041055, 14041056, 14041057, 14041058, 14041059, 14041060, 14041061, 14041062, 14041063, 14041064, 14041065, 14041067, 14041068, 14041069, 14041070, 14041071, 14041072, 14041074, 14041075, 14041076, 14041077, 14041078, 14041079, 14041080, 14041081, 14041082, 14041083, 14041084, 14041085, 14041086, 14041088, 14041089, 14041091, 14041092, 14041093, 14041094, 14041095, 14041097, 14041098, 14041099, 14041100, 14041101, 14041102, 14041103, 14041105, 14041107, 14041108, 14041109, 14041110, 14041111, 14041112, 14041113, 14041114, 14041115, 14041116, 14041117, 14041118, 14041119, 14041120, 14041121, 14041123, 14041124, 14041125, 14041126, 14041127, 14041128, 14041129, 14041130, 14041132, 14041133, 14041134, 14041135, 14041136, 14041137, 14041138, 14041141, 14041144, 14041146, 14041148, 14041150, 14041154, 14041155, 14041157, 14041164, 14041167, 14041171, 14041172, 14041174, 14041177, 14041178, 14041179, 14041180, 14041181, 14041182, 14041183, 14041184, 14041186, 14041187, 14041189, 14041190, 14041191, 14041192, 14041193, 14041194, 14041195, 14041196, 14041197, 14041198, 14041199, 14041200, 14041201, 14041203, 14041204, 14041205, 14041206, 14041207, 14041209, 14041211, 14041212, 14041215, 14041216, 14041217, 14041218, 14041220, 14041221, 14041223, 14041224, 14041225, 14041226, 14041227, 14041228, 14041229, 14041230, 14041231, 14041232, 14041233, 14041234, 14041235, 14041236, 14041237, 14041238, 14041239, 14041240, 14041241, 14041242, 14041243, 14041244, 14041245, 14041246, 14041247, 14041248, 14041249, 14041250, 14041251, 14041252, 14041253, 14041254, 14041255, 14041256, 14041257, 14041258, 14041259, 14041260, 14041261, 14041262, 14041263, 14041264, 14041265, 14041266, 14041267, 14041269, 14041270, 14041272, 14041273, 14041274, 14041275, 14041276, 14041277, 14041278, 14041279, 14041280, 14041281, 14041282, 14041283, 14041284, 14041285, 14041286, 14041288, 14041289, 14041290, 14041291, 14041292, 14041293, 14041294, 14041295, 14041296, 14041297, 14041298, 14041299, 14041300, 14041301, 14041302, 14041303, 14041304, 14041305, 14041306, 14041307, 14041308, 14041309, 14041310, 14041311, 14041312, 14041313, 14041314, 14041315, 14041316, 14041317, 14041318, 14041319, 14041320, 14041321, 14041322, 14041324, 14041325, 14041327, 14041328, 14041329, 14041330, 14041331, 14041332, 14041333, 14041334, 14041335, 14041337, 14041338, 14041339, 14041341, 14041342, 14041343, 14041344, 14041345, 14041346, 14041347, 14041348, 14041349, 14041350, 14041351, 14041352, 14041353, 14041354, 14041355, 14041356, 14041357, 14041358, 14041359, 14041360, 14041361, 14041362, 14041363, 14041364, 14041365, 14041366, 14041367, 14041369, 14041370, 14041371, 14041372, 14041373, 14041374, 14041375, 14041376, 14041377, 14041378, 14041379, 14041380, 14041381, 14041382, 14041383, 14041384, 14041385, 14041386, 14041387, 14041388, 14041389, 14041390, 14041391, 14041392, 14041393, 14041394, 14041395, 14041396, 14041397, 14041398, 14041399, 14041400, 14041401, 14041402, 14041403, 14041404, 14041405, 14041406, 14041407, 14041408, 14041409, 14041410, 14041411, 14041412, 14041413, 14041414, 14041415, 14041416, 14041417, 14041418, 14041418, 14041418, 14041419, 14041421, 14041422, 14041423, 14041424, 14041425, 14041426, 14041427, 14041428, 14041429, 14041430, 14041431, 14041432, 14041433, 14041434, 14041435, 14041437, 14041438, 14041439, 14041440, 14041441, 14041442, 14041443, 14041444, 14041445, 14041446, 14041447, 14041448, 14041449, 14041450, 14041451, 14041451, 14041451, 14041452, 14041453, 14041454, 14041455, 14041455, 14041455, 14041456, 14041457, 14041458, 14041459, 14041460, 14041461, 14041462, 14041463, 14041464, 14041466, 14041467, 14041468, 14041469, 14041470, 14041472, 14041473, 14041474, 14041475, 14041476, 14041477, 14041478, 14041479, 14041480, 14041481, 14041482, 14041483, 14041484, 14041485, 14041486, 14041487, 14041488, 14041489, 14041490, 14041491, 14041492, 14041493, 14041494, 14041495, 14041496, 14041497, 14041498, 14041499, 14041500, 14041501, 14041502, 14041503, 14041504, 14041505, 14041506, 14041507, 14041508, 14041509, 14041510, 14041511, 14041512, 14041513, 14041514, 14041515, 14041516, 14041517, 14041518, 14041519, 14041520, 14041521, 14041522, 14041523, 14041525, 14041526, 14041527, 14041528, 14041529, 14041530, 14041531, 14041532, 14041533, 14041534, 14041535, 14041536, 14041537, 14041538, 14041539, 14041540, 14041541, 14041542, 14041544, 14041545, 14041546, 14041547, 14041548, 14041549, 14041550, 14041551, 14041552, 14041553, 14041554, 14041555, 14041556, 14041557, 14041558, 14041559, 14041560, 14041561, 14041562, 14041563, 14041564, 14041565, 14041569, 14041574, 14041576, 14041580, 14041581, 14041583, 14041584, 14041585, 14041586, 14041588, 14041589, 14041590, 14041591, 14041592, 14041593, 14041594, 14041595, 14041596, 14041597, 14041598, 14041599, 14041601, 14041602, 14041603, 14041604, 14041605, 14041606, 14041607, 14041608, 14041609, 14041610, 14041611, 14041612, 14041613, 14041614, 14041615, 14041616, 14041617, 14041618, 14041619, 14041620, 14041621, 14041622, 14041623, 14041624, 14041625, 14041626, 14041627, 14041628, 14041629, 14041630, 14041632, 14041633, 14041634, 14041635, 14041637, 14041638, 14041641, 14041642, 14041643, 14041644, 14041645, 14041646, 14041647, 14041648, 14041650, 14041651, 14041652, 14041653, 14041654, 14041655, 14041656, 14041657, 14041661, 14041662, 14041663, 14041664, 14041665, 14041666, 14041667, 14041668, 14041669, 14041670, 14041672, 14041673, 14041674, 14041675, 14041676, 14041677, 14041679, 14041680, 14041681, 14041682, 14041683, 14041684, 14041685, 14041686, 14041687, 14041688, 14041689, 14041690, 14041691, 14041692, 14041693, 14041694, 14041695, 14041696, 14041698, 14041699, 14041700, 14041701, 14041702, 14041703, 14041704, 14041706, 14041707, 14041708, 14041709, 14041710, 14041711, 14041712, 14041713, 14041714, 14041715, 14041716, 14041717, 14041718, 14041719, 14041720, 14041721, 14041722, 14041723, 14041724, 14041725, 14041726, 14041727, 14041728, 14041729, 14041730, 14041731, 14041732, 14041733, 14041734, 14041735, 14041736, 14041737, 14041738, 14041739, 14041740, 14041741, 14041742, 14041743, 14041744, 14041746, 14041747, 14041748, 14041749, 14041750, 14041752, 14041753, 14041755, 14041756, 14041757, 14041758, 14041759, 14041760, 14041761, 14041762, 14041763, 14041764, 14041765, 14041766, 14041767, 14041768, 14041769, 14041770, 14041771, 14041772, 14041773, 14041774, 14041775, 14041776, 14041777, 14041778, 14041779, 14041780, 14041783, 14041784, 14041785, 14041786, 14041787, 14041788, 14041789, 14041791, 14041792, 14041793, 14041794, 14041795, 14041796, 14041797, 14041800, 14041803, 14041804, 14041805, 14041806, 14041807, 14041808, 14041809, 14041811, 14041814, 14041815, 14041816, 14041817, 14041818, 14041819, 14041820, 14041821, 14041822, 14041823, 14041824, 14041825, 14041826, 14041827, 14041829, 14041830, 14041831, 14041833, 14041834, 14041836, 14041837, 14041838, 14041839, 14041840, 14041842, 14041843, 14041844, 14041845, 14041846, 14041847, 14041848, 14041849, 14041850, 14041851, 14041852, 14041854, 14041856, 14041857, 14041859, 14041860, 14041861, 14041862, 14041863, 14041866, 14041869, 14041872, 14041873, 14041874, 14041875, 14041876, 14041877, 14041878, 14041879, 14041880, 14041881, 14041882, 14041883, 14041884, 14041885, 14041886, 14041887, 14041890, 14041891, 14041893, 14041894, 14041895, 14041896, 14041897, 14041899, 14041900, 14041901, 14041902, 14041904, 14041905, 14041906, 14041907, 14041908, 14041909, 14041910, 14041911, 14041912, 14041913, 14041914, 14041915, 14041916, 14041917, 14041918, 14041919, 14041920, 14041921, 14041922, 14041923, 14041924, 14041926, 14041927, 14041928, 14041929, 14041930, 14041931, 14041932, 14041933, 14041934, 14041935, 14041936, 14041937, 14041938, 14041939, 14041940, 14041941, 14041944, 14041945, 14041946, 14041947, 14041948, 14041949, 14041950, 14041952, 14041953, 14041954, 14041955, 14041956, 14041957, 14041959, 14041960, 14041961, 14041962, 14041963, 14041964, 14041965, 14041966, 14041967, 14041968, 14041969, 14041970, 14041971, 14041972, 14041973, 14041974, 14041975, 14041976, 14041977, 14041978, 14041979, 14041980, 14041981, 14041982, 14041983, 14041984, 14041985, 14041986, 14041987, 14041988, 14041989, 14041990, 14041993, 14041994, 14041995, 14041996, 14041997, 14041998, 14041999, 14042000, 14042001, 14042002, 14042003, 14042004, 14042005, 14042006, 14042007, 14042008, 14042009, 14042010, 14042011, 14042012, 14042013, 14042015, 14042016, 14042017, 14042018, 14042019, 14042021, 14042022, 14042023, 14042024, 14042026, 14042027, 14042028, 14042029, 14042030, 14042031, 14042032, 14042033, 14042034, 14042035, 14042036, 14042037, 14042038, 14042039, 14042040, 14042041, 14042042, 14042043, 14042044, 14042045, 14042046, 14042047, 14042049, 14042052, 14042053, 14042055, 14042056, 14042057, 14042058, 14042059, 14042060, 14042062, 14042063, 14042064, 14042065, 14042066, 14042067, 14042068, 14042069, 14042070, 14042072, 14042073, 14042074, 14042076, 14042077, 14042078, 14042079, 14042081, 14042082, 14042083, 14042084, 14042086, 14042087, 14042088, 14042089, 14042090, 14042091, 14042092, 14042093, 14042094, 14042095, 14042096, 14042097, 14042098, 14042099, 14042100, 14042101, 14042102, 14042104, 14042105, 14042106, 14042107, 14042108, 14042109, 14042110, 14042111, 14042112, 14042113, 14042115, 14042116, 14042117, 14042118, 14042119, 14042120, 14042121, 14042123, 14042124, 14042125, 14042126, 14042127, 14042128, 14042130, 14042132, 14042133, 14042134, 14042135, 14042136, 14042138, 14042139, 14042141, 14042142, 14042143, 14042147, 14042148, 14042149, 14042151, 14042152, 14042154, 14042155, 14042156, 14042157, 14042158, 14042160, 14042161, 14042162, 14042163, 14042164, 14042165, 14042167, 14042171, 14042172, 14042173, 14042174, 14042175, 14042176, 14042177, 14042178, 14042179, 14042180, 14042181, 14042182, 14042183, 14042185, 14042186, 14042187, 14042188, 14042190, 14042192, 14042194, 14042195, 14042196, 14042199, 14042201, 14042202, 14042204, 14042205, 14042208, 14042210, 14042211, 14042212, 14042213, 14042214, 14042215, 14042218, 14042219, 14042220, 14042221, 14042222, 14042224, 14042227, 14042228, 14042230, 14042231, 14042232, 14042233, 14042234, 14042236, 14042237, 14042238, 14042239, 14042243, 14042244, 14042245, 14042246, 14042248, 14042249, 14042250, 14042251, 14042253, 14042254, 14042255, 14042257, 14042260, 14042261, 14042262, 14042263, 14042265, 14042266, 14042267, 14042269, 14042270, 14042273, 14042275, 14042278, 14042280, 14042281, 14042282, 14042284, 14042285, 14042289, 14042293, 14042295, 14042296, 14042298, 14042299, 14042301, 14042309, 14042310, 14042311, 14042312, 14042313, 14042315, 14042316, 14042320, 14042323, 14042324, 14042326, 14042329, 14042331, 14042332, 14042336, 14042340, 14042341, 14042342, 14042346, 14042347, 14042350, 14042353, 14042355, 14042356, 14042358, 14042359, 14042360, 14042361, 14042362, 14042363, 14042364, 14042365, 14042366, 14042367, 14042369, 14042370, 14042371, 14042373, 14042374, 14042375, 14042376, 14042377, 14042378, 14042379, 14042380, 14042381, 14042382, 14042383, 14042384, 14042385, 14042386, 14042387, 14042388, 14042390, 14042391, 14042392, 14042393, 14042394, 14042396, 14042399, 14042400, 14042401, 14042402, 14042403, 14042404, 14042405, 14042406, 14042407, 14042408, 14042410, 14042411, 14042414, 14042415, 14042416, 14042417, 14042420, 14042421, 14042422, 14042424, 14042425, 14042426, 14042428, 14042429, 14042430, 14042432, 14042434, 14042436, 14042437, 14042438, 14042439, 14042442, 14042443, 14042444, 14042446, 14042447, 14042448, 14042450, 14042451, 14042451, 14042452, 14042452, 14042453, 14042454, 14042454, 14042455, 14042455, 14042459, 14042461, 14042462, 14042463, 14042464, 14042465, 14042466, 14042467, 14042468, 14042469, 14042470, 14042471, 14042472, 14042474, 14042475, 14042477, 14042478, 14042479, 14042481, 14042482, 14042483, 14042484, 14042485, 14042486, 14042487, 14042488, 14042490, 14042491, 14042492, 14042493, 14042494, 14042495, 14042496, 14042497, 14042498, 14042499, 14042500, 14042501, 14042502, 14042503, 14042504, 14042505, 14042506, 14042507, 14042508, 14042509, 14042510, 14042511, 14042512, 14042513, 14042514, 14042515, 14042516, 14042517, 14042518, 14042519, 14042520, 14042521, 14042522, 14042523, 14042524, 14042527, 14042528, 14042529, 14042530, 14042531, 14042532, 14042533, 14042534, 14042535, 14042536, 14042537, 14042538, 14042539, 14042540, 14042541, 14042542, 14042543, 14042544, 14042545, 14042546, 14042547, 14042548, 14042549, 14042550, 14042551, 14042552, 14042553, 14042554, 14042555, 14042556, 14042557, 14042558, 14042559, 14042560, 14042561, 14042562, 14042564, 14042565, 14042566, 14042567, 14042568, 14042569, 14042570, 14042571, 14042572, 14042573, 14042574, 14042575, 14042576, 14042577, 14042578, 14042579, 14042580, 14042581, 14042582, 14042583, 14042584, 14042586, 14042587, 14042588, 14042590, 14042591, 14042592, 14042593, 14042594, 14042595, 14042596, 14042597, 14042598, 14042599, 14042600, 14042601, 14042602, 14042603, 14042604, 14042605, 14042606, 14042607, 14042609, 14042610, 14042611, 14042612, 14042613, 14042614, 14042616, 14042617, 14042618, 14042619, 14042620, 14042621, 14042622, 14042623, 14042625, 14042626, 14042627, 14042628, 14042628, 14042628, 14042629, 14042630, 14042631, 14042632, 14042633, 14042634, 14042635, 14042636, 14042637, 14042638, 14042639, 14042640, 14042642, 14042643, 14042644, 14042645, 14042646, 14042647, 14042648, 14042649, 14042650, 14042651, 14042652, 14042653, 14042654, 14042655, 14042656, 14042659, 14042660, 14042661, 14042662, 14042663, 14042664, 14042666, 14042667, 14042668, 14042669, 14042670, 14042671, 14042672, 14042673, 14042674, 14042677, 14042678, 14042679, 14042680, 14042681, 14042682, 14042683, 14042684, 14042685, 14042686, 14042687, 14042688, 14042689, 14042690, 14042691, 14042692, 14042693, 14042694, 14042695, 14042696, 14042697, 14042698, 14042699, 14042700, 14042702, 14042703, 14042704, 14042705, 14042707, 14042708, 14042709, 14042710, 14042711, 14042712, 14042713, 14042714, 14042715, 14042717, 14042718, 14042719, 14042720, 14042722, 14042723, 14042724, 14042725, 14042726, 14042727, 14042728, 14042729, 14042730, 14042731, 14042732, 14042733, 14042734, 14042736, 14042737, 14042739, 14042740, 14042741, 14042742, 14042745, 14042746, 14042747, 14042749, 14042750, 14042751, 14042752, 14042755, 14042757, 14042758, 14042759, 14042760, 14042761, 14042763, 14042764, 14042767, 14042770, 14042771, 14042772, 14042774, 14042775, 14042776, 14042777, 14042780, 14042781, 14042783, 14042784, 14042785, 14042786, 14042788, 14042789, 14042790, 14042793, 14042794, 14042795, 14042798, 14042799, 14042800, 14042802, 14042803, 14042804, 14042805, 14042806, 14042807, 14042808, 14042809, 14042810, 14042811, 14042812, 14042813, 14042814, 14042815, 14042816, 14042817, 14042818, 14042819, 14042820, 14042821, 14042822, 14042823, 14042824, 14042825, 14042826, 14042827, 14042828, 14042829, 14042830, 14042831, 14042832, 14042833, 14042834, 14042835, 14042836, 14042837, 14042838, 14042839, 14042840, 14042842, 14042843, 14042844, 14042845, 14042846, 14042847, 14042848, 14042850, 14042851, 14042852, 14042853, 14042854, 14042855, 14042856, 14042857, 14042858, 14042859, 14042861, 14042862, 14042863, 14042864, 14042865, 14042866, 14042867, 14042868, 14042869, 14042870, 14042871, 14042873, 14042874, 14042875, 14042876, 14042877, 14042878, 14042879, 14042880, 14042881, 14042882, 14042883, 14042884, 14042885, 14042886, 14042887, 14042888, 14042889, 14042890, 14042891, 14042892, 14042893, 14042894, 14042895, 14042896, 14042897, 14042898, 14042899, 14042900, 14042902, 14042903, 14042904, 14042905, 14042906, 14042907, 14042908, 14042909, 14042910, 14042911, 14042912, 14042913, 14042915, 14042918, 14042919, 14042920, 14042921, 14042922, 14042923, 14042924, 14042924, 14042925, 14042927, 14042928, 14042929, 14042929, 14042930, 14042931, 14042932, 14042933, 14042935, 14042936, 14042937, 14042938, 14042939, 14042940, 14042941, 14042942, 14042943, 14042944, 14042945, 14042946, 14042947, 14042948, 14042949, 14042950, 14042951, 14042952, 14042953, 14042954, 14042955, 14042956, 14042957, 14042958, 14042959, 14042960, 14042961, 14042962, 14042963, 14042964, 14042965, 14042966, 14042967, 14042968, 14042969, 14042970, 14042971, 14042972, 14042973, 14042974, 14042975, 14042976, 14042977, 14042979, 14042980, 14042981, 14042982, 14042983, 14042984, 14042985, 14042986, 14042987, 14042988, 14042989, 14042990, 14042990, 14042991, 14042992, 14042993, 14042994, 14042995, 14042996, 14042997, 14042998, 14042999, 14043000, 14043001, 14043002, 14043003, 14043004, 14043005, 14043006, 14043008, 14043009, 14043010, 14043012, 14043013, 14043014, 14043015, 14043016, 14043017, 14043019, 14043020, 14043021, 14043022, 14043023, 14043024, 14043025, 14043026, 14043027, 14043028, 14043029, 14043030, 14043031, 14043032, 14043033, 14043034, 14043035, 14043036, 14043037, 14043038, 14043039, 14043040, 14043041, 14043042, 14043043, 14043044, 14043045, 14043046, 14043047, 14043048, 14043049, 14043050, 14043051, 14043052, 14043053, 14043054, 14043055, 14043056, 14043057, 14043058, 14043059, 14043060, 14043061, 14043062, 14043064, 14043065, 14043066, 14043067, 14043068, 14043069, 14043070, 14043071, 14043072, 14043073, 14043074, 14043075, 14043076, 14043077, 14043078, 14043079, 14043080, 14043081, 14043082, 14043083, 14043084, 14043085, 14043086, 14043087, 14043088, 14043089, 14043090, 14043091, 14043092, 14043093, 14043094, 14043095, 14043096, 14043097, 14043098, 14043101, 14043102, 14043104, 14043105, 14043106, 14043107, 14043108, 14043109, 14043110, 14043111, 14043112, 14043113, 14043116, 14043117, 14043120, 14043122, 14043124, 14043125, 14043126, 14043129, 14043130, 14043132, 14043133, 14043134, 14043135, 14043137, 14043138, 14043139, 14043140, 14043141, 14043142, 14043144, 14043145, 14043146, 14043149, 14043150, 14043154, 14043156, 14043157, 14043158, 14043159, 14043160, 14043161, 14043162, 14043163, 14043165, 14043166, 14043167, 14043168, 14043170, 14043172, 14043173, 14043174, 14043176, 14043180, 14043182, 14043183, 14043184, 14043186, 14043187, 14043188, 14043190, 14043193, 14043195, 14043197, 14043202, 14043203, 14043204, 14043206, 14043207, 14043209, 14043211, 14043214, 14043215, 14043217, 14043218, 14043219, 14043221, 14043222, 14043223, 14043225, 14043226, 14043227, 14043228, 14043229, 14043230, 14043231, 14043233, 14043234, 14043235, 14043236, 14043237, 14043238, 14043240, 14043241, 14043242, 14043243, 14043244, 14043246, 14043247, 14043249, 14043250, 14043251, 14043252, 14043253, 14043254, 14043255, 14043256, 14043257, 14043258, 14043259, 14043260, 14043261, 14043262, 14043263, 14043264, 14043266, 14043267, 14043268, 14043269, 14043271, 14043272, 14043273, 14043275, 14043276, 14043277, 14043277, 14043278, 14043279, 14043280, 14043281, 14043284, 14043285, 14043286, 14043287, 14043288, 14043289, 14043290, 14043291, 14043292, 14043293, 14043294, 14043295, 14043296, 14043297, 14043298, 14043299, 14043300, 14043301, 14043302, 14043303, 14043304, 14043305, 14043306, 14043307, 14043308, 14043309, 14043310, 14043311, 14043312, 14043314, 14043315, 14043316, 14043317, 14043318, 14043319, 14043320, 14043321, 14043322, 14043323, 14043324, 14043325, 14043326, 14043327, 14043328, 14043329, 14043330, 14043331, 14043332, 14043334, 14043335, 14043336, 14043338, 14043340, 14043341, 14043342, 14043343, 14043344, 14043345, 14043346, 14043347, 14043348, 14043349, 14043350, 14043351, 14043352, 14043354, 14043355, 14043356, 14043357, 14043360, 14043361, 14043362, 14043363, 14043366, 14043368, 14043369, 14043370, 14043371, 14043372, 14043374, 14043375, 14043376, 14043377, 14043378, 14043380, 14043381, 14043382, 14043383, 14043385, 14043386, 14043387, 14043389, 14043391, 14043392, 14043393, 14043394, 14043396, 14043397, 14043398, 14043399, 14043402, 14043403, 14043404, 14043405, 14043406, 14043407, 14043408, 14043411, 14043416, 14043417, 14043418, 14043419, 14043420, 14043421, 14043423, 14043424, 14043425, 14043426, 14043427, 14043434, 14043438, 14043440, 14043441, 14043442, 14043443, 14043444, 14043445, 14043446, 14043447, 14043448, 14043450, 14043451, 14043452, 14043453, 14043454, 14043455, 14043458, 14043460, 14043461, 14043462, 14043463, 14043464, 14043465, 14043466, 14043467, 14043468, 14043470, 14043471, 14043472, 14043473, 14043474, 14043475, 14043476, 14043477, 14043478, 14043479, 14043480, 14043481, 14043482, 14043483, 14043484, 14043485, 14043486, 14043487, 14043488, 14043489, 14043490, 14043491, 14043492, 14043493, 14043494, 14043495, 14043496, 14043497, 14043498, 14043499, 14043500, 14043501, 14043502, 14043503, 14043504, 14043505, 14043506, 14043507, 14043508, 14043509, 14043510, 14043511, 14043512, 14043513, 14043514, 14043515, 14043516, 14043517, 14043518, 14043519, 14043520, 14043521, 14043522, 14043523, 14043524, 14043525, 14043526, 14043527, 14043528, 14043529, 14043530, 14043531, 14043533, 14043534, 14043535, 14043536, 14043537, 14043538, 14043539, 14043540, 14043541, 14043542, 14043543, 14043544, 14043545, 14043546, 14043547, 14043548, 14043549, 14043550, 14043551, 14043552, 14043553, 14043554, 14043555, 14043556, 14043557, 14043558, 14043559, 14043560, 14043561, 14043562, 14043563, 14043564, 14043565, 14043566, 14043567, 14043568, 14043569, 14043570, 14043571, 14043572, 14043573, 14043574, 14043575, 14043576, 14043577, 14043578, 14043579, 14043580, 14043581, 14043582, 14043583, 14043584, 14043585, 14043586, 14043587, 14043588, 14043589, 14043590, 14043591, 14043592, 14043593, 14043594, 14043595, 14043596, 14043597, 14043598, 14043599, 14043600, 14043601, 14043602, 14043603, 14043604, 14043605, 14043606, 14043607, 14043608, 14043609, 14043610, 14043611, 14043612, 14043613, 14043614, 14043615, 14043616, 14043617, 14043618, 14043619, 14043620, 14043621, 14043622, 14043623, 14043624, 14043625, 14043626, 14043627, 14043628, 14043629, 14043630, 14043631, 14043633, 14043634, 14043635, 14043636, 14043637, 14043638, 14043639, 14043640, 14043641, 14043642, 14043643, 14043644, 14043645, 14043646, 14043647, 14043648, 14043649, 14043650, 14043651, 14043652, 14043653, 14043654, 14043655, 14043656, 14043657, 14043658, 14043659, 14043660, 14043661, 14043662, 14043663, 14043664, 14043665, 14043666, 14043667, 14043668, 14043669, 14043670, 14043671, 14043672, 14043673, 14043674, 14043675, 14043676, 14043677, 14043678, 14043679, 14043680, 14043681, 14043682, 14043683, 14043684, 14043685, 14043686, 14043687, 14043688, 14043689, 14043690, 14043691, 14043692, 14043693, 14043694, 14043695, 14043696, 14043697, 14043698, 14043699, 14043700, 14043701, 14043702, 14043703, 14043704, 14043705, 14043706, 14043707, 14043708, 14043709, 14043710, 14043711, 14043712, 14043713, 14043714, 14043715, 14043716, 14043717, 14043718, 14043719, 14043720, 14043721, 14043722, 14043723, 14043724, 14043725, 14043726, 14043727, 14043728, 14043729, 14043730, 14043731, 14043732, 14043733, 14043734, 14043735, 14043736, 14043737, 14043738, 14043739, 14043740, 14043741, 14043742, 14043743, 14043744, 14043745, 14043746, 14043747, 14043748, 14043749, 14043750, 14043751, 14043752, 14043753, 14043754, 14043755, 14043756, 14043757, 14043758, 14043759, 14043760, 14043761, 14043762, 14043763, 14043764, 14043765, 14043766, 14043767, 14043768, 14043769, 14043770, 14043771, 14043772, 14043773, 14043774, 14043775, 14043776, 14043777, 14043778, 14043779, 14043780, 14043781, 14043782, 14043783, 14043784, 14043785, 14043786, 14043787, 14043788, 14043789, 14043790, 14043791, 14043792, 14043793, 14043794, 14043795, 14043796, 14043797, 14043798, 14043799, 14043800, 14043801, 14043802, 14043803, 14043804, 14043805, 14043806, 14043807, 14043808, 14043809, 14043810, 14043811, 14043812, 14043813, 14043814, 14043815, 14043816, 14043817, 14043818, 14043819, 14043820, 14043821, 14043822, 14043823, 14043824, 14043825, 14043826, 14043827, 14043828, 14043829, 14043830, 14043831, 14043832, 14043833, 14043834, 14043835, 14043836, 14043837, 14043838, 14043839, 14043840, 14043841, 14043842, 14043843, 14043844, 14043845, 14043846, 14043847, 14043848, 14043849, 14043850, 14043851, 14043852, 14043853, 14043854, 14043855, 14043856, 14043857, 14043858, 14043859, 14043860, 14043861, 14043862, 14043863, 14043864, 14043865, 14043866, 14043867, 14043868, 14043869, 14043870, 14043871, 14043872, 14043873, 14043874, 14043875, 14043876, 14043877, 14043878, 14043879, 14043880, 14043881, 14043882, 14043883, 14043884, 14043885, 14043886, 14043887, 14043888, 14043889, 14043890, 14043891, 14043892, 14043893, 14043894, 14043895, 14043896, 14043897, 14043898, 14043899, 14043900, 14043901, 14043902, 14043903, 14043904, 14043905, 14043906, 14043907, 14043908, 14043909, 14043910, 14043911, 14043912, 14043913, 14043914, 14043915, 14043916, 14043917, 14043918, 14043920, 14043921, 14043922, 14043923, 14043924, 14043925, 14043926, 14043927, 14043928, 14043929, 14043930, 14043931, 14043932, 14043933, 14043934, 14043935, 14043936, 14043937, 14043938, 14043939, 14043940, 14043941, 14043942, 14043943, 14043944, 14043945, 14043946, 14043947, 14043948, 14043949, 14043950, 14043951, 14043952, 14043953, 14043954, 14043956, 14043957, 14043958, 14043959, 14043960, 14043961, 14043962, 14043963, 14043964, 14043965, 14043966, 14043967, 14043968, 14043969, 14043970, 14043971, 14043972, 14043973, 14043974, 14043975, 14043976, 14043977, 14043978, 14043979, 14043980, 14043981, 14043982, 14043983, 14043984, 14043985, 14043986, 14043987, 14043988, 14043989, 14043990, 14043991, 14043992, 14043993, 14043994, 14043995, 14043996, 14043997, 14043998, 14043999, 14044000, 14044001, 14044002, 14044003, 14044004, 14044005, 14044006, 14044007, 14044009, 14044010, 14044011, 14044012, 14044013, 14044014, 14044015, 14044016, 14044017, 14044018, 14044019, 14044021, 14044022, 14044023, 14044024, 14044025, 14044026, 14044028, 14044029, 14044030, 14044031, 14044032, 14044033, 14044034, 14044035, 14044037, 14044038, 14044039, 14044040, 14044041, 14044042, 14044043, 14044044, 14044046, 14044047, 14044048, 14044054, 14044055, 14044056, 14044057, 14044058, 14044059, 14044060, 14044061, 14044062, 14044063, 14044064, 14044065, 14044066, 14044067, 14044068, 14044069, 14044072, 14044073, 14044074, 14044075, 14044076, 14044078, 14044079, 14044080, 14044081, 14044082, 14044083, 14044084, 14044085, 14044086, 14044087, 14044088, 14044089, 14044090, 14044091, 14044092, 14044093, 14044094, 14044095, 14044096, 14044097, 14044098, 14044099, 14044100, 14044101, 14044102, 14044103, 14044104, 14044105, 14044106, 14044107, 14044108, 14044109, 14044110, 14044111, 14044112, 14044113, 14044114, 14044115, 14044116, 14044117, 14044118, 14044119, 14044120, 14044121, 14044122, 14044123, 14044124, 14044125, 14044126, 14044127, 14044128, 14044129, 14044130, 14044131, 14044132, 14044134, 14044135, 14044136, 14044137, 14044138, 14044139, 14044140, 14044141, 14044142, 14044143, 14044144, 14044145, 14044146, 14044147, 14044148, 14044149, 14044150, 14044151, 14044152, 14044153, 14044154, 14044155, 14044156, 14044158, 14044159, 14044160, 14044161, 14044162, 14044163, 14044164, 14044165, 14044166, 14044167, 14044168, 14044169, 14044170, 14044171, 14044173, 14044175, 14044176, 14044177, 14044178, 14044179, 14044180, 14044181, 14044182, 14044183, 14044188, 14044190, 14044192, 14044193, 14044194, 14044195, 14044196, 14044197, 14044202, 14044205, 14044207, 14044208, 14044209, 14044210, 14044211, 14044213, 14044215, 14044216, 14044219, 14044220, 14044221, 14044223, 14044224, 14044225, 14044227, 14044228, 14044230, 14044231, 14044232, 14044236, 14044237, 14044239, 14044241, 14044242, 14044243, 14044443, 14044444, 14044445, 14044446, 14044447, 14044448, 14044449, 14044450, 14044451, 14044452, 14044453, 14044454, 14044455, 14044456, 14044457, 14044458, 14044459, 14044460, 14044461, 14044462, 14044463, 14044464, 14044465, 14044466, 14044467, 14044468, 14044469, 14044470, 14044472, 14044473, 14044474, 14044475, 14044476, 14044478, 14044479, 14044480, 14044481, 14044482, 14044483, 14044484, 14044485, 14044486, 14044487, 14044488, 14044489, 14044490, 14044491, 14044492, 14044493, 14044494, 14044495, 14044496, 14044497, 14044498, 14044499, 14044500, 14044501, 14044502, 14044503, 14044504, 14044506, 14044507, 14044508, 14044509, 14044510, 14044511, 14044512, 14044513, 14044514, 14044515, 14044516, 14044517, 14044518, 14044519, 14044520, 14044521, 14044522, 14044523, 14044524, 14044525, 14044526, 14044527, 14044528, 14044529, 14044530, 14044531, 14044532, 14044533, 14044534, 14044535, 14044536, 14044537, 14044538, 14044539, 14044540, 14044541, 14044542, 14044543, 14044544, 14044545, 14044546, 14044547, 14044548, 14044549, 14044550, 14044551, 14044552, 14044553, 14044554, 14044555, 14044556, 14044557, 14044558, 14044559, 14044560, 14044561, 14044562, 14044563, 14044564, 14044565, 14044566, 14044567, 14044568, 14044569, 14044570, 14044571, 14044572, 14044573, 14044574, 14044575, 14044576, 14044577, 14044578, 14044579, 14044580, 14044581, 14044582, 14044583, 14044584, 14044585, 14044586, 14044587, 14044588, 14044589, 14044590, 14044591, 14044592, 14044593, 14044594, 14044595, 14044596, 14044597, 14044598, 14044599, 14044600, 14044601, 14044602, 14044603, 14044614, 14044615, 14044617, 14044618, 14044619, 14044620, 14044621, 14044622, 14044623, 14044624, 14044625, 14044626, 14044627, 14044628, 14044629, 14044630, 14044631, 14044632, 14044633, 14044634, 14044635, 14044636, 14044637, 14044638, 14044639, 14044640, 14044641, 14044642, 14044643, 14044644, 14044645, 14044646, 14044647, 14044648, 14044649, 14044650, 14044651, 14044652, 14044653, 14044654, 14044655, 14044656, 14044657, 14044658, 14044659, 14044660, 14044661, 14044662, 14044663, 14044664, 14044665, 14044666, 14044667, 14044668, 14044669, 14044670, 14044671, 14044672, 14044673, 14044674, 14044675, 14044676, 14044677, 14044678, 14044679, 14044680, 14044681, 14044682, 14044683, 14044684, 14044685, 14044686, 14044687, 14044688, 14044689, 14044690, 14044691, 14044692, 14044693, 14044694, 14044695, 14044696, 14044697, 14044698, 14044699, 14044700, 14044701, 14044702, 14044703, 14044704, 14044705, 14044706, 14044707, 14044708, 14044709, 14044710, 14044711, 14044712, 14044713, 14044714, 14044715, 14044716, 14044717, 14044718, 14044719, 14044720, 14044721, 14044722, 14044723, 14044724, 14044725, 14044726, 14044727, 14044728, 14044729, 14044730, 14044731, 14044732, 14044733, 14044734, 14044735, 14044736, 14044738, 14044739, 14044740, 14044741, 14044742, 14044744, 14044745, 14044746, 14044747, 14044748, 14044749, 14044750, 14044751, 14044752, 14044753, 14044754, 14044755, 14044756, 14044757, 14044758, 14044759, 14044760, 14044761, 14044762, 14044763, 14044764, 14044765, 14044766, 14044767, 14044768, 14044769, 14044771, 14044773, 14044774, 14044775, 14044776, 14044778, 14044779, 14044780, 14044782, 14044783, 14044784, 14044785, 14044785, 14044786, 14044787, 14044788, 14044789, 14044790, 14044791, 14044792, 14044793, 14044794, 14044795, 14044796, 14044797, 14044798, 14044799, 14044800, 14044801, 14044802, 14044803, 14044804, 14044805, 14044806, 14044807, 14044808, 14044809, 14044810, 14044811, 14044813, 14044814, 14044815, 14044816, 14044818, 14044819, 14044820, 14044821, 14044822, 14044823, 14044824, 14044825, 14044826, 14044827, 14044828, 14044829, 14044830, 14044831, 14044832, 14044833, 14044834, 14044835, 14044836, 14044837, 14044838, 14044839, 14044840, 14044841, 14044842, 14044843, 14044844, 14044845, 14044846, 14044847, 14044848, 14044849, 14044850, 14044851, 14044852, 14044853, 14044854, 14044855, 14044856, 14044857, 14044858, 14044859, 14044860, 14044861, 14044862, 14044863, 14044864, 14044865, 14044866, 14044868, 14044869, 14044870, 14044871, 14044872, 14044873, 14044874, 14044876, 14044877, 14044878, 14044879, 14044880, 14044881, 14044882, 14044883, 14044884, 14044885, 14044886, 14044888, 14044889, 14044890, 14044891, 14044892, 14044893, 14044894, 14044895, 14044896, 14044897, 14044898, 14044899, 14044900, 14044901, 14044902, 14044903, 14044904, 14044905, 14044906, 14044907, 14044908, 14044909, 14044910, 14044911, 14044912, 14044913, 14044914, 14044915, 14044916, 14044917, 14044918, 14044919, 14044920, 14044921, 14044922, 14044924, 14044925, 14044926, 14044927, 14044928, 14044929, 14044930, 14044931, 14044933, 14044934, 14044935, 14044936, 14044937, 14044938, 14044939, 14044940, 14044941, 14044942, 14044943, 14044944, 14044945, 14044946, 14044947, 14044948, 14044949, 14044950, 14044951, 14044952, 14044953, 14044954, 14044955, 14044956, 14044957, 14044958, 14044959, 14044960, 14044961, 14044962, 14044964, 14044965, 14044966, 14044967, 14044968, 14044969, 14044970, 14044971, 14044973, 14044974, 14044975, 14044977, 14044980, 14044981, 14044982, 14044983, 14044984, 14044985, 14044986, 14044987, 14044988, 14044989, 14044990, 14044991, 14044992, 14044993, 14044994, 14044995, 14044996, 14044997, 14044998, 14044999, 14045000, 14045001, 14045002, 14045003, 14045004, 14045005, 14045006, 14045007, 14045008, 14045009, 14045010, 14045011, 14045012, 14045013, 14045014, 14045015, 14045016, 14045017, 14045018, 14045019, 14045020, 14045021, 14045022, 14045023, 14045024, 14045025, 14045025, 14045026, 14045027, 14045027, 14045028, 14045029, 14045030, 14045031, 14045032, 14045033, 14045034, 14045035, 14045036, 14045037, 14045038, 14045039, 14045040, 14045041, 14045042, 14045043, 14045044, 14045045, 14045046, 14045047, 14045048, 14045049, 14045050, 14045051, 14045052, 14045053, 14045054, 14045055, 14045056, 14045057, 14045058, 14045059, 14045060, 14045061, 14045062, 14045063, 14045064, 14045065, 14045066, 14045067, 14045068, 14045069, 14045070, 14045071, 14045073, 14045074, 14045075, 14045076, 14045077, 14045078, 14045079, 14045080, 14045081, 14045082, 14045083, 14045084, 14045085, 14045086, 14045087, 14045088, 14045089, 14045090, 14045091, 14045092, 14045093, 14045094, 14045095, 14045096, 14045097, 14045098, 14045099, 14045100, 14045101, 14045102, 14045103, 14045104, 14045105, 14045106, 14045107, 14045108, 14045109, 14045110, 14045111, 14045112, 14045113, 14045114, 14045115, 14045116, 14045117, 14045117, 14045118, 14045119, 14045120, 14045121, 14045121, 14045122, 14045123, 14045124, 14045125, 14045126, 14045127, 14045128, 14045129, 14045130, 14045131, 14045132, 14045133, 14045134, 14045135, 14045136, 14045137, 14045138, 14045139, 14045140, 14045141, 14045142, 14045143, 14045144, 14045145, 14045146, 14045147, 14045148, 14045150, 14045151, 14045152, 14045153, 14045154, 14045155, 14045156, 14045157, 14045158, 14045159, 14045160, 14045161, 14045162, 14045163, 14045164, 14045165, 14045166, 14045167, 14045168, 14045169, 14045170, 14045171, 14045172, 14045173, 14045174, 14045175, 14045176, 14045177, 14045178, 14045179, 14045180, 14045181, 14045182, 14045183, 14045184, 14045185, 14045186, 14045187, 14045188, 14045189, 14045190, 14045192, 14045193, 14045195, 14045196, 14045197, 14045198, 14045199, 14045200, 14045201, 14045202, 14045203, 14045204, 14045205, 14045206, 14045207, 14045208, 14045209, 14045210, 14045211, 14045212, 14045213, 14045214, 14045215, 14045216, 14045217, 14045218, 14045219, 14045220, 14045221, 14045222, 14045223, 14045224, 14045225, 14045227, 14045228, 14045230, 14045231, 14045232, 14045233, 14045234, 14045235, 14045236, 14045237, 14045238, 14045239, 14045240, 14045241, 14045242, 14045243, 14045245, 14045246, 14045247, 14045248, 14045249, 14045250, 14045251, 14045252, 14045253, 14045254, 14045255, 14045256, 14045257, 14045258, 14045259, 14045260, 14045261, 14045262, 14045263, 14045264, 14045265, 14045266, 14045267, 14045268, 14045269, 14045270, 14045271, 14045272, 14045273, 14045274, 14045275, 14045276, 14045277, 14045278, 14045279, 14045280, 14045281, 14045282, 14045283, 14045284, 14045285, 14045286, 14045287, 14045288, 14045289, 14045290, 14045291, 14045292, 14045293, 14045294, 14045295, 14045296, 14045298, 14045299, 14045300, 14045301, 14045302, 14045303, 14045304, 14045305, 14045306, 14045307, 14045308, 14045309, 14045310, 14045311, 14045312, 14045313, 14045314, 14045315, 14045316, 14045317, 14045318, 14045319, 14045320, 14045321, 14045322, 14045323, 14045324, 14045325, 14045326, 14045327, 14045328, 14045329, 14045330, 14045331, 14045332, 14045333, 14045334, 14045336, 14045337, 14045338, 14045339, 14045340, 14045341, 14045343, 14045344, 14045346, 14045347, 14045348, 14045349, 14045351, 14045352, 14045353, 14045354, 14045355, 14045356, 14045357, 14045358, 14045359, 14045360, 14045361, 14045362, 14045363, 14045364, 14045365, 14045366, 14045367, 14045368, 14045369, 14045370, 14045371, 14045372, 14045373, 14045374, 14045375, 14045376, 14045377, 14045378, 14045379, 14045380, 14045381, 14045382, 14045383, 14045384, 14045385, 14045386, 14045387, 14045388, 14045389, 14045390, 14045391, 14045392, 14045393, 14045394, 14045395, 14045396, 14045397, 14045398, 14045399, 14045400, 14045401, 14045402, 14045403, 14045404, 14045405, 14045406, 14045407, 14045408, 14045409, 14045410, 14045411, 14045412, 14045413, 14045414, 14045415, 14045416, 14045417, 14045418, 14045419, 14045420, 14045421, 14045422, 14045423, 14045424, 14045425, 14045426, 14045427, 14045428, 14045429, 14045430, 14045431, 14045432, 14045433, 14045434, 14045435, 14045436, 14045437, 14045438, 14045439, 14045440, 14045441, 14045442, 14045443, 14045444, 14045445, 14045446, 14045447, 14045448, 14045449, 14045450, 14045451, 14045452, 14045453, 14045454, 14045455, 14045456, 14045457, 14045458, 14045459, 14045460, 14045462, 14045465, 14045466, 14045467, 14045469, 14045470, 14045471, 14045472, 14045473, 14045474, 14045475, 14045480, 14045486, 14045488, 14045489, 14045490, 14045491, 14045492, 14045493, 14045494, 14045497, 14045499, 14045500, 14045501, 14045503, 14045504, 14045506, 14045508, 14045509, 14045510, 14045511, 14045512, 14045514, 14045515, 14045516, 14045517, 14045518, 14045519, 14045520, 14045521, 14045522, 14045523, 14045524, 14045525, 14045526, 14045527, 14045528, 14045529, 14045530, 14045531, 14045532, 14045533, 14045534, 14045535, 14045536, 14045537, 14045538, 14045541, 14045542, 14045543, 14045544, 14045545, 14045546, 14045547, 14045548, 14045549, 14045551, 14045552, 14045553, 14045554, 14045555, 14045556, 14045557, 14045558, 14045559, 14045560, 14045561, 14045562, 14045563, 14045564, 14045565, 14045566, 14045567, 14045568, 14045569, 14045570, 14045571, 14045572, 14045573, 14045574, 14045575, 14045576, 14045577, 14045578, 14045579, 14045580, 14045581, 14045582, 14045583, 14045584, 14045585, 14045586, 14045587, 14045588, 14045589, 14045590, 14045592, 14045593, 14045594, 14045595, 14045596, 14045597, 14045599, 14045600, 14045601, 14045602, 14045603, 14045604, 14045605, 14045606, 14045607, 14045608, 14045609, 14045611, 14045612, 14045613, 14045614, 14045615, 14045616, 14045617, 14045618, 14045619, 14045620, 14045621, 14045622, 14045623, 14045624, 14045625, 14045626, 14045627, 14045628, 14045629, 14045630, 14045631, 14045632, 14045633, 14045634, 14045635, 14045636, 14045637, 14045638, 14045639, 14045640, 14045641, 14045642, 14045643, 14045644, 14045645, 14045646, 14045647, 14045648, 14045649, 14045650, 14045651, 14045652, 14045653, 14045654, 14045655, 14045656, 14045657, 14045658, 14045659, 14045660, 14045661, 14045662, 14045663, 14045664, 14045665, 14045666, 14045667, 14045668, 14045669, 14045670, 14045671, 14045673, 14045674, 14045675, 14045676, 14045677, 14045678, 14045679, 14045680, 14045681, 14045682, 14045683, 14045684, 14045685, 14045686, 14045687, 14045688, 14045689, 14045690, 14045691, 14045692, 14045695, 14045696, 14045697, 14045698, 14045699, 14045700, 14045701, 14045702, 14045703, 14045704, 14045705, 14045706, 14045707, 14045708, 14045710, 14045711, 14045712, 14045713, 14045714, 14045717, 14045718, 14045719, 14045720, 14045721, 14045722, 14045723, 14045724, 14045726, 14045727, 14045728, 14045729, 14045730, 14045731, 14045732, 14045733, 14045734, 14045735, 14045736, 14045737, 14045738, 14045739, 14045740, 14045741, 14045742, 14045743, 14045744, 14045745, 14045746, 14045747, 14045749, 14045750, 14045751, 14045752, 14045753, 14045754, 14045755, 14045756, 14045757, 14045758, 14045759, 14045760, 14045761, 14045762, 14045763, 14045764, 14045765, 14045766, 14045767, 14045768, 14045769, 14045770, 14045771, 14045772, 14045773, 14045774, 14045775, 14045776, 14045777, 14045778, 14045779, 14045780, 14045781, 14045782, 14045783, 14045784, 14045785, 14045786, 14045787, 14045788, 14045789, 14045790, 14045791, 14045792, 14045793, 14045794, 14045795, 14045796, 14045797, 14045798, 14045799, 14045800, 14045801, 14045802, 14045803, 14045804, 14045805, 14045806, 14045807, 14045808, 14045809, 14045810, 14045811, 14045812, 14045813, 14045814, 14045815, 14045816, 14045817, 14045818, 14045819, 14045820, 14045821, 14045822, 14045823, 14045824, 14045825, 14045826, 14045827, 14045828, 14045829, 14045830, 14045831, 14045832, 14045833, 14045834, 14045835, 14045836, 14045837, 14045838, 14045839, 14045840, 14045841, 14045842, 14045843, 14045844, 14045845, 14045846, 14045847, 14045848, 14045849, 14045850, 14045851, 14045852, 14045853, 14045854, 14045855, 14045856, 14045857, 14045858, 14045859, 14045860, 14045861, 14045862, 14045863, 14045864, 14045865, 14045866, 14045867, 14045869, 14045870, 14045871, 14045872, 14045873, 14045874, 14045875, 14045876, 14045877, 14045878, 14045879, 14045880, 14045881, 14045882, 14045883, 14045884, 14045885, 14045886, 14045887, 14045888, 14045889, 14045890, 14045891, 14045892, 14045893, 14045894, 14045895, 14045896, 14045897, 14045898, 14045899, 14045900, 14045901, 14045902, 14045903, 14045904, 14045905, 14045906, 14045907, 14045908, 14045910, 14045911, 14045912, 14045913, 14045914, 14045915, 14045916, 14045918, 14045919, 14045920, 14045921, 14045922, 14045923, 14045924, 14045925, 14045926, 14045927, 14045928, 14045929, 14045936, 14045937, 14045938, 14045939, 14045940, 14045941, 14045942, 14045943, 14045945, 14045946, 14045947, 14045948, 14045949, 14045950, 14045951, 14045952, 14045953, 14045954, 14045955, 14045956, 14045957, 14045958, 14045959, 14045960, 14045961, 14045962, 14045963, 14045964, 14045965, 14045966, 14045967, 14045968, 14045969, 14045970, 14045971, 14045972, 14045973, 14045974, 14045975, 14045976, 14045977, 14045978, 14045979, 14045980, 14045981, 14045982, 14045983, 14045984, 14045985, 14045986, 14045987, 14045988, 14045991, 14045992, 14045994, 14045995, 14045996, 14045997, 14045998, 14045999, 14046000, 14046001, 14046002, 14046003, 14046004, 14046005, 14046006, 14046007, 14046008, 14046009, 14046010, 14046011, 14046012, 14046013, 14046014, 14046016, 14046017, 14046018, 14046019, 14046020, 14046021, 14046023, 14046024, 14046025, 14046026, 14046027, 14046028, 14046029, 14046030, 14046031, 14046032, 14046033, 14046034, 14046035, 14046036, 14046038, 14046039, 14046041, 14046042, 14046043, 14046044, 14046045, 14046046, 14046047, 14046048, 14046049, 14046050, 14046051, 14046052, 14046053, 14046054, 14046055, 14046056, 14046057, 14046058, 14046059, 14046060, 14046061, 14046062, 14046063, 14046064, 14046065, 14046066, 14046067, 14046068, 14046069, 14046070, 14046071, 14046072, 14046073, 14046074, 14046075, 14046076, 14046077, 14046078, 14046079, 14046080, 14046081, 14046082, 14046083, 14046084, 14046085, 14046086, 14046087, 14046088, 14046089, 14046090, 14046091, 14046092, 14046094, 14046095, 14046097, 14046098, 14046099, 14046100, 14046101, 14046102, 14046103, 14046104, 14046106, 14046107, 14046108, 14046109, 14046110, 14046111, 14046112, 14046113, 14046114, 14046116, 14046117, 14046119, 14046120, 14046121, 14046122, 14046123, 14046124, 14046125, 14046126, 14046127, 14046128, 14046129, 14046131, 14046132, 14046133, 14046135, 14046136, 14046137, 14046138, 14046140, 14046142, 14046143, 14046144, 14046145, 14046146, 14046147, 14046148, 14046149, 14046150, 14046151, 14046153, 14046154, 14046157, 14046158, 14046159, 14046165, 14046166, 14046168, 14046170, 14046171, 14046172, 14046173, 14046174, 14046175, 14046176, 14046177, 14046178, 14046180, 14046185, 14046186, 14046187, 14046188, 14046190, 14046207, 14046211, 14046216, 14046221, 14046226, 14046236, 14046238, 14046240, 14046241, 14046242, 14046246, 14046264, 14046267, 14046273, 14046274, 14046282, 14046283, 14046287, 14046288, 14046289, 14046294, 14046301, 14046303, 14046321, 14046325, 14046327, 14046329, 14046332, 14046336, 14046337, 14046342, 14046343, 14046344, 14046345, 14046346, 14046348, 14046349, 14046351, 14046353, 14046354, 14046356, 14046357, 14046358, 14046359, 14046360, 14046361, 14046362, 14046363, 14046364, 14046365, 14046366, 14046367, 14046368, 14046369, 14046370, 14046371, 14046373, 14046374, 14046375, 14046376, 14046377, 14046379, 14046380, 14046381, 14046383, 14046384, 14046385, 14046386, 14046387, 14046388, 14046389, 14046391, 14046393, 14046394, 14046395, 14046397, 14046399, 14046400, 14046401, 14046402, 14046403, 14046405, 14046406, 14046408, 14046409, 14046410, 14046412, 14046414, 14046415, 14046418, 14046419, 14046420, 14046424, 14046427, 14046428, 14046429, 14046430, 14046431, 14046432, 14046433, 14046434, 14046435, 14046436, 14046438, 14046439, 14046440, 14046442, 14046443, 14046444, 14046445, 14046446, 14046447, 14046448, 14046449, 14046450, 14046452, 14046454, 14046455, 14046456, 14046457, 14046458, 14046459, 14046460, 14046461, 14046462, 14046463, 14046464, 14046466, 14046467, 14046468, 14046469, 14046470, 14046471, 14046472, 14046473, 14046475, 14046476, 14046478, 14046482, 14046483, 14046485, 14046487, 14046488, 14046490, 14046492, 14046493, 14046494, 14046495, 14046496, 14046498, 14046503, 14046515, 14046516, 14046519, 14046520, 14046522, 14046523, 14046524, 14046529, 14046531, 14046533, 14046535, 14046536, 14046542, 14046543, 14046544, 14046545, 14046546, 14046547, 14046549, 14046550, 14046551, 14046552, 14046559, 14046571, 14046580, 14046584, 14046585, 14046586, 14046592, 14046593, 14046595, 14046596, 14046597, 14046598, 14046600, 14046602, 14046604, 14046607, 14046609, 14046610, 14046612, 14046614, 14046615, 14046620, 14046621, 14046626, 14046628, 14046629, 14046632, 14046633, 14046634, 14046640, 14046641, 14046642, 14046643, 14046644, 14046645, 14046647, 14046648, 14046649, 14046650, 14046651, 14046652, 14046654, 14046655, 14046657, 14046659, 14046660, 14046661, 14046662, 14046663, 14046664, 14046665, 14046666, 14046667, 14046668, 14046669, 14046670, 14046671, 14046672, 14046673, 14046675, 14046676, 14046677, 14046678, 14046679, 14046680, 14046681, 14046682, 14046683, 14046684, 14046685, 14046686, 14046687, 14046688, 14046690, 14046692, 14046693, 14046694, 14046695, 14046696, 14046697, 14046698, 14046699, 14046700, 14046702, 14046704, 14046705, 14046707, 14046708, 14046709, 14046710, 14046711, 14046712, 14046713, 14046714, 14046715, 14046716, 14046717, 14046718, 14046722, 14046723, 14046724, 14046725, 14046726, 14046727, 14046728, 14046729, 14046730, 14046731, 14046732, 14046733, 14046734, 14046735, 14046736, 14046737, 14046738, 14046739, 14046740, 14046741, 14046742, 14046743, 14046744, 14046745, 14046746, 14046747, 14046748, 14046749, 14046750, 14046751, 14046752, 14046753, 14046754, 14046755, 14046756, 14046757, 14046758, 14046759, 14046760, 14046761, 14046762, 14046763, 14046764, 14046765, 14046766, 14046767, 14046768, 14046769, 14046770, 14046771, 14046772, 14046773, 14046774, 14046775, 14046776, 14046777, 14046778, 14046779, 14046780, 14046781, 14046782, 14046783, 14046784, 14046785, 14046786, 14046787, 14046788, 14046789, 14046790, 14046792, 14046793, 14046794, 14046795, 14046796, 14046797, 14046798, 14046800, 14046801, 14046802, 14046803, 14046804, 14046805, 14046806, 14046807, 14046808, 14046809, 14046810, 14046811, 14046812, 14046813, 14046814, 14046815, 14046816, 14046817, 14046818, 14046819, 14046820, 14046821, 14046822, 14046823, 14046824, 14046825, 14046826, 14046827, 14046828, 14046829, 14046830, 14046831, 14046832, 14046834, 14046835, 14046836, 14046837, 14046838, 14046839, 14046840, 14046841, 14046842, 14046843, 14046844, 14046845, 14046846, 14046847, 14046848, 14046849, 14046850, 14046851, 14046852, 14046853, 14046855, 14046856, 14046858, 14046859, 14046860, 14046861, 14046863, 14046865, 14046866, 14046867, 14046869, 14046870, 14046871, 14046872, 14046873, 14046875, 14046876, 14046878, 14046879, 14046880, 14046881, 14046882, 14046883, 14046884, 14046885, 14046886, 14046887, 14046888, 14046889, 14046890, 14046891, 14046892, 14046893, 14046894, 14046895, 14046896, 14046897, 14046899, 14046900, 14046901, 14046902, 14046903, 14046904, 14046905, 14046906, 14046907, 14046908, 14046909, 14046910, 14046911, 14046912, 14046913, 14046914, 14046915, 14046916, 14046917, 14046918, 14046919, 14046920, 14046921, 14046922, 14046923, 14046924, 14046925, 14046926, 14046927, 14046928, 14046929, 14046930, 14046931, 14046932, 14046934, 14046935, 14046936, 14046937, 14046938, 14046939, 14046943, 14046945, 14046946, 14046947, 14046948, 14046949, 14046950, 14046951, 14046952, 14046954, 14046955, 14046956, 14046958, 14046959, 14046960, 14046961, 14046962, 14046963, 14046964, 14046965, 14046966, 14046967, 14046968, 14046969, 14046970, 14046971, 14046972, 14046973, 14046974, 14046975, 14046977, 14046978, 14046979, 14046980, 14046981, 14046982, 14046983, 14046984, 14046985, 14046986, 14046987, 14046988, 14046989, 14046990, 14046991, 14046992, 14046993, 14046994, 14046995, 14046996, 14046997, 14046998, 14046999, 14047000, 14047001, 14047002, 14047003, 14047004, 14047005, 14047006, 14047007, 14047008, 14047009, 14047010, 14047011, 14047012, 14047013, 14047014, 14047015, 14047016, 14047018, 14047019, 14047020, 14047021, 14047022, 14047023, 14047024, 14047026, 14047027, 14047028, 14047029, 14047030, 14047031, 14047032, 14047034, 14047035, 14047036, 14047038, 14047039, 14047040, 14047041, 14047042, 14047043, 14047045, 14047046, 14047047, 14047048, 14047049, 14047050, 14047051, 14047052, 14047053, 14047055, 14047056, 14047057, 14047058, 14047060, 14047061, 14047062, 14047063, 14047064, 14047065, 14047066, 14047067, 14047068, 14047069, 14047070, 14047071, 14047072, 14047073, 14047074, 14047075, 14047076, 14047077, 14047078, 14047079, 14047080, 14047081, 14047082, 14047083, 14047084, 14047085, 14047086, 14047087, 14047088, 14047089, 14047090, 14047091, 14047092, 14047093, 14047094, 14047095, 14047096, 14047097, 14047098, 14047099, 14047100, 14047101, 14047102, 14047103, 14047104, 14047105, 14047106, 14047107, 14047108, 14047109, 14047110, 14047111, 14047112, 14047113, 14047114, 14047115, 14047116, 14047117, 14047118, 14047119, 14047120, 14047121, 14047122, 14047123, 14047124, 14047125, 14047126, 14047127, 14047128, 14047129, 14047130, 14047131, 14047133, 14047134, 14047135, 14047136, 14047137, 14047138, 14047140, 14047143, 14047145, 14047147, 14047148, 14047150, 14047151, 14047152, 14047153, 14047154, 14047155, 14047156, 14047158, 14047159, 14047160, 14047163, 14047164, 14047165, 14047166, 14047167, 14047168, 14047169, 14047170, 14047171, 14047172, 14047173, 14047174, 14047175, 14047176, 14047178, 14047179, 14047180, 14047181, 14047182, 14047184, 14047185, 14047186, 14047187, 14047188, 14047189, 14047190, 14047191, 14047192, 14047193, 14047194, 14047195, 14047196, 14047197, 14047198, 14047199, 14047200, 14047201, 14047202, 14047203, 14047204, 14047205, 14047207, 14047208, 14047209, 14047210, 14047211, 14047212, 14047213, 14047214, 14047215, 14047216, 14047217, 14047219, 14047220, 14047221, 14047222, 14047222, 14047223, 14047224, 14047225, 14047226, 14047227, 14047228, 14047229, 14047230, 14047231, 14047233, 14047234, 14047238, 14047239, 14047240, 14047241, 14047242, 14047243, 14047244, 14047245, 14047247, 14047248, 14047249, 14047250, 14047251, 14047252, 14047253, 14047254, 14047255, 14047256, 14047257, 14047258, 14047259, 14047260, 14047263, 14047264, 14047266, 14047267, 14047320, 14047321, 14047322, 14047323, 14047324, 14047325, 14047326, 14047327, 14047328, 14047331, 14047332, 14047333, 14047334, 14047335, 14047336, 14047337, 14047338, 14047339, 14047340, 14047341, 14047342, 14047343, 14047344, 14047345, 14047346, 14047347, 14047348, 14047349, 14047350, 14047351, 14047352, 14047353, 14047355, 14047356, 14047358, 14047360, 14047361, 14047362, 14047363, 14047364, 14047365, 14047366, 14047367, 14047369, 14047370, 14047371, 14047372, 14047373, 14047374, 14047375, 14047380, 14047383, 14047385, 14047386, 14047387, 14047388, 14047389, 14047391, 14047392, 14047393, 14047394, 14047395, 14047396, 14047397, 14047399, 14047400, 14047401, 14047402, 14047403, 14047404, 14047405, 14047407, 14047408, 14047409, 14047410, 14047411, 14047413, 14047414, 14047415, 14047416, 14047417, 14047418, 14047420, 14047421, 14047422, 14047423, 14047424, 14047425, 14047426, 14047427, 14047428, 14047429, 14047430, 14047431, 14047432, 14047433, 14047434, 14047435, 14047436, 14047437, 14047438, 14047439, 14047440, 14047441, 14047442, 14047443, 14047444, 14047445, 14047446, 14047447, 14047448, 14047449, 14047450, 14047451, 14047452, 14047453, 14047454, 14047455, 14047456, 14047457, 14047458, 14047459, 14047460, 14047461, 14047463, 14047464, 14047465, 14047466, 14047467, 14047468, 14047469, 14047470, 14047471, 14047472, 14047473, 14047474, 14047475, 14047476, 14047477, 14047479, 14047480, 14047481, 14047482, 14047483, 14047484, 14047485, 14047486, 14047487, 14047488, 14047489, 14047490, 14047491, 14047492, 14047493, 14047494, 14047495, 14047496, 14047497, 14047498, 14047499, 14047500, 14047501, 14047502, 14047503, 14047504, 14047505, 14047506, 14047507, 14047508, 14047509, 14047510, 14047511, 14047512, 14047514, 14047515, 14047516, 14047517, 14047518, 14047519, 14047520, 14047521, 14047522, 14047523, 14047524, 14047525, 14047526, 14047527, 14047528, 14047529, 14047530, 14047531, 14047532, 14047533, 14047534, 14047535, 14047536, 14047537, 14047538, 14047539, 14047540, 14047541, 14047542, 14047543, 14047544, 14047545, 14047546, 14047547, 14047548, 14047549, 14047550, 14047551, 14047552, 14047553, 14047554, 14047555, 14047556, 14047557, 14047558, 14047559, 14047560, 14047561, 14047562, 14047564, 14047565, 14047566, 14047568, 14047569, 14047570, 14047571, 14047572, 14047573, 14047574, 14047575, 14047576, 14047577, 14047578, 14047579, 14047580, 14047581, 14047582, 14047584, 14047585, 14047586, 14047587, 14047588, 14047589, 14047590, 14047591, 14047592, 14047671, 14047672, 14047673, 14047674, 14047675, 14047676, 14047677, 14047678, 14047679, 14047680, 14047681, 14047682, 14047683, 14047684, 14047685, 14047686, 14047687, 14047688, 14047689, 14047690, 14047691, 14047692, 14047693, 14047694, 14047695, 14047696, 14047697, 14047698, 14047699, 14047700, 14047701, 14047702, 14047704, 14047705, 14047706, 14047707, 14047708, 14047709, 14047710, 14047711, 14047712, 14047713, 14047714, 14047715, 14047716, 14047717, 14047718, 14047719, 14047720, 14047721, 14047723, 14047724, 14047725, 14047727, 14047728, 14047729, 14047730, 14047731, 14047732, 14047733, 14047734, 14047735, 14047736, 14047737, 14047738, 14047739, 14047740, 14047742, 14047743, 14047744, 14047745, 14047746, 14047747, 14047748, 14047749, 14047750, 14047751, 14047752, 14047754, 14047755, 14047756, 14047757, 14047759, 14047760, 14047761, 14047762, 14047763, 14047764, 14047765, 14047766, 14047767, 14047768, 14047769, 14047770, 14047771, 14047772, 14047773, 14047774, 14047775, 14047776, 14047777, 14047778, 14047779, 14047780, 14047781, 14047782, 14047783, 14047784, 14047785, 14047786, 14047787, 14047789, 14047790, 14047791, 14047792, 14047793, 14047794, 14047795, 14047796, 14047797, 14047798, 14047799, 14047800, 14047801, 14047803, 14047804, 14047805, 14047806, 14047807, 14047809, 14047810, 14047812, 14047813, 14047814, 14047815, 14047816, 14047818, 14047820, 14047822, 14047823, 14047825, 14047826, 14047827, 14047829, 14047831, 14047832, 14047833, 14047834, 14047837, 14047838, 14047840, 14047841, 14047842, 14047843, 14047845, 14047846, 14047848, 14047850, 14047851, 14047852, 14047854, 14047855, 14047858, 14047859, 14047860, 14047862, 14047863, 14047864, 14047868, 14047871, 14047881, 14047882, 14047883, 14047885, 14047886, 14047888, 14047891, 14047893, 14047894, 14047895, 14047905, 14047907, 14047908, 14047910, 14047911, 14047913, 14047916, 14047918, 14047922, 14047924, 14047925, 14047926, 14047927, 14047928, 14047929, 14047931, 14047932, 14047933, 14047934, 14047935, 14047938, 14047942, 14047943, 14047944, 14047945, 14047947, 14047948, 14047949, 14047950, 14047952, 14047953, 14047954, 14047958, 14047959, 14047960, 14047961, 14047963, 14047964, 14047968, 14047969, 14047971, 14047972, 14047973, 14047977, 14047978, 14047979, 14047980, 14047982, 14047983, 14047984, 14047985, 14047986, 14047987, 14047988, 14047989, 14047991, 14047992, 14047993, 14047994, 14047995, 14047996, 16412201, 16412202, 16412203, 16412204, 16412205, 16412206, 16412207, 16412208, 16412209, 16412210, 16412211, 16412212, 16412213, 16412214, 16412215, 16412216, 16412217, 16412218, 16412219, 16412220, 16412222, 16412223, 16412224, 16412225, 16412226, 16412227, 16412228, 16412229, 16412230, 16412231, 16412232, 16412233, 16412234, 16412235, 16412236, 16412237, 16412238, 16412239, 16412240, 16412241, 16412242, 16412243, 16412244, 16412245, 16412246, 16412247, 16412248, 16412249, 16412250, 16412251, 16412252, 16412253, 16412254, 16412255, 16412256, 16412257, 16412258, 16412259, 16412260, 16412261, 16412262, 16412263, 16412264, 16412265, 16412266, 16412267, 16412268, 16412269, 16412270, 16412271, 16412272, 16412273, 16412274, 16412275, 16412276, 16412277, 16412278, 16412279, 16412280, 16412281, 16412282, 16412283, 16412285, 16412286, 16412287, 16412288, 16412289, 16412290, 16412291, 16412292, 16412293, 16412294, 16412295, 16412297, 16412298, 16412299, 16412300, 16412301, 16412302, 16412303, 16412304, 16412305, 16412306, 16412307, 16412308, 16412309, 16412310, 16412311, 16412312, 16412313, 16412314, 16412315, 16412316, 16412317, 16412318, 16412319, 16412320, 16412321, 16412322, 16412323, 16412324, 16412325, 16412326, 16412327, 16412328, 16412329, 16412330, 16412331, 16412332, 16412333, 16412334, 16412335, 16412336, 16412337, 16412338, 16412339, 16412340, 16412341, 16412342, 16412343, 16412344, 16412345, 16412346, 16412347, 16412349, 16412350, 16412351, 16412352, 16412353, 16412354, 16412355, 16412356, 16412357, 16412358, 16412359, 16412360, 16412361, 16412362, 16412363, 16412364, 16412365, 16412366, 16412367, 16412368, 16412369, 16412370, 16412371, 16412372, 16412373, 16412374, 16412375, 16412376, 16412377, 16412378, 16412379, 16412380, 16412381, 16412382, 16412383, 16412384, 16412385, 16412386, 16412387, 16422201, 16422202, 16422203, 16422204, 16422205, 16422207, 16422208, 16422209, 16422210, 16422211, 16422212, 16422213, 16422214, 16422215, 16422216, 16422217, 16422218, 16422219, 16422220, 16422221, 16422222, 16422223, 16422224, 16422225, 16422226, 16422227, 16422228, 16422229, 16422230, 16422231, 16422232, 16422233, 16422234, 16422235, 16422236, 16422237, 16422238, 16422239, 16422240, 16422241, 16422242, 16422243, 16422244, 16422245, 16422246, 16422247, 16422248, 16422249, 16422250, 16422251, 16422252, 16422253, 16422254, 16422255, 16422256, 16422257, 16422258, 16422259, 16422260, 16422261, 16422262, 16422263, 16422264, 16422265, 16422266, 16422267, 16422268, 16422269, 16422270, 16422271, 16422272, 16422273, 16422274, 16422275, 16422276, 16422277, 16422278, 16422279, 16422280, 16422281, 16422282, 16422283, 16422284, 16422285, 16422286, 16422287, 16422288, 16422289, 16422290, 16422291, 16422292, 16422293, 16422294, 16422295, 16422296, 16422297, 16422298, 16422299, 16422300, 16422301, 16422302, 16422303, 16422304, 16422305, 16422306, 16422307, 16422308, 16422309, 16422310, 16422311, 16422312, 16422313, 16422314, 16422315, 16422316, 16422317, 16422318, 16422319, 16422321, 16422324, 16422327, 16422331, 16422334, 16422337, 16422341, 16422344, 16422347, 16422351, 16422354, 16422357, 16422361, 16422364, 16422367, 16422371, 16422374, 16422377, 16422381, 16422384, 16422387, 16422391, 16422394, 16422397, 16422401, 16422404, 16422407, 16422411, 16422414, 16422417, 16422421, 16422424, 16422427, 16422431, 16422434, 16422437, 16422441, 16422444, 16422447, 16422451, 16422454, 16422457, 16422461, 16422464, 16422467, 16422471, 16422474, 16422477, 16422481, 16422484, 16422487, 16422491, 16422494, 16422497, 16422501, 16422504, 16422507, 16422511, 16422514, 16422517, 16422521, 16422524, 16422527, 16422531, 16422534, 16422537, 16422541, 16422544, 16422547, 16422551, 16422554, 16422557, 16422561, 16422564, 16422567, 16422571, 16422574, 16422577, 16422581, 16422584, 16422587, 16422591, 16422594, 16422597, 16422601, 16422604, 16422607, 16422611, 16422614, 16422617, 16422621, 16422624, 16422627, 16422631, 16422634, 16422637, 16422641, 16422644, 16422647, 16422651, 16422654, 16422657, 16422661, 16422664, 16422667, 16422671, 16422674, 16422677, 16422681, 16422684, 16422687, 16422691, 16422694, 16422697, 16422701, 16794201, 16794202, 16794204, 16794205, 16794206, 16794207, 16794209, 16794210, 16794211, 16794212, 16794213, 16794214, 16794216, 16794217, 16794218, 16794220, 16794221, 16794222, 16794224, 16794225, 16794226, 16794227, 16794228, 16794230, 16794231, 16794232, 16794233, 16794234, 16794235, 16794236, 16794237, 16794238, 16794239, 16794240, 16794241, 16794242, 16794244, 16794246, 16794248, 16794249, 16794250, 16794252, 16794253, 16794254, 16794255, 16794256, 16794257, 16794259, 16794260, 16794262, 16794263, 16794264, 16794265, 16794266, 16794267, 16794269, 16794270, 16794271, 16794275, 16794276, 16794277, 16794278, 16794279, 16794280, 16794282, 16794284, 16794285, 16794286, 16794287, 16794289, 16794291, 16794292, 16794293, 16794294, 16794295, 16794296, 16794297, 16794299, 16794300, 16794302, 16794303, 16794304, 16794305, 16794306, 16794307, 16794309, 16794311, 16794312, 16794313, 16794316, 16794317, 16794318, 16794319, 16794320, 16794321, 16794322, 16794323, 16794324, 16794325, 16794326, 16794327, 16794328, 16794330, 16794331, 16794332, 16794333, 16794336, 16794337, 16794338, 16794339, 16794342, 16794344, 16794345, 16794346, 16794348, 16794349, 16794350, 16794351, 16794355, 16794356, 16794357, 16794358, 16794360, 16794361, 16794362, 16794363, 16794365, 16794366, 16794367, 16794368, 16794369, 16794370, 16794371, 16794372, 16794373, 16794374, 16794375, 16794377, 16794378, 16794379, 16794380, 16794381, 16794383, 16794385, 16794386, 16794388, 16794389, 16794390, 16794391, 16794392, 16794393, 16794394, 16794395, 16794396, 16794397, 16794398, 16794399, 16794400, 16794401, 16794402, 16794403, 16794404, 16794405, 16794407, 16794408, 16794409, 16794410, 16794411, 16794413, 16794414, 16794415, 16794416, 16794418, 16794419, 16794421, 16794423, 16794424, 16794425, 16794426, 16794427, 16794428, 16794429, 16794430, 16794433, 16794434, 16794436, 16794437, 16794440, 16794441, 16794442, 16794443, 16794444, 16794445, 16794447, 16794448, 16794449, 16794450, 16794451, 16794452, 16794454, 16794455, 16794456, 16794457, 16794458, 16794459, 16794460, 16794461, 16794462, 16794463, 16794464, 16794465, 16794466, 16794467, 16794468, 16794469, 16794470, 16794471, 16794473, 16794474, 16794475, 16794477, 16794478, 16794479, 16794480, 16794482, 16794483, 16794484, 16794485, 16794486, 16794487, 16794488, 16794489, 16794490, 16794491, 16794492, 16794493, 16794495, 16794497, 16794498, 16794499, 16794500, 16794501, 16794502, 16794503, 16794504, 16794505, 16794506, 16794507, 16794509, 16794510, 16794511, 16794512, 16794513, 16794514, 16794515, 16794521, 16794523, 16794525, 16794529, 16794530, 16794532, 16794536, 16794540, 16794541, 16794542, 16794543, 16794544, 16794545, 16794546, 16794552, 16794556, 16794561, 16794566, 16794567, 16794570, 16794573, 16794576, 16794577, 16794586, 16794587, 16794589, 16794590, 16794592, 16794594, 16794596, 16794597, 16794598, 16794604, 16794605, 16794607, 16794608, 16794613, 16794616, 16794628, 16794632, 16794635, 16794637, 16794638, 16794643, 16794644, 16794648, 16794653, 16794656, 16794657, 16794665, 16794668, 16794675, 16794676, 16794679, 16794683, 16794684, 16794688, 16794702, 16794703, 16794704, 16794705, 16794708, 16795201, 16795204, 16795207, 16795211, 16795214, 16795217, 16795221, 16795224, 16795227, 16795231, 16795234, 16795237, 16795241, 16795244, 16795247, 16795251, 16795254, 16795257, 16795261, 16795264, 16795267, 16795271, 16795274, 16795277, 16795281, 16795297, 16795301, 16795304, 16795307, 16795311, 16795314, 16795317, 16795324, 16795327, 16795331, 16795334, 16795341, 16795344, 16795351, 16795354, 16795357, 16795361, 16795364, 16795367, 16795371, 16795377, 16795381, 16795384, 16795387, 16795391, 16795394, 16795397, 16795401, 16795407, 16795411, 16795414, 16795417, 16795421, 16795424, 16795427, 16795434, 16795441, 16795444, 16795451, 16795454, 16795457, 16795461, 16795464, 16795467, 16795471, 16795477, 16795481, 16795484, 16795487, 16795491, 16795501, 16795507, 16795511, 16795521, 16838202, 16838203, 16838204, 16838205, 16838207, 16838208, 16838209, 16838210, 16838211, 16838212, 16838213, 16838214, 16838216, 16838217, 16838218, 16838219, 16838220, 16838221, 16838222, 16838223, 16838224, 16838225, 16838226, 16838227, 16838228, 16838229, 16838230, 16838231, 16838232, 16838233, 16838234, 16838235, 16838236, 18627408, 18627409, 18627410, 18627414, 18627415, 18627417, 18627418, 18627419, 18627420, 18627421, 18627422, 18627423, 18627424, 18627425, 18627426, 18627428, 18627429, 18627430, 18627431, 18627432, 18627435, 18627436, 18627439, 18627440, 18627442, 18627445, 18627447, 18627448, 18627449, 18627450, 18627451, 18627452, 18627453, 18627454, 18627455, 18627456, 18627459, 18627460, 18627461, 18627462, 18627463, 18627464, 18627465, 18627466, 18627467, 18627468, 18627469, 18627470, 18627471, 18627472, 18627474, 18627475, 18627476, 18627477, 18627478, 18627479, 18627481, 18627482, 18627483, 18627484, 18627485, 18627487, 18627488, 18627489, 18627490, 18627491, 18627492, 18627493, 18627494, 18627495, 18627496, 18627497, 18627498, 18627499, 18627500, 18627501, 18627502, 18627505, 18627507, 18627508, 18627509, 18627510, 18627511, 18627512, 18627517, 18627522, 18627525, 18627526, 18627530, 18627532, 18627533, 18627537, 18627542, 18627543, 18627544, 18627547, 18627554, 18627556, 18627558, 18627559, 18627563, 18627566, 18627572, 18627573, 18627581, 18627586, 18627589, 18627599, 18627607, 18627610, 18627612, 18627622, 18627629, 18627643, 18627644, 18627653, 18627661, 18627670, 18627687, 18627694, 18627695, 18627703, 18627704, 18627706, 18627708, 18627709, 18627710, 18627711, 18627713, 18627714, 18627715, 18627717, 18627718, 18627719, 18627720, 18627721, 18627723, 18627725, 18627728, 18627730, 18627731, 18627732, 18627733, 18627736, 18627745, 18627748, 18627749, 18627752, 18627754, 18627757, 18627759, 18627761, 18627762, 18627763, 18627765, 18627766, 18627767, 18627776, 18627777, 18627782, 18627783, 18627784, 18627787, 18627791, 18627801, 18627806, 18627810, 18627818, 18627819, 18627821, 18627822, 18627823, 18627824, 18627825, 18627826, 18627829, 18627830, 18627831, 18627832, 18627834, 18627836, 18627837, 18627840, 18627865, 18627867, 18627870, 18627871, 18627872, 18627874, 18627876, 18627877, 18627878, 18627880, 18627882, 18627884, 18627886, 18627891, 18627892, 18627897, 18627899, 18627901, 18627904, 18627905, 18627906, 18627908, 18627909, 18627910, 18627911, 18627912, 18627915, 18627916, 18627917, 18627918, 18627919, 18627922, 18627926, 18627927, 18627928, 18627930, 18627933, 18627935, 18627938, 18627939, 18627940, 18627941, 18627942, 18627943, 18627944, 18627945, 18627946, 18627947, 18627949, 18627950, 18627951, 18627952, 18627953, 18627954, 18627956, 18627957, 18627959, 18627960, 18627961, 18627962. |
| FEI Number |
3013319212
|
Recalling Firm/ Manufacturer |
Hospira Inc. 275 N Field Dr Lake Forest IL 60045-2579
|
| For Additional Information Contact | Chris Eustace 224-212-2000 |
Manufacturer Reason for Recall | The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Hospira sent an Urgent Device Field Correction letter, dated February 6, 2013, to customers. Customers were requested to return a reply form to Stericycle indicating that they have received the letter, and notified their users and/or customers. The customers were asked to supply the number of Plum A+ infusers at their facility. Product will be corrected by Hospira in the field. |
| Quantity in Commerce | 357,778 total |
| Distribution | United States nationwide distribution including PR, and USVI. Costa Rica, Turks and Caicos, and Dominican Republic. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = FRN
|
|
|
|