• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Philips Healthcare Ingenuity Core128

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Philips Healthcare Ingenuity Core128see related information
Date Initiated by FirmJuly 20, 2016
Create DateAugust 25, 2016
Recall Status1 Terminated 3 on July 02, 2018
Recall NumberZ-2658-2016
Recall Event ID 74820
510(K)NumberK033326 
Product Classification System, x-ray, tomography, computed - Product Code JAK
ProductIngenuity Core128, Model number 728323 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Code Information 728323 -- Ingenuity Core128  320004 320053 320056 
Recalling Firm/
Manufacturer
Philips Medical Systems (Cleveland) Inc
595 Miner Rd
Cleveland OH 44143-2131
For Additional Information ContactMrs. Holly Wright Lee
440-483-2015
Manufacturer Reason
for Recall
After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
FDA Determined
Cause 2
Software Design Change
ActionCustomers were sent a letter dated July 28, 2016 informing them of the problem, under what circumstances it can occur and the actions that the customer can take to minimize the effect of the problem. The letter also includes the actions planned by Philips to correct the problem. Philips plans on releasing a field change order (FCO) to correct the affected systems. If you need any further information or support concerning this issue, contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1 -800-722-9377).
Quantity in Commerce3
DistributionUS Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = JAK
-
-