• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Spacelabs Healthcare Xhibit Central Station

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Spacelabs Healthcare Xhibit Central Stationsee related information
Date Initiated by FirmAugust 25, 2016
Date PostedSeptember 26, 2016
Recall Status1 Terminated 3 on March 13, 2017
Recall NumberZ-2885-2016
Recall Event ID 75078
510(K)NumberK122146 
Product Classification Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
ProductSpacelabs Healthcare Xhibit Central Station, Model 96102, Version 1.1.2. Xhibit Telemetry Receivers (XTR) must be connected to an Xhibit Central Station for the display of patient vital signs.
Code Information Serial numbers of units distributed in the US   SLA0216F01822, SLA0216F01827, SLA0416F01924, SLA0516F01935, SLA0216F01771, SLA0216F01773, SLA0216F01784, SLA0216F01787, SLA0416F01897, SLA0416F01898, SLA0416F01899, SLA0416F01905, SLA0416F01909, SLA0416F01910, SLA0416F01911, SLA0416F01912, SLA0416F01914, SLA0416F01915, SLA0416F01916, SLA0416F01917, SLA0416F01918, SLA0416F01919, SLA0416F01920, SLA0516F01971, SLA0516F01995, SLA0314F00534, SLA0314F00535, SLA0314F00548, SLA0314F00561, SLA0314F00567, SLA0715F01310, SLA0814F00713, SLA1115F01593, SLA1115F01595, SLA1115F01596, SLA1115F01597, SLA1115F01602, SLA1115F01609, SLA0216F01806, SLA0216F01810, SLA0216F01817, SLA0216F01818, SLA0216F01832, SLA0216F01833, SLA0216F01835, SLA0216F01839, SLA0515F01104, SLA0515F01105, SLA0515F01107, SLA0515F01111, SLA0515F01115, SLA0515F01116, SLA0515F01119, SLA0515F01120, SLA0515F01121, SLA0515F01123, SLA0515F01129, SLA0515F01142, SLA0515F01144, SLA0515F01145, SLA0515F01146, SLA0515F01147, SLA0515F01148, SLA0515F01149, SLA0515F01150, SLA0515F01151, SLA0515F01152, SLA0515F01153, SLA0515F01154, SLA0515F01155, SLA0515F01156, SLA0515F01157, SLA0515F01158, SLA0716F02010, SLA0115F00919, SLA0115F00921, SLA0115F00927, SLA0115F00928, SLA0115F00934, SLA0515F01106, SLA0515F01108, SLA0515F01143, SLA0715F01327, SLA0715F01328, SLA0715F01329, SLA0915F01439, SLA0915F01442, SLA0915F01447, SLA0915F01451, SLA0915F01452, SLA0915F01457, SLA0915F01459, SLA0915F01468, SLA0915F01474, SLA0915F01496, SLA0915F01497, SLA0915F01498, SLA0915F01499, SLA0915F01501, SLA0915F01503, SLA0915F01506, SLA0915F01507, SLA0915F01508, SLA0915F01510, SLA0915F01511, SLA0915F01512, SLA0915F01513, SLA0915F01514, SLA0115F00926, SLA0115F00963, SLA0115F00964, SLA0115F00986, SLA0116F01711, SLA0214F00497, SLA0216F01796, SLA0216F01808, SLA0216F01811, SLA0216F01812, SLA0216F01834, SLA0216F01843, SLA0216F01845, SLA0516F01948, SLA0516F01950, SLA0516F01951, SLA1015F01536, SLA1015F01539, SLA1215F01636, SLA1215F01637, SLA1215F01638, SLA1215F01645, SLA1215F01646, SLA1215F01651, SLA1215F01652, SLA1215F01654, SLA1215F01657, SLA1215F01659, SLA1215F01695, SLA0715F01346, SLA0715F01347, SLA0815F01371, SLA0815F01372, SLA0815F01373, SLA0815F01374, SLA0815F01375, SLA0815F01376, SLA0815F01379, SLA0815F01380, SLA0815F01381, SLA0815F01382, SLA0815F01383, SLA0815F01384, SLA0815F01386, SLA0815F01387, SLA0815F01388, SLA0815F01390, SLA0815F01391, SLA0815F01395, SLA0815F01397, SLA0815F01400, SLA0815F01402, SLA0815F01404, SLA0815F01405, SLA0815F01406, SLA0815F01407, SLA0815F01408, SLA0815F01409, SLA0815F01412, SLA0815F01415, SLA0815F01418, SLA0815F01419, SLA0815F01424, SLA0114F00458, SLA0214F00484, SLA0214F00486, SLA0214F00527, SLA0416F01856, SLA0516F01967, SLA0516F01969, SLA0516F01970, SLA0516F01972, SLA0516F01974, SLA0516F01975, SLA0516F01976, SLA0516F01977, SLA0516F01981, SLA0516F01982, SLA0516F01985, SLA0516F01986, SLA0516F01987, SLA0516F01992, SLA0516F01996, SLA0516F02002, SLA0516F02005, SLA0615F01232, SLA0116F01722, SLA0915F01462, SLA0915F01463, SLA0216F01842, SLA0114F00445, SLA0114F00446, SLA0114F00447, SLA0116F01718, SLA0116F01720, SLA0414F00568, SLA0414F00585, SLA1214F00858, SLA1214F00871, SLA0615F01254, SLA0615F01255, SLA0615F01256, SLA0615F01262, SLA0615F01263, SLA1113F00355, SLA1113F00356, SLA1113F00357, SLA1113F00358, SLA1113F00363, SLA1113F00364, SLA1113F00365, SLA0615F01237, SLA0615F01239, SLA1015F01540, SLA1015F01551, SLA0416F01870, SLA0715F01314, SLA0716F02030, SLA0815F01426, SLA0115F00917, SLA0115F00918, SLA0115F00923, SLA0115F00924, SLA0115F00961, SLA0115F00968, SLA0115F00969, SLA0115F00972, SLA0115F00976, SLA0314F00529, SLA0314F00530, SLA0314F00531, SLA0314F00532, SLA0314F00536, SLA0314F00537, SLA0314F00538, SLA0314F00540, SLA0314F00542, SLA0314F00543, SLA0314F00544, SLA0314F00546, SLA0314F00547, SLA0314F00549, SLA0314F00550, SLA0314F00552, SLA0314F00553, SLA0314F00556, SLA0314F00565, SLA0416F01889, SLA0416F01895, SLA0216F01823, SLA0216F01828, SLA1015F01537, SLA0915F01471, SLA0915F01475, SLA0615F01188, SLA0615F01194, SLA0615F01197, SLA0615F01198, SLA0615F01199, SLA0615F01213, SLA0416F01866, SLA0416F01869, SLA0416F01881, SLA0515F01124, SLA0515F01125, SLA0515F01127, SLA0515F01128, SLA0515F01133, SLA0515F01137, SLA0416F01913, SLA1015F01534, SLA0416F01923, SLA0715F01334, SLA1015F01525, SLA1015F01528, SLA1015F01530, SLA1015F01558, SLA1015F01560, SLA1015F01535, SLA0715F01316, SLA0715F01317, SLA0715F01326, SLA0614F00655, SLA0116F01702, SLA0216F01783, SLA0216F01791, SLA0216F01794, SLA1015F01544, SLA0116F01692, SLA0516F01944, SLA0516F01949, SLA0516F01952, SLA0516F01957, SLA0516F01958, SLA0516F01959, SLA0516F01960, SLA0516F01961, SLA0516F01963, SLA1115F01576, SLA0416F01901, SLA0416F01904, SLA0416F01908, SLA0715F01275, SLA0715F01279, SLA0216F01824, SLA0216F01807, SLA0216F01809, SLA0416F01850, SLA0416F01860, SLA0416F01875, SLA0216F01732, SLA0216F01797, SLA0216F01800, SLA0216F01801, SLA0216F01802, SLA0815F01394, SLA0115F00948, SLA0115F00950, SLA0315F01047, SLA0416F01874, SLA0515F01159, SLA0515F01160, SLA0515F01161, SLA0515F01162, SLA0515F01163, SLA0515F01164, SLA0515F01165, SLA0515F01166, SLA0515F01167, SLA0515F01168, SLA0515F01169, SLA0515F01170, SLA0515F01171, SLA0515F01172, SLA0515F01173, SLA0515F01174, SLA0515F01175, SLA0515F01176, SLA0515F01177, SLA0515F01178, SLA0515F01179, SLA0515F01180, SLA0515F01181, SLA0515F01182, SLA0515F01183, SLA0614F00618, SLA0614F00619, SLA0614F00623, SLA0614F00624, SLA0614F00625, SLA0614F00627, SLA0614F00628, SLA0614F00629, SLA0614F00631, SLA0614F00633, SLA0614F00635, SLA0614F00636, SLA0614F00637, SLA0614F00641, SLA0614F00642, SLA0614F00644, SLA0614F00645, SLA0614F00650, SLA0614F00654, SLA0615F01226, SLA0615F01229, SLA0615F01231, SLA0615F01233, SLA0615F01234, SLA0615F01235, SLA0615F01236, SLA0615F01246, SLA0615F01248, SLA0714F00678, SLA0714F00687, SLA0715F01323, SLA0715F01324, SLA0715F01325, SLA0815F01331, SLA0815F01403, SLA0714F00688, SLA0714F00694, SLA0714F00696, SLA0714F00702, SLA0714F00703, SLA0516F01934, SLA0516F01962, SLA0614F00638, SLA0115F00955, SLA0216F01837, SLA0216F01840, SLA0216F01844, SLA1214F00876, SLA1214F00881, SLA1214F00885, SLA1214F00886, SLA0116F01693, SLA0116F01700, SLA0116F01707, SLA0116F01721, SLA0116F01729, SLA0416F01855, SLA0416F01891, SLA0416F01896, SLA1215F01622, SLA1215F01623, SLA1215F01625, SLA1215F01627, SLA1215F01629, SLA1215F01644, SLA1215F01789, SLA0116F01691, SLA0116F01699, SLA0116F01703, SLA0216F01772, SLA0216F01774, SLA0216F01775, SLA0216F01776, SLA0216F01778, SLA0216F01779, SLA0216F01780, SLA0216F01803, SLA0416F01861, SLA0416F01862, SLA0416F01863, SLA0416F01879, SLA0516F01941, SLA0516F01942, SLA0516F01943, SLA0516F01946, SLA0516F01980, SLA0516F01984, SLA0615F01212, SLA1014F00755, SLA1014F00760, SLA1014F00762, SLA1014F00767, SLA1014F00772, SLA1014F00773, SLA1014F00774, SLA1014F00775, SLA1014F00778, SLA1014F00779, SLA1014F00782, SLA1115F01572, SLA0216F01813, SLA0216F01821, SLA0416F01853, SLA0416F01884, SLA0416F01907, SLA0516F01945, SLA0516F01965, SLA0814F00707, SLA1115F01590, SLA1115F01600, SLA1115F01607, SLA0214F00489, SLA0615F01240, SLA0615F01244, SLA0615F01261, SLA0915F01493, SLA0516F01936, SLA0516F01953, SLA0516F01954, SLA0715F01274, SLA0116F01714, SLA0116F01715, SLA0116F01725, SLA0116F01726, SLA0216F01768, SLA0216F01769, SLA0216F01770, SLA0216F01777, SLA0216F01781, SLA0216F01788, SLA0216F01804, SLA0216F01805, SLA0115F00931, SLA0115F00937, and SLA0515F01131.   Serial Numbers of units distributed OUTSIDE the US   SLA0114F00430, SLA0114F00431, SLA0114F00432, SLA0114F00433, SLA0114F00436, SLA0114F00437, SLA0114F00438, SLA0114F00439, SLA0114F00440, SLA0114F00443, SLA0114F00450, SLA0114F00459, SLA0115F00951, SLA0115F00975, SLA0116F01701, SLA0116F01723, SLA0116F01724, SLA0214F00496, SLA0214F00500, SLA0214F00502, SLA0214F00505, SLA0214F00507, SLA0214F00508, SLA0214F00513, SLA0214F00517, SLA0214F00519, SLA0214F00523, SLA0214F00526, SLA0216F01733, SLA0216F01793, SLA0216F01795, SLA0216F01798, SLA0216F01814, SLA0216F01815, SLA0216F01816, SLA0216F01820, SLA0216F01825, SLA0216F01830, SLA0216F01838, SLA0314F00528, SLA0314F00541, SLA0414F00578, SLA0414F00579, SLA0414F00580, SLA0414F00581, SLA0414F00582, SLA0414F00586, SLA0414F00587, SLA0414F00595, SLA0414F00596, SLA0414F00605, SLA0414F00606, SLA0414F00607, SLA0414F00608, SLA0414F00609, SLA0414F00611, SLA0414F00612, SLA0416F01849, SLA0416F01865, SLA0416F01868, SLA0416F01877, SLA0416F01878, SLA0416F01880, SLA0416F01882, SLA0416F01890, SLA0416F01892, SLA0416F01893, SLA0416F01906, SLA0416F01925, SLA0512F00183, SLA0512F00185, SLA0512F00186, SLA0512F00187, SLA0512F00188, SLA0512F00189, SLA0512F00190, SLA0512F00191, SLA0512F00192, SLA0512F00195, SLA0512F00196, SLA0512F00203, SLA0512F00208, SLA0512F00209, SLA0512F00211, SLA0516F01926, SLA0516F01927, SLA0516F01928, SLA0516F01929, SLA0516F01930, SLA0516F01931, SLA0516F01932, SLA0516F01937, SLA0516F01938, SLA0516F01939, SLA0516F01940, SLA0516F01947, SLA0516F01964, SLA0516F01997, SLA0612F00217, SLA0612F00219, SLA0612F00221, SLA0612F00223, SLA0612F00224, SLA0612F00226, SLA0612F00227, SLA0612F00228, SLA0612F00229, SLA0612F00230, SLA0612F00232, SLA0612F00233, SLA0612F00235, SLA0612F00236, SLA0612F00237, SLA0612F00238, SLA0612F00240, SLA0612F00241, SLA0612F00245, SLA0612F00248, SLA0612F00249, SLA0612F00251, SLA0612F00253, SLA0612F00255, SLA0612F00256, SLA0612F00263, SLA0612F00264, SLA0612F00266, SLA0612F00268, SLA0612F00272, SLA0612F00273, SLA0614F00647, SLA0615F01186, SLA0615F01189, SLA0615F01195, SLA0615F01203, SLA0615F01205, SLA0615F01206, SLA0615F01207, SLA0615F01208, SLA0615F01215, SLA0615F01216, SLA0615F01218, SLA0615F01220, SLA0615F01221, SLA0615F01222, SLA0615F01224, SLA0615F01225, SLA0615F01227, SLA0615F01228, SLA0615F01230, SLA0714F00667, SLA0714F00668, SLA0714F00669, SLA0714F00670, SLA0714F00671, SLA0714F00672, SLA0714F00673, SLA0714F00674, SLA0714F00675, SLA0714F00676, SLA0714F00677, SLA0714F00679, SLA0714F00680, SLA0714F00681, SLA0714F00682, SLA0714F00683, SLA0714F00684, SLA0714F00690, SLA0714F00691, SLA0714F00692, SLA0714F00693, SLA0714F00695, SLA0714F00697, SLA0714F00698, SLA0714F00699, SLA0714F00700, SLA0714F00701, SLA0714F00706, SLA0715F01270, SLA0715F01271, SLA0715F01273, SLA0715F01276, SLA0715F01277, SLA0715F01278, SLA0715F01281, SLA0715F01282, SLA0715F01283, SLA0715F01289, SLA0715F01292, SLA0715F01298, SLA0715F01299, SLA0715F01300, SLA0715F01303, SLA0715F01309, SLA0715F01311, SLA0715F01312, SLA0715F01318, SLA0715F01319, SLA0715F01320, SLA0715F01333, SLA0715F01335, SLA0715F01343, SLA0715F01344, SLA0715F01345, SLA0715F01368, SLA0716F02024, SLA0716F02025, SLA0716F02032, SLA0716F02046, SLA0814F00710, SLA0814F00716, SLA0814F00731, SLA0814F00742, SLA0814F00745, SLA0815F01417, SLA0815F01423, SLA0815F01428, SLA0815F01433, SLA0915F01441, SLA0915F01445, SLA0915F01454, SLA0915F01456, SLA0915F01464, SLA0915F01467, SLA0915F01480, SLA0915F01481, SLA0915F01484, SLA0915F01485, SLA0915F01486, SLA0915F01487, SLA0915F01488, SLA0915F01489, SLA0915F01492, SLA0915F01495, SLA0915F01515, SLA0915F01517, SLA1014F00754, SLA1014F00757, SLA1014F00759, SLA1014F00764, SLA1014F00766, SLA1014F00769, SLA1014F00770, SLA1014F00776, SLA1014F00777, SLA1014F00780, SLA1014F00787, SLA1014F00791, SLA1014F00792, SLA1014F00793, SLA1015F01523, SLA1015F01526, SLA1015F01533, SLA1015F01545, SLA1015F01549, SLA1015F01550, SLA1015F01561, SLA1113F00348, SLA1113F00360, SLA1113F00361, SLA1113F00368, SLA1113F00373, SLA1113F00382, SLA1114F00831, SLA1115F01573, SLA1115F01580, SLA1115F01585, SLA1115F01589, SLA1115F01591, SLA1115F01601, SLA1115F01604, SLA1115F01605, SLA1213F00386, SLA1213F00389, SLA1213F00390, SLA1213F00395, SLA1213F00396, SLA1213F00397, SLA1213F00399, SLA1213F00408, SLA1213F00413, SLA1213F00419, SLA1214F00838, SLA1214F00854, SLA1214F00863, SLA1214F00867, SLA1214F00875, SLA1214F00879, SLA1214F00880, SLA1214F00883, SLA1214F00888, SLA1214F00892, SLA1214F00895, SLA1214F00900, SLA1214F00901, SLA1214F00904, SLA1214F00913, SLA1214F00914, SLA1215F01621, SLA1215F01626, SLA1215F01628, SLA1215F01631, SLA1215F01633, SLA1215F01642, SLA1215F01647, SLA1215F01648, SLA1215F01653, SLA1215F01655, SLA1215F01658, SLA123F00393, and SLA515F01141.   
Recalling Firm/
Manufacturer
Spacelabs Healthcare Inc
35301 SE Center St
Snoqualmie WA 98065-9216
For Additional Information ContactTechnical Support
800-522-7025 Ext. 2
Manufacturer Reason
for Recall
The firm received multiple reports of Xhibit Telemetry System going offline or locking up. In addition, short-duration asystole alarms may not display an audio and visual alarm as reported from a single Canadian hospital.
FDA Determined
Cause 2
Software design
ActionSpacelabs sent a letter dated August 25, 2016 to affected US consignees informing them of the going offline or locking up issue when Xhibit Central Station is used as part of an Xhibit Telemetry System. On September 1, 2016, a customer letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record. Spacelabs sent the second Urgent - Medical Device Correction letter, dated September 15, 2016, to affected US consignees informing them that they have received reports of certain short-duration asystole alarms generated by the Xhibit Telemetry Receiver (XTR) where there may be no audible or visual alarm. On September 22, 2016, a second letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record. Spacelabs Healthcare will contact consignees to schedule a convenient time for Spacelabs to update the recalled Xhibit Telemetry System at no cost. Customers with Xhibit Telemetry Receivers will get all Xhibit Central Stations updated regardless if connected to an Xhibit Telemetry Receiver. Consignees are directed to technical support for additional information or technical assistance: Technical Support Spacelabs Healthcare, Inc 35301 SE Center St. Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support
Quantity in Commerce798 units total (488 units in the US and 310 units outside US)
DistributionWorldwide distribution. US nationwide including Puerto Rico, Australia, Canada, China, France, Germany, Ireland, Italy, Netherlands, Philippines, Poland, Romania, Switzerland, and United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = MHX
-
-