• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Adult AirLife singlelimb

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall Adult AirLife singlelimb see related information
Date Initiated by Firm November 29, 2014
Date Posted January 24, 2018
Recall Status1 Terminated 3 on February 05, 2021
Recall Number Z-0345-2018
Recall Event ID 78636
510(K)Number K160764  
Product Classification Humidifier, respiratory gas, (direct patient interface) - Product Code BTT
Product Adult Air-Life single-limb circuit (Model AH202) with AirLife Humidification Chamber, Model Number AH290.

The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.
Code Information 0001115270 0001121993 0001124929 0001 124930 0001 124932 0001129063 0001 129064 0001131424 0001131425 0001 134347 0001134348 0001135401 0001135402 0001137252 0001137253 0001139266 0001139267 0001145147 0001145148 0001145149 0001156843
Recalling Firm/
Manufacturer
Vyaire Medical
75 N Fairway Dr
Vernon Hills IL 60061-1845
Manufacturer Reason
for Recall
Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.
FDA Determined
Cause 2
Process control
Action The firm, Vyaire Medical, sent an "URGENT RECALL NOTIFICATION" letter dated November 29th, 2017 to its customers. The recall letter described the product, problem and actions to be taken. The customers were instructed to do the following: "ALL CUSTOMERS: ~ Inspect current inventory on-hand. A 100% physical inventory should immediately be performed to identify and remove affected lots of the AirLife Humidification Chamber(s) and Heated Breathing Circuit Kit(s) due to the identified potential patient safety risk. ~ Destroy all affected product(s) in-stock in accordance with your facility's destruction protocol. ~ If you are a healthcare facility or distributor that has further distributed affected product you must further notify your customers of this URGENT RECALL NOTIFICATION. CUSTOMERS THAT HAVE PURCHASED DIRECT FROM VYAIRE: ~ Complete the Customer Response Form (Appendix 2) and return to GMB-GLBALFieldActions@CareFusion.com ~ If you wish to receive credit of your in-stock product affected by this recall or to place an order for the replacement Model/Part Numbers (as referenced per #6. Per the FAQ's) please contact Vyaire Medical's Customer Service Department at 800.323.9088 (Option #1, followed by Option #1) Monday-Friday 8:00 am CST - 5:00 pm CST. *Please have your original PO number available when calling to expedite the credit process. CUSTOMERS THAT HAVE PURCHASED THROUGH A DISTRIBUTOR: ~ Please work with your distributor to receive credit of your in-stock product affected by this recall. For any additional questions and support concerning this voluntary recall, please contact Clinical Risk Coordinator, at 827-757-0109 or lindy.schenning@vyaire.com.
Quantity in Commerce 16,670 units in total
Distribution US Nationwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = BTT and Original Applicant = CAREFUSION 2200, INC.
-
-