| | Class 2 Device Recall ACCOLADE MRI DR Pacemaker |  |
| Date Initiated by Firm | December 07, 2017 |
| Create Date | January 15, 2018 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0358-2018 |
| Recall Event ID |
78787 |
| PMA Number | P150012 |
| Product Classification |
Implantable pulse generator, pacemaker (non-CRT) - Product Code LWP
|
| Product | ACCOLADE MRI DR Pacemaker |
| Code Information |
Model L311 |
| FEI Number |
2124215
|
Recalling Firm/ Manufacturer |
Boston Scientific Corporation
4100 Hamline Ave N Bldg 3
Saint Paul
MN
55112-5700
|
| For Additional Information Contact | Technical Services 800-227-3422 |
Manufacturer Reason for Recall | Boston Scientific has received reports of intermittent over-sensing of the Minute Ventilation (MV) sensor signal with certain Boston Scientific pacemaker and cardiac resynchronization therapy pacemaker systems (pacemakers). |
FDA Determined Cause 2 | Device Design |
| Action | On December 19, 2017 the firm sent a letter to all their medical consignees stating the following:
December 2017
Dear Doctor,
Boston Scientific has received reports of intermittent oversensing of the Minute Ventilation (MV) sensor signal with certain Boston Scientific pacemaker and cardiac resynchronization therapy pacemaker systems (pacemakers). MV sensor signal oversensing may cause pre-syncope or syncope due to periods of pacing inhibition. This MV behavior may occur with any manufacturers pacing lead system, but Boston Scientific has determined it to be more likely for affected Boston Scientific pacemakers using Medtronic or Abbott/St. Jude
(Abbott) leads implanted in either the right atrium (RA) or right ventricle (RV).
Boston Scientific is actively developing a software update designed to automatically detect and resolve this MV sensor signal oversensing behavior. We anticipate submitting the software update to Regulatory Agencies in
March 2018 and pending approval, will release it as soon as possible thereafter. Until this software update is available, Boston Scientific has additional re commendations to mitigate this risk for affected pacemaker systems.
Root Cause Investigation
The MV sensor in Boston Scientific pacemakers can be used for RightRate" (rate adaptive pacing), Respiratory Rate Trend, or AP Scan"1. When the RA/RV pacing leads and lead terminal connections are operating as
intended, the MV sensor signal is appropriately filtered and therefore is not detected by the pacemaker or displayed on electrograms (EGMs). However, intermittency related to the lead or pacemaker-lead connection2 has
the potential to create a transient high impedance condition. A high impedance condition may subsequently alter the MV sensor signal such that it becomes visible on EGMs and potentially subject to oversensing on the RA or
RV channels. For a technical description of the Boston Scientifics MV sensor, please refer to Appendix A.
Engineering analysis and |
| Distribution | AK, AL, AR, AZ, CA , CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Puerto Rico
Worldwide |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = LWP
|
|
|
|