  Class 1 Device Recall AirLife Resuscitation see related information
Date Initiated by Firm April 27, 2017
Date Posted May 10, 2018
Recall Status1 Terminated 3 on May 09, 2019
Recall Number Z-1559-2018
Recall Event ID 78412
Product Classification Cardiopulmonary resuscitation aid kit - Product Code OEV
Product AirLife Resuscitation, Infant, with mask, 40inch oxygen reservoir tubing, pressure-relief valve, CO2 monitor, monometer

Product Usage:
The AirLife Resuscitation devices are manual resuscitation devices that are used to provide positive pressure ventilation to patients who are not breathing or cannot adequately breathe on their own.
Code Information Part Number: 2K8010. Lot: 0001007295, 0001015414, 0001018697, 0001022519, 0001022518, 0001034530, 0001040652, 0001040653, 0001043846, 0001043847, 0001044102, 0001064454, 0001067498
Recalling Firm/
Manufacturer
Vyaire Medical
75 N Fairway Dr
Vernon Hills IL 60061-1845
For Additional Information Contact Lindy Schenning
872-757-0109
Manufacturer Reason
for Recall
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
FDA Determined
Cause 2
Component design/selection
Action The firm, Vyaire Medical, sent an "URGENT RECALL NOTICE" letter on 04/27/2017 and again on 06/08/2017 for the expansion of the recall to its Customers. The letter described the product, problem and actions to be taken. The customers were instructed to inspect current inventory on-hand and immediately perform a 1OO% physical inventory of the product to identify and be removed; complete and return the Customer Response Form by Fax to: (312) 949.0972 or email to:GMB-GLB-VSFieldActions@CareFusion.com.; destroy all affected product in accordance with your facility's destruction protocol; and for replacements, contact your distributor Armstrong Medical direct at (800)323.4220 x129. For any additional questions and support concerning this voluntary recall, please contact Customer Advocacy Analyst at (224) 706.6829 or email: Kristina.Scheppa@CareFusion.com.
Quantity in Commerce 8,124 units
Distribution Worldwide Distribution - US (Nationwide) and in the countries of Canada, Puerto Rico, United Arab Emirates, Philippines, and Europe.