• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Artis zee

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Artis zee see related information
Date Initiated by Firm May 01, 2018
Create Date June 27, 2018
Recall Status1 Terminated 3 on May 16, 2019
Recall Number Z-2310-2018
Recall Event ID 80286
510(K)Number K141574  
Product Classification Interventional fluoroscopic x-ray system - Product Code OWB
Product Artis zee ceiling, Material no. 10094137, for angiography and whole body radiographic/fluoroscopic procedures.





Code Information 147763 147573 147608 147760 147636 147630 147208 147514 147165 147558 147304 146946 147585 147553 147645 147087 147203 147601 147659 147675 147516 147700 147745 147681 147795 147773 147789 147541 147577 147029 147625 147635 146152 147660 147753 147083 146861 147561 147771 147657 147364 147775 147790 147781 147734 146896 147559 147620 147287 147749 147633 147737 147621 147250 147253 147252 147251 147717 147778 147643 147670 147234 147557 147661 147235 147716 147739 147743 147752 147528 147687 146867 147186 
Recalling Firm/
Manufacturer
Siemens Medical Solutions USA, Inc
40 Liberty Blvd
Malvern PA 19355-1418
For Additional Information Contact Ms. Meredith Adams
610-219-4834
Manufacturer Reason
for Recall
After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, the system cannot be operated normally. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.
FDA Determined
Cause 2
Device Design
Action On May 1, 2018, the firm notified customers via Urgent Medical Device Correction letters. Users were warned that the Large Display may not show an image after it returns from power save mode. The Large Display stays black without showing an error message although X-Ray is still possible. The problem occurs sporadically, and does not occur during an ongoing procedure. The firm stated that there should be no influence on the previous treatment of patients. Users were instructed to follow standard emergency procedures in the event of system failures. It is recommended to consider reinforcing emergency procedures until the update has been completed. The firm's service organization will contact customers to arrange a date to correct the error by updating the firmware of the affected devices. Users were asked to promptly notify and instruct accordingly all staff within their organizations who need to be aware of the issue.
Quantity in Commerce 73
Distribution Worldwide distribution. US nationwide including Puerto Rico, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Luxembourg, Mexico, Myanmar, Netherlands, New Zealand, Norway, China, Paraguay, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = OWB and Original Applicant = SIEMENS MEDICAL SOLUTIONS USA, INC.
-
-