• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Incubator, Neonatal

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Incubator, Neonatalsee related information
Date Initiated by FirmSeptember 25, 2018
Create DateNovember 19, 2018
Recall Status1 Terminated 3 on July 13, 2020
Recall NumberZ-0479-2019
Recall Event ID 81224
510(K)NumberK152814 
Product Classification Incubator, neonatal - Product Code FMZ
ProductGiraffe OmniBed Carestation. Neonatal incubator. Device sold under the following product names: Carestation and Carestation CS1
Code Information Model Number, UDI (Serial Number): 2082844-001-526493, 010084068211686221TABX6226911180201 (TABX62269); 2082844-001-526493, 010084068211686221TABX6215911180122 (TABX62159); 2082844-001-526493, 010084068211686221TABX6219811180129 (TABX62198); 2082844-001-526493, 010084068211686221TABX6227011180201 (TABX62270); 2082844-001-526493, 010084068211686221TABX6229111180202 (TABX62291); 2082844-001-526493, 010084068211686221TABX6215611180122 (TABX62156); 2082844-001-526493, 010084068211686221TABX6226011180201 (TABX62260); 2082844-001-526493, 010084068211686221TABX6213511180119 (TABX62135); 2082844-001-526493, 010084068211686221TABX6218911180124 (TABX62189); 2082844-001-526493, 010084068211686221TABX6219611180124 (TABX62196); 2082844-001-526493, 010084068211686221TABX6218811180124 (TABX62188); 2082844-001-526493, 010084068211686221TABX6216011180122 (TABX62160); 2082844-001-487119, To be provided (TABV60550); 2082844-001-526493, 010084068211686221TABW6398711171213 (TABW63987); 2082844-001-526493, 010084068211686221TABX6214111180119 (TABX62141); 2082844-001-526493, 010084068211686221TABW6409911171219 (TABW64099); 2082844-001-526493, 010084068211686221TABW6400911171213 (TABW64009); 2082844-001-526493, 010084068211686221TABW6396611171211 (TABW63966); 2082844-001-526493, 010084068211686221TABW6400811171213 (TABW64008); 2082844-001-526493, 010084068211686221TABX6215411180122 (TABX62154); 2082844-001-526493, 010084068211686221TABW6408611171219 (TABW64086); 2082844-001-526493, 010084068211686221TABX6210811180117 (TABX62108); 2082844-001-526493, 010084068211686221TABX6213011180119 (TABX62130); 2082844-001-526493, 010084068211686221TABW6406911171219 (TABW64069); 2082844-001-526493, 010084068211686221TABW6408411171219 (TABW64084); 2082844-001-526493, 010084068211686221TABW6404311171218 (TABW64043); 2082844-001-526493, 010084068211686221TABW6409811171219 (TABW64098); 2082844-001-526493, 010084068211686221TABW6396411171211 (TABW63964); 2082844-001-526493, 010084068211686221TABW6404911171218 (TABW64049); 2082844-001-526493, 010084068211686221TABW6395411171211 (TABW63954); 2082844-001-526493, 010084068211686221TABW6408711171219 (TABW64087); 2082844-001-487119, Not Applicable (TABV60078); 2082844-001-526493, 010084068211686221TABX6232011180207 (TABX62320); 2082844-001-526493, 010084068211686221TABX6227211180201 (TABX62272); 2082844-001-526493, 010084068211686221TABX6211811180119 (TABX62118); 2082844-001-526493, 010084068211686221TABX6214611180119 (TABX62146); 2082844-001-526493, 010084068211686221TABW6398311171213 (TABW63983); 2082844-001-526493, 010084068211686221TABW6399411171213 (TABW63994); 2082844-001-526493, 010084068211686221TABX6220011180129 (TABX62200); 2082844-001-526493, 010084068211686221TABW6396211171211 (TABW63962); 2082844-001-526493, 010084068211686221TABX6227611180201 (TABX62276); 2082844-001-487119, Not Applicable (TABV60092); 2082844-001-526493, 010084068211686221TABW6393811171211 (TABW63938); 2082844-001-526493, 010084068211686221TABW6399511171213 (TABW63995); 2082844-001-526493, 010084068211686221TABX6213111180119 (TABX62131); 2082844-001-526493, 010084068211686221TABX6225711180201 (TABX62257); 2082844-001-526493, 010084068211686221TABX6214011180119 (TABX62140); 2082844-001-526493, 010084068211686221TABX6226411180201 (TABX62264); 2082844-001-526493, 010084068211686221TABX6231911180207 (TABX62319);  2082844-001-526493, 010084068211686221TABX6230111180202 (TABX62301);  2082844-001-526493, 010084068211686221TABX6228711180202 (TABX62287);  2082844-001-526493, 010084068211686221TABX6229611180202 (TABX62296);  2082844-001-526493, 010084068211686221TABX6230011180202 (TABX62300);  2082844-001-526493, 010084068211686221TABX6233311180207 (TABX62333);  2082844-001-526493, 010084068211686221TABX6233111180207 (TABX62331);  2082844-001-526493, 010084068211686221TABX6229011180202 (TABX62290);  2082844-001-526493, 010084068211686221TABX6229311180202 (TABX62293);  2082844-001-526493, 010084068211686221TABX6216311180123 (TABX62163);  2082844-001-526493, 010084068211686221TABX6223511180130 (TABX62235);  2082844-001-526493, 010084068211686221TABX6222711180130 (TABX62227);  2082844-001-526493, 010084068211686221TABX6216811180123 (TABX62168);  2082844-001-526493, 010084068211686221TABX6216411180123 (TABX62164);  2082844-001-526493, 010084068211686221TABX6214911180122 (TABX62149);  2082844-001-526493, 010084068211686221TABX6217811180123 (TABX62178);  2082844-001-526493, 010084068211686221TABX6220611180129 (TABX62206);  2082844-001-526493, 010084068211686221TABX6216511180123 (TABX62165);  2082844-001-526493, 010084068211686221TABX6225011180201 (TABX62250);  2082844-001-526493, 010084068211686221TABX6217311180123 (TABX62173);  2082844-001-526493, 010084068211686221TABX6216211180123 (TABX62162);  2082844-001-526493, 010084068211686221TABX6216611180123 (TABX62166);  2082844-001-526493, 010084068211686221TABX6223411180130 (TABX62234);  2082844-001-526493, 010084068211686221TABX6220811180129 (TABX62208);  2082844-001-526493, 010084068211686221TABW6395211171211 (TABW63952);  2082844-001-526493, 010084068211686221TABW6395111171211 (TABW63951);  2082844-001-526493, 010084068211686221TABW6407811171219 (TABW64078);  2082844-001-526493, 010084068211686221TABX6211011180117 (TABX62110);  2082844-001-526493, 010084068211686221TABX6211311180117 (TABX62113);  2082844-001-526493, 010084068211686221TABX6233211180207 (TABX62332);  2082844-001-526493, 010084068211686221TABX6228211180201 (TABX62282);  2082844-001-526493, 010084068211686221TABX6228311180201 (TABX62283);  2082844-001-526493, 010084068211686221TABW6404611171218 (TABW64046);  2082844-001-526493, 010084068211686221TABW6400011171213 (TABW64000);  2082844-001-526493, 010084068211686221TABW6409511171219 (TABW64095);  2082844-001-526493, 010084068211686221TABW6409611171219 (TABW64096);  2082844-001-526493, 010084068211686221TABW6393311171211 (TABW63933);  2082844-001-526493, 010084068211686221TABX6232411180207 (TABX62324);  2082844-001-526493, 010084068211686221TABX6229211180202 (TABX62292);  2082844-001-526493, 010084068211686221TABX6225411180201 (TABX62254);  2082844-001-526493, 010084068211686221TABX6225511180201 (TABX62255);  2082844-001-526493, 010084068211686221TABX6225811180201 (TABX62258);  2082844-001-487119, Not Applicable (TABV60433);  2082844-001-526493, 010084068211686221TABX6219711180124 (TABX62197);  2082844-001-526493, 010084068211686221TABX6212811180119 (TABX62128);  2082844-001-526493, 010084068211686221TABW6394411171211 (TABW63944);  2082844-001-526493, 010084068211686221TABW6398211171213 (TABW63982);  2082844-001-526493, 010084068211686221TABW6394811171211 (TABW63948);  2082844-001-526493, 010084068211686221TABW6394011171211 (TABW63940);  2082844-001-526493, 010084068211686221TABX6222611180130 (TABX62226);  2082844-001-526493, 010084068211686221TABW6393711171211 (TABW63937);  2082844-001-526493, 010084068211686221TABX6224011180130 (TABX62240);  2082844-001-526493, 010084068211686221TABX6223911180130 (TABX62239);  2082844-001-526493, 010084068211686221TABW6399611171213 (TABW63996);  2082844-001-526493, 010084068211686221TABX6224311180130 (TABX62243);  2082844-001-526493, 010084068211686221TABX6224211180130 (TABX62242);  2082844-001-526493, 010084068211686221TABX6224411180201 (TABX62244);  2082844-001-526493, 010084068211686221TABX6222911180130 (TABX62229);  2082844-001-526769, 010084068211686221TABW6397711171212 (TABW63977);  2082844-001-526493, 010084068211686221TABW6401311171213 (TABW64013);  2082844-001-526493, 010084068211686221TABW6395911171211 (TABW63959);  2082844-001-526493, 010084068211686221TABW6408211171219 (TABW64082);  2082844-001-526493, 010084068211686221TABW6402311171213 (TABW64023);  2082844-001-526493, 010084068211686221TABW6396511171211 (TABW63965);  2082844-001-526493, 010084068211686221TABX6233511180207 (TABX62335);  2082844-001-526493, 010084068211686221TABX6229511180202 (TABX62295);  2082844-001-526493, 010084068211686221TABX6224611180201 (TABX62246);  2082844-001-526493, 010084068211686221TABX6225311180201 (TABX62253);  2082844-001-526493, 010084068211686221TABX6222311180130 (TABX62223);  2082844-001-526493, 010084068211686221TABX6225111180201 (TABX62251);  2082844-001-526493, 010084068211686221TABX6225211180201 (TABX62252);  2082844-001-526493, 010084068211686221TABX6223311180130 (TABX62233);  2082844-001-526493, 010084068211686221TABX6223211180130 (TABX62232);  2082844-001-526493, 010084068211686221TABX6223611180130 (TABX62236);  2082844-001-526493, 010084068211686221TABX6224811180201 (TABX62248);  2082844-001-526493, 010084068211686221TABX6223811180130 (TABX62238);  2082844-001-526493, 010084068211686221TABX6224911180201 (TABX62249);  2082844-001-526493, 010084068211686221TABX6224511180201 (TABX62245);  2082844-001-526493, 010084068211686221TABX6210911180117 (TABX62109);  2082844-001-526493, 010084068211686221TABX6219411180124 (TABX62194);  2082844-001-526493, 010084068211686221TABX6219311180124 (TABX62193);  2082844-001-526493, 010084068211686221TABX6213911180119 (TABX62139);  2082844-001-526493, 010084068211686221TABX6216111180122 (TABX62161);  2082844-001-526493, 010084068211686221TABX6213611180119 (TABX62136);  2082844-001-526493, 010084068211686221TABX6212011180119 (TABX62120);  2082844-001-526493, 010084068211686221TABX6215811180122 (TABX62158);  2082844-001-526493, 010084068211686221TABX6213211180119 (TABX62132);  2082844-001-526493, 010084068211686221TABX6217111180123 (TABX62171);  2082844-001-526493, 010084068211686221TABX6221811180130 (TABX62218);  2082844-001-526493, 010084068211686221TABW6405911171219 (TABW64059);  2082844-001-526493, 010084068211686221TABX6224711180201 (TABX62247);  2082844-001-526493, 010084068211686221TABX6223111180130 (TABX62231);  2082844-001-403566, Not Applicable (TABU67011);  2082844-001-499499, 010084068211686221TABW6259811170606 (TABW62598);  2082844-001-430326, Not Applicable (TABV60027);  2082844-001-430326, Not Applicable (TABV60030);  2082844-001-430326, Not Applicable (TABV60028);  2082844-001-430326, Not Applicable (TABU67004);  2082844-001-462725, Not Applicable (TABU67040);  2082844-001-462725, Not Applicable (TABU67039);  2082844-001-462725, Not Applicable (TABU67056);  2082844-001-462725, Not Applicable (TABU67041);  2082844-001-462725, Not Applicable (TABU67053);  2082844-001-462725, Not Applicable (TABU67045);  2082844-001-462725, Not Applicable (TABU67047);  2082844-001-462725, Not Applicable (TABU67050);  2082844-001-462725, Not Applicable (TABU67043);  2082844-001-460023, Not Applicable (TABU67014);  2082844-001-510290, 010084068211685521TABV61236160902 (TABV61236);  2082844-001-558377, 010084068211686221TABX6238011180219 (TABX62380);  2082844-001-534467, 010084068211686221TABX6330711180628 (TABX63307);  2082844-001-534467, 010084068211686221TABX6330611180628 (TABX63306);  2082844-001-534467, 010084068211686221TABX6331411180628 (TABX63314);  2082844-001-534467, 010084068211686221TABX6236011180213 (TABX62360);  2082844-001-534467, 010084068211686221TABX6331211180628 (TABX63312);  2082844-001-534467, 010084068211686221TABX6331111180628 (TABX63311);  2082844-001-534467, 010084068211686221TABX6330911180628 (TABX63309);  2082844-001-534467, 010084068211686221TABX6331511180628 (TABX63315);  2082844-001-501649, 010084068211686221TABX6330411180626 (TABX63304);  2082844-001-501649, 010084068211686221TABX6330311180626 (TABX63303);  2082844-001-501649, 010084068211686221TABX6330511180626 (TABX63305);  2082844-001-467192, Not Applicable (TABV60033);  2082844-001-495642, Not Applicable (TABV60053);  2082844-001-508846, 010084068211686221TABX6247711180301 (TABX62477);  2082844-001-508846, 010084068211686221TABX6247611180301 (TABX62476);  2082844-001-508846, 010084068211686221TABX6246211180228 (TABX62462);  2082844-001-508846, 010084068211686221TABX6246311180228 (TABX62463);  2082844-001-508846, 010084068211686221TABX6248411180305 (TABX62484);  2082844-001-508846, 010084068211686221TABX6248511180305 (TABX62485);  2082844-001-498552, 010084068211686221TABX6248911180305 (TABX62489);  2082844-001-498552, 010084068211686221TABX6248611180305 (TABX62486);  2082844-001-498552, 010084068211686221TABX6248711180305 (TABX62487);  2082844-001-498552, 010084068211686221TABX6248811180305 (TABX62488);  2082844-001-508846, 010084068211686221TABX6332211180629 (TABX63322);  2082844-001-508846, 010084068211686221TABX6332111180629 (TABX63321);  2082844-001-508846, 010084068211686221TABX6221111180130 (TABX62211);  2082844-001-508846, 010084068211686221TABX6221211180130 (TABX62212);  2082844-001-485099, Not Applicable (TABV60013);  2082844-001-485099, Not Applicable (TABV60102);  2082844-001-485099, Not Applicable (TABV60055);  2082844-001-485099, Not Applicable (TABV60054);  2082844-001-485099, Not Applicable (TABV60065);  2082844-001-485099, Not Applicable (TABV60015);  2082844-001-485099, Not Applicable (TABV60072);  2082844-001-485099, Not Applicable (TABV60069);  2082844-001-485099, Not Applicable (TABV60016);  2082844-001-512022, 010084068211686221TABX6332411180702 (TABX63324);  2082844-001-531627, 010084068211686221TABX6254211180309 (TABX62542);  2082844-001-512022, 010084068211686221TABX6254111180307 (TABX62541);  2082844-001-495786, Not Applicable (TABV60056);  2082844-001-565598, 010084068211686221TABX63177180605 (TABX63177);  2082844-001-565598, 010084068211686221TABX63176180605 (TABX63176);  2082844-001-565598, 010084068211686221TABX63175180605 (TABX63175);  2082844-001-535110, 010084068211686221TABX6332011180629 (TABX63320);  2082844-001-534358, 010084068211686221TABX6237111180215 (TABX62371);  2082844-001-534358, 010084068211686221TABX6266511180313 (TABX62665);  2082844-001-523533, 010084068211686221TABW62153170215 (TABW62153);  2082844-001-547388, 010084068211686221TABX63316180628 (TABX63316);  2082844-001-498396, 010084068211686221TABX6331811180629 (TABX63318);  2082844-001-498396, 010084068211686221TABX6331911180629 (TABX63319);  2082844-001-505050, 010084068211686221TABX62459180227 (TABX62459);  2082844-001-505050, 010084068211686221TABX62460180227 (TABX62460);  2082844-001-505050, 010084068211686221TABW63210170815 (TABW63210);  2082844-001-507094, 010084068211685521TABV61052160804 (TABV61052);  2082844-001-510416, 010084068211686221TABX62540180307 (TABX62540);  2082844-001-500909, 010084068211686221TABX62457180227 (TABX62457);  2082844-001-500909, 010084068211686221TABX62458180227 (TABX62458);  2082844-001-477964, Not Applicable (TABU67058);  2082844-001-516038, 010084068211686221TABX6246111180227 (TABX62461);  2082844-001-501478, 010084068211686221TABX62431180222 (TABX62431);  2082844-001-471681, Not Applicable (TABU67028);  2082844-001-471681, Not Applicable (TABU67034);  2082844-001-471681, Not Applicable (TABU67032);  2082844-001-462005, Not Applicable (TABU67021);  2082844-001-462005, Not Applicable (TABU67026);  2082844-001-516778, 010084068211686221TABX6248011180302 (TABX62480);  2082844-001-516778, 010084068211686221TABX6248111180302 (TABX62481);  2082844-001-516778, 010084068211686221TABX6247811180302 (TABX62478);  2082844-001-516778, 010084068211686221TABX6247911180302 (TABX62479);  2082844-001-516778, 010084068211686221TABV6231611161114 (TABV62316);  2082844-001-500008, 010084068211686221TABX62365180215 (TABX62365);  2082844-001-500008, 010084068211686221TABX62363180215 (TABX62363);  2082844-001-500008, 010084068211686221TABX62362180215 (TABX62362);  2082844-001-500008, 010084068211686221TABX62367180215 (TABX62367);  2082844-001-500008, 010084068211686221TABX62364180215 (TABX62364);  2082844-001-500008, 010084068211686221TABX62366180215 (TABX62366);  2082844-001-500008, 010084068211686221TABX62369180215 (TABX62369);  2082844-001-500008, 010084068211686221TABX62368180215 (TABX62368);  2082844-001-500008, 010084068211685521TABV62289161114 (TABV62289);  2082844-001-500008, 010084068211685521TABV62294161114 (TABV62294);  2082844-001-499372, 010084068211686221TABX62377180215 (TABX62377);  2082844-001-499372, 010084068211686221TABX62376180215 (TABX62376);  2082844-001-499372, 010084068211686221TABX62373180215 (TABX62373);  2082844-001-499372, 010084068211686221TABX62374180215 (TABX62374);  2082844-001-517864, 010084068211686221TABX63236180619 (TABX63236);  2082844-001-517864, 010084068211686221TABX63252180619 (TABX63252);  2082844-001-403565, Not Applicable (TABV60039);  2082844-001-559202, 010084068211686221TABX6248211180302 (TABX62482);  2082844-001-403565, Not Applicable (TABV60097);  2082844-001-403565, Not Applicable (TABV60062);  2082844-001-403565, Not Applicable (TABV60003);  2082844-001-403565, Not Applicable (TABV60060);  
FEI Number 2126677
Recalling Firm/
Manufacturer
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information Contact
800-437-1171
Manufacturer Reason
for Recall
Potential electric leakage affecting the user.
FDA Determined
Cause 2
Process design
ActionGE Healthcare notified customers on about 09/25/2018 via URGENT MEDICAL DEVICE CORRECTION letter. The letter indicated that the device could continue to be used along with good clinical practices. GE Healthcare will visit each location to perform the necessary testing. Until GE Healthcare can visit each site, instructions are to disconnect devices that use the USB (Giraffe OmniBed Carestation and Giraffe Incubator Carestation only) and the RS232 Nurse call ports, and inspect all power cords to ensure there is no damage to the cord or ground pin. GE Healtchare will perform necessary electrical safety test on all affected products and correct if required. A GE Healthcare representative will contact each customer to arrange for this test. For questions or concerns, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in Commerce1,724 devices total
DistributionWorldwide distribution. US Nationwide, Algeria, Australia, Austria, Belgium, Canada, Chile, China, Costa Rica, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Honduras, India, Indonesia, Ireland, Italy, Japan, Kazakhstan, Korea, Lithuania, Mexico, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = FMZ
-
-