• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall LIFEPAK 15 Monitor/Defibrillator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall LIFEPAK 15 Monitor/Defibrillator see related information
Date Initiated by Firm February 01, 2019
Date Posted February 26, 2019
Recall Status1 Open3, Classified
Recall Number Z-0862-2019
Recall Event ID 81939
510(K)Number K142430  
Product Classification Automated external defibrillators (non-wearable) - Product Code MKJ
Product LIFEPAK 15 Monitor/Defibrillator

Product Usage:
The LIFEPAK(R) 15 Monitor/defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. The LP15 monitor/defibrillator is intended for use by trained medical personnel in out-of-doors and indoor emergency care settings within the environmental conditions specified in the Operating Instructions. The LP15 monitor/defibrillator is designed to be used during ground transportation except when specified otherwise. Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older.
Code Information Serial Numbers: 38344771 38523396 39370585 39747192 40266966 40284286 40284289 40284297 40758461 40767555 40918431 40991313 40991452 40991700 40991743 40991746 40991996 40995181 40995294 40995323 40995420 40995421 40995508 40995550 40995583 40995613 40995621 40995656 41216405 41216406 41216566 41216581 41216600 41216608 41216617 41216618 41216631 41216632 41216636 41216650 41216655 41216660 41216665 41216668 41216680 41216682 41216736 41216743 41216767 41216795 41216797 41216806 41216822 41216828 41216849 41216851 41216854 41216917 41216925 41216955 41216974 41216980 41217013 41217080 41217083 41217104 41217122 41217123 41217131 41217179 41217214 41217221 41217224 41217225 41217233 41217666 41217667 41217677 41217698 41217715 41217722 41217783 41217785 41217791 41217800 41217802 41217834 41217893 41217894 41217929 41218236 41218625 41218627 41218648 41218684 41218810 41218812 41218821 41218846 41218885 41218891 41218909 41218932 41218935 41218938 41218948 41218949 41218951 41219009 41219014 41220729 41220733 41220743 41220763 41220764 41220773 41220794 41220804 41220845 41220854 41220857 41220861 41220862 41220875 41220877 41220885 41220891 41220894 41220941 41220944 41220998 41221005 41221099 41221104 41221105 41221133 41221254 41221255 41221275 41221305 41221311 41221328 41221364 41221390 41221444 41221446 41221447 41221473 41221483 41221494 41221503 41221536 41221545 41221547 41221554 41221555 41221569 41221597 41221606 41221608 41221613 41221630 41221989 41221999 41222096 41222112 41222119 41222177 41222245 41222355 41222440 41222610 41222661 41222725 41222810 41222838 41222907 41222921 41225022 41225110 41225126 41225245 41225293 41226675 41231145 41235035 41238528 41238550 41238554 41238597 41238692 41238788 41238820 41238835 41238852 41238898 41238946 41238990 41238994 41239071 41239120 41239140 41239141 41239194 41239196 41239200 41239228 41239314 41239324 41239373 41239443 41239460 41239469 41239473 41239563 41239594 41239643 41239661 41239671 41239677 41239734 41239754 41239756 41239812 41239833 41239891 41239912 41239920 41239946 41240005 41241905 41241908 41241921 41241944 41241995 41242019 41242073 41242166 41242262 41242270 41242307 41242321 41242343 41242345 41242351 41242375 41242376 41242401 41242511 41242517 41242605 41242653 41242686 41242707 41242729 41242762 41242809 41242820 41242880 41242912 41242948 41242961 41242998 41243188 41243324 41243368 41243445 41243507 41243530 41243565 41243634 41243696 41243777 41244393 41244442 41244612 41244815 41244943 41245095 41246831 41246858 41246954 41246964 41246972 41246982 41246999 41247278 41247310 41247325 41247494 41247644 41248026 41248884 41250244 41250331 41250348 41250486 41250488 41250552 41250566 41250570 41250608 41251349 41251434 41251893 41251961 41251970 41253186 41253236 41253482 41253541 41253650 41253695 41253718 41253727 41253862 41254102 41254164 41254462 41254484 41254543 41254562 41254628 41254761 41254876 41257159 41257186 41257675 41266920 41266947 41267220 41267244 41267358 41267407 41267578 41267597 41267807 41267857 41267879 41267894 41267914 41267942 41267943 41267986 41268050 41270057 41270060 41270080 41270085 41270095 41270096 41270100 41270127 41270129 41270133 41270137 41270141 41270150 41270158 41270162 41270183 41270192 41270211 41270220 41270248 41270269 41270285 41270301 41270354 41270381 41270385 41270435 41270437 41270501 41270551 41270575 41270822 41270857 41270870 41270891 41270895 41271071 41271473 41271485 41271540 41271634 41271741 41271810 41272155 41272224 41272225 41272258 41272318 41272346 41272474 41274356 41274546 41274673 41274680 41274933 41275009 41276196 41276219 41276318 41276319 41276354 41276417 41276451 41276485 41276555 41278260 41278264 41278290 41278319 41278337 41278342 41278372 41278398 41278463 41278604 41278787 41278865 41278880 41279136 41279457 41279608 41280114 41280173 41280218 41280349 41280390 41280412 41283161 41283231 41283236 41283260 41283265 41283312 41283322 41283372 41283544 41283585 41283713 41283754 41283756 41283817 41283975 41284148 41284181 41284272 41284316 41284388 41284450 41284528 41284546 41284630 41284695 41284949 41284965 41284978 41284985 41285032 41285045 41285089 41285412 41286028 41286097 41287666 41287728 41287802 41287815 41287853 41287864 41287897 41287954 41288038 41288058 41288116 41288119 41288133 41288136 41288147 41288160 41288238 41288263 41288270 41288371 41288492 41288568 41288866 41288896 41288968 41289271 41289356 41289377 41289431 41289461 41289688 41289722 41289749 41291453 41291617 41291883 41291885 41292095 41292217 41292328 41292412 41292563 41292596 41292688 41295631 41295651 41295673 41295676 41295694 41295702 41295707 41295724 41295751 41295772 41295778 41295802 41295868 41295872 41295895 41295917 41295918 41295989 41296082 41296093 41296133 41296323 41296446 41296508 41296522 41296545 41296606 41296645 41296757 41296832 41296865 41296943 41296965 41296976 41296989 41297051 41297149 41297403 41297499 41300009 41300134 41300462 41300475 41300557 41300617 41300620 41300683 41300726 41301062 41301123 41303917 41303920 41303973 41303993 41303997 41304114 41304117 41304138 41304148 41304194 41305056 41305134 41305211 41305251 41305735 41305858 41305925 41305933 41309210 41309311 41310880 41310954 41312714 41312780 41312929 41313000 41313039 41313066 41313128 41313344 41314434 41314685 41315519 41318582 41318667 41319005 41325864 41326633 41328498 41328510 41328543 41328642 41328691 41328703 41328816 41329184 41329312 41329346 41329850 41329979 41330002 41330118 41330121 41330367 41330393 41330557 41330567 41330649 41330653 41330676 41330712 41330767 41332353 41332466 41332725 41332849 41332853 41332878 41332935 41332964 41333028 41333031 41333036 41333258 41333816 41334022 41335778 41335914 41336030 41336244 41336742 41336836 41336965 41341555 41341592 41341624 41341665 41341701 41341761 41341861 41341936 41342000 41342031 41357071 41357109 41359659 41359735 41359763 41359802 41359827 41359855 41359927 41360068 41360093 41360110 41360191 41360265 41360310 41360497 41360545 41365083 41365226 41365244 41365256 41365274 41365298 41365653 41365750 41365794 41371502 41373916 41373990 41374332 41374431 41374526 41374577 41374819 41376604 41378614 41378981 41379049 41379242 41379274 41379313 41379395 41379417 41379460 41379507 41379667 41381507 41387329 41387398 41390198 41390473 41390496 41390590 41390650 41390809 41393328 41395143 41395151 41396816 41396952 41397024 41397101 41397121 41404625 41404626 41404712 41404758 41405022 41405031 41405040 41405046 41405060 41405090 41405112 41405252 41405257 41405260 41405292 41405318 41405328 41405370 41405437 41405548 41405653 41405753 41405833 41406033 41406044 41406077 41406140 41407757 41407840 41407878 41407908 41407914 41408009 41408094 41408293 41408363 41412032 41412069 41412119 41413246 41413690 41419372 41419659 41419814 41419849 41419879 41420672 41423133 41423173 41423208 41426977 41427032 41427112 41427172 41427201 41427254 41427255 41427316 41427335 41427358 41427421 41427504 41427616 41429917 41429918 41429933 41429945 41429948 41430100 41430104 41430130 41430153 41430182 41430192 41430195 41430280 41430282 41430431 41430491 41430933 41431223 41431234 41431257 41431265 41431281 41431308 41431310 41431366 41431375 41431390 41431417 41431418 41431488 41431493 41431522 41431570 41431579 41431607 41431610 41431696 41436748 41436852 41437222 41437424 41437711 41443575 41443687 41443747 41443821 41443855 41443920 41443949 41444011 41444023 41445567 41445834 41445839 41445855 41445905 41446012 41446136 41446617 41446665 41446721 41446734 41452889 41452940 41453051 41453404 41454344 41454436 41454613 41454635 41454653 41457971 41458102 41458644 41458747 41458896 41458941 41459135 41459201 41459234 41459298 41461448 41462627 41462748 41464916 41465066 41465091 41465314 41465585 41465620 41465690 41465698 41465788 41465827 41465831 41465862 41465865 41465907 41465921 41465924 41465978 41465989 41466022 41466165 41466281 41466676 41466730 41468965 41468989 41469093 41469113 41469160 41469333 41469351 41469777 41470129 41470677 41470839 41473455 41473750 41476919 41476983 41482266 41482955 41483166 41483203 41483273 41483336 41483367 41485395 41485544 41485631 41485636 41486303 41486327 41486335 41486355 41486357 41486414 41486424 41486448 41486654 41486673 41486678 41486717 41486722 41486762 41486772 41486803 41486831 41486843 41486866 41486902 41487035 41487093 41487154 41487211 41487226 41487271 41489244 41489300 41489398 41489421 41489454 41489488 41489661 41489680 41489967 41489983 41490024 41490310 41490458 41490663 41497544 41497562 41497619 41497740 41497936 41498677 41498847 41500607 41501021 41501083 41501102 41501159 41501270 41501386 41501485 41501685 41501862 41503647 41503661 41503829 41503842 41504024 41504069 41504094 41504184 41504314 41504330 41504391 41504617 41504956 41505112 41511616 41511645 41511721 41511817 41511854 41511890 41511914 41511942 41512326 41512433 41512467 41512495 41512592 41512677 41512708 41512758 41512905 41515578 41515599 41515750 41515791 41515795 41515847 41515870 41516177 41516444 41519315 41519482 41519964 41520097 41520201 41520214 41520302 41520331 41520401 41520467 41520656 41520690 41520843 41520863 41520897 41520917 41520929 41520951 41520975 41520978 41521000 41521001 41522684 41522818 41522852 41522952 41522974 41522999 41523009 41523184 41523237 41523272 41523309 41523424 41523597 41523681 41523952 41524323 41524483 41524484 41524529 41524818 41524829 41525379 41525677 41525945 41525990 41526297 41526595 41526598 41527029 41527147 41527152 41536400 41536422 41536588 41536648 41536660 41536673 41536676 41536689 41536725 41538165 41538229 41538242 41538251 41538265 41538399 41538413 41540618 41540840 41540929 41541875 41542079 41542152 41542156 41542411 41542694 41545326 41545416 41546794 41550001 41550389 41550454 41550509 41550650 41552891 41557549 41557952 41558849 41565307 41565388 41566102 41566799 41570208 41571950 41572011 41572152 41572322 41573163 41573285 41573343 41573437 41573476 41573583 41574167 41577102 41578215 41578477 41579275 41589604 41589622 41589686 41589735 41589843 41598884 41599480 41601005 41601137 41601155 41601192 41601200 41601249 41601390 41601430 41601482 41601553 41601608 41601643 41601659 41601729 41601897 41602052 41602357 41602395 41602571 41602774 41602818 41602850 41602878 41604659 41604706 41604726 41604737 41604741 41604753 41604818 41605032 41605945 41606040 41606081 41606160 41606252 41606560 41609706 41609737 41609778 41610008 41610023 41610068 41610086 41610185 41610291 41610596 41610638 41610742 41618604 41618621 41618745 41618788 41618801 41618927 41618949 41618993 41618997 41619080 41619087 41619159 41619165 41619183 41619188 41619218 41619223 41619232 41619285 41619323 41621538 41621586 41621611 41621646 41621840 41621927 41623883 41624779 41631983 41635236 41635477 41635660 41635769 41638270 41638314 41638435 41638517 41638582 41639604 41640011 41642186 41642202 41642266 41642390 41642671 41643045 41644990 41645002 41645024 41645044 41645050 41645093 41645485 41646498 41646597 41646810 41646944 41653515 41653635 41653684 41653855 41653982 41654703 41656484 41656755 41663364 41663444 41663664 41664169 41664266 41664353 41664485 41664608 41664653 41664699 41665126 41665222 41665368 41671790 41671844 41671893 41671960 41671990 41672155 41672192 41674374 41674479 41674635 41675383 41675458 41675509 41675601 41676354 41677614 41678386 41678599 41679073 41679152 41679217 41679268 41679309 41679367 41679426 41684786 41684828 41684879 41685300 41686069 41686190 41686228 41686506 41686715 41686732 41686735 41688013 41688253 41688480 41689042 41689163 41689559 41689746 41690321 41691939 41692003 41692086 41692151 41692315 41692552 41692638 41692724 41693371 41693456 41694877 41695036 41695055 41695097 41695233 41695270 41695340 41695646 41695833 41695923 41703090 41703200 41703227 41703267 41704093 41704185 41704236 41704242 41704445 41704491 41704493 41704556 41704833 41704896 41704952 41704997 41705004 41705641 41705738 41706211 41706338 41706417 41707859 41708008 41708075 41708162 41708218 41708259 41708275 41708287 41708333 41708370 41708406 41708410 41708475 41708502 41708547 41708636 41708691 41708735 41708743 41708781 41710922 41711059 41711124 41711173 41711498 41711668 41712881 41713253 41718348 41718514 41718579 41718602 41718647 41718649 41718716 41718779 41718861 41718897 41718916 41718983 41719020 41719024 41719062 41719072 41719288 41719309 41719323 41719328 41719340 41719373 41719404 41719418 41719656 41719669 41719942 41719959 41720186 41720216 41720235 41720328 41720331 41720484 41720647 41720675 41720687 41720718 41721843 41721877 41721933 41721994 41722028 41722056 41722097 41722112 41722153 41722158 41722217 41722241 41722316 41722408 41722451 41722457 41722531 41722642 41722695 41722761 41722841 41723398 41723455 41723517 41723638 41723765 41723900 41724035 41724073 41724154 41725715 41726430 41726535 41727407 41727559 41727611 41727724 41727801 41727855 41728260 41729833 41730418 41731059 41731280 41734427 41734533 41734534 41734581 41736876 41736938 41736939 41736968 41737000 41737074 41737106 41737145 41737281 41737483 41737551 41737596 41737630 41737676 41737955 41737973 41738125 41738193 41738210 41738261 41738269 41738282 41743199 41747668 41747722 41747791 41748049 41748113 41748153 41748199 41748224 41748250 41748258 41748281 41748310 41748328 41748358 41748405 41748631 41748657 41749212 41749308 41749340 41749560 41749715 41749781 41749788 41749934 41756821 41756826 41756840 41756856 41756877 41756935 41757023 41757186 41757283 41757912 41758000 41758027 41758191 41758415 41758611 41758625 41759116 41760546 41760987 41761059 41761061 41761105 41761177 41761270 41761297 41761363 41761471 41761486 41761520 41761546 41761629 41761662 41761671 41761711 41761860 41761900 41761968 41761972 41762010 41764411 41764615 41765365 41765427 41765453 41765541 41765613 41765633 41765789 41765905 41767574 41767603 41767940 41767947 41767976 41768008 41768044 41768045 41768102 41768121 41768241 41768279 41768360 41768376 41768466 41768475 41768505 41768511 41768546 41768550 41768589 41768632 41768720 41768787 41768902 41769010 41769074 41769115 41769126 41769142 41769165 41769210 41769214 41769282 41769309 41769314 41769479 41769527 41769687 41775661 41775688 41775723 41775838 41775873 41775960 41775981 41776005 41776055 41776088 41776108 41776127 41776131 41776190 41776225 41776284 41776388 41776390 41776413 41776439 41776482 41776491 41776501 41776520 41776528 41776547 41776556 41776609 41776616 41776689 41777107 41777112 41777118 41777157 41777192 41777272 41777569 41777589 41777738 41777742 41777754 41778061 41778214 41780045 41780055 41780099 41780107 41780147 41780169 41780257 41780270 41780306 41780353 41780368 41780423 41780484 41780562 41780836 41780904 41780926 41780982 41780990 41781056 41781104 41781126 41781173 41781425 41781445 41781932 41782005 41782059 41782108 41782183 41782210 41782261 41782362 41782367 41782430 41782623 41782644 41782648 41782684 41782742 41785277 41785426 41785800 41786201 41786290 41786542 41786821 41786927 41786928 41786946 41787129 41787338 41788082 41788112 41789124 41791697 41791709 41794138 41798087 41798140 41798686 41798931 41807976 41808014 41809851 41810269 41810664 41810783 41812682 41813179 41814263 41814308 41814391 41814559 41814605 41814675 41814745 41815064 41816791 41817010 41817264 41817394 41817530 41817586 41818004 41818865 41819121 41819184 41819329 41819409 41820531 41821026 41821069 41821113 41821209 41821525 41821585 41821589 41821725 41822015 41822043 41822115 41822145 41822247 41822273 41822307 41822654 41824210 41824213 41824741 41824889 41824947 41825963 41826064 41826304 41826308 41826907 41830269 41844284 41844400 41844510 41844593 41844644 41844658 41844666 41844693 41844765 41844880 41844911 41846074 41846182 41846670 41850702 41850869 41850980 41851300 41851453 41852029 41853981 41866806 41869458 41871298 41871390 41873247 41873282 41874354 41874393 41874485 41874973 41878580 41878836 41878934 41879010 41879108 41879247 41879409 41879423 41880014 41880083 41881296 41881300 41881399 41881427 41881480 41881567 41881568 41881606 41881607 41882031 41882188 41882437 41882586 41888404 41888763 41888851 41889139 41889961 41890031 41890140 41890202 41890512 41892250 41892298 41892385 41892452 41892807 41892831 41893029 41893406 41893576 41894031 41894067 41895674 41896004 41896033 41896155 41896317 41896356 41896900 41896943 41896963 41898859 41898993 41899090 41899140 41899675 41899735 41899780 41900044 41900123 41900264 41900402 41900999 41902866 41907405 41907432 41907688 41907739 41907900 41908566 41908623 41908738 41908757 41908912 41909058 41909168 41909262 41909648 41909650 41909686 41909704 41909749 41909779 41909816 41909853 41911241 41911303 41911521 41911909 41912423 41912432 41912459 41913158 41913411 41913423 41913440 41913653 41913695 41913830 41913900 41913942 41914580 41914628 41914710 41914724 41917530 41917595 41918405 41918579 41924195 41924343 41924475 41924506 41924648 41925372 41925441 41926562 41926590 41927013 41927444 41928828 41929435 41929472 41932003 41933089 41934040 41934177 41935546 41935700 41936005 41936334 41936428 41936502 41936684 41937115 41937172 41937199 41937247 41937332 41937479 41937523 41938584 41938637 41938720 41938823 41940651 41940715 41940795 41940827 41940847 41941294 41941404 41941478 41941640 41942612 41942726 41943150 41943402 41951351 41951481 41951584 41951709 41951743 41951894 41951924 41951990 41952058 41952189 41952210 41952262 41952290 41952312 41952371 41952572 41952636 41952702 41952712 41952740 41952864 41952885 41952938 41953028 41953039 41953096 41953121 41953216 41953365 41954627 41954668 41954696 41954726 41954774 41954844 41954866 41954873 41954886 41954936 41954956 41954965 41954992 41955042 41955056 41955156 41955157 41955223 41955257 41955262 41955271 41955288 41955314 41955319 41955358 41955449 41955471 41955525 41955660 41955661 41956003 41956004 41956042 41956056 41956164 41956178 41956204 41956253 41956485 41958451 41958505 41958584 41958695 41958779 41959138 41959187 41959305 41961243 41964175 41964273 41964346 41964407 41964657 41965003 41965021 41965056 41965139 41965146 41965436 41965745 41965962 41965976 41966061 41966118 41966149 41966220 41967751 41967841 41967995 41968053 41968172 41968343 41968613 41968651 41968811 41973513 41973613 41973638 41973707 41976128 41976136 41977717 41983141 41985969 41985994 41985998 41986049 41986065 41986124 41986182 41986225 41986250 41986385 41986459 41986462 41986490 41986498 41986514 41986520 41986549 41986554 41986572 41986840 41986972 41987003 41987006 41987055 41987056 41987285 41987312 41987347 41987398 41987466 41987524 41987571 41987576 41987592 41987603 41987791 41987792 41987820 41987863 41988025 41988074 41988195 41988201 41988211 41988226 41988391 41990404 41990456 41990533 41990656 41990818 41991007 41991130 41991173 41991474 41991755 41991819 41998956 41999338 42000040 42001912 42001964 42002114 42002175 42003180 42005229 42005611 42005687 42006159 42006285 42006640 42006742 42006784 42006822 42007676 42008098 42008147 42008190 42008195 42008239 42008486 42008679 42008818 42008865 42009273 42009385 42009672 42018314 42018403 42018490 42018597 42019422 42019453 42019711 42020313 42020395 42020411 42020893 42022351 42024595 42026438 42026776 42026807 42026923 42027021 42027062 42027085 42027127 42027159 42027231 42027539 42027693 42027812 42027877 42028337 42028418 42029820 42029910 42029944 42030224 42030265 42030302 42030355 42030356 42030372 42030384 42030390 42030424 42030432 42030443 42030719 42030838 42030864 42030872 42030885 42030920 42030924 42030992 42031019 42031048 42031050 42031089 42031202 42031222 42031227 42031311 42031338 42035474 42035601 42035664 42035694 42035754 42035796 42036136 42036192 42036713 42036842 42036843 42036924 42036960 42036964 42037056 42037089 42037142 42037201 42039284 42039420 42039467 42039549 42039629 42039631 42039667 42039747 42039775 42039805 42039861 42039893 42039925 42039979 42039989 42040018 42040052 42040069 42040123 42040145 42040146 42040163 42040164 42040241 42040243 42040255 42040350 42040351 42040364 42040389 42040403 42040421 42040459 42040490 42040514 42040568 42040625 42040719 42040734 42040777 42040787 42040795 42040834 42040838 42042045 42042102 42042145 42042190 42042194 42042340 42042450 42042680 42042897 42043068 42043198 42043656 42043971 42045531 42045803 42045914 42046107 42046203 42046258 42046309 42046427 42046636 42046986 42046990 42047072 42048562 42048600 42048732 42048929 42049028 42049164 42049343 42049414 42049582 42052076 42053269 42055692 42055758 42055787 42055828 42055950 42055959 42058249 42059269 42059307 42062029 42062350 42062594 42062670 42062846 42068355 42068502 42068845 42068876 42068951 42068992 42069015 42069082 42069198 42069427 42070148 42070151 42070156 42070161 42070210 42070212 42070372 42070396 42070404 42070471 42070557 42070646 42070664 42070683 42070693 42070787 42071037 42071068 42071677 42071798 42071891 42071901 42071955 42076727 42077022 42077079 42077695 42082638 42082871 42082968 42083031 42083664 42091315 42097540 42100927 42100945 42101506 42101583 42101795 42101901 42101927 42101993 42102060 42102106 42102149 42102219 42102239 42102280 42102745 42103687 42103694 42103849 42104900 42110653 42110845 42110900 42110926 42111100 42111310 42111420 42111660 42112370 42112467 42112571 42112678 42113975 42114196 42114324 42114400 42115612 42115808 42116174 42117424 42117447 42117643 42118680 42118883 42118995 42119068 42119286 42121651 42121678 42121770 42121882 42122022 42122287 42122482 42122655 42122905 42122982 42123009 42123059 42123503 42133634 42133699 42133732 42133888 42133954 42134012 42134080 42134095 42134133 42134213 42134263 42134674 42137325 42137467 42137477 42137547 42137579 42137583 42137656 42137659 42137705 42137709 42137747 42137793 42137823 42137843 42137856 42137868 42137895 42137953 42137984 42137985 42138060 42138069 42138170 42138174 42138266 42138311 42138358 42138378 42138742 42138785 42139772 42140017 42142647 42142648 42142708 42142802 42142805 42142855 42143187 42143234 42143297 42143324 42143363 42143613 42143727 42144111 42144352 42144485 42144853 42146006 42146046 42146124 42146186 42146271 42146306 42146334 42146360 42146394 42146609 42146637 42146722 42146724 42146984 42146992 42147184 42147399 42147402 42147707 42147765 42147803 42147818 42148484 42151520 42151590 42151655 42151736 42151799 42151800 42151881 42151882 42151955 42151963 42152016 42152106 42152110 42152141 42152144 42152157 42152161 42152180 42152202 42152208 42152221 42152251 42152281 42152321 42152408 42152588 42152633 42152663 42152715 42152748 42152753 42152824 42152826 42152847 42152957 42153000 42153103 42153290 42153304 42153360 42153450 42153467 42153509 42154676 42154678 42154711 42154719 42154771 42154824 42154970 42155031 42155109 42155170 42155239 42155264 42155310 42155510 42155519 42155574 42155621 42155869 42155910 42156061 42156737 42156957 42158348 42158432 42158682 42158762 42158867 42159018 42159076 42159185 42159292 42159930 42160160 42173157 42173364 42173504 42173846 42174896 42174910 42174949 42175002 42175089 42175133 42175223 42175368 42175587 42175816 42175963 42176122 42176168 42176269 42176287 42176340 42176350 42176450 42176485 42176551 42176566 42176607 42178081 42178095 42178338 42178518 42178623 42178702 42178997 42179101 42179239 42179250 42179312 42179358 42179396 42179444 42179520 42179586 42179648 42179682 42179727 42179804 42181626 42181723 42181839 42181924 42181997 42182093 42182168 42182263 42182376 42182470 42182622 42183032 42183302 42183345 42183456 42183569 42194142 42194235 42194286 42194335 42194397 42194430 42195109 42195306 42195316 42195352 42195635 42195690 42195817 42195874 42196002 42197895 42198795 42199098 42200696 42200742 42200762 42202335 42202343 42202355 42202443 42202481 42202533 42202534 42202594 42202607 42202899 42202961 42203070 42203108 42203164 42204074 42204320 42210930 42210979 42211018 42211078 42211145 42211151 42211195 42211203 42211535 42211547 42211582 42211658 42211795 42212041 42212188 42212215 42212253 42212283 42212336 42212414 42212483 42212560 42212610 42212674 42212724 42212748 42212786 42214082 42214151 42214245 42214274 42214276 42214327 42214372 42214377 42214424 42214430 42214518 42214640 42214809 42214872 42214973 42215015 42215058 42215064 42215158 42215204 42215626 42215655 42215666 42215731 42215860 42215876 42215883 42215914 42215916 42215994 42215998 42216027 42216116 42216151 42216178 42216221 42216249 42216253 42217781 42217789 42217790 42217803 42217821 42217827 42217879 42217934 42217937 42217979 42217990 42218031 42218086 42218341 42218381 42218382 42218417 42218572 42219006 42219379 42220826 42221035 42221036 42221060 42221210 42221232 42221332 42221372 42221403 42221482 42221689 42221808 42221833 42221859 42221867 42221908 42221974 42222002 42222049 42222068 42222185 42222513 42222645 42222804 42222846 42224061 42224062 42224341 42224361 42224406 42224427 42224468 42224480 42224559 42224598 42224599 42230864 42230871 42230875 42230895 42230901 42230966 42230981 42231035 42231134 42231170 42231179 42231191 42231230 42231233 42231234 42231300 42231307 42231350 42231366 42231447 42231448 42231502 42231520 42231550 42231557 42231601 42231607 42231640 42231693 42231701 42231728 42231750 42231785 42231796 42231826 42231838 42231923 42231925 42231954 42231971 42232111 42232123 42232139 42232185 42232268 42232269 42232320 42232321 42232351 42232356 42232382 42232383 42232450 42232466 42232483 42232531 42232566 42232602 42232681 42232723 42232780 42232847 42234168 42234171 42234178 42234181 42234228 42234245 42234250 42234293 42234313 42234343 42234437 42234473 42234569 42234652 42234654 42234672 42234770 42234869 42234974 42235241 42235272 42235707 42235906 42236212 42236282 42236343 42236361 42236382 42236419 42238223 42238228 42238249 42238266 42238293 42238298 42238324 42238363 42238369 42238395 42238415 42238418 42238434 42238441 42238476 42238477 42238503 42238509 42238519 42238545 42238546 42238576 42238578 42239630 42239637 42239657 42239698 42239764 42239865 42239980 42240158 42240550 42240573 42240670 42240867 42241020 42243401 42243411 42243430 42243480 42243522 42243647 42243661 42243662 42243708 42243765 42245220 42245238 42245258 42245302 42245348 42245397 42245513 42245651 42245705 42245790 42247684 42247694 42247927 42247985 42248139 42248467 42248499 42248646 42249260 42249282 42249337 42249379 42249380 42249494 42249525 42249831 42253794 42253883 42254101 42254240 42254321 42254432 42254447 42254485 42254487 42254504 42254560 42254572 42254591 42254627 42254839 42254853 42254858 42254871 42254877 42254879 42255019 42255036 42255075 42255078 42255139 42255143 42255182 42255380 42255394 42255416 42255477 42255493 42257364 42257405 42257418 42257469 42257486 42257575 42257731 42257756 42257757 42257919 42258027 42258031 42258091 42258166 42258538 42258659 42258695 42258890 42258900 42258950 42258952 42258961 42259171 42259214 42259258 42259321 42259327 42259448 42259498 42259730 42261067 42261504 42261576 42261744 42261856 42261898 42261978 42262029 42263009 42263032 42263068 42263120 42263175 42263232 42263358 42263388 42279548 42279565 42279848 42279894 42280040 42280270 42280997 42281097 42281469 42281617 42282170 42282219 42282462 42282512 42282513 42282548 42282581 42282587 42282924 42283191 42283240 42284486 42284606 42284648 42284655 42284704 42284739 42284762 42284765 42284810 42284816 42284898 42284916 42284938 42284957 42285071 42285103 42285142 42285310 42285322 42285363 42285419 42285429 42285453 42285470 42285474 42285579 42285782 42285788 42286023 42286103 42286116 42286135 42286157 42286164 42286178 42286205 42286283 42286285 42286371 42286521 42286530 42286748 42286797 42286881 42286886 42286956 42286969 42287138 42287156 42291038 42291089 42291182 42291200 42291298 42291302 42291347 42291373 42291375 42291444 42291453 42291700 42291716 42291729 42291768 42291938 42291942 42291970 42292092 42292133 42292139 42293494 42293523 42293574 42293605 42293642 42293714 42294252 42294303 42294316 42294336 42294341 42294410 42294416 42294459 42294878 42294937 42294965 42294990 42295071 42295344 42295473 42296349 42296358 42296432 42296550 42296567 42296671 42296866 42296972 42296976 42297068 42297109 42297154 42297172 42297198 42297246 42297250 42297275 42297295 42297299 42297319 42297320 42297343 42297362 42297452 42297711 42297731 42297746 42297797 42297817 42297819 42297898 42298094 42298286 42298441 42298490 42298515 42298537 42298589 42302222 42302228 42302239 42302488 42302653 42302683 42302825 42303035 42303057 42303064 42303123 42303182 42303231 42303274 42303429 42303431 42303458 42303518 42303541 42303613 42303647 42303679 42303710 42303713 42303804 42303875 42303902 42303932 42303973 42304068 42305823 42306215 42306516 42306811 42306884 42307148 42307441 42309967 42310035 42310288 42310489 42310760 42310798 42310838 42310849 42310856 42311099 42311738 42311856 42312300 42312401 42312441 42314235 42314269 42314339 42314362 42314428 42314578 42314579 42314773 42315030 42315070 42315109 42315163 42315243 42315476 42315512 42315574 42315774 42315808 42316289 42316297 42316367 42316400 42316486 42316487 42316550 42316569 42316572 42316629 42316812 42316813 42316959 42316998 42317126 42317229 42317506 42318752 42318761 42318769 42318847 42318896 42319021 42319032 42319040 42319060 42319099 42319167 42319194 42319241 42319251 42319300 42319305 42319478 42319491 42319527 42319528 42319689 42319694 42319768 42319773 42319850 42319874 42319915 42319940 42320338 42320402 42320454 42320567 42320656 42320729 42320732 42320756 42320813 42320947 42321020 42321109 42321112 42321385 42321477 42321501 42321560 42321607 42321694 42321701 42321766 42321803 42323024 42323029 42323035 42323044 42323052 42323053 42323064 42323065 42323071 42323081 42323083 42323091 42323097 42323106 42323112 42323120 42323128 42323132 42323143 42323157 42323167 42323171 42323207 42323271 42323279 42323284 42323288 42323290 42323297 42323310 42323318 42323329 42323337 42323347 42323350 42323365 42323374 42323387 42323388 42323448 42323464 42323466 42323475 42323478 42323489 42323490 42323496 42323498 42323545 42323548 42323562 42323563 42323572 42323576 42323583 42323586 42323594 42323604 42323611 42323619 42323624 42324022 42324023 42324137 42324150 42324166 42324178 42324181 42324190 42324193 42324214 42324215 42324232 42324235 42324277 42324290 42324300 42324310 42324317 42324345 42324383 42324397 42324495 42324502 42324547 42324552 42324686 42324720 42324749 42324775 42324821 42324838 42324912 42324972 42324979 42325062 42325085 42325098 42325184 42325216 42325237 42325266 42325412 42325623 42326199 42327153 42327311 42327400 42327449 42329326 42329524 42329854 42329860 42330250 42332824 42332918 42332958 42333179 42333271 42333293 42333621 42334037 42337626 42341054 42341129 42341215 42341318 42341484 42341593 42341704 42341740 42341766 42343733 42345335 42345339 42345535 42345758 42345776 42345863 42345871 42345938 42346861 42346863 42346925 42347092 42353114 42353245 42353303 42353664 42354037 42354379 42355823 42355948 42361413 42361534 42361564 42361592 42363080 42363271 42363309 42363351 42363392 42374081 42374112 42374298 42374935 42374991 42375038 42375140 42375171 42375188 42375278 42375329 42375349 42375395 42375543 42376425 42376495 42376756 42376786 42376850 42376934 42376955 42376995 42377122 42377172 42377288 42382563 42382742 42388659 42388840 42388850 42388867 42389151 42389395 42389641 42389710 42389942 42390272 42390367 42392306 42393695 42394600 42394641 42394762 42394784 42394851 42395448 42395728 42395822 42396064 42396571 42396700 42396729 42397738 42397853 42398429 42398487 42398547 42398597 42398629 42398777 42399368 42399680 42399776 42399800 42401458 42401608 42402312 42402330 42402970 42403114 42403643 42407994 42408097 42408631 42408748 42409123 42411104 42411175 42411334 42411435 42411456 42411473 42414990 42415009 42415226 42415287 42415543 42415753 42415932 42416211 42416441 42416489 42416642 42418950 42419357 42419513 42419644 42419795 42420206 42420255 42426725 42426799 42426861 42426953 42427052 42427129 42427214 42427343 42427439 42427493 42428268 42428328 42428663 42428708 42429928 42429941 42429951 42429980 42430023 42430034 42430233 42430893 42431304 42431707 42431813 42432008 42433271 42433288 42433314 42433333 42433362 42433410 42433441 42433457 42433508 42433545 42433976 42433989 42434032 42434308 42434417 42434523 42434676 42434711 42434738 42434780 42434957 42441328 42441409 42441483 42441564 42441587 42441629 42441914 42442141 42442542 42442568 42442876 42444488 42445158 42445319 42445888 42446011 42446244 42448286 42448755 42449308 42452019 42452079 42452767 42452917 42462886 42462902 42462925 42462953 42462979 42462996 42463179 42463339 42463610 42463671 42465083 42466544 42468547 42469446 42469587 42469665 42469758 42470132 42470392 42470778 42470841 42471607 42473892 42475408 42477627 42478209 42478345 42478518 42478603 42478644 42478671 42479430 42479972 42481292 42481998 42482150 42484973 42485816 42486357 42487841 42489163 42489176 42489214 42489556 42492588 42492765 42493758 42494985 42494995 42495101 42495258 42495866 42498157 42498193 42501609 42501654 42501735 42501846 42501967 42502068 42502174 42502301 42502640 42502708 42503530 42503669 42504836 42504858 42504966 42505516 42505663 42507027 42507069 42507111 42507154 42509702 42512051 42519417 42530038 42530186 42530663 42530735 42531320 42531439 42531532 42531678 42531776 42531880 42531925 42532022 42534029 42534047 42534117 42534266 42534385 42534463 42534701 42534768 42534860 42534918 42535059 42535105 42535152 42535428 42538608 42538825 42538929 42540437 42545801 42546736 42547020 42547247 42547332 42547342 42547418 42547496 42547576 42547577 42547884 42550433 42552664 42552784 42553158 42553329 42553574 42553657 42553704 42553710 42553777 42553811 42553842 42553860 42553863 42553866 42554021 42554036 42565768 42565782 42565990 42566164 42566330 42566422 42566504 42566633 42566666 42566697 42566755 42566800 42566815 42566942 42567003 42567077 42567519 42567654 42569222 42569301 42569366 42569408 42569472 42569548 42569762 42569808 42569890 42570025 42570241 42570372 42570463 42572614 42573431 42573624 42575364 42575401 42575670 42575691 42575875 42575947 42576080 42576102 42576197 42576257 42576371 42576387 42576416 42576495 42576563 42576686 42588960 42588963 42588980 42589058 42589060 42589113 42589169 42589841 42590125 42590134 42590135 42590194 42590238 42590251 42590358 42590403 42590440 42590523 42590553 42590556 42591855 42591856 42591860 42591868 42591894 42591933 42591943 42591967 42592004 42592007 42592021 42592022 42592032 42592033 42592041 42592078 42592524 42592565 42592579 42592588 42592599 42592606 42592781 42592859 42593127 42593220 42593318 42593385 42593416 42593422 42593442 42593497 42594113 42594130 42594156 42594175 42594215 42594443 42594496 42595324 42595335 42595342 42595408 42595505 42595524 42595613 42595614 42595624 42595634 42595635 42595666 42595687 42595740 42595933 42595955 42595973 42595983 42596027 42596079 42596209 42596246 42596268 42596306 42596315 42596526 42596642 42596661 42596693 42596696 42596760 42596793 42596867 42596929 42597045 42597048 42597074 42597085 42597184 42597300 42597328 42597459 42597493 42597495 42599330 42599373 42599386 42599430 42599476 42599519 42599533 42599536 42599563 42599564 42599574 42599575 42599600 42599979 42600416 42600449 42600497 42600499 42600524 42600603 42600606 42600622 42600665 42600682 42600690 42600704 42600707 42600732 42600770 42600797 42600808 42600820 42601422 42601455 42601458 42601485 42601503 42601520 42601541 42601560 42601927 42602365 42602384 42602423 42602425 42602439 42602448 42602474 42602519 42602540 42602635 42602673 42602792 42602796 42603211 42603225 42603234 42603251 42603272 42603278 42603328 42604119 42604123 42604226 42604283 42604310 42604451 42604456 42604464 42604482 42604676 42604776 42604793 42604826 42604920 42604930 42604959 42607187 42609776 42609825 42609897 42609904 42609972 42610031 42610038 42610077 42610124 42610178 42610841 42612139 42612340 42612520 42612525 42612557 42612585 42612635 42612642 42612670 42612701 42612717 42612745 42612751 42612769 42613229 42613242 42613332 42613391 42613464 42613484 42613488 42616485 42616574 42616590 42616686 42616738 42616773 42616792 42616918 42616989 42616991 42617016 42617017 42617081 42617091 42617105 42617121 42617147 42617168 42617170 42617278 42617285 42617305 42617322 42617635 42623128 42623134 42623965 42624518 42624883 42625256 42625301 42625319 42625421 42625809 42625866 42625883 42625888 42627316 42627355 42627387 42628160 42628238 42630988 42631787 42631836 42632307 42632346 42632454 42632536 42633673 42634848 42635955 42636702 42638307 42639839 42640069 42640341 42640414 42658699 42658966 42659066 42659820 42659871 42665116 42665270 42667745 42667795 42667887 42667923 42667975 42668198 42668369 42668451 42668487 42668639 42669114 42669163 42669328 42685458 42692112 42698174 42698388 42698595 42698770 42698861 42699272 42699318 42699340 42699431 42699594 42699720 42699775 42699896 42699934 42701453 42701821 42702071 42702479 42702513 42704761 42709999 42716501 42716590 42720815 42720870 42720964 42721058 42721652 42722023 42722340 42722434 42722508 42723723 42723794 42724047 42724148 42724231 42724655 42726278 42727387 42728703 42728759 42728832 42728890 42730609 42730620 42730750 42730837 42730862 42731258 42731329 42731466 42732190 42760924 42763375 42763470 42765133 42768652 42768829 42768848 42768969 42769225 42793132 42793459 42793588 42796952 42798105 42798212 42798305 42798370 42798393 42798480 42798526 42798570 42798603 42798707 42798752 42798925 42798985 42799074 42799387 42799410 42799463 42799503 42799640 42799667 42799781 42800040 42806599 42806798 42807166 42808427 42808715 42808829 42808956 42819820 42819836 42819847 42820119 42820219 42820325 42820601 42827299 42827627 42828653 42829506 42829557 42829573 42829627 42829659 42829936 42829975 42831111 42831312 42831421 42831963 42832933 42833194 42833286 42833305 42833329 42833338 42833349 42833357 42833420 42833424 42833432 42833440 42835838 42835928 42835953 42835959 42835974 42835980 42835986 42835995 42836007 42836014 42836019 42836043 42836078 42836081 42836094 42836100 42836113 42836204 42836273 42836367 42836376 42836538 42836585 42836587 42836623 42836769 42836770 42836828 42836833 42836904 42836962 42837036 42837313 42837559 42837680 42837693 42837716 42837767 42837774 42837872 42837926 42837927 42837935 42837977 42837999 42838012 42838024 42838112 42838142 42838187 42838188 42838216 42838217 42839532 42839560 42839591 42839595 42839615 42839632 42839635 42839647 42839875 42839885 42840000 42840015 42840024 42840244 42840263 42841521 42841829 42851326 42851337 42851555 42851639 42852115 42852162 42852260 42852593 42853054 42853070 42853086 42853090 42853132 42853144 42854479 42854780 42854941 42854954 42855002 42855045 42855063 42855070 42855088 42855102 42855168 42855193 42855231 42855255 42855268 42855314 42855344 42855408 42855416 42855497 42856017 42856080 42856283 42859136 42859157 42859230 42859236 42859310 42859343 42859349 42859368 42859426 42859475 42859479 42859549 42859629 42859635 42859670 42859697 42859708 42859789 42859932 42860010 42860111 42860184 42860219 42860254 42866940 42867045 42867072 42867149 42867219 42867685 42867725 42868827 42868842 42868918 42869053 42869172 42869211 42869304 42869486 42869541 42869678 42869796 42870006 42870247 42870350 42870494 42870563 42870590 42870621 42871127 42871480 42872135 42877267 42879834 42893486 42895052 42895369 42895501 42895613 42895761 42895973 42896043 42896171 42896279 42896438 42896852 42898485 42898572 42898599 42918227 42920608 42921647 42922802 42925196 42939280 42939368 42939401 42939480 42945786 42945971 42945992 42946048 42948034 42993356 42993502 42993839 42994021 42998015 42998602 42998615 42998747 42998794 42998931 42999035 42999063 43002800 43002814 43003357 43003602 43007360 43007412 43017433 43040893 43041078 43041189 43041228 43082803 43083297 43110461 43133611 43137417 43138376 43138480 43157364 43158400 43158432 43158716 43158748 43158776 43158782 43158923 43159045 43159102 43159148 43159211 43159279 43159344 43159397 43159479 43159512 43160134 43160211 43160262 43160291 43160346 43161374 43161410 43161661 43161735 43161736 43163903 43164429 43203359 43203502 43203626 43203750 43205387 43205478 43209867 43219362 43269290 43285281 43285368 43311564 43311697 43311921 43337616 43499255 43559136 43715580 43715943 43716064 43716251 43716322 43716431 43716833 43727699 43727802 43727843 43728163 43743657 43743783 43744156 43744318 43744516 43783576 43783659 43783723 43783882 43783908 43808509 43808721 43808773 43808954 43809168 43809894 43810170 43810365 43878347 43878848 43878993 43879319 43879737 43879816 43884257 43884464 43926895 43927076 43927223 43927281 43927351 43927452 43927579 43927669 43927701 43928089 43928103 43928155 43928162 43928314 43928316 43928388 43928518 43928600 43928608 43928701 43928714 43928839 43928908 43964199 43964201 43964202 43964207 43965050 44013138 44678760 44678767 44678820 44678835 44679255 44679261 44679268 44679279 44679283 44679289 44679290 44680774 44680776 44680778 44680785 44680791 44680793 44681600 44681606 44681610 44682293 44682296 44682299 44682301 44682306 44682568 44682570 44682580 44682589 44682594 44685369 44685372 44685373 44685377 44685378 44685627 44685633 44685635 44685639 44685642 44685647 44685653 44719899 44721149 44721617 44721765 44721896 44721926 44721930 44721933 44722105 44722107 44722109 44722111 44722112 44724191 44724192 44724193 44724194 44724195 44724327 44724328 44724331 44724336 44724337 44724338 44724339 44724341 44724342 44724343 44808636 44808704 44808789 44809085 44810509 44810708 44810770 44810778 44810782 44810785 44810789 44811205 44811209 44811210 44811213 44811216 44811654 44811659 44811663 44811678 44811686 44812245 44812254 44812260 44812264 44813115 44813131 44814405 44814412 44814415 44814419 44814420 44814719 44814725 44814744 44814758 44814772 44815180 44815182 44815190 44816612 44816616 44816622 44816628 44816717 44817221 44819124 44819189 44819356 44819358 44819362 44819379 44819380 44819824 44819829 44819831 44819841 44819842 44820197 44820199 44820202 44820204 44820213 44820694 44820702 44820743 44820748 44820760 44821049 44821056 44821065 44821087 44822770 44822803 44822869 44823032 44823033 44823035 44823131 44823132 44823489 44823490 44823492 44823493 44823494 44823849 44823850 44823851 44823852 44823853 44824351 44824352 44824353 44824354 44824357 44825439 44825441 44825442 44825490 44825502 44827341 44827345 44827347 44827349 44827350 44827837 44827873 44827884 44827889 44828222 44828232 44828248 44828276 44828287 44828736 44828742 44828747 44828753 44829178 44829274 44829282 44829284 37930966 37937889 37937891 37964585 37964629 38053210 38053215 38053216 38127140 38177275 38200443 38200449 38200450 38200454 38200457 38200458 38200460 38200461 38200464 38200466 38200467 38200470 38315330 38324210 38332351 38332358 38332363 38553528 38611410 38611413 38611414 38611513 38611516 38611520 38611526 38620459 38620477 38624129 38624139 38624159 38626963 38992044 41381457 41461823 42030828 42031181 42035887 42234953 42236111 42254393 42254533 42341063 38725795 38726794 38996390 39155785 39179746 39183640 39605738 39789368 39899275 39945422 40007558 40007616 40007631 40510101 40637418 40936255 41226025 41226113 41226208 41226237 41231213 41233042 41233060 41271494 41279728 41296424 41305745 41318099 41427420 41429551 41446078 41446141 41487039 41497916 41512831 41515603 41516694 41519408 41540620 41540666 41570099 41604841 41604922 41654005 41654282 41654345 41654788 41656506 41674111 41674162 41674307 41678338 41684772 41691823 41691867 41692694 41695182 41695624 41705025 41705108 41705299 41706041 41711217 41711755 41712335 41718328 41718337 41718340 41718346 41718381 41718473 41718492 41718728 41718833 41718864 41722390 41724135 41725830 41726541 41729815 41734558 41734562 41737005 41737226 41737266 41737342 41743223 41743267 41743754 41744012 41744271 41744578 41744717 41745387 41745465 41745553 41747989 41757305 41757338 41757445 41757486 41757531 41757543 41757550 41757555 41757570 41757588 41757638 41757656 41757724 41757772 41758546 41768150 41786086 41786895 41787027 41787280 41789985 41798006 41798616 41812972 41813387 41813599 41813629 41817635 41817692 41824231 41824763 41824888 41825107 41830054 41830074 41830165 41830177 41830230 41889930 41913104 41913324 41964222 41968740 41968959 41973530 41973610 41974195 41974213 41974241 41974244 41974293 41974363 41974364 41974405 41974408 41974439 41974450 41976126 41976150 41976176 41976220 41976226 41976274 41976288 41976289 41976305 41976334 41976417 41976435 41976446 41976475 41976501 41976545 41976595 41976635 41976679 41976695 41976723 41976810 41976839 41976944 41977065 41977097 41977238 41977242 41977268 41977299 41977328 41977356 41977866 41977867 41977896 41977903 41977943 41977955 41978045 41978064 41978073 41978076 41978107 41978110 41978138 41978174 41978185 41978249 41982467 41982474 41982499 41982535 41982556 41982572 41982578 41982602 41982627 41982691 41982730 41982763 41982790 41982839 41982943 41982944 41983021 41983029 41983045 41983067 41983112 41983123 41983164 41983236 41983253 41983293 41983336 41983337 41983379 41983381 41983433 41983451 41983453 41985853 41999828 41999931 42001891 42001931 42001965 42002001 42027727 42027785 42091190 42097492 42117708 42117809 42117877 42117971 42118056 42118113 42118202 42118469 42151791 42160293 42160424 42164212 42164272 42164499 42171926 42172274 42172430 42172446 42172580 42172720 42172769 42172785 42172818 42172902 42172982 42173096 42173125 42173130 42234530 42311895 42312129 42325507 42333369 42333754 42337196 42392562 42393577 42408156 42409148 42409209 42416743 42433736 42463048 42478282 42502275 42510935 42511114 42536410 42538491 42566045 42593259 42605826 42610149 42610285 42610349 42610942 42610957 42610983 42612141 42612146 42612338 42612353 42612434 42612515 42612541 42612554 42612586 42612597 42612676 42612710 42612821 42623126 42623876 42623976 42624669 42629352 42688960 42689029 42697842 42702092 42702137 42707794 42707893 42717777 42724319 42746778 42763931 42791392 42791428 42792005 42792239 42795710 42795725 42795780 42795842 42827364 42827413 42829752 42829919 42830056 42830156 42832872 42837744 42837871 42839645 42839656 42839660 42839743 42839884 42840211 42840590 42841377 42841749 42841804 42841830 42850826 42850872 42850954 42851160 42851161 42851349 42851395 42851399 42851415 42851423 42851486 42851491 42851513 42851514 42851714 42851880 42851888 42851897 42851923 42852605 42856552 42856596 42856634 42857009 42859129 42891158 42903035 42907188 42907193 42907228 42907262 42907299 42907375 42907844 42907859 42907891 42907893 42907946 42907983 42907984 42908036 42908059 42908084 42908206 42908242 42925177 42933689 42965031 42965804 42965877 42966457 42966502 43174175 43182991 37923213 37924152 38046563 38046569 38051963 38051979 38059582 38120781 38124046 38126269 38135212 38136516 38150694 38150695 38150696 38155913 38161892 38161912 38162531 38162551 38162604 38164407 38164420 38164425 38164434 38166510 38166549 38166580 38166678 38166679 38214247 38214259 38217387 38226965 38253773 38253786 38263329 38272213 38281759 38281774 38287524 38287532 38290141 38324199 38338635 38340548 38341984 38344810 38346560 38353020 38353034 38355627 38359027 38359035 38359559 38368315 38368329 38368611 38368612 38368619 38369586 38370274 38370280 38370436 38371714 38392634 38441930 38523566 38523569 38523572 38523591 38523605 38523606 38523609 38536343 38536918 38537149 38537153 38537160 38537161 38537173 38537760 38537763 38537766 38537768 38537774 38537778 38538721 38538936 38538953 38538969 38540933 38540960 38540975 38543103 38544623 38544625 38544658 38544666 38544746 38548004 38548026 38553529 38553545 38559431 38559447 38560018 38560032 38560037 38560041 38561026 38564050 38569281 38610372 38610375 38611530 38611536 38626967 38626991 38626996 38631725 38725754 38733838 38733858 38733859 38733862 38743122 38784149 38784168 38784929 38784954 38784957 38784967 38784968 38784969 38784972 38785810 38785820 38790368 38790370 38790373 38790376 38790377 38790380 38790385 38790386 38790388 38790399 38790407 38790415 38793539 38793551 38793558 38793559 38793565 38793573 38793574 38793575 38795898 38798718 38798766 38798778 38798800 38800559 38802812 38805589 38805618 38807503 38807505 38807508 38809028 38809032 38809033 38811713 38817545 38817561 38819711 38821750 38839961 38852583 38852585 38852928 38856589 38875812 38878669 38888162 38897292 38897297 38897298 38897497 38897498 38898616 38898619 38898623 38898631 38898635 38898636 38898638 38898640 38898641 38898649 38898650 38898652 38898656 38898658 38898660 38934844 38934846 38934847 38934849 38934851 38934853 38988241 38989260 38989265 38992036 38996374 38996393 38996790 38996804 38999812 39029067 39029140 39040605 39050455 39054571 39054586 39054589 39054599 39054602 39054622 39054634 39054644 39054656 39054673 39054771 39054772 39059486 39084707 39108559 39126753 39133699 39133709 39139847 39139852 39139853 39139858 39150421 39150426 39150430 39150879 39152708 39153388 39186413 39186415 39187295 39221096 39229416 39229418 39230355 39231961 39239054 39242835 39246169 39246170 39246192 39246922 39251183 39251187 39258084 39258086 39263616 39263630 39263634 39263640 39264443 39264448 39332163 39341454 39346977 39347756 39347770 39347774 39348720 39371898 39373372 39373862 39374681 39374682 39375607 39381603 39381615 39382548 39382555 39382562 39383360 39383369 39383373 39383698 39392120 39392128 39392841 39394294 39410413 39410421 39415128 39415149 39420157 39423276 39423637 39423663 39423674 39426099 39426105 39426114 39426118 39427830 39430840 39430848 39430854 39430880 39430886 39430889 39430912 39461903 39463918 39476696 39519737 39519919 39519922 39520124 39520128 39520156 39520157 39520158 39520160 39524167 39524642 39524643 39524654 39526032 39527946 39528211 39528240 39528241 39529109 39529110 39529111 39529529 39530546 39530556 39530558 39530575 39531614 39531721 39536081 39536090 39536429 39536433 39536446 39536793 39537513 39537514 39537516 39537521 39541656 39541669 39541674 39583929 39584166 39587203 39590280 39593716 39596671 39605735 39605781 39607437 39607451 39608343 39608346 39608383 39608396 39610114 39610118 39610123 39610124 39611631 39611636 39612259 39612268 39612307 39612311 39613539 39615771 39615844 39616912 39616915 39616917 39618496 39618512 39618519 39628855 39643275 39647496 39647513 39648904 39650220 39650968 39650972 39651175 39651183 39651705 39656213 39656930 39658109 39659445 39663698 39663699 39665037 39665556 39667315 39682855 39684274 39684756 39684762 39685659 39685710 39685732 39703334 39709435 39710654 39710662 39710663 39710885 39720924 39733349 39735436 39739935 39747184 39749482 39751612 39756863 39757725 39764273 39766913 39767365 39767367 39771557 39771558 39772016 39772075 39772183 39777446 39788419 39789989 39789990 39790000 39790009 39790017 39790018 39793767 39797631 39798400 39798403 39798410 39799072 39803437 39806967 39808072 39808080 39808082 39808085 39808086 39808089 39814124 39814126 39814789 39814790 39816926 39818659 39820872 39820888 39820895 39820896 39820897 39821952 39821961 39826031 39826032 39833318 39841196 39850397 39850400 39852170 39853492 39853493 39859129 39863270 39865494 39870142 39871030 39887853 39902178 39902185 39902193 39902809 39904496 39913030 39920927 39945572 39957319 39964405 39972429 39976667 39985615 39986155 39987030 39988138 39988156 39988310 40002974 40009554 40009563 40010797 40010800 40010806 40017298 40023308 40026157 40033602 40033605 40042960 40052072 40058199 40065172 40083827 40090121 40110956 40110964 40125501 40125502 40134937 40134941 40135329 40136549 40140401 40179684 40182748 40185871 40185910 40185917 40186662 40186665 40187644 40190983 40215365 40222803 40228252 40256025 40269823 40277082 40278159 40284285 40304593 40304980 40321713 40321723 40330099 40330108 40335500 40342635 40344401 40344404 40354035 40356312 40359607 40361084 40390610 40390621 40391378 40397102 40397126 40397909 40397918 40401278 40414738 40414828 40415297 40420851 40422138 40441228 40445686 40450033 40456301 40460022 40460029 40465342 40476122 40488247 40489974 40502984 40507261 40511593 40517211 40519019 40519023 40525750 40525755 40530697 40542092 40542328 40542354 40555528 40555531 40558326 40567916 40567924 40567930 40573769 40574490 40576431 40576433 40576434 40576436 40576437 40576438 40576439 40582731 40589191 40591013 40622825 40636795 40643402 40644630 40654112 40669733 40688077 40688080 40698230 40708032 40710414 40717669 40717768 40717975 40717978 40765559 40765739 40769689 40786352 40814974 40819549 40844996 40871523 40875662 40878383 40881108 40890228 40904946 40905568 40905928 40906142 40908128 40908708 40908826 40909514 40909929 40914629 40959998 40975212 40975232 40975360 40990679 40990799 40991059 40991357 40991678 40991800 40991806 40991927 40991932 40995197 40995201 40995215 40995216 40995266 40995461 40995517 40995558 40995586 40995785 41002111 41002944 41005366 41013857 41016180 41116735 41142062 41142796 41143229 41159769 41163823 41181937 41182371 41204321 41204371 41208701 41208871 41221006 41222255 41222292 41222294 41222437 41222449 41222622 41222761 41222839 41222926 41222931 41223054 41223056 41225018 41225111 41225138 41225223 41225250 41225359 41225360 41225371 41225425 41225427 41225443 41225577 41225581 41225588 41225644 41225648 41225680 41225844 41225909 41225950 41226061 41226114 41226181 41226217 41226221 41226285 41226379 41226383 41226396 41226399 41226401 41226424 41226432 41226439 41226458 41226505 41226506 41226526 41226614 41226621 41226634 41226655 41226682 41226755 41226767 41226854 41226877 41226902 41226909 41226975 41226987 41226994 41227019 41227075 41227089 41227090 41227121 41227126 41227132 41227179 41227199 41227200 41227201 41227287 41227298 41227404 41227472 41227512 41227559 41227563 41227591 41227641 41227644 41227739 41227740 41227774 41229111 41229121 41229128 41229140 41229177 41229211 41229267 41229313 41229334 41229359 41229360 41229365 41229368 41229386 41229387 41229395 41229401 41229407 41229437 41229441 41229457 41229525 41229536 41229576 41229603 41229604 41229897 41229911 41230009 41230036 41230048 41230081 41230103 41230126 41230211 41230214 41230253 41230308 41230331 41230371 41230441 41230507 41230523 41230525 41230582 41230616 41230653 41231146 41231231 41231266 41231294 41231337 41231352 41231357 41231375 41231378 41231475 41231485 41231504 41231601 41231605 41231652 41231681 41231763 41231771 41231806 41231809 41231869 41231879 41231912 41232006 41232031 41232058 41232060 41232088 41232147 41232148 41232158 41232194 41232198 41232316 41232429 41232447 41232457 41232463 41232560 41232591 41232761 41232780 41232781 41232922 41232983 41232993 41233000 41233020 41233021 41233049 41233061 41233826 41233827 41233857 41233858 41233859 41233902 41233914 41233975 41233991 41234017 41234055 41234071 41234116 41234119 41234152 41234210 41234211 41234257 41234283 41234416 41234425 41234454 41234455 41234496 41234500 41234514 41234521 41234530 41234546 41234560 41234561 41234572 41234579 41234582 41234677 41234701 41234704 41234709 41234712 41234717 41234771 41234781 41234783 41234814 41235046 41235053 41235076 41235082 41235096 41235114 41235126 41235145 41235155 41235183 41235184 41235185 41235256 41235260 41235277 41235287 41235301 41235383 41235388 41235465 41235470 41235489 41235494 41235545 41235585 41235656 41235660 41235675 41235690 41235729 41235783 41235874 41235900 41235927 41235993 41235994 41236036 41236118 41236241 41236274 41236275 41236280 41236511 41236528 41236530 41236562 41237660 41237668 41237723 41237724 41237743 41237744 41237753 41237766 41237789 41237807 41237816 41237820 41237906 41237916 41237927 41237929 41237932 41237937 41237940 41237941 41238075 41238108 41238115 41238117 41238160 41238192 41238202 41238238 41238245 41238273 41238282 41238288 41238311 41238313 41238332 41238334 41238373 41238374 41238383 41238475 41238483 41238521 41238595 41238693 41238818 41238856 41238929 41238979 41239984 41242302 41243001 41243110 41243190 41243248 41243319 41243325 41243353 41243363 41243378 41243454 41243483 41243521 41243622 41243632 41243661 41243700 41243708 41243727 41243732 41243736 41243855 41243856 41243874 41243942 41243944 41243979 41244047 41244104 41244176 41244251 41244319 41244353 41244437 41244480 41244644 41244826 41244920 41245089 41245091 41246834 41246835 41246871 41246874 41246893 41246901 41246918 41246958 41246963 41246979 41247000 41247004 41247010 41247246 41247269 41247270 41247296 41247311 41247332 41247341 41247392 41247492 41247518 41247540 41247561 41247636 41247641 41247666 41247691 41247704 41247727 41247755 41247763 41247771 41247797 41247812 41247813 41248016 41248017 41248085 41248105 41248126 41248136 41248185 41248231 41248236 41248247 41248249 41248293 41248314 41248316 41248318 41248323 41248363 41248385 41248429 41248432 41248442 41248499 41248568 41248572 41248680 41248697 41248744 41248745 41248780 41248846 41248886 41248894 41248907 41248911 41248996 41249091 41249093 41250294 41250302 41250338 41250345 41250362 41250367 41250440 41250441 41250467 41250469 41250525 41250528 41250622 41250748 41250775 41251092 41251146 41251178 41251211 41251321 41251348 41251419 41251436 41251500 41251501 41251550 41251757 41251786 41251915 41251917 41251952 41252013 41253282 41253428 41253555 41253556 41253740 41253751 41253807 41253810 41253882 41254022 41254098 41254183 41254197 41254439 41254507 41254509 41254597 41254662 41254902 41257104 41257108 41257122 41257160 41266722 41266823 41267636 41267680 41268049 41270149 41270538 41270570 41270577 41270808 41270819 41270846 41270877 41270899 41270931 41270936 41270944 41270965 41270998 41271030 41271065 41271073 41271471 41271495 41271565 41271569 41271578 41271604 41271641 41271678 41271796 41271801 41271848 41271877 41271906 41272094 41272108 41272116 41272170 41272207 41272243 41272315 41272317 41272398 41273990 41273993 41274084 41274200 41274310 41274369 41274381 41274582 41274604 41274625 41274641 41274645 41274689 41274913 41275128 41275309 41275478 41275536 41275579 41275597 41275674 41275783 41275910 41276502 41278400 41278432 41278460 41278579 41278813 41278826 41278841 41278952 41279001 41279004 41279052 41279143 41279214 41279250 41279266 41279461 41279560 41279613 41279656 41279730 41279817 41279864 41279984 41280000 41280124 41280442 41282923 41283073 41283164 41283217 41283358 41283623 41283697 41283972 41284379 41284925 41285085 41285410 41285547 41287732 41287811 41287882 41287982 41288039 41288057 41288061 41288111 41288360 41288389 41288400 41288425 41288446 41288457 41288481 41288488 41288548 41288573 41288631 41288636 41288664 41288687 41288692 41288795 41288798 41288808 41288833 41288843 41288848 41288876 41288975 41289016 41289033 41289046 41289051 41289130 41289135 41289146 41289241 41289246 41289261 41289274 41289318 41289503 41289564 41289589 41289754 41291384 41291445 41291458 41291518 41291542 41291573 41291581 41291650 41291675 41291728 41291748 41291765 41291896 41292065 41292089 41292225 41292267 41292269 41292313 41292327 41292457 41292458 41292471 41292503 41292594 41296732 41296755 41296871 41296898 41297071 41297140 41297201 41297252 41297425 41299977 41300070 41300111 41300126 41300143 41300161 41300176 41300194 41300418 41300486 41300490 41300498 41300512 41300533 41300537 41300575 41300591 41300594 41300641 41300730 41300806 41300837 41300863 41300962 41301071 41301090 41301156 41301255 41301361 41301369 41304208 41304733 41304739 41304768 41304778 41304807 41304896 41304903 41304984 41305061 41305129 41305331 41305402 41305409 41305810 41306028 41309213 41309244 41309274 41309310 41310330 41310360 41310501 41310522 41310580 41310646 41310715 41310756 41310833 41310900 41311103 41313838 41313889 41314008 41314078 41314167 41314246 41314287 41314313 41314407 41314652 41314748 41314805 41314815 41315014 41315078 41315135 41315155 41315209 41315312 41315403 41317539 41317632 41317667 41317724 41317750 41317817 41317946 41317997 41318061 41318190 41318800 41318955 41319095 41320864 41321325 41321440 41321499 41321698 41321810 41321915 41325892 41325918 41326006 41326106 41326127 41326173 41326416 41326447 41326458 41326488 41326518 41326546 41326583 41326686 41327153 41327161 41327240 41328006 41328017 41328132 41328136 41328519 41328641 41328708 41328768 41328893 41329144 41329226 41329243 41329388 41329726 41329735 41329740 41329798 41329805 41329807 41329916 41329939 41330010 41330018 41330053 41330064 41330105 41330233 41330256 41330258 41330405 41330571 41330596 41330655 41330710 41332208 41332230 41332308 41332404 41332433 41332454 41332472 41332478 41332509 41332545 41332551 41332584 41332625 41332636 41332657 41332733 41332739 41332791 41333135 41333411 41333694 41333736 41333785 41333869 41333870 41333872 41333891 41334461 41334865 41334920 41334945 41334946 41334999 41335417 41335424 41335561 41335578 41335610 41335708 41335799 41335803 41335867 41336365 41365815 41370936 41371073 41371194 41371241 41371256 41371274 41371323 41371475 41371495 41371559 41371640 41371796 41371804 41371886 41371922 41371955 41371986 41372031 41373678 41373707 41373754 41373760 41373772 41373802 41373866 41374103 41374550 41374597 41376643 41378327 41378436 41378487 41378566 41378774 41378798 41381548 41381584 41381669 41381752 41387260 41387281 41387426 41387481 41390245 41390293 41390306 41390544 41390696 41392991 41393130 41393709 41393807 41393895 41393945 41394005 41394262 41394371 41394467 41394594 41395218 41396841 41396936 41396962 41396988 41397051 41397090 41397111 41397122 41397235 41397246 41397265 41397295 41397297 41397386 41397537 41397625 41398668 41398727 41398745 41405105 41405664 41405685 41405861 41406131 41407755 41407765 41407771 41407779 41407797 41407802 41407875 41407898 41408011 41408066 41408145 41408178 41408280 41408304 41408328 41408334 41408513 41408514 41408552 41408570 41408645 41408654 41408675 41408762 41408763 41408812 41408827 41408850 41409846 41409865 41409934 41409936 41410013 41410018 41410093 41410155 41410245 41411970 41412065 41412127 41412149 41412158 41412169 41412286 41412313 41412324 41412640 41412648 41412689 41412828 41412830 41412875 41412888 41412900 41412925 41412933 41412974 41412985 41413026 41413039 41413161 41413206 41413324 41413393 41413517 41413871 41419374 41419421 41419446 41419462 41419473 41419476 41419490 41419673 41419721 41419732 41419750 41419772 41419774 41419818 41419834 41419838 41419873 41419891 41420312 41420374 41420451 41420458 41420523 41420565 41420580 41420681 41420755 41420832 41423076 41423317 41423386 41424459 41424553 41426929 41427483 41430155 41431745 41436749 41436790 41436904 41436915 41437221 41437272 41437284 41437287 41437326 41437667 41437698 41437699 41437957 41437988 41437989 41438458 41438460 41438486 41438491 41438497 41438560 41438668 41439912 41439924 41440029 41440100 41440176 41440604 41440605 41440622 41440643 41440644 41440684 41440688 41440699 41440738 41440753 41440811 41440817 41440834 41440843 41440865 41440871 41440878 41440897 41440909 41440919 41440986 41440996 41441030 41441097 41441102 41441109 41441140 41441146 41441150 41441177 41441196 41441205 41441261 41441273 41442667 41442730 41442740 41442756 41442757 41442781 41442789 41442827 41442849 41442852 41442860 41442867 41442916 41442963 41442967 41443087 41443114 41443123 41443142 41443146 41443349 41443388 41443390 41443397 41443412 41443427 41443428 41443455 41443460 41443481 41443530 41443541 41443548 41443636 41443653 41443657 41443697 41443702 41443713 41443714 41443726 41443734 41443789 41443823 41443940 41443965 41443969 41444038 41444070 41445624 41445646 41445648 41445653 41445658 41445679 41445681 41445704 41445715 41445726 41445729 41445741 41445751 41445801 41445803 41445816 41445822 41445924 41445971 41446026 41446039 41446044 41446167 41446668 41446693 41446867 41446882 41446937 41446975 41446977 41447020 41447027 41452848 41452862 41452929 41452979 41453114 41453209 41453239 41453331 41454196 41454387 41454404 41458051 41458124 41458157 41458191 41458200 41458239 41458541 41461146 41461253 41461323 41461423 41461472 41461771 41461798 41461968 41462040 41462068 41462154 41462469 41462527 41462569 41464906 41465010 41465037 41465106 41465121 41465296 41465657 41466062 41468911 41469005 41469128 41469413 41469439 41469511 41469580 41470392 41470534 41473583 41473647 41473681 41474283 41474347 41474398 41474458 41474527 41474635 41474760 41474887 41474954 41475079 41476798 41476816 41476880 41476884 41476898 41476904 41476949 41477039 41477104 41477111 41477282 41477292 41477348 41477356 41477427 41477455 41477473 41477491 41477506 41477562 41477587 41477648 41477652 41477709 41477715 41477771 41477774 41477815 41477826 41477931 41477957 41478003 41478530 41478565 41478608 41478639 41478680 41478737 41478762 41478808 41478812 41478898 41478968 41479071 41479142 41479143 41479207 41479275 41479292 41479345 41481146 41481180 41481190 41481223 41481247 41481251 41481294 41481306 41481335 41481366 41481368 41481384 41481389 41481492 41481517 41481551 41481571 41481586 41481604 41481665 41481693 41481700 41481717 41481754 41481759 41481780 41481781 41481811 41481846 41481884 41481890 41482203 41482261 41482340 41482372 41482435 41482460 41482557 41482635 41482748 41482756 41482867 41482904 41483066 41483080 41483138 41483235 41483308 41483456 41483459 41485407 41485426 41485446 41485472 41485486 41485516 41485576 41485597 41485605 41485634 41485776 41486008 41486009 41486086 41486197 41486216 41486227 41486231 41486277 41486281 41486295 41486310 41486458 41487084 41487291 41487337 41487381 41489239 41489252 41489263 41489311 41489379 41489444 41489477 41489578 41489603 41489643 41489688 41489728 41489729 41489751 41489758 41489774 41489775 41489798 41489886 41489998 41490048 41490281 41490323 41490337 41490382 41490392 41490472 41490538 41490575 41490632 41490642 41490688 41490788 41497022 41497039 41497069 41497090 41497117 41497181 41497186 41497225 41497235 41497238 41497247 41497250 41497309 41497464 41497467 41497490 41497665 41497777 41498723 41500658 41501137 41501214 41501226 41501231 41501363 41501424 41501705 41501732 41501760 41501803 41503612 41503693 41504230 41504260 41504280 41504429 41504691 41512776 41515105 41515158 41515161 41515379 41515382 41515579 41515753 41516058 41516169 41516456 41516532 41516534 41516578 41516639 41516676 41516685 41516781 41516829 41516835 41516974 41516997 41517016 41517062 41517083 41517131 41517199 41517212 41517306 41518908 41518938 41518952 41519013 41519017 41519042 41519078 41519105 41519163 41519166 41519200 41519223 41519265 41519282 41519291 41519464 41519650 41520944 41522698 41522723 41523340 41523393 41523396 41525289 41525953 41525969 41526292 41526868 41526869 41527011 41527211 41527240 41527326 41527359 41536134 41536231 41536236 41536312 41536327 41536347 41536350 41536365 41536366 41536404 41536424 41536534 41536538 41536547 41536566 41536596 41536686 41536698 41536728 41536751 41536763 41536766 41538127 41538128 41538152 41538181 41538193 41538395 41538409 41538452 41538496 41538635 41538646 41540553 41540639 41540671 41540678 41540697 41540766 41540947 41541002 41541006 41541274 41541303 41541547 41541655 41541684 41541712 41541854 41542073 41542096 41542109 41542208 41542212 41542257 41542310 41542319 41542337 41542410 41542446 41542511 41542693 41542737 41542743 41545461 41545583 41545653 41545706 41545796 41545856 41545916 41546310 41546425 41546586 41546748 41546853 41546988 41549177 41549288 41549342 41549428 41549607 41549636 41549829 41549874 41549896 41549970 41550065 41550092 41550192 41550295 41552285 41552300 41552363 41552508 41552700 41552811 41552957 41553015 41553032 41553092 41553134 41553162 41553217 41553376 41553441 41553505 41553614 41553650 41553822 41553923 41553978 41554068 41554150 41557272 41557401 41557402 41557471 41557505 41557511 41557551 41557587 41557599 41557672 41557759 41557804 41557854 41557869 41557876 41557884 41557890 41557910 41557926 41557938 41558044 41558048 41558062 41558068 41558095 41558110 41558151 41558154 41558177 41558383 41558429 41558449 41558554 41558611 41558891 41558963 41559028 41559158 41560870 41560918 41560964 41561058 41561116 41561147 41561239 41561329 41561539 41561633 41561689 41561724 41561762 41561812 41561888 41561989 41562175 41562256 41562298 41562353 41562398 41562451 41562478 41562526 41562536 41562615 41562616 41562653 41562660 41562733 41562794 41565068 41565143 41565190 41565284 41565353 41565439 41565662 41565807 41565820 41565851 41565860 41565875 41565914 41565933 41565940 41565942 41565974 41565977 41566006 41566058 41566392 41566414 41566449 41566457 41566483 41566487 41566498 41566506 41566572 41566596 41566614 41566641 41566695 41566808 41566857 41566859 41566961 41567132 41567168 41567248 41567310 41570065 41570128 41570137 41570175 41570180 41570201 41570239 41570251 41570265 41570298 41570315 41570772 41571220 41571222 41571551 41571561 41571588 41571882 41572281 41572338 41572374 41572377 41572557 41572567 41572630 41572636 41572707 41572734 41572790 41572796 41572860 41572870 41572915 41572921 41573093 41573107 41573155 41573214 41573217 41573315 41573391 41573500 41573779 41573816 41573849 41573918 41573952 41574033 41574107 41574210 41576515 41576549 41576563 41576570 41576580 41576611 41576620 41576638 41576642 41576709 41576743 41576815 41576821 41576939 41576940 41576948 41576950 41576971 41577009 41577019 41577066 41577138 41577325 41577353 41577373 41577427 41577446 41577473 41577474 41577539 41577843 41577850 41578108 41578119 41578163 41578165 41578207 41578239 41578462 41578552 41578850 41578855 41578883 41578887 41579095 41579277 41585156 41585182 41589664 41590417 41590446 41590539 41591869 41591885 41591959 41591980 41591994 41592056 41592110 41592158 41592209 41592244 41592275 41598927 41599067 41599107 41599111 41599171 41599182 41599208 41599230 41599330 41599332 41599383 41599432 41599486 41601150 41601171 41601180 41601197 41601251 41601465 41601510 41601535 41601618 41601649 41601709 41601822 41601833 41601865 41601889 41602008 41602067 41602496 41602505 41604639 41604640 41604716 41604894 41605050 41605079 41605122 41605946 41606013 41606416 41611022 41618640 41618849 41618924 41619255 41619279 41619321 41619372 41619417 41619597 41619619 41619666 41619718 41619798 41619923 41619988 41620042 41620071 41621416 41621657 41621673 41623926 41624056 41624180 41624216 41624248 41624373 41624519 41624612 41629804 41629817 41629859 41629889 41629899 41629926 41630520 41630562 41630656 41630766 41630918 41631255 41631275 41631323 41631463 41631846 41631940 41632013 41634715 41634876 41635336 41635407 41635445 41635805 41635864 41636093 41636263 41636634 41636765 41636771 41636811 41636834 41637213 41638231 41638391 41638463 41638538 41638680 41638765 41638862 41638932 41639028 41639209 41639241 41639278 41639336 41639399 41639467 41639512 41639725 41639835 41639961 41641945 41641981 41642021 41642035 41642049 41642064 41642138 41642174 41642305 41642321 41642406 41642451 41642514 41642682 41642762 41642888 41643110 41643162 41643264 41643316 41643358 41643687 41643827 41644072 41644922 41645141 41645156 41645198 41645306 41645683 41645807 41646017 41646167 41646492 41646906 41653437 41653729 41653772 41653788 41653897 41654376 41654407 41654488 41654571 41656611 41656693 41656839 41659879 41659897 41663280 41663534 41663767 41663860 41664025 41664099 41664229 41664447 41664527 41665289 41665351 41670211 41670229 41670795 41670866 41670902 41670934 41671001 41671051 41671779 41671828 41671851 41671895 41671946 41671964 41671989 41672116 41672239 41672388 41672651 41673906 41674044 41674268 41674737 41675020 41675198 41675730 41676278 41677459 41677722 41678058 41678154 41678238 41678368 41678438 41678546 41679593 41679905 41679947 41684769 41684791 41684854 41684946 41684964 41685006 41685042 41685054 41685056 41685343 41685396 41685401 41685417 41685425 41685453 41685487 41685570 41685583 41685603 41685626 41685628 41685653 41685951 41685966 41685974 41686004 41686008 41686038 41686083 41686175 41686196 41686505 41686788 41688085 41688086 41688225 41688649 41689628 41692203 41692411 41692593 41694899 41703071 41703198 41703232 41703269 41704095 41704129 41704442 41704551 41704578 41704583 41704713 41704718 41704814 41704873 41704923 41705033 41705149 41705156 41705275 41705352 41705373 41705414 41705613 41705635 41705652 41705664 41705679 41705706 41705710 41705730 41705771 41705828 41705920 41705927 41706016 41706050 41706105 41706161 41706235 41706344 41706402 41707865 41707908 41708020 41708298 41708635 41708654 41708690 41708783 41708822 41708823 41708994 41709006 41709040 41709045 41709074 41709088 41709090 41709170 41709188 41709259 41709313 41709357 41709360 41709381 41709405 41709460 41709483 41709574 41709584 41709646 41709659 41709737 41710882 41710886 41710917 41710939 41711011 41711090 41711167 41711197 41711202 41711205 41711216 41711295 41711371 41711377 41711412 41711427 41711433 41711500 41711565 41711642 41711649 41711664 41711738 41711829 41711909 41711961 41712108 41712165 41712314 41712362 41718500 41718503 41718516 41718577 41718605 41718690 41718907 41718984 41722094 41722137 41722247 41722294 41722661 41722798 41723326 41723334 41723448 41723490 41723531 41723625 41723746 41723868 41723974 41724051 41724088 41724169 41724423 41724432 41725496 41725501 41725507 41725569 41725622 41725705 41725767 41725768 41725811 41725815 41725825 41725840 41725844 41726040 41726063 41726093 41726106 41726155 41726156 41726251 41726270 41726322 41726324 41726359 41726361 41726381 41726463 41726620 41726645 41726781 41726785 41726816 41727277 41727369 41727398 41727461 41728162 41729850 41729903 41729945 41730034 41730075 41730078 41730228 41730310 41730525 41730532 41730538 41730589 41730599 41730665 41730870 41730875 41730892 41730924 41730925 41730997 41731008 41731020 41731145 41731155 41733317 41733322 41733356 41733385 41733440 41733824 41733852 41733874 41733881 41733915 41733940 41734004 41734042 41734068 41734199 41734265 41734320 41734325 41734357 41734374 41734386 41734518 41734620 41734665 41734668 41734706 41734727 41734800 41734809 41734873 41734938 41734971 41735059 41735066 41736852 41737081 41737142 41737171 41737172 41737204 41737272 41737431 41738350 41738443 41738565 41738652 41738692 41743183 41743342 41743396 41743434 41743466 41743567 41743616 41743656 41743705 41743819 41743846 41743897 41743949 41744089 41744262 41744438 41744667 41745157 41745250 41747651 41747720 41747761 41747793 41747830 41747834 41747897 41747904 41748017 41748047 41748120 41748284 41748360 41748654 41748719 41748721 41748749 41748758 41748815 41749307 41749447 41749678 41749735 41749910 41757021 41757117 41757232 41757249 41757255 41757264 41757295 41757369 41757510 41757564 41757698 41757762 41757792 41757870 41757899 41757965 41757967 41758051 41758086 41758092 41758187 41758240 41758260 41758307 41758319 41758462 41758552 41759111 41760522 41760537 41760619 41760687 41760788 41760797 41760838 41760840 41760900 41760902 41760924 41760930 41760991 41761120 41761152 41761186 41761236 41761362 41761499 41761598 41761602 41761650 41761810 41761947 41762004 41762072 41762120 41764325 41764362 41764413 41764472 41764523 41764574 41764636 41764797 41764835 41764931 41765002 41765228 41765443 41765492 41765589 41765631 41765711 41766055 41766365 41766506 41767681 41767728 41767913 41768175 41768700 41768938 41768943 41769012 41769065 41769451 41775664 41775684 41775992 41776020 41776316 41777216 41777448 41777565 41777600 41777664 41777665 41778383 41780315 41780424 41780513 41780800 41781039 41781077 41781163 41781254 41781799 41781855 41781975 41782169 41782263 41782337 41785266 41785322 41785482 41785486 41785567 41785649 41785656 41785767 41785881 41785887 41786090 41786816 41786910 41786970 41787026 41787247 41787316 41788294 41788303 41788408 41788600 41789192 41789725 41789768 41790271 41790272 41790314 41791720 41792135 41793033 41793099 41793110 41793657 41793753 41793893 41793987 41794038 41794096 41795095 41795201 41795255 41795363 41795539 41795651 41795847 41795915 41796085 41796087 41797809 41797855 41797910 41798013 41798056 41798372 41798509 41798556 41798767 41798898 41798996 41799118 41799211 41799305 41799529 41800134 41800680 41801057 41801438 41808228 41808413 41808499 41808557 41808734 41808989 41809109 41809200 41809229 41809308 41809475 41810573 41810913 41810996 41811129 41812661 41812734 41812802 41812925 41812985 41813068 41813261 41813442 41813516 41813577 41813710 41813842 41813897 41814012 41814072 41814211 41814351 41814445 41814784 41816881 41816969 41817118 41817326 41818140 41818188 41818256 41818432 41818584 41818674 41818701 41818705 41818727 41818812 41818815 41818868 41818935 41818967 41819032 41819339 41819406 41819443 41820579 41820685 41820750 41820873 41820910 41820944 41820980 41821206 41821278 41821335 41821345 41821422 41821707 41821798 41822030 41822127 41822153 41822334 41822350 41822351 41822519 41822526 41822549 41822574 41822590 41822741 41822749 41822791 41822807 41824295 41824310 41824321 41824345 41824363 41824407 41824430 41824457 41824502 41824513 41824565 41824748 41825040 41825085 41825146 41825263 41825329 41825429 41825620 41825984 41826065 41826237 41826703 41826746 41826796 41826845 41826922 41827123 41827203 41827210 41845483 41845655 41845941 41846055 41846628 41847781 41848346 41849004 41849072 41849212 41850754 41850798 41850871 41850901 41851029 41851104 41851143 41851222 41851408 41851500 41851561 41851632 41851935 41851966 41852093 41852122 41852153 41852268 41852308 41852336 41852387 41852486 41852723 41853950 41854125 41854269 41854322 41854348 41866602 41866719 41866854 41866947 41867402 41874076 41874200 41874223 41874267 41874302 41874778 41875083 41877499 41877552 41877575 41877703 41877796 41877825 41877883 41878051 41878191 41878291 41878369 41878461 41878681 41878870 41881747 41881910 41881967 41882078 41882112 41882249 41882386 41882725 41882830 41882989 41883037 41883116 41883436 41888529 41888623 41888818 41888858 41888886 41888958 41889264 41889377 41889469 41889518 41889548 41889595 41889613 41889768 41889822 41889915 41890166 41890260 41890314 41890365 41890425 41892005 41892032 41892059 41892118 41892169 41892223 41892306 41892445 41892457 41892460 41892526 41892531 41892546 41892604 41892640 41892652 41892716 41892751 41892823 41892853 41892867 41892902 41893055 41893176 41893323 41893394 41893416 41893445 41893519 41893577 41893629 41893784 41893792 41893891 41893998 41894130 41894390 41895622 41895641 41895762 41895938 41895986 41896078 41896145 41896223 41896777 41896848 41896967 41896991 41897119 41897195 41897305 41897528 41897562 41898931 41899150 41899169 41899213 41899241 41899346 41899408 41899440 41899565 41899576 41899604 41899628 41899646 41899712 41900104 41900146 41900147 41900179 41900196 41900312 41900349 41900395 41900463 41900488 41900517 41900546 41900551 41900636 41900673 41902643 41902744 41902956 41903077 41903287 41907818 41908077 41908080 41908132 41908320 41908433 41908537 41908606 41908647 41908648 41908861 41908863 41908886 41908998 41909026 41909154 41909185 41909261 41909615 41909763 41909826 41911401 41911619 41911694 41911721 41911854 41911926 41912048 41912191 41912224 41912319 41912443 41912514 41913118 41913390 41913436 41917752 41917875 41917966 41918054 41918095 41918108 41918160 41918289 41918368 41918473 41918524 41918624 41923772 41923877 41923945 41923998 41924014 41924283 41924416 41924559 41925405 41925517 41926770 41926853 41927178 41927319 41928611 41928661 41928745 41928797 41928847 41928888 41928951 41928981 41929077 41929144 41929215 41929284 41929299 41929321 41929548 41929579 41932420 41933929 41935870 41935921 41935943 41936052 41936153 41936604 41941378 41942649 41942844 41942889 41943048 41943260 41955224 41955289 41955336 41956048 41956089 41956320 41956328 41956405 41958804 41958846 41958931 41959090 41959234 41959503 41960715 41961025 41961102 41961192 41961454 41961550 41961922 41961947 41962027 41964228 41964362 41964446 41964476 41964492 41964708 41964772 41964807 41964895 41964904 41964935 41964943 41964989 41965026 41965048 41965253 41965279 41965304 41965343 41965357 41965359 41965363 41965404 41965435 41965466 41965725 41966137 41966196 41966226 41967757 41967771 41968174 41968285 41968544 41968689 41968770 41968920 41969062 41969084 41969116 41969153 41969175 41969231 41969312 41969394 41969463 41969509 41969604 41969675 41972378 41972391 41972419 41972564 41972650 41972804 41972805 41972844 41972913 41972918 41972990 41972997 41973074 41973118 41973133 41973136 41973162 41973188 41973258 41973261 41973281 41973328 41973388 41973416 41973420 41973491 41973495 41973534 41973634 41973698 41973714 41974184 41974285 41976599 41976813 41976872 41976923 41977034 41977267 41977302 41978177 41982161 41982163 41982188 41982189 41982228 41982245 41982302 41982358 41982377 41982540 41982856 41983070 41983223 41983286 41983561 41983562 41983638 41983662 41983709 41983747 41983758 41983765 41983771 41983841 41983842 41983918 41983921 41984121 41985818 41985842 41985875 41985899 41985939 41987283 41987906 41988005 41988105 41990737 41990794 41990918 41991101 41991211 41991254 41991301 41991350 41991502 41991560 41991603 41991647 41991690 41991882 41991927 41991941 41992164 41992199 41998878 41998968 41998982 41998997 41999097 41999145 41999213 41999261 41999267 41999281 41999285 41999301 41999309 41999322 41999341 41999363 41999365 41999382 41999514 41999620 41999631 41999684 41999691 41999731 41999737 41999745 41999750 41999829 41999870 41999872 41999904 41999912 41999920 41999969 41999970 42000008 42000010 42000041 42000414 42000418 42000484 42000537 42000542 42000565 42000567 42001773 42001778 42001789 42001801 42001845 42001954 42002000 42002080 42002181 42002211 42002238 42002264 42002304 42002312 42002485 42002486 42002521 42002527 42002558 42002651 42002654 42002675 42002715 42003011 42003069 42003185 42003246 42003268 42003272 42003290 42003308 42003363 42003446 42003447 42003594 42003609 42003668 42004974 42005011 42005028 42005056 42005082 42005150 42005198 42005205 42005242 42005276 42005278 42005329 42005344 42005352 42005439 42005456 42005495 42005502 42005527 42005539 42005580 42005588 42005656 42005672 42005752 42005905 42006008 42006015 42006091 42006252 42006274 42006481 42006545 42006591 42006643 42007672 42007777 42007832 42007902 42007980 42008010 42008040 42008068 42008714 42009217 42009236 42009334 42009613 42009709 42009733 42009755 42018472 42018506 42018521 42018555 42018705 42018816 42018927 42019061 42019128 42019247 42019794 42020475 42022310 42022420 42022504 42022555 42022633 42022683 42022817 42022943 42022971 42023406 42023516 42023924 42024678 42026499 42026562 42026693 42026751 42026874 42027902 42028010 42028181 42029654 42029713 42030487 42031255 42031260 42031306 42031341 42035480 42035494 42035571 42035597 42035665 42035738 42035811 42035824 42035839 42035849 42035905 42035929 42035978 42036040 42036050 42036141 42036185 42036241 42036273 42036291 42036565 42036630 42036701 42036927 42036992 42036994 42037008 42037013 42037084 42037193 42037373 42037380 42037491 42037601 42037634 42037801 42039230 42039233 42039250 42039255 42039263 42039314 42039540 42039668 42039707 42040819 42040850 42042055 42042073 42042131 42042249 42042332 42042407 42042410 42042472 42042513 42042547 42042624 42042706 42042794 42042851 42042917 42042976 42043062 42043113 42043142 42043200 42043467 42043477 42043478 42043520 42043542 42043557 42043599 42043604 42043659 42043724 42043853 42043855 42043875 42043876 42043917 42043919 42043936 42044018 42044031 42044071 42045532 42045550 42045556 42045602 42045606 42045704 42045722 42045777 42045791 42045849 42045867 42046043 42046052 42046213 42046228 42046259 42046292 42046335 42046348 42046372 42046431 42046504 42046540 42046588 42046634 42046694 42046705 42046788 42046790 42046821 42046824 42046865 42046874 42046979 42047039 42047096 42047106 42047160 42047161 42048824 42049226 42049557 42051849 42051944 42052165 42053280 42053305 42053317 42053383 42053687 42056075 42056096 42056109 42056123 42057185 42057386 42057596 42057664 42057709 42057773 42057851 42057926 42058092 42058169 42058374 42059006 42059041 42059121 42059210 42059561 42059853 42061243 42061357 42061521 42061578 42061672 42061811 42061912 42061958 42062099 42062719 42062996 42063075 42063249 42064597 42064639 42064691 42064821 42065087 42065228 42065375 42065436 42065452 42065456 42065475 42065494 42065500 42065601 42065678 42065734 42065737 42065863 42065940 42066035 42066040 42066331 42066341 42066362 42066376 42066379 42066432 42066444 42066460 42066465 42066504 42066527 42066573 42066615 42066627 42066783 42066788 42066890 42066936 42066964 42066991 42067013 42067014 42067058 42067067 42067112 42067154 42067185 42067254 42067258 42067312 42067513 42067514 42067564 42067567 42067592 42067593 42067609 42067651 42067656 42067659 42067695 42067697 42067712 42067713 42067715 42067757 42067785 42067825 42068309 42068311 42068314 42068315 42068322 42068325 42068334 42068340 42068341 42068342 42068345 42068369 42068384 42068385 42068410 42068413 42068442 42068446 42068456 42068464 42068482 42068492 42068497 42068514 42068523 42068524 42068526 42068612 42068614 42068620 42068623 42068627 42068635 42068673 42068674 42068728 42068735 42068737 42068768 42068773 42068809 42068823 42068828 42068861 42068862 42069086 42069200 42069208 42069261 42070121 42070215 42070274 42070278 42070493 42070674 42070811 42071187 42071917 42076929 42083397 42083780 42090689 42090816 42090879 42090893 42090911 42090956 42091014 42091092 42091293 42091338 42097624 42097701 42101663 42101728 42102306 42103829 42103865 42103868 42103874 42104265 42104517 42104756 42104786 42104819 42104965 42104999 42105162 42105216 42105314 42105550 42110771 42110975 42111193 42111512 42112171 42112213 42112259 42112794 42112873 42114028 42114353 42115026 42115563 42115793 42115952 42116026 42117678 42118423 42118614 42118810 42122099 42122542 42122856 42123419 42123477 42123589 42133632 42133893 42133966 42134021 42134077 42134092 42134222 42134264 42134345 42134356 42134412 42134416 42134446 42134511 42134524 42134564 42134580 42134633 42134696 42135228 42135259 42135278 42135322 42135335 42135381 42135386 42135438 42135472 42135495 42135496 42135570 42135583 42135633 42135660 42135688 42135735 42135779 42135780 42137330 42137549 42137797 42138879 42138881 42138925 42138972 42139025 42139069 42139070 42139187 42139190 42139226 42139259 42139287 42139403 42142724 42142760 42142859 42142916 42142925 42142979 42143147 42143191 42143531 42143847 42143848 42144016 42144022 42144050 42144061 42144072 42144112 42144185 42144255 42144351 42144474 42144770 42144831 42144832 42144848 42144899 42146065 42146084 42146254 42146280 42146345 42146433 42146484 42146552 42146559 42147246 42147898 42148021 42148103 42148171 42148208 42148236 42148265 42148299 42148310 42148423 42148442 42148527 42148626 42148665 42148707 42148723 42148825 42151478 42151492 42151533 42151794 42152852 42152937 42154861 42154871 42154955 42155073 42155234 42155372 42155410 42155635 42155731 42156194 42156374 42159386 42159582 42159669 42160245 42172127 42173208 42173286 42173299 42173529 42173808 42175199 42175319 42175431 42175496 42175647 42175668 42175720 42175774 42176146 42176194 42178162 42178224 42178276 42178296 42178372 42178440 42178740 42178807 42179183 42182229 42182250 42182272 42183587 42183631 42183672 42183724 42183992 42190364 42190423 42190542 42190556 42190606 42190630 42190650 42190666 42190716 42190790 42190834 42190894 42190919 42191140 42191400 42191490 42191602 42191752 42191767 42191821 42191909 42191939 42191981 42192056 42192106 42192166 42192197 42192254 42192326 42193577 42193616 42193656 42193687 42193745 42193840 42193892 42193910 42193936 42193972 42194012 42194056 42195899 42195947 42196090 42197977 42198017 42198128 42198157 42198168 42198194 42198314 42198561 42198692 42198756 42198775 42198836 42199087 42199173 42199188 42199626 42200047 42200099 42200565 42200620 42202641 42202660 42203033 42203139 42203185 42203202 42203283 42203333 42203602 42203701 42203791 42203867 42203907 42203933 42203982 42204119 42204189 42204216 42204225 42204296 42210918 42211138 42211247 42211311 42211410 42211445 42211842 42212200 42212527 42212802 42214087 42214099 42214179 42214333 42214375 42214438 42214558 42214899 42215195 42215729 42217992 42218047 42218087 42218118 42218194 42218221 42218278 42218313 42218347 42218444 42218536 42218984 42218998 42219048 42219060 42219109 42219168 42219173 42219215 42219270 42219361 42219367 42219384 42219399 42219406 42219457 42219569 42219596 42219843 42220784 42220792 42220795 42220808 42220864 42220919 42220921 42220933 42220950 42220957 42220961 42220966 42221087 42221131 42221204 42221235 42221292 42221294 42221475 42221535 42221550 42221561 42231092 42232528 42232552 42234345 42234502 42234508 42234563 42234599 42234604 42234702 42234721 42234735 42234940 42235058 42235093 42235104 42235349 42235662 42235672 42235838 42235847 42235904 42235986 42235987 42236033 42236054 42236101 42236103 42236213 42236226 42236250 42236318 42238244 42239703 42239763 42239859 42239956 42240082 42240155 42240280 42240281 42240325 42240564 42240612 42240696 42240745 42240747 42240825 42240940 42240949 42241033 42241219 42241250 42241261 42241280 42241360 42241361 42241440 42241470 42241511 42241514 42241642 42243403 42243414 42243455 42243498 42243598 42243607 42243644 42243705 42243742 42243799 42243801 42243848 42243855 42243878 42243920 42243934 42244007 42244009 42244035 42244038 42244176 42244306 42244318 42244453 42244533 42244534 42244782 42245154 42245188 42245255 42245313 42245405 42245486 42245573 42245630 42245653 42245792 42245847 42245942 42247660 42247663 42247682 42247699 42247762 42247809 42247812 42247860 42247930 42247971 42247993 42248005 42248045 42248051 42248086 42248175 42248192 42248283 42248357 42248410 42248434 42248479 42248498 42248736 42249264 42249316 42249343 42249407 42249408 42249633 42249673 42249786 42249787 42249821 42249868 42249921 42249964 42253808 42253818 42253862 42253961 42253992 42253993 42254063 42254138 42254141 42254207 42254406 42254443 42254522 42254874 42255077 42255137 42255195 42255237 42255281 42255324 42255346 42257410 42257483 42257535 42257760 42257877 42257921 42257958 42258116 42258909 42259041 42259188 42259203 42259227 42259259 42259369 42259389 42259392 42259637 42259755 42260810 42260838 42260892 42260968 42261135 42261284 42261661 42280349 42280383 42280464 42280602 42281380 42281501 42281571 42282103 42282221 42282422 42282467 42282553 42282815 42282875 42282958 42283016 42283020 42283054 42283062 42284454 42284460 42284540 42284580 42285655 42292878 42293545 42293571 42293637 42294227 42294301 42294352 42294449 42294847 42294898 42294994 42295213 42295387 42296362 42296379 42296402 42296478 42296546 42296603 42296618 42296668 42296836 42296949 42297046 42297107 42297800 42297803 42297826 42297871 42297888 42297918 42298452 42298532 42298642 42302292 42302371 42302393 42302426 42302442 42302453 42302496 42302527 42302528 42302534 42302563 42302584 42302597 42302744 42302880 42302881 42302924 42302990 42303002 42303045 42303114 42303548 42303559 42303805 42304081 42304120 42304127 42305827 42305851 42305861 42305917 42305970 42306024 42306027 42306114 42306125 42306158 42306159 42306171 42306227 42306235 42306318 42306368 42306402 42306428 42306438 42306494 42306552 42306573 42306710 42306741 42306772 42306791 42306878 42306922 42307152 42307187 42307199 42307251 42307269 42307369 42307388 42307405 42307471 42307637 42307673 42307693 42307700 42307795 42307806 42307836 42307872 42307877 42307891 42307894 42309890 42309895 42309950 42310001 42310067 42310108 42310155 42310174 42310224 42310269 42310304 42310321 42310350 42310380 42310436 42310443 42310572 42310575 42310635 42310667 42310783 42310822 42310890 42310971 42311104 42311385 42311387 42311503 42311567 42311693 42311755 42311777 42311809 42311855 42311896 42312101 42312151 42312298 42312387 42312410 42312451 42314251 42314282 42314456 42314612 42314661 42314749 42314774 42315017 42315123 42315172 42315387 42315440 42316632 42316862 42316895 42317062 42317447 42318738 42318740 42318814 42318904 42319025 42319092 42319498 42320019 42320054 42320066 42323073 42323449 42324018 42324087 42324132 42324149 42324176 42324188 42324239 42324320 42324346 42324394 42324485 42324543 42324790 42324897 42325059 42325283 42325343 42325353 42325410 42325494 42325521 42325539 42325559 42325635 42325872 42325903 42325924 42325927 42325973 42325976 42326009 42326054 42326069 42326139 42326141 42326198 42327175 42327190 42327197 42327297 42327298 42327341 42327383 42327389 42327421 42327446 42327459 42327472 42327478 42327500 42327507 42327523 42327536 42327550 42327586 42327587 42327637 42327707 42327717 42327751 42327829 42327830 42327848 42327851 42327863 42327867 42327877 42327946 42327966 42327973 42328014 42328192 42328193 42328282 42328285 42328317 42328325 42328350 42328364 42328373 42328763 42328778 42329016 42329053 42329059 42329148 42329163 42329202 42329282 42329283 42329294 42329335 42329395 42329482 42329499 42329513 42329563 42329566 42329589 42329613 42329690 42329717 42329720 42329721 42329754 42329791 42329808 42329851 42330243 42330265 42332690 42332694 42332708 42332715 42332717 42332745 42332772 42332776 42332856 42332857 42332903 42332914 42332923 42332926 42332962 42333000 42333026 42333071 42333183 42333231 42333267 42333408 42333507 42333559 42333600 42333654 42333741 42333755 42333805 42334048 42334088 42334131 42334133 42336211 42336906 42336912 42336939 42337000 42337001 42337003 42337134 42337137 42337243 42337327 42337564 42337574 42337585 42337625 42337691 42337693 42337694 42337749 42337751 42337850 42337851 42337866 42340981 42340988 42341055 42341110 42341164 42341197 42341310 42341393 42341395 42343770 42343824 42343933 42343998 42344093 42346902 42347026 42347138 42352066 42352439 42352455 42352712 42352783 42352904 42353037 42354773 42354865 42354903 42355099 42355411 42355711 42377496 42382667 42388726 42389811 42390133 42390440 42390479 42390538 42390554 42390754 42390782 42390815 42392244 42392359 42392420 42392481 42392597 42396227 42396266 42396506 42396554 42396662 42397896 42398134 42398537 42398658 42398684 42398909 42399231 42399274 42399856 42401684 42402252 42402400 42402532 42402985 42403230 42408116 42408194 42408286 42408376 42408408 42408430 42408454 42408718 42408872 42408966 42408993 42409016 42409067 42409159 42409320 42411528 42411734 42412284 42412358 42412442 42412499 42412618 42412701 42412873 42412902 42412920 42412978 42413095 42415113 42415306 42415405 42415457 42415499 42415555 42415647 42415862 42416167 42416250 42416356 42416684 42416892 42418342 42418368 42418781 42418866 42418877 42419061 42419117 42419207 42419437 42419687 42419731 42419760 42419882 42420222 42426672 42428367 42429993 42430288 42430385 42430602 42430680 42432052 42432199 42433395 42433939 42441273 42442048 42442095 42442223 42442284 42442507 42442630 42442664 42442706 42442758 42442935 42443029 42444203 42444232 42444359 42444438 42444506 42444548 42444564 42444670 42444765 42444859 42445036 42445199 42445250 42445399 42445662 42445734 42445796 42446167 42446436 42448423 42448444 42448564 42448686 42448826 42448941 42449033 42449083 42449196 42449588 42449924 42450055 42451944 42452192 42452278 42452866 42453076 42453194 42453393 42453662 42463118 42463216 42463236 42463376 42463402 42463553 42463907 42464386 42464589 42465466 42465513 42466951 42467615 42471184 42471511 42471999 42473946 42474667 42474903 42475725 42477071 42477755 42478111 42480064 42481216 42482795 42482926 42484089 42484419 42484685 42485528 42485631 42485683 42495126 42497879 42501697 42501785 42501790 42501869 42502014 42502601 42502863 42502942 42503037 42505132 42505171 42505200 42505208 42505262 42505353 42505451 42505577 42505804 42505930 42506961 42507172 42507273 42507442 42507690 42509106 42509130 42509218 42509315 42509352 42509451 42509572 42509851 42510438 42511509 42519504 42519958 42530273 42530336 42530389 42530463 42530558 42530781 42530919 42531015 42535335 42536051 42536137 42536179 42536247 42536340 42536451 42536483 42538404 42538636 42538774 42538797 42539085 42539189 42539275 42539337 42539399 42539767 42540065 42540098 42540241 42540302 42540361 42541816 42541836 42541846 42541901 42542242 42542728 42543365 42543950 42544325 42545252 42545261 42545771 42547251 42547525 42547601 42550697 42551018 42551138 42551892 42551896 42552713 42553077 42553351 42553563 42553606 42553656 42553807 42553838 42553862 42554024 42554037 42565506 42565573 42565680 42565811 42565839 42565868 42565929 42566023 42566093 42566104 42570420 42570800 42572578 42572976 42573092 42573170 42573222 42573322 42573759 42575056 42575068 42575129 42575213 42575247 42575310 42575526 42575992 42576014 42576267 42591870 42592079 42592136 42592209 42592232 42592568 42592777 42593251 42593348 42593449 42593566 42594155 42594451 42595458 42595690 42595751 42595992 42596152 42596261 42596868 42600735 42600823 42601421 42601472 42601736 42602088 42602541 42603239 42603296 42604224 42604355 42604358 42604467 42604479 42604527 42604569 42604636 42604714 42604716 42604783 42604837 42604971 42605206 42605250 42605389 42605824 42605983 42606740 42607150 42607179 42609227 42609323 42609422 42609471 42609484 42609633 42609731 42609771 42609841 42609862 42609906 42609990 42609994 42610059 42610074 42610154 42610156 42610291 42610479 42610806 42610909 42610910 42610915 42610924 42610927 42610944 42610971 42612369 42612429 42612436 42612875 42613191 42613306 42613414 42613526 42613543 42613663 42613744 42613745 42613758 42613770 42613792 42613813 42613862 42613896 42616238 42616240 42616278 42616282 42616304 42616311 42616354 42616366 42616412 42616443 42616445 42616456 42616458 42616498 42616542 42616577 42616607 42616615 42616684 42616749 42617288 42617391 42617393 42617414 42617433 42617461 42617464 42617511 42617525 42617636 42617668 42617677 42617698 42617708 42622507 42623094 42623099 42623101 42623104 42623106 42623109 42623111 42623115 42623117 42623119 42623124 42623135 42623857 42623862 42623869 42623895 42623899 42623905 42623918 42623968 42624006 42624030 42624032 42624050 42624052 42624054 42624059 42624074 42624077 42624085 42624128 42624131 42624244 42624258 42624297 42624328 42624340 42624407 42624433 42624442 42624443 42624482 42624506 42624509 42624543 42624545 42624560 42624575 42624576 42624642 42624668 42624725 42624750 42624909 42624972 42625016 42625104 42625126 42625127 42625170 42625207 42625283 42625426 42625471 42625475 42625604 42625637 42625760 42625784 42625793 42625823 42625825 42625864 42627270 42627274 42627314 42627380 42627399 42627407 42627429 42627440 42627449 42627465 42627466 42627504 42627506 42627555 42627600 42627620 42627674 42627716 42627830 42627839 42627899 42628066 42628107 42628136 42628149 42628245 42628908 42628977 42629014 42629168 42629192 42630609 42630636 42630718 42630775 42630845 42630888 42630955 42631078 42631210 42631248 42631342 42631374 42631509 42631638 42631872 42631917 42632030 42632399 42633784 42633868 42633948 42634302 42634477 42634637 42634742 42635815 42635880 42635907 42635940 42636321 42636808 42636820 42636875 42636906 42636963 42637158 42638172 42638195 42638203 42638213 42638250 42638340 42638360 42638518 42638600 42638653 42639013 42639068 42639127 42639164 42639685 42639732 42639773 42639784 42639884 42639932 42639958 42639984 42640110 42640159 42640231 42640260 42640376 42658577 42658590 42667757 42667853 42667867 42668291 42668555 42669250 42669278 42681297 42681430 42681441 42681621 42681838 42682264 42682341 42684185 42684188 42684235 42684272 42684335 42684341 42684420 42684505 42684564 42684635 42684661 42684698 42684728 42684760 42684761 42684799 42684837 42684838 42684954 42684955 42685207 42685210 42685465 42685644 42685646 42685723 42686187 42686346 42686457 42686658 42686718 42688531 42688560 42688661 42688708 42688795 42688864 42688895 42688984 42689009 42689051 42689101 42689136 42689213 42689260 42689493 42689733 42689993 42690017 42690061 42690069 42690150 42690210 42690233 42690275 42690351 42691469 42691485 42691531 42691685 42691783 42691804 42691891 42691943 42691964 42692003 42692033 42692072 42692194 42692461 42692842 42692880 42692945 42693023 42693044 42693166 42693355 42697837 42697870 42697917 42697996 42698046 42698076 42698126 42698228 42698296 42698331 42698407 42698480 42698506 42698890 42699371 42699502 42699694 42699750 42701195 42701201 42701231 42701295 42701334 42701416 42701498 42701595 42701649 42701795 42701859 42701997 42702220 42702798 42702922 42703014 42703047 42703098 42703155 42703219 42703248 42703271 42703304 42703317 42704592 42704600 42704710 42704787 42704806 42704842 42704907 42705200 42705234 42705438 42705507 42705653 42705784 42705852 42705892 42706007 42706077 42706113 42706171 42706258 42706453 42706472 42706511 42707676 42709989 42710122 42716459 42716542 42716604 42716645 42716659 42716701 42716752 42716812 42716889 42716923 42717036 42717085 42717194 42717321 42717575 42717836 42717930 42717984 42718204 42718350 42718680 42720499 42720527 42720566 42720699 42720764 42721157 42721235 42721277 42721567 42721639 42721827 42721959 42721982 42722077 42722272 42723775 42723886 42723897 42723974 42723992 42724007 42724383 42724470 42724533 42727571 42727701 42728243 42728686 42730958 42731011 42731148 42746679 42746700 42746707 42746840 42746901 42746934 42746982 42747051 42747120 42747171 42747222 42747254 42747272 42747372 42747429 42747543 42747564 42747597 42747640 42747662 42747741 42747870 42747942 42747979 42759368 42759440 42759511 42759596 42759606 42759647 42759732 42759819 42759849 42759872 42759952 42760242 42760262 42760327 42760428 42760478 42760545 42760584 42760632 42760728 42760777 42760878 42760974 42761032 42761123 42763277 42763503 42763524 42763725 42763957 42764526 42765184 42768315 42768423 42768438 42768474 42768521 42768555 42768603 42768704 42768724 42768780 42768860 42768904 42769242 42791097 42791107 42791148 42791280 42791516 42791548 42791600 42791644 42791681 42791792 42791849 42791952 42792304 42792358 42792662 42792706 42792763 42792797 42792850 42792909 42792972 42793003 42793529 42793654 42793715 42795799 42795864 42795886 42795941 42796005 42796101 42796108 42796140 42796207 42796258 42796275 42796747 42796776 42796867 42796918 42797018 42797064 42797187 42797220 42797367 42798086 42798129 42798143 42798355 42798377 42798397 42798410 42798618 42798717 42798876 42798938 42798960 42798997 42799051 42806615 42806646 42806681 42806708 42806739 42806761 42806778 42806923 42806993 42807034 42807246 42807341 42807400 42807440 42807551 42807640 42808229 42808266 42808295 42819532 42819586 42819806 42819967 42820169 42820292 42820393 42820427 42820443 42820467 42820500 42820561 42820669 42827229 42827444 42827459 42827575 42827699 42827746 42827786 42827863 42827889 42828119 42828407 42828527 42829531 42829629 42829729 42829775 42829783 42829914 42829938 42829946 42830006 42831096 42831097 42831128 42831132 42831195 42831318 42831368 42831378 42831399 42831624 42831928 42832042 42832189 42832225 42832240 42832326 42832411 42832599 42832811 42832838 42832851 42832898 42832941 42833000 42833071 42833078 42833181 42833297 42836468 42839616 42839713 42839843 42839866 42839997 42840242 42840283 42840330 42840365 42840517 42840595 42840696 42841311 42841399 42841410 42841445 42841497 42841524 42841735 42841752 42841805 42850823 42850896 42850898 42850910 42851051 42851086 42851119 42851501 42851552 42851566 42851576 42851662 42851786 42851916 42852116 42852175 42852204 42852520 42852573 42852670 42852678 42852697 42852756 42852791 42852908 42852927 42852971 42853003 42854739 42855077 42856036 42856189 42856532 42856610 42857426 42857510 42859427 42859474 42859533 42859553 42859623 42867105 42867112 42867539 42868979 42869106 42869251 42869417 42870457 42871346 42872200 42872263 42872297 42872311 42872326 42872411 42873240 42873290 42873301 42873325 42873334 42873340 42873536 42873878 42874047 42874168 42874289 42874299 42874410 42874437 42874442 42874449 42874471 42874593 42874833 42875961 42876048 42876236 42876319 42876332 42876342 42876351 42876369 42877247 42877362 42877459 42877505 42877530 42877692 42877778 42877816 42877945 42877994 42878031 42878129 42878273 42878550 42879001 42879169 42879613 42879750 42879933 42880032 42880068 42880100 42880138 42880250 42882731 42882743 42882805 42882881 42883014 42883051 42883156 42891201 42891262 42891323 42891421 42891456 42891481 42891704 42891999 42892080 42892188 42892306 42892386 42892422 42892459 42892559 42892870 42892958 42893104 42893227 42893281 42893317 42893403 42893442 42893499 42893527 42893546 42893551 42893596 42893655 42895016 42895023 42895034 42895037 42895063 42895100 42895151 42895221 42895272 42895273 42895289 42895291 42895321 42895322 42895345 42895361 42895421 42895444 42895504 42895556 42895577 42895619 42895655 42895656 42895663 42895668 42895852 42896038 42896099 42896121 42896184 42896187 42896222 42896278 42896298 42896299 42896322 42896323 42896402 42896404 42896412 42896488 42896641 42896648 42896692 42896782 42896856 42896862 42896894 42897209 42897380 42898398 42898417 42898426 42898433 42898451 42898458 42898466 42898478 42898515 42898554 42898561 42898575 42898650 42898802 42899009 42899628 42900786 42901109 42901546 42901893 42901906 42901924 42901935 42901944 42901951 42901967 42902693 42902695 42902704 42902768 42902824 42902833 42902873 42902874 42902886 42902895 42902913 42902918 42902938 42902942 42902944 42903037 42903064 42903112 42903179 42903196 42903261 42903268 42903312 42903331 42903361 42903373 42903493 42903519 42903555 42903564 42903599 42903624 42903835 42903855 42903858 42903941 42903962 42903996 42904066 42904110 42904135 42904141 42904165 42904213 42904217 42904337 42904350 42904366 42904394 42904403 42904418 42904514 42904713 42904735 42904774 42906075 42906078 42906082 42906129 42906136 42906255 42906258 42906346 42906404 42906433 42906849 42906878 42906923 42906928 42906995 42907074 42907090 42907091 42907124 42907147 42907150 42907751 42908218 42908375 42908418 42908446 42908622 42908647 42908714 42912101 42912109 42912113 42912132 42912161 42912230 42912271 42912274 42912304 42912386 42912407 42912408 42912446 42912498 42912499 42912575 42912577 42912694 42912729 42912757 42912841 42912859 42912877 42912958 42912982 42912999 42913044 42913135 42913148 42913361 42913374 42913522 42913531 42913562 42913566 42913685 42913764 42914116 42914167 42917744 42917827 42917888 42917964 42917976 42918014 42918048 42918184 42918196 42918306 42918315 42918317 42918331 42918343 42918390 42918393 42918413 42918418 42918520 42918532 42918582 42918631 42918632 42918660 42918670 42918866 42918867 42918900 42918901 42918907 42918909 42918924 42918925 42918979 42918980 42920498 42920500 42920525 42920544 42920545 42920590 42920672 42920697 42920719 42920723 42920740 42920745 42920795 42920797 42920857 42920871 42920880 42920915 42920956 42920975 42921014 42921018 42921150 42921171 42921890 42921914 42921922 42921935 42921944 42921954 42921978 42922128 42922213 42922286 42922299 42922434 42922482 42922488 42922525 42922552 42922562 42922574 42922607 42922688 42922696 42922737 42922801 42924859 42924866 42924867 42924895 42924906 42924937 42924939 42924974 42924980 42925002 42925004 42925033 42925102 42925106 42925195 42925210 42925219 42933657 42933796 42933870 42933884 42933967 42934063 42934131 42934214 42934287 42934529 42934561 42934580 42945740 42945831 42945889 42945946 42946126 42947297 42947324 42947447 42947499 42947529 42947554 42947601 42948072 42948143 42948252 42950268 42950303 42950520 42950554 42953155 42953169 42953181 42953271 42953310 42953367 42953396 42953428 42953504 42953600 42953633 42953686 42953703 42953799 42953884 42953952 42954238 42954292 42954317 42954401 42954486 42954585 42954666 42954699 42954749 42954889 42954922 42961944 42961987 42962098 42962123 42962153 42962210 42962322 42962461 42962584 42962800 42962956 42963054 42963174 42963228 42963309 42963372 42963485 42963557 42963616 42963659 42963679 42963773 42963877 42965005 42965071 42965141 42965169 42965222 42965289 42965298 42965373 42965407 42965534 42965667 42965705 42965733 42965762 42965997 42966027 42966051 42966078 42966093 42966162 42966245 42966309 42966398 42966528 42966560 42969384 42969430 42985040 42985209 42985259 42985299 42985382 42985497 42985539 42985618 42985685 42985714 42985732 42988215 42988243 42988269 42988336 42988405 42988434 42988484 42988491 42988611 42988751 42988843 42988911 42989100 42989176 42989253 42989329 42989440 42989570 42989955 42990008 42990026 42990126 42990171 42990251 42990378 42990474 42990497 42990557 42990592 42992062 42992079 42992093 42992164 42992200 42992278 42992309 42992342 42992433 42992475 42992512 42992578 42992662 42992718 42992791 42992886 42993530 42993537 42993579 42993678 42993761 42993943 42994502 42994505 42994547 42994558 42994656 42994666 42994709 42994747 42994800 42994803 42994945 42994966 42995076 42995092 42997424 42997447 42997469 42997470 42997478 42997634 42997687 42997734 42997744 42997750 42997752 42997784 42997836 42997845 42997898 42997943 42997950 42997952 42998022 42998046 42998052 42998096 42998121 42998132 42998197 42998228 42998235 42998249 42998260 42998269 42998288 42998297 42998377 42998379 42998443 42998574 42998576 42998583 42998609 42998623 42998650 42998658 42998674 42998741 42998789 42998821 42998850 42998938 43002772 43002839 43002882 43003150 43003467 43003677 43003896 43003932 43003966 43004023 43004200 43004204 43004281 43004306 43004370 43004375 43004410 43004457 43004542 43004553 43004566 43007091 43007099 43007171 43007649 43007818 43007883 43007983 43017327 43017574 43040780 43040829 43041258 43041314 43041421 43041444 43041507 43041698 43043823 43043850 43043921 43043939 43043959 43044003 43044050 43044204 43049280 43050593 43050631 43050674 43050702 43050829 43050925 43050940 43050951 43051025 43053021 43053057 43053138 43053202 43053385 43053498 43053648 43053758 43053870 43053935 43054064 43054245 43054680 43055102 43055162 43055186 43055790 43056067 43056212 43056297 43056420 43057997 43058066 43058164 43058197 43058472 43058502 43058561 43058605 43058689 43059222 43059266 43059438 43059619 43059713 43060830 43060958 43060961 43061049 43061170 43061173 43070390 43070621 43070641 43070670 43070697 43070721 43070895 43071112 43076304 43076433 43076532 43076577 43076896 43077019 43077125 43077230 43077297 43077365 43077424 43077499 43077622 43077771 43077926 43077955 43078026 43078082 43078142 43078295 43078434 43078506 43078544 43078718 43078748 43078842 43079213 43079244 43079372 43082384 43082525 43082595 43082802 43082984 43082989 43083310 43083314 43084078 43098921 43099311 43099480 43099508 43099534 43099559 43099627 43099731 43100309 43100873 43100913 43100983 43101063 43101124 43101147 43101213 43101407 43101451 43101471 43101498 43101750 43101839 43104247 43104275 43104320 43104397 43104424 43104506 43104541 43104595 43104702 43104764 43104836 43104956 43105033 43105077 43105182 43105234 43105277 43105312 43109674 43109704 43109760 43109888 43109932 43110325 43110376 43110432 43110512 43110555 43110701 43132427 43132452 43132605 43132627 43132773 43132856 43132972 43133059 43133116 43133197 43133237 43133278 43133338 43133534 43133569 43133656 43133731 43133945 43134004 43134117 43134172 43134222 43137496 43137726 43137808 43138060 43138116 43154088 43156025 43156065 43156134 43156247 43156313 43156360 43156387 43156450 43156540 43156637 43156714 43158496 43158648 43158727 43158747 43158916 43158970 43159137 43159202 43159287 43159335 43159395 43159421 43159492 43159571 43159727 43159930 43159935 43159947 43159989 43159994 43160091 43160152 43160226 43160627 43161404 43161448 43161526 43161557 43161584 43161657 43161826 43161876 43163841 43163895 43163935 43163944 43163981 43164002 43164048 43164051 43164065 43164078 43164080 43164144 43164175 43164261 43164275 43164364 43164486 43164495 43164540 43164548 43164572 43164656 43164979 43165120 43165150 43165182 43165613 43165893 43165984 43165988 43166031 43166044 43166089 43166099 43166124 43166167 43166189 43166283 43166292 43166351 43166448 43166458 43166500 43166515 43166521 43166537 43166573 43166639 43166664 43166688 43166690 43166719 43166905 43167013 43167017 43167441 43167502 43167558 43167559 43167571 43167573 43167582 43168620 43168621 43168683 43168684 43168698 43168738 43168752 43168821 43168828 43168875 43168891 43168965 43169046 43169070 43169186 43169258 43169295 43169545 43169739 43169820 43169858 43169882 43169909 43169932 43169954 43169995 43170012 43170052 43170106 43170149 43170261 43170281 43170321 43170341 43170529 43170547 43170595 43170604 43170635 43170645 43170717 43170758 43170790 43170891 43170915 43170974 43170981 43171012 43171102 43171129 43171160 43171172 43171235 43171252 43171309 43171316 43172873 43172922 43172931 43172962 43172972 43173024 43173037 43173071 43173072 43173096 43173178 43173202 43173213 43173270 43173271 43173338 43173343 43173380 43173407 43173459 43173474 43173517 43173536 43173586 43173606 43173634 43173679 43173702 43173720 43173742 43173809 43173837 43173905 43173908 43174166 43174197 43174211 43174334 43174415 43174432 43174446 43174466 43174505 43174507 43174552 43174554 43174660 43174665 43174671 43174678 43174680 43174691 43174733 43174742 43174746 43174761 43177053 43177057 43177082 43177095 43177099 43177184 43177271 43177396 43177398 43177448 43177457 43177507 43177543 43177663 43177667 43177705 43177747 43177816 43177828 43177874 43177882 43177890 43177926 43177930 43177966 43177986 43178004 43178011 43178022 43178077 43178086 43178133 43178137 43178210 43178213 43178400 43178402 43178426 43178433 43178499 43178500 43178521 43178542 43178576 43178633 43178663 43178723 43179118 43179180 43179327 43179335 43179382 43179384 43179413 43179422 43179456 43179462 43179467 43179524 43179556 43179561 43181008 43181012 43181046 43181068 43181094 43181150 43181170 43181195 43181218 43181283 43181284 43181360 43181366 43181375 43181444 43181468 43181520 43181528 43181586 43181604 43181619 43181697 43181703 43181774 43181841 43181866 43181875 43181935 43181961 43181999 43182022 43182072 43182089 43182146 43182253 43182268 43182273 43182298 43182315 43182389 43182406 43182467 43182494 43182533 43182660 43182996 43183090 43183142 43185165 43185224 43185288 43185426 43185534 43185612 43185724 43185777 43185867 43186084 43186395 43186453 43186638 43187225 43187296 43187352 43187437 43187510 43187640 43187794 43187887 43190839 43190907 43191996 43192091 43192151 43194500 43194509 43194521 43194529 43194575 43194694 43194700 43194902 43194913 43195002 43195210 43206709 43206791 43206820 43206884 43206983 43207072 43207387 43209415 43209473 43209500 43209563 43209604 43209672 43209817 43209992 43210022 43210136 43210449 43210515 43210659 43218559 43218591 43218792 43218810 43218860 43218964 43219193 43219249 43219297 43219467 43219503 43269313 39233299 40879085 40976733 41378358 41387217 41387300 41441118 41446486 41446522 41446554 41446632 41446868 41446886 41447053 41465090 41465111 41465122 41465221 41465288 41465304 41465321 41465343 41465349 41465412 41465428 41465483 41465538 41500970 41566386 41748432 41760748 41761946 41764462 41764516 41764540 41764894 41765319 41765431 41765509 41767649 41767651 41767857 41978046 42178119 42235229 42315519 42315579 42315699 42315713 42315826 42316371 42316411 42316541 42316602 42448243 42550470 42551049 42759578 42763314 42763413 42763514 42763595 42763647 42763745 42763814 42763832 42763847 42763906 42763983 42764037 42764243 42764303 42764349 42764400 42764447 42764551 42764876 42764987 42765023 42765078 42768283 42850806 42855486 42856149 42856150 42856197 42856319 42856391 42856394 42856429 42856645 42857161 42857427 42857482 42857508 42857509 42857575 42857617 42857660 42857755 42857809 42857858 42859115 42859116 42859127 42859198 43157104 43157276 43157354 43157360 43158383 43158407 43159721 
Recalling Firm/
Manufacturer
Physio-Control Inc
11811 Willows Rd Ne
Redmond WA 98052-2003
Manufacturer Reason
for Recall
Certain LIFEPAK 15 Monitors/ Defibrillators were reported to experience a lockup condition after a shock was delivered. This condition is defined as a blank monitor display with LED lights on, indicating power on the device, but no response in keypad and device functions.
FDA Determined
Cause 2
Software design
Action Stryker sent an Urgent Medical Device Safety Notice & Correction Action Required letter dated February 1, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter requested customers. ¿ Review the attached impacted device list. Go to www.strykeremergencycare.com/fa281response to provide Stryker verification of the status of the devices listed. ¿ Upon confirmation of your device status, a member of our field service personnel will contact you to arrange for the correction of your device. The devices subject to this field action are planned to be serviced by December 31, 2019. ¿ If you have questions regarding this matter contact Customer Support team by calling 1 800 442 1142 and selecting option 7.
Quantity in Commerce 13,003 units
Distribution Worldwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database 510(K)s with Product Code = MKJ and Original Applicant = PHYSIO-CONTROL, INC.
-
-