• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Prodiva CS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Prodiva CSsee related information
Date Initiated by FirmDecember 21, 2018
Date PostedFebruary 22, 2019
Recall Status1 Terminated 3 on November 19, 2020
Recall NumberZ-0869-2019
Recall Event ID 82038
Product Classification System, nuclear magnetic resonance imaging - Product Code LNH
ProductProdiva CS, Model Number 781069. Nuclear Magnetic Resonance Imaging System
Code Information All systems
FEI Number 3002807880
Recalling Firm/
Manufacturer
Philips Medical Systems Nederlands
Veenpluis 4-6
Best Netherlands
Manufacturer Reason
for Recall
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
FDA Determined
Cause 2
Unknown/Undetermined by firm
ActionUrgent Medical Device Correction notification letters dated December 2018 were sent to customers.
Quantity in Commerce11226 total
DistributionU.S. Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.