• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Coorad Morrey Total Elbow

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Coorad Morrey Total Elbowsee related information
Date Initiated by FirmOctober 10, 2019
Create DateNovember 07, 2019
Recall Status1 Terminated 3 on May 19, 2020
Recall NumberZ-0301-2020
Recall Event ID 83962
510(K)NumberK001989 
Product Classification Prosthesis, total anatomic shoulder, uncemented metaphyseal humeral stem with no diaphyseal incursion, semi-co - Product Code JDC
ProductCoorad Morrey Total Elbow, Item Nos. 32810502501 32810502701 32810502900 32810503000 32810600012 32810600013 32855510796 32855510797 32855510799 32855510800 32855510801 32855510805 32855510822 32855510843 32855510844
Code Information All product manufactured prior to January 2014.
FEI Number 1000220733
Recalling Firm/
Manufacturer
Zimmer Biomet, Inc.
1800 W Center St
Warsaw IN 46580-2304
For Additional Information ContactZimmer Biomet Customer Service
574-371-3071
Manufacturer Reason
for Recall
Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
FDA Determined
Cause 2
Packaging change control
ActionThe firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, as per standard orthopedic procedure planning, to reduce the likelihood of an extension of surgery time. 3. Examine devices thoroughly for adherent LDPE bag residue prior to use. Use a backup device if LDPE bag residue is detected. 4. Return any devices which display adherent LDPE bag residue to Zimmer Biomet, consistent with the standard instructions for use regarding the disposition of damaged components. 5. Complete the Certificate of Acknowledgement and return to the firm. 6. Retain a copy of the acknowledgement form with your field action records in the event of a compliance audit of your facility's documentation. If you have further questions or concerns regarding this recall, please call the firm's customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
Quantity in Commerce5,459,583 total devices
DistributionDistributed nationwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = JDC
-
-