Class 2 Device Recall |
|
Date Initiated by Firm |
January 15, 2020 |
Date Posted |
January 24, 2020 |
Recall Status1 |
Terminated 3 on March 17, 2023 |
Recall Number |
Z-1181-2020 |
Recall Event ID |
84726 |
Product Classification |
Anesthesia Conduction Kit - Product Code CAZ
|
Product |
Epidural Catheterization Kit with FlexTip Plus, Open Tip, Single- Port Catheter, Product Code ASK-05502-NY
|
Code Information |
Lots: 23F19C0385 23F19G0310 23F19H0234 23F19K0248 23F19F0167 23F19G0434 23F19J0063 |
Recalling Firm/ Manufacturer |
Arrow International Inc 2400 Bernville Rd Reading PA 19605-9607
|
For Additional Information Contact |
Customer Service 866-396-2111
|
Manufacturer Reason for Recall |
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
|
FDA Determined Cause 2 |
Other |
Action |
An Urgent Medical Device Recall notification letter dated 1/15/20 was sent to customers.
Our records indicate you have received products that are subject to this field action. We are now notifying our customers to take the following actions:
1. If you have affected stock, immediately identify product, discontinue use and quarantine any products with the product codes and lot numbers listed above.
2. To acknowledge receipt of this Urgent Medical Device Recall Notice or to return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical.
3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter. |
Quantity in Commerce |
343,105 total |
Distribution |
The products were distributed to the following US states: AK, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. |