| Date Initiated by Firm | April 30, 2020 |
| Create Date | June 03, 2020 |
| Recall Status1 |
Terminated 3 on January 23, 2024 |
| Recall Number | Z-2246-2020 |
| Recall Event ID |
85610 |
| Product Classification |
Bed, ac-powered adjustable hospital - Product Code FNL
|
| Product | Centrella Smart+ Bed, Catalog Number P7900 |
| Code Information |
Serial codes anufactured between U290PF8145 through V015PF2473 plus the following serial codes outside this range that were updated during servicing: S237PF0789 T222PF6473 U067PF5719 U079PF6209 S248PF1062 T223PF6496 U067PF5722 U079PF6210 S249PF1117 T223PF6497 U067PF5725 U079PF6211 S254PF1218 T227PF6698 U071PF5783 U079PF6212 S255PF1262 T236PF7213 U071PF5788 U079PF6213 S256PF1300 T241PF7583 U071PF5789 U079PF6214 S257PF1353 T249PF8023 U071PF5790 U079PF6215 S300PF2244 T250PF8146 U072PF5798 U079PF6216 S303PF2287 T268PF9463 U078PF6163 U079PF6217 S303PF2306 T268PF9468 U078PF6166 U079PF6247 S305PF2366 T277PF9896 U078PF6169 U079PF6248 S331PF2510 T277PF9926 U078PF6171 U079PF6249 S332PF2554 T281PF0056 U078PF6173 U079PF6250 S333PF2570 T282PF0110 U078PF6174 U079PF6251 S339PF2673 T282PF0113 U078PF6175 U079PF6252 S340PF2717 T285PF0294 U078PF6176 U079PF6253 S346PF2865 T288PF0375 U078PF6177 U079PF6254 S347PF2912 T295PF0413 U078PF6178 U079PF6255 S347PF2919 T297PF0600 U078PF6179 U079PF6256 S354PF3029 T303PF0798 U078PF6180 U079PF6257 T002PF3078 T324PF1681 U078PF6181 U079PF6258 T003PF3090 T333PF2061 U078PF6182 U079PF6259 T003PF3099 T340PF2401 U078PF6183 U079PF6260 T003PF3101 T347PF2679 U078PF6184 U079PF6263 T003PF3102 T348PF2725 U078PF6185 U079PF6264 T003PF3111 T353PF2876 U078PF6186 U079PF6266 T003PF3113 T353PF2896 U078PF6187 U079PF6267 T004PF3119 U023PF3849 U078PF6188 U079PF6268 T004PF3153 U025PF4032 U078PF6189 U079PF6269 T009PF3187 U029PF4099 U078PF6190 U079PF6270 T078PF4163 U032PF4222 U078PF6191 U079PF6271 T079PF4173 U036PF4335 U078PF6192 U079PF6272 T079PF4177 U036PF4366 U078PF6193 U079PF6273 T079PF4182 U038PF4450 U078PF6194 U079PF6274 T079PF4185 U042PF4523 U078PF6195 U079PF6275 T079PF4187 U043PF4568 U078PF6196 U079PF6276 T099PF4537 U043PF4606 U079PF6197 U079PF6277 T102PF4616 U043PF4618 U079PF6198 U080PF6278 T103PF4638 U045PF4631 U079PF6199 U080PF6279 T106PF4657 U045PF4648 U079PF6200 U080PF6282 T135PF5209 U045PF4668 U079PF6201 U080PF6283 T138PF5276 U046PF4726 U079PF6203 U080PF6284 T192PF5530 U059PF5316 U079PF6204 U080PF6285 T199PF5706 U067PF5705 U079PF6205 U080PF6286 T214PF6131 U067PF5708 U079PF6206 U107PF7455 T221PF6408 U067PF5713 U079PF6207 U115PF7878 T222PF6429 U067PF5716 U079PF6208 U133PF8754 U291PF8261 U294PF8308 U295PF8365 U296PF8415 U291PF8262 U294PF8309 U295PF8366 U296PF8416 U291PF8263 U294PF8310 U295PF8367 U296PF8417 U291PF8264 U294PF8311 U295PF8368 U296PF8418 U291PF8265 U294PF8312 U295PF8369 U296PF8419 U291PF8266 U294PF8313 U295PF8370 U296PF8420 U291PF8267 U294PF8314 U295PF8371 U296PF8421 U291PF8268 U294PF8315 U295PF8372 U296PF8423 U291PF8269 U294PF8316 U295PF8373 U296PF8424 U291PF8270 U294PF8317 U295PF8374 U296PF8425 U291PF8271 U294PF8318 U295PF8375 U296PF8426 U291PF8272 U294PF8319 U295PF8376 U296PF8428 U291PF8273 U294PF8320 U295PF8377 U296PF8429 U291PF8274 U294PF8321 U295PF8378 U296PF8431 U294PF8275 U294PF8322 U295PF8379 U296PF8433 U294PF8276 U294PF8323 U295PF8380 U296PF8434 U294PF8277 U294PF8324 U295PF8381 U296PF8436 U294PF8278 U294PF8325 U295PF8382 U296PF8437 U294PF8279 U294PF8326 U295PF8383 U296PF8439 U294PF8280 U294PF8327 U295PF8384 U296PF8440 U294PF8281 U294PF8328 U295PF8385 U296PF8442 U294PF8282 U294PF8329 U295PF8386 U296PF8443 U294PF8283 U294PF8330 U295PF8387 U296PF8444 U294PF8284 U294PF8331 U295PF8388 U296PF8445 U294PF8285 U294PF8333 U295PF8389 U296PF8446 U294PF8286 U294PF8334 U295PF8390 U296PF8447 U294PF8287 U294PF8336 U295PF8391 U296PF8448 U294PF8288 U294PF8337 U295PF8392 U296PF8449 U294PF8289 U294PF8338 U295PF8393 U296PF8450 U294PF8290 U294PF8339 U295PF8394 U296PF8451 U294PF8291 U294PF8340 U295PF8395 U296PF8453 U294PF8292 U294PF8341 U295PF8396 U296PF8454 U294PF8293 U294PF8342 U295PF8397 U296PF8456 U294PF8294 U294PF8343 U295PF8398 U296PF8457 U294PF8295 U294PF8344 U295PF8400 U296PF8459 U294PF8296 U294PF8346 U295PF8401 U296PF8460 U294PF8297 U294PF8347 U295PF8402 U296PF8462 U294PF8298 U294PF8349 U295PF8403 U296PF8463 U294PF8299 U294PF8350 U295PF8404 U296PF8464 U294PF8300 U294PF8352 U295PF8405 U296PF8465 U294PF8301 U294PF8354 U295PF8406 U296PF8466 U294PF8302 U294PF8355 U296PF8407 U296PF8467 U294PF8303 U295PF8356 U296PF8409 U296PF8468 U294PF8304 U295PF8358 U296PF8410 U296PF8469 U294PF8305 U295PF8360 U296PF8412 U296PF8470 U294PF8306 U295PF8361 U296PF8413 U296PF8471 U294PF8307 U295PF8363 U296PF8414 U296PF8472 U298PF8614 U301PF8661 U301PF8708 U302PF8756 U298PF8615 U301PF8662 U301PF8709 U302PF8757 U298PF8616 U301PF8663 U301PF8710 U302PF8758 U298PF8617 U301PF8664 U301PF8711 U302PF8759 U298PF8618 U301PF8665 U301PF8712 U302PF8760 U298PF8619 U301PF8666 U301PF8713 U302PF8761 U298PF8620 U301PF8667 U301PF8714 U302PF8762 U298PF8621 U301PF8668 U301PF8715 U302PF8763 U298PF8622 U301PF8669 U301PF8716 U302PF8764 U298PF8623 U301PF8670 U301PF8717 U302PF8765 U298PF8624 U301PF8671 U301PF8718 U302PF8766 U298PF8625 U301PF8672 U301PF8719 U302PF8767 U298PF8626 U301PF8673 U301PF8720 U302PF8768 U298PF8627 U301PF8674 U301PF8721 U302PF8769 U298PF8628 U301PF8675 U301PF8722 U302PF8770 U298PF8629 U301PF8676 U302PF8723 U302PF8771 U298PF8630 U301PF8677 U302PF8724 U302PF8772 U298PF8631 U301PF8678 U302PF8725 U302PF8773 U298PF8632 U301PF8679 U302PF8726 U302PF8774 U298PF8633 U301PF8680 U302PF8727 U302PF8775 U298PF8634 U301PF8681 U302PF8728 U302PF8776 U298PF8635 U301PF8682 U302PF8729 U302PF8777 U298PF8636 U301PF8683 U302PF8730 U302PF8778 U298PF8637 U301PF8684 U302PF8731 U302PF8779 U298PF8638 U301PF8685 U302PF8732 U302PF8780 U298PF8639 U301PF8686 U302PF8733 U302PF8781 U298PF8640 U301PF8687 U302PF8734 U302PF8782 U298PF8641 U301PF8688 U302PF8735 U302PF8783 U301PF8642 U301PF8689 U302PF8736 U302PF8784 U301PF8643 U301PF8690 U302PF8737 U302PF8785 U301PF8644 U301PF8691 U302PF8738 U302PF8786 U301PF8645 U301PF8692 U302PF8739 U302PF8787 U301PF8646 U301PF8693 U302PF8740 U302PF8788 U301PF8647 U301PF8694 U302PF8741 U302PF8789 U301PF8648 U301PF8695 U302PF8743 U302PF8790 U301PF8649 U301PF8696 U302PF8744 U302PF8791 U301PF8650 U301PF8697 U302PF8745 U302PF8792 U301PF8651 U301PF8698 U302PF8746 U302PF8793 U301PF8652 U301PF8699 U302PF8747 U302PF8794 U301PF8653 U301PF8700 U302PF8748 U302PF8795 U301PF8654 U301PF8701 U302PF8749 U302PF8796 U301PF8655 U301PF8702 U302PF8750 U302PF8797 U301PF8656 U301PF8703 U302PF8751 U302PF8798 U301PF8657 U301PF8704 U302PF8752 U302PF8799 U301PF8658 U301PF8705 U302PF8753 U302PF8800 U301PF8659 U301PF8706 U302PF8754 U302PF8801 U301PF8660 U301PF8707 U302PF8755 U302PF8802 U304PF8951 U305PF8998 U308PF9046 U308PF9093 U304PF8952 U305PF8999 U308PF9047 U308PF9094 U304PF8953 U305PF9000 U308PF9048 U308PF9095 U304PF8954 U305PF9001 U308PF9049 U308PF9096 U304PF8955 U305PF9002 U308PF9050 U308PF9097 U304PF8956 U305PF9003 U308PF9051 U308PF9098 U304PF8957 U305PF9004 U308PF9052 U308PF9099 U304PF8958 U305PF9005 U308PF9053 U308PF9100 U304PF8959 U305PF9006 U308PF9054 U308PF9101 U304PF8960 U305PF9007 U308PF9055 U308PF9102 U304PF8961 U305PF9008 U308PF9056 U308PF9103 U304PF8962 U305PF9009 U308PF9057 U308PF9105 U304PF8963 U305PF9010 U308PF9058 U308PF9106 U305PF8964 U305PF9011 U308PF9059 U308PF9108 U305PF8965 U305PF9012 U308PF9060 U308PF9109 U305PF8966 U305PF9013 U308PF9061 U308PF9110 U305PF8967 U305PF9014 U308PF9062 U308PF9111 U305PF8968 U305PF9015 U308PF9063 U308PF9112 U305PF8969 U305PF9016 U308PF9064 U308PF9113 U305PF8970 U305PF9017 U308PF9065 U308PF9114 U305PF8971 U305PF9018 U308PF9066 U308PF9115 U305PF8972 U305PF9019 U308PF9067 U308PF9116 U305PF8973 U305PF9020 U308PF9068 U308PF9117 U305PF8974 U305PF9021 U308PF9069 U308PF9118 U305PF8975 U305PF9022 U308PF9070 U308PF9119 U305PF8976 U305PF9023 U308PF9071 U308PF9120 U305PF8977 U305PF9024 U308PF9072 U308PF9121 U305PF8978 U305PF9025 U308PF9073 U308PF9122 U305PF8979 U305PF9026 U308PF9074 U308PF9123 U305PF8980 U305PF9027 U308PF9075 U308PF9124 U305PF8981 U305PF9028 U308PF9076 U308PF9125 U305PF8982 U305PF9029 U308PF9077 U308PF9126 U305PF8983 U305PF9030 U308PF9078 U309PF9127 U305PF8984 U305PF9031 U308PF9079 U309PF9128 U305PF8985 U305PF9032 U308PF9080 U309PF9129 U305PF8986 U305PF9033 U308PF9081 U309PF9130 U305PF8987 U305PF9034 U308PF9082 U309PF9131 U305PF8988 U305PF9035 U308PF9083 U309PF9132 U305PF8989 U305PF9036 U308PF9084 U309PF9133 U305PF8990 U305PF9038 U308PF9085 U309PF9134 U305PF8991 U305PF9039 U308PF9086 U309PF9135 U305PF8992 U305PF9040 U308PF9087 U309PF9136 U305PF8993 U305PF9041 U308PF9088 U309PF9137 U305PF8994 U305PF9042 U308PF9089 U309PF9138 U305PF8995 U305PF9043 U308PF9090 U309PF9139 U305PF8996 U305PF9044 U308PF9091 U309PF9140 U305PF8997 U305PF9045 U308PF9092 U309PF9141 U310PF9287 U311PF9338 U312PF9386 U312PF9433 U310PF9288 U311PF9339 U312PF9387 U312PF9434 U311PF9289 U311PF9340 U312PF9388 U312PF9435 U311PF9290 U311PF9341 U312PF9389 U312PF9436 U311PF9291 U311PF9342 U312PF9390 U312PF9437 U311PF9292 U311PF9343 U312PF9391 U312PF9438 U311PF9293 U311PF9344 U312PF9392 U312PF9439 U311PF9294 U311PF9345 U312PF9393 U312PF9440 U311PF9296 U311PF9346 U312PF9394 U315PF9442 U311PF9297 U311PF9347 U312PF9395 U315PF9443 U311PF9299 U311PF9348 U312PF9396 U315PF9444 U311PF9300 U311PF9349 U312PF9397 U315PF9445 U311PF9302 U311PF9350 U312PF9398 U315PF9446 U311PF9303 U311PF9351 U312PF9399 U315PF9447 U311PF9304 U311PF9352 U312PF9400 U315PF9448 U311PF9305 U311PF9353 U312PF9401 U315PF9449 U311PF9306 U311PF9354 U312PF9402 U315PF9450 U311PF9307 U311PF9355 U312PF9403 U315PF9451 U311PF9308 U311PF9356 U312PF9404 U315PF9452 U311PF9309 U311PF9357 U312PF9405 U315PF9453 U311PF9311 U311PF9358 U312PF9406 U315PF9454 U311PF9312 U311PF9359 U312PF9407 U315PF9455 U311PF9313 U311PF9360 U312PF9408 U315PF9456 U311PF9314 U311PF9361 U312PF9409 U315PF9457 U311PF9315 U311PF9362 U312PF9410 U315PF9458 U311PF9316 U311PF9363 U312PF9411 U315PF9459 U311PF9317 U311PF9364 U312PF9412 U315PF9460 U311PF9318 U311PF9365 U312PF9413 U315PF9461 U311PF9319 U311PF9366 U312PF9414 U315PF9462 U311PF9320 U311PF9367 U312PF9415 U315PF9463 U311PF9321 U311PF9368 U312PF9416 U315PF9464 U311PF9322 U311PF9369 U312PF9417 U315PF9465 U311PF9323 U312PF9371 U312PF9418 U315PF9466 U311PF9324 U312PF9372 U312PF9419 U315PF9467 U311PF9325 U312PF9373 U312PF9420 U315PF9468 U311PF9326 U312PF9374 U312PF9421 U315PF9469 U311PF9327 U312PF9375 U312PF9422 U315PF9470 U311PF9328 U312PF9376 U312PF9423 U315PF9471 U311PF9329 U312PF9377 U312PF9424 U315PF9472 U311PF9330 U312PF9378 U312PF9425 U315PF9473 U311PF9331 U312PF9379 U312PF9426 U315PF9474 U311PF9332 U312PF9380 U312PF9427 U315PF9475 U311PF9333 U312PF9381 U312PF9428 U315PF9476 U311PF9334 U312PF9382 U312PF9429 U315PF9477 U311PF9335 U312PF9383 U312PF9430 U315PF9478 U311PF9336 U312PF9384 U312PF9431 U315PF9479 U311PF9337 U312PF9385 U312PF9432 U315PF9480 U317PF9631 U317PF9678 U318PF9726 U319PF9773 U317PF9632 U317PF9679 U318PF9727 U319PF9774 U317PF9633 U317PF9680 U318PF9728 U319PF9775 U317PF9634 U318PF9681 U318PF9729 U319PF9776 U317PF9635 U318PF9682 U318PF9730 U319PF9777 U317PF9636 U318PF9683 U318PF9731 U319PF9778 U317PF9637 U318PF9684 U318PF9732 U319PF9779 U317PF9638 U318PF9685 U318PF9733 U319PF9780 U317PF9639 U318PF9686 U318PF9734 U319PF9781 U317PF9640 U318PF9687 U318PF9735 U319PF9782 U317PF9641 U318PF9688 U318PF9736 U319PF9783 U317PF9642 U318PF9689 U318PF9737 U319PF9784 U317PF9643 U318PF9690 U318PF9738 U319PF9785 U317PF9644 U318PF9691 U318PF9739 U319PF9786 U317PF9645 U318PF9692 U318PF9740 U319PF9787 U317PF9646 U318PF9693 U318PF9741 U319PF9788 U317PF9647 U318PF9694 U318PF9742 U319PF9789 U317PF9648 U318PF9695 U318PF9743 U319PF9790 U317PF9649 U318PF9696 U318PF9744 U319PF9791 U317PF9650 U318PF9697 U318PF9745 U319PF9792 U317PF9651 U318PF9698 U318PF9746 U319PF9793 U317PF9652 U318PF9699 U318PF9747 U319PF9794 U317PF9653 U318PF9700 U318PF9748 U319PF9795 U317PF9654 U318PF9701 U318PF9749 U319PF9796 U317PF9655 U318PF9702 U318PF9750 U319PF9797 U317PF9656 U318PF9703 U318PF9751 U319PF9798 U317PF9657 U318PF9704 U318PF9752 U319PF9799 U317PF9658 U318PF9706 U318PF9753 U319PF9800 U317PF9659 U318PF9707 U318PF9754 U319PF9801 U317PF9660 U318PF9708 U318PF9755 U319PF9802 U317PF9661 U318PF9709 U318PF9756 U319PF9803 U317PF9662 U318PF9710 U318PF9757 U319PF9804 U317PF9663 U318PF9711 U319PF9758 U319PF9805 U317PF9664 U318PF9712 U319PF9759 U319PF9806 U317PF9665 U318PF9713 U319PF9760 U319PF9807 U317PF9666 U318PF9714 U319PF9761 U319PF9808 U317PF9667 U318PF9715 U319PF9762 U319PF9809 U317PF9668 U318PF9716 U319PF9763 U319PF9810 U317PF9669 U318PF9717 U319PF9764 U319PF9811 U317PF9670 U318PF9718 U319PF9765 U319PF9812 U317PF9671 U318PF9719 U319PF9766 U319PF9813 U317PF9672 U318PF9720 U319PF9767 U319PF9814 U317PF9673 U318PF9721 U319PF9768 U319PF9815 U317PF9674 U318PF9722 U319PF9769 U319PF9816 U317PF9675 U318PF9723 U319PF9770 U319PF9817 U317PF9676 U318PF9724 U319PF9771 U319PF9818 U317PF9677 U318PF9725 U319PF9772 U319PF9819 U323PF9961 U324PF0010 U324PF0057 U325PF0102 U323PF9962 U324PF0011 U324PF0058 U325PF0103 U323PF9963 U324PF0012 U324PF0059 U325PF0104 U323PF9964 U324PF0013 U324PF0060 U325PF0105 U323PF9965 U324PF0014 U324PF0061 U325PF0106 U323PF9966 U324PF0015 U324PF0062 U325PF0107 U323PF9967 U324PF0016 U324PF0063 U325PF0108 U323PF9968 U324PF0017 U324PF0064 U325PF0109 U323PF9969 U324PF0018 U324PF0065 U325PF0110 U323PF9970 U324PF0019 U324PF0066 U325PF0111 U323PF9971 U324PF0020 U324PF0067 U325PF0112 U323PF9972 U324PF0021 U324PF0068 U325PF0113 U323PF9973 U324PF0022 U324PF0069 U325PF0114 U323PF9974 U324PF0023 U324PF0070 U325PF0115 U323PF9975 U324PF0024 U324PF0071 U325PF0116 U323PF9976 U324PF0025 U324PF0072 U325PF0117 U323PF9977 U324PF0026 U324PF0073 U325PF0118 U323PF9978 U324PF0027 U324PF0074 U325PF0119 U323PF9979 U324PF0028 U324PF0075 U325PF0120 U323PF9980 U324PF0029 U324PF0076 U325PF0121 U323PF9981 U324PF0030 U324PF0077 U325PF0122 U323PF9982 U324PF0031 U324PF0078 U325PF0123 U323PF9983 U324PF0032 U324PF9998 U325PF0124 U323PF9984 U324PF0033 U324PF9999 U325PF0125 U323PF9985 U324PF0034 U325PF0079 U325PF0126 U323PF9986 U324PF0035 U325PF0080 U325PF0127 U323PF9987 U324PF0036 U325PF0081 U325PF0129 U323PF9988 U324PF0037 U325PF0082 U325PF0130 U323PF9989 U324PF0038 U325PF0083 U325PF0131 U323PF9990 U324PF0039 U325PF0084 U325PF0132 U323PF9991 U324PF0040 U325PF0085 U325PF0133 U323PF9992 U324PF0041 U325PF0086 U325PF0134 U323PF9993 U324PF0042 U325PF0087 U325PF0135 U323PF9994 U324PF0043 U325PF0088 U325PF0136 U323PF9995 U324PF0044 U325PF0089 U325PF0137 U323PF9996 U324PF0045 U325PF0090 U325PF0138 U323PF9997 U324PF0046 U325PF0091 U325PF0139 U324PF0000 U324PF0047 U325PF0092 U325PF0140 U324PF0001 U324PF0048 U325PF0093 U325PF0141 U324PF0002 U324PF0049 U325PF0094 U325PF0142 U324PF0003 U324PF0050 U325PF0095 U325PF0143 U324PF0004 U324PF0051 U325PF0096 U325PF0144 U324PF0005 U324PF0052 U325PF0097 U325PF0145 U324PF0006 U324PF0053 U325PF0098 U325PF0146 U324PF0007 U324PF0054 U325PF0099 U325PF0147 U324PF0008 U324PF0055 U325PF0100 U325PF0148 U324PF0009 U324PF0056 U325PF0101 U326PF0149 U329PF0292 U330PF0339 U331PF0386 U331PF0438 U329PF0293 U330PF0340 U331PF0387 U331PF0439 U329PF0294 U330PF0341 U331PF0388 U331PF0441 U329PF0295 U330PF0342 U331PF0389 U331PF0442 U329PF0296 U330PF0343 U331PF0390 U331PF0444 U329PF0297 U330PF0344 U331PF0391 U336PF0446 U329PF0298 U330PF0345 U331PF0392 U336PF0448 U329PF0299 U330PF0346 U331PF0393 U336PF0449 U329PF0300 U330PF0347 U331PF0394 U336PF0450 U329PF0301 U330PF0348 U331PF0395 U336PF0452 U329PF0302 U330PF0349 U331PF0396 U336PF0454 U329PF0303 U330PF0350 U331PF0397 U336PF0456 U329PF0304 U330PF0351 U331PF0398 U336PF0458 U329PF0305 U330PF0352 U331PF0399 U336PF0460 U329PF0306 U330PF0353 U331PF0400 U336PF0462 U329PF0307 U330PF0354 U331PF0401 U336PF0463 U329PF0308 U330PF0355 U331PF0402 U336PF0465 U329PF0309 U330PF0356 U331PF0403 U336PF0467 U329PF0310 U330PF0357 U331PF0404 U336PF0468 U329PF0311 U330PF0358 U331PF0405 U336PF0469 U330PF0312 U330PF0359 U331PF0406 U336PF0471 U330PF0313 U330PF0360 U331PF0407 U336PF0473 U330PF0314 U330PF0361 U331PF0408 U336PF0475 U330PF0315 U330PF0362 U331PF0409 U336PF0476 U330PF0316 U330PF0363 U331PF0410 U336PF0479 U330PF0317 U330PF0364 U331PF0411 U336PF0481 U330PF0318 U330PF0365 U331PF0412 U336PF0483 U330PF0319 U330PF0366 U331PF0413 U336PF0484 U330PF0320 U330PF0367 U331PF0414 U336PF0485 U330PF0321 U330PF0368 U331PF0415 U336PF0486 U330PF0322 U330PF0369 U331PF0416 U336PF0487 U330PF0323 U331PF0370 U331PF0417 U336PF0488 U330PF0324 U331PF0371 U331PF0418 U336PF0489 U330PF0325 U331PF0372 U331PF0419 U336PF0490 U330PF0326 U331PF0373 U331PF0420 U336PF0491 U330PF0327 U331PF0374 U331PF0421 U336PF0492 U330PF0328 U331PF0375 U331PF0422 U336PF0493 U330PF0329 U331PF0376 U331PF0423 U336PF0494 U330PF0330 U331PF0377 U331PF0424 U336PF0495 U330PF0331 U331PF0378 U331PF0425 U336PF0496 U330PF0332 U331PF0379 U331PF0426 U336PF0497 U330PF0333 U331PF0380 U331PF0428 U336PF0498 U330PF0334 U331PF0381 U331PF0429 U336PF0499 U330PF0335 U331PF0382 U331PF0430 U336PF0500 U330PF0336 U331PF0383 U331PF0432 U336PF0501 U330PF0337 U331PF0384 U331PF0435 U336PF0502 U330PF0338 U331PF0385 U331PF0436 U336PF0503 U338PF0665 U339PF0712 U339PF0759 U340PF0806 U338PF0666 U339PF0713 U339PF0760 U340PF0807 U338PF0667 U339PF0714 U339PF0761 U340PF0808 U338PF0668 U339PF0715 U339PF0762 U340PF0809 U338PF0669 U339PF0716 U339PF0763 U340PF0810 U338PF0670 U339PF0717 U339PF0764 U340PF0811 U338PF0671 U339PF0718 U339PF0765 U340PF0812 U338PF0672 U339PF0719 U339PF0766 U340PF0813 U338PF0673 U339PF0720 U339PF0767 U340PF0814 U338PF0674 U339PF0721 U339PF0768 U340PF0815 U338PF0675 U339PF0722 U339PF0769 U340PF0816 U338PF0676 U339PF0723 U339PF0770 U340PF0817 U338PF0677 U339PF0724 U339PF0771 U340PF0818 U338PF0678 U339PF0725 U339PF0772 U340PF0819 U338PF0679 U339PF0726 U339PF0773 U340PF0820 U338PF0680 U339PF0727 U339PF0774 U340PF0821 U338PF0681 U339PF0728 U339PF0775 U340PF0822 U338PF0682 U339PF0729 U339PF0776 U340PF0823 U338PF0683 U339PF0730 U339PF0777 U340PF0824 U338PF0684 U339PF0731 U339PF0778 U340PF0825 U338PF0685 U339PF0732 U339PF0779 U340PF0826 U338PF0686 U339PF0733 U339PF0780 U340PF0827 U338PF0687 U339PF0734 U339PF0781 U340PF0828 U338PF0688 U339PF0735 U339PF0782 U340PF0830 U338PF0689 U339PF0736 U339PF0783 U340PF0831 U338PF0690 U339PF0737 U339PF0784 U340PF0832 U338PF0691 U339PF0738 U339PF0785 U340PF0833 U338PF0692 U339PF0739 U339PF0786 U340PF0834 U338PF0693 U339PF0740 U339PF0787 U340PF0835 U338PF0694 U339PF0741 U339PF0788 U340PF0836 U338PF0695 U339PF0742 U339PF0789 U340PF0837 U338PF0696 U339PF0743 U339PF0790 U340PF0838 U338PF0697 U339PF0744 U339PF0791 U340PF0839 U338PF0698 U339PF0745 U339PF0792 U340PF0840 U338PF0699 U339PF0746 U339PF0793 U340PF0841 U338PF0700 U339PF0747 U339PF0794 U340PF0842 U338PF0701 U339PF0748 U339PF0795 U340PF0843 U338PF0702 U339PF0749 U339PF0796 U340PF0844 U338PF0703 U339PF0750 U339PF0797 U340PF0845 U338PF0704 U339PF0751 U339PF0798 U340PF0846 U338PF0705 U339PF0752 U339PF0799 U340PF0847 U338PF0706 U339PF0753 U340PF0800 U340PF0848 U338PF0707 U339PF0754 U340PF0801 U340PF0849 U338PF0708 U339PF0755 U340PF0802 U340PF0850 U339PF0709 U339PF0756 U340PF0803 U340PF0851 U339PF0710 U339PF0757 U340PF0804 U340PF0852 U339PF0711 U339PF0758 U340PF0805 U340PF0853 U344PF0997 U344PF1044 U345PF1091 U345PF1138 U344PF0998 U344PF1045 U345PF1092 U345PF1139 U344PF0999 U344PF1046 U345PF1093 U345PF1140 U344PF1000 U344PF1047 U345PF1094 U345PF1141 U344PF1001 U344PF1048 U345PF1095 U345PF1142 U344PF1002 U344PF1049 U345PF1096 U345PF1143 U344PF1003 U344PF1050 U345PF1097 U345PF1144 U344PF1004 U344PF1051 U345PF1098 U345PF1145 U344PF1005 U344PF1052 U345PF1099 U345PF1146 U344PF1006 U344PF1053 U345PF1100 U345PF1147 U344PF1007 U344PF1054 U345PF1101 U345PF1148 U344PF1008 U344PF1055 U345PF1102 U345PF1149 U344PF1009 U344PF1056 U345PF1103 U345PF1150 U344PF1010 U344PF1057 U345PF1104 U345PF1151 U344PF1011 U344PF1058 U345PF1105 U346PF1152 U344PF1012 U344PF1059 U345PF1106 U346PF1153 U344PF1013 U344PF1060 U345PF1107 U346PF1154 U344PF1014 U345PF1061 U345PF1108 U346PF1155 U344PF1015 U345PF1062 U345PF1109 U346PF1156 U344PF1016 U345PF1063 U345PF1110 U346PF1157 U344PF1017 U345PF1064 U345PF1111 U346PF1158 U344PF1018 U345PF1065 U345PF1112 U346PF1159 U344PF1019 U345PF1066 U345PF1113 U346PF1160 U344PF1020 U345PF1067 U345PF1114 U346PF1161 U344PF1021 U345PF1068 U345PF1115 U346PF1162 U344PF1022 U345PF1069 U345PF1116 U346PF1163 U344PF1023 U345PF1070 U345PF1117 U346PF1164 U344PF1024 U345PF1071 U345PF1118 U346PF1165 U344PF1025 U345PF1072 U345PF1119 U346PF1166 U344PF1026 U345PF1073 U345PF1120 U346PF1167 U344PF1027 U345PF1074 U345PF1121 U346PF1168 U344PF1028 U345PF1075 U345PF1122 U346PF1170 U344PF1029 U345PF1076 U345PF1123 U346PF1171 U344PF1030 U345PF1077 U345PF1124 U346PF1172 U344PF1031 U345PF1078 U345PF1125 U346PF1173 U344PF1032 U345PF1079 U345PF1126 U346PF1174 U344PF1033 U345PF1080 U345PF1127 U346PF1175 U344PF1034 U345PF1081 U345PF1128 U346PF1176 U344PF1035 U345PF1082 U345PF1129 U346PF1177 U344PF1036 U345PF1083 U345PF1130 U346PF1178 U344PF1037 U345PF1084 U345PF1131 U346PF1179 U344PF1038 U345PF1085 U345PF1132 U346PF1180 U344PF1039 U345PF1086 U345PF1133 U346PF1181 U344PF1040 U345PF1087 U345PF1134 U346PF1182 U344PF1041 U345PF1088 U345PF1135 U346PF1183 U344PF1042 U345PF1089 U345PF1136 U346PF1184 U344PF1043 U345PF1090 U345PF1137 U346PF1185 U350PF1336 U350PF1383 U351PF1430 U351PF1477 U350PF1337 U350PF1384 U351PF1431 U351PF1478 U350PF1338 U350PF1385 U351PF1432 U351PF1479 U350PF1339 U350PF1386 U351PF1433 U351PF1480 U350PF1340 U350PF1387 U351PF1434 U351PF1481 U350PF1341 U350PF1388 U351PF1435 U351PF1482 U350PF1342 U350PF1389 U351PF1436 U351PF1483 U350PF1343 U350PF1390 U351PF1437 U351PF1484 U350PF1344 U350PF1391 U351PF1438 U351PF1485 U350PF1345 U350PF1392 U351PF1439 U351PF1486 U350PF1346 U350PF1393 U351PF1440 U351PF1487 U350PF1347 U350PF1394 U351PF1441 U351PF1488 U350PF1348 U350PF1395 U351PF1442 U351PF1489 U350PF1349 U350PF1396 U351PF1443 U351PF1490 U350PF1350 U350PF1397 U351PF1444 U351PF1491 U350PF1351 U350PF1398 U351PF1445 U351PF1492 U350PF1352 U350PF1399 U351PF1446 U351PF1493 U350PF1353 U350PF1400 U351PF1447 U351PF1494 U350PF1354 U350PF1401 U351PF1448 U351PF1495 U350PF1355 U350PF1402 U351PF1449 U352PF1496 U350PF1356 U350PF1403 U351PF1450 U352PF1497 U350PF1357 U350PF1404 U351PF1451 U352PF1498 U350PF1358 U350PF1405 U351PF1452 U352PF1499 U350PF1359 U350PF1406 U351PF1453 U352PF1500 U350PF1360 U350PF1407 U351PF1454 U352PF1501 U350PF1361 U351PF1408 U351PF1455 U352PF1502 U350PF1362 U351PF1409 U351PF1456 U352PF1503 U350PF1363 U351PF1410 U351PF1457 U352PF1504 U350PF1364 U351PF1411 U351PF1458 U352PF1505 U350PF1365 U351PF1412 U351PF1459 U352PF1506 U350PF1366 U351PF1413 U351PF1460 U352PF1507 U350PF1367 U351PF1414 U351PF1461 U352PF1508 U350PF1368 U351PF1415 U351PF1462 U352PF1509 U350PF1369 U351PF1416 U351PF1463 U352PF1510 U350PF1370 U351PF1417 U351PF1464 U352PF1511 U350PF1371 U351PF1418 U351PF1465 U352PF1512 U350PF1372 U351PF1419 U351PF1466 U352PF1513 U350PF1373 U351PF1420 U351PF1467 U352PF1514 U350PF1374 U351PF1421 U351PF1468 U352PF1515 U350PF1375 U351PF1422 U351PF1469 U352PF1516 U350PF1376 U351PF1423 U351PF1470 U352PF1517 U350PF1377 U351PF1424 U351PF1471 U352PF1518 U350PF1378 U351PF1425 U351PF1472 U352PF1519 U350PF1379 U351PF1426 U351PF1473 U352PF1520 U350PF1380 U351PF1427 U351PF1474 U352PF1521 U350PF1381 U351PF1428 U351PF1475 U352PF1522 U350PF1382 U351PF1429 U351PF1476 U352PF1523 U353PF1670 U354PF1764 V006PF1876 V007PF1971 U353PF1671 U354PF1765 V006PF1877 V007PF1972 U353PF1672 U354PF1766 V006PF1878 V007PF1973 U353PF1673 U354PF1767 V006PF1879 V007PF1974 U353PF1674 U354PF1768 V006PF1880 V007PF1975 U353PF1675 U354PF1769 V006PF1881 V007PF1976 U353PF1676 U354PF1770 V006PF1882 V007PF1977 U353PF1677 U354PF1771 V006PF1883 V007PF1978 U353PF1678 U354PF1772 V006PF1884 V007PF1979 U353PF1679 U354PF1773 V006PF1885 V007PF1980 U353PF1680 U354PF1774 V006PF1886 V007PF1981 U353PF1681 U354PF1775 V006PF1887 V007PF1982 U353PF1682 U354PF1776 V006PF1888 V007PF1983 U353PF1683 U354PF1777 V006PF1890 V007PF1984 U353PF1684 U354PF1778 V006PF1891 V007PF1985 U353PF1685 U354PF1779 V006PF1892 V007PF1986 U353PF1686 U354PF1780 V006PF1893 V007PF1987 U353PF1687 U354PF1781 V006PF1894 V007PF1988 U354PF1688 U354PF1782 V006PF1895 V007PF1989 U354PF1689 U354PF1783 V006PF1896 V007PF1990 U354PF1690 U354PF1784 V006PF1897 V007PF1991 U354PF1691 U354PF1785 V006PF1898 V007PF1992 U354PF1692 U354PF1786 V006PF1899 V008PF1993 U354PF1693 U355PF1787 V006PF1900 V008PF1994 U354PF1694 U355PF1788 V006PF1901 V008PF1995 U354PF1695 U355PF1789 V006PF1902 V008PF1996 U354PF1696 U355PF1790 V006PF1903 V008PF1997 U354PF1697 U355PF1791 V006PF1904 V008PF1998 U354PF1698 U355PF1792 V006PF1905 V008PF1999 U354PF1699 U355PF1793 V006PF1906 V008PF2000 U354PF1700 U355PF1794 V006PF1907 V008PF2001 U354PF1701 U355PF1795 V006PF1908 V008PF2002 U354PF1702 U355PF1796 V006PF1909 V008PF2003 U354PF1703 U355PF1797 V006PF1910 V008PF2004 U354PF1704 U355PF1798 V006PF1911 V008PF2005 U354PF1705 U355PF1799 V006PF1912 V008PF2006 U354PF1706 U355PF1800 V006PF1913 V008PF2007 U354PF1707 U355PF1801 V006PF1914 V008PF2008 U354PF1708 U355PF1802 V006PF1915 V008PF2009 U354PF1709 U355PF1803 V006PF1916 V008PF2010 U354PF1710 U355PF1805 V006PF1917 V008PF2011 U354PF1711 U355PF1806 V006PF1918 V008PF2012 U354PF1712 U355PF1808 V006PF1919 V008PF2013 U354PF1713 U355PF1809 V007PF1920 V008PF2014 U354PF1714 U355PF1810 V007PF1921 V008PF2015 U354PF1715 U355PF1811 V007PF1922 V008PF2016 U354PF1716 U355PF1813 V007PF1923 V008PF2017 U354PF1717 U355PF1814 V007PF1924 V008PF2018 U354PF1718 U355PF1816 V007PF1925 V008PF2019 U354PF1719 U355PF1817 V007PF1926 V008PF2020 U354PF1720 U355PF1819 V007PF1927 V008PF2021 U354PF1721 U355PF1820 V007PF1928 V008PF2022 U354PF1722 U355PF1822 V007PF1929 V008PF2023 U354PF1723 U355PF1823 V007PF1930 V008PF2025 U354PF1724 U355PF1824 V007PF1931 V008PF2026 U354PF1725 U355PF1825 V007PF1932 V008PF2028 U354PF1726 U355PF1826 V007PF1933 V008PF2029 U354PF1727 U355PF1827 V007PF1934 V008PF2031 U354PF1728 U355PF1829 V007PF1935 V008PF2032 U354PF1729 U355PF1830 V007PF1936 V008PF2034 U354PF1730 U355PF1832 V007PF1937 V008PF2035 U354PF1731 U355PF1833 V007PF1938 V008PF2036 U354PF1732 U355PF1835 V007PF1939 V008PF2037 U354PF1733 U355PF1836 V007PF1940 V008PF2038 U354PF1734 V006PF1839 V007PF1941 V008PF2039 U354PF1735 V006PF1840 V007PF1942 V008PF2040 U354PF1736 V006PF1842 V007PF1943 V008PF2041 U354PF1737 V006PF1843 V007PF1944 V008PF2042 U354PF1738 V006PF1845 V007PF1945 V008PF2043 U354PF1739 V006PF1846 V007PF1946 V008PF2044 U354PF1740 V006PF1848 V007PF1947 V008PF2045 U354PF1741 V006PF1849 V007PF1948 V008PF2046 U354PF1742 V006PF1851 V007PF1949 V008PF2047 U354PF1743 V006PF1852 V007PF1950 V008PF2048 U354PF1744 V006PF1854 V007PF1951 V008PF2049 U354PF1745 V006PF1855 V007PF1952 V008PF2050 U354PF1746 V006PF1857 V007PF1953 V008PF2051 U354PF1747 V006PF1858 V007PF1954 V008PF2052 U354PF1748 V006PF1860 V007PF1955 V008PF2053 U354PF1749 V006PF1861 V007PF1956 V008PF2054 U354PF1750 V006PF1862 V007PF1957 V008PF2055 U354PF1751 V006PF1863 V007PF1958 V008PF2056 U354PF1752 V006PF1864 V007PF1959 V008PF2057 U354PF1753 V006PF1865 V007PF1960 V008PF2058 U354PF1754 V006PF1866 V007PF1961 V008PF2059 U354PF1755 V006PF1867 V007PF1962 V008PF2060 U354PF1756 V006PF1868 V007PF1963 V008PF2061 U354PF1757 V006PF1869 V007PF1964 V008PF2062 U354PF1758 V006PF1870 V007PF1965 V008PF2063 U354PF1759 V006PF1871 V007PF1966 V008PF2064 U354PF1760 V006PF1872 V007PF1967 V008PF2065 U354PF1761 V006PF1873 V007PF1968 V008PF2066 U354PF1762 V006PF1874 V007PF1969 V008PF2067 U354PF1763 V006PF1875 V007PF1970 V008PF2068 U141PF9106 U141PF9135 U154PF9670 U158PF9930 U165PF0400 U165PF0438 U173PF0959 U177PF1239 U178PF1302 U210PF2668 U220PF3462 U226PF3868 U231PF4135 U235PF4480 U236PF4663 U242PF5081 U256PF5999 U263PF6561 U266PF6740 U267PF6805 U275PF7295 U280PF7513 U280PF7535 U281PF7607 U283PF7790 |
| FEI Number |
1824206
|
Recalling Firm/ Manufacturer |
Hill-Rom, Inc. 1069 State Road 46 E Batesville IN 47006-7520
|
| For Additional Information Contact | Hill-Rom Technical Support 800-445-3720 |
Manufacturer Reason for Recall | The Bed Exit System may fail to send a remote alert through the nurse call system if a remote alert was previously sent and cancelled at the in-room nurse call wall unit. |
FDA Determined Cause 2 | Software design |
| Action | On April 30, 2020, the firm distributed Urgent Medial Device Correction letters to affected customers. Customers were informed that though the bed exit system will alert at the bed, it may fail to send a remote alert through the nurse call system.
The firm's letter states that the risk can be immediately mitigated by changing the nurse call setting from "SideCom" to "With A.C." For detailed instructions, please see "Instructions: Changing the Nurse Call Setting" section of the letter. Customers should also complete the response form and return it to the firm.
The latest software (version 1.28.000 or greater) resolves the communication issue and mitigates the potential patient risk. A software update to the latest software is required for affected devices.
Once Hill-Rom receives your response form and has confirmed that your device is affected, the firm will contact you to schedule your software update. If you have any questions or would like to proactively schedule your software update, please contact Hill-Rom Technical Support at 1-800-445-3720 (select option 2 and then select option 1) Monday through Friday, 8:00AM 5:00 PM EST. |
| Quantity in Commerce | 4381 beds |
| Distribution | Domestic distribution nationwide. Foreign distribution to Canada and Australia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|