• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall CARESCAPE R860

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall CARESCAPE R860see related information
Date Initiated by FirmJuly 14, 2020
Date PostedAugust 24, 2020
Recall Status1 Terminated 3 on September 23, 2024
Recall NumberZ-2883-2020
Recall Event ID 86107
510(K)NumberK142679 
Product Classification Ventilator, continuous, facility use - Product Code CBK
ProductCARESCAPE R860 Ventilator The CARESCAPE R860 ventilator is a microprocessor based, electronically controlled, pneumatically driven ventilator that includes integrated monitoring of FiO2, airway pressure, flow, and Volume.
Code Information CARESCAPE R860: 1506-8600-000 Spare parts: CARESCAPE R860: 1505-3215-000, 2071409-S, 2071410-001 Engstrom: M1115997-S, M1081816-S  Identifier # CBRZ03152, CBRZ03159, CBRZ03171, CBRZ03173, CBRZ03174, CBRZ03175, CBRZ03176, CBRZ03177, CBRZ03180, CBRZ03182, CBRZ03867, CBRZ03869, CBRZ03870, CBRZ03873, CBRZ03856, CBRZ03259, CBRZ03261, CBRZ03263, CBRZ03264

CBRZ03265, CBRZ03267, CBRZ03274, CBRZ03275, CBRZ03278, CBRZ03281, CBRZ03928, CBRZ03946, CBRZ03989, CBRZ03990, CBRZ03994, CBRZ03845, CBRZ03857, CBRZ03858, CBRZ03861, CBRZ03868, CBRZ03872, CBRZ03879, CBRZ03880, CBRZ03881, CBRZ03883, CBRZ03889, CBRZ03890, CBRZ03891, CBRZ03484, CBRZ03485, CBRZ03487, CBRZ03488, CBRZ03490, CBRZ03502, CBRZ03793, CBRZ03842, CBRZ03843, CBRZ03849, CBRZ03850, CBRZ03851, CBRZ03853, CBRZ03864, CBRZ03865, CBRZ03866, CBRZ03438, CBRZ03501, CBRZ03503, CBRZ03506, CBRZ03508, CBRZ03609, CBRZ03610, CBRZ03620, CBRZ03621, CBRZ03625, CBRZ03342, CBRZ03437, CBRZ03451, CBRZ03724, CBRZ03730, CBRZ03731, CBRZ03732, CBRZ03733, CBRZ03734, CBRZ03759, CBRZ03767, CBRZ03768, CBRZ03769, CBRZ03771, CBRZ03772, CBRZ03773, CBRZ03774, CBRZ03775, CBRZ03776, CBRZ03777, CBRZ03778, CBRZ03779, CBRZ03780, CBRZ03781, CBRZ03782, CBRZ03783, CBRZ03784, CBRZ03785, CBRZ03786, CBRZ03787, CBRZ03788, CBRZ03789, CBRZ03791, CBRZ03792, CBRZ03794, CBRZ03795, CBRZ03796, CBRZ03797, CBRZ03798, CBRZ03799, CBRZ03800, CBRZ03801, CBRZ03802, CBRZ03803, CBRZ03804, CBRZ03805, CBRZ03806, CBRZ03807, CBRZ03808, CBRZ03809, CBRZ03810, CBRZ03812, CBRZ03813, CBRZ03815, CBRZ03816, CBRZ03817, CBRZ03819, CBRZ03821, CBRZ03822, CBRZ03823, CBRZ03824, CBRZ03825, CBRZ03826, CBRZ03827, CBRZ03828, CBRZ03829, CBRZ03830, CBRZ03831, CBRZ03832, CBRZ03834, CBRZ03835, CBRZ03836, CBRZ03837, CBRZ03838, CBRZ03839, CBRZ03841, CBRZ03877, CBRZ03913, CBRZ03925, CBRZ03949, CBRZ03951, CBRZ03952, CBRZ03954, CBRZ03955, CBRZ03956, CBRZ03957, CBRZ03958, CBRZ03959, CBRZ03961, CBRZ03962, CBRZ03979, CBRZ03984, CBRZ03985, CBRZ03997, CBRZ03998, CBRZ04001, CBRZ04003, CBRZ04006, CBRZ04008, CBRZ04009, CBRZ04011, CBRZ04020, CBRZ04025, CBRZ04036, CBRZ04038, CBRZ04039, CBRZ04040, CBRZ04041, CBRZ04042, CBRZ04043, CBRZ04046, CBRZ04047, CBRZ04050, CBRZ04051, CBRZ04052, CBRZ04053, CBRZ04054, CBRZ04057, CBRZ04058, CBRZ04059, CBRZ04061, CBRZ04062, CBRZ04063, CBRZ04064, CBRZ04065, CBRZ04066, CBRZ04067, CBRZ04068, CBRZ04069, CBRZ04070, CBRZ04071, CBRZ04072, CBRZ04073, CBRZ04074, CBRZ04075, CBRZ04076, CBRZ04078, CBRZ04080, CBRZ04081, CBRZ04082, CBRZ04083, CBRZ04084, CBRZ04085, CBRZ04086, CBRZ04087, CBRZ04088, CBRZ04089, CBRZ04090, CBRZ04091, CBRZ04092, CBRZ04093, CBRZ04094, CBRZ04095, CBRZ04096, CBRZ04097, CBRZ04098, CBRZ04099, CBRZ04100, CBRZ04101, CBRZ04102, CBRZ04103, CBRZ04104, CBRZ04105, CBRZ04106, CBRZ04107, CBRZ04108, CBRZ04109, CBRZ04110, CBRZ04111, CBRZ04112, CBRZ04113, CBRZ04115, CBRZ04116, CBRZ04117, CBRZ04118, CBRZ04119, CBRZ04120, CBRZ04121, CBRZ04122, CBRZ04123, CBRZ04124, CBRZ04130, CBRZ04137, CBRZ04139, CBRZ04196, CBRZ04212, CBRZ04269, CBRZ04273, CBRZ04330, CBRZ04336, CBRZ04373, CBRZ04374, CBRZ04376, CBRZ04377, CBRZ04378, CBRZ04379, CBRZ04380, CBRZ04387, CBRZ04389, CBRZ04438, CBRZ04446, CBRZ04447, CBRZ04449, CBRZ04450, CBRZ04485, CBRZ04491, CBRZ04506, CBRZ04511, CBRZ04524, CBRZ04649, CBRZ04684, CBRZ04692, CBRZ04693, CBRZ04694, CBRZ04695, CBRZ04696, CBRZ04698, CBRZ04699, CBRZ04700, CBRZ04701, CBRZ04702, CBRZ04703, CBRZ04704, CBRZ04705, CBRZ04706, CBRZ04707, CBRZ04708, CBRZ04709, CBRZ04710, CBRZ04711, CBRZ04712, CBRZ04713, CBRZ04714, CBRZ04715, CBRZ04716, CBRZ04717, CBRZ04718, CBRZ04719, CBRZ04720, CBRZ04721, CBRZ04722, CBRZ04723, CBRZ04724, CBRZ04725, CBRZ04726, CBRZ04727, CBRZ04728, CBRZ04729, CBRZ04730, CBRZ04731, CBRZ04733, CBRZ04734, CBRZ04735, CBRZ04736, CBRZ04737, CBRZ04738, CBRZ04739, CBRZ04740, CBRZ04741, CBRZ04742, CBRZ04743, CBRZ04744, CBRZ04745, CBRZ04746, CBRZ04747, CBRZ04748, CBRZ04749, CBRZ04750, CBRZ04751, CBRZ04752, CBRZ04753, CBRZ04754, CBRZ04755, CBRZ04759, CBRZ04762, CBRZ04770, CBRZ04772, CBRZ04778, CBRZ04779, CBRZ04788, CBRZ04790, CBRZ04791, CBRZ04800, CBRZ04813, CBRZ04814, CBRZ04815, CBRZ04816, CBRZ04817, CBRZ04818, CBRZ04819, CBRZ04820, CBRZ04821, CBRZ04822, CBRZ04823, CBRZ04824, CBRZ04825, CBRZ04826, CBRZ04827, CBRZ04828, CBRZ04829, CBRZ04830, CBRZ04831, CBRZ04832, CBRZ04833, CBRZ04834, CBRZ04837,CBRZ04838, CBRZ04839, CBRZ04840, CBRZ04841, CBRZ04842, CBRZ04843, CBRZ04847, CBRZ04848, CBRZ04849, CBRZ04850, CBRZ04851, CBRZ04852, CBRZ04853, CBRZ04854, CBRZ04855, CBRZ04856, CBRZ04857, CBRZ04859, CBRZ04860, CBRZ04861, CBRZ04862, CBRZ04863, CBRZ04864, CBRZ04865, CBRZ04866, CBRZ04867, CBRZ04868, CBRZ04869, CBRZ04870, CBRZ04871, CBRZ04872, CBRZ04873, CBRZ04874, CBRZ04875, CBRZ04876, CBRZ04877, CBRZ04878, CBRZ04879, CBRZ04880, CBRZ04881, CBRZ04882, CBRZ04883, CBRZ04884, CBRZ04885, CBRZ04886, CBRZ04887, CBRZ04888, CBRZ04889, CBRZ04890, CBRZ04891, CBRZ04892, CBRZ04893, CBRZ04894, CBRZ04895, CBRZ04897, CBRZ04898, CBRZ04899, CBRZ04900, CBRZ04901, CBRZ04902, CBRZ04903, CBRZ04904, CBRZ04905, CBRZ04906, CBRZ04907, CBRZ04908, CBRZ04909, CBRZ04910, CBRZ04911, CBRZ04912, CBRZ04913, CBRZ04914, CBRZ04915, CBRZ04917, CBRZ04918, CBRZ04919, CBRZ04920, CBRZ04921, CBRZ04922, CBRZ04923, CBRZ04924, CBRZ04925, CBRZ04927, CBRZ04928, CBRZ04929, CBRZ04930, CBRZ04931, CBRZ04932, CBRZ04933, CBRZ04934, CBRZ04935, CBRZ04936, CBRZ04947, CBRZ04951, CBRZ05012, CBRZ05013, CBRZ05014, CBRZ05015, CBRZ05016, CBRZ05017, CBRZ05018, CBRZ05019, CBRZ05020, CBRZ05021, CBRZ05022, CBRZ05023, CBRZ05024, CBRZ05025, CBRZ05026, CBRZ05027, CBRZ05028, CBRZ05029, CBRZ05030, CBRZ05031, CBRZ05032, CBRZ05033, CBRZ05034, CBRZ05035, CBRZ05036, CBRZ05037, CBRZ05038, CBRZ05039, CBRZ05040, CBRZ05041, CBRZ05042, CBRZ05043, CBRZ05044, CBRZ05045, CBRZ05046, CBRZ05047, CBRZ05048, CBRZ05049, CBRZ05050, CBRZ05051,CBRZ05052, CBRZ05053, CBRZ05054, CBRZ05055, CBRZ05056, CBRZ05057, CBRZ05058, CBRZ05059, CBRZ05060, CBRZ05061, CBRZ05062, CBRZ05063, CBRZ05064, CBRZ05065, CBRZ05066, CBRZ05067, CBRZ05068, CBRZ05069, CBRZ05070, CBRZ05071, CBRZ05072, CBRZ05073, CBRZ05074, CBRZ05075, CBRZ05076, CBRZ05077, CBRZ05078, CBRZ05079, CBRZ05080, CBRZ05081, CBRZ05082, CBRZ05083, CBRZ05084, CBRZ05085, CBRZ05086, CBRZ05087, CBRZ05088, CBRZ05089, CBRZ05090, CBRZ05091, CBRZ05092, CBRZ05093, CBRZ05094, CBRZ05095, CBRZ05096, CBRZ05097, CBRZ05098, CBRZ05099, CBRZ05100, CBRZ05101, CBRZ05102, CBRZ05103, CBRZ05106, CBRZ05107, CBRZ05108, CBRZ05109, CBRZ05110, CBRZ05111, CBRZ05112, CBRZ05113 CBRZ05114, CBRZ05115, CBRZ05116, CBRZ05117, CBRZ05118, CBRZ05119, CBRZ05120, CBRZ05121, CBRZ05122, CBRZ05123, CBRZ05124, CBRZ05125, CBRZ05126, CBRZ05127, CBRZ05128, CBRZ05129, CBRZ05130, CBRZ05131, CBRZ05132, CBRZ05133, CBRZ05134, CBRZ05135, CBRZ05136, CBRZ05137, CBRZ05138, CBRZ05139, CBRZ05140, CBRZ05141, CBRZ05142, CBRZ05143, CBRZ05144, CBRZ05145, CBRZ05146, CBRZ05147, CBRZ05148, CBRZ05149, CBRZ05150, CBRZ05151, CBRZ05152, CBRZ05154, CBRZ05153, CBRZ05155, CBRZ05156, CBRZ05157, CBRZ05158, CBRZ05159, CBRZ05160, CBRZ05161, CBRZ05162, CBRZ05163, CBRZ05164, CBRZ05165, CBRZ05166, CBRZ05167, CBRZ05168, CBRZ05169, CBRZ05170, CBRZ05171, CBRZ05172, CBRZ05173, CBRZ05174, CBRZ05175, CBRZ05176, CBRZ05177, CBRZ05178 CBRZ05179, CBRZ05180, CBRZ05181, CBRZ05182, CBRZ05183, CBRZ05184, CBRZ05185, CBRZ05186, CBRZ05187, CBRZ05188, CBRZ05189, CBRZ05190, CBRZ05191, CBRZ05192, CBRZ05193, CBRZ05194, CBRZ05195, CBRZ05196, CBRZ05197, CBRZ05198, CBRZ05199, CBRZ05200, CBRZ05201, CBRZ05202, CBRZ05203, CBRZ05204, CBRZ05205, CBRZ05206, CBRZ05207, CBRZ05208, CBRZ05209, CBRZ05210, CBRZ05211, CBRZ05212, CBRZ05213, CBRZ05214, CBRZ05215, CBRZ05216, CBRZ05217, CBRZ05218, CBRZ05219, CBRZ05220, CBRZ05221, CBRZ05222, CBRZ05223, CBRZ05224, CBRZ05225, CBRZ05226, CBRZ05227, CBRZ05228, CBRZ05229, CBRZ05230, CBRZ05231, CBRZ05232, CBRZ05233, CBRZ05234, CBRZ05235, CBRZ05236, CBRZ05237, CBRZ05238, CBRZ05239, CBRZ05241, CBRZ05242, CBRZ05244, CBRZ05245, CBRZ05247, CBRZ05248, CBRZ05249, CBRZ05252, CBRZ05255, CBRZ05256, CBRZ05257, CBRZ05258, CBRZ05259, CBRZ05260, CBRZ05261, CBRZ05262, CBRZ05263, CBRZ05264, CBRZ05265, CBRZ05266, CBRZ05267, CBRZ05268, CBRZ05273, CBRZ05276, CBRZ05277, CBRZ05278, CBRZ05279, CBRZ05280, CBRZ05281, CBRZ05282, CBRZ05283, CBRZ05284, CBRZ05285, CBRZ05286, CBRZ05287, CBRZ05288, CBRZ05289, CBRZ05290, CBRZ05291, CBRZ05292, CBRZ05294, CBRZ05295, CBRZ05296, CBRZ05302, CBRZ05303, CBRZ05304, CBRZ05305, CBRZ05306, CBRZ05309, CBRZ05311, CBRZ05312, CBRZ05313, CBRZ05314 CBRZ05315, CBRZ05317, CBRZ05318, CBRZ05319, CBRZ05320, CBRZ05321, CBRZ05322, CBRZ05323, CBRZ05324, CBRZ05326, CBRZ05327, CBRZ05328, CBRZ05329, CBRZ05330 CBRZ05331, CBRZ05332, CBRZ05333, CBRZ05334, CBRZ05335, CBRZ05340, CBRZ40010, CBRZ40038, CBRZ40039, CBRZ40040, CBRZ03469, CBRZ03504, CBRZ03505, CBRZ03510, CBRZ03513, CBRZ03514, CBRZ03515, CBRZ03516, CBRZ03517, CBRZ03518, CBRZ03522, CBRZ03523, CBRZ03526, CBRZ03531, CBRZ03532, CBRZ03533, CBRZ03534, CBRZ03535, CBRZ03536, CBRZ03537, CBRZ03538, CBRZ03539, CBRZ03540, CBRZ03541, CBRZ03543, CBRZ03544, CBRZ03548, CBRZ03550, CBRZ03552, CBRZ02183, CBRZ03110, CBRZ03128, CBRZ03130, CBRZ03131, CBRZ03132, CBRZ03133, CBRZ03134, CBRZ03135, CBRZ03136, CBRZ03137, CBRZ03138, CBRZ03139, CBRZ03140, CBRZ03141, CBRZ03142, CBRZ03143, CBRZ03144, CBRZ03145, CBRZ03146, CBRZ03147, CBRZ03148, CBRZ03149, CBRZ03150, CBRZ03151, CBRZ03153, CBRZ03154, CBRZ03156, CBRZ03158, CBRZ03160, CBRZ03161, CBRZ03163, CBRZ03164, CBRZ03165, CBRZ03167, CBRZ03170, CBRZ03185, CBRZ03188, CBRZ03202, CBRZ03206, CBRZ03208, CBRZ03209, CBRZ03210, CBRZ03211, CBRZ03212, CBRZ03213, CBRZ03214, CBRZ03215, CBRZ03216, CBRZ03217, CBRZ03218, CBRZ03219, CBRZ03220, CBRZ03221, CBRZ03222, CBRZ03223, CBRZ03226, CBRZ03227, CBRZ03228, CBRZ03229, CBRZ03230, CBRZ03232, CBRZ03234, CBRZ03236, CBRZ03237, CBRZ03238, CBRZ03239, CBRZ03240, CBRZ03241, CBRZ03242, CBRZ03243, CBRZ03244, CBRZ03245, CBRZ03246, CBRZ03247, CBRZ03248, CBRZ03249, CBRZ03250, CBRZ03251, CBRZ03252, CBRZ03253, CBRZ03254, CBRZ03255, CBRZ03256, CBRZ03258, CBRZ03262, CBRZ03269, CBRZ03270, CBRZ03271, CBRZ03272, CBRZ03273, CBRZ03277, CBRZ03279, CBRZ03280, CBRZ03283, CBRZ03286, CBRZ03287, CBRZ03288, CBRZ03289, CBRZ03290, CBRZ03291, CBRZ03292, CBRZ03293, CBRZ03294, CBRZ03295, CBRZ03296, CBRZ03297, CBRZ03298, CBRZ03299, CBRZ03300,CBRZ03301, CBRZ03302, CBRZ03303, CBRZ03304, CBRZ03305, CBRZ03306, CBRZ03307, CBRZ03308, CBRZ03309, CBRZ03310, CBRZ03311, CBRZ03315, CBRZ03317, CBRZ03318, CBRZ03319, CBRZ03320, CBRZ03321, CBRZ03322, CBRZ03323, CBRZ03324, CBRZ03325, CBRZ03326 ,CBRZ03327, CBRZ03329, CBRZ03330, CBRZ03331, CBRZ03332, CBRZ03333, CBRZ03334, CBRZ03335, CBRZ03336, CBRZ03337, CBRZ03338, CBRZ03339, CBRZ03340, CBRZ03341, CBRZ03343, CBRZ03344, CBRZ03345, CBRZ03346, CBRZ03347, CBRZ03348, CBRZ03349, CBRZ03350, CBRZ03351, CBRZ03352, CBRZ03353, CBRZ03354, CBRZ03355, CBRZ03356, CBRZ03357, CBRZ03358, CBRZ03359, CBRZ03360, CBRZ03361, CBRZ03362, CBRZ03363, CBRZ03364, CBRZ03365, CBRZ03366, CBRZ03367, CBRZ03368, CBRZ03369, CBRZ03370, CBRZ03371, CBRZ03372, CBRZ03373, CBRZ03374, CBRZ03375, CBRZ03376, CBRZ03377, CBRZ03378, CBRZ03379, CBRZ03380, CBRZ03381, CBRZ03382, CBRZ03383, CBRZ03384, CBRZ03385, CBRZ03386, CBRZ03387, CBRZ03388, CBRZ03389, CBRZ03390, CBRZ03391, CBRZ03392, CBRZ03393, CBRZ03394, CBRZ03395, CBRZ03396, CBRZ03397, CBRZ03398, CBRZ03399, CBRZ03400, CBRZ03401, CBRZ03402, CBRZ03403, CBRZ03404, CBRZ03405, CBRZ03406, CBRZ03407, CBRZ03408, CBRZ03409, CBRZ03410, CBRZ03411, CBRZ03412, CBRZ03413, CBRZ03414, CBRZ03415, CBRZ03416, CBRZ03417, CBRZ03418, CBRZ03419, CBRZ03420, CBRZ03421, CBRZ03422, CBRZ03423, CBRZ03424, CBRZ03425, CBRZ03426, CBRZ03428, CBRZ03429, CBRZ03430, CBRZ03431, CBRZ03432, CBRZ03433, CBRZ03434, CBRZ03435, CBRZ03436, CBRZ03439, CBRZ03440, CBRZ03441, CBRZ03442, CBRZ03443, CBRZ03444, CBRZ03445, CBRZ03446, CBRZ03447, CBRZ03448, CBRZ03449, CBRZ03450, CBRZ03452, CBRZ03453, CBRZ03454, CBRZ03455, CBRZ03456, CBRZ03457, CBRZ03458, CBRZ03459, CBRZ03460, CBRZ03461, CBRZ03462, CBRZ03463, CBRZ03464, CBRZ03465, CBRZ03466, CBRZ03467, CBRZ03468 CBRZ03470, CBRZ03471, CBRZ03473, CBRZ03474, CBRZ03475, CBRZ03476, CBRZ03479, CBRZ03480, CBRZ03481, CBRZ03486, CBRZ03507, CBRZ03512, CBRZ03519, CBRZ03520, CBRZ03524, CBRZ03525, CBRZ03527, CBRZ03529, CBRZ03530, CBRZ03542, CBRZ03545, CBRZ03546, CBRZ03547, CBRZ03549, CBRZ03551, CBRZ03553, CBRZ03554, CBRZ03555, CBRZ03556, CBRZ03557, CBRZ03558, CBRZ03559, CBRZ03560, CBRZ03561, CBRZ03562, CBRZ03563, CBRZ03564, CBRZ03565, CBRZ03566, CBRZ03567, CBRZ03568, CBRZ03569, CBRZ03570, CBRZ03571, CBRZ03572, CBRZ03573, CBRZ03574, CBRZ03576, CBRZ03577, CBRZ03578, CBRZ03579, CBRZ03580, CBRZ03581, CBRZ03582, CBRZ03583, CBRZ03584, CBRZ03585, CBRZ03586, CBRZ03587, CBRZ03588, CBRZ03589, CBRZ03590, CBRZ03591, CBRZ03592, CBRZ03594, CBRZ03595, CBRZ03599, CBRZ03601, CBRZ03604, CBRZ03605, CBRZ03606, CBRZ03607, CBRZ03608, CBRZ03611, CBRZ03612, CBRZ03613, CBRZ03614, CBRZ03615, CBRZ03617, CBRZ03618, CBRZ03619, CBRZ03626, CBRZ03628, CBRZ03631, CBRZ03633, CBRZ03634, CBRZ03635, CBRZ03636, CBRZ03637, CBRZ03638, CBRZ03639, CBRZ03640, CBRZ03641, CBRZ03643, CBRZ03644, CBRZ03645, CBRZ03648, CBRZ03650, CBRZ03651, CBRZ03652, CBRZ03653, CBRZ03654, CBRZ03655, CBRZ03656, CBRZ03657, CBRZ03658, CBRZ03659, CBRZ03660, CBRZ03661, CBRZ03662, CBRZ03663, CBRZ03664, CBRZ03665, CBRZ03667, CBRZ03668, CBRZ03670, CBRZ03689, CBRZ03695, CBRZ03707, CBRZ03712, CBRZ03714, CBRZ03716, CBRZ03717, CBRZ03718, CBRZ03719, CBRZ03720, CBRZ03721, CBRZ03722, CBRZ03723, CBRZ03725, CBRZ03726, CBRZ03727, CBRZ03728, CBRZ03729, CBRZ03736, CBRZ03738, CBRZ03739, CBRZ03740, CBRZ03741, CBRZ03745, CBRZ03747, CBRZ03748, CBRZ03749, CBRZ03750, CBRZ03751, CBRZ03752, CBRZ03753, CBRZ03754, CBRZ03755, CBRZ03756, CBRZ03757, CBRZ03758, CBRZ03760, CBRZ03761, CBRZ03762, CBRZ03763, CBRZ03764, CBRZ03765, CBRZ03766, CBRZ03770, CBRZ03903, CBRZ03932, CBRZ03943, CBRZ03944, CBRZ03945, CBRZ03953, CBRZ03960, CBRZ03963, CBRZ03964, CBRZ03965, CBRZ03966, CBRZ03967, CBRZ03968, CBRZ03969, CBRZ03970, CBRZ03971, CBRZ03972, CBRZ03973, CBRZ03974, CBRZ03975, CBRZ03976, CBRZ03977, CBRZ03978, CBRZ03981, CBRZ03982, CBRZ03983, CBRZ03991, CBRZ03992, CBRZ03993, CBRZ03995, CBRZ03996, CBRZ03999, CBRZ04000, CBRZ04002, CBRZ04004, CBRZ04005, CBRZ04007, CBRZ04010, CBRZ04012, CBRZ04013, CBRZ04016, CBRZ04017, CBRZ04018, CBRZ04019, CBRZ04027, CBRZ04029, CBRZ04030, CBRZ04032, CBRZ04033, CBRZ04034, CBRZ04035, CBRZ04037, CBRZ04044, CBRZ04045, CBRZ04048, CBRZ04049, CBRZ04055, CBRZ04056, CBRZ04060, CBRZ04077, CBRZ04079, CBRZ04114, CBRZ04126, CBRZ04127, CBRZ04128, CBRZ04129, CBRZ04131, CBRZ04132, CBRZ04133, CBRZ04134, CBRZ04135, CBRZ04136, CBRZ04138, CBRZ04140, CBRZ04141, CBRZ04142, CBRZ04143, CBRZ04144, CBRZ04145, CBRZ04146, CBRZ04147, CBRZ04148, CBRZ04149, CBRZ04150, CBRZ04151,CBRZ04152, CBRZ04153, CBRZ04154, CBRZ04155, CBRZ04156, CBRZ04157, CBRZ04158, CBRZ04159, CBRZ04160, CBRZ04161, CBRZ04162, CBRZ04163, CBRZ04164, CBRZ04165, CBRZ04166, CBRZ04167, CBRZ04168, CBRZ04169, CBRZ04170, CBRZ04171, CBRZ04173, CBRZ04174, CBRZ04175, CBRZ04177, CBRZ04178, CBRZ04179, CBRZ04180, CBRZ04181, CBRZ04182, CBRZ04183, CBRZ04184, CBRZ04185, CBRZ04186, CBRZ04187, CBRZ04188, CBRZ04189, CBRZ04190, CBRZ04191, CBRZ04192, CBRZ04193, CBRZ04194, CBRZ04195, CBRZ04197, CBRZ04198, CBRZ04199, CBRZ04200, CBRZ04201, CBRZ04202, CBRZ04203, CBRZ04204, CBRZ04205, CBRZ04206, CBRZ04207, CBRZ04208, CBRZ04209, CBRZ04210, CBRZ04211, CBRZ04213, CBRZ04214, CBRZ04215, CBRZ04216, CBRZ04217, CBRZ04218, CBRZ04219, CBRZ04220, CBRZ04221, CBRZ04222, CBRZ04223, CBRZ04224, CBRZ04225, CBRZ04226, CBRZ04227, CBRZ04228, CBRZ04229, CBRZ04230, CBRZ04231, CBRZ04232, CBRZ04233, CBRZ04234, CBRZ04235, CBRZ04236, CBRZ04237, CBRZ04238, CBRZ04239, CBRZ04240, CBRZ04241, CBRZ04242, CBRZ04243, CBRZ04244, CBRZ04245, CBRZ04246, CBRZ04247, CBRZ04248, CBRZ04249, CBRZ04250, CBRZ04251, CBRZ04252, CBRZ04253, CBRZ04254, CBRZ04255, CBRZ04256, CBRZ04257, CBRZ04258, CBRZ04259, CBRZ04260, CBRZ04261, CBRZ04262, CBRZ04263, CBRZ04264, CBRZ04265, CBRZ04266, CBRZ04267, CBRZ04268, CBRZ04270, CBRZ04271, CBRZ04272, CBRZ04274, CBRZ04275, CBRZ04276, CBRZ04277, CBRZ04278, CBRZ04279, CBRZ04280, CBRZ04281, CBRZ04282, CBRZ04283, CBRZ04284, CBRZ04285, CBRZ04286, CBRZ04287, CBRZ04288, CBRZ04289, CBRZ04290, CBRZ04291, CBRZ04292, CBRZ04293, CBRZ04294, CBRZ04295, CBRZ04296, CBRZ04297, CBRZ04298, CBRZ04299, CBRZ04300, CBRZ04301, CBRZ04302, CBRZ04303, CBRZ04304, CBRZ04305, CBRZ04306, CBRZ04307, CBRZ04308, CBRZ04309, CBRZ04310, CBRZ04311, CBRZ04312, CBRZ04313, CBRZ04314, CBRZ04315, CBRZ04316, CBRZ04317, CBRZ04318, CBRZ04319, CBRZ04320, CBRZ04321, CBRZ04322, CBRZ04323, CBRZ04324, CBRZ04325, CBRZ04326, CBRZ04327, CBRZ04328, CBRZ04329, CBRZ04331, CBRZ04332, CBRZ04333, CBRZ04334, CBRZ04335, CBRZ04337, CBRZ04338, CBRZ04339, CBRZ04340, CBRZ04341, CBRZ04343, CBRZ04344, CBRZ04345, CBRZ04346, CBRZ04347, CBRZ04348, CBRZ04349, CBRZ04350, CBRZ04351, CBRZ04352, CBRZ04353, CBRZ04354, CBRZ04355, CBRZ04356, CBRZ04357, CBRZ04358, CBRZ04359, CBRZ04365, CBRZ04366, CBRZ04367, CBRZ04368, CBRZ04369, CBRZ04370, CBRZ04371, CBRZ04372, CBRZ04375, CBRZ04381, CBRZ04382, CBRZ04383, CBRZ04384, CBRZ04385, CBRZ04386, CBRZ04388, CBRZ04390, CBRZ04391, CBRZ04392, CBRZ04393, CBRZ04394, CBRZ04395, CBRZ04396, CBRZ04397, CBRZ04398, CBRZ04399, CBRZ04400, CBRZ04401, CBRZ04402, CBRZ04403, CBRZ04404, CBRZ04405, CBRZ04406, CBRZ04407, CBRZ04408, CBRZ04409, CBRZ04410, CBRZ04411, CBRZ04412, CBRZ04413, CBRZ04414, CBRZ04415, CBRZ04416, CBRZ04417, CBRZ04418, CBRZ04419, CBRZ04420, CBRZ04421, CBRZ04422, CBRZ04423, CBRZ04424, CBRZ04425, CBRZ04426, CBRZ04427, CBRZ04428, CBRZ04429, CBRZ04430, CBRZ04431, CBRZ04432, CBRZ04433, CBRZ04434, CBRZ04435, CBRZ04436, CBRZ04437, CBRZ04439, CBRZ04440, CBRZ04441, CBRZ04442, CBRZ04443, CBRZ04444, CBRZ04445, CBRZ04448, CBRZ04451, CBRZ04452, CBRZ04453, CBRZ04454, CBRZ04455, CBRZ04456, CBRZ04457, CBRZ04458, CBRZ04459, CBRZ04460, CBRZ04461, CBRZ04462, CBRZ04463, CBRZ04464, CBRZ04465, CBRZ04466, CBRZ04467, CBRZ04468, CBRZ04469, CBRZ04470, CBRZ04471, CBRZ04472, CBRZ04473, CBRZ04474, CBRZ04475, CBRZ04476, CBRZ04477, CBRZ04478, CBRZ04479, CBRZ04480, CBRZ04481, CBRZ04482, CBRZ04483, CBRZ04484, CBRZ04486, CBRZ04487, CBRZ04488, CBRZ04489, CBRZ04490, CBRZ04492, CBRZ04493, CBRZ04494, CBRZ04495, CBRZ04496, CBRZ04497, CBRZ04498, CBRZ04499, CBRZ04500, CBRZ04501, CBRZ04502, CBRZ04503, CBRZ04504, CBRZ04505, CBRZ04507, CBRZ04508, CBRZ04509, CBRZ04510, CBRZ04512, CBRZ04513, CBRZ04514, CBRZ04515, CBRZ04516, CBRZ04517, CBRZ04518, CBRZ04519, CBRZ04520, CBRZ04521, CBRZ04522, CBRZ04523, CBRZ04525, CBRZ04526, CBRZ04527, CBRZ04528, CBRZ04529, CBRZ04530, CBRZ04531, CBRZ04532, CBRZ04533, CBRZ04534, CBRZ04535, CBRZ04536, CBRZ04537, CBRZ04538, CBRZ04539, CBRZ04540, CBRZ04541, CBRZ04542, CBRZ04543, CBRZ04544, CBRZ04545, CBRZ04546, CBRZ04547, CBRZ04548, CBRZ04549, CBRZ04550, CBRZ04551, CBRZ04552, CBRZ04553, CBRZ04554, CBRZ04555, CBRZ04556, CBRZ04557, CBRZ04558, CBRZ04559, CBRZ04560, CBRZ04561, CBRZ04562, CBRZ04563, CBRZ04564, CBRZ04565, CBRZ04566, CBRZ04567, CBRZ04568, CBRZ04569, CBRZ04570, CBRZ04571, CBRZ04572, CBRZ04573, CBRZ04574, CBRZ04575, CBRZ04576, CBRZ04577, CBRZ04578, CBRZ04579, CBRZ04580, CBRZ04581, CBRZ04582, CBRZ04583, CBRZ04584, CBRZ04585, CBRZ04586, CBRZ04587, CBRZ04588, CBRZ04589, CBRZ04590, CBRZ04591, CBRZ04592, CBRZ04593, CBRZ04594, CBRZ04595, CBRZ04596, CBRZ04597, CBRZ04598, CBRZ04599, CBRZ04600, CBRZ04601, CBRZ04602, CBRZ04603, CBRZ04604, CBRZ04605, CBRZ04606, CBRZ04607, CBRZ04608, CBRZ04609, CBRZ04610, CBRZ04611, CBRZ04612, CBRZ04613, CBRZ04614, CBRZ04615, CBRZ04616, CBRZ04617, CBRZ04618, CBRZ04619, CBRZ04620, CBRZ04621, CBRZ04622, CBRZ04623, CBRZ04624, CBRZ04625, CBRZ04626, CBRZ04627, CBRZ04628, CBRZ04629, CBRZ04630, CBRZ04631, CBRZ04632, CBRZ04633, CBRZ04634, CBRZ04635, CBRZ04636, CBRZ04637, CBRZ04638, CBRZ04639, CBRZ04640, CBRZ04641, CBRZ04642, CBRZ04643, CBRZ04644, CBRZ04645, CBRZ04646, CBRZ04647, CBRZ04648, CBRZ04650, CBRZ04651, CBRZ04652, CBRZ04653, CBRZ04654, CBRZ04655, CBRZ04656, CBRZ04657, CBRZ04658, CBRZ04659, CBRZ04660, CBRZ04661, CBRZ04662, CBRZ04663, CBRZ04664, CBRZ04665, CBRZ04666, CBRZ04667, CBRZ04668, CBRZ04669, CBRZ04670, CBRZ04671, CBRZ04672, CBRZ04673, CBRZ04674, CBRZ04675, CBRZ04676, CBRZ04677, CBRZ04678, CBRZ04679, CBRZ04680, CBRZ04681, CBRZ04682, CBRZ04683, CBRZ04685, CBRZ04686, CBRZ04687, CBRZ04688, CBRZ04689, CBRZ04690, CBRZ04691, CBRZ04697, CBRZ04732, CBRZ04756, CBRZ04757, CBRZ04758, CBRZ04760, CBRZ04761, CBRZ04763, CBRZ04764, CBRZ04765, CBRZ04766, CBRZ04767, CBRZ04768, CBRZ04769, CBRZ04771, CBRZ04773, CBRZ04774, CBRZ04775, CBRZ04776, CBRZ04777, CBRZ04780, CBRZ04781, CBRZ04782, CBRZ04783, CBRZ04784, CBRZ04785, CBRZ04786, CBRZ04787, CBRZ04789, CBRZ04792, CBRZ04793, CBRZ04794, CBRZ04795, CBRZ04796, CBRZ04797, CBRZ04798, CBRZ04799, CBRZ04801, CBRZ04802, CBRZ04803, CBRZ04804, CBRZ04805, CBRZ04806, CBRZ04807, CBRZ04808, CBRZ04809, CBRZ04810, CBRZ04811, CBRZ04812, CBRZ04835, CBRZ04836, CBRZ04844, CBRZ04845, CBRZ04846, CBRZ04858, CBRZ04896, CBRZ04916, CBRZ04926, CBRZ04937, CBRZ04938, CBRZ04939, CBRZ04940, CBRZ04941, CBRZ04942, CBRZ04943, CBRZ04944, CBRZ04945, CBRZ04946, CBRZ04948, CBRZ04949, CBRZ04950, CBRZ04952, CBRZ04953, CBRZ05010, CBRZ05011,CBRZ05243, CBRZ05246, CBRZ05250, CBRZ05251, CBRZ05253, CBRZ05254, CBRZ05269, CBRZ05270, CBRZ05271, CBRZ05272, CBRZ05274, CBRZ05275, CBRZ05297, CBRZ05298, CBRZ05299, CBRZ05300, CBRZ05301, CBRZ05308, CBRZ05316, CBRZ05339, CBRZ40011, CBRZ40012, CBRZ40013, CBRZ40014, CBRZ40015, CBRZ40016, CBRZ40017, CBRZ40018, CBRZ40019, CBRZ40020, CBRZ40021, CBRZ40022, CBRZ40023, CBRZ40024, CBRZ40025, CBRZ40026, CBRZ40027, CBRZ40028, CBRZ40029, CBRZ40030, CBRZ40031, CBRZ40032, CBRZ40033, CBRZ40034, CBRZ40035, CBRZ40036, CBRZ40037, CBRZ03627, CBRZ03632, CBRZ03642, CBRZ03491, CBRZ03818, CBRZ03840, CBRZ03844, CBRZ03846, CBRZ03847, CBRZ03478, CBRZ03509, CBRZ03511, CBRZ03521, CBRZ03528, CBRZ03172, CBRZ03183, CBRZ03189, CBRZ03190, CBRZ03192, CBRZ03477, CBRZ03500, CBRZ03472, CBRZ03489, CBRZ03493, CBRZ03312, CBRZ03313, CBRZ03314, CBRZ03316, CBRZ03328, 70187546-2, 70187546-3, 70187742-5, 70187742-6, 70187742-7, 70187742-8, 70187742-9, 70189854-4, 70189854-5, CBRZ03181, CBRZ03184, CBRZ03186, CBRZ03191, CBRZ03194, CBRZ03933, CBRZ03986, CBRZ03988, CBRZ03947, CBRZ03950, CBRZ03980, CBRZ03987, CBRZ03622, CBRZ03623, CBRZ03624, CBRZ03629, CBRZ03630, CBRZ03427, CBRZ03482, CBRZ03483, CBRZ03492, CBRZ03494, CBRZ03742, CBRZ03705, CBRZ03708, CBRZ03710, CBRZ03715, CBRZ03073, CBRZ03074, CBRZ03076, CBRZ03080, CBRZ03082. 
Recalling Firm/
Manufacturer
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information ContactGE Healthcare Service
800-437-1171
Manufacturer Reason
for Recall
The oxygen sensor for certain CARESCAPE R860 devices has a potential issue that may result in an inaccurate display of FiO2 value from what is being delivered.
FDA Determined
Cause 2
Process control
ActionOn July 14th, 2020 the firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated 7/14/2020 to its customers with the following instructions: Safety Instructions You may continue to use the CARESCAPE R860 Ventilator. Ensure that the FiO2 alarm limit is set to the factory default limit +/- 6 % so that the clinician is aware if the measured value differs from set value. When making an extubation decision, use the set FiO2 value instead of the FiO2 measured by the oxygen sensor. The user can use other methods to measure FiO2 or switch to another device and call GE Healthcare. Product Correction: GE Healthcare will correct all affected units at no cost to you. A GE Healthcare representative will contact you to arrange for the correction. Contact Information: If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in Commerce2277 units
DistributionDomestic Distribution: USA International Distribution: Australia, Bahrain, Bermuda, Canada, Egypt, France, Germany, Honduras, India, Iraq, Italy, Lebanon, Mexico, Panama, Peru, Russia, Saudi Arabia, Singapore, South Africa, Spain, Taiwan and United Arab Emirates.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = CBK
-
-