  Class 2 Device Recall cobas 8000 Modular CORE System see related information
Date Initiated by Firm December 08, 2020
Create Date December 30, 2020
Recall Status1 Open3, Classified
Recall Number Z-0754-2021
Recall Event ID 86981
510(K)Number K100853  
Product Classification Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
Product Roche cobas 8000 modular analyzer series (i.e., cobas 8000 ISE module; cobas c 502, 701, and 702 modules; and cobas e 602 and 801 modules). Chemistry Analyzer (photometric discrete), for clinical use
Part Number: 08763662001
Code Information Serial No. 1017-01 1017-04 1019-09 1019-09 1019-09 1020-01 1022-05 1022-05 1029-09 1029-09 1030-08 1030-09 1030-09 1030-10 1034-09 1036-07 1036-07 1138-03 1138-03 1138-04 1140-09 1140-10 1142-01 1142-02 1142-02 1142-04 1142-04 1142-05 1142-06 1142-08 1142-08 1142-08 1142-09 1142-09 1142-09 1142-10 1143-01 1143-06 1143-09 1143-09 1143-10 1144-01 1146-02 1147-01 1147-03 1147-04 1147-07 1147-09 1149-01 1150-04 1150-05 1150-06 1150-06 1150-06 1150-06 1150-06 1161-05 1161-06 1161-08 1161-08 1161-08 1161-08 1161-08 1163-07 1163-07 1163-07 1163-07 1163-07 1163-10 1163-10 1164-01 1167-06 1167-06 1167-06 1167-07 1167-07 1167-08 1167-08 1167-09 1167-09 1167-10 1167-10 1168-02 1168-02 1168-02 1168-03 1168-03 1168-03 1168-03 1168-03 1168-03 1168-03 1168-03 1168-04 1168-04 1171-01 1171-01 1171-01 1171-02 1171-05 1206-02 1206-03 1206-04 1206-05 1206-05 1206-06 1206-06 1206-06 1206-06 1206-07 1207-07 1207-08 1207-08 1207-08 1207-09 1207-09 1214-17 1214-18 1214-19 1214-19 1214-19 1214-19 1214-19 1214-20 1214-20 1214-20 1215-01 1215-01 1215-01 1215-17 1215-17 1215-17 1215-18 1215-20 1215-20 1215-20 1215-20 1216-18 1216-18 1325-01 1325-02 1325-03 1325-13 1325-13 1325-13 1325-14 1325-14 1325-14 1325-14 1325-14 1325-15 1325-16 1326-01 1326-01 1326-01 1326-02 1326-02 1326-02 1326-03 1326-03 1326-03 1326-04 1326-13 1326-14 1326-15 1326-16 1326-17 1326-17 1326-17 1326-18 1326-19 1326-19 1326-19 1326-20 1332-01 1332-01 1332-02 1332-02 1332-03 1332-06 1332-13 1332-14 1332-15 1332-16 1343-10 1344-01 1344-02 1344-03 1344-04 1344-05 1344-05 1344-05 1344-05 1344-06 1344-07 1344-08 1344-09 1344-09 1344-09 1344-10 1344-10 1344-10 1345-01 1345-03 1345-04 1345-05 1345-05 1350-06 1350-06 1350-07 1350-08 1351-02 1353-01 1353-02 1353-03 1353-03 1353-04 1353-04 1360-02 1360-02 1360-02 1360-03 1360-03 1360-03 1360-04 1361-05 1361-06 1361-06 1361-06 1361-09 1361-09 1361-10 1361-10 1362-01 1362-02 1362-03 1362-04 1362-04 1362-04 1362-04 1362-04 1362-05 1362-05 1364-02 1364-03 1364-04 1364-05 1364-06 1364-07 1364-08 1364-08 1366-01 1366-01 1366-01 1366-02 1366-02 1366-02 1366-04 1366-05 1366-06 1366-07 1366-08 1372-01 1372-01 1372-02 1372-06 1475-08 1475-08 1475-09 1475-10 1475-10 1485-07 1485-08 1485-09 1487-06 1487-07 1487-07 1488-02 1488-07 1492-06 1492-06 1492-06 1492-06 1492-06 1492-07 1492-08 1492-08 1492-08 1492-08 1492-08 1497-01 1497-02 1497-03 1497-03 1497-04 1497-04 1497-05 1497-06 1497-07 1497-08 1497-08 1497-08 1497-08 1497-08 1497-08 1497-08 1498-01 14A1-10 14A2-05 14A2-05 14A2-06 14A2-07 14A2-07 14A2-08 14C2-03 14C4-06 14C6-01 14C6-02 14C6-02 14C6-03 14C9-06 14C9-09 14C9-10 14C9-10 14D1-09 14D1-10 14D2-01 14D2-02 14D2-03 14D3-07 14D3-08 14D4-02 14D4-09 14D4-10 14D5-01 14D5-02 14D5-10 14D6-01 14D6-02 14D6-03 15F3-04 15F3-04 15F3-05 15F3-05 15F3-06 15F3-07 15F3-07 15F3-08 15F3-08 15F3-09 15F3-09 15G6-09 15G6-09 15G6-10 15G6-10 15G7-01 15H4-01 15H4-01 15H4-02 15H4-02 15H4-03 15H4-05 15H4-06 15H4-06 15H4-08 15H4-08 15H4-09 15H4-10 15H5-01 15H5-02 15H5-04 15H5-05 15H5-06 15H6-06 15K9-08 15K9-09 15L0-02 15L9-09 15M6-07 15M6-09 15M6-10 15M7-01 15M7-02 15M7-03 15M7-03 15M7-04 15M7-04 15M7-05 15M7-05 15M7-06 15M7-06 15M9-03 15M9-04 15M9-07 15P0-02 15P2-08 15P5-03 15P5-03 15P5-04 15P5-04 15P6-03 15P6-03 15P8-09 15P8-10 15P9-01 15P9-01 15P9-02 15P9-02 15P9-03 15P9-05 15P9-06 15P9-07 15P9-07 15P9-08 15R8-01 15R8-01 15R8-04 15R8-04 15R8-05 15R8-06 15R8-07 15R8-10 15R8-10 15R9-01 15R9-02 15R9-03 15R9-04 15R9-05 15R9-06 15S3-03 15S3-04 15S3-08 15S3-10 1638-09 1639-01 1639-02 1639-03 1639-06 1639-07 1639-08 1639-09 1639-10 1640-01 1645-03 1647-09 1647-10 1647-10 1648-01 1648-01 1648-02 1648-03 1648-04 1648-05 1648-06 1648-07 1655-06 1655-07 1656-01 1656-02 16S8-03 16S8-04 16S8-05 16S8-07 16S8-08 16S8-09 16S9-01 16S9-02 16S9-02 16S9-03 16S9-03 16S9-04 16S9-05 16S9-08 16S9-08 16U1-05 16U1-09 16U1-10 16U2-01 16U2-02 16U2-03 16U2-04 16U2-09 16U2-10 16U5-02 16U5-03 16U5-04 16U5-10 16U6-01 16U6-01 16U6-02 16U6-02 16U6-03 16U6-05 16U6-06 16U6-07 16V9-03 16W0-06 16W0-07 16W0-08 16W0-09 16W4-07 16X4-07 16X4-08 16X4-09 16X4-10 16X5-01 16X5-02 16X5-03 16X5-04 16X5-05 16X5-06 16X5-07 16X5-08 16X5-09 16Y3-04 16Y3-05 16Y3-06 16Y3-07 16Y3-09 16Y3-10 16Y4-01 16Y4-01 16Y4-02 16Y4-02 16Y4-03 16Y4-03 16Y5-09 16Y5-10 16Y6-01 16Y6-02 16Y6-04 16Y6-05 16Y6-06 16Y6-10 16Y8-04 16Y8-05 16Y8-09 16Y8-09 16Y8-10 16Y9-01 16Y9-07 16Z2-01 16Z2-02 16Z2-03 16Z2-05 16Z2-06 16Z2-07 16Z2-08 16Z2-09 16Z8-05 1775-09 1779-02 1779-02 1779-03 1779-04 1779-05 1779-06 1779-07 1779-08 1795-01 1795-02 1795-03 1795-04 1795-05 1795-06 1795-07 1795-08 1795-10 1796-01 17A8-07 17A8-08 17A9-02 17A9-02 17B5-02 17B5-03 17B5-05 17B5-05 17B5-06 17B5-09 17B6-02 17D4-05 17D5-01 17D5-02 17D5-04 17D5-04 17D5-04 17D5-05 17D5-05 17D6-03 17D6-03 17D6-04 17D6-05 17D6-06 17D6-07 17D6-08 17D6-09 17E1-05 17E1-06 17E1-08 17E4-04 17E4-04 17E4-05 17E4-06 17E4-07 17E4-08 17E4-09 17E4-10 17F6-09 17F6-09 17F6-10 17F7-01 17F7-02 17F7-03 17G1-05 17G1-07 17G3-05 17G3-06 17G3-07 17G3-09 17G3-09 17G5-08 17G5-09 17G5-10 17G6-01 17G6-02 17G6-03 17H0-10 17H1-01 1814-06 1814-07 1814-08 1815-02 1815-04 1820-10 1820-10 1821-01 1821-06 1822-05 1822-08 1822-10 1823-02 1823-03 1829-05 1834-07 1835-03 18H5-03 18H5-03 18H5-03 18H5-04 18H5-05 18H5-06 18H5-07 18H5-08 18J6-03 18J6-04 18J6-05 18J6-06 18J6-07 18J6-08 18K0-04 18K0-05 18K0-06 18K3-01 18K3-02 18K3-03 18L0-09 18L0-10 18L1-01 18L1-02 18L1-03 18L1-04 18L1-05 18L1-06 18L1-08 18L1-09 18L1-10 18L2-02 18L2-03 18L2-04 18L5-04 18L5-05 18L5-06 18L5-06 18L5-07 18L5-08 18L5-09 18L5-10 18L6-01 18L6-02 18L6-02 18L6-03 18L6-04 18L6-04 18L8-09 18L8-10 18L9-04 18M3-03 18M3-04 18M3-05 18M3-06 18M3-07 18M3-08 18M3-09 18M3-10 18M4-01 18M4-02 18M4-03 18M4-04 18M4-05 18M4-06 18M5-02 18M5-03 18M5-05 18M6-09 18M6-10 18M7-01 18M7-02 18M7-03 18M7-05 18M7-06 18M7-07 18M7-08 18M7-09 18M8-02 18M8-05 18M8-06 18M8-07 18M8-09 18M9-04 18M9-04 18M9-05 18P5-03 18P5-04 18P5-05 18P5-07 18P5-08 18P5-09 18P5-10 18P6-01 18P6-02 18P6-04 18P6-05 18P6-06 18P6-07 18P6-08 18P6-09 18P6-10 18P7-01 18P7-02 18P7-09 18P8-02 18P8-04 18P8-05 18P8-06 18P8-07 18P8-08 18P8-09 18P8-10 18P9-01 18P9-02 18P9-03 18P9-04 18R2-04 18R2-05 18R2-07 18R5-09 18R5-10 18R6-01 18R6-02 18R6-03 18R6-04 18S3-07 18S3-09 18S3-10 18S4-01 18S4-02 18S4-03 18S4-04 18U8-05 18U8-06 18U8-07 18U8-08 18U8-09 18U8-10 18U9-02 18U9-03 18U9-04 18U9-05 18U9-06 18U9-08 18U9-09 18U9-10 18V0-06 18V0-07 18V0-08 18V0-09 18V0-10 18V1-01 18V1-02 18V1-03 18V1-04 18V1-06 18V3-02 18V3-03 18V3-04 18W1-07 18W1-08 18W1-09 18W1-10 18W2-01 18W2-05 18W2-06 18W2-07 18W3-05 18W3-06 18W3-07 18W3-08 18W3-09 18W3-10 18W7-01 18W7-02 18W7-04 18W7-05 18W7-06 18X0-06 18X0-07 18X1-03 18X1-04 18X7-04 18X7-05 18X7-06 18X7-07 18X7-08 18X7-09 18X7-10 18X8-01 18X8-02 18X8-03 18X8-04 18X8-05 18X8-06 18Y5-08 18Y5-09 19A3-09 19A3-10 19A4-01 19A4-02 19A4-03 19A4-04 19B2-05 19B2-08 19B2-09 19B9-07 19B9-08 19B9-09 19B9-10 19C0-01 19C0-02 19C0-03 19C0-04 19C0-05 19C0-06 19C0-07 19C0-08 19C0-09 19C6-01 19C6-02 19C6-03 19C6-03 19C6-04 19C6-05 19C6-06 19C6-07 19D1-02 19D1-03 19D5-03 19D5-04 19D5-05 19D5-06 19D5-07 19D5-08 19D5-09 19D5-10 19D6-01 19D6-02 19D6-03 19D6-04 19D6-05 19D6-06 19D6-07 19D6-08 19E1-01 19E1-02 19E1-03 19E1-04 19E1-05 19E1-06 19E1-07 19E1-08 19E6-01 19E6-02 19E6-03 19E6-04 19E6-05 19E6-06 19E6-07 19E6-08 19E6-08 19E6-09 19E6-10 19E7-01 19E7-02 19E7-03 19E7-04 19E7-05 19E7-06 19F0-10 19F1-02 19F1-03 19F1-04 19F1-05 19F1-07 19F1-08 19F5-09 19F5-10 19F6-01 19F6-02 19F6-03 19F6-04 19F6-09 19F6-10 19F7-01 19F7-02 19F7-03 19F7-04 19G1-05 19G1-07 19G1-09 19G1-10 19G2-01 19G2-02 19G2-03 19J0-09 19J0-10 20J1-02 20J1-03 20J1-04 20J5-02 20J5-03 20J5-04 20J5-05 20J9-07 20J9-08 20J9-09 20J9-10 20K0-01 20K0-02 20K0-04 20K2-10 20K3-01 20K3-02 20K3-03 20K3-04 20K3-05 20K3-06 20K3-07 20K3-08 20K3-09 20K3-10 20K4-01 20K4-02 20K4-03 20K4-04 20K4-05 20K4-06 20K4-07 20K4-08 20K4-09 20K5-01 20K5-02 20K5-03 20K5-04 20K9-06 20K9-07 20K9-08 20K9-09 20K9-10 20L0-02 20L0-03 20L4-06 20L4-07 20L4-08 20L4-09 20L5-01 20L5-02 20L5-05 20M0-04 20M0-05 20M0-06 20M0-06 20M0-06 20M0-06 20M0-07 20M0-08 20M0-09 20M0-10 20M1-01 20M5-09 20M5-10 20M6-01 20M6-02 20M6-09 20M6-10 20P1-09 20P1-10 20P2-01 20P2-02 20P2-03 20P2-05 20P2-06 20P2-07 20P2-08 20P3-03 20P3-04 20P7-02 20P7-03 20P7-04 20P7-05 20P7-06 20P7-07 20P7-08 20R1-06 20R1-07 20R1-08 20R1-09 20R2-09 20R7-03 20S1-03 20S1-04 20S1-05 20S3-06 20S3-06 20S3-06 20S3-06 20S3-07 20S3-08 20S3-09 20S7-02 20S7-03 20S7-04 
Recalling Firm/
Manufacturer
Roche Diagnostics Operations, Inc.
9115 Hague Rd
Indianapolis IN 46256-1025
For Additional Information Contact Roche Customer Support Center
317-521-4343
Manufacturer Reason
for Recall
Potential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may remain undetected due to the absence of associated data flags.
FDA Determined
Cause 2
Software design
Action Roche recall issued Urgent Medical Device Correction (UMDC TP-01165 and UMDC TP-01169) notifications on 12/8/2020 via UPS Ground (receipt required). Both UMDCs will also be posted to the diagnostics.roche.com website. Letter states reason for recall, health risk and action to take: Consignee Instructions: ¿ Inform any operators who use the cobas 8000 modular analyzer series or the cobas pro integrated solutions of the potential hazards associated with this issue and provide a copy of the UMDC notification, as appropriate. ¿ Follow the actions outlined in the Customer Actions to Determine if the Issue Has Occurred section of the UMDC. ¿ Confirm that you have backed up your most recent settings. Repeat this maintenance function whenever a system setting is changed. ¿ Consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. ¿ Complete the enclosed faxback form and return it following the instructions on the form. ¿ File this UMDC for future reference. Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Quantity in Commerce 1050 units
Distribution US Nationwide distribution.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.