| | Class 2 Device Recall Philips Respironics V60 Ventilator |  |
| Date Initiated by Firm | November 02, 2020 |
| Date Posted | January 22, 2021 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0911-2021 |
| Recall Event ID |
86697 |
| 510(K)Number | K082660 |
| Product Classification |
Ventilator, continuous, minimal ventilatory support, facility use - Product Code MNT
|
| Product | Philips Respironics V60 Ventilator Part Number 1076716 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories. |
| Code Information |
Model: V60 Part Number: 1076716 V60 Ventilator, CHINA OPT: CFLEX, AVAPS Serial Numbers: 100193935 100194169 100030124 100030335 100030359 100030360 100030761 100030762 100030763 100030764 100030766 100030767 100030768 100030769 100030770 100030771 100030773 100030775 100030776 100030777 100030779 100030782 100030783 100030784 100032461 100032686 100032687 100032688 100032689 100032690 100032691 100032692 100032695 100032697 100032698 100032701 100032702 100035773 100035778 100035779 100035896 100035897 100035898 100035899 100035900 100035901 100035902 100035903 100035904 100035907 100035909 100035910 100035911 100035913 100035914 100035918 100035919 100035926 100036047 100036302 100036306 100036307 100036310 100036311 100036313 100036314 100036316 100036317 100036320 100036432 100036433 100036434 100036439 100037365 100037366 100037367 100037368 100037369 100037370 100037371 100037372 100037373 100037374 100037375 100037376 100037377 100037378 100037379 100037380 100037382 100037383 100037384 100037385 100037386 100037387 100037388 100037389 100037390 100037391 100037392 100037393 100037394 100037395 100037541 100037542 100037548 100037551 100037559 100037561 100037562 100037563 100037564 100039784 100039789 100039973 100040077 100040088 100043091 100043304 100043306 100043307 100043308 100043309 100043310 100043311 100043312 100043318 100043320 100043321 100043941 100043942 100043943 100043946 100043947 100043950 100043953 100046144 100046153 100046155 100046158 100046160 100046161 100046162 100046163 100046546 100046745 100047058 100047068 100049056 100049057 100049058 100049059 100049060 100049061 100049062 100049063 100049064 100049065 100049068 100049069 100049070 100049072 100049076 100049077 100050076 100051086 100051087 100051090 100051091 100051092 100051093 100051094 100051095 100051096 100051097 100051098 100051320 100051321 100051322 100051323 100051324 100051325 100051326 100051327 100051328 100051329 100051330 100051332 100051333 100051334 100051335 100051336 100051337 100051338 100051339 100051340 100051341 100051342 100051343 100051344 100051345 100051346 100051347 100051348 100051349 100051965 100052401 100052403 100052407 100052411 100052413 100052414 100052415 100052416 100052417 100052524 100052525 100052526 100052527 100052528 100052529 100052530 100052531 100052532 100052533 100052534 100052535 100052536 100052537 100052538 100052747 100052841 100052843 100052853 100052942 100052943 100052948 100052952 100052953 100052954 100052955 100052956 100052957 100052958 100052959 100052960 100052961 100052962 100052963 100052964 100052965 100052966 100053047 100053048 100053049 100053051 100053052 100053053 100053054 100053055 100053056 100053057 100053058 100053059 100053060 100053061 100053062 100053063 100053065 100053066 100053067 100053068 100053069 100053071 100053072 100053073 100053074 100053075 100053581 100053582 100053583 100053584 100053585 100053587 100053588 100053589 100053590 100053591 100053592 100053593 100053595 100053597 100053599 100053601 100053678 100053681 100053683 100053687 100053689 100053690 100053691 100053693 100053695 100053778 100053787 100053789 100053790 100053791 100053792 100055717 100055808 100055809 100055810 100055811 100055812 100055813 100055814 100055816 100055817 100055818 100055819 100055820 100055821 100055822 100057204 100057207 100057208 100057209 100057210 100057211 100057212 100057215 100057352 100057379 100058322 100058323 100058324 100058325 100058326 100058327 100058328 100058329 100058330 100058331 100058332 100058333 100058334 100058335 100058336 100058337 100058338 100058339 100058340 100058341 100058473 100058474 100058475 100058476 100058477 100058478 100058479 100058480 100058481 100058482 100058483 100058484 100058485 100058486 100058488 100058492 100058558 100058560 100058564 100058567 100058569 100058570 100058571 100058833 100058834 100058836 100058838 100058841 100058912 100059167 100059170 100059171 100059247 100059250 100060392 100060393 100060405 100060415 100060585 100060591 100060593 100060594 100060597 100060599 100060600 100061623 100061624 100061628 100061630 100061631 100061632 100061633 100061634 100061643 100061644 100061647 100061648 100063165 100063172 100063173 100063174 100063175 100063176 100063178 100063179 100063271 100063272 100063273 100063274 100063275 100063276 100063277 100063278 100063279 100063280 100063281 100063282 100063283 100063284 100063285 100063286 100063287 100063288 100063290 100063291 100063292 100063294 100063297 100064513 100064514 100064515 100064516 100064517 100064518 100064519 100064520 100064521 100064522 100064523 100064534 100064564 100064569 100064573 100064668 100064671 100064672 100064673 100064674 100064675 100064676 100064677 100064698 100064699 100064703 100064720 100064721 100064722 100064831 100064832 100064833 100064835 100064836 100065341 100065497 100065499 100065508 100065509 100065511 100065515 100065591 100065592 100065598 100065600 100065602 100065605 100067434 100067452 100067453 100067534 100068186 100068189 100068195 100068196 100068206 100068209 100068365 100068366 100068368 100068369 100068374 100068378 100069509 100069510 100069511 100069516 100069517 100069519 100069522 100069611 100069699 100069745 100069865 100069957 100069959 100069960 100069961 100069962 100069963 100069964 100069966 100069967 100069969 100069986 100069987 100069989 100069990 100069991 100069992 100069993 100069994 100069995 100070235 100070237 100070238 100070239 100070241 100070242 100070244 100070245 100070246 100070247 100070296 100070297 100070298 100070299 100070300 100070302 100070303 100070304 100070564 100070565 100070566 100070568 100070571 100071390 100071516 100071521 100071596 100071864 100071868 100071871 100071872 100071873 100071878 100071921 100071922 100071923 100071924 100071925 100071926 100071927 100071928 100071929 100071930 100071960 100071961 100071962 100071963 100071964 100071965 100071966 100071967 100071968 100071969 100071970 100071971 100071972 100071973 100071974 100071975 100071976 100071977 100072013 100072014 100072015 100072016 100072017 100072018 100072019 100072020 100072021 100072022 100072023 100072024 100072025 100072026 100072027 100072028 100072029 100072031 100072032 100072047 100072048 100072049 100072050 100072051 100072052 100072053 100072054 100072055 100072056 100072057 100072058 100072059 100072060 100072061 100072062 100072063 100072064 100072113 100072114 100072116 100072117 100072118 100072119 100072121 100072122 100072123 100072124 100072125 100072126 100072128 100072129 100072130 100072296 100072407 100072500 100072567 100072568 100072569 100073258 100073260 100075096 100075366 100075374 100075376 100075393 100075394 100075397 100075399 100075400 100075401 100075403 100075404 100075405 100075406 100075410 100078371 100078382 100078383 100078384 100078386 100078690 100078695 100078697 100078698 100078700 100078701 100078702 100078703 100078704 100078705 100078706 100078707 100078708 100078709 100078710 100078711 100078712 100078713 100078714 100078715 100078716 100078717 100078718 100078719 100078720 100078721 100078816 100078817 100078818 100078819 100078840 100078841 100078842 100078843 100078844 100078845 100078846 100078847 100078848 100078849 100078851 100078852 100078853 100078854 100078855 100078856 100078857 100078858 100078859 100078860 100078861 100078862 100078863 100078864 100078865 100078866 100078867 100078868 100078869 100078870 100078963 100078964 100078965 100078966 100079371 100079372 100079373 100079374 100079375 100079376 100079377 100079378 100079379 100079380 100079381 100079382 100079383 100079384 100079385 100079386 100079388 100079389 100079559 100079560 100079561 100079562 100079563 100079564 100079565 100079567 100079568 100079569 100079570 100079571 100079572 100079573 100079574 100079575 100079576 100079577 100079578 100079617 100079618 100079619 100079701 100079702 100079703 100079704 100079705 100079706 100079707 100079708 100079709 100079712 100079714 100079715 100079716 100079717 100079718 100079719 100079720 100079721 100079723 100079724 100080281 100080420 100080551 100080554 100080555 100080556 100080557 100080558 100080562 100080564 100080565 100080566 100080702 100081711 100081713 100081746 100081747 100081753 100082466 100083145 100083146 100083147 100083148 100083149 100083150 100083151 100083152 100083155 100083157 100083159 100083160 100083161 100083162 100083212 100083214 100083217 100083218 100083221 100083229 100083230 100083231 100083562 100083568 100083572 100083575 100083578 100085085 100085099 100085340 100085389 100085399 100085467 100085468 100085469 100085470 100085471 100085472 100085473 100085475 100085476 100085483 100085967 100085968 100085969 100085970 100085971 100085972 100085973 100085974 100085975 100085976 100085977 100085978 100085979 100085980 100085981 100085982 100085983 100085984 100085985 100085986 100085987 100085988 100085989 100085990 100085991 100085992 100085993 100085994 100085995 100085996 100086049 100086050 100086051 100086052 100086053 100086054 100086055 100086056 100086057 100086058 100086059 100086060 100086061 100086062 100086063 100086064 100086065 100086066 100086067 100086068 100086233 100086234 100086235 100086238 100086239 100086441 100086442 100086443 100086444 100086445 100086446 100086447 100086448 100086449 100086450 100086451 100086452 100086453 100086455 100086456 100086457 100086458 100086459 100086460 100086774 100086775 100086776 100086777 100086778 100086779 100086780 100086781 100086782 100086785 100086786 100086787 100086788 100086789 100086791 100086792 100086820 100086821 100086822 100086823 100086824 100086825 100086826 100086828 100086829 100086830 100086832 100086884 100086885 100086888 100086890 100086893 100086897 100086898 100086900 100086901 100087012 100087016 100087017 100087019 100087020 100087021 100087023 100087027 100087029 100087053 100087054 100087057 100087089 100087090 100087091 100087092 100087096 100087098 100087099 100087100 100087101 100087102 100087106 100087107 100087108 100087109 100087161 100087163 100087164 100087165 100087166 100087167 100087168 100087169 100087170 100087171 100087173 100088053 100089265 100089273 100089648 100089651 100089805 100089820 100089822 100089823 100089825 100089826 100089827 100089828 100089829 100089830 100089831 100089893 100089894 100089895 100089896 100089897 100089898 100089899 100089900 100089901 100089902 100089904 100089905 100089906 100089907 100089908 100089909 100089910 100089911 100089912 100090127 100090128 100090129 100090130 100090131 100090132 100090133 100090134 100090136 100090137 100090138 100090139 100090140 100090141 100090532 100091382 100092645 100092647 100092648 100092649 100092871 100092885 100092886 100092888 100092921 100092922 100092923 100092924 100092925 100092970 100092971 100092972 100092973 100092974 100092980 100092981 100092982 100092983 100092984 100092985 100092986 100092987 100092989 100092990 100092992 100092993 100093325 100093326 100093327 100093328 100093329 100093332 100093333 100093334 100093335 100093336 100093340 100093341 100093342 100093343 100093574 100093575 100093576 100093577 100093578 100093579 100093580 100093581 100093582 100093583 100093584 100093585 100093586 100093587 100093588 100093589 100093590 100093591 100093592 100093593 100093594 100093595 100093596 100093597 100093598 100094070 100094071 100094072 100094073 100094074 100094075 100094076 100094077 100094078 100094079 100094080 100094081 100094082 100094084 100094085 100094086 100094087 100094088 100094089 100094096 100094097 100094098 100094099 100094127 100094128 100094129 100094130 100094131 100094132 100094133 100094134 100094135 100094136 100094139 100094140 100094141 100094142 100094143 100094144 100094145 100094148 100094149 100094150 100094151 100094152 100094153 100094154 100094155 100094230 100094231 100094232 100094233 100094234 100094235 100094236 100094237 100094238 100094239 100094240 100094241 100094242 100094243 100094244 100094245 100094320 100094321 100094322 100094323 100094350 100094353 100094426 100096915 100096916 100096917 100096919 100096920 100096921 100096922 100096923 100096924 100096925 100096926 100096929 100096930 100097026 100097027 100097030 100097040 100097042 100097089 100097090 100097091 100097092 100097093 100097094 100097095 100097096 100097097 100097133 100097135 100097136 100097137 100097138 100097273 100097274 100097275 100097277 100097278 100097279 100097280 100097281 100097282 100097382 100097384 100097385 100097386 100097388 100097389 100097390 100097391 100097396 100097982 100097983 100097985 100097987 100097996 100097998 100097999 100098002 100098077 100098078 100098079 100098093 100098435 100098437 100098438 100098460 100098461 100098462 100098464 100098465 100098468 100098471 100098476 100098478 100098500 100098501 100098999 100099020 100099021 100099022 100099023 100099024 100099025 100099026 100099027 100099028 100099029 100099030 100099032 100099033 100099034 100099035 100099036 100099037 100099038 100099039 100099040 100099041 100099042 100099043 100099135 100099136 100099137 100099139 100099140 100099141 100099143 100099144 100099145 100099146 100099148 100099149 100099151 100099158 100099159 100099265 100099266 100099267 100099268 100099269 100099270 100099271 100099272 100099273 100099274 100099275 100099276 100099277 100099278 100099279 100099280 100099281 100099282 100099283 100099284 100099285 100099286 100099291 100099292 100099293 100099440 100099441 100099442 100099443 100099444 100099445 100099446 100099447 100099448 100099449 100099450 100099451 100099518 100099519 100099521 100099524 100099525 100099527 100099530 100099533 100099535 100099536 100100133 100100134 100100136 100100137 100100138 100100167 100100170 100100176 100100182 100100184 100100185 100100186 100100188 100100189 100100281 100100282 100100285 100100286 100100288 100100289 100100290 100100291 100100292 100100308 100100311 100101942 100101950 100101951 100101952 100101954 100101955 100101956 100101957 100101959 100101960 100101961 100101962 100101963 100102035 100102037 100102038 100102044 100102048 100102143 100102144 100102146 100102147 100102148 100102149 100102151 100102153 100102154 100102156 100102157 100102158 100102159 100102160 100102161 100102162 100102163 100103673 100104051 100104075 100104629 100104630 100104631 100104632 100104635 100104639 100104640 100104642 100104643 100104644 100104647 100104649 100104650 100104651 100104652 100105068 100105074 100105077 100105247 100105360 100105361 100105362 100105363 100105364 100105365 100105366 100105367 100105549 100105552 100105555 100106415 100107007 100107008 100107009 100107011 100107013 100107016 100107017 100107018 100107020 100107021 100107029 100107031 100107032 100107033 100107034 100107127 100107129 100107130 100107131 100107132 100107134 100107135 100107136 100107137 100107138 100107140 100107142 100107144 100107296 100107298 100107299 100107301 100107303 100107306 100107307 100107311 100107315 100107316 100107317 100107439 100107440 100107441 100107444 100107445 100107446 100107447 100107448 100107449 100107450 100107451 100107453 100107454 100107455 100107456 100109493 100109496 100109500 100109502 100109505 100109506 100109507 100109508 100109509 100109512 100109513 100109514 100109515 100109516 100109517 100109519 100114469 100114475 100114476 100114477 100114478 100114479 100114766 100114997 100114998 100114999 100115201 100115203 100115204 100115205 100115206 100115209 100115212 100115213 100115217 100115218 100115219 100115260 100115261 100115262 100115263 100115264 100115265 100115267 100115268 100115269 100115270 100115271 100115272 100115273 100115274 100115275 100115277 100115278 100115279 100115280 100115281 100115282 100115283 100115284 100115285 100115286 100115403 100115460 100115462 100115465 100115466 100115467 100115468 100115469 100115470 100115471 100115472 100115473 100115474 100115475 100115476 100115477 100115478 100115479 100115480 100115481 100115482 100115483 100115484 100115485 100115486 100115487 100115489 100115490 100115491 100115492 100115493 100115494 100115495 100115496 100115497 100115795 100115796 100115797 100115799 100115801 100115805 100115807 100115808 100116682 100116685 100117986 100117987 100117988 100117989 100117990 100117991 100117992 100117993 100117994 100117995 100117996 100117997 100117998 100117999 100118081 100118266 100118267 100118268 100118269 100118270 100118271 100118272 100118273 100118274 100118275 100118276 100118277 100118278 100118279 100118280 100118281 100118282 100118283 100118284 100118285 100118286 100118287 100118340 100118341 100118344 100118347 100118350 100118533 100118534 100118535 100118536 100118537 100118538 100118542 100118551 100118559 100118669 100118672 100118674 100118678 100118682 100118696 100118794 100118795 100118796 100118797 100118798 100118799 100118921 100118922 100118923 100118924 100118925 100118926 100118928 100118930 100118933 100118934 100118935 100118936 100118937 100118938 100118939 100118940 100118941 100118942 100118943 100118944 100118945 100118946 100118947 100118948 100118950 100118951 100118952 100119458 100119459 100119460 100119461 100119462 100119463 100119464 100119465 100119466 100119467 100119468 100119469 100119470 100119471 100119472 100119473 100119474 100119475 100119476 100119477 100119478 100119479 100119480 100119481 100119482 100119483 100119484 100119486 100119487 100119488 100119489 100119492 100119493 100119494 100120103 100120104 100120105 100120106 100120107 100120108 100120109 100120110 100120111 100120112 100120113 100120114 100120115 100120116 100120120 100120121 100120123 100120125 100120126 100120766 100122330 100122332 100122333 100122339 100122345 100122575 100122580 100122591 100122595 100122596 100122597 100122598 100122685 100122688 100123367 100123368 100123369 100123370 100123371 100123372 100123373 100123374 100123375 100123376 100123377 100123378 100123379 100123392 100123393 100123395 100123545 100123546 100123547 100123548 100123549 100123550 100123551 100123552 100123553 100123554 100123555 100124864 100125745 100125747 100125845 100125928 100125930 100125931 100125932 100125934 100125935 100125936 100125937 100125938 100125939 100125960 100125961 100125963 100125964 100125966 100125967 100125968 100125969 100125970 100125974 100125975 100125976 100125977 100125978 100125979 100125980 100125981 100125982 100125983 100125984 100126015 100126016 100126017 100126018 100126019 100126160 100126166 100126169 100126171 100126174 100126175 100126176 100126177 100126178 100126179 100126321 100126322 100126323 100126324 100126325 100126326 100126327 100126328 100126331 100126333 100126380 100126381 100126382 100126383 100126384 100126658 100126659 100126660 100126661 100126662 100126663 100126664 100126665 100126666 100126667 100126668 100126669 100126670 100126671 100126672 100126673 100126674 100126675 100126676 100126677 100126678 100126680 100126681 100126682 100137211 100137494 100137782 100137837 100137920 100138563 100139809 100139810 100139811 100139818 100139821 100139897 100143597 100143725 100143733 100143734 100143738 100143740 100143741 100143743 100143746 100143747 100143749 100143751 100143870 100144117 100144118 100144119 100144163 100144180 100144181 100144182 100144183 100144184 100144185 100144186 100144187 100144188 100144189 100144190 100144191 100144192 100144193 100144194 100144195 100144215 100144216 100144217 100144218 100144219 100144356 100144357 100144358 100144359 100144380 100144381 100144382 100144383 100144384 100144386 100144388 100144389 100144390 100144391 100144392 100144393 100144394 100144395 100144396 100144397 100144398 100144399 100144400 100144401 100144402 100144404 100144405 100144406 100144407 100144408 100144409 100144410 100144480 100144481 100144482 100144483 100144484 100144485 100144486 100144487 100144488 100144489 100144490 100144491 100144492 100144493 100144494 100144495 100144497 100144498 100144527 100144528 100144531 100144532 100144533 100144534 100144535 100144536 100144537 100144964 100145121 100145124 100145125 100145128 100145131 100145135 100145140 100145141 100145142 100145144 100145145 100145146 100145147 100145148 100145149 100145150 100145151 100145152 100145153 100145154 100145155 100145156 100145157 100145158 100145159 100145206 100145207 100145208 100145209 100145210 100145211 100145212 100145213 100145214 100145215 100145216 100145217 100145218 100145219 100145220 100145221 100145222 100145224 100145225 100145226 100145227 100145228 100145229 100145230 100145231 100145232 100145233 100145234 100145235 100145236 100145237 100145238 100145239 100145240 100145241 100145242 100145243 100145286 100145287 100145288 100145289 100145290 100145291 100145292 100145293 100145294 100145295 100145296 100145297 100145298 100145299 100145360 100145361 100145362 100145363 100145364 100145365 100145366 100145367 100145368 100145369 100145370 100145371 100145372 100145373 100145374 100145375 100145376 100145377 100145378 100145379 100145380 100145381 100145382 100145383 100145384 100145385 100145386 100145387 100145556 100145557 100145558 100145559 100145560 100145561 100145562 100145563 100145564 100145565 100145566 100145567 100145568 100145569 100145570 100145571 100145572 100145573 100145574 100145575 100145576 100145577 100145578 100145579 100145580 100145581 100145582 100145583 100145740 100145741 100145742 100145743 100145744 100145745 100145746 100145747 100145748 100145749 100145750 100145751 100145752 100145753 100145754 100145755 100145756 100145757 100145758 100145759 100145760 100145761 100145762 100145763 100145764 100145765 100145766 100145767 100145768 100145769 100145770 100145771 100145772 100145773 100145774 100145775 100145933 100145935 100145936 100145937 100145938 100145939 100145960 100145961 100145963 100145964 100145965 100145967 100145968 100145969 100145970 100145971 100145972 100145973 100145974 100145975 100145976 100145977 100145978 100145979 100145980 100145981 100145982 100145983 100145984 100145985 100147108 100147109 100147110 100147111 100147112 100147113 100147114 100147116 100147117 100147118 100147119 100147181 100147182 100147183 100147185 100147186 100147187 100147188 100147189 100147191 100147192 100147193 100147194 100147195 100147196 100147198 100147201 100147287 100147290 100147291 100147294 100147296 100147297 100147298 100147380 100149618 100149619 100149758 100149759 100149820 100149823 100149825 100149827 100149828 100149831 100149833 100149838 100149839 100149841 100149842 100149843 100149844 100149845 100149847 100149848 100149850 100149851 100149852 100149853 100149854 100149855 100149856 100149857 100149858 100149859 100149980 100149981 100149982 100149983 100149984 100149985 100149986 100149987 100149988 100149989 100149990 100149991 100149992 100149993 100149994 100149995 100149996 100149997 100149998 100149999 100150000 100150001 100150002 100150003 100150004 100150005 100150006 100150007 100150009 100150010 100150011 100150012 100150013 100150014 100150015 100150016 100150017 100150163 100150164 100150165 100150166 100150167 100150168 100150169 100150170 100150171 100150172 100150180 100150181 100150182 100150183 100150184 100150185 100150187 100150188 100150189 100150191 100150193 100150203 100150204 100150205 100150206 100150207 100150209 100150210 100150212 100158460 100160690 100160824 100160831 100160832 100160837 100161455 100161456 100161457 100161458 100161459 100161576 100161577 100161578 100161579 100161680 100161681 100161682 100161683 100161684 100161685 100161686 100161687 100161688 100161689 100161690 100161691 100161692 100161693 100161694 100161695 100161696 100161697 100161698 100161699 100161700 100161701 100161702 100161703 100161704 100161705 100161706 100161707 100161708 100161709 100161710 100161711 100161712 100161713 100161714 100161715 100161716 100161717 100161718 100161719 100161720 100161924 100161925 100161926 100161927 100161928 100161929 100161931 100161932 100161933 100161934 100161935 100161936 100161937 100161938 100161939 100161940 100161941 100161942 100161943 100161944 100161945 100161946 100161947 100161948 100161949 100161950 100161951 100161952 100161953 100161954 100161955 100161956 100161957 100161961 100161963 100161964 100161965 100161966 100161967 100162407 100162408 100162410 100162411 100162415 100162444 100162445 100163589 100163711 100163717 100163718 100163720 100163724 100163727 100163728 100163729 100163730 100163731 100163739 100163743 100163747 100163748 100163752 100163753 100163754 100163927 100163928 100163929 100163932 100163933 100163935 100163936 100163937 100163938 100163939 100164001 100164002 100164003 100164004 100164005 100164006 100164007 100164008 100164009 100164010 100164039 100164080 100164081 100164083 100164084 100164085 100164086 100164090 100169177 100169179 100169220 100169221 100169228 100169231 100169233 100169235 100169237 100169238 100169239 100169240 100169241 100169329 100169330 100169331 100169332 100169333 100169334 100169335 100169336 100169337 100169338 100169339 100169340 100169341 100169342 100169343 100169344 100169345 100169346 100169347 100169348 100169349 100169350 100169351 100169352 100169353 100169354 100169355 100169357 100169358 100169362 100169363 100169365 100169367 100169368 100169370 100169373 100170687 100173085 100173087 100173088 100173091 100173092 100173093 100173094 100173099 100173205 100173220 100173221
100173222 100173223 100173226 100173229 100173232 100173238 100173243 100173249 100173255 100173291 100173292 100173293 100173295 100173296 100173297 100173298 100173299 100173441 100173444 100173445 100173446 100173447 100173448 100173449 100173450 100173452 100173453 100173454 100173456 100173459 100173463 100173466 100173470 100173589 100173590 100173592 100173596 100178918 100179380 100179383 100179384 100179408 100179410 100179420 100179421 100179422 100179423 100179424 100179462 100179463 100179466 100179469 100179470 100179471 100179472 100179475 100179476 100179477 100179479 100179560 100179561 100179562 100179563 100179564 100179565 100179602 100179603 100179604 100179605 100179606 100179607 100179608 100179609 100179610 100179611 100179612 100179613 100179614 100179615 100179616 100179617 100179618 100179619 100179620 100179621 100179622 100179623 100179624 100179625 100179626 100179627 100179628 100179629 100179630 100179631 100179632 100179633 100179634 100179635 100179636 100179637 100179638 100179639 100179640 100179641 100179642 100179643 100179644 100179645 100179646 100179647 100179648 100179650 100179651 100179652 100179653 100179654 100180187 100180188 100180190 100180191 100180192 100180193 100180194 100180195 100180196 100180198 100180199 100180200 100180203 100180204 100180205 100180211 100180216 100180220 100180221 100180225 100180227 100180228 100180230 100180231 100180232 100180233 100180234 100180235 100180236 100180386 100180387 100180388 100180389 100180390 100180391 100180392 100180393 100180394 100180395 100180396 100180397 100180398 100180399 100180400 100180401 100180402 100180403 100180404 100180405 100180406 100180407 100180408 100180409 100180410 100180412 100180414 100180419 100180424 100180426 100180429 100180430 100180431 100180432 100180433 100180434 100180435 100181887 100181898 100182121 100182122 100182123 100182125 100182126 100182128 100182129 100182130 100182131 100182133 100182135 100182136 100182137 100182138 100182139 100182141 100182143 100182145 100182147 100182148 100182149 100182150 100182388 100182390 100182391 100182392 100182393 100182394 100182395 100182398 100182420 100182424 100182427 100182435 100182439 100182440 100182441 100182445 100182446 100182447 100182449 100182450 100182451 100182452 100182460 100182468 100182469 100182471 100182472 100182473 100182476 100182477 100182479 100182480 100184596 100184597 100184598 100184599 100184600 100184601 100184602 100184603 100184604 100184605 100184606 100184607 100184609 100184610 100184611 100184612 100184613 100184614 100184615 100184616 100184617 100184618 100184619 100184620 100184621 100184622 100184623 100184624 100184625 100184626 100184627 100184628 100184629 100184630 100184631 100184632 100184633 100184634 100184635 100184636 100184638 100184639 100185373 100185374 100185375 100185376 100185377 100185378 100185379 100185380 100185383 100185384 100185385 100185595 100185596 100185801 100185802 100185803 100185804 100185805 100185806 100185807 100185808 100185809 100185810 100185811 100185812 100185813 100185814 100185815 100185816 100185817 100185818 100185819 100185820 100185821 100185822 100185823 100185824 100185825 100185826 100185827 100185828 100185829 100185830 100185831 100185832 100185833 100185834 100185835 100185836 100185838 100185839 100185840 100185962 100185964 100186775 100186779 100186789 100186796 100186797 100186798 100186799 100186800 100186801 100186802 100186803 100186806 100186808 100186809 100186900 100186901 100186902 100186903 100186904 100187084 100187088 100187090 100187091 100187092 100187093 100187094 100187096 100187097 100187098 100187099 100187100 100187101 100187102 100187103 100187104 100187105 100187106 100187107 100187108 100187109 100187110 100187111 100187112 100187113 100187115 100187116 100187118 100187119 100195310 100195314 100195315 100195317 100195318 100195443 100195450 100195459 100195460 100195462 100195466 100195469 100195644 100195646 100195652 100195658 100195661 100195663 100195664 100195666 100195667 100195669 100195671 100195673 100195674 100195863 100195865 100195866 100195868 100195871 100195877 100195879 100195880 100195881 100195883 100195887 100195888 100195890 100195895 100195967 100195972 100195974 100195975 100195977 100195979 100196433 100196435 100196438 100196621 100196626 100223190 100223281 100223285 100223292 100223480 100224884 100224885 100224886 100224887 100224888 100224889 100224890 100224891 100224892 100224893 100224894 100224895 100224896 100224897 100224898 100224899 100225260 100225261 100225262 100225264 100225266 100225268 100225269 100225272 100225274 100225275 100225276 100225277 100225278 100226341 100226342 100226343 100226344 100226345 100226346 100226347 100226349 100226350 100226352 100226353 100226355 100226356 100226360 100226361 100226362 100226365 100226385 100226681 100226683 100226684 100226685 100226698 100226699 100226720 100226721 100226722 100226723 100226724 100226725 100226727 100226728 100226729 100226730 100226731 100226732 100226733 100226734 100226735 100226736 100226737 100226738 100226739 100226740 100226741 100226742 100226860 100226862 100226863 100226864 100226865 100226866 100226867 100226868 100226871 100226872 100226873 100226874 100226875 100226879 100226880 100226882 100226885 100226886 100226889 100226890 100226892 100226893 100227093 100227120 100227121 100227308 100227309 100227310 100227311 100227312 100227313 100227314 100227315 100227316 100227317 100227318 100227319 100227509 100227510 100227512 100227513 100227515 100227516 100227517 100227518 100227519 100227520 100227521 100227522 100227523 100227524 100227525 100227526 100227527 100227528 100227730 100227780 100227781 100227782 100227783 100227785 100227786 100227787 100227788 100227790 100227792 100227793 100227794 100227795 100227796 100230072 100230169 100231064 100231065 100231103 100231104 100231105 100231109 100231110 100231111 100231112 100231113 100231114 100231115 100231116 100231117 100231118 100231169 100231174 100231179 100231200 100231201 100231204 100231205 100231206 100231207 100231208 100231209 100231210 100231211 100231405 100231406 100231407 100231408 100231409 100231410 100231411 100231412 100232460 100232461 100232462 100232463 100232464 100232465 100232466 100232467 100232468 100232469 100232470 100232471 100232472 100232478 100232480 100232487 100232740 100232744 100232753 100232765 100232770 100233064 100233940 100233941 100233943 100233944 100233945 100233946 100233947 100233948 100233950 100233951 100233952 100233953 100233954 100233955 100233956 100233957 100233958 100233959 100233960 100233961 100237066 100237067 100237068 100237070 100237071 100237072 100237073 100237074 100237075 100237076 100237077 100237078 100237079 100237184 100237185 100237186 100237187 100237188 100237193 100237199 100237220 100237221 100237224 100237225 100237226 100237227 100237228 100237229 100237230 100237231 100237233 100237234 100237235 100237237 100237238 100237301 100237304 100237305 100237307 100237311 100237313 100237314 100237315 100237316 100237317 100237318 100237322 100237325 100237326 100246476 100246477 100246478 100246479 100246480 100246481 100246482 100246483 100246485 100246486 100246487 100246488 100246489 100246490 100246492 100246493 100246494 100246495 100246496 100246497 100246499 100246500 100246501 100246504 100246505 100246508 100246509 100246510 100246511 100246512 100246514 100246515 100246518 100246519 100246520 100246521 100246522 100246523 100246525 100246526 100246527 100246528 100247080 100247081 100247083 100247084 100247088 100247089 100247090 100247099 100247100 100247103 100247104 100247106 100247109 100247110 100247111 100247112 100247254 100247256 100247363 100247364 100247365 100247366 100247367 100247370 100247371 100247372 100247373 100247374 100247375 100247376 100247377 100247378 100247379 100247400 100247404 100247408 100247414 100247415 100247484 100247485 100247486 100247488 100247489 100247520 100247521 100247522 100247523 100247524 100247525 100247526 100247527 100247529 100247530 100247531 100247533 100247534 100247535 100247536 100247537 100247539 100247540 100247541 100247542 100247543 100247544 100247545 100247547 100247548 100247549 100247550 100247551 100247610 100247611 100247612 100247613 100247614 100247615 100247617 100247729 100247730 100247732 100247734 100247736 100247737 100247738 100247739 100247760 100247761 100247765 100247766 100247921 100247923 100247924 100247925 100247926 100247927 100247928 100247929 100247931 100247932 100247934 100247935 100247936 100247938 100247944 100248615 100249454 100249720 100249721 100249722 100249723 100249725 100249726 100249727 100249728 100249729 100249730 100249731 100249732 100249733 100249734 100249735 100249737 100249738 100249739 100249740 100249741 100249742 100249744 100249745 100249746 100249748 100249749 100249751 100249752 100249753 100249754 100249755 100249756 100249757 100249758 100249759 100249761 100249762 100249763 100249765 100249895 100249896 100249898 100249899 100249900 100249901 100249902 100249903 100249904 100249905 100249906 100249908 100249909 100249911 100249914 100249928 100249931 100249936 100250328 100250331 100250333 100250338 100250339 100250340 100250341 100250342 100250344 100250346 100250347 100250348 100250349 100250350 100250351 100250352 100250353 100250354 100250355 100250356 100250357 100250358 100250359 100250360 100250361 100250363 100250364 100251168 100251170 100251171 100251172 100251175 100251177 100251179 100251240 100251241 100251242 100251243 100251244 100251245 100251246 100251247 100251248 100251249 100251250 100251251 100251252 100251253 100251254 100251255 100251256 100251257 100251258 100251259 100251260 100251261 100251262 100251263 100251264 100251265 100251266 100251267 100251268 100251269 100251270 100251271 100251272 100251464 100251465 100251466 100251467 100251468 100251469 100251470 100251471 100251472 100251473 100251474 100251475 100251476 100251477 100251478 100251479 100251520 100251521 100251522 100251523 100251526 100251527 100251528 100251529 100251530 100251531 100251532 100251533 100251534 100251535 100251536 100251538 100251539 100251540 100251541 100251542 100251543 100251544 100251545 100251546 100251649 100251650 100251652 100251653 100251654 100251655 100251656 100251657 100251658 100251659 100251660 100251661 100251662 100251663 100251664 100251665 100251666 100251667 100251668 100251669 100251670 100251671 100251672 100251673 100251674 100251675 100251676 100251677 100251678 100251679 100251680 100251681 100251682 100256844 100256848 100256850 100256859 100256980 100256982 100256983 100256984 100256985 100256986 100256988 100257085 100257086 100257087 100257089 100257090 100257091 100257093 100257094 100257096 100257097 100257098 100257099 100257162 100257163 100257164 100257166 100257167 100257171 100257172 100257173 100257174 100257175 100257176 100257200 100257201 100257202 100257203 100257204 100257205 100257207 100257208 100257209 100257211 100257212 100257216 100257219 100257260 100257261 100257263 100257264 100257265 100257267 100257268 100257269 100257271 100257311 100257312 100257313 100257314 100257315 100257316 100257318 100257319 100257377 100257378 100257379 100257380 100257381 100257382 100257383 100257384 100257385 100257386 100257387 100257395 100257397 100257398 100257400 100257402 100257403 100257404 100257580 100257581 100257583 100257584 100257587 100257588 100257589 100257591 100257592 100257672 100257673 100257674 100257675 100257676 100257677 100257678 100257679 100257680 100257681 100257682 100257683 100257684 100257685 100257686 100257687 100257688 100257689 100257690 100257693 100257694 100257695 100257696 100257697 100257698 100257800 100257801 100257803 100257804 100257805 100257806 100257807 100257808 100257810 100257813 100257814 100257815 100257817 100257819 100257921 100257923 100257924 100258020 100258021 100258023 100258024 100258025 100258026 100258027 100258028 100258029 100258030 100258031 100258033 100258034 100258035 100259726 100259737 100259783 100259785 100259786 100259788 100259791 100259796 100259797 100259873 100259998 100260032 100260034 100260174 100260176 100260186 100260231 100260234 100260235 100260236 100260237 100260238 100260239 100260240 100260241 100260242 100260243 100260244 100260245 100260246 100260247 100260248 100260249 100260250 100260251 100260252 100260253 100260254 100260255 100260256 100260257 100260259 100260440 100260451 100260452 100260525 100260526 100260527 100260528 100260529 100260530 100260531 100260532 100260533 100260534 100260535 100260536 100260537 100260538 100260539 100260540 100260541 100260542 100260543 100260544 100260545 100260546 100260547 100260549 100260551 100260552 100260553 100260554 100260707 100260708 100260709 100260710 100260712 100260713 100260714 100260951 100260953 100260954 100260955 100260957 100260958 100260959 100260960 100260961 100260962 100260963 100260965 100260966 100260967 100260968 100260974 100261115 100261117 100267528 100267564 100267570 100267572 100267573 100267689 100267692 100267697 100267699 100267700 100267701 100267703 100267704 100267706 100267707 100267708 100268244 100268245 100268246 100268247 100268248 100268249 100268250 100268251 100268252 100268253 100268254 100268255 100268256 100268257 100268258 100268259 100268320 100268321 100268322 100268323 100268324 100268325 100268326 100268327 100268328 100268329 100268330 100268331 100268332 100268333 100268334 100268335 100268336 100268337 100268338 100268339 100268426 100268427 100268428 100268429 100268430 100268431 100268432 100268433 100268434 100268435 100268436 100268437 100268438 100268439 100268480 100268481 100268482 100268483 100268484 100268485 100268486 100268487 100268488 100268489 100268490 100268491 100268492 100268493 100268494 100268495 100268496 100268497 100268498 100268499 100268500 100268501 100268502 100268503 100268504 100268505 100268506 100268507 100268508 100268509 100268510 100268511 100268512 100268513 100268600 100268601 100268602 100268603 100268604 100268605 100268606 100268607 100268608 100268609 100268610 100268611 100268612 100268613 100268614 100268615 100268616 100268617 100268618 100268619 100268620 100268621 100268622 100268623 100268624 100268625 100268626 100268627 100268628 100268629 100268630 100268631 100268860 100268861 100268862 100268863 100268864 100268865 100268866 100268867 100268868 100268869 100268870 100268871 100268872 100268873 100268874 100268875 100268876 100268877 100268878 100268879 100268880 100268881 100268882 100268883 100268884 100268885 100268886 100268887 100268888 100268889 100268890 100268891 100268892 100268893 100268894 100268895 100268896 100268897 100268911 100268912 100268913 100268914 100268915 100268916 100268917 100268918 100268919 100269080 100269081 100269082 100269083 100269084 100269085 100269086 100269087 100269088 100269089 100269090 100269091 100269092 100269093 100269094 100269095 100269096 100269097 100269098 100269099 100269100 100269101 100269102 100269103 100269104 100269105 100269106 100269107 100269108 100269109 100269110 100269111 100269112 100269113 100269114 100269115 100269116 100269117 100269118 100269119 100269127 100269128 100269129 100269130 100269131 100269132 100269133 100269134 100269135 100269136 100269137 100269138 100269139 100269140 100269141 100269142 100269143 100269144 100269145 100269146 100269147 100269148 100269149 100269150 100269151 100269152 100269153 100269154 100269155 100269156 100269157 100269158 100269159 100269160 100269161 100269162 100269163 100269164 100269165 100269166 100269167 100269168 100269169 100269170 100269300 100269301 100269302 100269303 100269304 100269305 100269306 100269307 100269308 100269309 100269310 100269311 100269312 100269313 100269314 100269315 100269316 100269317 100269318 100269319 100269320 100269321 100269322 100269323 100269326 100269327 100269328 100269329 100269330 100269331 100269332 100269333 100269334 100269335 100269336 100269582 100269583 100269584 100269585 100269586 100269587 100269588 100269589 100269590 100269591 100269592 100269593 100269594 100269595 100269596 100269597 100269598 100269599 100269600 100269601 100269602 100269603 100269604 100269605 100269606 100269607 100269608 100269609 100269610 100269611 100269612 100269613 100269614 100269615 100269616 100269617 100269618 100269619 100269620 100269621 100269622 100269623 100269624 100269625 100269626 100269627 100269628 100269629 100269630 100269631 100269632 100269633 100269634 100269635 100269636 100269637 100269638 100269768 100269769 100269770 100269771 100269772 100269773 100269774 100269775 100269776 100269777 100269778 100269779 100269780 100269781 100269782 100269783 100269784 100269785 100269786 100269787 100269788 100269789 100269790 100269791 100269792 100269793 100269794 100269795 100269796 100269797 100269798 100269799 100269800 100269801 100269802 100269803 100269804 100269805 100269806 100269807 100269808 100269809 100269810 100269811 100269812 100269813 100269814 100269815 100269816 100269817 100270000 100270001 100270003 100270004 100270005 100270006 100270007 100270008 100270009 100270010 100270011 100270012 100270013 100270014 100270015 100270016 100270017 100270018 100270019 100270020 100270021 100270022 100270023 100270024 100270025 100270026 100270027 100270028 100270029 100270030 100270031 100270032 100270033 100270034 100270035 100270036 100270037 100270038 100270039 100270040 100270252 100270253 100270254 100270255 100270256 100270258 100270259 100270260 100270261 100270262 100270263 100270264 100270265 100270266 100270267 100270268 100270269 100270270 100270271 100270272 100270273 100270274 100270275 100270277 100270279 100270280 100270281 100270282 100270284 100270285 100270286 100270288 100270289 100270292 100272768 100274440 100274456 100274586 100274595 100274694 100274799 100280977 100280982 100280984 100280990 100280991 100280992 100280993 100280994 100280995 100280996 100280997 100280998 100280999 100281000 100281001 100281002 100281003 100281004 100281005 100281006 100281007 100282833 100282882 100286941 100286942 100286944 100286946 100286951 100286953 100286954 100286955 100286958 100286959 100286960 100286961 100286962 100286964 100286965 100286966 100286967 100286968 100286970 100286971 100286977 100286979 100287349 100287350 100287351 100287352 100287353 100287354 100287355 100287356 100287357 100287358 100287359 100287401 100287402 100287403 100287404 100287405 100287406 100287407 100287408 100287409 100287410 100287411 100287412 100287413 100287414 100287415 100287416 100287417 100287418 100287419 100287420 100287422 100287423 100287424 100287425 100287426 100287427 100287428 100287430 100287431 100287432 100287433 100287435 100287436 100287437 100287438 100287439 100287440 100287480 100287481 100287482 100287483 100287484 100287485 100287486 100287487 100287488 100287489 100287490 100287491 100287492 100287493 100287494 100287495 100287496 100287497 100287498 100287499 100287500 100287501 100287502 100287503 100287504 100287505 100287506 100287507 100287508 100287739 100287740 100287741 100287742 100287743 100287744 100287745 100287746 100287747 100287748 100287749 100287750 100287751 100287752 100287753 100287754 100287755 100287756 100287757 100287758 100287759 100287760 100287761 100287762 100287763 100287764 100287765 100287766 100287767 100287768 100287769 100287770 100287771 100287772 100287773 100287774 100287775 100287776 100287778 100287873 100287874 100287875 100287876 100287877 100287878 100287879 100288011 100288012 100288013 100288014 100288015 100288016 100288017 100288018 100288019 100288120 100288121 100288122 100288123 100288124 100288125 100288126 100288127 100288128 100288129 100288130 100288131 100288132 100288133 100288134 100288135 100288136 100288137 100288138 100288139 100288140 100288141 100288142 100288143 100288154 100288155 100288156 100288157 100288158 100288159 100288259 100288260 100288261 100288262 100288300 100288301 100288302 100288303 100288304 100288305 100288307 100288308 100288309 100288310 100288312 100288313 100288314 100288315 100288316 100288317 100288318 100288320 100288322 100288324 100288326 100288329 100288331 100288332 100288334 100288335 100288336 100288994 100288995 100288996 100288997 100288998 100288999 100289062 100289063 100289064 100289065 100289066 100289067 100289068 100289069 100289070 100289071 100289072 100289073 100289074 100289075 100289076 100289077 100289078 100289079 100289080 100289081 100289082 100289083 100289084 100289085 100289086 100289088 100289089 100289090 100289091 100289092 100289093 100289094 100289095 100289096 100289097 100289098 100289099 100289101 100289280 100289281 100289429 100289432 100289435 100289439 100289483 100289484 100289486 100289487 100289488 100289489 100289490 100289491 100289493 100289494 100289497 100289499 100289500 100289501 100289681 100289682 100289683 100289685 100289687 100289688 100289693 100291319 100291454 100291461 100291463 100291476 100291480 100291826 100291827 100291828 100291830 100291835 100291836 100291840 100291841 100291843 100291845 100293608 100294321 100294328 100294329 100294330 100294331 100294332 100294333 100294334 100294335 100294336 100294337 100294338 100294339 100294340 100294341 100294342 100294343 100294344 100295729 100295730 100295731 100295732 100295734 100295735 100295736 100295737 100295738 100295739 100295800 100295801 100295802 100295803 100295804 100295805 100295806 100295807 100295809 100295810 100295811 100295812 100295813 100295814 100295815 100295816 100295817 100295818 100295819 100295977 100295978 100295979 100296000 100296001 100296002 100296003 100296004 100296005 100296006 100296007 100296008 100296009 100296010 100296011 100296012 100296013 100296014 100296015 100296016 100296017 100296018 100296019 100296020 100296021 100296022 100296024 100296025 100296026 100296027 100296028 100296029 100296030 100296031 100296119 100296120 100296121 100296122 100296123 100296124 100296125 100296126 100296127 100296128 100296129 100296130 100296131 100296132 100296133 100296134 100296135 100296136 100296137 100296138 100296139 100296140 100296141 100296142 100296143 100296144 100296145 100296146 100296147 100296148 100296354 100296355 100296356 100296357 100296358 100296440 100296441 100296442 100296443 100296444 100296445 100296446 100296447 100296448 100296449 100296450 100296451 100296452 100296453 100296454 100296455 100296456 100296457 100296458 100296519 100296541 100296542 100296543 100296544 100296545 100296552 100296553 100296558 100296560 100296562 100296563 100296564 100296565 100296566 100296567 100296568 100296569 100296570 100296571 100296572 100296747 100296748 100296749 100296750 100296751 100296752 100296756 100296758 100296759 100296781 100296783 100296784 100296785 100296786 100296787 100296788 100296789 100296790 100296791 100296792 100296793 100296794 100296795 100296796 100296827 100296828 100296829 100296830 100296831 100296832 100296833 100296834 100296835 100296836 100296837 100296838 100296840 100296841 100296842 100296844 100296982 100296983 100296984 100296985 100296986 100296987 100296988 100296989 100296990 100296991 100296992 100296993 100296994 100296996 100296997 100296998 100296999 100297140 100297143 100297144 100297145 100297146 100297149 100297152 100297155 100297158 100297222 100297225 100297228 100297231 100297235 100297238 100297241 100297245 100297248 100297251 100297254 100297258 100297259 100297260 100297261 100297262 100297263 100297264 100297265 100297266 100297267 100297268 100297269 100297270 100297271 100297272 100297273 100297275 100297276 100297277 100297278 100297279 100297360 100297361 100297362 100297363 100297364 100297365 100297366 100297367 100297368 100297369 100297370 100297371 100297372 100297373 100297374 100297375 100297376 100297377 100297378 100297379 100297380 100297381 100297382 100297383 100297384 100297820 100297821 100297822 100297823 100297824 100297825 100297826 100297827 100297828 100297829 100297830 100297831 100297832 100297833 100297835 100297836 100297838 100297839 100297841 100297843 100297844 100297845 100297854 100297857 100297859 100298642 100298643 100298644 100298645 100298646 100298647 100298648 100298649 100298650 100298651 100298652 100298653 100298654 100298656 100298657 100298658 100298768 100298769 100298770 100298771 100298772 100298773 100298774 100298775 100298776 100298777 100298778 100298779 100298780 100298781 100298782 100298783 100298784 100298785 100298786 100298787 100298788 100298789 100298790 100298791 100298792 100298793 100298794 100298795 100298796 100298797 100298798 100298800 100298801 100298802 100298803 100298804 100298805 100298921 100298922 100298923 100298924 100298925 100298926 100298927 100298930 100298932 100298933 100298934 100300418 100300419 100300420 100300421 100300422 100300423 100300424 100300425 100300426 100300427 100300428 100300429 100300430 100300431 100300432 100300433 100300435 100300436 100300437 100300438 100300439 100300524 100300525 100300526 100300528 100300529 100300530 100300532 100300536 100301390 100301392 100301394 100301395 100301396 100301397 100301398 100301399 100301440 100301441 100301442 100301443 100301444 100301445 100301995 100301996 100301997 100301998 100301999 100302001 100302002 100302010 100302013 100302017 100302264 100302265 100302267 100302268 100302269 100302270 100302271 100302272 100302273 100302274 100302275 100302480 100302481 100302482 100302483 100302484 100302486 100302487 100302488 100302490 100302491 100302492 100302493 100302494 100304058 100304059 100304086 100304087 100304088 100304089 100304090 100304092 100304093 100304094 100304096 100304098 100304099 100304118 100304119 100304220 100304221 100304222 100304260 100304261 100304262 100304265 100304440 100304441 100304442 100304443 100304444 100304445 100304449 100304450 100304452 100304453 100304454 100304455 100304456 100304457 100304458 100304459 100304460 100304461 100304463 100304464 100304465 100304466 100304467 100304518 100304519 100304537 100304538 100304539 100304560 100304561 100304562 100304563 100304564 100304565 100304566 100304567 100304568 100304569 100304570 100304571 100304572 100304574 100304575 100304576 100304603 100304604 100304605 100304606 100304608 100304609 100304610 100304611 100304614 100304615 100304616 100304618 100304619 100304680 100304681 100304682 100304683 100304684 100304685 100304686 100304687 100304688 100304690 100304691 100304692 100304693 100304694 100304695 100304696 100304698 100304700 100304702 100304703 100304704 100304705 100304706 100304708 100304709 100304710 100304711 100304712 100304713 100304714 100304715 100304716 100304717 100304718 100304719 100304760 100304761 100304763 100304765 100304767 100304768 100304769 100304771 100304773 100304774 100304777 100304778 100304780 100304781 100304782 100304783 100304784 100304785 100304787 100304900 100304901 100304902 100304903 100304904 100304905 100304906 100304907 100304945 100304946 100304947 100304948 100304951 100304952 100304954 100304955 100304957 100304959 100305040 100305041 100305042 100305580 100305669 100305670 100305671 100305672 100305673 100305674 100305675 100305676 100305679 100305741 100306470 100306471 100306472 100306473 100306476 100306479 100306500 100306507 100306624 100306626 100306629 100306630 100306633 100306634 100306635 100306636 100306637 100306639 100306640 100306642 100309412 100309414 100309415 100309416 100309417 100309418 100309419 100309480 100309482 100309483 100309484 100309485 100309486 100309487 100309488 100309489 100309490 100309493 100309495 100311492 100311879 100312445 100312459 100312520 100312524 100312525 100312526 100312527 100312529 100312534 100312535 100312536 100312537 100313061 100313065 100313070 100313071 100313072 100313073 100313074 100313076 100313101 100313102 100313103 100313105 100314484 100314485 100314487 100314488 100314489 100314490 100314491 100314492 100314493 100314494 100314495 100314496 100314498 100314499 100314537 100314538 100314539 100314620 100314621 100314622 100314623 100314624 100314625 100314627 100314635 100314875 100314879 100315037 100315039 100315151 100315159 100315185 100315187 100315194 100315197 100315198 100315199 100315361 100315363 100315365 100315366 100315367 100315368 100315369 100315370 100315371 100315372 100315373 100315374 100315375 100315376 100315378 100315379 100315380 100315381 100315382 100315383 100315384 100315385 100315386 100315387 100315388 100315389 100315392 100315393 100315394 100315395 100315396 100315599 100315601 100315603 100315605 100315606 100315608 100315609 100315612 100315613 100315616 100315871 100315873 100315874 100315878 100315879 100315891 100315892 100315893 100315896 100315898 100315899 100316061 100316062 100316064 100316067 100316068 100316069 100316907 100316917 100316918 100316919 100317128 100317129 100317133 100317134 100317136 100317137 100317138 100317139 100317260 100317261 100317262 100317263 100317265 100317266 100317276 100317380 100317382 100317384 100317388 100317497 100318308 100318310 100318311 100318313 100318314 100318317 100318318 100318319 100318400 100318401 100318403 100318404 100318409 100318410 100318412 100318417 100318419 100318420 100318421 100318423 100318425 100318428 100318429 100318430 100318595 100318596 100318598 100318606 100318611 100318617 100318618 100318619 100318626 100318627 100318628 100319094 100319095 100319096 100319097 100319098 100319099 100319220 100319221 100319222 100319223 100319265 100319266 100319267 100319268 100319269 100319273 100319275 100319276 100319277 100319278 100319279 100319280 100319281 100319282 100319283 100319285 100319287 100319288 100319297 100319301 100319580 100319581 100319582 100319583 100319584 100319585 100319586 100319587 100319588 100319590 100319591 100319592 100319593 100319594 100319595 100319596 100319597 100319598 100319599 100319600 100319601 100319602 100319603 100319604 100319605 100319606 100319607 100319608 100319609 100319755 100319757 100319758 100319759 100319760 100319761 100319762 100319763 100319765 100319767 100319791 100319795 100319796 100320045 100320046 100320053 100320054 100320055 100320058 100320059 100320145 100320146 100320147 100320148 100320151 100320155 100320156 100320157 100320158 100320360 100320363 100320364 100320365 100320366 100320368 100320369 100320370 100320372 100320374 100320375 100320376 100320378 100320379 100320380 100320381 100320382 100320383 100320384 100320385 100320386 100320388 100320389 100320390 100320391 100320392 100320393 100320394 100320395 100320396 100320397 100320402 100320632 100327210 100327532 100327537 100327607 100327608 100327609 100327612 100327770 100327882 100327883 100327885 100327886 100327887 100327888 100327889 100327890 100327891 100327892 100327893 100327894 100327895 100327896 100327898 100327912 100328006 100328007 100328008 100328010 100328011 100328012 100328014 100328015 100328016 100328017 100328018 100328039 100328040 100328041 100328042 100328043 100328044 100328045 100328047 100328049 100328050 100328051 100328052 100328053 100328140 100328141 100328142 100328143 100328144 100328220 100328266 100328267 100328268 100328269 100328270 100328271 100328272 100328273 100328274 100328275 100328277 100328278 100328279 100328318 100328319 100328360 100328361 100328362 100328363 100328364 100328365 100328366 100328367 100328368 100328369 100328370 100328371 100328372 100328460 100328461 100328462 100328463 100328464 100328465 100328466 100328467 100328468 100328479 100328601 100328602 100328603 100328605 100328606 100328607 100328608 100328610 100328620 100328622 100328623 100328624 100328626 100328629 100328631 100328786 100328787 100328788 100328908 100329960 100329970 100329973 100329989 100330121 100330123 100330125 100330127 100330130 100331715 100331716 100331717 100331718 100331719 100331820 100331821 100331822 100331823 100331824 100331825 100331826 100331827 100331828 100331829 100331830 100331831 100331832 100331833 100331834 100331835 100331836 100331837 100331838 100331839 100331840 100331841 100331842 100331843 100331844 100331845 100331846 100331847 100331848 100331849 100331850 100331851 100331852 100331853 100331854 100331855 100331856 100331857 100331858 100331859 100332002 100332003 100332004 100332005 100332006 100332007 100332008 100332009 100332010 100332011 100332012 100332013 100332014 100332015 100332016 100332017 100332018 100332019 100332020 100332021 100332022 100332023 100332024 100332025 100332026 100332027 100332028 100332029 100332030 100332031 100332032 100332033 100332034 100332035 100332036 100332037 100332038 100332039 100332040 100332041 100332042 100332043 100332044 100332045 100332046 100332047 100332196 100332197 100332198 100332199 100332260 100332261 100332262 100332263 100332264 100332265 100332266 100332267 100332268 100332269 100332270 100332271 100332272 100332273 100332274 100332275 100332276 100332277 100332278 100332279 100332280 100332281 100332282 100332283 100332284 100332285 100332286 100332287 100332288 100332289 100332290 100332291 100332292 100332293 100332294 100332295 100332320 100332321 100332322 100332323 100332324 100332325 100332326 100332327 100332328 100332329 100332330 100332331 100332332 100332333 100332334 100332335 100332336 100332337 100332338 100332339 100332340 100332341 100332342 100332343 100332344 100332345 100332346 100332347 100332348 100332349 100332350 100332351 100332352 100332353 100332354 100332355 100332356 100332358 100332359 100332360 100332361 100332362 100332363 100332364 100332555 100332556 100332557 100332558 100332559 100332680 100332681 100332682 100332683 100332684 100332685 100332686 100332688 100332689 100332691 100332692 100332693 100332694 100332695 100332696 100332697 100332698 100332699 100332700 100332701 100332702 100332703 100332704 100332705 100332706 100332707 100332708 100332709 100332710 100332711 100332712 100332713 100332714 100332715 100332716 100332717 100332718 100332719 100332720 100332721 100332722 100332723 100332725 100332726 100332727 100332728 100332729 100332730 100332731 100332732 100332733 100332734 100332735 100332736 100332737 100332738 100332739 100332740 100332741 100332742 100332743 100332744 100332745 100332746 100332747 100332881 100332882 100332883 100332884 100332885 100332886 100332887 100332889 100332890 100332891 100332893 100332894 100332895 100332896 100332897 100332898 100332899 100333001 100333002 100333003 100333004 100333006 100333007 100333008 100333009 100333010 100333011 100333012 100333013 100333014 100333016 100333017 100333018 100333019 100333020 100333021 100333022 100333023 100333024 100333025 100333026 100333027 100333029 100333030 100333032 100333033 100333034 100333096 100333097 100333099 100333276 100333277 100333278 100333279 100333281 100333282 100333283 100333284 100333285 100333286 100333287 100333288 100333289 100333290 100333291 100333292 100333293 100333294 100333295 100333296 100333297 100333298 100333300 100333301 100333302 100333304 100333306 100333307 100333308 100333309 100333310 100333311 100333312 100333313 100333314 100333315 100333316 100333320 100333321 100333530 100333531 100333532 100333535 100333537 100333540 100333541 100333543 100333544 100333545 100333546 100333548 100333553 100333558 100333560 100333567 100333570 100333575 100333577 100333579 100333741 100333744 100333745 100333746 100333747 100333749 100333751 100333753 100333761 100333767 100333768 100333776 100333778 100333779 100333781 100333782 100333783 100333784 100333785 100333786 100333788 100333789 100333790 100333791 100333792 100333794 100333795 100333796 100333797 100333798 100333799 100333800 100333801 100333802 100333803 100333804 100334000 100334001 100334002 100334003 100334004 100334005 100334006 100334007 100334008 100334009 100334010 100334011 100334012 100334013 100334014 100334015 100334016 100334017 100334018 100334019 100334020 100334021 100334022 100334023 100334024 100334025 100334026 100334027 100334028 100334029 100334030 100334031 100334032 100334033 100334034 100334035 100334036 100334037 100334038 100334039 100334040 100334041 100334042 100334043 100334044 100334045 100334046 100334047 100334048 100334049 100334050 100334051 100334052 100334952 100334953 100334954 100334955 100334956 100334957 100334958 100334959 100335349 100335350 100335351 100335352 100335353 100335354 100335355 100335356 100335357 100335358 100335359 100335420 100335421 100335422 100335423 100335424 100335425 100335426 100335427 100335428 100335429 100335430 100335431 100335432 100335433 100335434 100335435 100335436 100335437 100335438 100335439 100335440 100335441 100335442 100335443 100335444 100335445 100335446 100335447 100335448 100335449 100335450 100335451 100335452 100335453 100335454 100335455 100335456 100335457 100335458 100335459 100335460 100335461 100335462 100335463 100335464 100335465 100335466 100335467 100335468 100335580 100335581 100335582 100335584 100335585 100335586 100335587 100335588 100335589 100335590 100335591 100335592 100335593 100335594 100335595 100335596 100335597 100335598 100335599 100335600 100335601 100335602 100335603 100335604 100335605 100335606 100335607 100335608 100335609 100335610 100335611 100335612 100335613 100335614 100335615 100335616 100335617 100335618 100335619 100335620 100335621 100336024 100336026 100336027 100336028 100336029 100336030 100336031 100336032 100336033 100336034 100336035 100336036 100336037 100336038 100336039 100336060 100336061 100336062 100336063 100336064 100336065 100336066 100336067 100336068 100336069 100336070 100336071 100336072 100336073 100336074 100336132 100336133 100336134 100336135 100336137 100336138 100336139 100336264 100336265 100336266 100336267 100336268 100336269 100336270 100336271 100336272 100336273 100336274 100336275 100336276 100336277 100336278 100336279 100336280 100336281 100336282 100336283 100336284 100336285 100336286 100336287 100336288 100336290 100336291 100336292 100336293 100336294 100336295 100336297 100336298 100336300 100336302 100336303 100336304 100336305 100336306 100336308 100336309 100336310 100336311 100336312 100336314 100336315 100336316 100336317 100336319 100336320 100336321 100336322 100336323 100336324 100336325 100336327 100336328 100336580 100336581 100336582 100336583 100336585 100336586 100336587 100336588 100336589 100336590 100336591 100336592 100336593 100336594 100336595 100336596 100336597 100336598 100336599 100336600 100336601 100336602 100336604 100336605 100336606 100336607 100336608 100336609 100336610 100336611 100336612 100336614 100336615 100336616 100336617 100336618 100336619 100336620 100336621 100336622 100336623 100336624 100336625 100336850 100336851 100336852 100336853 100336854 100336856 100336857 100336858 100336859 100336860 100336861 100336862 100336863 100336864 100336865 100336866 100336868 100336869 100336870 100336871 100336872 100336873 100336874 100336875 100336876 100336877 100336878 100336879 100336880 100336881 100336883 100336884 100336885 100336886 100336887 100336890 100336891 100336892 100336893 100336894 100336897 100337094 100337257 100337262 100337264 100337265 100337266 100337648 100337649 100337650 100337651 100337652 100337654 100338488 100338489 100338490 100338491 100338492 100338493 100338494 100338495 100338496 100338497 100338498 100338541 100338542 100338543 100338544 100338545 100338546 100338548 100338549 100338550 100338551 100338552 100338554 100338555 100338556 100338557 100338558 100338561 100338562 100338564 100338565 100338566 100338567 100338568 100338569 100338570 100338571 100338572 100338573 100338574 100338652 100338674 100338954 100338958 100338978 100339212 100339225 100339226 100339242 100339244 100339247 100339248 100339250 100339252 100339253 100339308 100339309 100339310 100339311 100339316 100339321 100339322 100339323 100339325 100339326 100339464 100339465 100339466 100339467 100339468 100339469 100339470 100339471
100339472 100339473 100339474 100339475 100339476 100339477 100339478 100339479 100339480 100339481 100339482 100339483 100339484 100339485 100339486 100339488 100339492 100339494 100339495 100339496 100339498 100339500 100339501 100339502 100339503 100339505 100339506 100339507 100339509 100339510 100339511 100339512 100339513 100339514 100339515 100339516 100339517 100339518 100339519 100339520 100339521 100339522 100339523 100339524 100339525 100339526 100339527 100339528 100339529 100339530 100339532 100339534 100339535 100339536 100339537 100339538 100339539 100339540 100339541 100339542 100339543 100339544 100339545 100339546 100339547 100339548 100339549 100339550 100339551 100339799 100339880 100339881 100339882 100339883 100339884 100339920 100339921 100339922 100339923 100339924 100339925 100339926 100339927 100339928 100339929 100339930 100339931 100339932 100339933 100339934 100339935 100339936 100339937 100339938 100339939 100339940 100339941 100339942 100339943 100339944 100339945 100339946 100339947 100339948 100339949 100339950 100339951 100339952 100339955 100339956 100339957 100339959 100339960 100339963 100339966 100339970 100339972 100339977 100339978 100339979 100339980 100339981 100339983 100340356 100340358 100340359 100340361 100340364 100340365 100340367 100340368 100340369 100340371 100340373 100340377 100340378 100340381 100340386 100340387 100340392 100340393 100340395 100340396 100340397 100340399 100340400 100340405 100340406 100340407 100340408 100340409 100340911 100340915 100341050 100341052 100341053 100341054 100341055 100341056 100341057 100341059 100341124 100341125 100341126 100341127 100341128 100341129 100341130 100341131 100341132 100341133 100341134 100341135 100341136 100341137 100341138 100341139 100341160 100341161 100341162 100341163 100341164 100341165 100341167 100341168 100341170 100341173 100341174 100341175 100341176 100341177 100341178 100341179 100341180 100341181 100341182 100341183 100341184 100341185 100341186 100341188 100341189 100341190 100341192 100341193 100341194 100341196 100341197 100341199 100341264 100341269 100341361 100341363 100341364 100341365 100341370 100341372 100341373 100341376 100341377 100341378 100341379 100341380 100341382 100341383 100341384 100341386 100341387 100341389 100341390 100341393 100341394 100341395 100341396 100341399 100341400 100341401 100341402 100341403 100341404 100341405 100341406 100341407 100341408 100341409 100341410 100341411 100341413 100341415 100341416 100341417 100341421 100341423 100341424 100341425 100341427 100341428 100341838 100341839 100341861 100341863 100341867 100341868 100341871 100341885 100341887 100350183 100350184 100350185 100350291 100350665 100350696 100350705 100350710 100350721 100350726 100350729 100350731 100350736 100350739 100350741 100350749 100350750 100350753 100350754 100350755 100350881 100350882 100350883 100350884 100350886 100350888 100350889 100350890 100350891 100350892 100350893 100350894 100350895 100350897 100350898 100350899 100350900 100350901 100350902 100350904 100350905 100350906 100350907 100350908 100350909 100350910 100350911 100350913 100350915 100350916 100350917 100350918 100350922 100350924 100350925 100350926 100350927 100350928 100350933 100351486 100351558 100352786 100352787 100352791 100352794 100352796 100352797 100352798 100352799 100352802 100352836 100352837 100352838 100352839 100352840 100352841 100352843 100352945 100352948 100352950 100353260 100353311 100353909 100353912 100353914 100353916 100353917 100353918 100354005 100354063 100354066 100354068 100354075 100354077 100354420 100354421 100354425 100354427 100354466 100354478 100354479 100354482 100354483 100354484 100354488 100354703 100354709 100354713 100354827 100354829 100354830 100354842 100354843 100354845 100354848 100354849 100354850 100354851 100354852 100354859 100354865 100354869 100354879 100354883 100354888 100354892 100355145 100355155 100355158 100355161 100355163 100355166 100355171 100355382 100355383 100355385 100355389 100355390 100355392 100355393 100355394 100355397 100355399 100355676 100355679 100355797 100355799 100355801 100355804 100355806 100355807 100355810 100355811 100355901 100355906 100355907 100355912 100355914 100355916 100355921 100355930 100355932 100355933 100355934 100356011 100356013 100356014 100356018 100356080 100356081 100356084 100356089 100356093 100356096 100356099 100356202 100356211 100356240 100356242 100356243 100356244 100356245 100356248 100356249 100356251 100356252 100356255 100356264 100356265 100356266 100356267 100356268 100356270 100356272 100356274 100356282 100356287 100356290 100356291 100356292 100357130 100357132 100357133 100357134 100357457 100357738 100357739 100357740 100357742 100357743 100357744 100357747 100357751 100357759 100357760 100357761 100357762 100357764 100357765 100357774 100357775 100357779 100357780 100357782 100357784 100357793 100357798 100357800 100358122 100358123 100358144 100358152 100358175 100358177 100358308 100358310 100358317 100358325 100358330 100358332 100358334 100358335 100358339 100358346 100358352 100359505 100359643 100359657 100359770 100359771 100359772 100359773 100359774 100359775 100359776 100359777 100359878 100359950 100359952 100359953 100359954 100360085 100360090 100360092 100360093 100360094 100360095 100360097 100360098 100360150 100360188 100360189 100360193 100360194 100360196 100360197 100360201 100360202 100360205 100360210 100360255 100360256 100360281 100360283 100360284 100360285 100360288 100360289 100360290 100360292 100360293 100360298 100360356 100360368 100361219 100361228 100361229 100361231 100361233 100361393 100361397 100361402 100361407 100361414 100361415 100361420 100361423 100361427 100361433 100361436 100361442 100361443 100361457 100361486 100361492 100361510 100361517 100361523 100361526 100361535 100361557 100361574 100361579 100361586 100361588 100361757 100361771 100361772 100362003 100362004 100362007 100362015 100362017 100362020 100362021 100362024 100362025 100362033 100362035 100362037 100362038 100362040 100362042 100362043 100362044 100362045 100362047 100362048 100362049 100362050 100362057 100362058 100362059 100362060 100362061 100362068 100362381 100362384 100362388 100362391 100362392 100362395 100362396 100362442 100362449 100362452 100362512 100362540 100362542 100362545 100362546 100362547 100363027 100363070 100363088 100363649 100363653 100364561 100364609 100365168 100365171 100365181 100365189 100365190 100365191 100365193 100365195 100365196 100365641 100365642 100365643 100365650 100365654 100365734 100365737 100365739 100366644 100366645 100366647 100366648 100366650 100366651 100366655 100366703 100366706 100366708 100366709 100366710 100366715 100366915 100366916 100366918 100366919 100366921 100366924 100366925 100366926 100366928 100366929 100366930 100366931 100366932 100366934 100366935 100366938 100366939 100366942 100366944 100366947 100366954 100366957 100367017 100367021 100367023 100367025 100367026 100367028 100367040 100367041 100367143 100367150 100367156 100367167 100367169 100367262 100367265 100367268 100367272 100367276 100368009 100368014 100368015 100368061 100368063 100368064 100368065 100368066 100368068 100368069 100368070 100368071 100368072 100368078 100368082 100368083 100368084 100368085 100368086 100368087 100368089 100368181 100368183 100368184 100368185 100368186 100368188 100368189 100368191 100368192 100368194 100368195 100368196 100368197 100368198 100368199 100368371 100368372 100368373 100368374 100368375 100368376 100368378 100368379 100368500 100368501 100368503 100368504 100368505 100368506 100368507 100368508 100368511 100368512 100368513 100368515 100368516 100368517 100368518 100368519 100368520 100368521 100368523 100368524 100368525 100368527 100368528 100368529 100368530 100368533 100368534 100368535 100368550 100368551 100368555 100368559 100368601 100368605 100368609 100368885 100370287 100370757 100370758 100370759 100370790 100370793 100370794 100370796 100370798 100370799 100370800 100370825 100370831 100370832 100370833 100370836 100370838 100370839 100370843 100370886 100370887 100370888 100370890 100370891 100370899 100370987 100370988 100370991 100370993 100370995 100370997 100371000 100371101 100371104 100371106 100371108 100371110 100371111 100371112 100371113 100371114 100371115 100371116 100371118 100371119 100371125 100371126 100371129 100371130 100371131 100371132 100371134 100371142 100371166 100371168 100371169 100371170 100371171 100371173 100371175 100371177 100371178 100371205 100371206 100371207 100371209 100371210 100371213 100371373 100371374 100371376 100371377 100371378 100371379 100371407 100371408 100371413 100371414 100371416 100371417 100371420 100371422 100371423 100371424 100371426 100371427 100371430 100371431 100371432 100371434 100371436 100371437 100371438 100371439 100371600 100371601 100371603 100371604 100371605 100371606 100371607 100371608 100371610 100371611 100371615 100371616 100371617 100371619 100371653 100371654 100371655 100371656 100371660 100371662 100371663 100371664 100371666 100371671 100371672 100371674 100371675 100371676 100371677 100371679 100371691 100371693 100371694 100371696 100371697 100371698 100371700 100371703 100371706 100371707 100372280 100372282 100372284 100372285 100372288 100372290 100372294 100372295 100372296 100372297 100372899 100373206 100373210 100373212 100373213 100373214 100373222 100373225 100373226 100373227 100373230 100373231 100373431 100373432 100373434 100373435 100373436 100373437 100373438 100373439 100373440 100373441 100373442 100373444 100373446 100373447 100373448 100373450 100373451 100373452 100373454 100373455 100373456 100373457 100373458 100373460 100373466 100373467 100373468 100373469 100373470 100373503 100373504 100373505 100373506 100373508 100373509 100373511 100373514 100373600 100373601 100373602 100373681 100373685 100373686 100373687 100375014 100375278 100375332 100376393 100376394 100376395 100376396 100376398 100376399 100376400 100376401 100376402 100376403 100376404 100376405 100376406 100376407 100376408 100376409 100376411 100376412 100377481 100377482 100377483 100377484 100377485 100377486 100377487 100377488 100377489 100377490 100377491 100377492 100377493 100377494 100377495 100377496 100377497 100377498 100377499 100377541 100377542 100377543 100377544 100377545 100377546 100377547 100377548 100377549 100377551 100377552 100377553 100377554 100377555 100377556 100377557 100377559 100377580 100377720 100377721 100377722 100377723 100377724 100377725 100377727 100377728 100377729 100377730 100377731 100377732 100377733 100377734 100377735 100377736 100377737 100377738 100377739 100377740 100377741 100377742 100377743 100377744 100377745 100377746 100377747 100377748 100377749 100377878 100377879 100377882 100377883 100377884 100377886 100377890 100377897 100377899 100377900 100377901 100377902 100377903 100377905 100378900 100380448 100380449 100380452 100380453 100380454 100380456 100380457 100380651 100380652 100380653 100380654 100380655 100380656 100380780 100380782 100380783 100380784 100380785 100380786 100380787 100380788 100380789 100380790 100380791 100380792 100380793 100380794 100380795 100380796 100380797 100380799 100380800 100380801 100380802 100380803 100380804 100380805 100380806 100380807 100380808 100380810 100380824 100380831 100380832 100380833 100380834 100380835 100380836 100380837 100380838 100380839 100380900 100380901 100380902 100380903 100380904 100380905 100380906 100380907 100380908 100380909 100380910 100380911 100380912 100380913 100380914 100380915 100380916 100380917 100380918 100380919 100380920 100380921 100380922 100380924 100380925 100380926 100380927 100380928 100380962 100380963 100380966 100380967 100380968 100380969 100380970 100380974 100380976 100380977 100380978 100380979 100380980 100380981 100380985 100380986 100380987 100380988 100380991 100380992 100380993 100380994 100380996 100380998 100380999 100381000 100381003 100381274 100381424 100381730 100381854 100382194 100382202 100382203 100382204 100382205 100382206 100382207 100382208 100382209 100383183 100383184 100383185 100383187 100383188 100383189 100383190 100383191 100383192 100383193 100383194 100383285 100383286 100383287 100383288 100383289 100383290 100383291 100383293 100383294 100383296 100383297 100383298 100383299 100383320 100383321 100383323 100383324 100383325 100383328 100383329 100383330 100383331 100383333 100383334 100383335 100383336 100383337 100383338 100383339 100383340 100383341 100383342 100383343 100383344 100383345 100383346 100383347 100383348 100383349 100383350 100383351 100383352 100383353 100383354 100383355 100383356 100383357 100383358 100383359 100383361 100383362 100383363 100383364 100383365 100383368 100383370 100383371 100383373 100383374 100383375 100383376 100383377 100383378 100383379 100383381 100383382 100383383 100383385 100383386 100383387 100383388 100383389 100383390 100383391 100383392 100383393 100383394 100383396 100383397 100383398 100383401 100383404 100383405 100383407 100383409 100383412 100383620 100383621 100383623 100383624 100383626 100383629 100383630 100383631 100383632 100383634 100383636 100383637 100383639 100383640 100383643 100383644 100383645 100383648 100383651 100383652 100383894 100383895 100383896 100383897 100383906 100383927 100383935 100383981 100383982 100383983 100383986 100383988 100383989 100383990 100383991 100383992 100383993 100383994 100383995 100383996 100383998 100383999 100384080 100384083 100384084 100384086 100384096 100384099 100384223 100384224 100384231 100384233 100384234 100384236 100384244 100384245 100384249 100384250 100384253 100384254 100384258 100384280 100384288 100384290 100384770 100384803 100384839 100385322 100385954 100387125 100387144 100387170 100387275 100387311 100387319 100387320 100387526 100387538 100387543 100387558 100387823 100387828 100387832 100387833 100387836 100387841 100387844 100387846 100387847 100387850 100387852 100387853 100387856 100387858 100387861 100387867 100387883 100387942 100387956 100387957 100387959 100388062 100388067 100388070 100388071 100388074 100388076 100388077 100388082 100388083 100388089 100388093 100388096 100388107 100388109 100388595 100388599 100388620 100388621 100388627 100388632 100388633 100388635 100388636 100388637 100388639 100388642 100388645 100388647 100388650 100388658 100388660 100388662 100388663 100388664 100388666 100388667 100388671 100388672 100388673 100388674 100388677 100388681 100388683 100388685 100388690 100388692 100388694 100388695 100388696 100388697 100388698 100388699 100388700 100388701 100388702 100388703 100388704 100388705 100388706 100388707 100388708 100388709 100388710 100388711 100388712 100388713 100388715 100388716 100388717 100388718 100388719 100388730 100388732 100388736 100388738 100388741 100389229 100389230 100389232 100389233 100389236 100389239 100389380 100389401 100389403 100389406 100389411 100389414 100389420 100389422 100389425 100389428 100389429 100389430 100389433 100389438 100389440 100389441 100389468 100389470 100389474 100389477 100389480 100389487 100389494 100389498 100389994 100389995 100389996 100389997 100389999 100390084 100390085 100390086 100390087 100390122 100390128 100390132 100390133 100390134 100390135 100390136 100390137 100390138 100390180 100390181 100390184 100390185 100390186 100390187 100390188 100390189 100390190 100390193 100390195 100390196 100390200 100390406 100390408 100390409 100390418 100390419 100390485 100390486 100390493 100390494 100390496 100390498 100391054 100391378 100391470 100392259 100392612 100392631 100392639 100393578 100393655 100393659 100393714 100393864 100393896 100394375 100394391 100394398 100394460 100394499 100394500 100394501 100394503 100394504 100394511 100394519 100394522 100394526 100394527 100394528 100394532 100394533 100394534 100394536 100394537 100394538 100394539 100394541 100394542 100394543 100394545 100394546 100394549 100394966 100394967 100394970 100394971 100394975 100394976 100394977 100394979 100394982 100395070 100395071 100395073 100395076 100395077 100395079 100395322 100395324 100395325 100395326 100395328 100395332 100395334 100395335 100395336 100395338 100395339 100395341 100395349 100395352 100395407 100395409 100395410 100395412 100395414 100395415 100395419 100395561 100395563 100395565 100395573 100395574 100395575 100395585 100395586 100395587 100395591 100395594 100395595 100395596 100395599 100395603 100395784 100395787 100395788 100395789 100395792 100395793 100395795 100395796 100395798 100395980 100395981 100395982 100395983 100395984 100395985 100395990 100396622 100396624 100396625 100396765 100396766 100396769 100396773 100396774 100396775 100396778 100396801 100396805 100396807 100396809 100396812 100396814 100396819 100396820 100396821 100396822 100396823 100396824 100396825 100396826 100396827 100396828 100396830 100396831 100396832 100396833 100396834 100396835 100396836 100396841 100396843 100396846 100396848 100396849 100396851 100396852 100396853 100396854 100396858 100396859 100396860 100396863 100396864 100396865 100396867 100396868 100396874 100397383 100397384 100397385 100397386 100397387 100397388 100397389 100397390 100397391 100397392 100397394 100397395 100397397 100397464 100397465 100397467 100397469 100397474 100397476 100397478 100397479 100397480 100397482 100397484 100397485 100397486 100397487 100397491 100397492 100397493 100397495 100397499 100397500 100397502 100397503 100397504 100397511 100397513 100397514 100397515 100397516 100397520 100397523 100397524 100397526 100397528 100397529 100397530 100397531 100397532 100397533 100397534 100397536 100397672 100397674 100397678 100397841 100397842 100397844 100397845 100397846 100397847 100397848 100397849 100397852 100397857 100397859 100398206 100398211 100398213 100398217 100398218 100398363 100398365 100398367 100398368 100398370 100398371 100398374 100398398 100398421 100398426 100398434 100398440 100398441 100398442 100398444 100398447 100398449 100398514 100398515 201002903 201002905 201002906 201002907 201002908 201002911 201002913 201002914 201003088 201003090 201003093 201003097 201003102 201003104 201003105 201003106 201003108 201003109 201003369 201003372 201003373 201003378 201003379 201003384 201003392 201003394 201003401 201003405 201003410 201003411 201003412 201003413 201003422 201003423 201003425 201003428 201003430 201003436 201003439 201003445 201003461 201003936 201003937 201003938 201003939 201003941 201003942 201003943 201003952 201003959 201003960 201003961 201004009 201004010 201004011 201004012 201004013 201004014 201004015 201004016 201004017 201004018 201004019 201004020 201004022 201004023 201004050 201004053 201004247 201004249 201004252 201004253 201004255 201004257 201004258 201004357 201004368 201004377 201004379 201004382 201004383 201004385 201004386 201004683 201004685 201004688 201004689 201004690 201006541 201006543 201006940 201006941 201006942 201006943 201006944 201007444 201007446 201007448 201007452 201007453 201007454 201007527 201007539 201007560 201007561 201007562 201007563 201007564 201007565 201007581 201007585 201007610 201007626 201007628 201007629 201007632 201007633 201007634 201007644 201007646 201007654 201007657 201007658 201007659 201007660 201007661 201007666 201007667 201007672 201007673 201007674 201007675 201007676 201007677 201007781 201007782 201007783 201007784 201007785 201007786 201007787 201007788 201007789 201007790 201007791 201007792 201007793 201007794 201007795 201007796 201007797 201007798 201007799 201007805 201007806 201007807 201007808 201007809 201007811 201007812 201007813 201007814 201007815 201007816 201007817 201007818 201007819 201007820 201007821 201007822 201007823 201007824 201007825 201007826 201007827 201007828 201007829 201007830 201007831 201007832 201007833 201007834 201007840 201007842 201007843 201007844 201007845 201007846 201007847 201007848 201007849 201007850 201007851 201007852 201007853 201007854 201007855 201007856 201007857 201007858 201007859 201007860 201007861 201007862 201007863 201007864 201007866 201007867 201007868 201007869 201007880 201007881 201007882 201007883 201007884 201007885 201007886 201007887 201007888 201007889 201007890 201007891 201007892 201007893 201007894 201007895 201007896 201007897 201007898 201007899 201007900 201007901 201007902 201007903 201007904 201007905 201007906 201007907 201007908 201007909 201007910 201007911 201007912 201007913 201007914 201007915 201007920 201007921 201007922 201007923 201007924 201007925 201007926 201007927 201007928 201007929 201007930 201007931 201007932 201007933 201007934 201007935 201007936 201007937 201007938 201007939 201007940 201007941 201007942 201007943 201007944 201007945 201007960 201007961 201007962 201007966 201007967 201007968 201007969 201007970 201007971 201007972 201007973 201007974 201007975 201007976 201007977 201007978 201007979 201007980 201007981 201007982 201007983 201007984 201007985 201007987 201007988 201007989 201007990 201007991 201007992 201007993 201007994 201007995 201007996 201007997 201007998 201007999 201008001 201008002 201008003 201008004 201008005 201008006 201008007 201008008 201008009 201008010 201008011 201008018 201008042 201008052 201008065 201008071 201009205 201009290 201009291 201009292 201009293 201009294 201009295 201009296 201009297 201009298 201009299 201009359 201009473 201009474 201009475 201009476 201009477 201009478 201009479 201009500 201009501 201009502 201009503 201009504 201009505 201009506 201009507 201009508 201009509 201009510 201009511 201009513 201009514 201009515 201009516 201009520 201009521 201009522 201009523 201009540 201009541 201009542 201009543 201009544 201009545 201009546 201009547 201009574 201009575 201009576 201009577 201009578 201009579 201009580 201009581 201009582 201009583 201009584 201009585 201009586 201009622 201009627 201009731 201009732 201009733 201009734 201009735 201009736 201009737 201009738 201009739 201009760 201009761 201009762 201009763 201009764 201009801 201009802 201009803 201009804 201009805 201009806 201009807 201009808 201009809 201009810 201009811 201009812 201009813 201009814 201009815 201009866 201009867 201009868 201009876 201009877 201009878 201009879 201009950 201009951 201009952 201009953 201009954 201009955 201009956 201009957 201009958 201009959 201009960 201009961 201009963 201009964 201009966 201009968 201009969 201009970 201009972 201009973 201009976 201009977 201009978 201009979 201009984 201009985 201009989 201010000 201010007 201010008 201010009 201010010 201010011 201010012 201010013 201010014 201010015 201010016 201010017 201010018 201010019 201010045 201010114 201010115 201010118 201010119 201010125 201010128 201010129 201010130 201010131 201010132 201010135 201010136 201010137 201010139 201010140 201010141 201010142 201010143 201010144 201010145 201010146 201010147 201010148 201010150 201010152 201010163 201010167 201010169 201010170 201010172 201010174 201010176 201010177 201010178 201010179 201010186 201010187 201010188 201010190 201010192 201010194 201010196 201010197 201010198 201010212 201010213 201010214 201010215 201010220 201010240 201010242 201010243 201010244 201010246 201010248 201010264 201010265 201010266 201010267 201010268 201010269 201010270 201010271 201010272 201010280 201010281 201010282 201010283 201010304 201010305 201010306 201010307 201010319 201010342 201010343 201010345 201010346 201010347 201010348 201010349 201010350 201010351 201010352 201010353 201010354 201010355 201010356 201010357 201010358 201010359 201010361 201010362 201010363 201010368 201010376 201010377 201010380 201010381 201010382 201010383 201010384 201010385 201010441 201010442 201010519 201010523 201010568 201010572 201010573 201010574 201010575 201010576 201010580 201010581 201010583 201010584 201010585 201010586 201010591 201010592 201010593 201010594 201010595 201010597 201010600 201010605 201010606 201010610 201010611 201010612 201010613 201010614 201010615 201010616 201010618 201010619 201010760 201010761 201010762 201010763 201010764 201010765 201010766 201010767 201010768 201010769 201010770 201010771 201010772 201010774 201010775 201010776 201010777 201010781 201010782 201010783 201010784 201010788 201010790 201010792 201010793 201010794 201010800 201010801 201010803 201010820 201010829 201010838 201010880 201010893 201010894 201010895 201010897 201010899 201010924 201011169 201012772 201012773 201012774 201012775 201012776 201012778 201012779 201012888 201012889 201012890 201012891 201012892 201012893 201012894 201012895 201012896 201012897 201012898 201012919 201012920 201012921 201012922 201012923 201012924 201012925 201012926 201012927 201012928 201012929 201012930 201012931 201012932 201012933 201012934 201012935 201012936 201012937 201012938 201012939 201012940 201012941 201012942 201012943 201012944 201012945 201012946 201012947 201012948 201012974 201012975 201012976 201012977 201012978 201012979 201012980 201012981 201012983 201012984 201012985 201012986 201012987 201012988 201012989 201012991 201012992 201012993 201012994 201012995 201012996 201013040 201013042 201013043 201013044 201013045 201013047 201013048 201013049 201013050 201013052 201013053 201013054 201013055 201013056 201013057 201013058 201013059 201013060 201013061 201013062 201013085 201013086 201013088 201013089 201013090 201013091 201013093 201013094 201013095 201013096 201013097 201013098 201013099 201013100 201013102 201013103 201013104 201013105 201013106 201013107 201013108 201013109 201013110 201013111 201013112 201013113 201013115 201013116 201013118 201013119 201013120 201013123 201013124 201013125 201013126 201013127 201013129 201013130 201013141 201013142 201013143 201013144 201013145 201013146 201013147 201013148 201013149 201013157 201013160 201013161 201013162 201013165 201013648 201013792 201013824 201014000 201014319 201014367 201014368 201014369 201014370 201014371 201014372 201014400 201014401 201014402 201014403 201014404 201014405 201014406 201014407 201014408 201014409 201014410 201014411 201014412 201014413 201014414 201014415 201014416 201014417 201014418 201014433 201014434 201022723 201022726 201022733 201022735 201022736 201022760 201022762 201022764 201022767 201022769 201022770 201022962 201022964 201022965 201022970 201022971 201022972 201022973 201022975 201022977 201022981 201022982 201022986 201022988 201022990 201022992 201022994 201022997 201023000 201023001 201023002 201023003 201023004 201023005 201023047 201023050 201023051 201023054 201023056 201023058 201023080 201023081 201023082 201023084 201023085 201023088 201023093 201023095 201023104 201023106 201023107 201023108 201023111 201023266 201023267 201023269 201023273 201023274 201023275 201023277 201023627 201023628 201023629 201023630 201023631 201023632 201023633 201023634 201023635 201023636 201023637 201023638 201023639 201023675 201023833 201023834 201023835 201023836 201023838 201023839 201023840 201023842 201023848 201023851 201023853 201023854 201023855 201023857 201023859 201023860 201023865 201023866 201023867 201023869 201023870 201023872 201023873 201023874 201023875 201023877 201023878 201023880 201023882 201023884 201023891 201023893 201023894 201023895 201023897 201023901 201023903 201023904 201023906 201023907 201023908 201023909 201023910 201023912 201023913 201023915 201023916 201023917 201023918 201023919 201023920 201023921 201023923 201023924 201023925 201023927 201023930 201023993 201023994 201023995 201023996 201023997 201023998 201023999 201024001 201024002 201024007 201024009 201024011 201024012 201024013 201024017 201024018 201024019 201024020 201024021 201024022 201024023 201024024 201024025 201024026 201024027 201024028 201024029 201024031 201024032 201024034 201024040 201024041 201024042 201024043 201024044 201024045 201024046 201024047 201024048 201024049 201024050 201024051 201024052 201024053 201024054 201024055 201024056 201024057 201024059 201024060 201024062 201024064 201024065 201024066 201024067 201024068 201024069 201024071 201024072 201024074 201024076 201024080 201024081 201024082 201024083 201024084 201024087 201024088 201024089 201024090 201025180 201025185 201025194 201025265 201025267 201025268 201025269 201025270 201025271 201025272 201025273 201025275 201025276 201025277 201025278 201025279 201025296 201025305 201026250 201026253 201026413 201026416 201026430 201026447 201026448 201026454 201026458 201026460 201026464 201026467 201026469 201026473 201026474 201026479 201026510 201026511 201026512 201026513 201026514 201026515 201026517 201026518 201026519 201026520 201026526 201026527 201026528 201026529 201026530 201026535 201026536 201026537 201026538 201026539 201026540 201026541 201026542 201026543 201026544 201026545 201026546 201026547 201026548 201026549 201026550 201026551 201026553 201026554 201026555 201026556 201026557 201026558 201026559 201026564 201026565 201026566 201026571 201026580 201026581 201026583 201026584 201026585 201026587 201026588 201026589 201026590 201026591 201026593 201026594 201026595 201026596 201026597 201026599 201026603 201026604 201026606 201026607 201026608 201026610 201026611 201026612 201026613 201026654 201026664 201026666 201026670 201026683 201026713 201027192 201027193 201027194 201027195 201027196 201027197 201027198 201027199 201027306 201027311 201027312 201027313 201027315 201027316 201027317 201027319 201027407 201027408 201027409 201027410 201027411 201027412 201027413 201027414 201027415 201027416 201027417 201027418 201027419 201027421 201027422 201027423 201027424 201027427 201027428 201027429 201027430 201027431 201027435 201027438 201027441 201027442 201027443 201027444 201027445 201027446 201027447 201027448 201027449 201027450 201027451 201027452 201027453 201027454 201027455 201027476 201027477 201027478 201027479 201027500 201027501 201027502 201027503 201027504 201027505 201027506 201027507 201027509 201027512 201027518 201027519 201027521 201027522 201027526 201027527 201027529 201027531 201027582 201027583 201027592 201027593 201027595 201027596 201027597 201027598 201027599 201027600 201027602 201027603 201027604 201027606 201027607 201027608 201027609 201027610 201027611 201027612 201027613 201027614 201027615 201027680 201027681 201027682 201027683 201027684 201027686 201027687 201027689 201027690 201027691 201027692 201027693 201027694 201027695 201027696 201027699 201027700 201027701 201027702 201027703 201027704 201027705 201027706 201027707 201027708 201027709 201027710 201027711 201027712 201027713 201027714 201027715 201027716 201027744 201027745 201027746 201027747 201027748 201027749 201027750 201027751 201027752 201027753 201027756 201027758 201027759 201027760 201027761 201027762 201027763 201027764 201027765 201027766 201027767 201027768 201027769 201027770 201027771 201027772 201027773 201027774 201027775 201027776 201027777 201027778 201027779 201027780 201027781 201027782 201027783 201027800 201027801 201027803 201027804 201027806 201027807 201027810 201027811 201027812 201027813 201027814 201027816 201027818 201027820 201027827 201027828 201027829 201027868 201027869 201027872 201027879 201027884 201027907 201028008 201028011 201028014 201028019 201028023 201028032 201028036 201028038 201028039 201028040 201028041 201028042 201028043 201028287 201028291 201028482 201028484 201028485 201028487 201028488 201028489 201028491 201028492 201028493 201028494 201028498 201028520 201028524 201028529 201028530 201028533 201028534 201028535 201028541 201028552 201028556 201028557 201028600 201028601 201028602 201028604 201028607 201028608 201028610 201028615 201028622 201028681 201028796 201028797 201028799 201028901 201028902 201028906 201028907 201028908 201028910 201028912 201028913 201028914 201028915 201028918 201028919 201028920 201028922 201028923 201028924 201028926 201028927 201028929 201028931 201028933 201028934 201028935 201028938 201028939 201028940 201028941 201028942 201028943 201028944 201028945 201028946 201028947 201028949 201028963 201028964 201028965 201028966 201028967 201028969 201028970 201028971 201028972 201028973 201028974 201028976 201028978 201028979 201028981 201028982 201028983 201028985 201028987 201028993 201030432 201030436 201030723 201030724 201030725 201030726 201030727 201030729 201030730 201030731 201030732 201030733 201030734 201030735 201030736 201030737 201030747 201030748 201030754 201030756 201030757 201030764 201030769 201030776 201030777 201030779 201030780 201030781 201030784 201030785 201030786 201030787 201030789 201030792 201030794 201030795 201030799 201030828 201030832 201030833 201030834 201030835 201030836 201030838 201030841 201030842 201030843 201030846 201030860 201030861 201030862 201030864 201030865 201030883 201030887 201030888 201030892 201030893 201030894 201030895 201030898 201030905 201030916 201030919 201030921 201030925 201030927 201030929 201030930 201030941 201030943 201030944 201030948 201030954 201030958 201030960 201030962 201031061 201031065 201031066 201031069 201031072 201031076 201031077 201031078 201031079 201031154 201031155 201031156 201031160 201031162 201031194 201031203 201031208 201031209 201031211 201031216 201031219 201031220 201031222 201031223 201031374 201031375 201031376 201031377 201031378 201031379 201031453 201031457 201031536 201031537 201031539 201031598 201031599 201031640 201031641 201031642 201031643 201031676 201031677 201031679 201031681 201031683 201031685 201031689 201031692 201031693 201031694 201031696 201031698 201031699 201031700 201031701 201031702 201031703 201031710 201031711 201031712 201031713 201031748 201031793 201031794 201031795 201031796 201031797 201031798 201031799 201031818 201031819 201031857 201031858 201031861 201031865 201031866 201031868 201031871 201031872 201031874 201031875 201031878 201031879 201031881 201031882 201031883 201031884 201031885 201031886 201031887 201031888 201031889 201031890 201031891 201031900 201031901 201031902 201031903 201031904 201031905 201031906 201031907 201031908 201031920 201031921 201031922 201031923 201031924 201031925 201031926 201031927 201031928 201031929 201031930 201031931 201031932 201031935 201031936 201031940 201031941 201031942 201031943 201031944 201031945 201031946 201031947 201031948 201031949 201031950 201031951 201031952 201031953 201031954 201031955 201031956 201031957 201031958 201031959 201031960 201031961 201031962 201031963 201031964 201031965 201031966 201031967 201031968 201031969 201031970 201031971 201031972 201031973 201031974 201031975 201031976 201031977 201031978 201031979 201031980 201031981 201031982 201031983 201031984 201031985 201031986 201031988 201031989 201031990 201031991 201031992 201031993 201031994 201031995 201031996 201031997 201031998 201031999 201032000 201032001 201032002 201032003 201032004 201032005 201032006 201032007 201032070 201032160 201032162 201032165 201032166 201032167 201032168 201032169 201032170 201032172 201032173 201032176 201032177 201032179 201032180 201032181 201032182 201032187 201032248 201033000 201033210 201033260 201033267 201033273 201033293 201033301 201033302 201033303 201033304 201033306 201033307 201033308 201033309 201033310 201033323 201033328 201033340 201033342 201033349 201033375 201033378 201033400 201033401 201033403 201033405 201033406 201033407 201033408 201033409 201033410 201033411 201033412 201033413 201033414 201033415 201033416 201033418 201033419 201033420 201033421 201033422 201033423 201033424 201033425 201033426 201033427 201033428 201033429 201033430 201033432 201033433 201033434 201033435 201033560 201033561 201033562 201033563 201033564 201033565 201033566 201033567 201033568 201033569 201033581 201033582 201033583 201033584 201033585 201033586 201033614 201033615 201033617 201033619 201033620 201033628 201033707 201033747 201033785 201033809 201033817 201033843 201033850 201033851 201033852 201033853 201033855 201033857 201033867 201033874 201033880 201033882 201033883 201033885 201033886 201033887 201033888 201033889 201033890 201033891 201033892 201033893 201033894 201033895 201033896 201033897 201033898 201033899 201033940 201033941 201033942 201033943 201033944 201033945 201033946 201033947 201033948 201033949 201033950 201033951 201033952 201033953 201033954 201033955 201033956 201033957 201033958 201033959 201033960 201033961 201033962 201033963 201033964 201033965 201033966 201033967 201033968 201033969 201033970 201033971 201033972 201033973 201033974 201033975 201033976 201033977 201033978 201033979 201034001 201034002 201034003 201034004 201034005 201034006 201034007 201034008 201034010 201034011 201034012 201034024 201034025 201034026 201034027 201034028 201034029 201034030 201034031 201034032 201034033 201034034 201034035 201034037 201034084 201023674 201025176 201025190 201025193 201025197 100046717 100094146 100144967 100163931 100164000 100164082 100169236 100179412 100182467 100228837 100237636 100246516 100247532 100259724 201007836 201007837 201007838 201007839 201007965 201007986 201008012 201008013 201008014 201008015 201008016 201008017 201008019 201008020 201008021 201008022 201008023 201008024 201008040 201008041 201008043 201008044 201008045 201008046 201008047 201008048 201008049 201008050 201008051 201008053 201008055 201008057 201008059 201008061 201008062 201008063 201008064 201008066 201008067 201008068 201008069 201008072 201009512 201022969 201026444 201026516 201026582 201026586 201031221 |
Recalling Firm/ Manufacturer |
Respironics California, LLC 2271 Cosmos Ct Carlsbad CA 92011-1517
|
| For Additional Information Contact | 800-722-9377 |
Manufacturer Reason for Recall | Potential issues with ventilator navigation ring interface and liquid ingress:
1) The navigation ring may become inoperative or operate intermittently;
2) The navigation ring may cause ventilator parameter values to change erratically during setting entries and adjustments. |
FDA Determined Cause 2 | Device Design |
| Action | On 11/02/2020, the firm sent an "URGENT-MEDICAL DEVICE CORRECTION" Notification Letter via priority mail informing customer that the V60 Ventilators may experience two potential issues with the navigation ring interface (see table below for additional information).
1. The navigation ring may become inoperative or begin operating intermittently.
2. The navigation ring may not behave as expected, making it difficult for the user to change ventilator settings via the touchscreen.
and if clinicians accept an incorrect setting and leaves the patient on the ventilator without responding to the resulting alarms, the patient could sustain serious injuries.
Customers are instructed:
1. When cleaning V60 ventilators, do not allow cleaning agents to pool on the
ventilator surfaces. This is especially important on the device interface.
2. Check setting changes before and after accepting the changes.
3. If the navigation ring becomes inoperative or operates intermittently during
clinical use, use the touchscreen for all settings changes and navigation,
and contact Philips to schedule service following clinical use.
4. If the navigation ring interferes with your ability to adjust therapy settings via
touchscreen, remove the ventilator from service and contact Philips for
repair.
and to please review the following sections of the user manual:
1. Chapter 10: Care and maintenance. To prevent possible damage to the
ventilator when cleaning, use only those cleaning and disinfecting agents
listed, and do not drip or spray any liquids directly onto any ventilator
surface.
2. Chapter 1: Warnings, cautions, and notes. During clinical use of the
ventilator, monitor patients for changes in respiratory status with each
ventilator setting adjustment. Use an external oxygen monitor to verify
oxygen concentration, and have an alternative means of ventilation available
whenever a ventilator is in use.
Customers are also instructed to complete, sign and return the Acknowledgement and Receipt |
| Quantity in Commerce | 10,291 systems |
| Distribution | Worldwide distribution - US Nationwide including Guam and Puerto Rico; Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = MNT
|
|
|
|