Class 2 Device Recall Trilogy Evo Universal

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Trilogy Evo Universal see related information
Date Initiated by Firm July 26, 2021
Create Date August 13, 2021
Recall Status1 Open3, Classified
Recall Number Z-2290-2021
Recall Event ID 88333
510(K)Number K181170  
Product Classification Ventilator, continuous, facility use - Product Code CBK
Product Trilogy Evo Universal, Product number DS2000X11B
Code Information Software version numbers 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06, 1.05.01, and 1.06.02
Recalling Firm/
Manufacturer
Philips Respironics, Inc.
1001 Murry Ridge Ln
Murrysville PA 15668-8517
For Additional Information Contact Philips Customer Service
800-722-9377
Manufacturer Reason
for Recall
Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).
FDA Determined
Cause 2
Labeling design
Action On July 26, 2021, the firm issued Urgent Medical Device Correction letters to affected customers. Customers were advised of the two software issues that the firm had identified related to pressure increase. Customers received guidance on how to identify affected products: Devices with software versions devices with software versions 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06, 1.05.01 and 1.06.02 are impacted. Customers were provided with actions to take until the upcoming software fix is implemented on your device. Philips will be releasing a software correction to resolve this issue, and will contact customers when the software correction is released. If you need any further information or support concerning this issue, please contact Philips Customer Care Solutions Center Trilogy Evo: +1-800-345-6443 Option 2, Option 1 (Available 24/7) Trilogy EV300: +1-800-722-9377 (Available 24/7)
Quantity in Commerce 10,000 (FEMA Stockpile Only)
Distribution Worldwide distribution - US nationwide distribution.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database 510(K)s with Product Code = CBK and Original Applicant = Respironics Inc.
-
-