Class 2 Device Recall Philips

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Philips see related information
Date Initiated by Firm June 08, 2021
Create Date August 18, 2021
Recall Status1 Terminated 3 on August 10, 2023
Recall Number Z-2316-2021
Recall Event ID 88336
510(K)Number K160743  
Product Classification System, x-ray, tomography, computed - Product Code JAK
Product 728327 Ingenuity CT Upgrades-Computed Tomography X-ray system
Code Information Serial Number: 300099 300106 300143 Software Version 4.1.10 SynchRight Option
Recalling Firm/
Manufacturer
Philips North America Llc
222 Jacobs St
Cambridge MA 02141-2289
Manufacturer Reason
for Recall
When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
FDA Determined
Cause 2
Software design
Action Philips issued Urgent Medical Device Correction letter dated 6/8/21 to consignees via certified mail. Letter states reason for recall, health risk and action to take: following steps: Go to Preferences> Regional Settings > Weight units, select the Weight units to Kg. Step 1: Verify that the patient weight unit is set to Kg, rather than Lbs. Step 2: Change the weight unit to Kg if weight unit is set to Lbs, referring to the screenshot below Step 3: Log out and then log back in to make the changes effective. Please share this notice with all those who need to be aware within your organization. Please complete and return the attached response form to Philips Healthcare promptly upon receipt and no later than 30 days from receipt, via email to CTNM.QARA@PHILIPS.COM The software update is planned to be released in the 3rd quarter of 2021. Once the software update is released, a Philips representative will contact you to schedule a software update to permanently resolve the issue. For North America, contact the Customer Care Solutions Center (1-800-722-9377).
Quantity in Commerce 3 units
Distribution US Nationwide distribution in the states of AZ, FL, LA, MN, NE, NV, NY, PA.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = JAK and Original Applicant = Philips Medical Systems (Cleveland), Inc.
-
-