| | Class 1 Device Recall CADDSolis VIP ambulatory infusion pump |  |
| Date Initiated by Firm | March 24, 2020 |
| Date Posted | October 19, 2021 |
| Recall Status1 |
Terminated 3 on December 22, 2023 |
| Recall Number | Z-0107-2022 |
| Recall Event ID |
88701 |
| 510(K)Number | K111275 |
| Product Classification |
Pump, infusion - Product Code FRN
|
| Product | CADD-Solis VIP ambulatory infusion pump, Model Numbers:
a. 21-2120-0105-02L
b. 21-2120-0105-03L
c. 21-2120-0105-06L
d. 21-2120-0105-07L
e. 21-2120-0105-08L
f. 21-2120-0105-12L
g. 21-2120-0105-13L
h. 21-2120-0105-14L
i. 21-2120-0105-15L
j. 21-2120-0105-17L
k. 21-2120-0105-50L
l. 21-2127-0105-02L
m. 21-2127-0105-06L
n. 21-2127-0105-50L |
| Code Information |
a. 21-2120-0105-02L, Serial Numbers: 1171502, 1171536, 1171537, 1171538, 1171539, 1171540, 1171541, 1171542, 1171543, 1171544, 1171545, 1171546, 1171547, 1171548, 1171549, 1171550, 1171551, 1171552, 1171553, 1171554, 1171555, 1171556, 1171557, 1171558, 1171559, 1171560, 1171561, 1171562, 1171563, 1171564, 1171565, 1171566, 1171567, 1171568, 1171569, 1171570, 1171571, 1171572, 1171573, 1171574, 1171575, 1171576, 1171577, 1171578, 1171579, 1171580, 1171581, 1171582, 1171583, 1171584, 1171632, 1171633, 1171649, 1171650, 1171665, 1171666, 1171667, 1171668, 1171669, 1171670, 1171671, 1171672, 1171673, 1171674, 1171675, 1171676, 1171677, 1171678, 1171679, 1171680, 1171681, 1171682, 1171683, 1171700, 1171701, 1171702, 1171703, 1171704, 1178267, 1178268, 1178269, 1178270, 1178271, 1178272, 1178273, 1178274, 1178275, 1178277, 1178278, 1178279, 1178280, 1178281, 1178282, 1178283, 1178284, 1178285, 1178286, 1178287, 1178288, 1178289, 1178291, 1178292, 1178293, 1178294, 1178295, 1178296, 1178297, 1178298, 1178299, 1178300, 1178301, 1178302, 1178303, 1178304, 1178305, 1178306, 1178307, 1178308, 1178309, 1178310, 1178311, 1178312, 1178313, 1178314, 1178315, 1178316, 1178317, 1178318, 1178319, 1178320, 1178321, 1178322, 1178323, 1178324, 1178325, 1178326, 1178327, 1178328, 1178329, 1178330, 1178331, 1178332, 1178333, 1178334, 1178335, 1178336, 1178337, 1178338, 1178339, 1178340, 1178341, 1178342, 1178343, 1178344, 1178345, 1178346, 1178347, 1178348, 1178349, 1178350, 1178351, 1178352, 1178353, 1178899, 1178900, 1178901, 1178902, 1178903, 1178904, 1178905, 1178907, 1178909, 1178910, 1178911, 1178912, 1178913, 1178914, 1178915, 1178916, 1178917, 1178918, 1178919, 1178920, 1178921, 1178922, 1178923, 1178924, 1178926, 1178927, 1178928, 1178929, 1178930, 1178931, 1178933, 1178934, 1178935, 1178936, 1178938, 1178939, 1178940, 1178942, 1178943, 1178945, 1178946, 1178947, 1178948, 1178949, 1178950, 1178951, 1178953, 1178956, 1178957, 1178958, 1178960, 1178961, 1178962, 1178963, 1178964, 1178965, 1178966, 1178967, 1178968, 1178969, 1178970, 1178971, 1178972, 1178973, 1178974, 1178975, 1178976, 1178977, 1178978, 1178980, 1179160, 1179162, 1179163, 1179164, 1179166, 1179167, 1179169, 1179170, 1179172, 1179173, 1179174, 1179175, 1179176, 1179178, 1179180, 1179181, 1179182, 1179185, 1179186, 1179187, 1179188, 1179189, 1179191, 1179192, 1179193, 1179195, 1179196, 1179197, 1179198, 1179199, 1179200, 1179202, 1179204, 1179205, 1179211, 1179212, 1179213, 1179214, 1179215, 1179216 b. 21-2120-0105-03L, Serial Numbers: 1171616, 1171618, 1171619, 1171620, 1171622, 1171631, 1171634, 1171635, 1171636, 1171651, 1171652, 1171653, 1171684, 1171685, 1171686, 1171687, 1171688, 1171689, 1171690, 1171691, 1171692, 1171693, 1178356, 1178357, 1178358, 1178359, 1178360, 1178361, 1178362, 1178859, 1178860, 1178862, 1178863, 1178864, 1178865, 1178866, 1178867, 1178868, 1178869, 1178870, 1178871, 1178872, 1178873, 1178874, 1178875, 1178876, 1178877, 1178878, 1178879, 1178880, 1178881, 1178882, 1178883, 1178884, 1178981, 1178982, 1178983, 1178984, 1178985, 1178986, 1178987, 1178989, 1178990, 1178991, 1178992, 1178993, 1178994, 1178995, 1178996, 1178998, 1178999, 1179000, 1179001, 1179002, 1179005, 1179006, 1179007, 1179008, 1179009, 1179010, 1179011, 1179012, 1179013, 1179015, 1179016, 1179018, 1179019, 1179020, 1179023, 1179025, 1179027, 1179028, 1179029, 1179030, 1179031, 1179032, 1179033, 1179034, 1179036, 1179037, 1179038, 1179039, 1179040, 1179041, 1179042 c. 21-2120-0105-06L, Serial Numbers: 1171405, 1171406, 1171407, 1171408, 1171409, 1171410, 1171411, 1171412, 1171413, 1171414, 1171415, 1171416, 1171417, 1171418, 1171419, 1171420, 1171421, 1171422, 1171423, 1171424, 1171425, 1171426, 1171427, 1171428, 1171429, 1171430, 1171431, 1171432, 1171433, 1171434, 1171435, 1171436, 1171437, 1171438, 1171439, 1171440, 1171441, 1171442, 1171443, 1171444, 1171445, 1171446, 1171447, 1171448, 1171449, 1171450, 1171451, 1171452, 1171453, 1171454, 1171455, 1171456, 1171457, 1171458, 1171459, 1171460, 1171461, 1171462, 1171463, 1171464, 1171465, 1171466, 1171467, 1171468, 1179142, 1179148, 1179149, 1179151, 1179152, 1179155, 1179157 d. 21-2120-0105-07L, Serial Numbers: 1171626, 1171627, 1171628, 1171629, 1171630 e. 21-2120-0105-08L, Serial Numbers: 1171469, 1171470, 1171471, 1171472, 1171473, 1171474, 1171475, 1171476, 1171477, 1171478, 1171479, 1171485, 1171486, 1171487, 1171488, 1171489, 1171490, 1171491, 1171492, 1171493, 1171494, 1171495, 1171496, 1171497, 1171498, 1171499, 1171500, 1171501, 1171503, 1171504, 1171585, 1171586, 1171587, 1171588, 1171589, 1171590, 1171591, 1171592, 1171593, 1171594, 1171598, 1171599, 1171600, 1171601, 1171602, 1171603, 1171604, 1171605, 1171606, 1171607, 1171608, 1171609, 1171610, 1171623, 1171624, 1171625, 1171643, 1171644, 1171645, 1171646, 1171647, 1171648, 1171654, 1171655, 1171656, 1171657, 1171658, 1171659, 1171660, 1171661, 1171662, 1171663, 1171664, 1171694, 1171695, 1171699 f. 21-2120-0105-12L, Serial Numbers: 1171480, 1171481 g. 21-2120-0105-13L, Serial Numbers: 1171611, 1171612, 1171613, 1171614, 1171615, 1179158 h. 21-2120-0105-14L, Serial Numbers: 1178363, 1178366, 1179096, 1179097, 1179132, 1179133, 1179135, 1179136, 1179137, 1179217, 1179242 i. 21-2120-0105-15L, Serial Numbers: 1171595, 1171596, 1171597, 1178887, 1178888, 1178889, 1178890, 1178891, 1178893, 1178894, 1178895, 1178896, 1178897, 1178898, 1179103, 1179105, 1179106, 1179107, 1179108, 1179109, 1179110, 1179112, 1179113, 1179114, 1179115, 1179116, 1179117, 1179118, 1179119, 1179120, 1179121, 1179122, 1179123, 1179124 j. 21-2120-0105-17L, Serial Numbers: 1171505, 1171506, 1171507, 1171508, 1171509, 1171510, 1171511, 1171512, 1171513, 1171514, 1171526, 1171527, 1171528, 1171529, 1171530, 1171531, 1171532, 1171533, 1171534, 1171535, 1179138, 1179139, 1179140 k. 21-2120-0105-50L, Serial Numbers: 1171516, 1171518, 1171520, 1171524, 1179098, 1179099, 1179100, 1179101, 1179102 l. 21-2127-0105-02L, Serial Numbers: 1179125, 1179127, 1179128, 1179129, 1179218, 1179219, 1179220, 1179222, 1179223, 1179224, 1179225, 1179226, 1179228, 1179230, 1179232, 1179233, 1179234, 1179235, 1179236, 1179237, 1179238, 1179239, 1179240, 1179241 m. 21-2127-0105-06L, Serial Numbers: 1171621 n. 21-2127-0105-50L, Serial Numbers: 1178354, 1178355 |
| FEI Number |
3012307300
|
Recalling Firm/ Manufacturer |
Smiths Medical ASD Inc. 6000 Nathan Ln N Minneapolis MN 55442-1690
|
| For Additional Information Contact | Dave Halverson 763-383-3310 |
Manufacturer Reason for Recall | Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line event may not be recognized by the pump and may not alarm to notify the clinician. |
FDA Determined Cause 2 | Component change control |
| Action | The firm notified its consignees by email beginning 03/24/2020. The notice explained the issue, risk to health, and requested a response. Distributors/pharmacy suppliers were directed to notify their customers. The firm is seeking return of the units for correction. |
| Quantity in Commerce | 642 units |
| Distribution | Distribution to European countries: BE, BH, CH, DE, DK, ES, FI, FR, GB, IT, HL, NO, PT, SE |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = FRN
|
|
|
|